You are on page 1of 30

User Manual

REF 66800059, 66800697


and 66801243 only
EN

User Manual
*smith&nephew
only RENASYS™ EZ
RENASYS™ EZ
PLUS
Negative Pressure Wound
Therapy

Table of Contents
Introduction 4 Information signals and safety alarms 13

Device description 4 Troubleshooting guide 16

Device orientation 6 Maintenance 21

Important information Cleaning 21


Monitoring NPWT 6
Battery operation 21
Indications for use 7
Returning device 21
Contraindications 7
Storage 21
Warnings 7
Electromagnetic compatibility
Precautions 8 RENASYS EZ REF 66800059 22

Physician orders 9 Electromagnetic compatibility


RENASYS EZ PLUS
Dressing changes 10 REF 66800697, 66801243 24
Canister selection and installation 10 Caution statements 26
Connect RENASYS Soft Port Essential performance 26
dressing and canister 11
Specifications 27
Disconnect RENASYS Soft Port
dressing and canister 11 Glossary of symbols 29

Operating RENASYS EZ and Limited warranty 31


RENASYS EZ PLUS devices 12
Before NPWT 12 Global customer assistance 351
Set NPWT 12
Adjust NPWT 12
Lock NPWT 12
Start NPWT 12
Stop NPWT 12
EN
Introduction
This user manual contains important information regarding the safe and effective operation of RENASYS™ EZ
(REF 66800059) and RENASYS EZ PLUS (REF 66800697; REF 66801243) Negative Pressure Wound Therapy [NPWT]
devices. These devices are intended for use by or on the direction of a trained and licensed physician. This manual
is intended to aid in training of personnel and to provide a reference for experienced users.

Front view of device


(RENASYS EZ PLUS REF 66801243 shown)
Device description
The RENASYS EZ and RENASYS EZ PLUS
suction devices are designed to provide Handle
NPWT to a closed environment over a
wound, in order to drain exudates from Vacuum port
the wound site to a disposable container,
Pressure
which may promote wound healing via
Canister gauge
removal of irrigation and body fluids, wound
holder bracket
exudates and infectious materials. 90 100

AC mains 80 120

Pressure
The closed environment is created by power and
70 140

60 160
selector knob
applying a RENASYS sterile wound alarm indicators 50 180

dressing to the wound site and


40 200

Pressure
connecting the sealed wound to the Audio pause selector lock
suction device. The wound dressing kit is button and
considered to be the patient applied part. indicator Mode of
operation
The RENASYS EZ and RENASYS EZ PLUS run on AC switch
mains power or internal battery power. The internal
battery charges when device is plugged into AC mains
power both during operation and when turned off and
not in use.
The RENASYS EZ and RENASYS EZ PLUS should be
positioned at or below wound height and face user/ Rear view of device
operator’s location when in use.
For proper and effective use of RENASYS EZ (REF
66800059) and RENASYS EZ PLUS (REF 66800697, Handle
66801243) devices, ensure use only with RENASYS EZ
PLUS/EZ MAX canisters, RENASYS dressing kits and
RENASYS ancillary componentry.
IV pole
Refer to instructions for use provided with dressing knob
kits for additional information on dressing use and
maintenance.
Bed hooks

Specification
badge

IV pole pad

4
EN
Rear view with bed hooks extended Left view of device

AC mains power
inlet
Fuse cover

Rear view with IV pole knob in use Underside of device

Air exhaust outlet


Rubber feet Rubber feet

5
EN
Device orientation Important information
The device is designed to operate only in the upright
Monitoring NPWT
position while delivering NPWT. The device can stand
Carefully monitor the patient, device, and dressing
on a surface, hang on a bedrail or hang on an IV pole.
frequently to determine if there are any signs of
When standing upright, ensure placement is on a flat
bleeding, exudate accumulation (pooling), infection,
hard surface.
maceration, or loss of Negative Pressure Wound
Caution: Any orientation other than described could Therapy (NPWT). The frequency should be determined
result in device malfunction, compromising patient by the clinician based on individual characteristics of
safety and/or damage the device. the patient and wound. NPWT devices are not designed
to detect or issue an alarm condition based on the
Device and system tubing should be positioned at or presence of bleeding or pooling. These conditions may
below wound height and face user/operator's location only be detected by frequent monitoring.
when in use.
Special attention to the risks of bleeding or loss of
Attach to bedrail: Device can be attached to the NPWT should be considered when prescribing for
bedrail of the patient's bed using the device bed hooks. use in the Home Environment.
This is achieved by pulling the 2 metal hooks from the
recess on the back of the device to a 90° angle. The NPWT may be impacted by various conditions
hooks should be placed over the bedrail to allow the related to system configuration, set-up and individual
device housing to hang from the bedrail. characteristics of the patient and wound (e.g. exudate
characteristics, patient anatomy). Alignment of the
Attach to IV pole: Device can be attached to an IV port to the opening in the drape, use of a bridging
pole by rotating IV pole clamp knob enough to allow technique and choice of dressing configuration based
the IV pole to be located between the upper rubber on wound characteristics may impact NPWT vacuum
pad and clamp. Ensure IV pole alignment with the lower delivery over the course of therapy. Exudate volume,
rubber pad. Once the IV pole is correctly located, the viscosity and consistency may influence fluid removal
knob must be carefully tightened to ensure the pads or occlusion formation. A full canister, incorrect canister
grip and hold the IV pole securely. orientation and device/tubing height relative to the
wound can contribute to loss of NPWT and exudate
Caution: When not in use, ensure IV pole knob and
accumulation within the wound, which could lead to
bed hooks are completely closed.
maceration, infection, or unrecognized bleeding.
Monitor the wound for infection and ensure that all
wound filler is removed at each dressing change to
reduce the risk of infection.
Skin grafts should be closely monitored to ensure
NPWT is being delivered.
Review Contraindications, Warnings & Precautions
before use.

6
EN
Indications for use Warnings
RENASYS™ EZ and RENASYS EZ PLUS are indicated 1. Carefully monitor patients for signs of bleeding,
for patients who would benefit from a suction device which may lead to interruption in therapy and
(Negative Pressure Wound Therapy), as it may promote hemodynamic instability. If such symptoms are
wound healing via removal of fluids, including irrigation observed, immediately discontinue therapy, take
and body fluids, wound exudates and infectious appropriate measures to control bleeding, and
materials. contact treating clinician.
Appropriate wound types include: 2. Patients suffering from difficult hemostasis or
• Chronic who are receiving anticoagulant therapy have an
increased risk of bleeding. During therapy, avoid
• Acute using hemostatic products that, if disrupted, may
• Traumatic increase the risk of bleeding.

• Sub-Acute and dehisced wounds 3. Do not use directly on exposed blood vessels or
organs. Sharp edges such as bone fragments must
• Ulcers (such as pressure or diabetic) be covered or removed prior to initiating therapy,
• Partial-thickness burns due to risk of puncturing organs or blood vessels
drawn closer under the action of negative pressure.
• Flaps and grafts
4. NPWT has not been studied on pediatric patients.
Patient size and weight should be considered
Contraindications when prescribing the device.

Use of device is contraindicated in presence of: 5. Foam or gauze must not be tightly packed or forced
into any wound area. Over-packing may interfere
• Untreated osteomyelitis
with distribution of NPWT evenly across the wound.
• Exposed arteries, veins, organs or nerves This may decrease the ability of the wound to
properly contract and permit exudate to remain in
• Necrotic tissue with eschar present wound.
• Malignancy in wound (with exception of palliative 6. In the event defibrillation is required, disconnect
care to enhance quality of life) device from the wound dressing prior to
• Non-enteric and unexplored fistulas defibrillation. Remove the wound dressing only if its
location will interfere with defibrillation.
• Anastomotic sites
7. The device is not MRI compatible. Do not bring
device into MRI suite. Prior to entering MRI suite,
disconnect device from dressing. The dressing can
remain intact on patient.
8. Device is unsuitable for use in areas where there is
danger of explosion (e.g., hyperbaric oxygen unit).
9. When operating, transporting or disposing of
device and accessories, there is risk of infectious
liquids being aspirated or contamination of
device assembly through incorrect use. Universal
precautions should be observed whenever working
with potentially contaminated components or
equipment.
10. Device and canister are provided non-sterile and
should not be placed within a sterile field.

7
EN
Precautions
1. More frequent device and wound dressing Cover wound bed, including fistula opening, with
monitoring, should be taken for patients who are non-adherent gauze or with one layer of saline
or may be: moistened gauze. During the course of treatment,
• Suffering from infected blood vessels patient’s fluid levels must be closely monitored.
• Receiving anticoagulant therapy or platelet 9. Avoid use of circumferential dressings except in
aggregation inhibitors, in addition to patients cases of oedema or heavily exuding extremities,
with intrinsic coagulation problems such as low where this technique may be necessary to
platelet counts maintain a seal. Consider using multiple drapes
• Actively bleeding or have friable blood vessels to minimize risk of decreased distal circulation.
or organs Regularly assess distal pulses, and discontinue
therapy if changes in circulation are detected.
• Suffering from difficult wound hemostasis
• Untreated for malnutrition 10. Monitor patient for any signs of local or systemic
infection. Infected wounds may require more
• Noncompliant or combative
frequent dressing changes. As NPWT is not
• Suffering from wounds in close proximity to intended to directly treat infection, if there are any
blood vessels or delicate fascia signs of systemic infection or advancing infection at
When monitoring patients for delivery of therapy, wound area, contact treating clinician immediately.
ensure wound dressing is free of air leaks, fully
11. If multiple pieces of foam or gauze are needed
compressed and firm to the touch. Ensure that
to fill the wound profile, count and record how
pressure indicated on pressure gauge reflects set
many pieces are present to ensure all pieces are
pressure on pressure selector knob.
removed at a dressing change to minimize the risk
2. As a condition of use, device should only be used of retention and possible infection.
by qualified and authorized personnel. User must
12. NPWT should remain on for duration of treatment.
have necessary knowledge of the specific medical
The length of time a patient may be disconnected
application for which NPWT is being used.
from device is a clinical decision based on individual
3. For patients with high risk of bleeding use 250ml characteristics of patient and wound. Factors to
canister. Ensure the 250ml canister viewing window consider include location of wound, volume of
is checked frequently for signs of bleeding, as the drainage, integrity of dressing seal, assessment of
canister holder obscures the viewing window. bacterial burden and patient’s risk of infection.

4. In the event of heavy or viscous drainage, drainage 13. Do not use a dressing kit with breached or
with sediment or when blood is present, regular damaged packaging.
monitoring and more frequent dressing changes
14. Use of NPWT presents risk of tissue in-growth.
may be required to reduce the risk of interruption of
Tissue in-growth may be reduced by decreasing
therapy, maceration, infection, and ensure proper
therapy pressure, using a wound contact layer or
exudate removal.
increasing the frequency of dressing changes.
5. Underlying structures, such as bone, tendons, 15. NPWT should not be painful. If patient reports
ligaments and nerves should be covered with
discomfort, consider reducing pressure setting and
natural tissue or a non-adherent dressing layer
use of a wound contact layer. Pressure setting is a
prior to applying the NPWT dressing to ensure
clinical decision based on individual characteristics
protection and minimize the risk of damage from
of patient and wound. Factors to consider include
direct contact with the dressing.
location of wound, volume of drainage and integrity
6. To minimize risk of bradycardia, do not place NPWT of dressing seal.
in proximity to the vagus nerve.
16. Maintain regular monitoring of device and wound
7. In the event a patient with spinal cord injury site during therapy to ensure therapeutic treatment
experiences autonomic dysreflexia, discontinue use and patient comfort.
of NPWT and immediately seek medical assistance.
17. Device is only to be used with Smith & Nephew
8. When treating enteric fistulas, do not place NPWT authorized components. Use of any other products
dressing in direct contact with exposed bowel. have not been proven safe and effective with
RENASYS™ EZ and RENASYS PLUS devices.

8
EN

18. Ensure canister tubing and RENASYS Soft Port are 28. Canisters are single-use devices. Do not reuse.
installed completely and without any kinks to avoid
leaks or blockages in vacuum circuit. Position device
29. Do not apply SECURA™ No-sting barrier film wipes
directly to open wounds. SECURA No-sting barrier
and tubing appropriately to avoid risk of a trip hazard.
film is flammable. Use in a well ventilated area.
Device and system tubing should be positioned
Avoid using around flames and sources of ignition.
level with or below wound to ensure optimization of
Keep out of reach of children. For external use only.
therapy and prevent therapy interruption.
19. When bathing or showering patient must 30. As with all adhesive products, apply and remove
dressing carefully from sensitive or fragile skin to
disconnect from device, protecting both ends
avoid blistering and skin stripping, especially after
of Soft Port tubing using tethered caps. Ensure
frequent dressing changes. Use of skin sealant
aeration disc located near quick click connector
may assist with protection of periwound skin.
is free of moisture before reactivation of therapy
to ensure proper alarm functionality and prevent 31. If patient must be disconnected, the ends of the
interruption in therapy. RENASYS Soft Port and canister tubing should be
protected using tethered caps to avoid leakage of
20. If any liquids penetrate device, discontinue use and
fluid and cross contamination.
return to your Smith & Nephew authorized provider
for service. 32. When using RENASYS F/P or G/P wound dressing
kits, ensure dressing clamp is engaged before
21. CT scans and x-ray have the potential to interfere
turning off device. For reconnection to device,
with some electronic medical devices. Keep device
ensure that device is active prior to opening
out of x-ray or scanner range.
dressing clamp.
22. Use caution if device is used in presence of a
flammable anesthetic mixture with air or with Physician orders
oxygen or nitrous oxide.
Prior to placement of RENASYS EZ and RENASYS EZ
23. Ensure device is placed on a stable level surface.
PLUS, the medical professional treating the wound must
When placed on an uneven surface, device can
assess how to best use the system for an individual
become unbalanced as exudate fills canister.
wound. It is important to carefully assess the wound and
24. AC mains power can only be removed by patient to ensure clinical indications for NPWT are met.
disconnecting power cord. Take care in positioning All orders should include:
device to allow access to cord receptacle.
• Wound location, size and type
25. If power cord is damaged, wires are frayed or • Smith & Nephew wound dressing kit
exposed, do not use power cord. Contact your
• Pressure settings
Smith & Nephew representative for a replacement
cord. • Frequency of dressing changes
• Adjunctive dressings
26. Canisters should be changed at least once a
week, whenever there is a change of patient or in
the event that canister contents reach maximum
volume indication (250ml or 800ml fill line).
Do not wait for canister over-capacity alarm
activation to change canister.
27. The in-line bacterial overflow guard is designed for
single patient use and is provided as part of the
single-use canister. The guard is replaced each
time canister is changed. Check guard for any
changes in color, liquid in guard or poor vacuum
performance. Replace canister, inclusive of guard, if
any of these occur.

9
EN
Dressing changes Canister selection and installation
Refer to instructions for use provided with dressing Device is to be used with RENASYS™ EZ/EZ PLUS/EZ
kits for additional informaton on dressing use and MAX 250ml and 800ml canisters.
maintenance.
Canisters are non-sterile and should not be used in a
1. Foam dressings should be changed every 48 to 72 sterile field.
hours after initial application of therapy. If no leak
Canisters are single-use devices. Do not re-use.
is present and the patient is comfortable, dressing
changes should occur no less than 3 times per
week. 250ml canister
CAUTION:
2. Gauze dressings should be changed 48 hours after
initial application of therapy. If no leak is present Ensure
and patient is comfortable, dressing changes the 250ml
should occur 2–3 times per week. canister
viewing
3. In event of heavy or viscous drainage, drainage window is
with sediment, or when blood is present, regular checked
monitoring and more frequent dressing changes frequently for signs of bleeding, as the canister
may be required. holder obscures the viewing window.
4. When dressing a wound involving difficult to seal For patients with high risk of bleeding, use 250ml
anatomy or exposure to external moisture, frequent canister.
inspection of dressing is recommended to ensure a
seal is maintained. Ensure wound dressing is fully 800ml canister
compressed and firm to the touch. CAUTION:
Ensure the 800ml canister
5. Ensure all wound filler material placed in wound viewing window is checked
800ml

700ml

has been removed before redressing wound. If frequently for signs of bleeding. It
600ml

500ml

foam dressing adheres to wound, apply normal may be necessary to lift canister
400ml

saline into wound dressing and let it set for


300ml

from canister holder to view 200ml

15–30 minutes before gently removing foam. canister contents.


100ml

Appropriately discard used wound dressings


observing your institution’s protocol for medical
waste handling. RENASYS EZ/EZ PLUS/EZ MAX canisters use an in-
line bacterial overflow guard for protection of device
6. As with all adhesive products, apply and remove against overflow and spread of aspirated micro-
dressing carefully from sensitive or fragile skin organisms. Guard is designed for single-patient use
to avoid skin stripping, especially after frequent and is provided as part of single-use canister. Guard
dressing changes. Use of skin sealant may assist is replaced each time canister is changed. Check
with protection of periwound skin. guard for any changes in color, liquid in guard,
or poor vacuum performance. Replace canister,
7. Monitor patient for any signs of local or systemic inclusive of guard, if any one of these occur.
infection by checking dressing regularly. Infected
wounds may require more frequent dressing Canisters should be changed at least once a week,
changes. As NPWT is on intended to directly whenever there is a change of patient or in event that
treat infection, if there are any signs of systemic canister contents reach maximum volume indication
infection or advancing infection at wounded area, (250ml or 800ml fill line). Do not wait for canister
contact treating clinician immediately. over-capacity alarm activation to change canister.
Canisters may have to be changed regularly within
single-patient treatment episodes if exudate levels
are high. Check canisters regularly to monitor exudate
levels, ensuring they are below maximum volume
indication (250ml or 800ml fill line).
Check canister for any signs of cracks or damage. If
noted, discard and replace canister.

10
EN
Installing canister holder Connect RENASYS™ Soft Port dressing
Slide canister holder down onto canister holder bracket and canister
on the front of device housing until it stops.
Connect Soft Port
Canister installation to canister tubing
by pushing orange
1. Ensure device is off. quick click connectors
2. Prior to placing canister into canister holder on together.
device, connect blue end of canister tubing to An audible click
canister lid port labeled with patient symbol: indicates connection is
secure.

3. Place canister into canister holder and insert in-line


bacterial overflow guard into vacuum port on front
of device. Note: Do not cover aeration disc.
Important! The in-line bacterial overflow guard
must be inserted past the fourth ridge to a point
approximately halfway between the fourth ridge and
Disconnect RENASYS Soft Port
the face of in-line bacterial overflow guard. Failure to dressing and canister
fully insert may result in a leak in the system, which 1. Hold orange quick click connector above wound to
may impact device alarm functionality. help ensure exudate does not leak from tubing.
2. Turn device off.
3. Disconnect canister tubing from Soft Port.

432 1
90 100
100

80 120
120

70 140

60 160

In-line bacterial
50 180
40 200

overflow guard
PLUS

Canister holder

4. Insert tethered cap into both sides of orange quick


Note: Inserting in-line bacterial overflow guard click connector.
completely may require force.
Note: Ensure in-line bacterial overflow guard (filter)
component comes to a "hard stop" when fully
inserted past the fourth ridge as described above.
4. To verify functionality of complete blockage alarm,
turn device on and insert tethered cap of canister
5. Disconnect in-line bacterial overflow guard from
tubing into connector to simulate a blockage. A
device.
functional device will activate a complete blockage
alarm within 5 minutes. 6. Lift canister out of holder.
In the case that complete blockage alarm does not 7. Discard canister, in-line bacterial overflow guard
activate, check canister installation and contact and tubing as one piece.
your Smith & Nephew authorized representative.
8. Disposal of used canisters should follow
5. Securely connect opposite end of canister tubing facility protocols or local ordinances
to Soft Port. relating to handling of potentially infected
or bio-hazardous materials.

11
EN
Operating RENASYS™ EZ and RENASYS
EZ PLUS devices
1. Before NPWT Continuous mode: Device will maintain preset
If battery operation is required for first use of device, pressure setting without stopping until switched off.
battery must be charged by connecting device to AC
mains power until battery indicator illuminates solid Continuous
green. During charging process, battery indicator will
flash green. Intermittent mode:

Intermittent
2. Set NPWT
• RENASYS EZ (REF 66800059): Device cycles ON and
Pressure setting is a decision that the healthcare
OFF in increments of approximately 32 seconds ON
provider must make based on an individual
(active vacuum) and approximately 16 seconds OFF
assessment of the particular wound.
(no vacuum).
40–120mmHg is recommended therapeutic pressure
• RENASYS EZ PLUS (REF 66800697, 66801243): Device
range.
cycles ON and OFF in increments of approximately 5
In the event you have a wound that is difficult to seal, minutes ON (active vacuum) and approximately
you may increase the pressure setting to achieve 2 minutes OFF (no vacuum).
desired negative pressure at the wound bed.
Note: RENASYS EZ/EZ PLUS devices offer both
continuous and intermittent modes. Selection of
3. Adjust NPWT therapy mode setting is a clinical desicion based on
Device features a 12-position adjustable pressure physician judgement, wound type, exudate level,
selector knob. Pressure level is indicated on an analog exudate consistancy and patient comfort.
pressure gauge above pressure selector knob when
Intermittent therapy is not recommended for:
device is actively creating vacuum. Pressure setting is
increased by turning knob clockwise. When not in use, • Highly exudating wounds
pressure selector knob should be turned to minimum • Wounds with tunnels or undermining
setting. • Wounds in difficult areas where maintaining a seal is
problematic
4. Lock NPWT • Patients who experience pain during intermittent
Pressure selector knob can be locked to ensure no therapy
accidental adjustment in pressure is made. Turn
pressure selector lock to LOCK position as shown Caution: Before starting NPWT ensure that device and
below: system tubing is positioned horizontal with or below
the wound and is away from any direct sources of heat.
Caution: After starting NPWT, confirm that device
is operating at pressure setting that was selected
by observing pressure level on pressure gauge.
The needle on gauge may teeter around set point
(approximately ±10mmHg from set point).
Note: Sealed wound dressings connected to active
NPWT should be free of air leaks, fully compressed
5. Start NPWT and firm to the touch.
To begin NPWT, turn pressure selector knob to desired
6. Stop NPWT
pressure setting and press mode of operation switch
To stop NPWT, return mode of operation switch to
ON for either continuous or intermittent mode. NPWT
center position. Turning OFF device will stop vacuum
will commence as soon as the switch is selected.
and end delivery of NPWT.
Check device’s pressure gauge periodically as this
provides an indication of pressure level created,
keeping in mind there will be 2 minute intervals when OFF
device is not actively delivering pressure if set in
intermittent mode.

12
EN
Information signals and safety alarms
Devices are equipped with both information signals If device has been exposed to temperatures outside its
and alarms, which are described below. recommended temperature range, let device return to
room temperature, prior to activation.
Alarms for RENASYS™ EZ PLUS (REF 66801243) are
determined to be "Low Priority," indicating operator If alarm persists, contact your Smith & Nephew
awareness is required (IEC 60601-1:2005, 3rd edition authorized representative.
and IEC 60601-1-8:2006).
Over vacuum alarm:
AC mains power information signal
If system encounters an excessively high vacuum (of
(Feature available only on RENASYS EZ: REF 66800059)
>235mmHg), device will stop delivering NPWT.
When the system is connected to AC mains power, the
status light will illuminate solid green. This does not • REF 66800059 and 66800697:
indicate the device is turned on. Audible alarm will sound and indicator will flash
amber.
ON/OFF information signal • REF 66801243:
(Feature available only on Audible alarm will sound and indicator will illuminate
RENASYS EZ PLUS: REF 66800697, 66801243) solid yellow.
When mode of operation switch is ON in continuous or Pressing audio pause button will pause alarm for
intermittent mode, the status light will illuminate solid approximately 2 – 4 minutes.
green. To reset alarm function, device must be switched off by
returning mode of operation switch back to
Low battery alarm: OFF position, then turned back ON.
Battery has up to one hour of therapy time remaining.
If alarm occurs again, there is potentially a fault with
• REF 66800059 and 66800697: device. Contact your Smith & Nephew authorized
Audible alarm will sound and battery indicator will representative.
flash amber.
Note: After device stops delivering therapy due to an
• REF 66801243: over vacuum alarm, pressure in wound will begin to
Audible alarm will sound and battery indicator will drop. The low vacuum alarm will also be activated.
illuminate solid yellow.
Cautions – Lack of alarms:
Pressing audio pause button will pause alarm for
approximately 2 – 4 minutes. Devices are equipped with high flow/leak, low vacuum
Plug device into AC mains power to charge device and complete blockage/canister over-capacity
battery immediately when alarm occurs. alarms to alert users to these events occurring during
therapy. These alarms are designed to activate based
Battery fault alarm: on changes in pressure status detected by device.
(Feature available only on However, there are scenarios that may occur during
RENASYS EZ PLUS: REF 66800697, 66801243) therapy that can impact alarm functionality. Therefore,
Battery within device has failed to charge. it is important that patient, device and wound dressing
are monitored at an appropriate interval to ensure
• REF 66800697: therapy is being delivered.
Battery indicator will illuminate solid amber,
indicating that device has encountered a fault within
battery/charging system. There is no audible alarm.
• REF 66801243:
Battery indicator will illuminate solid cyan (light blue),
indicating that device has encountered a fault with
battery/charging system. There is no audible alarm.
Connect device to AC mains power to continue
therapy; device will not operate on battery power if
battery fault alarm has activated.

13
EN
High flow/leak alarm: Complete blockage/canister
Device detects a significant leak.
over-capacity alarm:
Device detects that canister has exceeded maximum
• REF 66800059 and 66800697: volume capacity or there is a complete blockage in the
Audible alarm will sound and indicator will flash system. There will be no negative pressure at wound
amber. site.
• REF 66801243:
• REF 66800059 and 66800697:
Audible alarm will sound and indicator will illuminate
solid yellow. Audible alarm will sound and indicator will flash
amber.
Pressing audio pause button will pause alarm for
approximately 2 – 4 minutes. • REF 66801243:
Audible alarm will sound and indicator will illuminate
Once system is sealed, alarm(s) will automatically reset.
solid yellow.
Note: An air leak may result in a pressure drop Pressing audio pause button will pause alarm for
occurring within the system. As a result, the low approximately 2 – 4 minutes.
vacuum alarm may also activate while the high flow/
leak alarm is active. Once corrected, the audible alarm and indicator will
reset.
Cautions – High flow/leak alarm:
Cautions – Complete blockage alarm:
• When a significant air leak is present in system, the
high flow/leak alarm will activate. However, if a partial • If partial blockage occurs, the change in pressure
blockage is present within the system it may inhibit status detected by device may not be significant
device's detection of a significant air leak, resulting in no enough to trigger alarm activation. Over time if
alarm activation. Potential sources of blockage include: blockage reaches point of full occlusion, complete
blockage alarm will activate .
–– Physical occlusion in wound dressing (coagulated
blood or purulent material in filler, compacted • Blockage formation within wound dressing will
filler, high volume viscous fluid). not be detected by the system as it has occured
outside the monitored vacuum circuit, but can affect
–– Physical occlusion in tubing (kink in canister pressure status at the wound. Appropriately frequent
tubing, clot in tubing). monitoring of wound dressing is recommended to
–– Soft Port aperture misaligned to dressing opening. confirm adequate delivery of therapy.

• Check wound dressing regularly to ensure it is fully • If a complete blockage is present in system, but an
compressed and firm to the touch. air leak occurs between blockage and device, the
alarm may not activate. Ensure that all connections
are secure and there are no air leaks present in
Low vacuum alarm: system. Potential sources of air leaks include:
Vacuum level is lower than set point of –– Tear in Soft Port
NPWT by >15mmHg.
–– Misplaced or worn O-ring within orange quick
• REF 66800059 and 66800697: click connection
Audible alarm will sound and indicator will flash –– Loose blue canister tubing attachment (to port on
amber. top of canister lid)
• REF 66801243: –– Cracked or damaged canister
Audible alarm will sound and indicator will illuminate
solid yellow. –– Partially inserted in-line bacterial overflow guard

Pressing audio pause button will pause alarm for Note: For RENASYS F/P or G/P wound dressing kits,
approximately 2 – 4 minutes. the complete blockage alarm will only activate if a
complete blockage occurs with the canister tubing.
Once system is sealed, alarm(s) will automatically reset.
There is no blockage detection in the dressing tubing.
Note: If low vacuum alarm is due to a leak in the
system, the high flow/leak alarm may also activate
while low vacuum alarm is active.

14
EN
Cautions – Canister over-capacity alarm:
• Canister over-capacity alarm activation is driven
by occlusion of filter located at top of canister. As
canister fills with fluid, the filter will begin to occlude
as fluid makes continuous contact, creating a
complete blockage in vacuum circuit. A completely
occluded filter will result in canister over-capacity
alarm activation.
• Canister over-capacity alarm will not activate until
filter is completely occluded. Canister orientation,
rate at which fluid enters canister and how exudate
solidifies can impact filter occlusion and canister
over-capacity alarm functionality.
• Canisters should be kept in an upright orientation to
optimize canister over-capacity alarm functionality
while maximizing canister volume. Do not tip over,
orient on its side or turn canister upside down.
• To optimize canister over-capacity alarm functionality,
canister must be changed in the event the canister
contents reach maximum volume indication (250ml
or 800ml fill line). Do not wait for alarm activation
to change the canister.

Audio pause:

• REF 66800059 and 66800697

• REF 66801243

Pressing audio pause button will pause alarm for


approximately 2 – 4 minutes.

If cause of alarm is not resolved during this time, alarm


will recommence sounding.

15
EN
Troubleshooting guide
Symptom Cause Remedy

No AC mains power Electrical power interrupted 1. Check wall outlet.


indication from wall outlet.

Electrical power cord may 1. Check AC mains power cord at device and
be loose. at wall socket.
RENASYS™ EZ (REF
66800059) only. Blown fuses. 1. Disconnect system power and inspect both fuses. If
Not available on fuses need to be replaced, contact your
RENASYS EZ PLUS Smith & Nephew authorized representative.
(REF 66800697,
66801243)

Over vacuum alarm Device has reached high 1. Reset alarm by turning mode of operation switch OFF.
vacuum levels
(of >235mmHg). 2. Replace canister.
3. Inspect connections and tubing to ensure they are
free of obstructions. Ensure there are no kinks in
canister tubing.
4. Turn device ON. If alarm condition continues, device
may have malfunctioned. Contact your
Smith & Nephew authorized representative.
Note: After device stops delivering therapy, due to over
vacuum alarm, pressure in wound will begin to drop. Low
vacuum alarm will also activate.

16
EN

Symptom Cause Remedy

High flow/leak There is a significant leak 1. With device actively creating vacuum, check wound
alarm in the system. This could dressing for air leaks. Look for loose or decompressed
include an air leak around dressing appearance, listen for air movement around
the dressing or a poor seal dressing and feel for areas less compressed or
at one of the connectors cooler in temperature. Address identified leaks with
between the dressing, transparent film or adhesive gel patch.
Soft Port, canister or device.
2. Ensure the following connections are secure:
• Orange quick click connector between Soft Port
and canister tubing
• Canister tubing (blue end) to canister (port on top
of canister)
• Canister in-line bacterial overflow guard to vacuum
port on device
• Y-connector orange quick click connectors, if
present
3. Disconnect Soft Port from canister tubing at orange
quick click connector and insert tethered cap into
both connectors.
• If alarm condition continues, there is a potential
issue with the device or canister. Replace canister
and contact your Smith & Nephew authorized
representative.
• If alarm condition resolves, an air leak is present
within the wound dressing or Soft Port. Reassess
and replace as needed.
Note: An air leak may result in a pressure drop occurring
within the system. As a result, the low vacuum alarm may
also activate while the high flow/leak alarm is active.
Caution – Lack of alarms: Under specific circumstances,
a significant air leak may occur in system without device
activating a high flow/leak alarm. This may be due to
partial blockage between source of air leak and device,
prohibiting detection of the leak by device. As a result,
alarm will not activate. Over time, if blockage reaches
point of full occlusion, the complete blockage alarm will
activate. Read complete blockage/canister over-capacity
alarm troubleshooting for instruction on locating the
system blockage.

17
EN

Symptom Cause Remedy

Low vacuum alarm - Device is unable to achieve 1. With device actively creating vacuum, check wound
vacuum level is lower selected vacuum level due to dressing for air leaks. Look for loose or decompressed
than set point of internal device malfunction dressing appearance, listen for air movement around
NPWT by >15mmHg or a significant leak in the dressing and feel for areas less compressed or
system. cooler in temperature. Address identified leaks with
transparent film or adhesive gel patch.
2. Ensure the following connections are secure:
• Orange quick click connector between Soft Port
and canister tubing
• Canister tubing (blue end) to canister (port on top
of canister)
• Canister in-line bacterial overflow guard to vacuum
port on device
• Y-connector orange quick click connectors, if
present
3. Disconnect Soft Port from canister tubing at orange
quick click connector and insert tethered cap into
both connectors.
• If alarm condition continues, there is a potential
issue with the device or canister. Replace canister
and contact your Smith & Nephew authorized
representative.
• If alarm condition resolves, an air leak is present
within the wound dressing or Soft Port. Reassess
and replace as needed.
Note: If low vacuum alarm is due to a leak in system, the
high flow/leak alarm may also activate while low vacuum
alarm is active.
Note: If over vacuum alarm indicator is illuminated along
with low vacuum indicator, follow troubleshooting steps
for over vacuum alarm.

18
EN

Symptom Cause Remedy

Complete blockage/ There is a complete blockage 1. Check canister. If contents have reached maximum
canister over- in system; this includes a volume indication (250ml or 800ml fill line), replace
capacity alarm canister where contents have canister.
- No negative exceeded maximum volume
pressure at the capacity. 2. Inspect connections, tubing and Soft Port aeration
wound site disc (located on Soft Port near the orange quick click
connector) to ensure they are free of obstructions.
Ensure there are no kinks in canister tubing.
3. Disconnect Soft Port from canister tubing at orange
quick click connector. Insert tethered cap into Soft Port
connector. Allow air to flow freely into canister tubing.
• If alarm condition continues, there is a potential
issue with the device or canister. Replace canister
and contact your Smith & Nephew authorized
representative.
• If alarm condition resolves, a blockage is present
within the Soft Port. Reassess and replace as
needed.
Caution – Lack of alarms: If partial blockage occurs, the
change in pressure status detected by device may not
be significant enough to activate complete blockage/
canister over-capacity alarm. Over time, if blockage
reaches point of full occlusion, the complete blockage/
canister over-capacity alarm will activate.
Blockage formation within wound dressing will not be
detected by the system as occlusion has formed outside
of monitored vacuum circuit. However, it can impact
pressure status at wound site. Appropriately frequent
monitoring of wound dressing is recommended to
confirm adequate delivery of therapy.
In event of heavy or viscous drainage, drainage with
sediment or when blood is present, regular monitoring
and more frequent dressing changes may be required.
If a complete blockage is present in the system but an
air leak occurs between the blockage and device, the
complete blockage/canister over-capacity alarm may
not activate. Ensure all connections are secure and no
leaks are present in system. Potential sources of air leaks
include:
• Tear in Soft Port
• Misplaced or worn O-ring within orange quick click
connector (between the Soft Port and canister tubing)
• Loose blue canister tubing attachment (to port on top
of canister lid)
• Cracked or damaged canister
• Partially inserted in-line bacterial overflow guard

19
EN

Symptom Cause Remedy

When using a Y-connector to connect two Soft Ports


to the same device, a complete blockage alarm will not
activate unless a complete blockage is present in both
Soft Ports. If a partial blockage is present in one or both
Soft Ports, the complete blockage alarm will not activate.
When Y-connecting within a NPWT system, regular
monitoring of both wound dressings is recommended.
Ensure that all dressings are fully compressed and firm
to the touch.
NOTE: For RENASYS F/P or G/P wound dressing kits, the
complete blockage alarm will only activate if a complete
blockage occurs within the canister tubing. There is no
blockage detection in the dressing tubing.

Low battery alarm Battery depleted. Up to 1 hour Plug unit into AC mains power to charge battery.
active therapy remains.

Battery fault alarm Battery/charger fault. If device has been exposed to temperatures outside its
recommended temperature range, let device return to
room temperature. If alarm persists, contact your
Smith & Nephew authorized provider.

20
EN
Maintenance
Inspect device for visible signs of damage before each Device will indicate when battery is low.
use, including in-line bacterial overflow guard and
canister inclusive of tubing. • REF 66800059 and 66800697:
Audible alarm will sound and indicator will flash
Do not use if device has been dropped or shows signs amber.
of damage. Return damaged device to your • REF 66801243:
Smith & Nephew authorized representative in the Audible alarm will sound and indicator will illuminate
original packaging supplied. solid yellow.
As there are no serviceable parts in the device, do not Plug into AC mains power immediately when low
attempt to open device housing. Contact your battery alarm occurs.
Smith & Nephew representative, distributor or an
authorized provider if service is required. Returning device
Prior to returning device to your Smith & Nephew
Cleaning authorized representative, it must be cleaned
Adherence to facility directives concerning device according to steps outlined under cleaning section of
cleaning is of prime importance. The instructions user manual.
supplied with all cleaning agents as well as sterilization
and/or disinfection units must be followed. The device should be returned within the original
packaging.
Cleaning of device outer casing should be done
in accordance with guidelines below: Storage
• Wipe down device surface, including vent on Prior to storage, ensure that the battery is
underside of unit and protection feature of mode of approximately 40 – 60% charged. Some battery
operation switch, with a damp soft cloth. Use a low- discharge may occur in storage.
level cleaning agent or disinfectant and ensure that it
The device should be stored between 41 to 104°F /
is compatible with plastics.
5 to 40°C for optimal battery performance but can
• Dampen another soft cloth with clean water and be stored between 14 to 131°F / -10 to 55°C for short
use to wipe down all surfaces to remove any excess periods of time.
solution.
If the device is stored for longer than 6 months, the
• Dry with a separate soft cloth. battery may need to be charged before it will operate
on battery power. If the device indicates that the battery
• Do not use plastic solvents or abrasives. is still charging after more than 8 hours of continuous
• Do not immerse any part of device in fluid or use a charge, contact your Smith & Nephew distributor or
cloth dripping fluid. No fluids should enter device. If authorized provider.
any liquids penetrate device contact your Caution:
Smith & Nephew distributor or authorized provider.
If device has been stored at temperatures below
freezing, it must be brought to room temperature prior
Battery operation to use or device may be damaged.
The device contains a lithium-ion rechargeable battery
good for approximately 300 to 500 recharges.
Device battery life will vary depending on pressure setting
selected. A fully charged battery will typically deliver
therapy for approximately 30 to 50 hours if pressure
setting is set between 80 to 120mmHg. At a pressure
setting of 200mmHg, a fully charged battery will typically
deliver approximately 18 hours of therapy.
Battery charges when plugged into AC mains power both
during operation and when not in use.
If device is fully charged and is not going to be used
further, unplug power cord from AC mains power and
device AC mains power inlet.

21
EN
Electromagnetic compatibility RENASYS™ EZ (REF 66800059)
This equipment has been tested and found to comply with the limits for medical devices to IEC 60601-1-2-2001. These limits and test
levels are intended to provide reasonable safety with regard to electromagnetic disturbances when the device is used in a typical
medical installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not
occur in a particular installation.
Guidance and manufacturer’s declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidelines

Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or ceramic


discharge (ESD) ±8 kV air ±8 kV air tile. If floors are covered with synthetic
IEC 61000-4-2 material, the relative humidity should be at
least 30%

Electrical fast ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that of a
transient/burst typical commercial or hospital environment
IEC 61000-4-4

Surge ±1 kV differential mode ±1 kV line to line Mains power quality should be that of a
IEC 61000-4-5 ±2 kV common mode ±2 kV line to earth typical commercial or hospital environment

Voltage <5% UT (>95% dip in UT) for 0.5 cycles >95% for 10ms Mains power quality should be that of a
dips, short 40% UT (60% dip in UT) for 5 cycles 60% for 100ms typical commercial or hospital environment.
interruptions 70% UT (30% dip in UT) for 25 cycles 30% for 500ms If the user of the device requires continued
and voltage <5% UT (>95% dip in UT) for 5 sec >95% for 5000ms operation during power mains interruptions,
variations on it is recommended that the device be
power supply powered from an uninterruptable power
input lines supply or battery
IEC 61000-4-11

NOTE UT is the a.c. mains voltage prior to application of the test level

Portable and mobile RF communications


equipment should be used no closer to
any part of device, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
Conducted RF 3 Vrms 150 kHz to 80 MHz 3 Vrms d = 1.2 √P
IEC 61000-4-6
Radiated RF IEC 3 V/m 80 MHz to 2.5 GHz 3 V/m d = 1.2 √P (80 MHz to 800 MHz)
61000-4-3 d = 2.3 √P (800 MHz to 2.5 GHz)
where P is the maximum output power rating
NOTE 1: At 80 MHz, the higher frequency range applies. of the transmitter in watts (W) according
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is to the transmitter manufacturer and d is
affected by absorption and reflection from structures, objects and people.
the recommended separation distance
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) in meters (m). Field strengths from fixed
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV RF transmitters, as determined by an
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic electromagnetic site survey, a should be less
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the device is used exceeds 3V/m, the
than the compliance level in each frequency
device should be observed to verify normal operation. If abnormal performance is observed, rangeb. Interference may occur in
additional measures may be necessary, such as reorienting or relocating the device. the vicinity of equipment marked
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. with the following symbol:

22
EN
Guidance and manufacturer’s declaration – electromagnetic emissions RENASYS™ EZ (REF 66800059)

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidelines
RF emissions CISPR 11 Group 1 The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The device is suitable for use in all establishments other
than domestic and those directly connected to the public
Harmonic emissions IEC 61000-3-2 Not applicable low-voltage power supply network that supplies buildings
Voltage fluctuations/flicker emissions Comply used for domestic purposes.
IEC 61000-3-2

WARNING: The device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use
is necessary, the device should be observed to verify normal operation in the configuration in which it will be used.

Recommended separation distances between portable and mobile RF communications equipment and the device

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output Separation distance according to frequency of transmitter (m):
power of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
(W) d = 1.2√P d = 1.2√P d = 2.3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1.0 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum power rating of the transmitter
in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.

23
EN
Electromagnetic compatibility RENASYS EZ PLUS (REF 66800697, 66801243)
This equipment has been tested and found to comply with the limits for medical devices to IEC 60601-1-2-2001. These limits and test
levels are intended to provide reasonable safety with regard to electromagnetic disturbances when the device is used in a typical
medical installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not
occur in a particular installation.
Guidance and manufacturer’s declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidelines

Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or ceramic


discharge (ESD) ±8 kV air ±8 kV air tile. If floors are covered with synthetic
IEC 61000-4-2 material, the relative humidity should be at
least 30%

Electrical fast ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that of a
transient/burst typical commercial or hospital environment
IEC 61000-4-4

Surge ±1 kV differential mode ±1 kV line to line Mains power quality should be that of a
IEC 61000-4-5 ±2 kV common mode ±2 kV line to earth typical commercial or hospital environment

Voltage <5% UT (>95% dip in UT) for 0.5 cycles >95% for 10ms Mains power quality should be that of a
dips, short 40% UT (60% dip in UT) for 5 cycles 60% for 100ms typical commercial or hospital environment.
interruptions 70% UT (30% dip in UT) for 25 cycles 30% for 500ms If the user of the device requires continued
and voltage <5% UT (>95% dip in UT) for 5 sec >95% for 5000ms operation during power mains interruptions,
variations on it is recommended that the device be
power supply powered from an uninterruptable power
input lines supply or battery
IEC 61000-4-11

NOTE UT is the a.c. mains voltage prior to application of the test level

Portable and mobile RF communications


equipment should be used no closer to
any part of device, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
Conducted RF 3 Vrms 150 kHz to 80 MHz 3 Vrms d = 1.2 √P
IEC 61000-4-6
Radiated RF IEC 3 V/m 80 MHz to 2.5 GHz 3 V/m d = 1.2 √P (80 MHz to 800 MHz)
61000-4-3 d = 2.3 √P (800 MHz to 2.5 GHz)
where P is the maximum output power rating
NOTE 1: At 80 MHz, the higher frequency range applies. of the transmitter in watts (W) according
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
to the transmitter manufacturer and d is
affected by absorption and reflection from structures, objects and people.
the recommended separation distance
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) in meters (m). Field strengths from fixed
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV RF transmitters, as determined by an
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
electromagnetic site survey, a should be less
If the measured field strength in the location in which the device is used exceeds 3V/m, the than the compliance level in each frequency
device should be observed to verify normal operation. If abnormal performance is observed, rangeb. Interference may occur in
additional measures may be necessary, such as reorienting or relocating the device. the vicinity of equipment marked
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. with the following symbol:

24
EN
Guidance and manufacturer’s declaration – electromagnetic emissions RENASYS EZ PLUS (REF 66800697, 66801243)

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidelines
RF emissions CISPR 11 Group 1 The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The device is suitable for use in all establishments other
than domestic and those directly connected to the public
Harmonic emissions IEC 61000-3-2 Not applicable low-voltage power supply network that supplies buildings
Voltage fluctuations/flicker emissions Comply used for domestic purposes.
IEC 61000-3-2

WARNING: The device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use
is necessary, the device should be observed to verify normal operation in the configuration in which it will be used.

Recommended separation distances between portable and mobile RF communications equipment and the device

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output Separation distance according to frequency of transmitter (m):
power of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
(W) d = 1.2√P d = 1.2√P d = 2.3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1.0 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum power rating of the transmitter
in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.

25
EN
Caution statements Essential Performance
In order to ensure safe and proper performance, the (IEC 60601-1:2005, 3rd Edition)
following conditions must be met: Essential performance of the RENASYS™ EZ PLUS
• All assembly, operation, adjustment, maintenance (REF 66801423), for safe operation, is to maintain
and/or repair should be carried out by qualified vacuum delivered by device within its specification for
personnel authorized by Smith & Nephew. pressure selected, to provide Negative Pressure Wound
Therapy (NPWT).
• Any modification of this equipment is prohibited.
• To avoid risk of electric shock, equipment must only
be connected to a supply mains with protective
earth.
• If device is damaged, performance could be affected;
do not use device. Contact your Smith & Nephew
authorized representative.
• Use only AC mains power cord provided with device
to prevent the potential for electrical shock hazard.
• If power cord is damaged, wires are frayed or
exposed, do not use power cord.
• When necessary, device may be isolated from AC
mains by removing detachable AC mains power
cord.
• Electrical installation of the room must comply with
appropriate electrical wiring standards.
• Product must be used in accordance with this user
manual and all applicable labeling.
Failure to comply with these conditions will void any
pertinent warranties.
This user manual is not intended as a guarantee or
warranty. It is intended only as a guide. For medical
questions, please consult a physician. For additional
product information, or a specific product question,
please refer to the numbers listed in the customer
assistance section of this user manual.

26
EN
Specifications
Pressure
Setting 40, 50, 60, 70, 80,90, 100, 120, 140, 160, 180, 200mmHg:
tolerance +/-10mmHg

Display (gauge) 0–200mmHg; tolerance +/-5mmHg


Maximum 200mmHg
Alarms
General (all alarms)
Priority Low
Auditory sound level 53 dB
Over vacuum
Indicator color REF 66800059, 66800697 Amber; REF 66801243 Yellow
Alarm overall delay Mean: <1 second, Stand. Dev. : <1 second
High flow/leak
Indicator color REF 66800059, 66800697 Amber; REF 66801243 Yellow
Alarm condition delay Mean: <1 second, Stand. Dev. : <1 second
Alarm overall delay Mean: 68 seconds, Stand. Dev. : 6 seconds
Low vacuum
Indicator color REF 66800059, 66800697 Amber; REF 66801243 Yellow
Alarm overall delay Mean: <1 second, Stand. Dev. : <1 second
Complete blockage/canister
over-capacity
Indicator color REF 66800059, 66800697 Amber; REF 66801243 Yellow
Alarm condition delay Mean: <1 second, Stand. Dev. : <1 second
Alarm overall delay Mean: 81 seconds, Stand. Dev. : 3 seconds
Low battery
Indicator color REF 66800059, 66800697 Amber; REF 66801243 Yellow
Alarm overall delay Mean: <1 second, Stand. Dev. : <1 second
Battery fault
Indicator color REF 66800059, 66800697 Amber; REF 66801243 Cyan (light blue)
Alarm overall delay Mean: <1 second, Stand. Dev. : <1 second

27
EN
AC mains power
Supply voltage 100–240 VAC
Supply frequency 50 / 60 Hz
Input power 90 VA
Fuses (replaceable) 2 each, Dual quick blow fuses 3.15 A / 250 VAC
Physical
Dimensions 14.5 x 9.5 x 7 in. / 361 x 240 x 170 mm
Weight 8.14 lbs / 3.7 kg
Battery
Operating time ~ 30–50 hours (therapy) when operating from 80mmHg to 120mmHg;
~ 18 hours (therapy) at 200mmHg
Type Lithium ion
Safety Protection
Earth protection Class I
Patient protection Type BF
Ingress protection IP2X
Environmental
Storage and transport 14 to 131°F / -10 to 55°C
Operational temperature 41 to 104°F / 5 to 40°C
Relative humidity 30% to 70% RH
Atmospheric pressure 700 mbar to 1,060 mbar
Compliance
UL 60601-1
IEC 60601-1
IEC 60601-1-2
CAN/CSA C22.2

28
EN
Glossary of symbols Over vacuum When system encounters
excessively high vacuum (of >235mmHg),
Operation switch device will stop delivering NPWT. In order
to restart therapy, mode of operation switch
ON continuous therapy Device will must be turned OFF and back ON.
maintain pressure setting without stopping
REF 66800059, 66800697: Audible alarm
until switched off.
and indicator will flash amber.
OFF Device is not delivering NPWT. REF 66801243: Audible alarm and indicator
will illuminate solid yellow.
ON intermittent therapy High flow/leak When system detects a
RENASYS™ EZ – REF 66800059: Device significant leak:
cycles ON and OFF in increments of REF 66800059, 66800697: Audible alarm
approximately 32 seconds ON (active and indicator will flash amber.
therapy) and 16 seconds OFF (no vacuum). REF 66801243: Audible alarm and indicator
RENASYS EZ PLUS – REF 66800697, will illuminate solid yellow.
66801243: Device cycles ON and OFF in Low vacuum If vacuum level is lower than
increments of approximately 5 minutes set point of therapy by >15mmHg:
ON (active therapy) and approximately 2
REF 66800059, 66800697: Audible alarm
minutes OFF (no vacuum).
and indicator will flash amber.
REF 66801243: Audible alarm and indicator
will illuminate solid yellow.
Indicators
Complete blockage/canister over-
AC mains power (Feature available only on capacity When system detects that
RENASYS EZ – REF 66800059) When system canister has exceeded maximum volume
is connected to AC mains power, indicator capacity or that there is a complete
will illuminate solid green; does not indicate blockage in the system. To optimize
device is turned on. canister over-capacity alarm functionality,
the canister must be changed in the event
ON/OFF (Feature available only on the canister contents reach maximum
RENASYS EZ PLUS – REF 66800697, volume indicaiton (250ml or 800ml fill line).
66801243) When mode of operation switch Do not wait for alarm activation to
is ON in continuous or intermittent mode, change canister.
the indicator will illuminate solid green. REF 66800059, 66800697: Audible alarm
and indicator will flash amber.
Battery indicator
REF 66801243: Audible alarm and indicator
• Battery full: solid green indicator. will illuminate solid yellow.
• Battery charging: flashing green indicator.
Audio pause Pressing audio pause button
• Battery low: will pause alarm for approximately
REF 66800059, 66800697: Audible 2 – 4 minutes. The indicator will illuminate
alarm and indicator will flash amber. REF
solid yellow.
REF 66801243: Audible alarm and 66800059,
indicator will illuminate solid yellow. 66800697
• Battery fault (Feature available only
on RENASYS EZ PLUS – REF 66800697,
66801243):
REF 66800697: Indicator will illuminate REF
solid amber. 66801243
REF 66801243: Indicator will illuminate
solid cyan (light blue).

29
EN

Equipment classification
Keep dry European representative
Isolation type BF applied part

CSA international
Storage temperature Batch code
classification

EU:
Date of manufacture Serial number
Not for general waste

Refer to instruction Product catalogue


Manufacturer
manual/booklet number

Caution: Earth
CE mark
Follow operating instructions (ground)

Do not use if package is


damaged
Fuse Keep upright

Non-ionizing
Biological risk Alternating current
electromagnetic radiation

Single use Caution: U.S. Federal


Lock position
Do not reuse
law restricts this device
Patient to sale by or on the
Patient Unlock position order of a physician
number

MR MRI Unsafe

30