Professional Documents
Culture Documents
Peek JC
Peek JC
www.elsevier.com/locate/jmbbm
Short Communication
Article history: A modified Johnson–Cook (JC) model was proposed to describe the flow behaviour of
Received 17 July 2015 polyether-ether-ketone (PEEK) with the consideration of coupled effects of strain, strain
Received in revised form rate and temperature. As compared to traditional JC model, the modified one has better
28 August 2015 ability to predict the flow behaviour at elevated temperature conditions. In particular, the
Accepted 30 August 2015 yield stress was found to be inversely proportional to temperature from the predictions of
Available online 9 September 2015 the proposed model.
Keywords: & 2015 The Authors. Published by Elsevier Ltd. This is an open access article under the CC
Flow stress
Modelling
Elevated temperature
1. Introduction and Schlegel, 2001; Wang et al., 2010). As a result, there has
been an increased demand of PEEK for medical applications.
Polyether-ether-ketone (PEEK) is a semi-crystalline polyro- In so doing, it is necessary to understand the mechanical
mantic linear polymer with a good combination of strength, properties of PEEK not only at room temperature but also
stiffness, toughness and environmental resistance (Lu et al., under elevated temperature for favourable processing condi-
1996; Jenkins, 2000). In recent years, with the confirmation of tions. In the past two decades, there has been an increasing
biocompatibility (Rivard et al., 2002), PEEK has been increas- interest in mechanical properties of PEEK (Boyce and Arruda,
ingly employed as an effective biomaterial for implantable 1990; Dahoun et al., 1995; Hamdan and Swallowe, 1996; Jaekel
medical devices such as orthopaedic, spinal and cranial et al., 2011). A series of material models were developed to
implants (Toth et al., 2006; Kurtz and Devine, 2007; EI Halabi quantify mechanical behaviours of PEEK (El Halabi et al., 2011;
et al., 2011). Compared to stainless steel and titanium, an Jaekel et al., 2011; El-Qoubaa and Othman, 2015; Garcia-
implant made of PEEK has clear benefits on temperature Gonzalez et al., 2015). However, most of these work focused
sensitivity, weight reduction and radiology advantage (Green on the mechanical properties at room temperature. Little
n
Corresponding author.
E-mail address: H.Ou@nottingham.ac.uk (H. Ou).
http://dx.doi.org/10.1016/j.jmbbm.2015.08.037
1751-6161/& 2015 The Authors. Published by Elsevier Ltd. This is an open access article under the CC BY license
(http://creativecommons.org/licenses/by/4.0/).
428 journal of the mechanical behavior of biomedical materials 53 (2016) 427 –433
thermal softening exponent, respectively. Therefore, the total where ρ is the density, Cp is the heat capacity, and ɛ is the
effect of strain hardening, strain rate hardening and thermal strain. Assuming ρ and Cp are constants, therefore, Eq. (3) can
softening on the flow stress can be calculated by multi- be rearranged to,
plication of these three terms in Eq. (1).
Z T Z εp
The temperature increase caused by deformation cannot 0:9
T ¼ T0 þ ΔT ¼ T0 þ dT ¼ T0 þ s dε ð4Þ
be neglected when the strain rate is relatively high. The T0 ρCp 0
200 200
180 180
True stress (MPa)
140 140
120 120
100 100
0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45
True strain True strain
200
160
True stress (MPa)
120
80
40
0
0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45
True strain
Fig. 2 – Comparisons of stress–strain of PEEK (a) and (b) at different strain rates at room temperature and (c) at different
temperatures and strain rate of 10 3 s 1.
430 journal of the mechanical behavior of biomedical materials 53 (2016) 427 –433
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0
0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55
T T T T
e −e
T T
e−e
At the reference strain rate of 10 3 s 1, Eq. (1) reduces to, 2.2. Modified JC model
n
sðεp ; ε̇p ; TÞ ¼ A þ Bðεp Þ ð5Þ
Similar to the case of the traditional JC model, 296 K and
10 3 s 1 are taken as the reference temperature Treference and
The value of A is calculated from the yield stress (i.e. the
strain rate ε_ reference , respectively, in deriving the modified JC
stress at strain of 2 10 3) of the flow curve at 296 K and
model. By substituting the modified temperature term in the
10 3 s 1. Substituting the value of A in Eq. (5) and using the
traditional JC model, the modified JC model is proposed as
flow stress data at various strains for the same flow curves, ln
follows:
(s A) vs ln ɛ was plotted. B was calculated from the intercept
of this plot while n was obtained from the slope. n ε̇p
sðεp ; ε̇p ; TÞ ¼ A þ Bðεp Þ 1 þ C ln
At the reference temperature, there is no flow softening ε̇reference p
term, and so Eq. (1) can be expressed as Troom =Tmelting
eT=Tmelting
e
1 λ ð7Þ
n e eTroom =Tmelting
sðεp ; ε̇p ; TÞ ¼ A þ Bðεp Þ 1 Tm ð6Þ
where A, B, n, C and λ are materials parameters. Troom is the
Using the flow stress data for a particular
h istrain at room temperature, 296 K. Adopting the same method as
different temperatures, the graph of ln 1 AþBsðεp Þn vs ln T* mentioned above, the material constants can be obtained
was plotted. The material constant m was obtained from the as, A ¼ 132 MPa, B¼ 1.0797, n ¼0. 06,802, C¼ 0.0207. It is note-
slope of this graph. worthy that the values of B, n and C are different from the
By using the experimental data (Rae et al., 2007), the values obtained by Garcia-Gonzalez et al. (2015). This is
parameters of the traditional JC model were obtained by Gar- mainly due to the mathematical treatment of the
cia-Gonzalez et al. (2015), as shown in Table 1. experimental data.
Fig. 2 shows the comparisons between the predictions of From Eq. (7), the following equation can be obtained:
JC model and experimental data. As can be seen from Fig. 2a
s eT=Tmelting eTroom =Tmelting
and b, in the range of strain rates from 10 4 s 1 to 102 s 1, 1 h
i ¼ λ ð8Þ
A þ Bðεp Þn 1 þ C ln ε̇ε̇0 p
p
e eTroom =Tmelting
the maximum deviation between the experimental data and
JC model are less than 7%. Thus the developed JC model
By using the
experimental data, the graph of
by Garcia-Gonzalez et al. (2015) can give an accurate predic- 8 9
< =
tion of the flow stress at room temperature. However, at hs
T=T
i vs e meltingTroom
T =T
e room melting
1 was plotted,
elevated temperature, as shown in Fig. 2c, the maximum : ½AþBðεp Þn 1þC ln ε_pp ; ee
=Tmelting
ε_ 0
difference between the experimental data and JC model is as shown in Fig. 3. The material constant is obtained from
38%. Therefore, the traditional JC model cannot give good the slope of the graph to be 1.5343. Thus, the following
enough predictions under elevated temperatures. Hence it is modified JC model is obtained as
highly desirable to develop new constitutive models that can h i
ε̇p
0:06802
be used to give improved prediction of the flow behaviour of sðεp ; ε̇p ; TÞ ¼ 132 þ 1:0797ðεp Þ 1 þ 0:0207 ln
ε̇reference p
PEEK at both room and elevated temperatures.
journal of the mechanical behavior of biomedical materials 53 (2016) 427 –433 431
200 200
180 180
160 160
140 140
120 120
100 100
0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45
True strain True strain
200
160
True stress (MPa)
120
80
40
0
0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45
True strain
Fig. 4 – Comparisons of stress–strain of PEEK given by the experimental data and the modified JC model (a) and (b) at different
strain rates and room temperature and (c) at different temperatures and strain rate of 10 3 s 1.
eT=Tmelting eTroom =Tmelting the modified JC model is that it gives a better prediction at
1 1:5343 Troom =Tmelting
ð9Þ
e e elevated temperatures with a maximum deviation of 13%, as
shown in Fig. 4c.
Fig. 4 shows the comparisons between the predictions of
the modified JC model and experimental data. As can be seen
from Fig. 4a and b, in the range of strain rates from 10 4 s 1
3. Strain rate and temperature sensitivity
to 102 s 1, the differences between the experimental data and
the modified JC model are less than 5% when the strain is
Fig. 5 shows the strain rate and the temperature sensitivity of
about 0.4. The maximum difference is 12% when the strain is
PEEK 450G material. From Fig. 5a, it is obvious that the yield
relative small, less than 0.1. However, the main advantage of stress is sensitive to the strain rate. In general, the yield
432 journal of the mechanical behavior of biomedical materials 53 (2016) 427 –433
280 4. Summary
80
Acknowledgements
r e f e r e n c e s
120
Lu, S.X., Cebe, P., Capel, M., 1996. Thermal stability and thermal Toth, J.M., Wang, M., Estes, B.T., Scifert, J.L., Seim Ш, H.B., Turner,
expansion studies of PEEK and related polyimides. Polymer 37 A.S., 2006. Polyetheretherketone as a biomaterial for spinal
(14), 2999–3009. applications. Biomaterials 27, 324–334.
Rae, P.J., Brown, E.N., Orler, E.B., 2007. The mechanical properties Wang, H., Xu, M., Zhang, W., Kwok, D.T., Jiang, J., Wu, Z., Chu, P.K.,
of poly(ether-ether-ketone) (PEEK) with emphasis on the large 2010. Mechanical and biological characteristics of diamond-
compressive strain response. Polymer 48, 598–615. like carbon coated poly aryl-ether-ether-ketone. Biomaterials
Rivard, C.H., Rhalmi, S., Coillard, C., 2002. In vivo biocompatibility 31, 8181–8187.
testing of peek polymer for a spinal implant system: a study in
rabbits. J. Biomed. Mater. Res. 62, 488–498.