You are on page 1of 2

Service Report Service Request Type:

Corrective Repair
Service Request Number:
2021-11/0250

Customer Details

Customer Name: RSU.Sentra medika Location: Jakarta-Indonesia


PO #: cisalak Contact Phone: +6282292264455
Contact Nome: Mrs.Tiara Email Address:
Initial Problem Description: Artefac of
result emage from
panoramic issue

Equipment Details

System ID:
Panoramic Initial Equipment Status: Iintermitten on panoramic machine
Serial Number:
Entitlement: -
Product Description:

Activity Description

Activity #: Activity Close


Activity Type: Field Support Dote: Engineer Minal Aidin saputra
Tube Slice Reading: Name:
Problem Description: - Artefac emage

Couse/ Diagnosis: Nov 19-11-2020 :


- Chek casset on x-ray result OK
- Cassete tray not working properly

Nov 19-11-2020 :
Resolution/ Service
Performed:
- Check and inspect all fuse (OK)
- Chek and inspect LCD and keypad OK
- Chek and inspect voltage suplay OK
- Chek and inspect try cassete OK
- Adjustment rotary try cassete OK
- Running test machine On X-ray result OK
- Panoramic ready to use

Recommendotion :

“carry out maintenance according to schedule”

Page No. 1 of 2
Service Request Type:
Request Number: Corrective Repair
Service Report 2320030-55

Date Time Hours Labour Type


Nov 19, 2021 00:00-01:44 Onsite Repair
Nov 19,2021 01:50-00:00 Zona Travel

Ports Used Details

Port
Number
0tY Description Asset Serial No

Expense Tracker

Expense Type 0tY Comments


Dote

Engineer Customer Signoture:


Signoture: 19-Nov-2021 01:00
Minal Aidin saputra

Page No.2 of
2