MILESTONE

H E L P I N G
C H E M I S T S

duoPUR - subPUR

Operator Manual
MA180
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Thank you for selecting our distillation system.


We are sure that you will be completely satisfied with the performance of
this new unit entering your laboratory. We invite you to read carefully this
user manual and to keep it in reach for convenient and fast consulting. The
person who will be using this unit should have received an appropriate
training from a Milestone trained technician. For any possible clarification or
any request for assistance please contact either our Representative in your
country:

or Milestone at the following address:

Publication date (gg/mm/aaaa): 09/10/2020

Milestone Srl
Via Fatebenefratelli, 1/5
24010 Sorisole (BG) Italy
Tel +39 035 573857
Fax +39 035 575498
Website www.milestonesrl.com
E-mail analytical@milestonesrl.com

1
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Sommario

1 SAFETY INFORMATION ....................................... 4


1.1 General ................................................................................. 4
1.2 Safety instructions for start-up.............................................. 4
1.3 Safety instructions for operations.......................................... 4
1.4 Safety instructions for service operations .............................. 5
1.5 No manipulations to the device ............................................. 5
1.6 Parts replacement ................................................................. 5
1.7 Customer support ................................................................. 5
1.8 Further information ............................................................... 5
1.9 Repair/defective parts ........................................................... 6

2 INSTALLATION ...................................................... 7
2.1 Unit unpacking ...................................................................... 7
2.2 Scope of supply..................................................................... 7
2.3 Unit description ................................................................... 10
2.4 Technical data..................................................................... 11
2.5 Sub-boiling principle............................................................ 11
2.6 Connections and control elements ...................................... 12
2.6.1 Terminal control system T480 ............................................................ 12
2.6.2 Basic unit subPUR/duoPUR ................................................................ 12
2.6.3 Cooling water connections ................................................................. 13

2.7 Unit assembly ..................................................................... 14


2.7.1 Quartz tube installation ...................................................................... 14
2.7.2 Heating element installation ............................................................... 16
2.7.3 Cooling water connections (isolation and silicon tubes)...................... 17
2.7.4 Safety protection shield...................................................................... 19

3 SOFTWARE DESCRIPTION ................................ 20


3.1 Operations .......................................................................... 20
3.2 Setup .................................................................................. 21
3.3 Program control .................................................................. 22

2
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

4 OPERATIONS WITH THE UNIT ......................... 23


4.1 Function outline .................................................................. 23
4.2 Filling .................................................................................. 24
4.2.1 Manual filling...................................................................................... 24
4.2.2 Automatic with pump station (option) ................................................ 24

4.3 Heating program ................................................................. 27


4.4 Pre-conditioning of the boilers ............................................ 27
4.5 Distillate quality ....................................................................... 27
4.6 Cleaning procedure ............................................................. 28
4.7 Draining and Cleaning ............................................................. 29
4.7.1 Manual ................................................................................................. 29
4.7.2 Automatic with pump station (Optional) ............................................... 30

3
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

1 SAFETY INFORMATION
1.1 General
The instruments commercialized by Milestone are intended for the use in strict compliance with
all applicable laws and in a manner that promotes public safety and/or linked to approved
scientific or medical research purposes.
NOTE Milestone does not endorse, encourage or promote the use of its instruments for any illegal
activities.
No liability for any consequences, direct or indirect, is accepted if the subPUR/duoPUR system is
used in any way in which the instrument is not intended to be used.

▪ Please read the user manual carefully before start-up and follow the instructions.
▪ If additional devices have been delivered with this device (resp. this device is an additional device),
the respective user manuals must also be considered.
▪ This Operator Manual is part of the delivery, hence must be always kept together with the instrument
on its working place.
▪ It is imperative that every person operating this system has read and fully understood this manual.
▪ The non-observance of the instructions contained herein or improper use may involve damages/injuries
that are not covered by product liability!
▪ The national ruling and international safety norms in working environment shall in any case be observed
under user’s responsibility.
▪ It is also user’s responsibility to keep constantly updated all safety norms ruling in the country or
operation.

Information means an important instruction to be followed.

Attention means that failure to follow instructions may result in property damage.

Warning means that failure to follow instructions may result in danger to life and health.

1.2 Safety instructions for start-up


▪ The first start-up of the system must exclusively be carried out by authorized and trained service
engineers appointed by Milestone.

1.3 Safety instructions for operations


▪ The device may only be used and operated in accordance with its intended purpose.
▪ The device may only be operated within its performance limits.
▪ The device may only be operated by trained personnel.

4
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

▪ All parts of the electrical system must be protected against moisture and humidity.
▪ The device must be kept clean.
▪ Basic rules of chemistry will always apply.
The basic rules of chemistry are valid also for handling chemicals in this unit. Working with
chemicals, always take all those safety measures that are usually required (i.e. lab coat, protective
gloves, protective glasses mask, exhausts hood, etc.)

1.4 Safety instructions for service operations


▪ In case of damage of the device a trained service technician must be called. The device must not be
used until the successful repair has been completed.
▪ In the slightest doubt about the proper function of the device, you may not be working with it.
▪ Service can only be carried out by authorized and trained service engineers of Milestone company.
▪ Works on the electrical and electronic components of the system must be carried out exclusively by
specialized electrical/electronic engineers.
▪ Before repairs / service / installation work, switch off the main breaker, disconnect the plug and secure
it against being switched on again.

1.5 No manipulations to the device


▪ Constructive modifications to the device are forbidden.

1.6 Parts replacement


▪ In order to guarantee the good functioning of the unit and user’s safety, all replacement parts must
be exclusively original supplied by Milestone. Original spare parts are granted by high quality control
on the material and during the production.

1.7 Customer support


▪ The Warranty and the Service Report are included in the documentation delivered together with the
unit.
▪ After the installation, you are kindly requested to send us the filled in warranty card via fax or e-mail
(form on the website).
▪ In case of required unit reparation, you should send us the filled in Service Report including precise
description of the errors.
▪ For any request or order, please contact your local area representative or directly Milestone.

1.8 Further information


When working with subPUR/duoPUR, please follow the below additional safety instructions, in
order to ensure a longer life cycle of the unit and to prevent pollution of lab air.

▪ The unit must be installed in a protected atmosphere, in order to prevent damages to the quartz parts.
In case of such damages, the unit operator will be consider responsible.
▪ The use of electronic components in a fume hood or in a room with aggressive atmosphere can result
in a considerable limited life cycle of the unit, due to corrosion of the electronic components. For this
reason the SubPur/DuoPur is divided in a terminal control unit (installation outside the fume hood) and
a distillation unit (possible installation in a fume hood).
▪ Pressure balance without gas loading

5
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

The hole of PTFE adapter above the PTFE bottle is important to avoid any pressure generation inside
the system. If the following points are matched, the unit can be operated as well without gas loading
outside a fume hood.
The PTFE and the silicone hoses are used to lead dangerous and corrosive gases into the fume hood,
in the event that the distillation unit is not directly installed into a fume hood. The length of the tube
should be kept as shortest as possible. If the distillation unit is operated under a fume hood, the short
PTFE tube is enough.

 The 4 cm short PTFE tube (A) is introduced in


the hole of the collector to compensate pressure.
 The 2 m long PTFE tube (C) is pushed onto the
PTFE tube (A).
 The connection is locked by using the short
silicon tube (B), which is pushed onto both PTFE
tubes.

Pressure balance of the system

▪ System cooling before removal


Before removing the distillate, the system must be cooled down, otherwise acid vapors and gases such a
NOx could be released.
Always wear gloves, protective clothes and glasses.
The cooling temperature must be set by the user in a way that no condensation occurs on the cooling
hoses. The condensation on the hoses is affected by the humidity of the environment.
The refrigeration lines must be checked periodically for wear and leakage. The material of the hose lines
can be affected by environmental factors, which can lead to damage. The hoses must be replaced at regular
intervals.
▪ Additional PTFE bottle (option)
An additional PTFE bottle should be used for the following reasons:
Reduced gas loading when emptying the bottle
If the distillate is removed, a new PTFE bottle should be screwed. In this way the lab air load is significantly
reduced, since the residual acid vapors remain in the system.
Easier and faster handling
The additional bottle can be immediately screwed and the following acid can be distillated without any
interruption. Once completed the distillation, the condensate outlet must be connected to a new bottle.

1.9 Repair/defective parts


All parts are examined before delivery for full operability. In case of defective parts, return them only with
Milestone’s original packing, as the fragile quartz systems can be damaged during transportation. If
necessary Milestone original packing can be requested.
During transportation, the packaging has an appropriate marking (fragile, glass of etc). Milestone is not
responsible in case the system is dropped from a height of more than 30 cm. Any damages must be
notified immediately. A warranty claim exists exclusively for the shipper (transport insurance during original
packing) as proof of an inappropriate treatment.

6
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

2 INSTALLATION
2.1 Unit unpacking
Check within 24 hours that the delivery is complete and, in case some of the items previously
listed are missing, contact Milestone or Milestone authorized distributor in your country.
Check eventual transport damages of the unit. Damaged unit must not be operated. Any damage
should be reported immediately in writing.

Keep the packing safely. In case of unit return to the factory for repair purpose, use only the original
packing and secure it with protective belts. Inadequately packed unit could be considerably damaged during
transport.
1. Loosen the packing tapes and open up the box.
2. First, take out the overhead, smaller boxes. These includes the terminal and the accessories of the
subPur/DuoPur.
3. Remove the cardboard cover, the foam inserts and the protective film.

The unit weight is ca. 30 kg and must be lifted from two people.

4. Reach underneath the unit and lift it out of the box on the table.

2.2 Scope of supply


subPUR
Here are listed the items installed in the unit which are separately or not directly packed. The other items
are listed under p/n SUB02010.

Art. Nr. Picture Description


SUB00300 subPUR – system with terminal T480
System with an evaporation and condensation unit for
production of high pure chemicals according to the
„sub-boiling effect“.

84699 Quartz boiler complete (1 piece)

84690 Heating elements DuoPur (2 pieces)

7
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

DSS0053 Safety protection shield

DSS0042 Cover for dosing funnel NS 19

DSS0043 PTFE pierced stopper Sub/DuoPur NS 19 with


connection DB

BEH0040 PFA bottle 500 ml

DSS0052 Adapter for PFA bottle (1 pieces)

933676 Silicon tube Ø 10x6 mm (2 Meters)

SL0050 Tube PU Ø 8x6 mm (5 Meters)

933920 Power cable 230V

duoPUR

8
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Here are listed the items installed in the unit which are separately or not directly packed. The other items
are listed under p/n DUO02010

Art. Nr. Picture Description


DUO00300 duoPUR – system with terminal T480
System with an evaporation and condensation
unit for production of high pure chemicals
according to the „sub-boiling effect“.
System with two evaporation and condensation
units

84699 Quartz boiler complete (2 pieces)

84690 Heating elements duoPUR (4 pieces)

DSS0053 Safety protection shield

DSS0042 Cover for dosing funnel NS 19 (2 pieces)

DSS0043 PTFE pierced stopper Sub/DuoPur NS 19 (2


pieces)

BEH0040 PFA bottle 500 ml (2 pieces)

DSS0052 Adapter for PFA bottle (2 pieces)

933676 Silicon tube Ø 10x6 mm (4 Meters)

9
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

SL0050 Tube PU Ø 8x6 mm (5 Meters)

933920 Power cable 230V

2.3 Unit description


The unit SubPur/DuoPur is used for chemicals purification in lab operation.
It is designed as a complete workstation for distillations with ease-of-use.
The communication between the base unit and the operator occurs via the terminal.
▪ The distillation unit can be used directly in a fume hood.
▪ The terminal could be positioned outside the fume hood.

A Main switch

B Quartz boiler

C PTFE adapter

D Distillate collector

E Heating element

F Cooling finger

G Housing for elettric


connections

H Pump station (optional)

I Safety protection shield

10
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

2.4 Technical data


Power supply 230V; 50 Hz
At 100 % Power Current consumption 6 A
Power each heating element 680W
Weight subPUR 24 kg
Weight duoPUR 27 kg
Weight Terminal 0,5 kg
Dimensions Height: 520 mm
subPUR / duoPUR Width: 350 mm
Depth: 615 mm
Dimensions Height: 65 mm
terminal T480 Width: 140 mm
Depth: 140 mm

2.5 Sub-boiling principle


The principle of distillation is based on the sub-
boiling procedure.
The high purity of the distilled liquid is achieved by
the fact that liquids are evaporated and condensed
by heating of the surface and the fluid temperature
is below the boiling point. It does not boil all the
liquid, but the liquid evaporates without spraying
directly from the surface. This means that virtually
no metal ions are entrained by the vapor as droplets.

Advantages of sub-boiling principle

11
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

2.6 Connections and control elements


2.6.1 Terminal control system T480

Connection back terminal T480

Terminal Model T480, 4.3’’ – Color – Touch Screen

A USB 1 USB port for Service updates

Connection to SubPUR/DuoPUR unit


Cable required: RJ45 with following scheme

PIN SIGNAL
B Device port (RJ45)
2 RxT (In)

5 GND

6 TxD (Out)

2.6.2 Basic unit subPUR/duoPUR

CONNECTION

A RJ45 port for the connection with the


control terminal (RJ45 cable)

B Power supply

Basic unit subPUR/duoPUR connections

12
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Connecting the unit to the T480 terminal


1. Connect the unit to the power supply via the power cable. Check that the main switch on the
front side is set to OFF position.

Never connect the terminal when the unit is switched on, otherwise this could damage the unit
interface board.

2. Introduce the interface cable RJ45 in the DEVICE connection on the terminal back side.
3. Introduce the other end of the cable in the terminal connection on the microwave unit back side.

1. 2. 3.

2.6.3 Cooling water connections

CONNECTION

C Outlet cooling water

D Inlet cooling water

Basic unit subPUR/duoPUR connections

A cooling water unit connection is required for the operations with subPUR/duoPUR. The cooling water
connection takes place by PVC and silicon tubes with related insulating tubes included in the delivery.
The water pressure must not exceed 4 bar.
The distillation unit is connected after the transport with the cooling water supply according to the following
scheme. The cooling water inlet is represented by IN, the cooling water oulet by OUT.

Cooling water connection subPUR Cooling water connection duoPUR

13
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

During the operations, the cooling water temperature must be kept between 5°C and 15°C and
constantly monitored. The flow rate of cooling water must be set at least at 4,5 L/Min.

The cooling temperature and humidity must be set by the operator, so that no condensation
occurs in the cooling tube. The cooling lines must be checked periodically for wear and leakage.
The material of pipe line can be affected by environmental factors leading to damages. The tubes
must be replaced at regular intervals.

2.7 Unit assembly


The unit is delivered without the quartz distillation units (quartz tubes).
The quartz distillation units are included in a breakage-protected packing. The SubPur is
equipped with a distillation unit, while the DuoPur with two distillation units.

Unit assembly

2.7.1 Quartz tube installation

1. Remove the cover, by loosening the screws on the top and back of the cover. Pay attention to
disconnect the ground wire from the cover.

Housing removal

14
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Backside of the unit (open)

2. Carefully remove the quartz tubes from the transport packing and introduce it in the housing, as shown
in the following chapter.

Remove each above PP pin located on the external side of the unit back side. Once introduced,
these pins are used to fix the quartz tubes.

Quartz tube introduction

15
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

2.7.2 Heating element installation

1. Carefully remove the heating elements from the


packing and take out eventual packing
residues.
2. Introduce the heating element into the quartz
tube through the PTFE holder.
The distance of the front edge of the heating elements up
to the end of the tube should be ca. 1 cm.

3. Fix the heating element by tightening the PTFE


locking screw.

4. After the introduction of two or four heating


elements, introduce the connection plug oft he
heating elements in the intended network
sockets.

16
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

2.7.3 Cooling water connections (isolation and silicon tubes)

1. Connect the shorter silicon tube (OUT


connection) to the upper connection of the
closest quartz tube (left side).

Lock the silicon tube using the cable tie.

2. Connect the longer silicon tube (IN connection)


to the lower connection of the quartz tube on
the right side (cold finger inlet).

17
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

3. Use the spare silicon/isolation tube to connect


the quartz tubes among each other.

Slide carefully. The cooling fingers are very fragile.

You can find the outline for the correct connection of the corresponding silicon tube to the cooling
finger in chapter Cooling water connections.

4. Once connected all silicon tubes to the


corresponding cooling finger, slide the isolation
tubes to their ending position; the silicon tubes
should be covered.
Make sure that the isolation tubes are not twisted.

Once connected the cooling water lines, it’s recommended to let flow the cooling water and test
the line connections for eventual leaks. Thereafter it’s possible to re-install the housing cover.

18
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

2.7.4 Safety protection shield

During operations, the quartz tube could become


extremely hot.
The purpose of the safety protection shield is to
prevent the user to get in direct contact with the
hot quartz tubes.

In order to facilitate periodic cleaning or


maintenance operations, the protection shield can
be easily opened by hand.

19
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

3 SOFTWARE DESCRIPTION

3.1 Operations
The control of the SubPur and DuoPur is managed by the touch screen of the terminal T480. Use the touch-
pen included which is located in its holder on the right side of the terminal.

Click this button to open another menu.

By the left arrow you go back to the previous level


and by the hook you confirm your entry.

After switching on the terminal, the start screen


appears.
In the upper part of the screen you can see the
current program name and below the set control
parameters. The green button is used to start the
respective program.
Optionally, a pumping unit for filling and emptying
is available. When this pump is installed in the
device, the button >Fill Clean< appears on the
screen.

By touching the triangle in the upper left corner,


you can get into the Setup page.

20
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

3.2 Setup
In the setup, only few settings can be made by the administrator. Most parameters are preset
by the manufacturer or service.

Via > Settings < you return to the login menu. Via
the button > Administrator < you have access to
the setup menu.

The password assigned at the moment of delivery


is 123456. Log in and confirm your entry by the
arrow in the upper right corner.

In the tab > Access / Settings < you can manage


the key tone via the button > Bell <. With the
button > background < you can change the
background color.

Set the background color of the terminal with the


help of a slider. Confirm the settings by the right
arrow.

In the tab > Access / Users < you can create new
users, edit existing users or delete them. Also the
administrator password can be changed here.
Users of higher authorization cannot be edited.

In the tab > Configuration / Power < you can test


the functionality of the heating cartridge.

In the tab > Configuration / Pump < you can test


functionality of the pump (option).

21
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

3.3 Program control


By touching the button in the upper right corner you
enter the program editing mode.

In the upper part of the screen you have the


possibility to create, delete or edit methods.

By opening the scroll menu, a method can be


selected from the list (methods created by the
user).

To create, delete or modify methods Administrator


login is needed.

Log in with your Admin password.

When you delete a method, an additional safety


query needs to be confirmed.

In the menu > New < create a new method by


touching the empty rectangular field and entering a
new method name via the visual keyboard.

In the lower part, you can specify time and


percentage of power of the individual program
steps.

In the menu > Change < you can modify existing


methods by touching the appropriate field to make
the changes.

22
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

4 OPERATIONS WITH THE UNIT


4.1 Function outline

Function outline

A Filling opening D Distillate connecting funnel

B Cooling finger E Collecting vessel (Distillate)

C Heating elements F Ventilation system

23
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

4.2 Filling
4.2.1 Manual filling

The sub-boiling system can be filled max. up to 2 cm below the connecting funnel (D) (ca. 500-
600 ml).

1. Introdue a suitable funnel into the filling


opening (A) and introduce the acid to be
cleaned.
2. After the filling, seal the opening with the PTFE
stopper included.
The stopper should be left there, until it expands
due to the heating.

NO DRYING RUN! Never heat the unit without any liquid.

A refilling should be carried our before reaching a minimum filling level of 1-2 cm over the
distillation vessel bottom.

4.2.2 Automatic with pump station (option)

Before the first use, carry out a run test with water.

Always check before filling and emptying the tightness of hose connections and plugs to
prevent splashing out of the acid.
Make sure that the distillation vessel is residues-free before the filling.

Before starting the pump for filling or emptying, the port on the buffer bottle must be plugged into
the corresponding position (P for filling and V for emptying)!

24
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Filling of distillation vessel (Function principle)


For filling, the fluid from the supply bottle (SB) is pressed into the distillation boiler.
The buffer bottle (BB) is directly connected to the pump (P).
For safety reasons, only a small pressure is established.
The supply bottle (SB) with the chemicals is placed in the change holder to prevent tipping.
The supply bottle has two connectors shown in the diagram and is always connected to the pump via
the buffer bottle (BB).
For safety reasons, the distillate collecting bottle remains connected, so that no chemicals run out, in case
of overfilling.

Loosen the two hoses from the supply bottle and


remove the supply bottle.

Fill the supply bottle with the acid to be distilled


and re-screw the PFA cover with both connectors
on the flask.
Re-introduce the flask with the cover in the unit.

Slide the tubes into the supply bottle. Pay attention


to the tightness of hoses and that the silicone
gaskets seals the PTFE hoses and coupling.

25
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Introduce the tube into the connection P of the


pump unit.

Press the button>Fill Clean< until the supply


bottle is completely empty and the distillation
boiler is filled with the corresponding amount.

In case of duoPUR, the system is equipped with a


unique pump with two separate tubing lines, red
and yellow.
This means that the acids cannot be supplied to
the two boilers at the same time but one at a time.

IMPORTANT NOTE:
Before the distillation operation, the PTFE hoses must be
plugged into B position!

26
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

4.3 Heating program


Distillation speed and purity level are strictly dependent on the heating program.
The higher is the distillation power, the higher will be the productivity but, on the other hand, the quality
of distillate will be lower.
The quality of the distillate also depends on the quality of the starting acid.

The system must not be operated with sulfuric acid (H2SO4), perchloric acid (HClO4),
hydrogen peroxide (H2O2), hydrofluoric acid (HF) or other flammable solvents.

Keeping in mind all the above mentioned considerations, here are suggested distillation programs:
• HNO3: 45% power to have 50 ml/hour.
• HCl: 35% power to have 40 ml/hour.

4.4 Pre-conditioning of the boilers


Once received a new unit, please note that the boilers are not cleaned. As such it is important to initialize
and acclimatize the boilers, before collecting the purified acid.
With a new system, it is suggested to follow the below pre-conditioning procedure:
1. Add 500 ml of reagent that has to be purified
2. Perform a method of 4 hours at 40% power. If the acid boils, reduce the power to 20-30%.
3. Collect only the last few milliliters of purified reagent and analyze it. If the reagent quality is enough,
cleaning is no longer required.
4. Repeat this operation 2-3 times, until the quality achieved is good.

4.5 Distillate quality


You can obtain the heighest cleaning quality by low power and high temperature difference in the area of
the cooling finger.
You can obtain the highest cleaning quality with the distillate liquids obtained by sub-boiling procedure,
since acids/water are evaporated by heating of the surfaces and condensed. This means that the liquid
doesn’t boil, but evaporates without spraying directly on the surface.
According to the acid concentration, then lower or higher concentrations are obtained, until the azeotropic
mixtures are adjusted, developing in this way constant concentrations during the distillation.

HCl HNO3

Concentration 20 % 69,2 %

Boiling point 110°C 122 °C

Table: azeotropic misture, boiling point

The system must not be operated with sulfuric acid (H2SO4), perchloric acid (HClO4), hydrogen
peroxide (H2O2) or other flammable solvents.

27
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

ACIDS (Standard quality)

ELEMENTS Al Cd Cu Fe Mg Pb Zn

Water (Quality analysis) 0,5-2 0,2-0,5 0,1-0,4 0,1-2 1-5 0,2-0,6 1-5

Water (Milestone “Sub- 0,04 0,01 0,02 0,3 0,01 0,02 0,04
boiling)*

HCl-12 N (Quality analysis) 10-50 0,1-10 1-20 >100 >10 0,5-20 5-50

HCl-10 N (Cleaning quality) 1-10 0,05-1 0,2-1 10 0,5-20 0,1-5 0,5-2

HCl-10 N (Milestone “Sub- 0,05 0,01 0,07 0,5 0,1-5 <0,03 0,2
boiling)*

HNO3 15 N (Quality analysis) 10-50 0,1 2 10-50 0,1 1-10 10-100

HNO3 15 N (Cleaning quality) 5-50 0,05 1-3 10-50 10-100 0,5-5 5-10

HNO3 15 N (Milestone “Sub- 0,03 0,001 0,15 0,1 0,1 0,01 0,03
boiling)*

Table: element content in µg/L


*The above mentioned “Milestone sub-boiling“ value has been compared to usual acids. For test have been used samples with
“analysis” and ultra-pure quality. The distillation power of the unit can be optimized by introducing the heating power for each
corresponding acid. To obtain 1 liter ultra-pure acid, 1,5 to 6 hours are required at a power of 300-1000 watt.

4.6 Cleaning procedure


When you have to change the acid type to purify, please clean the boiler by following the below procedure:
1. Remove the reagent inside the boiler through a pump
2. Introduce DI water inside the boiler
3. Remove DI water from the boiler through a pump
4. Add one more time water inside the boiler and start a heating program of 1 hour at 40% power.
Once the distillation ends, waste the collected water and remove the DI water remained inside the
boiler through a pump
5. Add the new reagent to distil to the boiler
6. Remove acid from the boiler through a pump
7. Add new acid to the boiler and start the distillation to produce purified acid.

Waste the first 25 ml of purified acid, then start to collect the cleaned acid.

28
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

4.7 Draining and Cleaning


4.7.1 Manual

Introduce a resistant tube in the desired PTFE stopper


with connector and connect the tube to a vacuum
pump.
Then introduce the PTFE stopper in the filling opening.
Flush the sub-boiling unit several times with e.g.
distilled H2O, in order to clean it.

If there are liquids in the exhaust tube, the tube end can be placed in a collection vessel below
the unit level (without pump).

If any resistant pump isn’t introduced, a Woulf bottle should be installed between the pump and
the system.

Woulf bottle draining system

29
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

4.7.2 Automatic with pump station (Optional)

Draining (cleaning) of the distillation vessel (Function principle)

To empty the distillation boiler, the vacuum side (V) of the pump is directly connected to the buffer bottle
(BB).
By starting the pump (touch the button fill/clean) the residual liquid is pumped out from the distillation
boiler into an empty supply bottle. Therefore, the lid and all connections must be tightly closed.
Another rinse with a new (fresh) template and repeated aspiration is advised only in the event of very
strong contamination.
For filling and emptying the respective procedure is repeated.

Introduce the PTFE tube in the PTFE stopper and


lock the connection using the short silicon tube.

Empty the supply bottle and restore the connections


with the buffer bottle and the quartz boiler.

30
Milestone Srl MA180-005 - SubPUR and DuoPUR T480 – User Manual

Insert the hose from the supply bottle into to the


terminal V of the pump unit.

Press the button >Fill Clean< until the distillation


vessel is completely emptied.

In case of duoPUR, the system is equipped with a


unique pump with two separate tubing lines, red
and yellow.
This means that the acids cannot be removed from
the two boilers at the same time but one at a time.

IMPORTANT NOTE:
Before the distillation operation, the PTFE hoses must be
plugged into B position!

31
MILESTONE
Milestone Srl - via Fatebenefratelli 1/5 - 24010 Sorisole (BG) - Italy
Phone +39 035 573857 - Fax +39 035 575498
H E L P I N G
e-mail: analytical@milestonesrl.com - web: www.milestonesrl.com
C H E M I S T S