You are on page 1of 178

Document No.

: M517-E429
Revision code: R

Digital Radiography System


DAR-8000
Remote Maintenance Installation Manual

This manual is intended for technical service engineers.


Its contents are not directly related to daily operation.
This precaution and prohibitions seen through the manual are classified as follows:

States a direct danger that may cause death or serious injury if


it is not avoided.

States an indirect or potential danger that may cause death or


serious injury if it is not avoided.

States a danger that may cause slight or medium injury or may


cause damage in equipment or fire if it is not avoided.

States the information that helps to use the system correctly.

M517- E429R
Revision Record
Revision Date Content
- Create first edition
Change the installation procedure by changing MPC.
Change the name of unit "PPWS" to "Side Station".
Change the port for the LAN cable of DAR-8000.
Add the description on the installation site for the data logger.
Add the note when the same segment is used in on-site
A 2006.08 network.
Change the procedure for connecting with Remote Desktop.
Change the procedure for DAR-8000 Network Settings.
Add "5-2-10 Setting of PC automatic start when power supply
returns."
Add "3 Preparation for Setup".
B 2007.03
Add "9 Special Operation".
C 2007.08 The remote maintenance supports the broadband.
D 2007.12 Revised according to the change of production system.
E 2009.05 Revised according to the change of MPC.
Change the Xcat setting in "9.3.3. Remote Maintenance Setting".
F 2010.04
Add "11 BBRM Upgrade Procedure".

G 2010.07 Revised according to the supports for Super safire


Add "11. Update of PCU Error Monitoring Tool (Only
H 2011.02
DAR-8000f)".
I - -

J - -

Divide schedule setting according to the Air-cooling FPD and the


Water Cooling FPD.
Support the Water-cooling FPD and add the processing for
monitoring No.2 in PCU Error Monitoring Tool.
Add settings of the Side Station (Windows7).
Revised according to the support of the BBRM version 3.0.1
K 2012.09 (Agent Service Link5.2)
Revised "9.2 How to Use Service Tools for BBRM-Agent".
- Move the common tool independent of the version to section
9.2.1.
- Separate the tool depending on versions into sections 9.2.2
and 9.2.3.
Revised "10 BBRM Upgrade Procedure".

M517- E429R
- Add "10.2 How to Upgrade BBRM to Version 3.0.1".
(Move the upgrading method to 2.3.0. to section 12.1.)
Changes the installation procedure of section 8.2.2. "Installing
and setting the UltraVNC in the DAR-8000" according to the
upgrade to UltraVNC1.0.9.6.2.
Correct the setting method in section 1.5 according to installed
OS.
Amendment of the following in accordance with the addition of
UPS made by OMRON.
"2. System Configuration"
Adding the cable (13) made by OMRON on "Table 1-1 Remote
Maintenance Cables and Connection Destinations".
"7.3. Connecting the UPS and MPC (DAR-8000f Only) "
Each connection of UPS made by APC and UPS made by
OMRON is indicated(7.3.1., 7.3.2.).
"8.1.2. Settings and Function of UPS control S/W(DAR-8000f
L 2013.11
Only) "
The title is changed.
Each UPS control SW of UPS made by APC and UPS made
by OMRON is indicated.
"8.1.3. Checking UPS Operation and Restarting the System
(DAR-8000f Only)"
Adding the check steps of UPS made by OMRON.
Clerical error correction, change of the termination parts of
manufacture
Description of UPS is divided to OMRON and APC.
“8.1. Setting and Checking Devices”
Description of temperature logger command is devided to for
M 2015-2
Air-cooling FPD and for Water-cooling FPD.
“10.1. How to Upgrade BBRM to Version 3.0.1”
Note is added. (for the sites BBRM isn’t installed yet)
UPS model change (BN220S to BN220T) and PCU model
change (PCU3 to PCU4)
“1.4 Network (Remote Maintenance) Model”
N 2015-2 Change diagrams.
"2. System Configuration"
Change diagrams, wiring and add LAN cable.
“5 Installing HUB”

M517- E429R
Add section.
“8.3.2 In the case of UPS made by OMRON”
Change photo.
“8.5.1 Connecting the MPC and the DAR-8000 Power
Transformer”
Change the connection destination of MPC power cable.
“8.5.2 Connecting Units to the Network”
Change wiring according to HUB added.
“9.1.2 Settings and Function of UPS control S/W(DAR-8000f
Only)”
Change settings for PCU4.
“9.1.3 Checking UPS Operation and Restarting the System
(DAR-8000f Only)”
Change checking contents and photo for PCU4.
“11.1.3 Install BBRM Version 3.0.1”
Modify UVNC setting.(checking “Allow Loopback Connections”)
Add caution to UPS operation and modify description of UVNC
settings.
“6 Network Setting”
Delete PCAnyware
“6.1. Remote Maintenance Settings”
Modify description.
P 2015-3 “9.1.3 Checking UPS Operation and Restarting the System
(DAR-8000f Only)”
Add caution to UPS operation.
“9.2.2. Installing and Setting the UltraVNC in the DAR-8000”
Modify UVNC settings.
“9.5.8. Installing and Setting the UltraVNC to the Side Station”
Modify UVNC settings.
Supported UD CONT-2002X board.
R 2015-4 “9.1.6. remolog UD Configuration” added.
“10.1.1. Operation in UD150B-40/L-40” revised.

M517- E429R
<This page is intentionally left blank.>

M517- E429R
CONTENTS
1 Introduction .........................................................................1
1.1. Tools Required for Installation .......................................................................................... 1
1.2. Checking the On-Site LAN Settings ................................................................................. 1
1.3. Overview of Installation Procedure ................................................................................... 2
1.4. Network (Remote Maintenance) Model ............................................................................ 3
1.5. Settings for Connecting with the Remote Center ............................................................. 4
2 System Configuration .........................................................8
3 Preparation for Setup .......................................................10
3.1. Add LAN Port (Only for DAR-8000i) ............................................................................... 10
4 MPC Setup ....................................................................... 11
4.1. Installing the Adapter Driver (DAR-8000f Only) ............................................................. 12
4.1.1. Installing of the Adapter Software ....................................................................... 13
4.1.2. Checking after Installation ................................................................................... 19
5 Installing HUB ...................................................................20
6 Network Setting ................................................................22
6.1. Remote Maintenance Settings ....................................................................................... 22
7 Connecting the MPC to X-ray High-Voltage Generator ....24
8 Connecting the Units ........................................................25
8.1. Connecting the Maintenance Cable to the X-ray Diagnostic Table ............................... 25
8.2. Connecting the RS-232C Cable to the X-ray High-Voltage Generator .......................... 26
8.3. Connecting the UPS and MPC (DAR-8000f Only) ......................................................... 27
8.3.1. In the case of UPS made by APC ....................................................................... 27
8.3.2. In the case of UPS made by OMRON ................................................................. 28
8.4. Connecting Room Temperature Gauges (DAR-8000f Only) ......................................... 29
8.5. Connecting Units to the MPC ......................................................................................... 33
8.5.1. Connecting the MPC and the DAR-8000 Power Transformer ............................ 35
8.5.2. Connecting Units to the Network ......................................................................... 37
9 Setting MPC and Checking Operations ............................38
9.1. Setting and Checking Devices........................................................................................ 38
9.1.1. Setting the Room Temperature Gauges and Check Its Operation (DAR-8000f
Only) 38
9.1.2. Settings and Function of UPS control S/W(DAR-8000f Only) ............................ 46
9.1.3. Checking UPS Operation and Restarting the System (DAR-8000f Only) .......... 63
9.1.4. Settings for Remote Maintenance Storage Server ............................................. 68
9.1.5. "Osiris" Installation ............................................................................................... 70
9.1.6. remolog UD Configuration ................................................................................... 70
9.1.7. BB Remote Maintenance Setting and Checking ................................................. 72

M517- E429R i
9.1.8. Checking the UltraVNC ........................................................................................73
9.2. DAR-8000 Settings ..........................................................................................................74
9.2.1. DAR-8000 Network Settings ................................................................................74
9.2.2. Installing and Setting the UltraVNC in the DAR-8000 ..........................................76
9.2.3. Installing the DAR-8000 Automatic Log Transfer Batch File ...............................83
9.2.4. Checking the Storage of DAR-8000 Data ............................................................85
9.3. PCU Settings (DAR-8000f Only) .....................................................................................86
9.3.1. PCU Settings ........................................................................................................86
9.3.2. PCU Network Settings ..........................................................................................93
9.3.3. Remote Maintenance Setting ...............................................................................95
9.4. Side Station Settings (Only Required If Side Station Is Used) .......................................96
9.4.1. Side Station Network Settings ..............................................................................96
9.4.2. Setting the Remote Desktop of the Side Station..................................................98
9.4.3. Installing the Automated log transfer batch file of Side Station ...........................99
9.4.4. Checking the Storage of Side Station Data ...................................................... 101
9.5. Side Station (Windows 7) Settings (Only Required If Side Station Is Used) ............... 102
9.5.1. Side Station (Windows 7) Network Setting ....................................................... 102
9.5.2. Setting the Remote Desktop of the Side Station............................................... 105
9.5.3. Setting the Power Options................................................................................. 107
9.5.4. Turning off the Internet time .............................................................................. 108
9.5.5. Turning off the Windows Auto Update .............................................................. 109
9.5.6. Turning off the Windows Firewall ...................................................................... 110
9.5.7. Turning off the System Restoration................................................................... 111
9.5.8. Installing and Setting the UltraVNC to the Side Station .................................... 112
9.6. Checking after Installation ............................................................................................ 118
10 Special Operation ........................................................... 119
10.1. Operation by TeraTerm .............................................................................................. 119
10.1.1. Operation in UD150B-40/L-40 ........................................................................... 119
10.1.2. Operation in SGIM ............................................................................................. 122
10.1.3. Operation in ZS-100 .......................................................................................... 124
10.2. How to Use Service Tools for BBRM-Agent............................................................... 126
10.2.1. Common Tool .................................................................................................... 126
10.2.2. BBRM Version 2.3.0 or earlier........................................................................... 128
10.2.3. BBRM Version 3.0.1 or Later ............................................................................ 135
11 BBRM Upgrade Procedure ............................................ 140
11.1. How to Upgrade BBRM to Version 3.0.1.................................................................... 140
11.1.1. Confirm Serial Number and Proxy Connection Setting ..................................... 141
11.1.2. Uninstall the already installed version ............................................................... 142

ii M517- E429R
11.1.3. Install BBRM Version 3.0.1................................................................................ 143
11.1.4. Checking Version............................................................................................... 156
11.1.5. Error Window ..................................................................................................... 157
12 Update of PCU Error Monitoring Tool (Only DAR-8000f)
158
12.1. How to Update ............................................................................................................ 158
13 Appendix.........................................................................162
13.1. How to Upgrade BBRM to Version 2.3.0 .................................................................... 162
13.1.1. How to Upgrade ................................................................................................. 162
13.1.2. How to Check BBRM Version............................................................................ 165
13.1.3. Error Window ..................................................................................................... 166

M517- E429R iii


<This page is intentionally left blank.>

iv M517- E429R
1 Introduction

This manual explains the carry-in, setup, and wiring procedures required to install the
DAR-8000 remote maintenance option. It also provides precautionary information related to
installation. In order to ensure trouble-free, efficient operation that complies with performance
specifications, read this manual carefully before installation.

This manual applies to "Rev 04.00.00" or later versions of the DAR-8000 software and
"DAR-8000 Remote Maintenance S/W Rev 03.00.04" or later.
The setup procedures in this manual may also include items previously set at the
factory prior to shipment.

1.1. Tools Required for Installation


Prepare the following tools for installation:
 Phillips screwdrivers (large, small, and short)
 CD for X-Ray High-Voltage Generator: UD, Remote Maintenance S/W(502-24708)
 Setup CD: DAR-8000, Remote Maintenance S/W (502-24707)
 Notebook PC: Runs Windows software and can connect to a network
When preparation is complete, perform installation using the procedures described in this
manual.

1.2. Checking the On-Site LAN Settings


In order to facilitate installation, obtain the following information from the administrator of the
on-site LAN.
1) DAR-8000 (PC) Network Settings
 IP address
 Subnet mask
 Default gateway
 Preferred DNS server (if required)
 Alternate DNS server (if required)
2) Settings for All Connected DICOM Devices
 IP address
 Subnet mask
 Default gateway
 Host name
 AE title
 Port number

When starting or re-starting DAR-8000, use [Shimadzu] for the login account name.

M517- E429R 1
1.3. Overview of Installation Procedure

The procedure for installing the MPC is outlined below.

1) Install the MPC and connect it to the power, and then make the additional settings.
2) Make the connections required for remote maintenance between the units.
3) Make the remote maintenance settings for the PCU (DAR-8000f only).
4) Make the remote maintenance settings for the DAR-8000.
5) Make the remote maintenance settings for the Side Station (if a Side Station is used).
6) Check operation.

In principle, the remote maintenance option is installed after installation of the RF


system is completed. However, if the remote maintenance option and a new unit are
installed at the same time, the network cables running between the examination room
and the operation room may be connected.

2 M517- E429R
1.4. Network (Remote Maintenance) Model

Explanations given in this manual are based on the network model shown below.

1) ISDN Line

Hospital
Public Shimadzu MS
Network
ISDN Dialup Router Remote
Maintenance
Center

HUB

Local network closed

SS(*2) DAR PCU(*1) MPC with RF system

(*1) Incorporated when using the DAR-8000f


On-site LAN (e.g., DICOM) (*2) SS: Side Station

2) BB Line

Hospital
Public Shimadzu MS
Network
BB Modem Remote
BB Router Maintenance
Center

HUB

Local network closed

SS(*2) DAR PCU(*1) MPC with RF system

(*1) Incorporated when using the DAR-8000f


On-site LAN (e.g., DICOM) (*2) SS: Side Station

M517- E429R 3
1.5. Settings for Connecting with the Remote Center
In order to connect with the Remote Maintenance Center, input the following settings in the
DAR-8000, MPC, Side Station and PCU.
* For items on the reference pages, refer to the corresponding pages.
For items without reference pages, refer to the following information and enter the settings.

* In the "Corresponding Unit" columns, "Done" indicates setting is completed. "Yes" indicates setting
is required, "No" indicates setting is not required.
Reference
Corresponding Unit
Page
Description Setting Method
DAR-
PCU MPC SS
8000
Click [Start] and right-click
[Computer].
Set [Allow Remote Click [Properties] to display the
Assistance [System] window.
connections to this Click [Remote settings] to
computer] to OFF display the [System Properties]
and [Allow SS(Windo
window.
1 connections from Yes No Done Yes ws7):
computers running In the [Remote] tab, clear [Allow
Page 105
any version of Remote Assistance connections
Remote Desktop to this computer] check box and
(less secure)] to select [Allow connections from
ON. computers running any version
of Remote Desktop (less
secure)] check box.
Select [Control Panel] - [User
Account]. Select "Administrator"
for the account. Select "Make
may password" and enter
"Shimadzu".
(If another password has been
previously set, "Change my
password" is displayed.)
Enter the account * Caution
name If the PC setup fails, -
2 "Administrator" and Yes No Done No
the password
"Administrator" will disappear
"Shimadzu". from the display.
If a different account is then used to
install the software, the software will
need to be re-installed when the
user returns to "Administrator"
mode. Therefore, be sure to set up
the "Administrator" account before
installing software. If this setup
method is unclear, please contact
the remote maintenance center.
Select [Start] - [Run...] and enter
"control userpasswords2".
In the [User Accounts] window,
clear the [Users must enter a
user name and password to use PCU:
Page 86
Set for automatic this computer.]check box and DAR-
3 Yes Yes Done No
login at startup. click [Apply]. 8000: -
In the [Automatically Log On]
window, enter "Administrator"
into [User name], "Shimadzu"
into [Password] and [Confirm
Password].

4 M517- E429R
Reference
Corresponding Unit
Page
Description Setting Method
DAR-
PCU MPC SS
8000
Select [Start] - [Control Panel]. In
the [All Control Panel Items]
Set [Never] as window, click [Power Options].
[Turn off the In the [Power Options] window,
SS(Windo
display] and [Put click [High performance] and
4 Yes Yes Done Yes ws7):
the computer to then click [Change plan settings].
Page 107
sleep] in [Power In the [Edit Plan Settings]
Options]. window, set [Never] as [Turn off
the display] and [Put the
computer to sleep].
Click [Start] - [Control Panel]. In
the [All Control Panel Items]
window, click [Date and Time].
In the [Internet Time] tab of the SS(Windo
Set the Internet
5
clock to OFF.
[Date and Time] window, click Yes No Done Yes ws7):
[Change settings... ]. In the Page 108
[Internet Time settings] window,
clear the [Synchronize with an
Internet time server] check box.
PCU
Click [Start] - [Control
Panel]-[Security Center].
In the [Windows Security Center]
window, click [Automatic
Updates]. In the [Automatic
Updates] window, click [Turn off
Automatic Updates.]. PCU:
Page 78
Set the Windows
6 Automatic Update SS Yes No Done Yes SS(Windo
to OFF. Select [Start] - [Control Panel]. In ws7):
the [All Control Panel Items] Page 108
window, click [Windows
Update]. In the [Windows
Update] window, click [Change
settings]. In the [Change
settings] window, select [Never
check for updates] as [Important
updates].
PCU
Click [Start] - [Control
Panel]-[Security Center]. In the
[Windows Security Center]
window, click [Windows Firewall].
Click the [General] tab and
select [Off (not recommended)]. PCU:

Page 78
Set the Windows SS
7 Yes No Done Yes
Firewall to OFF. Select [Start] - [Control Panel]. SS(Windo
In the [All Control Panel Items] ws7):
window, click [Windows Firewall]. Page 110
In the [Windows Firewall]
window, click [Change
notification settings]. In the
[Customize Settings] window,
click [Turn off Windows
Firewall].

M517- E429R 5
Reference
Corresponding Unit
Page
Description Setting Method
DAR-
PCU MPC SS
8000
Click [Start] and right-click
[Computer]. Then click
[Properties] to display the
[System] window.
Click [System protection] to
SS(Windo
Disable the system display [System Properties]
8 Yes No Done Yes ws7):
recovery feature window.
Page 111
Click [OS (C:)] as [Protection
settings] and then [Configure...].
In the [System Protection for OS
(C:)], click [Turn off system
protection].
Click [Control Panel]-[System].
Click [Advanced] tab and click
Set the display of [Settings] of [Startup and
9 the list of operating Recovery]. No No Done No -
systems to OFF. Turn off [Time to display list of
operating system:] of [System
startup].
DAR-8000
Double-click [My Network
Places] on the desktop, open the
"My Network Places" window,
and click [Network and Dial-up
Connections]. In the [Network
and Dial-up Connections]
window, double-click [Local Area
Connection] to display the [Local
Area Connection Status]
window.
Click [Properties] to display
[Local Area Connection
Properties] window. DAR-
Click [Internet Protocol (TCP/IP)] 8000:
and [Properties] to display
Make [Default [Internet Protocol (TCP/IP) Page 74
Gateway] empty. Properties] window.
(Except when using PCU:
10
Click [Use the following IP Yes Yes Done Yes Page 93
an on-site LAN's
local area
address] and make [Default
connection) gateway] empty. SS(Windo
ws7):
PCU Page 102
Click [Start] - [Control Panel] and
double-click [Network
Connections]. Double-click
[Local Area Connection].
- When connected to the
network: The [Local Area
Connection Status] window is
displayed. Click the [General] tab
and [Properties]. The [Local Area
Connection Properties] window
is displayed.
- When not connected to the
network: The [Local Area
Connection Properties] window

6 M517- E429R
Reference
Corresponding Unit
Page
Description Setting Method
DAR-
PCU MPC SS
8000
is displayed.
Click the [General] tab and
[Internet Protocol (TCP/IP)] -
[Properties] to display the
[Internet Protocol (TCP/IP)
Properties] window. Click [Use
the following IP address], and
make [Default gateway] empty.

SS
Select [Network address setting]
of the [Service Tools] menu .
Double-click [Local Area
Connection] in the [Network
Connections] window. In the
[Local Area Connection Status]
window, select [Properties]. In
the [Networking] tab of the [Local
Area Connection Properties]
window, click the [Internet
Protocol Version 4 (TCP/IPv4)]
and [Properties]. In the [Internet
Protocol Version 4 (TCP/IPv4)
Properties], click [Use the
following IP address] of and
make [Default gateway] empty.
At the American
Power Conversion
Corp. PowerChute Refer to 9.1.2. "Settings and
BusinessEdition
11 Function of UPS control No No Yes No Page 46
login, enter
"service" and S/W(DAR-8000f Only)".
"Shimadzu" for the
password.
Set the room
temperature
thermometer's Refer to step 8 in section 9.1.1.
measurement time
"Setting the Room Temperature
12 period to 30 No No Yes No Page 38
minutes. The Gauges and Check Its Operation
thermometer's time (DAR-8000f Only)".
is set in synch with
the PC.

M517- E429R 7
2 System Configuration

With reference to Figure 2-1 “Remote Maintenance Connections” and Table 2-1 "Remote
Maintenance Cables and Connection Destinations", connect the MPC, the router, the hub, and
all the other units.
Refer to 7 "Connecting the MPC to X-ray High-Voltage Generator", 8 "Connecting the Units"
and 8.5.1. "Connecting the MPC and the DAR-8000 Power Transformer" for details.

*1 Prepare a Router, HUB, or an ISDN and BB line at the site. (They are indicated by dotted line in
the figure below.) (Four LAN ports of the ISDN/BB router is available as the HUB.)
*2 This gauge is required for the Water-cooling FPD.

*2
(4) (5)
(7) *1
(16)
(6) (13)
COM4

(16)
(4) (5) USB2
USB1 ON board
(10)
COM3 (17) ISDN/BB Router ISDN
COM2 BB modem BB
appropriate device
COM1 GND or others

(8)

(9)

DAR-8000
Control cabinet

Figure 2-1 Remote Maintenance Connections

8 M517- E429R
Table 2-1 Remote Maintenance Cables and Connection Destinations

Cable Cable DAR- DAR-


Part number Part name Connected from Connected to
number length 8000f 8000i

MPC,
(1) 088-81104-92 CABLE, LD-CT/BU30 30 m HUB  
ONboard LAN port
DAR-8000,
(2) 088-81104-87 CABLE, LD-CTT/BU10/RS 10 m HUB  
ONboard LAN port
Room
CONNECTOR,
(3) 088-50674-01 CABLE, AOM6S-20 20 m temperature 
MODULAR PMJ-6
gauge
CONNECTOR, MODULAR
(4) 070-08127-01 
PMJ-6

(5) 088-50892-73 CABLE, BSUSRC0610BS ○

MPC, Expansion
(6) 511-15196-10 CABLE A, 09-2A-10M-2D 10 m 
serial port 4

(7) 037-38785-18 SPACER, SBA-305(E) ○

CABLE, MPC, ONboard UD150B-40


(8) 070-02160-19 5.0 m  
RS232C-09FEMALE-5M serial port 1 control cabinet
UD150B-40
CABLE, MPC, ONboard
(9) 070-02160-19 5.0 m control cabinet,  
RS232C-09FEMALE-5M serial port 2
SGIM board
X-ray
diagnostic
MPC, Expansion table
(10) 502-24345-01 Maintenance cable 15 m  
serial port 3 Control
Cabinet
Power
(11) 502-24966-20 AC power cable 20 m MPC  
Transformer
Ground cable,
(12) 502-24426-03 20 m MPC  
#10
UPS
INTERFACE, AP9824LJ
(13) (made by APC ) 
074-81311-33 The attached cable(Maker
(*1) UPS 
P/N: BUC26)
(made by OMRON )
Side Station,
(14) 088-81104-87 CABLE, LD-CTT/BU10/RS 10 m HUB [] []
ONboard LAN port

PCU,
(15) 088-81104-87 CABLE, LD-CTT/BU10/RS 10 m HUB 
ONboard LAN port

Cable attached to Room


(16) - 
Temperature Gauge

ISDN/BB
(17) 088-81104-87 LD-CTT/BU10/RS 10 m HUB
Router

[ ]: Option
*1 This cable is used by alternative by UPS to be used.
The case of UPS made by APC is indicated on the uppwer row, and the case of UPS made
by OMRON is indicated in the lower row.

M517- E429R 9
3 Preparation for Setup

3.1. Add LAN Port (Only for DAR-8000i)

There are two type of DAR-8000i; one type has one LAN port and another type has two LAN
ports. The number of LAN port of DAR-8000i depends on the shipment time.

If there is only one LAN port, install an expansion LAN card to PC inside the DAR-8000 cabinet.

Refer to Appendix F "Installing the LAN Card" in "M517-E380 DAR-8000i Installation Manual
(Rev.B or later)" to check the number of LAN port and install it when the LAN port is only one.

10 M517- E429R
4 MPC Setup

In order to set up the MPC alone, use the attached power cable.

The power requirement of the MPC is AC 100V. Make sure that the power source is
AC 100V before connection. If you connect the MPC to other than AC 100V, the MPC
will be damaged.

The attached power cable is not used after the completion of this setup. Disconnect
and keep the cable.

Check that the main switch on the back of the MPC is tuned OFF (0), and connect the MPC to a
commercial power source in the hospital (AC100V).
A display, keyboard, and mouse are not included in the standard equipment.
Connect those of the DAR-8000 to the MPC temporarily to perform settings described below.
This section descries the setup of the adapter driver.

M517- E429R 11
4.1. Installing the Adapter Driver (DAR-8000f Only)

Install additional software after connecting the adapter. Install the driver and software using the
procedures described below. If the USB port to which the adapter is connected is changed after
installation of the driver and software, the software must be installed again. Use the USB port 1 or
2 indicated in the figure below.

USB Port 1

USB Port 2

USB port of PC SMALL

12 M517- E429R
4.1.1. Installing of the Adapter Software

After starting MPC, when "BSUSRC0610BS" ((5) in Figure 2-1;088-50892-73) is connected to


MPC, the window to install additional software appears automatically.
Use the procedure described below to connect "BSUSRC0610BS" (088-50892-73) to USB
ports 1 and 2 to perform installation.

In cases where the window does not appear, and the COM number is already displayed,
correlation with the driver has been completed normally. Therefore this section is not needed.
Go to the section 5 "Installing HUB

Install the HUB (088-80110-01 HUB, ES1008MTP) in the PSU rack as shown below:
The HUB is magnetic-backed.

M517- E429R 13
14 M517- E429R
The HUB’s power cable is connected to UPS as follows.

M517- E429R 15
Network Setting".

Check the following items, in cases where the window of Step 1 in the next page is
not displayed.
1. Check whether the adapter is attached to MPC.
2. Click [Start]-[Control panel]-[System]-[hardware] Tab. Click [Device
manager]-[Ports (COM&LPT)] to display the list of the COM ports.

As shown in the following figure, in cases where "USB-serial Controller" is


selected, click and highlight "USB-serial Controller". Then right-click "USB-serial
Controller" and select [Update Driver…]. The window of Step 1 will be displayed.

16 M517- E429R
- Connecting "BSUSRC0610BS" (088-50892-73) to USB Port 1

1 Connect "B" connector of provided USB


cable to "BSUSRC0610BS"
(088-50892-73) .
2 Turn on MPC. After Windows starts,
connect "A" connector of USB cable to
USB port 1 of MPC. The window on the
left is displayed.
3 Select [No, not this time] and click
[Next].

4 Select [Install the software automatically


[Recommended]] and click [Next].

5 When the window on the left is


displayed, click [Continue Anyway].
The installation starts. The following
window is displayed when the
installation is completed.

6 Click [Finish]. The window closes.

M517- E429R 17
- Connecting "BSUSRC0610BS" (088-50892-73) to USB Port 2 (for Water Cooling FPD)

1 Connect "B" connector of provided USB


cable to "BSUSRC0610BS"
(088-50892-73).
2 Turn on MPC. After Windows starts,
connect "A" connector of USB cable to
USB port 2 of MPC. The window on the
left is displayed.
3 Select [No, not this time] and click
[Next].

4 Select [Install the software automatically


[Recommended]] and click [Next].

5 When the window on the left is


displayed, click [Continue Anyway].
The installation starts. The following
window is displayed when the
installation is completed.

6 Click [Finish]. The window closes.

18 M517- E429R
4.1.2. Checking after Installation

Check that the driver and software for the adapter have been installed properly.
1 Click [Start].
2 Right-click [My Computer] on the
Start menu, and click [Properties].
The [System Properties] window is
displayed.
3 Click the [Hardware] tab and click
[Device Manager]. The window on
the left is displayed.
4 Double-click [Ports (COM & LPT)]
and check that the item shown
below is displayed.
"USB Serial Port (COM5)"
"USB Serial Port (COM6)"
*The Air-cooling FPD has one
temperature-and-humidity logger, so
only one COM number is recognized.
*The COM number differs depending
on the environment.
5 After checking the display, close
the [Device Manager] window.
6 Shutdown MPC.

M517- E429R 19
5 Installing HUB

Install the HUB (088-80110-01 HUB, ES1008MTP) in the PSU rack as shown below:
The HUB is magnetic-backed.

20 M517- E429R
The HUB’s power cable is connected to UPS as follows.

To HUB

M517- E429R 21
6 Network Setting

Set the network of all units.

6.1. Remote Maintenance Settings


Make the remote maintenance settings as shown in "Table 6-1 Remote Maintenance Settings".
Standard network setting of the MPC has been completed as shown in the table below.
Except when there is need to change an IP address, the setting is not required.

Refer to section 9.2.1. "DAR-8000 Network Settings", 9.3.2. "PCU Network Settings", 9.4.1.
"Side Station Network Settings", 9.5.1. "Side Station (Windows 7) Network Setting" and the
Remote Maintenance Center manual for the router.
Table 6-1 Remote Maintenance Settings
Item Router MPC*1 PCU*2 DAR-8000 Side Station*3

IP address 192.168.100.1 192.168.100.10 192.168.100.11 192.168.100.13 192.168.100.14

Subnet mask 255.255.255.0 255.255.255.0 255.255.255.0 255.255.255.0 255.255.255.0

ISDN - - - - -
Default
gateway (192.168.100.1)
BB
*4

ISDN - - - - -
Priority
DNS (192.168.100.1)
BB
*4

Account Administrator Administrator Shimadzu Administrator


*5
Password Shimadzu Shimadzu Dar8k Shimadzu

Windows XP or
Windows 2000 or
Windows XP Windows 7
OS - Windows XP Windows XP
English version English version
English version
*6

Connection UltraVNC or UltraVNC or UltraVNC or


- Remotedesktop
method Remotedesktop Remotedesktop Remotedesktop

*1: MPC settings have been completed based on the settings in the table as standard.
*2: DAR-8000f only.
*3: Option
*4: It is need to be set to connect to the BB.
*5: Refer to the Remote Maintenance Center manual.
*6: It is Windows 7 English version when the Side Station is Revision. 03.00.00 or later.

22 M517- E429R
■ Connecting Two or More DAR-Units to the Same Network
Specify the last digits of the IP addresses as shown in the table below. Settings other than IP
addresses are the same as those shown in the table above. Even if two or more DAR-8000
units are installed, no more than one router is used in the same network. If the number of ports
on the router is insufficient, add hubs as appropriate and connect all units to the network.
Item Router MPC PCU*1 DAR-8000 Side Station*2
IP address:
192.168.100.1 192.168.100.10 192.168.100.11 192.168.100.13 192.168.100.14
first unit

IP address:
- 192.168.100.20 192.168.100.21 192.168.100.23 192.168.100.24
second unit
IP address:
- 192.168.100.30 192.168.100.31 192.168.100.33 192.168.100.34
third unit

Set the fourth and subsequent units in a similar way.


*1: DAR-8000f only.
*2: Option

■When the same segment has been used in the existing on-site network.

When the same segment (192.168.*.*) has been used in the existing on-site network, the said
IP address may not be available for the remote maintenance. In this case, ask the administrator
of the on-site network to assign a proper IP address.

■IP address setting for the software displaying the temperature warning. (Only for DAR-8000f)

IP address is set to the above described address as a default setting. In case that more than
two DAR-8000 are connected in the same network or the IP address is changed to the specified
address, please change the IP address setting in “C:\PCUErrSend\PCUErrorSend.ini “ on MPC.

"PcuErrorSend.ini"
[Network]
IP=192.168.100.13  Change the IP address.
PORT=3000

Don’t change except for IP address.

M517- E429R 23
7 Connecting the MPC to X-ray High-Voltage Generator
Top cover
Eyebolt 1 Remove the two eyebolts and then
remove the top cover from the X-ray
high-voltage generator's control
cabinet.

2 Remove the four legs of the MPC.


(They are attached with two sided tape.)
*After removal, the legs are not used. Store
them separately.

Legs

3 Attach the "Fuzzy loops" (502-25867)


on the bottom surface of MPC.
*Refer to the left figure for their size.
4 Attach the "Spikey hooks" (502-25866)
on the "PC mounting board"
Hook and loop (502-23936).
fastener
*Refer to the left figure for their size.
5 Place the MPC on the "PC mounting
board" (502-23936).
Check the MPC has been firmly secured
to the "PC mounting board" (502-23936)
by tilting it.
Then secure it using the eyebolts
Fuzzy loops
(P/N: 502-25867) removed in step 1 to the top of the
control cabinet.

Spikey hooks
(P/N: 502-25866)

PC mounting board
(P/N: 502-23936)

6 Insert provided cables into 4 serial


ports and firmly secure them with the
screws.

24 M517- E429R
8 Connecting the Units

Connect the MPC and each unit by following section 2 "System Configuration" and 8.5.
"Connecting Units to the MPC".

8.1. Connecting the Maintenance Cable to the X-ray Diagnostic Table


1 With reference to the photo on the left,
connect "Maintenance cable" ((10) ;
502-24345-01) to connector J4 on NEX
SH board in X-ray diagnostic table's
control cabinet.
*Refer to 0 "

Maintenance cable

M517- E429R 25
8.2. Connecting the RS-232C Cable to the X-ray High-Voltage Generator

Connect the RS-232C cable to the X-ray high-voltage generator's control cabinet. Refer to 8.5.
"Connecting Units to the MPC" for details on connecting to the MPC.

1 With reference to the photo on the left,


connect "CABLE, RS232C-09FEMALE-5M"
((8) ;070-02160-19) to connector J15 on
the UDCONT-2002 board in the X-ray
high-voltage generator's control cabinet.
Connector

Board inside control


cabinet

RS-232C cable

UDCONT-2002

2 Connect "CABLE, RS232C-09FEMALE-5M"


((9) ;070-02160-19) to connector CN11 on
the SGIM board in the X-ray high-voltage
generator's control cabinet.
RS-232C cable *Refer to 2 “System Configuration” for
details on connecting to the MPC.

SGIM board

26 M517- E429R
8.3. Connecting the UPS and MPC (DAR-8000f Only)
8.3.1. In the case of UPS made by APC
Attach an expansion board (made by APC) to the UPS and connect the MPC connection cable.

Spacers Interface cable 1 Attach two "SPACER, SBA-305(E)"


((7) ;037-38785-18) to the connector at
the male end of "CABLE A,
09-2A-10M-2D" ((6) ;511-15196-10).
2 Connect "INTERFACE, AP9824LJ"
Cable A, 09-2A-10M-2D, male end
((13) ;074-81311-33) to the male end of
"CABLE A, 09-2A-10M-2D"
((6) ;511-15196-10).

3 Connect "INTERFACE, AP9824LJ"


((13) ;074-81311-33) to the UPS serial
port.

Interface cable

M517- E429R 27
8.3.2. In the case of UPS made by OMRON
Attach an expansion board (made by OMRON) to the UPS and connect the MPC connection cable.
Spacers The attached cable of UPS 1 Attach two "SPACER, SBA-305(E)"
(Maker P/N: BUC26)
((7) ;037-38785-18) to the connector at
the male end of "CABLE A,
09-2A-10M-2D" ((6) ;511-15196-10).
Cable A, 09-2A-10M-2D, male end
2 Connect " The attached cable of
UPS"(Maker’s P/N:BUC26) to the male
end of "CABLE A, 09-2A-10M-2D"
((6) ;511-15196-10).

3 Connect " The attached cable of


UPS"(Maker’s P/N:BUC26) to the UPS
serial port.
The attached cable of UPS
(Maker P/N: BUC26)

28 M517- E429R
8.4. Connecting Room Temperature Gauges (DAR-8000f Only)

Set up TH-101 room temperature gauges (made by Microtechno Corporation) and set them to
record data every 30 minutes.
■ Names and Functions of Parts

1 Power Switch
This is the TH-101 room temperature gauge's power switch. The knob of the switch is inside a
window on the side of the case. Slide it with the tip of a ballpoint pen or some other appropriate
implement.

2 Record Switch
Used as the execution switch for each mode.

3 Interval Switch
Used to change parameters for each mode.

4 Mode Switch
Used to switch between TH-101 modes.
5 COM Connector
Used to connect to the PC. Connect it with the cable provided and set times, measurement
conditions or retrieve data as necessary.

6 Alarm Output
If any two parameters out of the temperature, the humidity, or the dew point become equal to
previously set upper or lower limits, a contact turns ON, and an alarm is output.
7 LCD
Indicates the temperature and humidity. It also indicates the status of data recording, displays
messages indicating that, for example, the battery is being replaced or that measurement is in
progress, and indicates alarms (upper/lower limit).

8 Sensor
Incorporates temperature and humidity sensors.

M517- E429R 29
■ Connection and Setup

1 Connect the RS-232C (D-Sub9 pin male) connector CABLE,


of " CABLE, BSUSRC0610BS " ((5) ; BSUSRC0610BS
088-50892-73) and Cable attached to Room
Temperature Gauge ((16)).
2 Connect the USB side of "CABLE, CABLE,
BSUSRC0610BS " ((5) ; 088-50892-73) to the
USB port of MPC.

3 Insert batteries.
a. Remove the battery cover on the back of the
gauge by sliding it in the direction indicated by
the arrow.
b. Insert AAA alkaline batteries as shown on the
right. Be sure to insert the batteries with the
correct orientation. The orientation is the same
for both batteries.
c. Attach the battery cover.

4 Turn on the MPC to start Windows.


5 Connect the gauge's COM connector to the MPC
using the communications cable provided with the
gauge. If the cable is too short, connect with
"CABLE, AOM6S-20" ((3) ;088-50674-01) and "
CONNECTOR, MODULAR PMJ-6" ((4) ;
070-08127-01).
Near FPD Unit: COM5
1
Near Water Cooler Unit: COM6*

1
* COM6 is connected only for the Water-cooling
FPD.

6 Slide the knob of the gauge's power switch with the


tip of a ballpoint pen or some other appropriate
implement to turn on the power to the gauge.

30 M517- E429R
7 Place this gauge in an appropriate location.
 Set one unit around the FPD.
 Set the another unit around the Water Cooler unit
for the Water Cooling FPD.
 Secure the gauge with Velcro attached to the
position arrowed as shown on the right.
 When installing a room temperature gauge near
the FPD, be sure it is installed at approximately
the same height.
 Run the cable so that it does not interfere with
examinations.

The service life of the batteries is approximately 3 months. When the batteries near the end
of their service life, a battery symbol is displayed in the gauge's LCD. Replace both AAA
alkaline batteries with reference to the procedure described in "Connection and Setup".

The Air-cooling FPD has 1 gauge and the Water-cooling FPD has 2 gauges.

M517- E429R 31
■ Installation Site(For around the FPD)

The data logger installed for around FPD is installed near the pole of ZS-100 as shown below to
measure the temperature around the ZS-100.

Pole of ZS-100

■Installation Site(For around the Cooling Unit)

The another data logger installed for around the cooling unit is installed on the cooling unit
inside the sub cabinet as shown below to measure the temperature around the cooling unit.

32 M517- E429R
8.5. Connecting Units to the MPC

Connect the serial port and USB port of MPC to each device. The connectors correspond to
each port as follows:

MPC Connected to
COM1 UD CONT-2002 board in X-ray high-voltage generator’s control cabinet
COM2 SGIM board in X-ray high-voltage generator’s control cabinet
COM3 NEX SH board in X-ray diagnostic table’s control cabinet
COM4 UPS
USB Port 1 (COM5) Room temperature gauge for vicinity of FPD Unit
USB Port 2 (COM6) Room temperature gauge for vicinity of Water Cooler Unit

*Step 1 and 2 are for DAR-8000f


unit only.
1 Connect USB Port 1 of MPC
To Room temperature gauge (Near Water Cooler and Room temperature
unit) (Only for Water-cooling type FPD)
gauge near FPD unit.
To Room temperature gauge 2 Connect USB Port 2 of MPC
(Near FPD unit)
and Room temperature
gauge near Water Cooler
Unit.
*Refer to 8.4. "Connecting
Room Temperature Gauges
(DAR-8000f Only)" for
details.

M517- E429R 33
To SGIM board in X-ray high-voltage To X-ray diagnostic table’s 3 Connect UD CONT-2002
generator’s control cabinet control cabinet board in X-ray high-voltage
UPS generator's control cabinet
To UD CONT-2002 board for X-ray
high-voltage generator’s control cabinet and onboard COM1 port of
MPC with "CABLE,
RS232C-09FEMALE-5M"
((8) ;070-02160-19).
*Refer to 8.2. "Connecting the
RS-232C Cable to the X-ray
High-Voltage Generator" for
details.
4 Connect SGIM board in X-ray
high-voltage generator’s
control cabinet and onboard
COM2 port of MPC with
"CABLE,
RS232C-09FEMALE-5M"
((9) ;070-02160-19).
*Refer to 8.2. "Connecting the
RS-232C Cable to the X-ray
High-Voltage Generator" for
details.
5 Connect NEX SH board in
X-ray diagnostic table’s
control cabinet and onboard
COM3 port in MPC with
"Maintenance cable
"((10) ;502-24345-01). *Refer
to 8.1. "Connecting the
Maintenance Cable to the
X-ray Diagnostic Table" for
details.
6 Connect UPS and onboard
COM4 port of MPC with
"CABLE A, 09-4A-10M-4D"
((6) ;070-02273-08).
*Refer to 8.3. "Connecting the
UPS and MPC (DAR-8000f
Only)" for details.

34 M517- E429R
8.5.1. Connecting the MPC and the DAR-8000 Power Transformer

This section describes how to connect MPC's AC power cable. Follow the procedure below
according to the UPS to be used.

- UPS by APC

With reference to the procedure described below, connect MPC's "AC power cable"
((11) ;502-24966-20) and "Ground cable, #10" ((12) ;502-24426-03) to the DAR-8000 power
transformer. The AC power cable's three terminals are respectively marked with the characters
"0V", "100V", and "G". Be sure to connect the cable correctly.

1. Be sure to check that the breaker of the power switchboard of the unit is
turned off to prevent electric shock.
2. Be sure to check that the breaker of each unit is turned off.
When it is tuned ON, be sure to turn it off before starting connection.

Check that the main switch on the back of the MPC is turned OFF (0). If it is turned ON,
be sure to turn it off before starting connection.

1 Connect the terminals of the


MPC's "AC power cable"
((11) ;502-24966-20) as follows:
Ground bar  0V→L OUT1 terminal
 100V→L OUT2 terminal
 G→E terminal
2 Connect MPC's "Ground cable,
(1) #10" ((12) ;502-24426-03) to the
(2) ground bar inside the power
transformer.

DAR-8000 power transformer

M517- E429R 35
3 Use any screw from the back of
MPC, and secure "Ground cable,
#10" ((12) ;502-24426-03) to
MPC.

1. After the power transformer ASSY installation, if the breaker of the power
switchboard of unit is turned off, the MPC does not work.
2. When the breaker of the power switchboard is turned off unavoidably, turn on
the MPC again to restart it.
Be sure to inform a customer about this note.

- UPS by OMRON

Connect MPC's power cable (511-15134 CORD SET, KP300CLS16FS) to the power socket on
the UPS as shown below:

MPC Power Socket

36 M517- E429R
8.5.2. Connecting Units to the Network

When the connections between the units are completed, connect the Hub and the LAN ports of
the units and then connect the HUB and ISDN/BB router. Refer to the connection diagram in
section 2 "System Configuration".

The LAN ports of the units and the HUB are shown below:

MPC

② ③ ④
(for Maintenance) (for Maintenance)

LAN LAN
(Expansion Board) (Expansion Board)
(for DICOM) (for DICOM)


④ ③ ② ④ To router

M517- E429R 37
9 Setting MPC and Checking Operations
Set devices and check the operation of them that connect to the MPC.

9.1. Setting and Checking Devices

9.1.1. Setting the Room Temperature Gauges and Check Its Operation (DAR-8000f
Only)

Set the room temperature gauge near FPD unit, and if Water Cooler Unit is used, also set the
gauge near the unit.
1 Turn on the gauge's power switch.
2 Turn on MPC. The Windows starts.

3 Click [Start] – [All Programs] –


[TH-101]-[TH-101 Manager].
The application starts. [TH-101
Manager] window is displayed.

38 M517- E429R
4 Select [Communication(C)] on the main
menu and then select [Serial port
setting(M)…]. The [Communication port]
window is displayed.

5 Set the communication port as follows


and click [OK].
 Port: COM5
A confirmation window is displayed.
Click [OK].
*Set the Port Numbers to COM5 for
Room temperature gauge near FPD unit,
and COM6 for Room temperature gauge
near Water Cooler unit.

The baud rate, parity, and stop bit


settings are fixed at the following values
and cannot be changed.
 Baud rate: 9600
 Parity: Odd
 Stop: 1

6 Select [Communication(C)] on the main


menu and then select [Setup(O)…]. The
[Various setups] window is displayed.
*When the writing stops,
[Setup(O)…]can be available.

M517- E429R 39
7 Click the [Time] tab and input the
correct date and time. Select the [Adjust
to the PC time] check box. Click
[Setting].

8 Click the [Measuring condition] tab,


make the settings shown below, and
click [Setting].
 Recording mode: Overwrite (LOOP)
 Unit of Temperature: Centigrade
 Recording interval: 30 min
 Instrument number: 1
*Set a unique Instrument number for
each instrument.

9 Click the [Upper/lower limit] tab, make


the settings shown below, and click
[Setting].
 Upper limit temp: No
 Lower limit temp: No
 Upper limit hum: No
 Lower limit hum: No

40 M517- E429R
10 Click the [Dew point alarm] tab, make
the setting shown below, and click
[Setting].
 Upper limit: No
11 When all the settings are completed,
click [Close]. The window closes.

M517- E429R 41
■ Checking Operation of the Room Temperature Gauges and Starting Recording
After checking the operation of the room temperature gauges, start recording. Perform this for
each instrument. With reference to "Setting the Room Temperature Gauges", select the COM
ports to which the gauges are connected.
1 Select [Communication] on the main
menu and then select [Realtime
indication]. The temperature and
humidity are indicated as graphs.
*The display interval setting is fixed at 2 sec.
*When the graphs reach the right-hand side,
they automatically shift to the left.
*Observe the temperature graph for a few
minutes to check that the correct
temperature is displayed, and then click
[Stop]. A dialog box is displayed to confirm
that the real-time display data is to be

saved. Click [No].

2 Start recording.
Click [Communication(C)] on the main
menu and then click [Start
recording(S)].
Click [Yes] in [Do you want to
Proceed ?].
Click [OK] in [A measurement condition
was set.].
The gauges start recording immediately.

42 M517- E429R
■ Installing the Automatic Storage Tool of Room Temperature Logs

Note that the command differs between the Air-cooling FPD and the Water-cooling FPD.

Air-cooling FPD Water-cooling FPD

Installation location Around FPD Around FPD Around cooling unit

Log collection frequency Once a day Once a week Once a week

Logger name (menu name) No1(TempLog-1) No1(TempLog-1) No2(TempLog-2)

Start log Start Recording Start Recording Start Recording


collection
Command
Register Start Scheduling_Daily Start Scheduling_Weekly Start Scheduling
schedule

- Air-cooling FPD

1. Click the [Start] button, and point to [All Programs].


Open the [TempLog-1] menu to select the [Start Recording]. Recording the room temperature
starts.

2. Click the [Start] button, and point to [All Programs].


Open the [TempLog-1] menu to select the [Start Scheduling Daily]. Storing the room temperature
logs is registered into the scheduler.
(It is executed early on every morning.)

- Water-cooling FPD

1. Click the [Start] button, and point to [All Programs].


Open the [TempLog-1] menu to select the [Start Recording]. Recording the room temperature
starts.

2. Click the [Start] button, and point to [All Programs].


Open the [TempLog-1] menu to select the [Start Scheduling Weekly]. Storing the room temperature
logs is registered into the scheduler.
(It is executed early on every Sunday.)

3. Click the [Start] button, and point to [All Programs].


Open the [TempLog-2] menu to select the [Start Recording]. Recording the room temperature
starts.

4. Click the [Start] button, and point to [All Programs].


Open the [TempLog-2] menu to select the [Start Scheduling]. Storing the room temperature logs is
registered into the scheduler.
(It is executed early on every Sunday.)

M517- E429R 43
■Operation Confirmation of the Automatic Storage Tool of the Room Temperature Logs

Click the [Start] button, and point to [All Programs]. Open the [TempLog-1] menu to select the
[Loading the Data]". Check the following items:
 The log is created under the "C: \Temp_rlog\No1\logs" folder.
 Then, after a while, this log is transmitted to the server of the RMC, and is
moved to the "sent_log" folder.

The procedure below is performed for the Water-cooling FPD.

Click the [Start] button, and point to [All Programs]. Open the [TempLog-2] menu to select the
"Loading the Data". Check the following items:
 The log is created under the "C: \Temp_rlog\No2\logs" folder.
 Then, after a while, this log is transmitted to the server of the RMC, and is moved to the
"sent_log" folder.

44 M517- E429R
■Confirming the Task Schedule
Click the [Start] button, and point to [All Programs]. Open the [Scheduled Tasks] folder and
check that there are the following tasks.

In case of Air-cooling FPD

In case of Water -cooling FPD

M517- E429R 45
9.1.2. Settings and Function of UPS control S/W(DAR-8000f Only)

This section describes how to configure the UPS control S/W and check its function. Please
check the desktop icon in order to confirm which type of UPS control S/W is installed .

Desktop icon of PowerChute Business Editon(APC inc.)

Desktop icon of PowerActPro Master Agent(OMRON inc.)

・ Settings and Function of PowerChute Business Edition windows (In the case of UPS made by
APC)
Function of Power Chute Business Edition windows
The PowerChute Business Edition windows are explained here. Double-click the [APC
PowerChute Business Edition] icon on the MPC desktop to start APC PowerChute Business
Edition.

● Logging on to PowerChute Console

1 If the login window is displayed when


PowerChute is started up, enter the
following and click [Connect].
User Name: service
Password: Shimadzu
Server: localhost

* Password: "S" of Shimadzu is a capital


letter and enter the Password with "Caps
Lock" released.
*If the [Save User Name and Password]
check box is selected, this window is
subsequently not displayed.

2 Check that the UPS is connected


normally. Double-click [localhost]
under [Hostname]. The status is
displayed in lower of the window.

*It may take about one minute to


communicate with UPS, retrying
communication.

46 M517- E429R
● Device Properties
Check the device settings and change them if necessary. In the APCPowerChute Business
Edition Basic, on "view" menu, click [Device Properties]. The [Device Properties] list is
displayed. Click an item on the left to display the settings. Select the [Show advanced items]
check box at the bottom left of the window to display the [Diagnostics] and [Communications]
items.

[Summary]
No setting is required.

[Admin Information]
No setting is required.

M517- E429R 47
[UPS Identification]
No setting is required.

[UPS Status]
The UPS status is displayed. No setting is
required.

[Battery Status]
The battery status is displayed. Check that
the [Battery Status] is "Normal".

48 M517- E429R
[Power Parameters]
Make the following settings and click
[Apply].
 High-Transfer Voltage: 110
 Low-Transfer Voltage: 090
 Sensitivity: Medium
 Nominal Output Voltage: 100
 Audible Warning: On Battery

[Shutdown Sequence]
Click [Configure Shutdown Sequence] to
display the setting window. No setting is
required.

Setting window

M517- E429R 49
[Shutdown Schedule]
No setting is required.

[Power Failure]
- Select [After UPS has been on battery
for]
- Set [6] minutes

After selecting, click [Apply].

[Shutdown Type]
Select [Shutdown] and click [Apply].

50 M517- E429R
[Self-Test]
The self-test results are displayed. No
setting is required.

[UPS Calibration]
No setting is required.

[Data Log]
The recorded data is displayed. Nothing is
displayed if the UPS is not operating.

M517- E429R 51
[Event Log]
The recorded events are displayed.
Nothing is displayed if the UPS is not
operating.

[Log Options]
Make the following settings and click
[Apply].
Max. Log Size: 1000
Delete Entries After: 3 Months
Data Logging Interval: 10 minutes

[UPS Communications]
The communication port settings are
displayed. No setting is required.

52 M517- E429R
[SNMP Communications]
No setting is required.

When the settings and changes are completed for all the windows, click [Close].

M517- E429R 53
● Power Event Summary

Select [Power Event Summary] and click


the refresh button in the top right of the
window. The power event types are
displayed together with numbers of
occurrences.

Power Event Summary

54 M517- E429R
● Voltage Analysis

Select [Voltage Analysis] and click the


refresh button in the top right of the
window. The UPS input line voltage is
displayed.

APC UPS Input Line Voltage

M517- E429R 55
● Status
The operating status of the host and the failure status are displayed. Click the [Status] icon on
the left of the window.
The status of the host is displayed.

[Power Failure Alert]


If the UPS power supply stops for some
reason, a power failure message is
displayed. If power is restored, the system
continues operation.

[System Shutdown]
When the time set in the device properties
([Shutdown] – [Power Failure]) is reached,
system shutdown starts. In this state, the
shutdown procedure continues even if
power is restored. The shutdown signal is
output continuously for approximately
2 minutes.

56 M517- E429R
・ Settings and Function of Power windows (In the case of UPS made by OMRON)

Please perform this section After checking that the units connected to UPS is turned off.
When equipment is not turned OFF, the electric supply from UPS stops during a setup,
and there is a danger that equipment will break down.

PowerPro Master Agent windows are explained here.

● Logging on to PowerPro Master Agent(PowerAct Pro Monitor)

1 Double Click[ConfigShutdownParameter]
icon on the MPC desktop.

Right screen is displayed.


Set following settings.
●Shutdown Parameter
・Select UPS Outlet : Outlet A (No Control)
・AC Fail Delay Time : 240 sec
・Shutdown Start Delay Time : 240 sec
・External Command Line : (blank)
・Run External Command Executed
Time : 60 sec
・Shutdown Need Time : 60 sec
・Select OS Shutdown Mode :
Shutdown
・UPS Auto Stop : Enable
●Other Device
・Outlet A(Control): 420 Sec
・Outlet B(Control): 420 Sec
・Outlet C(Control): 420 Sec
●Other Device
・Maximum backup time for AC fail: 0 Min

M517- E429R 57
3 Click [Boot/Reboot] Tab.
Set following settings.
●Output Start Time of UPS Boot Up
・Send Outlet Delay Command : Enable
・Outlet A(Control) : 0 Sec
・Outlet B(Control) : 0 Sec
・Outlet C(Control) : 0 Sec
●UPS Reboot Setting
・UPS Start Delay Time : 0 Sec
・Battery Capacity : 0%
●UPS Cold Start
・UPS Cold Start Function : Disable

4 Click [Boot/Reboot] Tab.


No settings.
Click [OK] button.

5 Double Click [PowerActProConsole] icon.

58 M517- E429R
6 Right Screen of Console is displayed.
① [NetSearch Window Menu]
Enter "2" and Press Enter key. ①
② [Input ID and Password Menu]
Enter "3" and Press Enter key.
Enter "127.0.0.1" and Press Enter key.
③ [User Login Window]

Enter following value.
Login ID: Admin
Press Enter key.
Enter following value.
Password: Shimadzu ③
Press Enter key.

7 Right screen is displayed.

Check following settings.


UPS Model:BN220T
UPS Status:On Line
Battery Status:Normal

After checking, Enter "Q".

8 Right screen is displayed.


Begin settings of UPS from step 9.

M517- E429R 59
9 This step is settings of UPS Battery
Change Date.
・Screen of step 8
Enter "2" and Press Enter key.
・[Reconfirm The “UPS” Setting] Screeen 9-1
Enter "4" and Press Enter key.
・[Reconfirm The “UPS” Battery Change
Date]
[Reconfirm The “UPS” Battery Change

Date] is displayed.(Screen 9-1)。
1 Start Date of UPS
2 New battery replace date
3 Last battery replace date(Read
Only)

4 Next battery replace date(Plan)

The setting value of 1,2 and 3 is UPS


installation date. ③
The setting value of 4 is the date after two Screen 9-2
year from UPS installation day.
(Screen9-2 is the scrren in the case of
setting step of [4 Next battery replace
date(Plan)])

Set the following settings on 1-4.


①Enter the number of settings(1-4).
Press Enter key.
②Enter “yyyy-mm-dd”.(yyyy:year,
mm:month, dd:day)
Press Enter key.
③[Reconfirm The “UPS” Battery Change
Date] is displayed again.
Check that the setting of ② is set.

The following steps are repeated and are


performed until the screen of step 8 is
displayed.
Enter "0" and Press Enter key.

60 M517- E429R
10 This step is settings of EventLog Option.
・Screen of step8 (Screen 10-1)
①Enter "1" and Press Enter key.
・[“System” menu of PowerAct Pro
Monitor(MPC10)] (Screen 10-1) ①
②Enter "1" and Press Enter key.

・[“Configuration” menu of PowerAct Pro


Monitor(MPC10)](Screen 10-2) ②
①Enter "3" and Press Enter key.

Screen 10-1
・[“Log Option” menu of PowerAct Pro
Monitor(MPC10)] (Screen10-2)
②Enter "1" and Press Enter key.

・[“Event Log Option” menu of PowerAct
Pro Monitor(MPC10)] (Screen10-2)
Set the settings in the [2. The maximum
Event Log Record Time(Week):].

Set the following settings.
③Enter "2" and Press Enter key.
④Enter "12" and Press Enter key.

⑤[“Event Log Option” menu of PowerAct
Pro Monitor(MPC10)] is displayed again. ④
Check that the value in the [“Event Log
Option” menu of PowerAct Pro
Monitor(MPC10)] is [2. The maximum
Event Log Record Time(Week):12]. ⑤
Screen 10-2

M517- E429R 61
11 The following steps are repeated and are
performed until the screen of [“Log Option” ①
menu of PowerAct Pro Monitor(MPC10)] in
step 10 is displayed.
Enter "0" and Press Enter key.
・[“Log Option” menu of PowerAct Pro

Monitor(MPC10)]
①Enter "2" and Press Enter key. ③
・[“Data Log Option” menu of PowerAct Pro
Monitor(MPC10)]
The following menu is displayed.
1. Maximum Data Log Record Time(Week)
2.The record interval time <On line><Sec> ④
3.The record interval time <On

Battery><Sec>

Set [1. Maximum Data Log Record


Time(Week)].
Set the following settings.
②Enter "1" and Press Enter key.
③Enter "12" and Press Enter key. ⑥


Set [2.The record interval time <On
line><Sec>].
Set the following settings.
④Enter "2" and Press Enter key. ⑧
⑤Enter "600" and Press Enter key.

Set [3.The record interval time <On


Battery><Sec>].
Set the following settings.
⑥Enter "3" and Press Enter key.
⑦Enter "600" and Press Enter key.
⑧[“Data Log Option” menu of PowerAct
Pro Monitor(MPC10)] is displayed again.
Check the settings of step 11.

The following steps are repeated and are


performed until the window of
PowerActProConsole is closed.
Enter "0" and Press Enter key.

62 M517- E429R
9.1.3. Checking UPS Operation and Restarting the System (DAR-8000f Only)

When the connection between the MPC and the UPS control S/W and configuration check of
the UPS are completed, check the UPS operation using the either procedure described below:

After checking the UPS operation, return to the original status within 10 minutes irrespective of
whether the UPS operation is normal or not. If not, the system may be damaged.

■ Checking UPS Operation


The procedure differs depending on the type of UPS (made by APC or made by OMRON).

・In the case of UPS made by APC

1) Turn ON the power to the system (i.e., X-ray high-voltage generator and DAR-8000).
2) From the PCU Xcat "power control", check that the HV and DC power supplies are both
ON.
3) With the "power control" window displayed, turn OFF the power supply on the side of the
PSU rack. (See photo below.) (The power supply from "200 / 220 / 230 / 240 V" * to
100 VAC is turned OFF.)
4) Check that the HV and DC power are turned OFF. The UPS operates and an audible
warning is generated.
*The power voltage is different according to region.

・In the case of UPS made by OMRON

1) Turn ON the power to the system (i.e., X-ray high-voltage generator and DAR-8000).
2) From the PCU Xcat "power control", check that the HV and DC power supplies are both
ON.
3) With the "power control" window displayed, turn OFF the power supply on the side of the
PSU rack. (See photo below.) (The power supply from "200 / 220 / 230 / 240 V" * to
100 VAC is turned OFF.) The UPS operates and an audible warning is generated.
4) Check that the PCU and MPC automatically shuts down and the HV and DC power are
turned OFF 10 minutes after turning off the power supply.
5) Check that the UPS automatically shuts down 12 minutes after turning off the power
supply.

* The time in above steps varies slightly.

M517- E429R 63
The trans may make oscillation noise in case UPS has no loads, but it isn’t problem.
Its noise ceases if loads are applied. (Any loads are applied under normal use.)

電源

Plugs

Air-cooling FPD PSU RACK (PSU RACK4)


Plugs

64 M517- E429R
Plugs
Power supply Power supply

Plugs
Water-cooling FPD PSU RACK (PSU RACK2)

UPS model:SU2200J UPS model:SUA2200JB

Basic Monitoring Ports


To PCU
Basic Monitoring Ports (34-d)+(34-e)
To PCU
(34-d)+(34-e) To PCU
(34-a) To HV MAIN
OUTPUT C OUTPUT
100V 50/60Hz To PS CTRL
A Board B A 100V 50/60Hz (30-b)

(33-a) (33-b)
To DC MAIN (30-b)
D
B D (34-a)
To HV MAIN (33-a) To PCU
C
Overcurrent Protection Overcurrent Protection
Reset button To DC MAIN Reset button
(33-b)
INPUT INPUT
To PS CTRL 120V 50/60Hz 120V 50/60Hz
Board To PSU TRANS To PSU TRANS
(33-c) (33-c)

UPS Rear Panel(In the case of UPS made by APC)

M517- E429R 65
To PS Ctrl To PCU
To PCU (33-b) (34-a)
(34-d)+(34-e)

To HUB

To DC Main(48)

To PSU Trans
To MPC
(33-c)

UPS Rear Panel(In the case of UPS made by OMRON)

In 10 minutes after checking the operation, restore the state of power to what it was
before the check regardless of whether the operation is correct or wrong.
If you do not so, the system may be damaged.

UPS Audible Warnings


If the power supply is interrupted, the UPS operates. At first, four beeps are sounded
once every 30 seconds. Then, if the battery comes closer to expiry, the audible warning
sounds continuously. If the power supply is restored, the UPS starts charging operation,
and the audible warning stops.

66 M517- E429R
■ Restarting the System
After checking UPS operation, restart the system using the following procedure.

・In the case of UPS made by APC


1) Turn on the power to the PSU rack and supply power.
2) The shutdown signal from the UPS is maintained. Wait at least 2 minutes, and then turn on
the power to the MPC. Check that PowerChute starts.
3) Connect the plugs for the DC and HV power supplies to the UPS.
4) Start the PCU and check that Xcat starts. Because the DC power supply is turned OFF,
approximately 3 minutes is required for startup. The temperature monitor indicates a
temperature of 94.4 C.
5) Check that the PCU does not shut down for at least 5 minutes.
6) Turn on the DC power supply from the Xcat "power control".
7) Select "Synchronize with the host system" check box that enables ON/OFF in
synchronization with the DAR-8000 and close the power control dialog box.
8) Turn ON the power to the system (DAR-8000) and check that the HV power supply turns
ON. Because the DC power supply was OFF, it takes time for the system to stabilize.
Perform calibration after at least 30 minutes has elapsed.
9) When the system (DAR-8000) starts, check that it is operating normally. (Check that
images are displayed with both fluoroscopy and radiography.)

・In the case of UPS made by OMRON


1) Turn on the power to the PSU rack and supply power.
2) Check that the MPC and PCU start.
3) Reffering the PCU, check that Xcat starts. Because the DC power supply and the
DAR-8000 are OFF, approximately 3 minutes is required for startup. The temperature
monitor indicates a temperature of 94.4 C.
4) Check that the PCU does not shut down for at least 5 minutes.
5) Turn on the DC power supply from the Xcat "power control".
6) Select "Synchronize with the host system" check box that enables ON/OFF in
synchronization with the DAR-8000 and close the power control dialog box.
7) Turn ON the power to the system (DAR-8000) and check that the HV power supply turns
ON. Because the DC power supply was OFF, it takes time for the system to stabilize.
Perform calibration after at least 30 minutes has elapsed.
8) When the system (DAR-8000) starts, check that it is operating normally. (Check that
images are displayed with both fluoroscopy and radiography.)

M517- E429R 67
After installation, contact the Remote Maintenance Center to request connection and
operation confirmation.

9.1.4. Settings for Remote Maintenance Storage Server


Make the settings that allow problematic images to be viewed with the remote maintenance
option. If a storage server function is incorporated in the MPC and the MPC server is registered
in the DAR-8000, problematic images transferred by the user to the MPC can be viewed with
the remote maintenance option.
■ Installation in the MPC
Use the following procedure to install [Simple_Storage] in the MPC and allow the MPC to
function as a DICOM image storage server.
(1) Connect a keyboard, a mouse, and monitor to the MPC to make it operable. If the MPC
cannot be operated directly because, for example, it is installed on the X-ray high-voltage
generator, make it operable via a remote desktop connection from the PCU.

Connecting with Remote Desktop


A Switch to the PCU screen using the PC Selector.
B From the [Start] menu, select [All Programs] – [Accessories] -
[Communications] - [Remote Desktop Connection].
C Enter "192.168.100.10" (*1) as the MPC's IP address and connect to the
MPC.
D Enter "Administrator" as the account and "Shimadzu" as the password.
E After performing the required operations quit the remote desktop
connection.
(*1) This IP address is a standard. Specify the IP address assigned to the MPC at the installation site.

After the following network settings are entered in a notebook PC, remote desktop
connection from this PC is possible.
IP Address: 192.168.100.200
Subnet mask: 255.255.255.0
Default gateway: ___. ___. ___ . ___ (Empty spaces, no numbers)

68 M517- E429R
(2) From the Start menu, select [All Programs] – [Startup] – [Simple_Storage] to execute the
program.

*With the MPC factory setting, when the power is turned on, [Simple_Storage] starts
automatically. The task of [Simple_Storage] appears in the task bar.

(3) Check that [Simple_Storage] appears in the task bar.


[Simple_Storage]

M517- E429R 69
9.1.5. "Osiris" Installation

Send images of DAR-8000 to MPC and display it on the screen of MPC in order to capture a
problematic image on MPC.
(1) Send images of DAR-8000 to the MPC. (Select the Simple_MPC to send images.)
(2) Check that there are files that are sent from DAR-8000 in the C:\Image\XRF folder of the MPC.
(3) Start the Osiris4 and check that the image file in the C:\Image\XRF folder is displayed.

9.1.6. remolog UD Configuration


The main control board of High Voltage Generator is changed after shipping of April 2015.
Before: UD CONT-2002 (P/N: 502-22500 / 562-22500)
After: UD CONT-2002X (P/N: 565-11900)
The configuration of serial port communication varies depending on the board type.

(1) Check board type in generator’s cabinet.

UD CONT-2002 UD CONT-2002X

70 M517- E429R
(2) Run configuration tool in ‘C:\B40_rlog\ChangeUdRemolog’ folder.
For UD CONT-2002X: Run ‘ChangeUdRemolog_2002X.exe’.
For UD CONT-2002: Run ’ChangeUdRemolog_2002.exe’.

(3) The serial port configuration is changed and following screen appeares. Press [OK] to close.

(4) Click [X] to close TeraTerm window named ‘UD Terminal – COM1 VT’.
(5) Restart ‘C:\B40_rlog\remolog.exe’ from Windows Explorer.
(6) Click [UD Terminal – COM1 VT] on the task bar to show TeraTerm window.
(7) Press [Enter] key and check the prompt "UD>" is displayed.

M517- E429R 71
9.1.7. BB Remote Maintenance Setting and Checking

This setting is required only for the BB remote maintenance. Do not set it for the
ISDN remote maintenance.

1 Click the [Start] button, point to [All


Program] -> [Start_Agent] to execute it.

2 The icon appears in the task bar.

3 The execution is completed for about 10


seconds. Click the [Start] -> [Turn off
Computer] -> [Restart] to restart the
computer.

72 M517- E429R
9.1.8. Checking the UltraVNC

Check that the VNC icon appears after starting MPC.

VNC icon

When all installations are finished, request connection confirmation to the remote maintenance
center.

M517- E429R 73
9.2. DAR-8000 Settings
9.2.1. DAR-8000 Network Settings

1 Turn on the DAR-8000's control cabinet. The


Windows starts.
2 Log on to the DAR-8000 by entering "shimadzu"
as the username and "Dar8k" as the password.

3 Double-click [My Network Places] on the


desktop. The [My Network Places] window
opens.
4 Click [Network and Dial-up Connections] on the
left of the window. The [Network and Dial-up
Connections] window opens.

(DAR-8000f)
5 Double-click [Local Area Connection]. The [Local
Area Connection Status] window opens.

(DAR-8000i)
5 Double-click [Local Area Connection2]. The
[Local Area Connection2 Status] window opens.

(DAR-8000f)
6 Click [Properties]. The [Local Area Connection
Properties] window opens.

(DAR-8000i)
6 Click [Properties]. The [Local Area Connection2
Properties] window opens.

74 M517- E429R
7 Select [Internet Protocol (TCP/IP)] and click
[Properties]. The [Internet Protocol (TCP/IP)
Properties] window opens.

8 Select [Use the following IP address] and make


the following settings.
 IP address: 192.168.100.13
 Subnet mask: 255.255.255.0
Do not enter anything in [Default gateway].
9 After entering the settings, click [OK] and close
all windows.
10 Restart the system.

M517- E429R 75
9.2.2. Installing and Setting the UltraVNC in the DAR-8000

Install the UltraVNC in the DAR-8000 and make the required settings.

■ Preparation

1 In the My Documents folder of the PCU


(MPC), create a new file named "VNC".

2 Set the setup CD "DAR-8000, Remote


Maintenance S/W (502-24707)" in
DAR-8000 and open the "D8_VNC"
folder in setup CD using Windows
Explorer. Copy the
UltraVNC_1.0.9.6.2_Setup.exe to the
"VNC" folder created.

76 M517- E429R
■ Install UltraVNC

Double-click the UltraVNC_1.0.9.6.2_Setup.exe and follow the procedure below.

1 Click [Next>].

2 Click [I accept the agreement] and


[Next>].

3 Click [Next>].

M517- E429R 77
4 Click [Next>].

5 Check that the [Full Installation] is


selected and Click [Next>].

6 Click [Next>].

78 M517- E429R
7 Clear the[Download Vista addons files
now]check box and click [Next>].

8 Clear the[Download the mirror driver]


check box and click [Next>].

9 Select the check box below and click


[Next>].

-"Register UltraVNC Server as a system


service"
-"Start or restart UltraVNC service"
-"Create UltraVNC desktop icons"
-"Associate UltraVNC Viewer with the .vnc
file extension"

10 Click [Install].

M517- E429R 79
11 The installation is started and the
window below appears. Wait for a
while.

12 The window shows information after


installation. Click [Next>].

13 Click [Finish] and restart the PC.

14 Right-click the VNC icon. The menu is


displayed.

80 M517- E429R
15 Click [Admin Properties]. The
[UltraVNC Server Property Page]
window is displayed.

16 Input "Shimadzu" to [VNC Password:].

Select the [Allow Loopback


Connections].

Check that the [User impersonation


(for Service only)] check box is
cleared, and click [OK].

M517- E429R 81
■Confirmation
(1) The VNC icon is displayed on the Task bar.

(2) After the set-up is completed, the "UltraVNC Server" and "UltraVNC Viewer" icons are
displayed on the desktop.

(3) Drag and drop the "UltraVNC Server" and "UltraVNC Viewer" icons into the "VNC" folder in
the My Documents folder.

(4) When each terminal installation is completed and checked, access to the remote
maintenance center to check the connection.

82 M517- E429R
9.2.3. Installing the DAR-8000 Automatic Log Transfer Batch File

At startup, the automatic log transfer batch file (D8K_logcopy.bat) automatically connects the
MPC and the DAR-8000, and creates a folder named using the current day's date/time in the
MPC's C:\Diag\DAR8k folder. It copies the entire contents of the DAR-8000's C:\Diag folder to
this folder. This makes it easy, in subsequent maintenance, to refer to logs for days on which
problems occur.

(1) Close all applications and return to the Windows display. (If the auto log-off and auto
power-off settings are enabled, display the system settings window from the [Start] dialog
box and disable these settings. Refer to Appendix B "Various Basic Settings" in DAR-8000f
Installation Manual / M517-E383.)
(2) Set the setup CD "DAR-8000, Remote Maintenance S/W (502-24707)" in the DAR-8000
and open the [D8_LogCopy] folder in setup CD using Windows Explorer.
(3) Double-click [InstLogcopy.bat] to run the installation program. The following window is
displayed.

(4) Type the MPC's IP address (usually "192.168.100.10") and press the ENTER key.
[LogCopy has installed!] is displayed in the window.

M517- E429R 83
(5) When [LogCopy has installed!] is displayed, press any key. Installation is completed.
(6) Press the  key (corresponds to the  (Windows) key) to display the task bar. Select
[Programs] – [Startup] on the Start menu and check that [D8K_logcopy.bat] is displayed on
the [Startup] menu.
(7) After restarting the system and DAR-8000 starts, check that the log file is copied to the
MPC with reference to the description [Checking the Copying of DAR-8000 Log Files].

The log file is not copied over if connection to the MPC is not possible via the network.

■ Checking the Copying of DAR-8000 Log Files


Turn on the MPC.
After checking that the MPC startup operation is completed, turn on the power to the
DAR-8000.
After checking that the DAR-8000 startup operation is completed, check that a folder named
using the current day's date/time is created in the MPC under the sent_DAR8k folder, as shown
below, and that the DAR-8000 log file has been copied over.

84 M517- E429R
9.2.4. Checking the Storage of DAR-8000 Data

Check that data is sent from the DAR-8000 to the MPC's storage server.
■ Registration at the DAR-8000
With reference to "Settings for DICOM Transmission and DICOM CD Writing" in the DAR-8000i
Installation Manual (M517-2135 (Japanese)/M517-E380 (English)) or the DAR-8000f Installation
Manual (M517-2139 (Japanese)/M517-E383 (English)), register [SimpleStorage], which is
installed in the MPC, at the DAR-8000 using the information given in the following table.

Setting name MaintenancePC


IP address 192.168.100.10 (*1)
AE title simple_storage
Port number 4000
(*1) This IP address is a standard. Specify the IP address assigned to the MPC at the installation site.

■ Transmission from the DAR-8000


With reference to "Exporting Files" in the DAR-8000i Operation Manual (M517-1113 (Japanese)
/M517-E059 (English)) or the DAR-8000f Operation Manual (M517-1114 (Japanese)
/M517-E060 (English)), send image data from the DAR-8000 to the MPC.
Check that the image file sent from the DAR-8000 is contained in the MPC's C:\Image folder.

M517- E429R 85
9.3. PCU Settings (DAR-8000f Only)
9.3.1. PCU Settings

Make the settings for enabling remote maintenance of the PCU.

■ Password Setting for the Remote Desktop

1 Turn on the power to the PCU. The


Windows starts.
2 Click [Run] on the Start menu. The
next window is displayed.

3 In the [Open] box, type "control


userpasswords" and click [OK]. The
[User Accounts] window is displayed.

4 Click [Administrator]. The next window


is displayed.

86 M517- E429R
5 Click [Change my password]. The next
window is displayed.

6 Enter a new password and click


[Change Password].

7 Click [X] on the upper right corner of


the [User Accounts] window to close
the window.

M517- E429R 87
8 Click [Run] on the Start menu. The
next window is displayed.

9 In the [Open] box, type "control


userpasswords2" and click [OK]. The
[User Accounts] window is displayed.

10 Click the [User] tab and clear the


[Users must enter a user name and
password to use this computer.] check
box.
11 Click [Apply]. The [Automatically Log
On] window is displayed.

12 Make the following settings and click


[OK].
 User name: Administrator
 Password: Shimadzu
 Confirm Password: Shimadzu
Returns to the [User Accounts] window.
*When entering "Shimadzu", only use upper
case for the "S". Disable [Caps Lock].

88 M517- E429R
■ Extension Display Settings

1 Right-click [Start] on the task bar.


2 Click [Explore] on the pop-up
menu. Windows Explorer is
displayed.

3 On the main menu, select [Tools]


– [Folder Options]. The [Folder
Options] window is displayed.

4 Click the [View] tab and clear the


[Hide extensions for known file
types] check box under
[Advanced settings].
5 Click [Apply] and then click [OK].
6 Quit Windows and turn OFF the
power to the PCU.
7 Restart the PCU and check that
Xcat starts without any request
for a password to be entered.

M517- E429R 89
■ Stopping the Display of Security Center Alerts

1 Open the [Control Panel] window


and double-click [Security
Center]. The [Windows Security
Center] window opens.

2 Check that [Firewall] and


[Automatic Updates] are both set
to [OFF]. If they are set to [ON],
set them to [OFF]. (Refer to the
next page.)
3 Click [Change the way Security
Center alerts me] under
[Resources]. The [Alert Settings]
window opens.

4 Clear all the check boxes under


[Alert Settings] and click [OK].
5 Close the [Windows Security
Center] window.

90 M517- E429R
■ Disabling the Security Center Settings
If [Firewall] and [Automatic Updates] in the Windows Security Center settings are set to [ON],
use the procedure described below to set them to [OFF].
1 Click [Automatic Updates] in the
[Windows Security Center]
window. The [Automatic Updates]
window is displayed.

2 Select [Turn off Automatic


Updates.] and click [OK].

3 Click [Windows Firewall] in the


[Windows Security Center]
window. The [Windows Firewall]
window is displayed.

M517- E429R 91
4 Click the [General] tab and select
[Off (not recommended)].
5 Click [OK].

92 M517- E429R
9.3.2. PCU Network Settings
1 Turn on the power to the PCU. The
Windows starts.
2 Open the [Control Panel] window and
double-click [Network Connections].
The [Network Connections] window is
displayed.

3 Double-click [Local Area Connection].


 When connected to the network:
The [Local Area Connection Status]
window is displayed. Click the
[General] tab and then click
[Properties]. The [Local Area
Connection Properties] window is
displayed.
 When not connected to the
network:
The [Local Area Connection Properties]
window is displayed.

[Local Area Connection Status] window

M517- E429R 93
4 Click the [General] tab and select
[Internet Protocol (TCP/IP)].
5 Click [Properties]. The [Internet Protocol
(TCP/IP) Properties] window is
displayed.

6 Select [Use the following IP address],


make the following settings, and click
[OK].
 IP address: 192.168.100.11
 Subnet mask: 255.255.255.0

*Do not enter anything in [Default gateway].


7 After entering the settings, click [OK]
and close all windows.
8 Restart the system.

94 M517- E429R
9.3.3. Remote Maintenance Setting
1 Turn on the PCU.
The Windows starts.
2 Click the [Start] button, point to [All
Program] -> [Shimadzu PCU Tools], and
then click [Remote Maintenance
Config]. "Remote Maintenance Config
Tool" is displayed.
3 Input as follows:
・ PCU Unit No: 1
・ Host Name: MPC address
Example 192.168.100.10
・ User Name: Administrator
・ Password: Shimadzu
4 Click the "Check" button to check the
connection with MPC. The dialog of
"Check completed [OK]." is displayed.
Click [OK] to close the dialog.

5 Start the Xcat.


6 On the "Tool" menu, click "Option" to
open the option setting window. Click
the arrow button to display [MPC] tab.
7 Select "Enable remote maintenance"
check box and "HV power off by host
system synchronization".
* When HV power is kept ON due to
uninterrupted power supply system,
contact the remote maintenance center
to select "After calibration".
8 Select the check boxes of "Send
recovery from alert" and "Transfer all
correction data on alert".
9 Click [OK] to close the window.
10 Restart the Xcat.
11 Turn off the system and wait until the
HV power of the FPD is turned off
(usually 30 minutes). When the HV
power is turned off, a log is transmitted
to MPC.
12 Check that three kinds of logs
(ApplicationLog_X.txt, TempLog_X.log,
DefectMap_X_Y.cab) are transmitted to
the C:\Diag\PCU\Log folder of the PC.

M517- E429R 95
9.4. Side Station Settings (Only Required If Side Station Is Used)

When the OS of the Side Station is Windows 7, refer to section 9.5. "Side Station
(Windows 7) Settings (Only Required If Side Station Is Used)".

9.4.1. Side Station Network Settings


1 Turn on the power to the Side Station. The
Windows starts.

After the windows starts, when the display


changes to that shown on the right, press the
DELETE key twice while holding down the
CTRL and ALT keys. The [Log On to Windows]
window is displayed. Make the following
settings and click [OK].

User name: Shimadzu


Password: service
*When entering "Shimadzu", only use upper case for
the "S". Disable [Caps Lock].

The Windows desktop is displayed.

2 Select [All Programs] on the Start menu. Then


select [Service Tools] – [Network address
setting].

96 M517- E429R
3 Double-click [Local Area Connection].

4 Click the [General] tab and select [Internet


Protocol (TCP/IP)].
5 Click [Properties]. The [Internet Protocol
(TCP/IP) Properties] window is displayed.

6 Select [Use the following IP address], make


the following settings, and click [OK].
 IP address: 192.168.100.14
 Subnet mask: 255.255.255.0
*Do not enter anything in [Default gateway].

7 After entering the settings, click [OK] and


close all windows.
8 Restart the system.

M517- E429R 97
9.4.2. Setting the Remote Desktop of the Side Station

For the connection establishment, the computer accessed from remote locations must
open port 3389 to the outside. If the firewall or broadband router of the company
exclusive line blocks this port, or if the firewall function of Windows XP or other
commercially available firewall software does not allow data to pass through the port
3389, user cannot access the computer from remote locations using the Remote
Desktop function. In those cases, ask the network administrator to open the port 3389
or change the firewall settings to open it.

1 Right-click [My Computer] on the Start


menu, and click [Properties]. The
[System Properties] window is
displayed.

*Allow user to connect remotely to


this computer using Remote Desktop
function.

2 Click the [Remote] tab.

3 Select the [Allow users to connect


remotely to this computer] check box.

4 Click [OK] to apply the settings and


close the [System Properties] window.

98 M517- E429R
9.4.3. Installing the Automated log transfer batch file of Side Station

Automated log transfer batch file(PPWS_logcopy.bat)automatically connects the MPC and the
Side Station on startup and create a folder named using the current day's date/time in
C:\Diag\ PPWS folder in the MPC. To this folder, all contents of D:\Shimadzu\Log folder in the
Side Station is copied, making it easier on maintenance later to find out the log of the day when
an error occurs.

(1) Log on the Side Station as user "shimadzu". Refer to section 4.4.1 "Logging On" in
"M517-E078 Side Station Installation Manual".
(2) Insert the setup CD "DAR-8000 Remote Maintenance S/W" (502-24707)into the CD drive
of Side Station. Open the "PPWS_LogCopy" folder inside this CD using Windows Explore.
(3) Double click [InstLogcopyPPWS.bat] to execute it.
The following window is displayed.

(4) Enter the IP address of MPC(Normally"192.168.100.10").


[LogCopy has installed!] is displayed on the screen.

M517- E429R 99
(5) Press any key after [LogCopy has installed!] is displayed.
The install is done.
(6) After restarting the system, confirm the log file is copied to MPC on the startup of the Side
Station, referring the following "Checking the Copying of Side Station Log Files".
1)

The log file is not copied over if connection to the MPC is not possible via the network.

■ Checking the Copying of Side Station Log Files

(1) Turn on the MPC.


(2) After checking that the MPC startup operation is completed, turn on the power to the Side
Station.
(3) After checking that the Side Station starts up, check that a folder named using the current
day's date/time is created in the MPC under the sent_PPWS folder, as shown below, and
that the Side Station log file has been copied over.

100 M517- E429R


9.4.4. Checking the Storage of Side Station Data

Check that data is sent from the Side Station to the MPC's storage server.
■ Registration at the Side Station
With reference to "Settings for DICOM Transmission" in "M517-E078 Side Station Installation
Manual", register [SimpleStorage], which is installed in the MPC, at the Side Station using the
information given in the following table.

Setting name MaintenancePC


IP address 192.168.100.10 (*1)
AE title simple_storage
Port number 4000

(*1) This IP address is a standard. Specify the IP address assigned to the MPC at the installation site.

■ Transmission from the Side Station


With reference to "DICOM Transmission" in "M517-E451 Side Station Operation Manual", send
image data from the Side Station to the MPC.
Check that the image file sent from the Side Station is contained in the MPC's C:\Image folder.

M517- E429R 101


9.5. Side Station (Windows 7) Settings (Only Required If Side Station Is Used)

Perform confirmation of 8.4.3 and 8.4.4 after settings of 8.5.1 to 8.5.8.

When the OS of the Side Station is Windows XP, refer to section 9.4. Side Station
Settings (Only Required If Side Station Is Used)".

Settings of 8.5.3, 8.5.4, 8.5.5, and 8.5.7 were completed at the time of shipment of the
Side Station. Perform these setting, only when a change is made.

9.5.1. Side Station (Windows 7) Network Setting


1 Turn on the power to the Side Station.
The Windows starts.

Input the User name and Password


below to log in.

User name: shimadzu


Password: service
*Disable [Caps Lock].

The Windows desktop is displayed.

2 Select [Service Tools] - [Network


address setting]. The [Network
Connections] window is displayed.

102 M517- E429R


3 Double-click [Local Area Connection].
The [Local Area Connection Status]
window is displayed.

4 Click [Properties].
The [Local Area Connection
Properties] window is displayed.

5 Click the [Networking] ] tab and select


[Internet Protocol Version 4
(TCP/IPv4)].
6 Click [Properties].
The [Internet Protocol Version 4
(TCP/IPv4) Properties] window is
displayed.

M517- E429R 103


7 Click [Use the following IP address],
make the following settings, and click
[OK].
- IP address:192.168.100.14
- Subnet mask:255.255.255.0

*Do not enter anything in [Default gateway].

8 After entering the settings, click [OK] to


close all windows.
9 Restart the system.

104 M517- E429R


9.5.2. Setting the Remote Desktop of the Side Station

For the connection establishment, the computer accessed from remote locations must
open port 3389 to the outside. If the firewall or broadband router of the company
exclusive line blocks this port, or if the firewall function of Windows XP or other
commercially available firewall software does not allow data to pass through the port
3389, user cannot access the computer from remote locations using the Remote
Desktop function. In those cases, ask the network administrator to open the port 3389
or change the firewall settings to open it.

1. Right-click [Computer] on the Start


menu, and click [Properties].
The [System] window is displayed.

*Allow user to connect remotely to


this computer using Remote Desktop
function.

2. Click [Remote settings].


The [System Properties] window is
displayed.

M517- E429R 105


3. Click the [Remote] tab of the [System
Properties] window.

4. Clear the [Allow Remote Assistance


connections to this computer] check
box of [Remote Assistance].

5. Click [Allow connections from


computers running any version of
Remote Desktop (less secure)] of
[Remote Desktop].

6. Click [OK] to apply the settings and


close the [System Properties] window.

106 M517- E429R


9.5.3. Setting the Power Options
1. Click [Control Panel] on the Start
menu.
The [All Control Panel Items] window
is displayed.
* When [Control Panel] is displayed,
set [Small icons] as [View by].

2. Click [Power Options].

3. Click [High performance].

4. Click [Change plan settings].

5. Select [Never] as [Turn off the


display].

6. Select [Never] as [Put the computer to


sleep].

7. Click [Save changes].

M517- E429R 107


9.5.4. Turning off the Internet time
1. Click [Control Panel] on the Start
menu.
The [All Control Panel Items] window
is displayed.
* When [Control Panel] is displayed,
set [Small icons] as [View by].

2. Select [Date and Time].


The [Date and Time] window is
displayed.

3. Click the [Internet Time] tab and


[Change settings...].

4. Clear the [Synchronize with an


Internet time server] check box.

5. Click [OK].

6. Click [OK] to apply the setting and


close the [Date and Time] window.

108 M517- E429R


9.5.5. Turning off the Windows Auto Update
1. Click [Control Panel] on the Start
menu.
The [All Control Panel Items] window
is displayed.
* When [Control Panel] is displayed,
set [Small icons] as [View by].

2. Click [Windows Update].


The [Windows Update] window is
displayed.

3. Click [Change settings].

4. Select [Never check for updates] as


[Important updates].

5. Click [OK].

M517- E429R 109


9.5.6. Turning off the Windows Firewall
1. Click [Control Panel] on the Start
menu.
The [All Control Panel Items] window
is displayed.
* When [Control Panel] is displayed,
set [Small icons] as [View by].

2. Click [Windows Firewall].


The [Windows Firewall] window is
displayed.

3. Click [Change notification settings].

4. Select [Turn off Windows Firewall] as


[Home or work (private) network
location settings].

5. Click [OK].

110 M517- E429R


9.5.7. Turning off the System Restoration
1. Right-click [Computer] on the Start
menu, and click [Properties].
The [System] window is displayed.

2. Click [System protection].


The [System properties] window is
displayed.

3. Click [OS (C:)] of [Protection settings].


4. Click [Configure...].
The [System Protection for OS (C:)]
window is displayed.

5. Click [Turn off system protection].

6. Click [OK].
The confirmation dialog is displayed.

7. Clicks [Yes]. The confirmation dialog


is closed.

8. Click [OK].

M517- E429R 111


9.5.8. Installing and Setting the UltraVNC to the Side Station
■ Preparation

1 Create the "VNC" folder in the My


documents.

2 Set the setup CD "DAR-8000, Remote


Maintenance S/W (502-24707)" to the
Side Station, and open the
"PPWS_VNC" folder in the setup CD in
the Windows Explorer. Copy the
UltraVNC_1.0.9.6.2_Setup.exe in this
folder to the created "VNC" folder.

■ Installing the UltraVNC

Double-click the UltraVNC_1.0.9.6.2_Setup.exe and follow the procedure below.

1 Click [Next].

112 M517- E429R


2 Click "I accept the agreement" and
[Next].

3 Click [Next].

4 Click [Next].

M517- E429R 113


5 Check that [Full Installation] is selected
and click [Next].

6 Click [Next].

7 Clear the [Download Vista addons files


now] check box and click [Next].

114 M517- E429R


8 Clear the [Download the mirror driver]
check box and click [Next].

9 Select all check boxes and click [Next].


Start the installation.

10 Click [OK].
*The right window is the error
displayed when not downloading the
mirror driver.

11 Click [Next].

M517- E429R 115


12 Click [Finish].

13 Click [Show hidden icons] of the task


bar, and right-click the VNC icon.
The Pop-up menu is displayed.

14 Click [Admin Properties]. The


[UltraVNC Server Property Page]
window is displayed.

116 M517- E429R


15 Input "Shimadzu" to [VNC Password:].

Select the [Allow Loopback


Connections].

Check that the [User impersonation


(for Service only)] check box is
cleared, and click [OK].

■ Confirmation

(1) Display the VNC icon on [Show hidden icons] of the task bar.

2)
(2) The "UltraVNC Server" and "UltraVNC Viewer" icons are displayed on the desktop after the
setup is completed.

3)
(3) Dragging and dropping the "UltraVNC Server" and "UltraVNC Viewer" icons to the VNC
folder created in the My documents.

4)
(4) After installation and confirmation are completed, contact the Remote Maintenance Center
to request connection and operation confirmation.

M517- E429R 117


9.6. Checking after Installation

After the MPC installation is completed, check the following items.

-Ping reply from all PC to the ISDN/BB router.

-LEDs of LAN and ISDN/BB of the ISDN/BB router are turned on.

-The Windows authentication of MPC is completed. (Contact to Microsoft Corp. and input
authentication ID.)

118 M517- E429R


10 Special Operation

10.1. Operation by TeraTerm

This section describes how to set UD150B-40/L-40, ZS-100, and SGIM connected to MPC with
the serial cable to use them other than the normal operation.

When the shortcut to Ttermpro is not on the desktop, create the shortcut to C:\Program
files\TTERMPRO\ttermpro.exe and move it to the desktop.

This book describes the connection procedures. Refer to each document about the
operation of each unit.

10.1.1. Operation in UD150B-40/L-40

(1) Click [UD Terminal – COM1] on the task bar.

(2) The window of TeraTerm, which store the UD log, is displayed.

(3) Click [X] to close [UD Terminal – COM1 VT] window.

M517- E429R 119


(4) Double-click [Tera Term Pro] shortcut on the desktop.

(5) [Network Connection] is displayed. Click [Serial] and select [COM1] from the list as "Port".
Click [OK].

(6) Click [Setup]-[Serial port...].

(7) In "Serial port setup" window, set as follows and click [OK].

The configuration varies depending on the board type.


(Reference: section 9.1.6. remolog UD Configuration)

120 M517- E429R


UD CONT-2002 UD CONT-2002X
9600 38400
8 bit 8 bit
odd none
2 bit 2 bit
none none

(8) Press [Enter] key and check the prompt "UD>" is displayed.

(9) Click [X] to close [COM1 VT] window.

(10) Start "C:\B40_rlog\remolog.exe" from Windows Explore.

Make sure to start "remolog.exe". If "remolog.exe" does not start, the log of
UD150B-40/L-40 will be not stored until MPC restarts.

The connection to UD150B/L-40 using TeraTerm is complete.

M517- E429R 121


10.1.2. Operation in SGIM

(1) Double-click [Tera Term Pro] shortcut on the desktop.

(2) In [New Connection] window, click [Serial] and select [COM2] from the list as "Port".
Click [OK].

(3) Click [Setup]-[Serial port…].

(4) In "Serial port setup" window, set as follows and click [OK].

122 M517- E429R


(5) Press [Enter] key and check the prompt ">" is displayed.

(6) Click [X] to close [COM2 VT] window.

(7) Start "C:\SGIM_rlog\SGIMRlog.exe" using Windows Explore.

Do not forget the start of SGIMRlog.exe. If you do not start it, SGIM log is not saved
until the MPC restarts.

The connection to SGIM using TeraTerm is complete.

M517- E429R 123


10.1.3. Operation in ZS-100

(1) Double-click [Tera Term Pro] shortcut on the desktop.

(2) In [New Connection] window, click [Serial] and select [COM3] from the list as "Port".
Click [OK].

(3) Click [Setup]-[Serial port…].

(4) In "Serial port setup" window, set as follows and click [OK].

124 M517- E429R


(5) Press [Enter] key and check the prompt "T>" is displayed.

(6) Click [X] to close [COM3 VT] window.

(7) Start "C:\ RF_log\RF_LOG_Start.exe" using Windows Explore.

Do not forget the start of RF_LOG_Start.exe. If you do not start it, RF_Table log is not
saved until the MPC restarts.

The connection to RF_Table using TeraTerm is complete.

M517- E429R 125


10.2. How to Use Service Tools for BBRM-Agent

Do not perform "Start Agent" until you submit "Site-ViewBB Customer Registration /
Connection Consent Request".

10.2.1. Common Tool

This section describes the common tool used for all version of BBRM.

10.2.1.1. Starting BBRM-Agent (Start Agent)

At the time of shipment of the MPC, the proxy connection conditions under the LAN
environment where MPC is installed are unknown. Therefore, the BBRM-Agent is unavailable.

When installing the MPC, the proxy connection setting may be needed before starting the
BBRM-Agent. If proxy connection is needed, set up the proxy connection settings by referring to
the section below according to the BBRM version installed.

BBRM Version Section for Proxy connection setting


2.3.0 or earlier 10.2.2.1.
3.0.1 or later 10.2.3.1.

To start the BBRM-Agent, double-click the "Start Agent" shortcut in the Start menu. As shown in
the following figure, the command prompt appears and disappears after a while.

After that, In the case of BBRM Version 2.3.0 or earlier, both the Axeda-Connector and the
Axeda-Remote will be available and started automatically at the time of starting the system.
In the case of BBRM Version 3.0.1 or later, only the Axeda-Connector will be available and
started automatically at the time of starting the system.

Refer to the table below for the function of Axeda-Connector and the Axeda-Remote.
Title Function
Axeda-Connector Used in BBRM "2.3.0 or earlier" and "3.0.1 or later".
Notify the status of MPC to the BBRM-Server periodically.
Axeda-Remote Used in BBRM "2.3.0 or earlier".
Allow user to refer the console window of MPC from the
outside via the BBRM-Server.

126 M517- E429R


10.2.1.2. Stopping BBRM-Agent (Stop Agent)

Double-click the "Stop Agent" shortcut in the Start menu. In the case of BBRM Version 2.3.0 or
earlier, both the Axeda-Connector and the Axeda-Remote, in the case of BBRM Version 3.0.1 or
later, the Axeda-Connector will be stopped.
The following command prompt appears and disappears after a while.

Both the Axeda-Connector and the Axeda-Remote will not be started at the time of starting the
system.

The operation methods of the tools except the common above differ depending on the BBRM
versions installed. Refer to either section in the table below according to the BBRM version.

BBRM Version Agent Version Reference Section


2.3.0 or earlier Service Link 3.7 9.2.2
3.0.1 or later Service Link 5.2 9.2.3

M517- E429R 127


10.2.2. BBRM Version 2.3.0 or earlier

You can see starting icons of various services tools, including the proxy connection conditioning
settings, for performing introduction, operation, and administration of the BBRM-Agent in the
Start menu of the MPC. (Refer to the following figure.)
To display the Start menu, click the [Start] button at the lower left corner of the window, and then
point to the "All Programs".

The following part explains how to use these service tools.

128 M517- E429R


10.2.2.1. Setting Proxy Connection Condition for BBRM-Agent (Change ProxyAddress)

When the line for remote maintenance is available, no setting in this section is needed.
Please go to next section (9.2.2.2.).

Proxy connection that requires a user ID and password is not supported.

The BBRM-Agent has communication with BBRM-Server through the Internet.


If the proxy server exists, the BBRM-Agent needs the proxy connections settings in the LAN
environment where MPC has been installed. (The browser setting is also required.) To start
the service tool for the proxy connections settings, double-click the [Change ProxyAddress]
shortcut in the "start" menu.

The current proxy setting status and buttons are displayed. (Refer to above figure.)
Clicking [No] does not change the proxy setting, and exit the tool.
Clicking [Yes] will open the proxy setting. (Refer to the following figure.)

When clicking [No], the proxy setting will not be used (Disable), and the proxy setting
confirmation window will be displayed.
When clicking [Yes], the proxy setting will be used (Enabled), and the following proxy IP
address setting window will be displayed.
When clicking [Cancel], the proxy setting will not be changed, and the proxy setting tool will be
closed.

M517- E429R 129


In the following proxy IP address setting window, input the proxy IP address and Port number
with separating them with ":". After clicking [OK], the following window will be displayed.

In the above proxy setting confirmation window, settings are displayed.


When you apply settings, click [OK]. When you change settings, click [Cancel].

After clicking [OK], the following completion window will be displayed.

After clicking [OK], the completion window will be closed.

The proxy setting above will be applied to the BBRM-Server communication in both the
Axeda-Connector and the Axeda-Remote.

130 M517- E429R


10.2.2.2. Checking BBRM-Agent Station and Serial Number (Display SerialNumber)

Double-click the "DisplaySerialNumber" shortcut in the Start menu. The model name and serial
numbers as the Agent station of MPC are displayed.

Clicking [OK] will close the serial number display window (above figure).

M517- E429R 131


10.2.2.3. Changing BBRM-Agent Station and Serial Number (Change SerialNumber)

The serial number as the Agent station of MPC was set at the time of shipment of MPC. When
you need to change the serial number when replacing the MPC or with other reasons, follow
procedures below.
Double-click the [Change SerialNumber] shortcut in the Start menu. The serial number of the
current Agent station is displayed.

Clicking [NO] does not change the serial number, and exit the tool.
Clicking [YES] will open the following window to input the serial number.

Arbitrary character strings can be inputted into a serial number.


Clicking [OK] displays the following window.

In the above setting confirmation window, settings are re-displayed.


When you apply settings, click [OK]. When you change settings, click [Cancel].
Clicking [OK] opens the following completion window of serial number setting.

132 M517- E429R


Clicking [OK] closes the window.

M517- E429R 133


10.2.2.4. Displaying Proxy Connection Setting for Axeda-Remote (and Application to
Axeda-Connector) (Display ProxyAddress for RemoteAccess)

In normal operation, proxy setting will be common between the Axeda-Connector function and
the Axeda-Remote function by execution of "Change ProxyAddress" at the time of installation.

However, the proxy setting to the Axeda-Remote can be changed in the properties during
operation. Therefore, when the Axeda-Remote does not work correctly, the actual setting detail
can be checked.
To check the setting detail, double-click the "Display ProxyAddress for RemoteAccess" shortcut
in the Start menu. The current proxy settings in current Axeda-Remote and copy instruction
button are displayed. (Refer to the following figure.) If the settings displayed differ from those
set in section 10.2.2.1. "Setting Proxy Connection Condition for BBRM-Agent (Change
ProxyAddress)", repeat the setting in section 10.2.2.1.

When clicking [No], the proxy setting will not be changed, and the proxy setting tool will be
closed.
When clicking [Yes], copy the displayed proxy settings of Axeda-Remote to the proxy setting of
the Axeda-Connector.

After copying it, the above window is displayed. Click [OK].

134 M517- E429R


10.2.3. BBRM Version 3.0.1 or Later

The Axeda-attached tool, Deployment, is used for various setting for introduction, operation, and
administration of the BBRM-Agent. Click [Axeda] of the Start menu -> [Deployment] -> [Axeda
Deployment] to start the Deployment. (Refer to the figure below.)

The start screen of the Axeda Deployment appears. Please wait for a while.

The window below appears. Click [Agent is installed on this computer], and [OK].

(1) Click [Agent is installed on this computer].

(2) Click [OK].

The Axeda Deployment starts.

M517- E429R 135


10.2.3.1. Changing BBRM-Agent Terminal and Serial Number, and Setting Proxy Connection
Condition

The BBRM-Agent communicates with the BBRM-Server via the Internet. When the proxy exists for
connection to the Internet, proxy connection setting is necessary on MPC. (The browser setting is
also required.) Follow the procedure below.

(1) Starts the Deployment. (Refer to P.135.)

(2) Click [Basic Information] and select the "Update the Agent's identification" check box.

(1) Click [Basic Information].

(2) Select the "Update the Agent's identification"


check box.

(3) Execute "Display SerialNumber.vbs" on the Start menu.

136 M517- E429R


(4) The serial number is displayed. When this window is displayed, the serial number is copied
to the clipboard. Click [OK] to close the window.

(5) Move the cursor to "Serial" of Deployment and press [Ctrl]+[v] of the Keyboard.

Press [Ctrl]+[v].

M517- E429R 137


Next, set up the proxy connection condition. When the proxy setting is unnecessary, skip (6) and
(7).

When the line for remote maintenance is available, the proxy setting here is not need.

Proxy connection that requires a user ID and password is not supported.

(6) Click [Enterprise Servers] and select [Enable HTTP Proxy] check box.

(1) Click [Enterprise Servers].

(2) Select [Enable HTTP Proxy] check box.

(7) Input data into the "Host:" and "Port:".

138 M517- E429R


(8) Click [Deploy Changes to Agent] to save settings.

(9) The window asking whether to continue processing or not appears. Clicks [Yes].

*When clicking [Yes], the changes are saved, and the Agent is restarted
and connected with the BBRM-Server. When neither the router nor
others are set up, or when connection with the remote maintenance
center is not established, the "connection failed" message appears.
Click [OK] to close it. (Even in this case, the changes are saved.)

(10) Click [Agent] -> [Exit] to end the Deployment.

M517- E429R 139


11 BBRM Upgrade Procedure

The PCU Xcat must be 1.20.513.0 or later. If the Xcat is earlier than it, upgrade the
Xcat.

This section describes how to upgrade the BBRM to 3.0.1.


Refer to 13.1. "How to Upgrade BBRM to Version 2.3.0" for the upgrade to 2.3.0.

11.1. How to Upgrade BBRM to Version 3.0.1

If BBRM isn’t installed yet, you have to execute “D:/D3_SETUP/setup.bat” before


Upgrade to version 3.0.1 or later. In case it isn’t done, an error will occur after
installation is finished and deployment software is started, and BBRM Agent
doesn’t work.

Upgrade BBRM to Version 3.0.1 by following the steps below.


Steps Description Reference Section
1 Confirm Serial Number and Proxy Connection Setting 10.1.1
2 Uninstall the already installed version 10.1.2
3 Install BBRM Version 3.0.1 10.1.3

140 M517- E429R


11.1.1. Confirm Serial Number and Proxy Connection Setting

1) Click [Start]-[All Programs]-[Display SerialNumber].


*If click [Set Default SerialNumber] by mistake, different value than S/N on the
remote application may be displayed. Click [Display SerialNumber] and check
that the value is the number applied to the remote maintenance application form.

2) The following window will display. Write down the value of Serial Number (S/N).

*Go to step 3) only if setting Proxy server.

If serial number is wrong, disable to recognize correctly on remote maintenance


server.

3) Click [Start]-[All Programs]-[Display ProxyAddress for RemoteAccess] and write


down the value of Proxy IP-Address and Port-Number.

M517- E429R 141


11.1.2. Uninstall the already installed version

1) Insert "DAR8000 Remote Maintenance Master CD" (Revision.03.00.01 or later) in


the drive of the MPC. Execute "D:\BBRM\AgentUninstaller\AgentUnInstall.bat".

2) “Axeda Access Remote Service Successfully unregistered” message is displayed


and will disappear after 15 seconds. Enable to start next processing by clicking
[Close] button.

3) Uninstall is completed when “All operations Completed!” is displayed on command


prompt. Press any key to close the command window.

4) If “Operations Not Completed!” is displayed on the command prompt, uninstall is


not completed normally. Press any key to close the command window and reboot
MPC to uninstall again.

142 M517- E429R


11.1.3. Install BBRM Version 3.0.1

(1) Insert "DAR8000 Remote Maintenance Master CD" (Revision.03.00.01 or later) in the drive
of the MPC. Execute "D:\BBRM\Install Service Link 5.2 for DAR-8000f.vbs".

(2) Execution of the Install Service Link 5.2 for DAR-8000f.vbs will display the following
window. Click [OK] to continue processings and [Cancel] to stop it.

(3) The following window appears when [OK] is selected in step (2). Wait for a while until the
next window appears.

M517- E429R 143


(4) After a while, the window below appears. Click [Next].

(5) The window to check the license appears. Click [I accept the terms of the License
Agreement], and [Next].

(1) Click [I accept the terms of the License Agreement].

(2) Click [Next].

144 M517- E429R


(6) The window to specify the installation destination appears. Just click [Next].

(7) The window to specify the location of the shortcut appears. Just click [Next].

M517- E429R 145


(8) The windows to show installation information appears. Click [Install].

(9) After starting the installation, the window below appears. Wait for a while.

146 M517- E429R


(10) When installation is completed, the window below appears. Click [Done] to close the
window.

(11) After a while, the window to start the installation of UltraVNC appears. Click [Next>].

The window above may be hidden behind other window. Check the task bar and click
the UltraVNC icon.
Click it.

M517- E429R 147


(12) The window to check the license appears. Click [I accept the agreement] and [Next>].

(1) Click [I accept the agreement].

(2) Click [Next>].

(13) The window to show installation information appears. Click [Next>].

148 M517- E429R


(14) The window to specify the installation destination appears. Just click [Next>].

(15) The window to select the component to be installed appears. Just click [Next>].
(The "Full installation" is selected in the example below.)

M517- E429R 149


(16) The window to select the Start menu appears. Just click [Next>].

(17) The window to install the add-on appears. Clear the check box and click [Next>].

(1) Clear the check box.

(2) Click "Next>".

150 M517- E429R


(18) The window to install the mirror driver appears. Clear the check box and click [Next>].

(1) Clear the check box.

(2) Click "Next>".

(19) The window to select the additional function appears. Just click [Next>].

M517- E429R 151


(20) The window to show installation information appears. Click [Install].

(21) The installation is started and the window below appears. Wait for a while.

The window below may appear after starting the installation. Click [Yes].

152 M517- E429R


(22) The window to show the information after installation appears. Click [Next>].

(23) When installation is completed, the window below appears. Click [Finish] to close the
window.

(24) The window below appears when all the processings are completed normally. Click [OK] to
close the window.

M517- E429R 153


(25) After MPC restart, set mouse cursor on the following icon and right-click on it.

(26) The window below appears. Click [Admin Properties].

(27) The window below appears. Check [Allow Loopback Connections] check box and clear
[User impersonation (for Service only)]. Then click [Apply] and close this dialog.

Check

Clear

154 M517- E429R


(28) Then, upgrade the remolog.exe, the log monitor tool of UD150B-40.
Execute \MPC_UD_LOG_TOOL\setup.bat. The remolog.exe may run at the execution.

(29) When the installation is completed, the window below appears.

M517- E429R 155


(30) Close the window and start C:\B40_rlog\remolog.exe. The update is completed.

11.1.4. Checking Version

The version of BBRM installed in the MPC can be checked in the procedure below.
(1) Open the "C:\Axeda\BBRMVersion.dat" file in the Notepad.

(2) The version is displayed as shown below.


BBRM Version

Agent Version

156 M517- E429R


11.1.5. Error Window

11.1.5.1. Error Window by Installation Failure

The error window below appears to notify an application that has not been installed. Nothing is
displayed when all the applications have been installed normally.

The name of the application that was


not installed is displayed.

* When this error occurs, it is necessary to install only the application not installed again. Select
the execution file in the table below according to the application name displayed on the window
and execute it to again install it.

Application name Execution file (Location in installation CD)

Axeda Deployment \BBRM\Installer\Deployment\Deployment(VM).exe


Axeda Agent \BBRM\Installer\Agent\Agent for DAR-8000f.msi
EtcInstaller \BBRM\Installer\EtcInstaller.msi
UltraVNC \BBRM\Installer\UltraVNC\UltraVNC_Installer.vbs (*1)
(*1) The "UltraVNC_1.0.9.6.1_Setup.exe" in the same location cannot be available
because it is older version. Please be sure to execute "UltraVNC_Installer.vbs".

M517- E429R 157


12 Update of PCU Error Monitoring Tool (Only DAR-8000f)
Updates the PCU Error Monitoring Tool of MPC in the following procedure at the same time of
DAR-8000f update.

12.1. How to Update

(1) From the Start menu of PCU, select [All Programs] - [Accessories] - [Communications] -
[Remote Desktop Connection].

(2) The window below appears. Click [Connect].

158 M517- E429R


(3) Log on as user "Administrator", password "shimadzu" to remote-login the MPC.

(4) Exit the running PCU Error Monitoring Tool. Right-click the "PCU" icon displayed on the
right side of the task bar, and then click [Exit] on the popup menu to exit the program.

(5) With the Windows Explorer, move to "C:\PCUErrSend" and delete all the files in the folder.

(6) Insert "DAR8000 Remote Maintenance Master CD" (Revision.02.50.00 or later) in the drive
of the MPC. Execute "\MPC_PCU_ALERT_TOOL\PCUErrSendInst.bat".

(7) The next window appears. Click [Yes] for the Water-cooling FPD and click [No] for the
Air-cooling FPD.

M517- E429R 159


(8) Move to "C:\PCUErrSend" and check that "PCUErrorSend.exe" is created.

(9) From the Start menu of PCU, select [All Programs] - [Startup] - [PCU Error Send] to start
the PCU Error Monitoring Tool.

(10) Check that the "PCU" icon is displayed on the right side of the task bar. When it is not
displayed, repeat from the step (5).

(11) Return to "DAR8000 Remote Maintenance Master CD", and execute


"\MPC_TEMP_LOG_TOOL\Setup.bat ".

160 M517- E429R


(12) Next, execute "\MPC_TEMP_LOG_TOOL\CreateShortcut.vbs".

For the Air-cooling FPD, execute the steps (13) and (14). Move to step (15) for the
Water-cooling FPD.

(13) Click the [Start] button, and point to [All Programs]. Open the [TempLog-1] menu to select
the [Stop Scheduling].
The scheduler of No1 is canceled.

(14) Click the [Start] button, and point to [All Programs]. Open the [TempLog-1] menu to select
the [Start Scheduling Daily].
The scheduler of No1 is registered. (Executed early on every morning.)

For the Water-cooling FPD, execute the steps (15) and (16).

(15) Click the [Start] button, and point to [All Programs]. Open the [TempLog-1] menu to
select the "Stop Scheduling".
The scheduler of No1 is canceled.

(16) Click the [Start] button, and point to [All Programs]. Open the [TempLog-1] menu to
select the [Start Scheduling Weekly].
The scheduler of No1 is registered. (Executed early on every Sunday.)

(17) No execution is needed on [TempLog-2] menu.

The update is completed.

M517- E429R 161


13 Appendix

13.1. How to Upgrade BBRM to Version 2.3.0

13.1.1. How to Upgrade

(1) Execute the VERSIONUP.vbs in the status with the BBRM installed. The Agent may run at
the execution.

(2) Execution of VERSIONUP.vbs will display the following window. Click [Yes] to upgrade.
Click [No] to cancel upgrade.

(3) The following window appears when [Yes] is selected in step (2). Click [Close] at the
bottom of the window to complete the processing as early as possible. You may not click it.
(After predetermined time (15 seconds), the next processing will automatically start.)

162 M517- E429R


(4) The following window appears in several seconds. Click [Close] to complete the processing
as early as possible. You may not click it. (After predetermined time (15 seconds), the next
processing will automatically start.)

(5) The following window appears when all the processings are completed normally. Click [OK]
to exit upgrade processing.

M517- E429R 163


(6) The Agent does not start automatically after upgrade. Click [Start] -> [All Programs] ->
[Start Agent] to start the Agent.

The upgrade processing and restart of the Agent are completed.

164 M517- E429R


13.1.2. How to Check BBRM Version

Click [Start] -> [All Programs] -> [Display BBRM version] to check the version of BBRM installed
in the MPC.

M517- E429R 165


13.1.3. Error Window
13.1.3.1. Absence of File and Folder
The following error windows will appear due to absence of any file and folder required for
upgrade. No upgrade processing is performed.
The error window below is the example for absence of required folder.

* The BBRM may not be installed in the PC when this error occurs. When the BBRM is not
installed, execute the SETUP.bat file to set up it. In addition, when setup is performed with
SETUP.bat, execution of the VERSIONUP.vbs is unnecessary.

The error windows below are the example for absence of a file required for upgrade in the
MPC.

166 M517- E429R


13.1.3.2. File Error
The error windows below will appear due to detection of an invalid file at the time of upgrade
processing.

* When a window above appears, an error may have occurred during the upgrade
processing, resulting in data mismatching.
Therefore, once execute the UNINSTALL.bat manually to clear data, then, execute
SETUP.bat. In this case, re-execution of the VERSIONUP.vbs is unnecessary.

M517- E429R 167


<This page is intentionally left blank.>