Models TXR 325D, TXR325M and TXR525SFQ MIDWEST SUPER SPEED BUCKY
(Terminal Strip Identifies Table or Wall) (Terminal Strip)
B1 (Blue) B1 (Blue)
B2 (Orange) B2 (Orange)
No wire is required from the x-ray control terminal, but a
Jumper is required Between B2 and B3 on the Midwest
B3 Bucky Terminal.
B4 (White) B4 (White)
B8 (Brown) B8 (Brown)
Earth Ground (Green) Earth Ground (Green)
NOTICE: Connecting Midwest Super Speed Buckys to TXR X-Ray Controls require four wire conductors. An external
Earth Ground is required for Patient Safety.
NOTICE: DO NOT CONNECT PAR SPEED BUCKYS TO HIGH FREQUENCY X-RAY CONTROLS – SERIOUS
DAMAGE TO CONTROL CIRCUITS WILL OCCUR
NOTICE ALL BUCKYS REQUIRE - BIH Harness (35 ft. long 7 Conductor) – Sold Separately.