Professional Documents
Culture Documents
MV/TV
15 0.9 10 0.9 15
14 9 14
0.8 0.8
13 13
3 4 12 8 12
5 0.7 0.7
2 11 11
6 7
10 0.6 0.6 10
1
9 6 9
1
7
8
0.5
5
0.5
8 Tec 7 Tec 7
8 7 0.4 7
2 0.4
6 4 6
5 0.3 0.3 5
4 3 4
3 0.2 0.2 3
2
2 2
0.1 0.1
1 1 1
ISO SEV
AIR N 2O O2
4 4 4
5 5 5
300
3 3 3
58 73 58 73 58 73
43 6 43 6 43 6
87 87 87
kPa kPa kPa
x x x
600
100 100 100
8 8 8
10
AIR N20 02
900 8
L/min O2
6
1200 4
150
200
150
200
150
200
2 100 2175
100 2175 100 2175
1500
1450 2900 1450 2900 1450 2900
kPa kPa kPa
50 725 psi x 50 725 psi x
50 725 psi x
100 100 100
N20 AIR 02
20 70
02+
max=560 cmH2O
Education Services
Clinical Development
Aespire View 6.x
Participant Guide
Version A
Education Services
C l i n i c a l D e v e l o p m e n t
Notice
The materials contained in this document are intended for educational purposes only.
This document does not establish specifications, operating procedures or maintenance
methods for any of the products referenced. Always refer to the official written materials
(labeling) provided with the product for specifications, operating procedures and
maintenance requirements.
Note! This participant guide is not intended to replace the User’s Reference
Manuals that you received with the machine. Please refer to the disclaimer notice
at the end of this participant guide for more information.
This course is intended for Aespire software level 6.X. The material contained in this
course is intended for educational purposes only. Always refer to the official written
materials provided with the Aespire anesthesia system for specifications, operating
procedures, and maintenance requirements.
i
Contents
1 Welcome . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1 4 Clinical Training Documents . . . . . . . 4.1
Gas Regulation & Flow Control . . . . 2.14 Non-clinical Training Quiz . . . . . . . . . 5.4
3 Resources . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
Aespire View User Operational . . . .
Maintenance Schedule . . . . . . . . . . . 3.1
Aespire View Anesthesia Supplies
and Accessories . . . . . . . . . . . . . . . . . 3.2
Aespire View Troubleshooting
Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3
ii
Aespire View 6.x 1 Welcome
1 Welcome
1 Welcome
We would like to take a moment to thank you for choosing GE Healthcare for your
anesthesia carestation needs. Our goal now is to provide you with the best training service
available while continuing to support you as you use our products in your workplace.
Class Description
This course is designed to give Participants the information and product knowledge needed
to proficiently operate the Aespire View. This class will use discussion, scenarios and return
demonstrations. The application of problem-solving techniques will also be incorporated.
1.1
Aespire View 6.X 2 The Aespire View
MV/TV
15 0.9 10 0.9 15
14 9 14
0.8 0.8
13 13
3 4 12 8 12
5 0.7 0.7
2 11 11
6 7
10 0.6 0.6 10
1
9 6 9
1
7
8
0.5
5
0.5
8 Tec 7 Tec 7
8 7 0.4 7
2 0.4
6 4 6
5 0.3 0.3 5
4 3 4
3 0.2 0.2 3
2
2 2
0.1 0.1
1 1 1
ISO SEV
AIR N 2O O2
4 4 4
5 5 5
300
3 3 3
58 73 58 73 58 73
43 6 43 6 43 6
87 87 87
kPa kPa kPa
x x x
600
100 100 100
8 8 8
10
AIR N20 02
900 8
L/min O2
1200 4
150
200
150
200
150
200
2 100 2175
100 2175 100 2175
1500
1450 2900 1450 2900 1450 2900
kPa kPa kPa
50 725 psi x 50 725 psi x
50 725 psi x
100 100 100
N20 AIR 02
20 70
02+
max=560 cmH2O
2.1
Aespire View 6.X Participant Guide
2.2
Aespire View 6.X 2 The Aespire View
Display I J Flowmeters
Datex-Ohmeda Aespire View
MV/TV
1
7
8
0.5
5
0.5
8 Tec 7 Tec 7
8 7 0.4 7
2 0.4
6 4 6
5 0.3 0.3 5
4 3 4
K Vaporizers
3 0.2 0.2 3
2
2 2
0.1 0.1
1 1 1
ISO SEV
AIR N 2O O2
Auxiliary O2 flowmeter G
43 6 43 6 43 6
87 87 87
kPa kPa kPa
x x x
600
100 100 100
8 8 8
10
AIR N20 02
900 8
L/min O2
6
1200
1500
4
2 100
50
1450
725
150
2175
2900
psi
200
300
kPa
x
100
50
100
1450
725
150
2175
2900
psi
200
300
kPa
x
100
100
50
1450
725
150
2175
2900
psi
200
300
100
kPa
x
A System switch
N20 AIR 02
20 70
02+
max=560 cmH2O
F B Secondary gas
O2 flush supply pressure
button gauges (cylinder
gauges)
Brakes C
2.3
Aespire View 6.X Participant Guide
2.4
Aespire View 6.X 2 The Aespire View
Breathing system F
O2 A Bellows assembly
C
Rebreathing bag
D Ezchange and
Condenser Options
2.5
Aespire View 6.X Participant Guide
2.6
Aespire View 6.X 2 The Aespire View
Suction B
regulator
280-600kPa
Pipeline C 280-600kPa
inlets
280-600kPa
Cylinder D
yokes
G Flexible yoke
Main circuit E
breaker
Cylinders F
2.7
Aespire View 6.X Participant Guide
2.8
Aespire View 6.X 2 The Aespire View
selection button is pressed again or the ventilation, the user simply moves the
com wheel can be pressed (similar to Bag/Vent Switch to the Vent position. If
pressing the Enter key on a computer). the Aespire is turned On with the switch
in the Vent position, the switch will need
J Ventilation Settings: These buttons
to be moved to the bag mode and then
MV/TV
...........................
Paw
cmH2O D Alarm silence
Measured
120 30 Ppeak Pmean
I/min . Pa 25
. . . . 15. . . . . . . . . . . . . . . . . . . . . . Ppeak 40
7
cmH2O
33 2
Paw w
15 PEEP
Pmean 24
RR /min
1032Off
40
. . . . 0. . . . . . . . . . . . . . . . . . . . . . . PpauseOff
30
10 . . . . . . . . . . . . . . . . . . . . . . . . . . O2
.
Flow 60 MV I/min %
E Measured
72 F
6.25 values
20 9.0 100
5 . . . . 0. . . . . . . . . . . . . . . . . . . . . . . l
o
TV exp ml
3.0
25
10 w
1
TVexp
625 1000 ml
590
0 ........................... Off
-60 1450
0.5 35C
EtCO2 mmHg FiCO2 mmHg
60
450
0 Off
F Menu
0 2
. 4 . . . .6 . Time
. . (sec)
8
. . . . 10. . . .12 . . . 14. . . .16. . . MV
. O2 l/min
23
RR l/min
CO2 40 /
EtO2 % FiO2 %
N2O O2
5.9 50 10
50
O
3.0
Off Off
3.0
15.0
2 2.0 Off Off
0
Vent OffOn Gases
Vent VCV - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV
%
590
RR
I/min
10
I:E
%
1:1.5
Pmax
cmH2O
40
PEEP
5
/min cmH2O O2
More
Settings
3.1
Air Sev 0.10
Off
5.4 Off 4.3
5.9
8.0
Off
G End case
50 ml 6.00 /min 6.0 23
cmH2O 10
cmH2O Off N20N2O 1.148 59
Total
H Total flow
sensing
I Command
wheel
2.9
Aespire View 6.X Participant Guide
Main Menu
1. Push the Menu key to show the Main Menu.
2. Turn the ComWheel counterclockwise to highlight the next menu item. Turn the
ComWheel clockwise to highlight the previous menu item.
3. Push the ComWheel to enter the highlighted window or a sub menu.
4. Turn the ComWheel clockwise or counterclockwise to highlight the desired selection.
5. Push the ComWheel to confirm the selection.
6. Push the Menu key to exit the menu and return to the normal monitoring screen.
39 2
Main
Paw
Menu
w
Ventilation
Paw 15
Mode VCV PEEP
Pmean 24
RR /min
1032
40
40Alarm Setup
. . . . 0. . PCV
..................... Ppause
Off
Off
30
10
Setup/Calibration . . . . . . SIMV/PSV
. . . . . . . . . . . . . . . . . . . . O2
. %
Flow 60 MV I/min
726.25
30
Screen and Audio Setup PSVPro
20 F 9.0 100
5
20Cardiac Bypass 0
SIMV-PC l
o
TV exp ml
3.0
25
10 PCV-VG w
10
1
TVexp
625 1000 ml
590
0
Normal Screen
........................... Off
-60 1450
0 EtCO2 mmHg FiCO2 mmHg
450 Menu keys
0.5 35
C 60
0 Off
0 2 4
. . . .6 . . . 8. . . .10. . . 12. . . . 14. . . .16. . . . . MV O
2
l/min
23
RR l/min
CO2 40 /
EtO2 % FiO2 %
N2O O2 Time (sec)
5.9 50 10
50
O
3.0
Off Off
3.0
15.0
2 2.0 Off Off
0
Vent OffOn Gases
Vent PSVPro - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV RR I:E Pmax PEEP
cmH2O O2 3.1
4.3
% I/min % cmH2O /min More Off 8.0
590 10 1:1.5 40 5 Settings 5.4 Off
Air Sev 0.10 5.9 Off
50 ml 6.00/min 6.0 23
cmH2O 10
cmH2O Off N20N2O 1.148 59
Total
ComWheel
The Main Menu options will be discussed in greater detail in the Ventilation section.
The options that are accessed through the Main Menu include:
Ventilation Mode: Change the current mode of ventilation to a new mode.
Alarm Setup: Change individual alarm limits.
Setup/Calibration: Perform an O2 sensor calibration and access advanced
ventilation settings.
Screen and Audio Setup: Change the audio and visual appearance of the screen.
Cardiac Bypass: Activate and deactivate Cardiac Bypass mode.
Normal Screen: Back to normal view of the display.
2.10
Aespire View 6.X 2 The Aespire View
Ppeak cmH2O
30
Paw
Pmean 7
40
30 O2 %
20
10
49 Off
18
TVexp ml
372
0
Off
Off
0 2 4 6 8 10 12 14 16
Time (sec)
MV l/min RR l/min
4.5 12
6.0
3.1
1 3
2
1. Press the Quick Key associated with Tidal Volume (located just below the Tidal
Volume value displayed on the monitor).
2. Turn the ComWheel to change the setting (turning clockwise increases the level,
turning counter-clockwise decreases the level).
3. Press the Tidal Volume quick key to confirm the new setting.
MV/TV
Ppeak cmH2O
30
Paw
Pmean 7
40
30 O2 %
20
10
49 Off
18
TVexp ml
372
0
Off
Off
0 2 4 6 8 10 12 14 16
Time (sec)
MV l/min RR l/min
4.5 12
6.0
3.1
1
2 3
Figure 2.7 Changing Settings Example Two
1. Press the Quick Key associated with Tidal Volume (located just below the Tidal
Volume value displayed on the monitor).
2. Turn the ComWheel to change the setting (turning clockwise increases the level,
turning counter-clockwise decreases the level).
3. Press the ComWheel to confirm the new setting.
2.11
Aespire View 6.X Participant Guide
System Switch
Set the System switch to the On (I) position to permit gas flow and turn on the system.
The mains indicator will come on when the AC power is connected.
MV/TV
........................... cmH2O
Measured
120 Paw 30 Ppeak Pmean
I/min . Pa 25
. . . . 15. . . . . . . . . . . . . . . . . . . . . . Ppeak 40
7
cmH2O
33 2
Paw w
15 PEEP
Pmean 24
RR /min
1032Off
40
. . . . 0. . . . . . . . . . . . . . . . . . . . . . . PpauseOff
30
10 . . . . . . . . . . . . . . . . . . . . . . . . . . O2
. %
15 0.9 10 0.9 15
Flow 60 MV I/min
72
14 9 14
0.8 0.8
13
6.25 9.0
13
20 F 100 3 4 12 8
0.7 12
5
5
. . . . 0. . . . . . . . . . . . . . . . . . . . . . . l
o
3.0
25 2
6
11
0.7
7
0.6
11
10
w TV exp ml 10 0.6
10 1
625 1000
9 6 9
TVexp 7 0.5
Tec 7
1 Tec 7
ml
590
8
0.5 8
0 ........................... Off 1
8 7
5
0.4 7
-60 1450 2
6
0.4
4 6
EtCO2 mmHg FiCO2 mmHg
450
0.5 35 0
5 0.3 0.3 5
C 60 3 4
4
Off
0 2
. 4 . . . .6 . Time
. . (sec)
8
. . . . 10. . . .12 . . . 14. . . .16. . . .
MV O
2 EtO2 %
23
l/min RR l/min 3 0.2 0.2 3
CO2 40 /
FiO2 % 2
N2O O2 2 2
5.9 5 5
10
O 15.0 Off 0.1
3.0 3.0
Off 0.1
2 2.0 Off Off
1 1 1
0
Vent OffOn Gases
Vent VCV - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV RR I:E Pmax PEEP
cmH2O O2 3.1
4.3
% I/min % cmH2O /min More Off 8.0
590 10 1:1.5 40 5 Settings 5.4 Off
Air Sev 0.10 5.9 Off ISO SEV
50 ml 6.00 /min 6.0 23
cmH2O 10
cmH2O Off N20N2O 48
1.1 59
Total
AIR N 2O O2
4 4 4
5 5 5
300
3 3 3
58 73 58 73 58 73
43 6 43 6 43 6
87 87 87
kPa kPa kPa
x x x
600
100 100 100
8 8 8
10
AIR N20 02
900 8
L/min O2
6
1200 4
150
200
150
200
150
200
2 100 2175
100 2175 100 2175
1500
1450 2900 1450 2900 1450 2900
kPa kPa kPa
50 725 psi x 50 725 psi x
50 725 psi x
100 100 100
Indicator
20 70
System
switch
02+
Battery Capacity
The battery will provide power to the Aespire View for approximately 1.5 hours when fully
charged. If the battery is not fully charged, the Battery charging message will appear on
the ventilator display. If the mains supply is not connected or has failed, the message On
battery will appear on the ventilator display.
39
Paw ventilator
Pmean 24
40 display
Ppause 32
30 O2 %
20
10
72 100
25
TVexp ml
590
0
1450
450
0 2 4 6 8 10 12 14 16
Time (sec)
MV l/min RR l/min
5.9 10
Push the knob to confirm the change. Turn the 15.0
knob to change the setting. 2.0
2.12
Aespire View 6.X 2 The Aespire View
Alarm Management
Overview
Note! Refer to your User’s Reference Manual, Section 6: Alarms and Troubleshooting
for additional alarm information, including ranges and default values, and alarm tests.
2.13
Aespire View 6.X Participant Guide
O2 Sensor
The O2 Sensor, located within the Advanced Breathing System (ABS). The Galvanic Cell
measures the circuit O2. The O2 sensor is calibrated daily during the system checkout,
and can be expected to last approximately 18 months. It CANNOT be disinfected.
O2 Cell
2.14
Aespire View 6.X 2 The Aespire View
O2 Flush Button
The O2 Flush Button delivers a high flow (35 - 50 L/min) of 100% O2 to the breathing
system, bypassing the vaporizers.
MV/TV
15 0.9 10 0.9 15
14 9 14
0.8 0.8
13 13
3 4 12 8 12
5 0.7 0.7
2 11 11
6 7
10 0.6 0.6 10
1
9 6 9
1
7
8
0.5
5
0.5
8 Tec 7 Tec 7
Flow meters
8 7 0.4 7
2 0.4
6 4 6
5 0.3 0.3 5
4 3 4
3 0.2 0.2 3
2
2 2
0.1 0.1
1 1 1
ISO SEV
300
3 3 3
58 73 58 73 58 73
43 6 43 6 43 6
87 87 87
kPa kPa kPa
x x x
600
100 100 100
8 8 8
10
AIR N20 02
900 8
L/min O2
1200 4
150
200
150
200
150
200
2 100 2175
100 2175 100 2175
1500
1450 2900 1450 2900 1450 2900
kPa kPa kPa
50 725 psi x 50 725 psi x
50 725 psi x
100 100 100
N20 AIR 02
20 70
02+
max=560 cmH2O
O2 Flush button
2.15
Aespire View 6.X Participant Guide
Vaporization
Halothane
s
e
g
m
in
ti
.
ce
ll
rn
re rv ic rs n a fo r a
n n o rm .
ce n
ti te rf s
n a
t
a in e n
S ct o tio a t a
a
d a p tio
W
F e ra a n g h
e m ng uc
O e m p ri
m n d ti tr
!
s a
m a ec s
l
co e a ff l In
u
u
S p
e
e
K
o
e
p
te
da
a
e
ic
e
rv
se
t
ex
N
Tec 5
5
4
3 0%
Red
2 1
.8 .6 .4 .2
Isoflurane
Purple
s
e
g
m
in
ti
.
ce
ll
rn
re rvic rs na for a
n n orm .
ce n
ti te rf s
n a
t
a in e n
S cto tio a t a
a
d a p tio
W
F era an gh
e m ng uc
O e m pri
tr
!
s a
m a ec s
co e aff l In
l
u
u
S p
ex m d ti
e
e
K
o
e
n
p
te
da
a
e
ic
e
rv
se
t
N
Tec 7
5
4
3 0%
2 1
Enflurane
.8 .6 .4 .2
Tec 7
Sevoflurane
Yellow
g im e s
.
Tec 6 and 6 Plus
Desflurane
of
ce t s .
in at all .t r e s ce nag e nct io n
W arn h t a v e h e p h e t ic lu n
p r ig t ocl t st con .
p u au u s e in an aee d in or. s . an ce
e e ot s io n r m
k • K D o n fo r ab leb e u on it r u ct r fo ce
L oc fore
m
• N ot m m t o
• fla ly O al fo I n s t r g p e e n an
n an u al in t .
be se • O w it h e m anat io n aff e ct m ainio n s
u Se p e r or s an d d at
• O F act r v ic e m e
n
• Se com d at
e
Saf-T-fill
• re v ice r
se
xt
Ne
18
16
14
12
. 10
g im e s of
ce t s . 8 7
in at all .t r e s ce nag e nct io n 6 5
W arn h t a v e h e p h e t ic lu n 4 3 2 1
Blue
p r ig t ocl t st con .
p u au u s e in an aee d in or. s . an ce
e e ot s io n r m
k • K D o n fo r ab leb e u on it r u ct r fo ce
L oc fore
m
• N ot m m t o
• fla ly O al fo I n s t r g p e e n an
n an u al in t .
be se • O w it h e m anat io n aff e ct m ainio n s
sm g
ay
u Se p e r or s an d d at
• O F act r v ic e m e
n flo w h in se of
• Se com
• re d at
e
g as b r e att h e u
e
h low t h n d s
v ice e r
ts
w it h in m e
x
Ne
g g s w it
n in t t in r e s r e com an d .
W ar iz e r sem ix t ur on g ly in . 13 m b
18 L /m
16 ap or ox
ic a st
. se . 10 C at 10
14 h v h y p m e d in g e u to
H ig ad t o . O h on it or e for 0.2 t o 30
12
le cu it n m al b s of 18 C
10
8 7 cir y g e m an u flo w s of
ox at
t o se at u r e
6 5 fe r u r
4 3 2 1 R e d for m p e
e te
b r at ie n t
C ali am b
at
ay
sm g
flo w h in se of
g as b r e att h e u
e
h low t h n d s
g w it h in m e
n in g s w it
t t in r e s r e com an d .
W ar iz e r sem ix t ur on g ly in . 3m b
L /m
ap or ox
ic a st
. se . 10 at 101
h v h y p m e d in g e u t o 30C
H ig ad t o . O h on it or b e for of 0.2 t o
le cu it n m u al w s 18C
cir y g e m an flo s of
ox t o se atat u r e
fe r u r
R e d for m p e
e e
r at ie n t t
C alibam b
at
E
FLURAN
Y DES
USE ONL
L
TIONA
OPERA
E d
TIC nde
NO me rate
om flow
rec
hin and
wit ion id thering T
Usecentrat ive
to avodel s
NO OUTPU
conting s of trationting
set sibilitycen set l. AGENT
pos nt con n the dia E LOW
ageher thaon the FLURAN
hig icated edan.
ind to Ohmatio
Y DES WA RM
UP
er ific
Refall not USE ONL
rec RM LOW
A L A T E RY
BA T
TIONAL
OPERA
T
NO OUTPU
AGENT
LOW
UP
WA RM
RM LOW
A L A T E RY
BA T
2.16
Aespire View 6.X 2 The Aespire View
3
Vaporizer outlet port 4
5
Interlock rod
2.17
Aespire View 6.X Participant Guide
Tec 5 Filling
Keyed cap
2.18
Aespire View 6.X 2 The Aespire View
Datex-Ohmeda
Tec6 Plus
DESFLURANE
Green Operational
OPERATIONAL
indicator
Warm-up indicator
WARM-UP
Use only
Desflurane
2.19
Aespire View 6.X Participant Guide
Datex-Ohmeda
Tec6 Plus
DESFLURANE
OPERATIONAL 2
NO OUTPUT
LOW AGENT
WARM-UP
ALARM
BATTERY LOW
Use only
Desflurane
2.20
Aespire View 6.X 2 The Aespire View
Forane ®
(Isoflurane,
Easy-fil adapter
Liquid for Inhalation
Keyed cap
Agent bottle
Forane ®
(Isoflurane,
Liquid for Inhalation
2.21
Aespire View 6.X Participant Guide
30
300
600
900
1200
1500
(mL)
! ~ VT
! AGSS
10
4-1
00 %
0-0 02
05
F6
RE
4
53
51
2.22
Aespire View 6.X 2 The Aespire View
Bag/Vent Switch
The Bag/Vent Switch is a bistable mechanical switch that selects between manual
ventilation (bag) and mechanical ventilation (vent). When the switch is changed from bag
to vent mode, the ventilator is automatically switched ON.
300
600
900
1200
1500
(mL)
! ~ VT
APL Valve
REF
15
P7
! SN
The APL Valve allows you to change the settings from minimum to 70 cmH2O. When the
APL is between 30 and 70 cmH2O, you will notice that the valve becomes a tactile
indicator, meaning that the knob clicks at each additional cmH2O.
Note! There is no need for a quick release valve because the APL can go from 70
cmH2O to minimum with one turn.
At minimum – there are 2-3 cm of Intrinsic PEEP, caused by resistance of flow in the system.
300
600
900
1200
30 1500
(mL)
! ~ VT
2.23
Aespire View 6.X Participant Guide
Bag Hose
Mask holder
Rebreathing Bag
2.24
Aespire View 6.X 2 The Aespire View
Absorber Canister
!
The canisters hold 950 ml and are available as disposable or reusable. The reusable
canister is autoclavable. The Advanced
!
Breathing System also includes an EZChange
option for the canister. The EZChange sealing component seals the system so you can
O2
These canisters can be exchanged simply by pushing on the teal-colored release latch
O
and disconnecting the canister, removing it for replacement.
2
EZChange
Sealing
Component
! AGSS ! AGSS
2.25
Aespire View 6.X Participant Guide
300
600
900
1200
30
1500
(mL)
! ~ VT
! AGSS
2 3 4-1
10
00 %
0-0 02
05
2.26
Aespire View 6.X 2 The Aespire View
3. Tilt the Cradle down, and slide the canister up and out of the cradle.
!
Note! If the EZChange sealing component is not present, when the canister is
removed the system is open. It does not self-seal; you do have a leak within the
patient circuit until the canister is replaced.
2
O2
EZchange canister manifold
10
4-1
00 %
0-0 02
05
F6
RE
4
53
51
P71
! SN
! AGSS
Condenser
Drain button
2.27
Aespire View 6.X Participant Guide
Condenser
The condenser is an optional component that may be used alone or in addition to the
EZChange canister. Gas flows from the absorber canister through the condenser, where it
Save Print Change
Loop Loop Loop
allows moisture to condense out of the gas before entering the inspiratory limb. This
Spirometry
moisture is then drained out at the bottom of the condenser, and should be attended to.
Ref.
Gas
Note! Visually check the condenser reservoir daily, and drain if needed.
AGSS
! AGSS EZchange
canister
manifold
Condenser
2.28
Aespire View 6.X 2 The Aespire View
Flow Sensors
Flow Sensors are differential pressure pneumotachs that measure inspiratory pressures
and both inspiratory and expiratory volumes. Removing the Flow Sensor Module while
the Aespire View is powered ON zeros the pressure transducers.
Expiratory Flow
Sensor Inspiratory Flow
Sensor
300
600
900
1200
1500
! ~ VT
(mL)
Flow Sensor Location
P7
11
15
0
15
34
%
02
! SN
2.29
Aespire View 6.X Participant Guide
max=560 cmH2O
max=560 cmH2O
ACGO ACGO
selected
Figure 2.29 ACGO
2.30
Aespire View 6.X 2 The Aespire View
• Mainly used in operating rooms with no active gas extraction system for waste
Note! A 30 mm passive scavenging hose is also available that does not require the
purple adapter.
! AGS S
30 mm Passive
Scavenging Connector
Purple Adapter
19 mm Hose End
2.31
Aespire View 6.X Participant Guide
Active Systems:
• Open Reservoir System: Used with high vacuum (36 L/min) disposal systems. No
needle valve or visual indicator bag required. Visual flow indicator present.
• Active Adjustable Flow: Provides the capability to adjust the flow with a needle
valve and a visual indicator bag which should be properly inflated. To ensure
adequate scavenging, adjust the needle valve so the visual indicator bag puffs
out slightly with each breath. The bag should not be completely collapsed (close
the needle valve slightly) nor completely inflated (open the needle valve slightly.)
AGSS
! AGSS
Scavenging
Hose
Connector
Scavenging
Flow Indicator
Scavenging hose
connects to
hospital vacuum
system
2.32
Aespire View 6.X 2 The Aespire View
• Problems with Active Adjustable Flow gas scavenging can cause a “High Peep”
alarm. Make sure the scavenge hose is connected to suction. Make sure that the
needle valve is open enough that the scavenger bag moves with inspiration and
expiration.
! AGS S
Scavenging hose
connects to hospital
scavenging system
Adjustable
Needle Valve
3-Liter Bag
2.33
Aespire View 6.X Participant Guide
System Checkout
Note! After completing a System Checkout, always return the APL to the fully open
position (MIN).
2.34
Aespire View 6.X 2 The Aespire View
2.35
Aespire View 6.X Participant Guide
Test Procedures
Pipeline Test
1. Disconnect the pipeline supplies and close all cylinder valves.
4 4 4
5 5 5
3 3 3
58 73 58 73 58 73
43 6 43 6 43 6
87 87 87
15 0.9 10 0.9 15
14 9 14
0.8 0.8
13 13
3 4 12 8 12
5 0.7 0.7
11 11
6 7
10
9
0.6
6
0.6 10
9
Tec 7
kPa kPa kPa
x
7
x x
0.5
8
8
0.5
5
8
7
Tec 7
100
7
100 100
0.4 0.4
2
6 4 6
5 0.3 0.3 5
8
3
8 8
4 4
3 0.2 0.2 3
2
2 2
0.1 0.1
1 1 1
N20 02
ISO
AIR
SEV
AIR N 2O O2
4
5
4 4
Primary Supply (Pipeline) Pressure Gauges
5 5
3 3 3
58 73 58 73 58 73
43 6 43 6 43 6
87 87 87
kPa kPa kPa
x x x
100 100 100
8 8 8
10
AIR N20 02
8
L/min O2
6
4
150 150 150
200 200 200
2 100
1450
2175
2900
100
1450
2175
2900
100
1450
2175
2900
kPa kPa kPa
50 725 psi x 50 725 psi x
50 725 psi x
100 100 100
N20 AIR 02
150 150 150
200 100 200 100 200
100 2175 2175 2175
1450 2900 1450 2900 1450 2900
kPa kPa kPa
50 725 psi x 50 725 psi x
50 725 psi x
02+ 100 100 100
max=560 cmH2O
300 300 300
N20 AIR 02
Secondary Supply (Cylinder) Pressure Gauges
Figure 2.34 Pressure Gauges
2.36
Aespire View 6.X 2 The Aespire View
Cylinder Test
1. Disconnect the pipeline supplies and
close all cylinder valves.
2. If the pipeline and cylinder pressure
2.37
Aespire View 6.X Participant Guide
8. Adjust the flow of all gases through the full range and make sure that the flow tube
floats move smoothly.
9. Disconnect the O2 pipeline supply or close the O2 cylinder valve.
Note! Nitrous oxide (N2O), if available, flows through the system during this test.
Use a safe and approved procedure to collect and remove the N2O.
2.38
Aespire View 6.X 2 The Aespire View
15 0.9 10 0.9 15
14 9 14
13 0.8 0.8
13
12 8 12
0.7 0.7
11 7 11
10 0.6 0.6 10
9 6 9
0.5 0.5
8 8
5
7 0.4 7
0.4
6 4 6
5 0.3 0.3 5
4 3 4
3 0.2 0.2 3
2
2 2
0.1 0.1
1 1 1
AIR N 2O O2
2.39
Aespire View 6.X Participant Guide
4 3 4 3 2 1 .8 .6 .4 .2 0
Tec 7
Figure 2.38 Adjust the vaporizer concentration
• If the O2 flow decreases more than 1 l/min, install a different vaporizer and repeat
the steps 1 through 4
• If the O2 flow decreases less than 1 l/min when testing a different vaporizer, the
malfunction is in the vaporizer that failed the test
• If the O2 flow decreases more than 1 l/min with the different vaporizer, the
malfunction is most likely in the system. Do not use the system
5. Repeat steps 1 through 4 for each vaporizer.
15 0.9 10 0.9 15
14 9 14
0.8 0.8
13 13
3 4 12 8 12
5 0.7 0.7
2 11 11
6 7
10 0.6 0.6 10
1
9 6 9
1
7
8
0.5
5
0.5
8 Tec 7 Tec 7
8 7 0.4 7
2 0.4
6 4 6
5 0.3 0.3 5
4 3 4
3 0.2 0.2 3
2
2 2
0.1 0.1
1 1 1
ISO SEV
AIR N 2O O2
4 4 4
5 5 5
300
3 3 3
58 73 58 73 58 73
43 6 43 6 43 6
87 87 87
kPa kPa kPa
x x x
600
100 100 100
8 8 8
10
AIR N20 02
900 8
L/min O2
1200 4
150
200
150
200
150
200
2 100 2175
100 2175 100 2175
1500
1450 2900 1450 2900 1450 2900
kPa kPa kPa
50 725 psi x 50 725 psi x
50 725 psi x
100 100 100
N20 AIR 02
20 70
02+
max=560 cmH2O
2.40
Aespire View 6.X 2 The Aespire View
1 0
Tec 7
ISO
max=560 cmH2O
ACGO position
2.41
Aespire View 6.X Participant Guide
4. Compress and release the bulb until all air is removed from the bulb.
• Occlude the inlet of the test device. Make sure it is a tight seal
• If the bulb of the test device inflates in less than 60 seconds,
use a different test device
max=560 cmH2O
• The floats will move. If the bulb inflates in 30 seconds or less, there
is a leak in the low-pressure circuit. See Pneumatic problems in the
Alarms and Troubleshooting section of the User’s Reference Manual
• Disconnect the test device
2.42
Aespire View 6.X 2 The Aespire View
Tec 7
• Perform step 5
• Repeat this test with each vaporizer
• If a low-pressure leak occurs while testing any of the vaporizers, see
Pneumatic problems in the Alarms and Troubleshooting section of the
User’s Reference Manual
• Turn the vaporizer off
7. Turn all flow controls fully clockwise for minimum flow. Do not over tighten.
8. Clear the system of agent.
• Set the System switch to On
Note! Agent mixtures from the low-pressure leak test stay in the system. Clear the
system by flowing O2 at 1 l/min for one minute.
2.43
Aespire View 6.X Participant Guide
Alarm Tests
1. Connect a test lung to the patient connection.
2. Set the Bag/Vent switch to Vent.
2.44
Aespire View 6.X 2 The Aespire View
2
0.1
1
NO22O
2.45
Aespire View 6.X Participant Guide
1032Off
40
O2
40
. . . 21
. 0. . . . . . 99
................. PpauseOff
MV
30
10 . . . 3.0
. . . . . . . 7.0
. . . . . . . . . . . . . . . . O2
. %
Flow 60 MV I/min
726.25
30
TVexp 200 600
20 F 9.0 100
Leak Audio
20 5 On
0
l
o
TV exp ml
3.0
25
10 w
10
1
TVexp
625 1000 ml
590
0 ........................... Off
-60 1450
0 EtCO2 mmHg FiCO2 mmHg
450 Menu
0.5
Go to Main Menu
35
C 60
Off 0 key
0 2 4
. . . .6 . . . 8. . . .10. . . 12. . . . 14. . . .16. . . . . MV O
2 EtO2 %
23
l/min RR l/min
CO2 40 /
FiO2 %
N2O O2 Time (sec)
5.9 50 10
50
O 15.0 Off
3.0
Off
3.0 2 2.0 Off Off
0
Vent OffOn Gases
Vent VCV - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV RR I:E Pmax PEEP
cmH2O O2 3.1
4.3
% I/min % cmH2O /min More Off 8.0
590 10 1:1.5 40 5 Settings 5.4 Off
Air Sev 0.10 5.9 Off
50 ml 6.00/min 6.0 23
cmH2O 10
cmH2O Off N20N2O 1.148 59
Total
12. Test the O2 monitor and alarms (alarms other than Low O2 and High O2 may occur).
• Remove the O2 cell, and make sure that the cell measures approximately
21% O2 in room air.
• Push the Menu key.
• Select Alarm Setup from the Main menu.
Main Menu
Ventilation Mode
Alarm Setup
Setup/Calibration
Screen and Audio Setup
Cardiac Bypass Off
Normal Screen
• Set the Low O2 alarm to 50%, and make sure that a Low O2 alarm occurs.
• Set the Low O2 alarm to 21%, and make sure that the Low O2 alarm stops.
This will create a latched alarm, acknowledge this by pushing the
Alarm silence key.
• Put the O2 cell back into the circuit.
• Set the High O2 alarm to 50%.
2.46
Aespire View 6.X 2 The Aespire View
• Push the O2 flush button to fill the breathing system, and make sure that the
High O2 alarm occurs.
• Set the High O2 alarm to Off, and make sure that the alarm stops.
• Flow 100% O2 for 2 minutes, and make sure that the O2 cell measures 100% O2.
20
APL
30
! ~ cm H2O
70 MIN
20. Occlude the patient connection and push the O2 flush button.
21. Make sure that the Ppeak high. Blockage? (sustained airway pressure) alarm occurs
after approximately 15 seconds at the sustained pressure limit.
2.47
Aespire View 6.X Participant Guide
Note! After performing the breathing system tests, make sure that there are no test
plugs or other objects caught in the breathing system.
2.48
Aespire View 6.X 2 The Aespire View
5. Push the O2 flush button to fill the bellows. Release the O2 flush button.
6. Make sure that the pressure does not increase to more than 15 cmH2O on the
pressure gauge.
7. If the bellows falls lower than the top of the indicator, there is a leak. See Breathing
20
APL
30
! ~ cm H2O
70 MIN
2.49
Aespire View 6.X Participant Guide
20
APL
30
! ~ cm H2O
70 MIN
5. Set the O2 flow to 3 l/min. Make sure that the value on the inspiratory pressure gauge
does not exceed 85 cmH2O. Some pressure fluctuation is normal.
6. Set the APL valve to MIN.
7. Make sure that the value on the inspiratory pressure gauge is less than
approximately 5 cmH2O.
8. Push the O2 flush button. Make sure that the value on the inspiratory pressure gauge
stays near zero.
9. Set the O2 flow to minimum and make sure that the value on the inspiratory pressure
gauge does not decrease below 0 cmH2O.
2.50
Aespire View 6.X 2 The Aespire View
Ventilation
7900 Ventilator
Modes
In addition to Volume Control Ventilation (VCV), the Aespire View can provide Pressure
Control Ventilation (PCV), Synchronized Intermittent Mandatory Ventilation (SIMV) with
pressure support, Pressure Support Ventilation with Apnea protection backup mode
(PSVPro), SIMV with Pressure Control, and Pressure Control Ventilation with Volume
Guarantee (PCV-VG).
Each of these modes will be described in greater detail on the following pages.
2.51
Aespire View 6.X Participant Guide
Pressure
Time
Figure 2.43 Pressure Waveform Shark fin Pattern
Flow
Time
Figure 2.44 Flow Waveform Square Inspiratory Pattern
2.52
Aespire View 6.X 2 The Aespire View
Set Pressure
Time
Figure 2.45 Pressure Waveform
Flow
Time
Figure 2.46 Flow Waveform
2.53
Aespire View 6.X Participant Guide
SIMV/PSV
Synchronized Intermittent Mandatory Ventilation with Pressure Support (SIMV/PSV) is a
mode in which periodic volume breaths are delivered to the patient at preset intervals
(time-triggered). Between the machine delivered breaths, the patient can breathe
spontaneously at the rate, tidal volume, and timing that the patient desires.
At the specified time interval, the ventilator will wait for the next inspiratory effort from
the patient. The sensitivity of this effort is adjusted using the flow trigger level. When the
ventilator senses the beginning of inspiration it synchronously delivers a volume breath
using the set tidal volume, and inspiratory time that is set on the ventilator. If the patient
fails to make an inspiratory effort during the trigger window time interval, the ventilator
will deliver a machine breath to the patient. The ventilator will always deliver the specific
number of breaths per minute that the clinician has set.
In SIMV/PSV, the spontaneous breaths can be pressure supported to assist the patient in
overcoming the resistance of the patient circuit and the artificial airway. When the
Psupport level is set, the ventilator will deliver the pressure support level to the patient
during inspiration. PEEP can also be used in combination with this mode.
Window
I E
2.54
Aespire View 6.X 2 The Aespire View
PSVPro
PSVPro is pressure supported ventilation with apnea backup. PSVPro is a spontaneous mode
of ventilation that provides a constant pressure once the ventilator senses that the patient
has made an inspiratory effort. In this mode, the clinician sets the Pressure Support (Psupport)
2.55
Aespire View 6.X Participant Guide
SIMV-PC
Synchronized Intermittent Mandatory Ventilation with Pressure Controlled Ventilation is a
mode in which a relatively slow mandatory breathing rate is set with pressure-controlled
breathing. This mode combines mandatory breaths with spontaneous breath support.
If a trigger event occurs within the synchronization window, a new pressure-controlled
breath is initiated. If a trigger event occurs elsewhere during the expiratory phase, a
support for a spontaneous breath is provided with pressure support added as set by
the clinician.
Spontaneous Trigger
Tinsp
breathing period
window
Paw waveform
Pinsp
Pressure
supported
breath
Spontaneous Trigger
Tinsp
breathing period
window
Flow waveform
2.56
Aespire View 6.X 2 The Aespire View
PCV-VG
In PCV-VG, a tidal volume is set and the ventilator delivers that volume using a
decelerating flow and a constant pressure. The ventilator will adjust the inspiratory
pressure needed to deliver the set tidal volume breath-by-breath so that the lowest
Tinsp Texp
Paw waveform
Variable pressure to
deliver desired TV
Peep
Time
Tinsp Texp
Flow waveform
TV TV
Time
2.57
Aespire View 6.X Participant Guide
Ventilation Mode
7
cmH2O
39 2
Main
Paw
Menu
w
Ventilation
Paw 15
Mode VCV PEEP
Pmean 24
RR /min
1032
40
. . . . 0. . PCV
.....................
4
Off
40Alarm Setup Ppause Off
30
10
Setup/Calibration . . . . . . SIMV/PSV
. . . . . . . . . . . . . . . . . . . . O2
. %
Flow 60 MV I/min
726.25
30
Screen and Audio Setup PSVPro
20 F 9.0 100
5
20Cardiac Bypass 0
SIMV-PC l
o
TV exp ml
3.0
25
PCV-VG w
1
10
10
1
TVexp
625 1000 ml
590
0
Normal Screen
........................... Off
-60 1450
0 EtCO2 mmHg FiCO2 mmHg
450
0.5 35
C 60
0 Off
. . . .6 . . . 8. . . .10. . . 12. . . . 14. . . .16. . . . MV
. O2
5
23
0 2 4
l/min RR l/min
CO2 40 /
EtO2 % FiO2 %
N2O O2 Time (sec)
5.9 50 10
50
O
3.0
Off Off
3.0
15.0
2 2.0 Off Off
0
Vent OffOn Gases
Vent PSVPro - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV RR I:E Pmax PEEP
cmH2O O2 3.1
4.3
% I/min % cmH2O /min More Off 8.0
590 10 1:1.5 40 5 Settings 5.4 Off
Air Sev 0.10 5.9 Off
50 ml 6.00/min 6.0 23
cmH2O 10
cmH2O Off N20N2O 1.148 59
Total
2.58
Aespire View 6.X 2 The Aespire View
Alarm Setup
Use the volume alarms key to turn the volume alarms on and off. When the alarms are
off, an X covers the alarm limits. Use this control during manual ventilation when constant
attention is on the patient.
Note! Do not turn off volume alarms for a spontaneously breathing patient. The
system will not alarm for low volume.
39 2
Paw Setup
Alarm w
Paw 15 PEEP
Pmean 24
RR /min
1032Off
40
O2
40
. . . 21
. 0. . . . . . 99
................. PpauseOff
. . . 3.0
. . . . . . . 7.0
. . . . . . . . . . . . . . . . O2
.
3 MV
30
10 Flow 60 MV I/min %
726.25
30
TVexp 200 600
20 F 9.0 100
Leak Audio
20 5 On
0
l
o
TV exp ml
3.0
25
10 w
10
1
TVexp
625 1000 ml
590
0 ........................... Off
-60
1
1450
0
5
EtCO2 mmHg FiCO2 mmHg
450
0.5
Go to Main Menu
35
C 60
Off 0
0 2 4
. . . .6 . . . 8. . . .10. . . 12. . . . 14. . . .16. . . . . MV O
2 EtO2 %
23
l/min RR l/min
CO2 40 /
FiO2 %
N2O O2 Time (sec)
5.9 50 10
50
O 15.0 Off
3.0
Off
3.0 2 2.0 Off Off
0
Vent OffOn Gases
Vent VCV - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV RR I:E Pmax PEEP
cmH2O O2 3.1
4.3
% I/min % cmH2O /min More Off 8.0
590 10 1:1.5 40 5 Settings 5.4 Off
Air Sev 0.10 5.9 Off
50 ml 6.00/min 6.0 23
cmH2O 10
cmH2O Off N20N2O 48
1.1 59
Total
2.59
Aespire View 6.X Participant Guide
Setup/Calibration
1. Push the Menu key.
2. Select Setup/Calibration from the Main Menu. The following window will appear:
39 2
a
Ventilation Mode Paw w
Paw 15
Setup/Calibration PEEP
Pmean 24
RR /min
1032
40
Alarm Setup More
40
. . . . 0. . . . . . . . . . . . . . . . . . . . . . .
Vent Settings Ppause
Off
Off
2 Setup/Calibration
10 . . . . . . . . . . . . . . . . . . . . . . . . . . O2
30 Sensor Calibration
O2 Flow 60
.
MV I/min %
726.25
30
Screen and Audio Setup
About
20
Ventilator . . . F 9.0 100
Cardiac Bypass 20 5 0
l
o
TV exp ml
3.0
25
10 w
10
1
TVexp
625 1000 ml
590
0 ........................... Off
5.9 50 10
50
O
3.0
Off Off
3.0
15.0
2 2.0 Off Off
0
Vent OffOn Gases
Vent VCV - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV RR I:E Pmax PEEP
cmH2O O2 3.1
4.3
% I/min % cmH2O /min More Off 8.0
590 10 1:1.5 40 5 Settings 5.4 Off
Air Sev 0.10 5.9 Off
50 ml 6.00/min 6.0 23
cmH2O 10
cmH2O Off N20N2O 1.148 59
Total
3. Use the ComWheel to scroll to the desired submenu. Push the ComWheel to confirm
the selection. Select More Vent Settings to set the following ventilator values:
• Pmax
• Trig Window
• Trigger
• End of Breath
• Backup Mode Active
• Tpause
• Rise Rate
4. After selecting a ventilator setting, set it to the desired value by turning the
ComWheel. Confirm the value is correct by pushing the ComWheel.
5. Select Go to Setup/Calibration Menu to return to the Setup/Calibration, or push the
Menu key to return to the normal monitoring screen.
2.60
Aespire View 6.X 2 The Aespire View
Cardiac Bypass
Cardiac Bypass suspends alarms for patients on cardiac bypass when the ventilator is
not mechanically ventilating. Mechanical ventilation must be off. When mechanical
ventilation is turned on, Cardiac Bypass is automatically turned off, the alarms are
Note! Cardiac Bypass mode should only be used when the patient is receiving
extra-corporeal oxygenation by means of a heart-lung machine. This mode of
ventilation is not intended to provide metabolic levels of ventilation to the patient.
1
MV/TV Apnea alarm standby
. . . . . . . . . .Battery
. . .charger
. . .fail. . . . . . . . . . . cmH2O
Measured
120 Paw 30 Ppeak Pmean
I/min . Pa 25
. . . . 15. . . . . . . . . . . . . . . . . . . . . . Ppeak 40
7
cmH2O
39 2
Main Menu
Paw w
Ventilation
Paw 15
Mode PEEP
Pmean 24
RR /min
1032Off
40
40
Alarm Setup
. . . . 0. . . . . . . . . . . . . . . . . . . . . . . PpauseOff
10
Setup/Calibration
30 . . . . . . . . . . . . . . . . . . . . . . . . . . O2
. %
Flow 60 MV I/min
726.25
30
Screen and Audio Setup
3 20
20
Cardiac Bypass
5
On
0
F
l
o
9.0 100
3.0
25
w TV exp ml
10
4 10
1
TVexp
625 1000 ml
590
Normal
0 Screen ........................... Off
2
-60 1450
0 EtCO2 mmHg FiCO2 mmHg
450
0.5 35
C 60
Off 0
0 2 4
. . . .6 . . . 8. . . .10. . . 12. . . . 14. . . .16. . . . . MV O
2 EtO2 %
23
l/min RR l/min
CO2 40 /
FiO2 %
N2O O2 Time (sec)
5.9 50 10
50
O 15.0 Off
3.0
Off
3.0 2 2.0 Off Off
0
Vent OffOn Gases
Vent VCV - Volume Control Circle Fresh Gas Flow l/min
O2 Total Flow Sev Pinsp RR PEEP % Et Fi
TV RR I:E Pmax PEEP
cmH2O O2 3.1
4.3
% I/min % cmH2O /min More Off 8.0
590 10 1:1.5 40 5 Settings 5.4 Off
Air Sev 0.10 5.9 Off
50 ml 6.00/min 6.0 23
cmH2O 10
cmH2O Off N20N2O 1.148 59
Total
2.61
Aespire View 6.X Participant Guide
20 70
30
1 ! AGSS
30
30
! AGSS
! AGSS
3 4
Figure 2.54 Removing the Breathing System
2.62
Aespire View 6.X 2 The Aespire View
10
4-1
00 %
0-0 02
05
F6
RE
34
15
15
P7
! SN
2.63
Aespire View 6.X Participant Guide
20 70
Thumbscrew
2.64
Aespire View 6.X 2 The Aespire View
2.65
Aespire View 6.X Participant Guide
1. If applicable - Remove the O2 cell cable from the cell. Unscrew the O2 cell
counterclockwise and remove it. Remove the O2 cell cable by pressing on the
connector button while pulling the connector out.
2. Rotate the breathing circuit module counterclockwise. After rotating, pull the
breathing circuit module apart from the bellows assembly by lifting up on the
breathing circuit module.
3. On the breathing circuit module, remove the check valve lens by squeezing the
latches together and pulling up on the lens. Lift out the check valve assemblies.
Turn O2 Sensor
counterclockwise
Check valve 0
4-11
lens 6050
-000
REF
release latches
! R7
SN
Pins in bulkhead
wall insert through
these holes O2 Sensor Cable
Absorber Canister
Release Lever O2 Sensor cable
connector button
Figure 2.59 Removing the Breathing Circuit Module
2.66
Aespire View 6.X 2 The Aespire View
2. Insert the breathing circuit module into the bellows assembly aligned as shown.
3. Rotate the breathing circuit module clockwise at the point shown by the dotted line to
attach it to the bellows assembly.
4. If applicable – Replace the cell by screwing it in clockwise. Reconnect the O2 cell cable.
2.67
Aespire View 6.X Participant Guide
Bellows Assembly
Disassemble the bellows assembly:
1. Turn the bellows housing counterclockwise and lift.
2. Remove the bottom edge of the bellows and lift.
3. Push the two tabs toward the center and remove the rim.
4. Remove the pressure relief valve.
5. Push the latch toward the center and remove the locking tabs.
6. Remove the seal.
300
Bellows housing
600
900
1200
1500
(mL)
! ~ VT
Bellows
Locking tabs
Seal
Rim
2.68
Aespire View 6.X 2 The Aespire View
300
600
900
1200
1500
(mL)
! ~ VT
2.69
Aespire View 6.X Participant Guide
Exhalation Valve
Remove the exhalation valve assembly:
1. With the breathing system removed, the exhalation valve assembly can be removed.
2. Loosen the thumbscrews and lift the assembly off.
Relief
Valve
Thumbscrews
Vented
Cover
Base
Ventilator housing
2.70
Aespire View 6.X 3 Resources
3 Resources
Aespire View User Operational Maintenance Schedule
Activity Frequency
3 Resources
Suction Overflow Trap PRN: (URM sect. 7-3)
3.1
Aespire View 6.X 3 Resources
8004515 Reusable, hytrel patient tubing 36” (for use on bag arm)
8004459 ABS Bag arm connector elbow (for use on bag arm)
Vaporizer Supplies
1100-3025-000 Easy Fill Bottle Adapter, Isoflurane
**These items are supplies that will be used regularly and should be stocked. The others are for reserve or
back-up and are optional.
3.2
Aespire View 6.X 3 Resources
Aespire View Troubleshooting Guide Taken from pages 6-4 through 6-8 of the User’s Reference Manual
3 Resources
volume for at least 30 seconds (mechanical connections. Patient circuit leak audio can be
ventilation). turned off in the Alarm Setup menu.
No exp flow sensor; Electrical signals show the flow sensor is not Connect the flow sensors. Make sure the flow
No insp flow sensor connected. sensor module is in all the way.
PEEP high. Paw ≥ sustained limit for 15 seconds. Check tubing for kinks, blockages, disconnects.
Blockage? Check scavenging. Calibrate the flow sensors.
Plug in power The mains supply is not connected or has Ventilate manually to save power. At full
cable. failed and the system is using battery power. charge, the battery permits approximately 90
On battery. minutes of mechanical ventilation. Make sure
power is connected and circuit breakers are
closed.
Reverse exp flow. Flow toward the patient seen in the expiratory Look at the check valves. Check for water
flow sensor during inspiration for six breaths in buildup in the flow sensor tubes. Check the
a row. flow sensor condition. Replace the expiratory
check valve.
Reverse insp flow. Flow through inspiratory sensor during Look at the check valves. Check for water
expiration for six breaths in a row. buildup in the flow sensor tubes. Check the
flow sensor condition. Replace the expiratory
check valve.
TV not achieved Tidal volume measured by inspiratory flow Adjust controls to supply adequate tidal vol-
sensor is less than the set value for six breaths umes. Check I:E, Pmax, and volume settings.
in a row after the first minute of mechanical Possible leak. Modify settings or check for
ventilation. system leaks.
Unable to drive Manifold pressure is greater than Paw. Check the drive gas. Increase fresh gas flow (or
bellows push O2 flush button) to fill bellows.
Vol vent only. Manifold pressure error. Pressure Control Continue to use volume control ventilation or
No PEEP or PSV. unavailable. ventilate manually. Shut down system as soon
as possible. Contact a Datex-Ohmeda trained
service representative.
3.3
Aespire View 6.X 4 Clinical Training Documents
The purpose of this document is to verify that the following items are covered during
your Aespire View training. Training sessions have been separated into two sections.
The lecture portion of the training is approximately 60 minutes in duration. The hands-
on portion for practice and return demonstrations is approximately 60 minutes in
duration, as well. Please plan to be in the training class for the entire time that your
manager has assigned.
You will be given a Training Quiz and a Course & Instructor Evaluation at the end of the
class. The evaluation needs to be returned with the completed quiz to your manager.
Your manager will give you all of the correct answers to the quiz once you complete it.
4 Clinical Training
You should be prepared to operate the Aespire View by the end of the training time.
Documents
If you still believe that you are not prepared, please talk to your trainer or your
manager. Please refer to the Aespire View User’s Reference Manuals for complete
instructions on the operation of the Aespire View Anesthesia System.
Vaporization
• Filling the Tec 7 vaporizers
• Remove and replace the Tec 7 vaporizers
4.1
Aespire View 6.X 4 Clinical Training Documents
Ventilation
• Basic Modes: Volume Control, Pressure Control
Hands-on exercises
Training Quiz
4.2
Aespire View 6.X 4 Clinical Training Documents
Demonstrate the ability to calibrate (remove and replace) the flow sensor module.
Demonstrate the ability to remove and reconnect the CO2 absorber canister.
4 Clinical Training
• APL Valve test
Documents
Demonstrate the ability to start a case, starting with the machine that is turned off.
Demonstrate the ability to completely set ventilator parameters and then change from
one mode of ventilation to another.
Demonstrate the ability to change alarm limits, use the Alarm Silence and Volume
Alarms keys.
Turn suction on to maximum. Change to regulated suction and adjust suction pressure.
4.3
Aespire View 6.X 4 Clinical Training Documents
After completion of the test, compare your answers to the answer key. Ask the
instructor to explain any answers you do not understand.
1. How long with the battery provide power to the machine when fully charged and
under normal use?
A. 15 minutes
B. 90 minutes
C. 4 hours
D. 8 hours
5. It is possible to delivery 100% N2O to the patient through the Aespire View.
A. True
B. False
6. How is patient gas carried between the rebreathing bag and the ABS (Advanced
Breathing System)?
A. Through the bag arm
B. Through the bag hose
4.4
Aespire View 6.X 4 Clinical Training Documents
8. During Pressure Control Ventilation (PCV), what is the typical pressure waveform pattern?
A. Square wave pattern
B. “Shark fin” wave pattern
C. Decelerating wave pattern
4 Clinical Training
11. High priority warning alarms display white text and what color background?
Documents
A. Red
B. Yellow
C. White
12. PSVPro is a mode of ventilation that works well with paralyzed patients.
A. True
B. False
14. At the end of the machine checkout, the APL valve (pop off) should be placed in what
position?
A. 70 cmH2O
B. Minimum
C. 30 cmH2O
15. Flow sensors measure airway flow, calculate pressure changes, and display both
inspiratory and expiratory volumes on the ventilator screen.
A. True
B. False
4.5
Aespire View 6.X 4 Clinical Training Documents
19. What is the purpose of the Volume Alarms key on the ventilator display?
A. Temporarily silences any active alarms
B. Permanently silences any active alarms
C. Mutes the audio tone for alarms
D. Turns the volume alarms (tidal volume and minute volume) on and off
4.6
Aespire View 6.X 4 Clinical Training Documents
1. How long with the battery provide power to the machine when fully charged and
under normal use?
Answers B. 90 minutes
2. Where is the pipeline and tank pressures displayed?
C. Front of the machine
3. A Preoperative Checkout should be performed
C. Every day before your first patient
4. What is the purpose of the End Case key?
A. Minimizes alarms between cases
5. It is possible to delivery 100% N2O to the patient through the Aespire View.
B. False
6. How is patient gas carried between the rebreathing bag and the ABS (Advanced
4 Clinical Training
Breathing System)?
Documents
B. Through the bag hose
8. During Pressure Control Ventilation (PCV), what is the typical pressure waveform
pattern?
A. Square wave pattern
9. An active alarm can be silenced for _________ seconds.
D. 120
14. At the end of the machine checkout, the APL valve (pop off) should be placed in
what position?
B. Minimum
4.7
Aespire View 6.X 4 Clinical Training Documents
15. Flow sensors measure airway flow, calculate pressure changes, and display both
inspiratory and expiratory volumes on the ventilator screen.
A. True
17. The ventilator does not have to be on to provide Pressure Support Ventilation.
B. False
4.8
Aespire View 6.X 5 Non-clinical Training Documents
The purpose of this document is to verify that the following items are covered during
your Aespire View training. Training sessions have been separated into two sections.
The lecture portion of the training is approximately 60 minutes in duration. The hands-
on portion for practice and return demonstrations is approximately 60 minutes in
duration, as well. Please plan to be in the training class for the entire time that your
manager has assigned.
You will be given a Training Quiz and a Course & Instructor Evaluation at the end of the
class. The evaluation needs to be returned with the completed quiz to your manager.
Your manager will give you all of the correct answers to the quiz once you complete it.
You should be prepared to operate the Aespire View by the end of the training time.
If you still believe that you are not prepared, please talk to your trainer or your
manager. Please refer to the Aespire View User’s Reference Manuals for complete
instructions on the operation of the Aespire View Anesthesia System.
5 Non-clinical Training
Electrical and pneumatic systems including battery back-up power
Documents
Gas regulation and flow control
• Flow controls, Auxiliary O2
Vaporization
• Filling the Tec 7 vaporizers
• Remove and replace the Tec 7 vaporizers
5.1
Aespire View 6.X 5 Non-clinical Training Documents
5.2
Aespire View 6.X 5 Non-clinical Training Documents
Demonstrate the ability to calibrate (remove and replace) the flow sensor module.
Demonstrate the ability to remove and reconnect the CO2 absorber canister.
Demonstrate the ability to empty water from the condenser reservoir.
Demonstrate the ability to power on the machine when the power is off.
Demonstrate the ability to end a case to put alarms into standby.
Demonstrate the ability to perform a complete machine check.
• Pipeline test
• Cylinder test
• Flow Controls test
• Vaporizer Back Pressure test
• Negative Low-Pressure Leak test
• Alarm test
• Bellows test
• Circuit test
5 Non-clinical Training
• APL Valve test
Documents
Turn suction on to maximum. Change to regulated suction and adjust suction pressure.
Demonstrate the ability to replace/clean suction overflow trap.
5.3
Aespire View 6.X 5 Non-clinical Training Documents
After completion of the test, compare your answers to the answer key. Ask the
instructor to explain any answers you do not understand.
1. How long with the battery provide power to the machine when fully charged and under
normal use?
A. 15 minutes
B. 90 minutes
C. 4 hours
D. 8 hours
3. What will happen on a machine with a charged battery if the system switch is turned on,
and the system circuit breaker (main switch on the rear of the machine) is turned off?
A. Nothing, the machine will not power up
B. The message ‘Plug in power cable. On battery’ will be displayed
C. The machine will run on battery
D. Both B and C
6. How is patient gas carried between the rebreathing bag and the ABS (Advanced
Breathing System)?
A. Through the bag arm
B. Through the bag hose
5.4
Aespire View 6.X 5 Non-clinical Training Documents
10. High priority warning alarms display white text and what color background?
A. Red
B. Yellow
C. White
5 Non-clinical Training
12. If a vaporizer was improperly installed and had a leak, which of the following tests would
most likely find the problem?
A. Pipeline test
Documents
B. Cylinder test
C. Flow Control test
D. Negative Low-Pressure Leak test
13. At the end of the machine checkout, the APL valve (pop off) should be placed in what
position?
A. 70 cmH2O
B. Minimum
C. 30 cmH2O
14. Flow sensors measure airway flow, calculate pressure changes, and display both
inspiratory and expiratory volumes on the ventilator screen.
True False
5.5
Aespire View 6.X 5 Non-clinical Training Documents
16. What is the purpose of the Volume Alarms key on the ventilator display?
A. Temporarily silences any active alarms
B. Permanently silences any active alarms
C. Mutes the audio tone for alarms
D. Turns the volume alarms (tidal volume and minute volume) on and off
17. The gas cylinder valves should be in what position if the pipeline supply is in use?
A. Closed
B. Fully open
C. 50%
18. It is OK to fill a vaporizer while the vaporizer is in use and turned on.
True False
5.6
Aespire View 6.X 5 Non-clinical Training Documents
1. How long with the battery provide power to the machine when fully charged and under
normal use?
Answers B. 90 minutes
2. Where is the pipeline and tank pressures displayed?
C. Front of the machine
3. What will happen on a machine with a charged battery if the system switch is turned on,
and the system circuit breaker (main switch on the rear of the machine) is turned off?
D. Both B and C
5 Non-clinical Training
8. An active alarm can be silenced for _________ seconds.
Documents
D. 120
12. If a vaporizer was improperly installed and had a leak, which of the following tests would
most likely find the problem?
D. Negative Low-Pressure Leak test
13. At the end of the machine checkout, the APL valve (pop off) should be placed in what
position?
B. Minimum
5.7
Aespire View 6.X 5 Non-clinical Training Documents
14. Flow sensors measure airway flow, calculate pressure changes, and display both
inspiratory and expiratory volumes on the ventilator screen.
True
17. The gas cylinder valves should be in what position if the pipeline supply is in use?
A. Closed
18. It is OK to fill a vaporizer while the vaporizer is in use and turned on.
False
5.8
Aespire View 6.X 6 Course Evaluation
4 Course Evaluation
Aespire View 6.X
Course Name: Date:
Location: Instructor:
Please complete the following survey. The information you provide will help us to improve the course for
future learners.
Rate the training by circling the appropriate number
The knowledge checks and/or assessments in this course were effective in helping
4 3 2 1 NA
me validate my existing and acquired knowledge.
Comments:
6 Course Evaluation
Course Experience 10 = Strongly agree 1 = Strongly disagree
Would you like to be contacted in the future for further inputs on our training development, delivery, and
operations processes?
6.1
© 2008 General Electric Company – All rights reserved.
GE and GE Monogram are trademarks of General Electric Company.
Healthcare Re-imagined
GE is dedicated to helping you transform healthcare delivery
by driving critical breakthroughs in biology and technology.
Our expertise in medical imaging and information technologies,
medical diagnostics, patient monitoring systems, drug discovery,
and biopharmaceutical manufacturing technologies is enabling
healthcare professionals around the world to discover new ways
to predict, diagnose and treat disease earlier. We call this model
of care “Early Health.” The goal: to help clinicians detect disease
earlier, access more information and intervene earlier with more
targeted treatments, so they can help their patients live their lives
to the fullest. Re-think, Re-discover, Re-invent, Re-imagine.
GE Healthcare
3030 Ohmeda Drive
Madison, WI 53718
USA
www.gehealthcare.com
EDU-LSS-C-Aespire 6.X-PG-revA
DOC0679577
Printed in U.S.A.