630 Peña Drive, Suite 800

ANEMOMETER CALIBRATION REPORT Davis, CA 95618-7726


Office: (530) 757-2264
Test Date: 29 March 2010 Revision No: 0
http://www.otechwind.com
Customer Information Instrument Under Test (IUT)
NRG Systems, Inc. Model No: NRG #40 Sine
110 Riggs Road Serial No: 179500149082
Hinesburg, VT 05461 Output: Sine Wave
USA IUT Power: 0 VDC
Heater Power: 0 VDC
Mount Diameter: 12.7 mm
Test Procedure: OTECH-CP-001
Wind Tunnel Test Facility Data Acquisition
Otech Tunnel ID: WT2B Hardware : National Instruments CDAQ-9172 USB 2.0 chassis
Type : Eiffel (open circuit, suction) with NI 9205 32-chan 16-bit AI module
Test Section Size : 0.61 m x 0.61 m x 1.22 m Software : National Instruments LabVIEW 8.5
Manufacturer : Engineering Laboratory Design, Inc. Signal Reduction Method for IUT: FFT to determine frequency
Measuring Equipment Test Conditions
Reference Speed : Four United Sensor Type PA Pitot-static Reference Speed Position Correction = 1
tubes sensed by an MKS Barotron Type 220D Reference Speed Blockage Correction = 1
Differential Pressure Transducer (NIST traceable) Mean Ambient Pressure = 101,162 Pa
Amb. Pressure : Setra Model 270 Barometer (NIST traceable) Mean Ambient Temperature = 24.1 deg C
Amb. Temperature : OMEGA HX94 SS Probe (NIST traceable) Mean Relative Humidity = 43.2% RH
Relative Humidity : OMEGA HX94 SS Probe (NIST traceable) Mean Density = 1.1800 kg/cubic meter

30 0.40
Reference Speed, V [m/s]

Speed Residual, ΔV [m/s]

25 0.30
0.20
20 0.10
15 0.00
10 -0.10
-0.20
5 -0.30
0 -0.40
0 10 20 30 40 0 10 20 30 40
Anemometer Signal, f [Hz] Anemometer Signal, f [Hz]

Transfer Function Test Results: V [m/s] = 0.756 f [Hz] + 0.39


Reference Anemometer Residual Ref. Speed
Regression Parameters Speed [m/s] Output [Hz] [m/s] Uncertainty
r = 0.99996 std. err. estimate =0.0673 m/s 3.961 4.903 -0.133 0.487%
slope =0.756 m/s per Hz std. err. slope =0.00213 m/s per Hz 7.957 9.998 0.011 0.479%
offset =0.39 m/s std. err. offset =0.04541 m/s 11.960 15.223 0.062 0.474%
15.933 20.496 0.048 0.487%
19.941 25.842 0.013 0.467%
23.932 31.198 -0.046 0.477%
25.928 33.802 -0.019 0.472%
21.942 28.606 -0.076 0.466%
17.946 23.190 0.024 0.478%
13.937 17.832 0.067 0.483%
9.939 12.523 0.083 0.477%
5.974 7.434 -0.034 0.480%

Note: Generic photo of test set-up


Approved by: Adam Havner,
Lab Manager

This document reports that the above IUT was tested at Otech Engineering, Inc., a wind tunnel laboratory accredited in accordance
with the recognised International Standard ISO/IEC 17025:2005 (Certificate number CL-126). This accreditation demonstrates
technical competence for a defined scope and the operation of a laboratory quality management system (refer joint ISO-ILAC-IAF
Communiqué dated January 2009). Uncertainties estimated at 95 % confidence level. This report shall not be reproduced except in
full, without written approval from Otech Engineering, Inc.

References available upon request. Page 1 of 1 179500149082_2010-03-29.pdf