Professional Documents
Culture Documents
Service Manual
(V1.0, Issue in 2019.06)
Page 1 of 65
AC 310 Service Manual
Caution
Biochemical hazard
Page 2 of 65
AC 310 Service Manual
Contents
CONTENTS ........................................................................................................ 3
Page 3 of 65
AC 310 Service Manual
Page 4 of 65
AC 310 Service Manual
Notice: The packed accessories of the product shall be subject to the packing list.
Page 6 of 65
AC 310 Service Manual
Page 7 of 65
AC 310 Service Manual
1.7 Startup
After confirming that all the connections related to the analyzer are correct, firstly switch on the
power socket, and then turn on the power switch of the AC 310 analyzer. After the analyzer displays the
login interface, input the user name of “admin” and input the corresponding password which originally
is also ”admin”, press “OK” to enter the interface of waiting for connection, and enter the report
interface on completion of initiation, perfusion, cleaning and blank best.
Click “ ” on the software interface, and select the blank test mode as shown in the figure
below:
Then press the count start key for blank test; if the blank fails to reach the standard, perform the blank
Page 8 of 65
AC 310 Service Manual
test again; if it still fails to reach the standard, enter “ →clean” for “clean fluid ”, and then
perform the blank test again until the blank meets the requirements shown in the table below.
WBC ≤0.2ⅹ109/L
RBC ≤0.02ⅹ1012/L
PLT ≤10ⅹ109/L
HGB ≤1g/L
HCT ≤0.5%
Page 9 of 65
AC 310 Service Manual
From the view of system, the hardware system is between the software system and the mechanical
driving system of fluid path, and it plays a role as a connecting link. The ultimate user manages the
whole resources of the analyzer through the software of the upper computer. Specifically, the software
performs the related actions and data acquisition through the hardware.
PCBA layout:
Page 10 of 65
AC 310 Service Manual
Page 11 of 65
AC 310 Service Manual
J1
J3 J2
J4
Definition of interface signal:
Page 12 of 65
AC 310 Service Manual
Position Name
J1 CBC signal pre-amplification interface
J2 WBC Apertures interface
J3 RBC Apertures interface
J4 HGB interface
PCBA layout:
Page 13 of 65
AC 310 Service Manual
Pre-amplifying board
Interface board
Main Board
Lyse Sensor
Hard Disk
IPC board
Page 14 of 65
AC 310 Service Manual
Touch panel
control board
LED display
Board
Touch screen
control board
Page 15 of 65
AC 310 Service Manual
The software architecture is divided into four layers: upper computer, middle computer, master
lower computer and slave lower computer. The lower computers are mainly responsible for the control
of partial time sequences, the execution of control command sent by the upper computer, and the
driving control of mechanical components; they also have the functions of temperature and pressure
detection, etc. The middle computer realizes the functions of storing the time sequence files of the
analyzer, executing the control command sent by the upper computer, controlling partial time sequences,
acquiring signal data, etc. The upper computer is responsible for the operations of user interaction,
operation and control of analyzer running, result display, printing output, etc. Since only the upper
machine computer system is involved in the user interaction, no more instructions will be given for the
Page 16 of 65
AC 310 Service Manual
AC 610 software
AC310 Software
Page 17 of 65
AC 310 Service Manual
Page 18 of 65
AC 310 Service Manual
LH Lyse Diluent
SV9
WBC RBC
SV4
Press
SV10 Bath
SV3
2
1
SV6
Isolation
Isolation
SV2
SV1
SV8 SV7
ASP Syringe
DIL Syringe
SV12
Waste
Pump
SV11
4.2.2 Syringe
Page 19 of 65
AC 310 Service Manual
Isolation chamber Isolation chamber Prevent foaming and backflow of waste fluid;
Air pressure sensor Air pressure sensor Monitor the air pressure in the pressure chamber
Pressure chamber Pressure chamber Store positive pressure and negative pressure
Diluent 1940μl
Test end
Page 20 of 65
AC 310 Service Manual
Diluent 200μl
Mix samples
Diluent
1719.73μl
Distribute 7.273μl of
blood to the WBC bath
Aspirate 42μl of blood
Mix WBC bath
Diluent
Lyse 300μl and
2965μl
Dilent 500ul
Mix WBC bath for
Mix RBC bath
the second time
WBC counts
RBC, PLT counts
HGB measurement
Test end
Page 21 of 65
AC 310 Service Manual
1、 Drain the RBC chamber: SV7、SV12 valve are open, and the pump drains the RBC chamber, see
WBC RBC
Press
SV10 Bath
2
1
Isolation
Isolation
SV8 SV7
SV12
Waste
Pump
SV11
Page 22 of 65
AC 310 Service Manual
WBC RBC
Press
SV10 Bath
2
1
Isolation
Isolation SV7
SV8
SV12
Waste
Pump
SV11
3、 Drain the WBC chamber(6-8.8S): SV8、SV12 valve are open, and the pump drains the WBC
4、 Create foaming positive pressure: SV11 is open, and the pump starts to create the pressure of
Press
SV10 Bath
2
1
SV8 SV7
SV12
Waste
Pump
SV11
Page 23 of 65
AC 310 Service Manual
5、 Add base solution to the WBC chamber: SV1, SV3 and SV4 valve are open, and the diluent
6、 Add blood to the WBC chamber: The sampling probe moves into the bottom of the WBC
chamber, SV1 and SV4 valves are open, and sample syringe injects the sample to the chamber; after
complete adding blood, the sampling probe moves to the upper position to clean the inner and
outer walls of the sampling probe; at the same time, SV8 valve open, and the sample is mixed by
bubbles.
7、 Create positive pressure: SV11 open, and the pump starts to create the pressure of 6.5KPA in
8、 Aspirate the sample after dilution: After mixing, the sampling probe moves into the bottom of
the WBC chamber, the sample syringe aspirate a certain volume of sample, after sampling, the
sampling probe moves up, and clean the outer wall of the sampling probe.
9、 Aspirate the Lyse: SV1, SV3, SV4 and SV9 valves open, and the diluent syringe aspirate the
Lyse.
10、Add reagents to the WBC chamber (diluent and Lyse): SV1, SV3 and SV4 valves open, and the
diluent syringe injects a certain volume of diluent and Lyse to the WBC chamber; Then SV8 valve
open, and mix the blood samples in the WBC chamber by bubbles.
11、Add base solution to the RBC chamber: SV1, SV3 valves open, the diluent syringe injects a
12、Add the sample after dilution to the RBC chamber: The sampling probe moves into the RBC
chamber, SV1 and SV2 valves open, and the diluent syringe injects a certain volume of diluent for
dilution and sampling a certain volume from WBC chamber to the RBC chamber; the sampling
probe moves to the upper position, and clean the outer wall of the sampling probe at the same
time; and then SV8 valve open, mix the samples in the RBC chamber by bubbles.
13、Create negative pressure: SV11 and SV12 valves open, and the pump starts to create the
14、Count the WBC and RBC/PLT: SV10 valve open, and it begins to count and measure the WBC
Page 24 of 65
AC 310 Service Manual
WBC RBC
Press
SV10 Bath
2
1
SV6
Isolation
Isolation
SV8 SV7
SV12
Waste
Pump
SV11
15、Measure the HGB: At 1s before the WBC measurement is finished, it begins to measure the
16、Clean the WBC chamber and RBC chamber: After blood tests, the WBC chamber and the RBC
17、Add base solution to the WBC chamber and the RBC chamber: SV1, SV3 and SV4 valve open,
and the diluent syringe injects a certain volume of diluent as the base solution into the WBC
chamber; and then SV1 and SV3 valve open, and the diluent syringe injects a certain volume of
Functional names of
Functional descriptions Remark
Fluidics
Page 25 of 65
AC 310 Service Manual
blood
Used for providing diluent to dilute the Capillary Blood sample in
Diluent
the pre-dilution mode。
All the pipelines of the analyzer are perfused with the
Fluid path perfusion corresponding reagents, and the status for sample test can be
reached after completion.
Replace Lyse Replace the Lyse in the analyzer.
Replace diluent Replace the diluent in the analyzer.
Clean the WBC chamber using the diluent; foaming for
Clean WBC chamber
strengthened cleaning is available in the process of cleaning.
Clean the RBC chamber using the diluent; foaming for strengthened
Clean RBC chamber
cleaning is available in the process of cleaning.
Clean sampling probe Clean the inner and outer walls of the sampling probe.
Clean all the devices and pipelines in contact with the blood
Cleaning fluidics
samples in the analyzer using the diluent.
Drain WBC chamber Drain the WBC chamber.
Drain RBC chamber Drain the RBC chamber.
Drain all the fluid in the analyzer, including those in the devices and
Drain fluidics
in the pipelines.
Firstly, drain all the fluid in the analyzer, then perfuse the analyzer
Shutdown after
with purified water or deionized water, then drain the fluid in the
packaging
analyzer, and finally shut down the analyzer.
Drain the pressure
Eliminate all the fluid in the pressure chamber.
chamber
Back flush the WBC and RBC Apertures by high pressure at the
Flush Apertures
same time, so as to eliminate the pollutants.
Zap the WBC and RBC Apertures at the same time, so as to
Zap Apertures
eliminate the pollutants.
Unclog Apertures Combination of back flushing and burning
Startup for normal Startup cleaning process after the analyzer is shut down through
shutdown the “shutdown” function on the software
Startup cleaning process after the analyzer is shut down not
Startup for abnormal through the “shutdown” function on the software but by directly
shutdown powering off, etc.; the cleaning degree for this process is greater
than that for the startup after normal shutdown.
The startup cleaning for shutdown after packaging equals to adding
Startup after pack-up the function of fluid path perfusion before the cleaning procedure
for normal shutdown.
Shut down the analyzer software through the “shutdown” function
on the software; before shutting down the analyzer, the analyzer
Shutdown
executes the immersing process for each chamber and pipeline
using the probe cleanser.
Page 26 of 65
AC 310 Service Manual
After the probe fluid enters the analyzer and then diluted, the
Sock Fluidics with devices and pipelines in contacted with blood samples shall be
probe Cleanser perfused, immersed and cleaned, the immersion shall last for 5
minutes, and the immersion cannot be canceled.
After the analyzer is not in used for longer time, recollect the
Enter the sleep sampling probe, and fill the WBC/RBC chamber with diluent for
immersion.
For the sleep time is less than 1 hour after sleep, when exiting from
Exit from sleep 1 (less sleep, maintain the fluid path mainly by flushing each chamber and
than 1 hour) pipeline, in order to ensure the measurement accuracy of the
analyzer.
For the sleep time is more than 1 hour and less than 5 hours after
Exit from sleep 1
sleep, when exiting from sleep, maintain the fluid path mainly by
(greater than 1 hour,
flushing each chamber and pipeline, in order to ensure the
less than 5 hours)
measurement accuracy of the analyzer.
For the sleep time is more than 5 hours after sleep, when exiting
Exit from sleep 3
from sleep, maintain the fluid path mainly by flushing each
(greater than or
chamber and pipeline, in order to ensure the measurement
equal to 5 hours)
accuracy of the analyzer.
Positive pressure
Inspect whether the process of creating positive pressure is normal
self-inspection
Vacuum pressure
Inspect whether the process of creating vacuum is normal
self-inspection
Pressure release Release the pressure in the pressure chamber and the pipelines
Recover all the fluid path devices to the prepared status for
Initialize Fluidics
measurement
Page 27 of 65
AC 310 Service Manual
Touch Screen
Aspirate Key
USB port
LAN port
Fan
Waste sensor
Power Switch
Waste
Power socket
Diluent
Grounding rod
Page 28 of 65
AC 310 Service Manual
Sampling probe
assembly
Sample syringe
Swa
b
Diluent Syringe
Sampling Probe
Diluent Sensor
Pressure Chamber
WBC chamber
RBC chamber
Page 29 of 65
AC 310 Service Manual
Pre-amplifying board
Interface board
Main board
Hard Disc
IPC board
Power Supply
Lyse sensor
Page 30 of 65
AC 310 Service Manual
Perform the sample analysis using the normal control in this interface for 3 times successively, and
calculate the mean values of “NEU peak” and “RBC peak”, respectively; the mean values of “NEU peak”
and “RBC peak” are required to be within the ranges of 265±3 and 85±1.5, respectively, and “HGB
blank” is required to range from 2800 to 3000, if:
the “NEU peak” is greater, adjust to reduce the “WBC gain” value; the “NEU peak” is smaller,
adjust to increase the “WBC gain” value.
the “RBC peak” is greater, adjust to reduce the “RBC gain” value; the “RBC peak” is smaller, adjust
to increase the “RBC gain” value.
the “HGB blank ” is greater, adjust to reduce the “HGB Gain” value; the “HGB blank AD” is smaller,
adjust to increase the “HGB Gain” value.
Page 31 of 65
AC 310 Service Manual
Enter “ → Service →Debug → Machinery” to confirm each position of the sampling probe.
【1】 The Horizontal Position of Sample is fixed with 0, Vertical Position of Origin is fixed with 0 ,Make
Bubble Motor step 100, Wbc/Rbc Depth 296.
【2】 In the process of position detection, it is prohibited to click the “Get Steps” button at random,
otherwise it will lead to the problem of error in pointer position.
【3】 Please perform “Initialize” before adjustment, and make sure to perform “Initialize” after
detecting each position.
【4】 Firstly, detect whether the horizontal position is correct; taking the detection of “WBC Cell”
position as an example, click the “Confirm Pos.” button under the “WBC Cell” position, and then
judge whether the sampling probe assembly is moved to the “WBC Cell” position; the standard
for judgment is that: click the “VM Disable” button, and then manually move the sampling
probe down until the sampling probe is inserted into the WBC chamber; the sampling probe is
required to be near the middle position of the chamber in the horizontal direction; if the
deviation is greater, firstly inspect whether the horizontal belt is loose; if yes, adjust the
horizontal belt, and then perform the detection; Make sure to perform “Initialize” before the
next detection. If it is not the problem of looseness in the horizontal belt, it is necessary to
adjust the probe position according to the following methods:
Click “HM Disable” and “VM Disable”, manually move the movable needle frame horizontally to the WBC
chamber position, then insert the sampling probe into the WBC chamber for about 30mm, adjust the
movable needle frame to make the sampling probe in the middle position at the port of WBC chamber
in the horizontal direction, and then press the “Get Steps” button corresponding to the “WBC Cell”
Page 32 of 65
AC 310 Service Manual
position in the horizontal position; on completion, the steps of the WBC chamber in the horizontal
direction will appear in the input field of the WBC chamber position; click the “Setup” button to finish
the adjustment. On completion of position adjustment, perform confirmation detection for the
debugging results, and see steps [3] and [4] for the methods.
【5】 The detection steps for other positions are the same as [4].
【6】 For detection in the vertical direction, the standard for judging whether it is in the correct
direction is that: the distance from the tip of the sampling probe to the bottom of each chamber
is about 2mm. if the deviation is greater, adjust the needle position in the vertical direction;
taking the position of WBC chamber in the vertical direction as an example according to the
following methods: click “HM Disable” and “VM Disable”, manually move the movable needle
frame horizontally to the WBC chamber position, then insert the sampling probe into the WBC
chamber for about 30mm, adjust the movable needle frame to make the sampling probe at
about 1mm on the right of the middle position at the port of WBC chamber in the horizontal
direction, insert the sampling probe to the bottom of the WBC chamber (just contact with the
chamber bottom), and then click the “Get Steps” button corresponding to the “WBC Cell”
position in the vertical position; on completion, the steps of the WBC chamber in the vertical
direction will appear in the input field of the WBC chamber position, and the steps shall be
rewritten in the input field after subtracting the steps by 16; click the “Setup” button to finish
the adjustment.
【7】 The sampling position in the vertical direction shall be confirmed according to the following
method: it is required to extend the sampling probe from the swab for 91~92mm under the
condition that the swab closely adheres to the lower edge of the mounting position, if it is not
within the range, it can be adjusted by increasing or reducing the steps in the position.
Page 33 of 65
AC 310 Service Manual
1、 Take the connecting pipelines of all reagents out of the respective reagent bottle or bucket,
enter interface of reagent replacement to replace each reagent, drain the reagents in the
pipelines in the diluent sensor and the Lyse sensor, and then input the value by adding 50 to the
readings corresponding to the “current value” in the input fields behind diluent deficient and
Lyse deficient.
2、 Click the “Setup” button to save.
3、 If a single sensor is involved, it is only necessary to debug the corresponding parameters.
6.4 Test
6.4.1 Functional test
Page 34 of 65
AC 310 Service Manual
Detection requirements:
The devices must be detected in turn, and it is not allowed to detect the next device before
finishing the detection of the former device.
Each device shall be in steady running without stagnation and abnormal noise.
For the detection of valves, 11 sounds for opening and closing the valves shall be heard.
No alarm is allowed for the detections of positive pressure and vacuum.
2. Grounding inspection
Use the 200 gear of the multimeter to test whether the resistance between all screws of the back
plate and the grounding terminal of the power supply is less than 0.5 ; it is unqualified if it is greater
than 0.5 .
3. Inspection of software version
Inspect whether the versions of the upper computer software, middle computer software, driver
board software, fluid path time sequence and FPGA are conforming with BOM; it is unqualified if they
are not conforming.
4. Inspection of waste fluid alarm function
In the testing process, when the floating ball on the waste fluid sensor is pushed to the upper part,
the analyzer interface will prompt for full waste fluid; it is unqualified if the prompt is not heard.
5. Inspection of prompt tone and alarm sound
In the report interface, press the suction key for normal test; when the sampling probe is moved up
on completion of sample suction, there will be voice prompt, and it is unqualified if the voice prompt is
not heard.
6.4.2 Performance test
1. Blank counting: In the report interface, in the CBC whole blood mode, it is required to perform blank
counting for 3 times successively, and the results shall all meet: WBC≤0.2, RBC≤0.02, PLT≤10 and HGB≦1;
if it still fails to meet the requirements, it is necessary to inspect whether the reagent is polluted, and
Page 35 of 65
AC 310 Service Manual
the analyzer shall be cleaned and maintained; if it still fails to meet the requirements, it will be
processed according to the processing for faulty machine;
2. Test of carried contamination: In the report interface, in the CBC whole blood mode, firstly perform
the test with high-value quality control for 3 times, then perform the blank test for 3 times, and record
the test results; calculate the rate of carried contamination according to the results. The rate of carried
contamination is required to be: WBC≦0.5%, RBC≦0.5%, PLT≦1% and HGB≦0.5%. Note: Requirements for
high-value sample: WBC>15, RBC>6, PLT>200 and HGB>300; low-value sample is blank test.
3. Repeatability test: In the report interface, select the CBC whole blood mode, perform the test using
more than 1mL of normal fresh human blood for 10 times successively, and calculate the CV value of the
10 test results. Test 3 human blood samples successively; the CV value of each blood sample is required
4. Analyzer calibration: Use more than 2mL of normal control blood, perform the test for 3 times on the
standard machine, calculate the mean values of WBC, RBC, HGB, PLT, MCV and MPV as the target values
for quality control, respectively; or use the fresh blood samples which have been tested on other blood
cell analyzer and printed with reports, and take the test results on the reports as the target values.
①Calibration in whole blood mode: In the interface of fresh blood calibration, select the whole blood
mode, input the target values of the normal control blood in WBC, RBC, HGB, PLT, and MCV, and perform
the test on the machine using normal control (or fresh blood samples) for at least 3 times successively;
the CV value of the test results must meet: WBC≦2% RBC≦1.5% PLT≦4% HGB≦1.5% MCV≦1%,
and the results are saved; if the CV value of the test results cannot meet the requirements, it is
necessary to re-calibrate.
② Calibration in pre-dilution mode: In the interface of fresh blood calibration, select the pre-dilution
mode, input the target values of the normal control blood (or fresh blood sample) in WBC, RBC, HGB, PLT,
and MCV; press “Add diluent”, place the test tube under the sampling needle, press the suction key for 4
times to add the diluent by the analyzer, take 80uL of blood sample using the 100uL pipette, wipe off the
outer wall of the pipette head, then add the blood sample into the test tube which has been added with
the diluent, perform the test after fully mixing, and repeat the test for at lest 3 times; the CV value of the
test results must meet: WBC≦2% RBC≦1.5% PLT≦4% HGB≦1.5% MCV≦1%, and the results are
saved; if the CV value of the test results cannot meet the requirements, it is necessary to re-calibrate.
5. Test of analyzer comparability: Perform the test using more than 1mL of normal control blood (or
fresh blood) on the standard machine for 5 times, and calculate the mean value of each item as the
reference value.
① In the test interface, select the CBC whole blood mode, test the sample for 3 times, and calculate the
mean value of each item; the percentage of deviation between the mean value and the reference value
for each item shall meet: WBC±3% RBC±2% HGB±2% PLT±5% MCV±2%。
② In the test interface, select the CBC pre-dilution mode, click “OK” and then press “Add diluent”, place
the test tube under the sampling needle, press the suction key for 4 times to add the diluent by the
analyzer, take 80uL of the sample using the 100uL pipette, wipe off the outer wall of the pipette head,
then add the sample into the test tube which has been added with the diluent, perform the test on the
machine after fully mixing, test the sample for 3 times, and calculate the mean value of each item; the
percentage of deviation between the mean value and the reference value for each item shall meet:
Page 36 of 65
AC 310 Service Manual
Page 37 of 65
AC 310 Service Manual
Cleaning function
2. When there is serious pollution, you can choose to use the various functions under the
Page 38 of 65
AC 310 Service Manual
Maintenance function
*When the analyzer is moved or transported for long distances or unused for more than 6 days, perform
the "Pack-up" action under the "Service - Maintain the whole Device" menu to complete the cleaning
and emptying of the analyzer fluid circuit and place it at a clean place.
Page 39 of 65
AC 310 Service Manual
Sign Description
Parameter is invalid or cannot be
---
calculated
+++ Parameter is beyond the display range
Parameter value alarm sign appears with the parameter result, whose specific meaning is as shown
in following table.
Sign Description
+ Parameter is beyond the measurement range
Parameter suspicious sign represents that the reliability of the parameter result
is not high. Specific to all parameters
Parameter Possible reasons
1. Poor hemolysis
2. Erythrocytes, giant platelet and platelet aggregation
WBC interference
3. Electrical noise, bubble interference
4. Hole blockage
1. Analyzer is uncalibrated
HGB 2. High leukocyte interference and lipemia sample
* 3. Abnormal HGB voltage
1. Giant platelet / platelet aggregation interference
2. Microcyte is interfered by platelet
RBC 3. Abnormal distribution of erythrocyte
4. Electrical noise, bubble interference
5. Hole blockage
1. Giant platelet / platelet aggregation
2. Microcyte
PLT 3. Abnormal distribution of platelet
4. Electrical noise, bubble interference
5. Hole blockage
::: Hole blockage
Page 40 of 65
AC 310 Service Manual
Alarm prompt message is the explain information of the alarm result, whose specific content is
shown as the following table:
Page 41 of 65
AC 310 Service Manual
Page 42 of 65
AC 310 Service Manual
Page 43 of 65
AC 310 Service Manual
Page 44 of 65
AC 310 Service Manual
Page 45 of 65
AC 310 Service Manual
Page 46 of 65
AC 310 Service Manual
Page 47 of 65
AC 310 Service Manual
Liquid
Order Possible
Circuit Solutions
Number reasons
failure
1. Check the pipeline from the swab to the liquid pump
through SV11 is connected normally and whether there
1. Connecting is leakage.
pipe is not 2. Enter "Service→Self-test" interface, and check whether
Liquid falls connected. SV11 valve can work normally.
1 out of the 2. SV11 valve 3. Enter "Service→Self-test" interface, and check the pump; if
swab failure. the pump cannot work, replace the pump. If the pump
3. Pump failure. can operate, open the pump head, check whether there
4. Swab wear. are particles or debris in the pump head diaphragm and
other positions; if so, clear them and re-test.
4. If the fault still exists, replace the swab.
1. Enter "Service→Self-test" interface, and check whether
1.SV8, SV11,SV12
SV8, SV11, SV12 valve, especially SV8 valve are working
valve failure.
WBC chamber properly.
2 2. Pump failure.
liquid overflow 2. Enter "Service→Self-test" interface, and check the
3.Drain pipeline
pump; if the pump cannot work, replace the pump. If
is detached.
the pump can operate, open the pump head, check
Page 48 of 65
AC 310 Service Manual
Page 49 of 65
AC 310 Service Manual
contaminate
d or liquid
circuit
system has
bubbles.
4. Reagent is
contaminated or
expired.
5.Noise
interference.
1. Sample
selection is
bad (WBC
abnormal 1. Select the appropriate sample.
blood). 2. Enter "Service→Maintenance→Maintain" interface and
2. Apertures perform "Unclog Aperture" action.
blockage in 3. Enter "Service→Maintenance→Clean" interface and
WBC perform "Clean WBC chamber" action; if the vacuole
chamber overflows the cup mouth after blending, clean the inner
3. WBC chamber wall of WBC chamber with soft, dustless cloth manually.
Poor WBC pollution 4. Observe whether the bubble blending process of the WBC
9
repeatability 4. RBC chamber chamber is normal. If no bubble comes out, refer to the
without correct error method above "No Bubble Comes Out in
blending the Bubble Blending Process".
bubbles 5. Check whether the Lyse is expired, and whether the action
5. Lyse fault to add Lyse is correct.
6. WBC gain 1、 Enter "Service→Debug→Basic Parameters→CBC gain"
setting is interface and re-adjust WBC gain.
inappropriat 7. Eliminate the interference source.
e
7. Noise
interference.
1. Sample
1. Select the appropriate sample.
selection is
2. Enter "Service→Maintenance→Maintain" interface and
not good
perform "Unclog Aperture" action.
(RBC/PLT
3. Enter "Service→Maintenance→Clean" interface and
abnormal
perform "Clean RBC chamber" action.
RBC/PLT blood).
4. Observe whether the bubble blending process of the RBC
10 Poor 2. Apertures
chamber is normal. If no bubble comes out, refer to the
reproducibility blockage
correct error method above "No Bubble Comes Out in
3. RBC chamber
the Bubble Blending Process".
pollution
5. Enter "Service→Debug→Basic Parameter→CBC gain"
5. RBC chamber
interface and readjust WBC gain.
without
6. Eliminate the interference source.
bubbles
Page 50 of 65
AC 310 Service Manual
4. RBC gain
setting is not
appropriate.
6. Noise
interference.
1. During the test
process,
whether
there is
small 1. Enter "Service→Maintenance→Clean" interface and
Single HGB bubbles perform "Clean WBC Chamber" action; if the vacuole
11 repeatability is attached on overflows the cup mouth after blending, clean the inner
bad. the WBC wall of WBC chamber with soft, dustless cloth manually.
chamber 2. Check whether the luminous diode is normal.
wall
2. Luminous
diode is
broken.
1. WBC blending
1. Observe that during the measurement process, whether
failure.
bubble appears in two times of beating bubbles in WBC
Erythrocytes have
Normal sample chamber. If no bubble appears, refer to the according to
not been
12 WBC test the above correct error "No Bubble Comes Out in the
dissolved
value >20 Bubble Blending Process".
completely.
2. Check whether the pipeline has bubbles when adding
2. Lyse addition
Lyse and SV9 valve is working properly.
is not enough.
Page 51 of 65
AC 310 Service Manual
Page 52 of 65
AC 310 Service Manual
16. Use default to select “Backup system to AOMEI OneKey Recovery Partition”, and click “Next”
Page 53 of 65
AC 310 Service Manual
18. The backup takes about 15-20 minutes, and the backup is completed when the following interface is
displayed
Page 54 of 65
AC 310 Service Manual
2. Click file, and create a new task, input “explore” then Click “OK”
Page 55 of 65
AC 310 Service Manual
3. When machine start, it automatically start from AC 310 of D disk, the E disk is for backup,
once the application get error or files lost, you can copy the AC 310 of E disk .
Page 56 of 65
AC 310 Service Manual
4. Copy the AC 310 folder from E disk backup, and cover the AC 310 folder in D disk.
Page 57 of 65
AC 310 Service Manual
Page 58 of 65
AC 310 Service Manual
Click Capture options icon, and select which network connection you want to
capture.
In our analyzer, capture the network which connect with main board, the name
may “Local area Connection”
After you press start, it will start capture IP, and several seconds, you can click stop icon
,and then you check the TCP protocol, and that is what we need.
Page 59 of 65
AC 310 Service Manual
In our analyzer because that network connection only connect with main board, so in analyzer it will be
only one TCP, and the source IP is IPC address, and the Destination IP is main board IP address . Then set
the network IP same as destination IP of main board.
1. Get the Update package from Wheisman Service. And unzip it to folder.
2. Enter “Service --Software Update”, Select “Analyzer Software”, then Click Update.
Page 60 of 65
AC 310 Service Manual
Page 61 of 65
AC 310 Service Manual
Page 62 of 65
AC 310 Service Manual
Set the “Choose when to turn off the display” and “Change when the computer sleeps”to never
turn off display and sleep as below:
Page 63 of 65
AC 310 Service Manual
Page 64 of 65
AC 310 Service Manual
Page 65 of 65