You are on page 1of 1

Date 17/11/2021

Operator ID SYUKUR Equipment Number KUL001/001041


Calibration Tech NABILAH Serial Number
Calibration Date 10/02/2021 Manufacturer Ezcare Medical
Firmware Version 3.01.03 Model Mediaire Deluxe 3
Serial Number 3903559 Location ICU
Date & Time 2021/11/17 & 12:33pm Other PPM
JOB Name

Template Name MATTRESS SYSTEMS Standard IEC62353-Direct


Pause after Power ON NO Pause before Power OFF NO
Power ON delay 2 Power OFF delay 0
Test Speed NORMAL Test Mode AUTO
Halt on Test Failure YES Multi PE Test NO
Include Time YES Multi Resstore WORST/LAST
Insulation Resistance Voltage 500V Reverse Polarity YES
Multi Enclosure Test NO Classification II

MATTRESS BF 1

Insulation Resistance
Mains to Applied Parts 999 MOhm - - P
P
P
Mains Voltage
- P
Live to Neutral 240.4 V -
P
Equipment Current 0.6 A - - P
Direct Applied Part Leakage
Normal Condition
MATTRESS 0.6 uAAC 5000 - P
Normal Condition-Reversed mains P
MATTRESS 0.6 uAAC 5000 - P
Direct Equipment Leakage P
Open Earth 70.5 uA 500 - P
Open Earth-Reversed Mains 94.6 uA 500 - P