Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

INSTALLATION - 8

[File: E3014_Rev1.doc]
Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

SUMMARY
UNPACKING 2
GROUPS ASSEMBLY: IMAGE INTENSIFIER TUBE AND COLLIMATOR ................................................. 4
INTENSIFIER GROUP ASSEMBLY ............................................................................................................. 5
MONOBLOC GROUP ASSEMBLY............................................................................................................. 10
SETUP 14
POWER SUPPLY ....................................................................................................................................... 14
115V power supply ................................................................................................................................ 14
230V power supply ................................................................................................................................ 14
TUBE SETTING IN ..................................................................................................................................... 15

[File: E3014_Rev1.doc] Pag. 1/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

UNPACKING
To unpack the unit, please refer to picture 1 (page 4), the letters indicate the parts of the packing that have to
been disposed of, as they are not component parts of the system, but assure a safe transportation from the
factory to the destination site; the numbers indicate the component parts of the system.

The packing is composed by :

Qty. Description Material code


1. N°1 monitor trolley *
2. N°2 monitor *
3. N°1 mobile image intensifier *
4. N°1 x-ray emission lamp *
5. N°1 x-ray foot switch 62308A
6. N°1 sterilizable drapes set *
7. N°1 single extended board 01605
double extended board 01500A
8. N°1 English user’s manual *
Italian user’s manual *
English service manual *
Italian service manual *
9. N°1 cassette holder *
10. N°1 paint kit 62313A

*: the code for these parts depends on the unit’s configuration.

[File: E3014_Rev1.doc] Pag. 2/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

Picture 1

[File: E3014_Rev1.doc] Pag. 3/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

GROUPS ASSEMBLY: IMAGE INTENSIFIER TUBE AND COLLIMATOR


To perform this operation correctly, act as follow:

Verify that each handle and foot brake is well locked.

ATTENTION, COLLISION DANGER


Never loosen handles during the following operation. In case, get ready to support
any kind of weight, as the C arm could reach a speed dangerous for persons, and
harmful for components.

Loosen the central handle very carefully, while


supporting the arm. Move the arm as show in the
picture, and lock the handle properly.
Withdraw the arm holder totally, in order to have the
barycentre as safe and withdrawn as possible.

[File: E3014_Rev1.doc] Pag. 4/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

INTENSIFIER GROUP ASSEMBLY


WARNING
Check again that each handle is well locked, and the
pedal brake well positioned

Take the intensifier group out from the box carefully.


Position it on a table, protecting it with a cloth and, in
order to withdraw the upper cope (cone stub) at the
top. Clean it from package residues

Take the cope (cone stub) out, by unscrewing the


washer. Take the covercables sheet out from the
intensifier group support, by unscrewing the four
screws.

[File: E3014_Rev1.doc] Pag. 5/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

Take the covercables sheet out from the intensifier


group support, by unscrewing the four screws.

Take the six thumbscrews out from the intensifier


group base.

Keep them carefully close, as they are necessary for


the next operation.

[File: E3014_Rev1.doc] Pag. 6/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

WARNING The following operation needs a two technicians intervention. Being a


quite complex phase it has to be read and comprehended carefully.

1. Approach the thumbscrews well.


2. Grip the thumbscrews well.
3. Let the cables pass through the plate hole.

Insert the plastic plugs which cover the hexagonal wrenches.

Insert the connector with the red gate in correspondence of


the red gate of the tube intensifier.
After passing the CCD camera connector through the hole, let
it carefully pass between U bolt lower space and CCD camera.

[File: E3014_Rev1.doc] Pag. 7/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

Connect the two CCD camera connectors.


Fix the CCD camera connectors with the screws previously
positioned.

By means of a lever washer, a flexible washer and a nut,


ensure the cable lead on the threaded pivot of the intensifier
group.

Fix the cables by two bands.

Assembly the little case, and fix it by the four lever washers
and screws as shown in the picture.

[File: E3014_Rev1.doc] Pag. 8/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

Re-assembly the cover cable sheet and lock the four


thumbscrews well.

Assembly the cope by means of a washer and the screw


previously taken off.

[File: E3014_Rev1.doc] Pag. 9/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

MONOBLOC GROUP ASSEMBLY


WARNING Before assembling the monobloc group,
make sure that each screw of the intensifier group is
well locked, and that handles are in block position.

Take the collimator-monobloc group out from the


box very carefully.

Position it carefully on a table, protecting it with a


cloth, and take package residues off.

Take the black cope which covers the collimator off,


by unscrewing the three screws.

[File: E3014_Rev1.doc] Pag. 10/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

WARNING The following operation needs a two technicians intervention. Being


a complex operation it has to be carefully read and well comprehended before
its performing.

This operation needs a two technicians intervention. One has to


support the collimator-monobloc group near its own location, the
second has to let the cables pass one by one very carefully.
Once all the cables have passed, place the collimator-monobloc
group, on the supporting plate.

Fix the group by the three hexagonal wrenches, lever washers


and flexible washers. The hexagonal wrenches have not to be
completely locked, as they are necessary for the aligning
adjustment with the intensifier and collimator-monobloc, as
described in the following paragraphs.

[File: E3014_Rev1.doc] Pag. 11/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

Connect the cables to the collimator-monobloc group.


(See Chapter Electrical Drawing for the monobloc topographic
board)
Once this operation has been completed, block cables by using
bands.

Insert the connector of the collimator, and fix the collimator


cable to the group by using belts.

Re-assemble the black cope of the collimator, by using the three


screws as shown in the picture.

[File: E3014_Rev1.doc] Pag. 12/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

ATTENTION DANGER OF X-RAY RADIATION


Before x-ray exposure, activate each x-ray protection

Connect the unit to the mains.


(See User’s Manual)

Switch the unit on and set 40KV – 0.2mA in manual fluoroscopy.


(See User’s Manual)
1. Control fluoroscopy.
2. Close the iris diaphragm and parallel shutters of the collimator..
3. The situation shown in the central picture has to be displayed on the monitor.

In case the images are different from


the ones in the picture, it’s necessary
to adjust the collimator-monobloc
group, operating on its own fixing
screws, and its own fixing screws, and
repeat the test twice.

Once the image is proper, make sure


that the monobloc-collimator group
screws are well locked.

Assemble the covering cases on the


monobloc-collimator group using the
three screws as described in the
picture.

[File: E3014_Rev1.doc] Pag. 13/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

SETUP
POWER SUPPLY
Before inserting the supply plug into the mains, perform the following checks:
• verify that the plug is supplied with the earth connection:
• verify that the mains voltage corresponds to the one of the unit.

IF ALL THESE CONDITIONS ARE NOT VERIFIED AND NOT CERTAIN, DO NOT CONNECT THE
SUPPLY PLUG TO THE MAINS.

During the purchase order procedure, it is already specified whether the unit has to operate either at 230V or
115V, therefore, no modification is required during the installation phase. For any further verification, please
refer to the electrical drawings and to the set up tables.

ATTENTION: For any check, the intervention of qualified and authorized personnel is
required.

115V power supply


Refer to electrical drawings

device drawing pos. set


auto-transformer TR1 MONITOR TROLLEY / ACTIVATION A3 TB2-2 clamp connected to VTC1 wire
(monitor trolley) CIRCUITS TB2-3 clamp not connected
jumpers JP3, JP2, JP1 MONITOR TROLLEY / ACTIVATION B4 JP1 open
(monitor trolley ) CIRCUITS JP2 closed
JP3 closed
fuse F1 MONITOR TROLLEY / ACTIVATION B3 1A fuse
(monitor trolley) CIRCUITS
transformer TR1 DC SUPPLY / WATCH-DOG B31 TR1-5 clamp connected to power supply
(mobile image TR1-6 clamp not connected
intensifier )
transformer TR2 SUPPLY FILAMENT A91 TR2-5 clamp connected to power supply
(mobile Image TR2-6 clamp not connected
intensifier)

230V power supply


Refer to electrical drawings

device drawing pos. set


auto-transformer TR1 MONITOR TROLLEY / ACTIVATION A3 TB2-3 clamp connected to VTC1wire
(monitor trolley ) CIRCUITS TB2-2 clamp not connected
jumpers JP3, JP2, JP1 MONITOR TROLLEY / ACTIVATION B4 JP1 closed
(monitor trolley) CIRCUITS JP2 open
JP3 open
fuse F1 MONITOR TROLLEY / ACTIVATION B3 500mA fuse
(monitor trolley) CIRCUIT
transformer TR1 DC SUPPLY / WATCH-DOG B31 TR1-6 clamp connected to power supply
(mobile image TR1-5 clamp not connected
intensifier )
transformer TR2 SUPPLY FILAMENT A91 TR2-6 clamp connected to power supply
(mobile image TR2-5 clamp not connected
intensifier )

[File: E3014_Rev1.doc] Pag. 14/15


Villa Sistemi Medicali ARCOVIS 2000 R - Installation - 8

TUBE SETTING IN
It’s necessary to proceed with the X-RAY TUBE SETTING IN before using the unit for the very first time and
after any long period of non-use (3 months or more).

To warm up, carefully follow the following procedure:


1. Set the unit in manual fluoroscopy (see users’ manual)
2. Perform the exposures described here below , in the sequence and time frames indicated in the table.

Functioning parameters exposure time temporal


frequency
kV mA ON time OFF time
60 3 2.5' 2.5' 5 times
80 2 2.5' 2.5' 3 times
100 1 1' 1' 2 times

ATTENTION:
If during this procedure, any discrepancies or anomalies are detected, it’s necessary
to interrupt it for at least half an hour and start again from the beginning.

[File: E3014_Rev1.doc] Pag. 15/15