You are on page 1of 1

Date 29/11/2021

Operator ID SYUKUR Equipment Number KUL001/001261


Calibration Tech NABILAH Serial Number
Calibration Date 10/02/2021 Manufacturer Nihon Kohden
Firmware Version 3.01.03 Model TEC 5631K
Serial Number 3903559 Location ICU
Date & Time 2021/11/29 & 02:40pm Other PPM
JOB Name

Template Name DEFIBRILATOR Standard IEC62353-Direct


Pause after Power ON NO Pause before Power OFF NO
Power ON delay 2 Power OFF delay 0
Test Speed NORMAL Test Mode AUTO
Halt on Test Failure YES Multi PE Test NO
Include Time YES Multi Resstore WORST/LAST
Insulation Resistance Voltage 500V Reverse Polarity YES
Multi Enclosure Test NO Classification I

PADDLE CF 2
ECG LEAD CF 3

Protective Earth Resistance 0.151 Ohm 0.3 - P


Insulation Resistance P
Mains to Protective Earth 999 MOhm - - P
Mains to Applied Parts 999 MOhm - - P
Applied Parts to Non-Earth Accessible Conductive Part 999 MOhm - - P
Mains Voltage P
Live to Neutral 240.8 V - - P
Live to Earth 242.3 V - - P
Neutral to Earth 0.3 V - - P
Equipment Current 0.9 A - - P
Direct Applied Part Leakage P
Normal Condition P
PADDLE 6.2 uAAC 50 - P
ECG LEAD 6.2 uAAC 50 - P
Normal Condition-Reversed mains P
PADDLE 6.2 uAAC 50 - P
ECG LEAD 6.2 uAAC 50 - P
Direct Equipment Leakage P
Open Earth 70.5 uA 500 - P
Open Earth-Reversed Mains 94.6 uA 500 - P

_________
Signature