This document summarizes the results of an electrical safety test performed on equipment with identification number E01215-HCB. The test was performed on April 22, 2022 by calibration technician 27992 and included measurements of protective earth resistance, insulation resistance, mains voltage, equipment current, earth leakage current, and enclosure leakage current. All test results were within specified limits and passed the test.
This document summarizes the results of an electrical safety test performed on equipment with identification number E01215-HCB. The test was performed on April 22, 2022 by calibration technician 27992 and included measurements of protective earth resistance, insulation resistance, mains voltage, equipment current, earth leakage current, and enclosure leakage current. All test results were within specified limits and passed the test.
This document summarizes the results of an electrical safety test performed on equipment with identification number E01215-HCB. The test was performed on April 22, 2022 by calibration technician 27992 and included measurements of protective earth resistance, insulation resistance, mains voltage, equipment current, earth leakage current, and enclosure leakage current. All test results were within specified limits and passed the test.
Calibration Tech 27992 Serial Number Calibration Date 04/22/2022 Manufacturer Firmware Version 3.01.03 Model Serial Number 5775743 Location Date & Time 2022/10/19 & 01:49am Other JOB Name
Template Name RESPIRADOR Standard IEC60601-1-3rd Ed
Pause after Power ON NO Pause before Power OFF NO Power ON delay 2 Power OFF delay 0 Test Speed RAPID Test Mode AUTO Halt on Test Failure YES Multi PE Test NO Include Time YES Multi Resstore WORST/LAST Insulation Resistance Voltage 500V Reverse Polarity YES Multi Enclosure Test NO Classification I
Protective Earth Resistance 0.196 Ohm 0.2 - P
Insulation Resistance P Mains to Protective Earth 999 MOhm - - P Mains to Non-Earth Accessible Conductive Part 999 MOhm - - P Mains Voltage P Live to Neutral 232.7 V - - P Neutral to Earth 1.6 V - - P Live to Earth 234.3 V - - P Equipment Current 0.2 A - - P Earth Leakage Current P Normal Condition 224 uA 5000 - P Open Neutral 443 uA 10000 - P Open Neutral-Reversed Mains 448 uA 10000 - P Normal Condition-Reversed mains 222 uA 5000 - P Enclosure Leakage Current P Normal Condition 0.2 uA 100 - P Open Earth 224 uA 500 - P Open Neutral 0.2 uA 500 - P Open Neutral-Reversed Mains 0.2 uA 500 - P Normal Condition-Reversed mains 0.2 uA 100 - P Open Earth-Reversed Mains 223 uA 500 - P