Professional Documents
Culture Documents
Sul uu
LESSON 4
I. SUPPLIES
The current chemistry laboratory has shifted mostly to automation. Some of these
supplies may not anymore used, but it is still important that a medical
technologist to be
knowledgeable of their operation and use.
A. Laboratory Balance
A properly operating balance is vital in producing high-quality reagents. Its use
has
minimized because most chemistry laboratories are now purchasing commercially
prepared reagents.
o The most popularly used balances are analytic and electronic balances:
E-_-_-_-_-_-_-_-_-_-_-_-_-_-_-_-_-_-_-_->->->->->->->-:->:-:>:>-:>::-:-:-::— ee
prepared for CSA-B MT students by gpbdeloreyes 1
MLS 111A, Lesson 4
B. pH meters
C. Thermometers
D. Centrifuge
—re---:-:-:>:>->-EE->->->-EE—-:-:-:-:--:->:>:E---ee
prepared for CSA-B MT students by gpbdeloreyes 2
MLS 111A, Lesson 4
o Centrifugal force depends on 3 factors: mass, speed, radius. Speed is
expressed in revolutions per minute (rpm). The faster the speed and the
longer the radius, the better is the quality of the filtrate obtained
o RCF means relative centrifugal force
where:
1.118x 10° =aconstant, determined from angular velocity
r = radius in cm, measured from the center of the
1. Principle of balance.
¥ tubes and carriers must be of equal weight, shape, size, and must
be placed in opposing positions in the centrifuge head.
¥ The principle of balance is necessary to avoid noise, vibration, or
shaking of centrifuge.
v¥ An unbalanced load also decreases the speed of centrifuge leading
to improper separation of phases.
2. Always operate centrifuges with lids closed to avoid aerosol
contamination.
3. Allow the rotor to stop spinning before opening the centrifuge lid.
4. Use only tubes that are specified as appropriate for that particular
centrifuge
5. Tachometer (Speed indicator) should be used as a guide for reproducible
centrifugation.
6. Clean any spills or debris ( glass or blood )
Nowadays, plastics are becoming more popular and eventually have replaced some
glassware; there are certain characteristics of plasticware that could be better
than
glassware. Table 3 shows the comparison between plastic and glassware.
Regardless of type, the supplies must satisfy tolerances of accuracy and must
belong into 2 classes of precision tolerances ( class A or class B) given by the
American Society for Testing and Materials (ASTM).
Plastic Glass
-Leaching of surface-
bound constituents
-Permeability water vapor
-Evaporation through
breathing of plastic
-Absorption of dyes,
stains, or proteins
e Flasks have flat bottom and sloping sides that gradually narrow in diameter
so that the top opening is bottle-like. They are to hold liquids, mix solutions
on non-critical values. There are kinds of flasks:
a. Erlenmeyer flasks have a flat bottom and sloping sides that gradually
narrow in diameter. They are often used for preparing reagents and for
titration procedures. They also come in various sizes: 10,mL-4,000 mL.
b. Florence flask has a flat bottom and rounded sides that give rise to a long
cylindrical neck. The only marking seen is the total capacity in mL. Size
range: 50-2,000 mL
c. Volumetric flasks have round bulb at the bottom. This tapers to a long
neck, on which the calibration mark is found. They are calibrated “to
contain” (TC) the exact volume but are not calibrated “to deliver” (TD) the
exact volume. For each use of volumetric flask, there are certain
allowable limits within which its volume must lie. This is called the
tolerance of the flask.
e Test tubes come in various sizes. Tubes without lips are most satisfactory
because there is less chance of chipping and breakage.
Erlenmeyer laboratory
flask bottle
measuring ee
cylinder
laboratory glassware
wwwinchbyinchde ©O@Q
deliver (TD)
a. TC pipette holds a particular volume but does not dispense that exact
volume.
—re---:-:-:>:>->-EE->->->-EE—-:-:-:-:--:->:>:E---ee
prepared for CSA-B MT students by gpbdeloreyes 7
MLS 111A, Lesson 4
o Transfer
¥ .volumetric
¥ ostwald-folin
v Pasteur
¥ Automatic macropipettes or micropipettes
Mohr pipet has the same measuring technique as the serological pipet. It is
however, a TD/"no blowout” pipet. The end-calibration line for measuring the full
volume of the pipet is before the tip of the pipet. The pipet is held vertical and
the
liquid is allowed to drain from the beginning calibration line to the end
calibration
line to measure the full volume. The receiving receptacle is tilted so that the
pipet
tip touches the inside wall of the container. Mohr is more accurate than the
serological.
Volumetric pipet has a bulb in the middle of the pipet and two slender pieces of
glass on either side. It is not graduated. It measures only one volume. It is a
TD/’no
blowout” pipet. It is more accurate than serological and is used to add diluent to
a
lyophilized control or measure standards and reagents.
Micropipettes are pipets that measure small amount of a liquid, less than 1 mL.
They can be deigned either as Mohr or a serologic but they are usually “to contain”
(TC) pipets, To deliver the contained volume, the pipet contents are usually pushed
out by a small pipetting bulb and the inner walls are rinsed out with the diluent
in the
receiving receptacle using an aspiration technique and a small aspirating bulb.
This
pipet is more precise if used properly.
—re---:-:-:>:>->-EE->->->-EE—-:-:-:-:--:->:>:E---ee
prepared for CSA-B MT students by gpbdeloreyes 8
MLS 111A, Lesson 4
Pasteur and Disposable Transfer pipets are used for transferring liquid from one
receptacle to another. Disposable pipets are made of plastic, while, Pasteur
usually
have a reusable suction top and disposable glass pipet bottom.
Figure 2
TYPES OF PIPETTES
lI
a
SSS Se
°-Volumetric
-Measuring
-Mohr
Serological google.com,
. Check the pipet to ascertain its correct size, being careful also to check for
broken
delivery or suction tips. Volume indication and type of pipette can be seen near
the
mouth of the pipette.
. Wearing protective gloves, hold the pipette lightly between the thumb and the
last
three fingers.
. Place the tip of the pipet well below the surface of the liquid to be pipetted.
. Using mechanical suction or an aspirator bulb, carefully draw the liquid up into
the
pipette until the level of the liquid is well above the calibration mark.
. Quickly cover the suction opening at the top of the pipet with the index finger
. Wipe the outside of the pipet dry with a piece of gauze or tissue to remove
excess
fluid.
. Held the pipette in a vertical position with the delivery tip against the inside
of the
original vessel. Carefully allow the liquid in the pipet to drain by gravity until
the
bottom of the meniscus is exactly at the calibration mark. (The meniscus is the
concave or convex surface of a column of liquid as seen in a laboratory pipette,
buret, or other measuring devise). To do this, do not entirely remove the index
finger from the suction-hole end of the pipette; rather , by rolling the finger
slightly
over the opening allow slow drainage to take place.
—re---:-:-:>:>->-EE->->->-EE—-:-:-:-:--:->:>:E---ee
prepared for CSA-B MT students by gpbdeloreyes 9
MLS 111A, Lesson 4
8. While still holding the pipette in a vertical position, touch the tip of the
pipette to the
inside wall of the receiving vessel. Remove the index finger from the top of the
pipette to permit free drainage. Remember to keep the pipette in a vertical
position
for correct drainage. In TD pipettes, a small amount of fluid will remain in the
delivery tip.
10.Remove the pipette from the receiving vessel, and place it in the appropriate
place
for washing
A. C.
Serologic/Mohr The effect of the pipetting position
(e.g. using a 2-10 ml pipette)
hnttps://ph.lmages.search yanco.com/
se 2 At
f i}
Correct [=| / A f
aan lime fsff
Volumetric/Ostwald-Folin i_o 7 Peal f
a= J f
' Accuracy ——# Accuracy ie
4] \_! 020.4% ‘ 0.6-0.8% i ao
1
Lm f Accuracy
, | [my 1-1.2%
LH |
| { f
pee Ly Z
1H rt 7 ]
icm ) 3 fe 3 3 J
Correct
prepared for CSA-B MT students by gpbdeloreyes 10
MLS 111A, Lesson 4
B.
D.
ee
How to “take” pipette.
Bo
A. Meniscus level
google.com
Read here...
Eye level
MLS 111A, Lesson 4
Convex
Concave
Ca
Jj ¥ Low reading
By
Meniscus
“D High reading
Tip Ejector
(PA)
Push button rod
(ABS)
Display cover a
(P.
A)
Handle
(ASAIPC)
2000-10000yu1(PPS) '
—.. . e
1.
2.
3.
4.
Attach the proper tip to the pipettor, and set the delivery volume.
Place the tip into the solution, and allow the piston to rise slowly back to its
original
position.
Some tips are wiped with a dry gauze at this step, and some are not wiped. Follow
the manufacturer’s directions.
Place the tip on the wall of the receiving vessel, and depress the piston, first to
a
stop position where the liquid is allowed to drain, then to a second stop position
where the full dispensing of the liquid takes place.
Dispose of the tip in the waste disposal receptacle. Some pipettors automatically
eject the used tips, thus minimizing biohazard exposure.
Method 1:
Forward pipetting
Upper stop
ees oot Tt
Second stop
mm
Electronic pipetters or automatic micropipettors provide a variety of programmable
application parameters. The applications include the pipetting mode, the fixed-
volume
mode, and the dispensing mode. The pipetting mode allows the user to create
programs
for reverse pipetting, a method used for highly viscous fluid. The pipetting mode
includes:
The fixed volume allows the user to program a number of the most frequently used
volumes for quick recall. The dispensing mode provides repetitive dispensing of a
constant volume.
REAGENTS
Reagent is defined as any substance employed to produce a chemical reaction.
With automation, most instruments use their own ready-to-use form or “kit”.
—re---:-:-:>:>->-EE->->->-EE—-:-:-:-:--:->:>:E---ee
prepared for CSA-B MT students by gpbdeloreyes 13
MLS 111A, Lesson 4
A. Chemicals
A.1. Analytic chemicals are in varying grades of purity.
e Analytic reagent (AR) grade are of a high grade of purity and are used often in
the preparation of reagents.
e ultrapure
e Chemically Pure (CP) Grade indicate that the impurity limitations are not stated
and that the preparation of these chemicals is not uniform. Clinical laboratories
cannot use these chemicals in reagent preparation unless with reagent blank.
However, the CP designation does not reveal the limits of impurities that are
tolerated.
e United Sates Pharmacopeia (USP) and National Formulary (NF) grade are
chemicals that are generally less pure that CP-grade chemicals, because the
tolerances specified are such that USP and NF chemicals are not injurious to
B. Reagent Water
Water is the most frequently used reagent in the laboratory.
require maximum water purity. It must be used immediately after it is produced and
not to be stored.
Type Ii reagent water is used for qualitative chemistry procedures and for most
procedures done in Hematology, immunology, microbiology, and other clinical test
area.
Type Iii reagent water can be used for some qualitative laboratory tests, such as
done in general urinalysis. It is used as a source for type 1 or type II water. It
is also
used for washing and rinsing glassware.
—re---:-:-:>:>->-EE->->->-EE—-:-:-:-:--:->:>:E---ee
prepared for CSA-B MT students by gpbdeloreyes 14
MLS 111A, Lesson 4
Deionized water is obtained from passing the water through a resin column
containing positively and negatively charged particles.
C. Buffers are weak acids or bases and their related salts that, as a result of
their dissociation characteristics, minimize changes in the hydrogen ion
concentration.
Henderson-Hasselbalch equation:
pH = -log [H]