TXR Tingle X-Ray Products, Inc.


Technical Pamphlet
M-2008-3-3
BUCKY INTERFACE (BIH)
TO
SHF SERIES HIGH FREQUENCY X-RAY CONTROLS
NOTICE: DO NOT CONNECT PAR SPEED BUCKYS TO HIGH FREQUENCY X-RAY CONTROLS – SERIOUS
DAMAGE TO CONTROL CIRCUITS WILL OCCUR

PROGENCY BUCKY INTERFACE (TABLE)


HIGH FREQUENCY MODULE
(Terminal Strip 3TS1 is located on the outside of the
swing out door by the top hinge ) PROGENCY TRUE SPEED BUCKY (Terminal Strip)
Connect a Jumper between 3TS1-2 and 3TS1-3
Connect a Jumper between 3TS1-1 and 3TS1-27
3TS1-5 (Blue) 1 (Blue)
3TS1-6 (Orange) 2 (Orange)
3TS1-4 (White) 3 (White)
3TS1-3 (Brown) L (Brown)
3TS1-1 (Green) N (Green)
Earth Ground Earth Ground
In case of noise due to Bucky: Add R2=22 ohm, ½ watt,
NOTICE: Connect Black and Red to 24 VAC on top of 5%; and C2=470nF, 250 VAC . Connect between #3 and
the Module N on the Bucky. See manual for additional information.

PROGENCY BUCKY INTERFACE (WALLSTAND)


HIGH FREQUENCY MODULE
(Terminal Strip 3TS1 is located on the outside of the
swing out door by the top hinge ) PROGENCY TRUE SPEED BUCKY (Terminal Strip)
Connect a Jumper between 3TS1-9 and 3TS1-10
Connect a Jumper between 3TS1-8 and 3TS1-27
3TS1-12 (Blue) 1 (Blue)
3TS1-13 (Orange) 2 (Orange)
3TS1-11 (White) 3 (White)
3TS1-10 (Brown) L (Brown)
3TS1-8 (Green) N (Green)
Earth Ground Earth Ground
In case of noise due to Bucky: Add R2=22 ohm, ½ watt,
NOTICE: Connect Black and Red to 24 VAC on top of 5%; and C2=470nF, 250 VAC . Connect between #3 and
the Module N on the Bucky. See manual for additional information.

IMAGO BUCKY (24VAC) INTERFACE (TABLE)


HIGH FREQUENCY MODULE IMAGO TABLE TERMINALS
(Terminal Strip 3TS-1 is located on the outside of the Terminal Strip inside Table Base (The numbers in
swing out door by the top hinge ) parenthesis ( ) identify the connections on the Bucky
Connect a Jumper from 3TS1-2 to TB7-4 NOTICE; This table covers Imago Table Connections
(TB7 is located on the SHF power module lock board ) Only.
3TS1-5 CS1-6 (B1)
3TS1-6 CS1-5 (B2)
3TS1-4 CS1-4 (B4)
CS1-3 (B5)
TB7-3
Earth Ground B3 and B6 is Earth Ground (Connect to Earth Ground).
B7 and B8 is provided internally from the table base
power supply. No external wiring is necessary.
NOTICE ALL BUCKYS REQUIRE - BIH Harness (35 ft. long 7 Conductor) – Sold Separately.
2

BUCKY INTERFACE (BIH)


TO
LINE FREQUENCY X-RAY CONTROLS

Models TXR 325D, TXR325M and TXR525SFQ


(Terminal Strip Identifies Table or Wall) PROGENCY TRUE SPEED BUCKY (Terminal Strip)
B1 (Blue) 1 (Blue)
B2 (Orange) 2 (Orange)
No wire is required from the x-ray control terminal, but a
Jumper is required Between 2 and 3 on the Progency
B3 Bucky Terminal.
B4 (White) N (White)
B8 (Brown) L (Brown)
Earth Ground (Green) Earth Ground (Green)
NOTICE: Connecting True Speed Buckys to TXR X-Ray Controls require four wire conductors. An external Earth
Ground is required for Patient Safety.

Models TXR 325D, TXR325M and TXR525SFQ MIDWEST SUPER SPEED BUCKY
(Terminal Strip Identifies Table or Wall) (Terminal Strip)
B1 (Blue) B1 (Blue)
B2 (Orange) B2 (Orange)
No wire is required from the x-ray control terminal, but a
Jumper is required Between B2 and B3 on the Midwest
B3 Bucky Terminal.
B4 (White) B4 (White)
B8 (Brown) B8 (Brown)
Earth Ground (Green) Earth Ground (Green)
NOTICE: Connecting Midwest Super Speed Buckys to TXR X-Ray Controls require four wire conductors. An external
Earth Ground is required for Patient Safety.

IMAGO TABLE BUCKY (24VAC) INTERFACE TO SINGLE PHASE

NOTICE: REQUIRES A 120/24 VAC 75 VA DOUBLE INSULATED STEP DOWN TRANSFORMER


IMAGO TABLE TERMINALS
Models TXR 325D, TXR325M and TXR525SFQ Terminal Strip inside Table Base (The numbers in
(Terminal Strip Identifies Table or Wall) parenthesis ( ) identify the connections on the Bucky
NOTICE; This table covers Imago Table Connections
Only.
B1 CS1-5
B2 CS1-6
B3 Connects to one side of 120 VAC input of the step CS1-4 Connects to one side of the 24 VAC output of the
down transformer step down transformer
CS1-3 Connects to the other side of the 24 VAC output of
B4 Connects to other side of the 120 VAC input of the the step down transformer
step down transformer
Earth Ground B3 and B6 is Earth Ground (Connect to Earth Ground).
B7 and B8 are provided internally from the table base
power supply. No external wiring is necessary.
NOTICE ALL BUCKYS REQUIRE - BIH Harness (35 ft. long 7 Conductor) – Sold Separately.
3

NOTICE: DO NOT CONNECT PAR SPEED BUCKYS TO HIGH FREQUENCY X-RAY CONTROLS – SERIOUS
DAMAGE TO CONTROL CIRCUITS WILL OCCUR

Models TXR 325D, TXR325M and


TXR525SFQ MIDWEST PAR SPEED BUCKY
(Terminal Strip) (Terminal Strip)
B1 B1
B2 B2
Place a Jumper Between B2 and B3 in the X- Connect B3 to A2 in the X-ray Control
Ray Control
B4 B4
Earth Ground Earth Ground
NOTICE: Par Speed Buckys to TXR X-Ray Controls require four wire conductors. An external Earth Ground is
required for Patient Safety.

NOTICE ALL BUCKYS REQUIRE - BIH Harness (35 ft. long 7 Conductor) – Sold Separately.