You are on page 1of 16

Customer Name ST. JOSEPH HOSPITAL Certificate No.

SJHD008
Date of Calibration 20-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 19-10-2022
CUSTOMER Address
624001
Date of Issue 21-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 200331091100104
Make/Model No. YONKER/YK-8000C Location NICU
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.5 1 0 MΩ P
  5.LEAD 5 0.4 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.5 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.4 1 0 MΩ P
  5.LEAD 5 0.1 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD029
Date of Calibration 20-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 19-10-2022
CUSTOMER Address
624001
Date of Issue 21-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. CN92381717
Make/Model No. PHILIPS/EFFICIA CM12 Location CCU
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.5 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.2 1 0 MΩ P
  4.LEAD 4 0.6 1 0 MΩ P
  5.LEAD 5 0.1 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.6 1 0 MΩ P
  2.LEAD 2 0.5 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.1 1 0 MΩ P
  5.LEAD 5 0.6 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD030
Date of Calibration 20-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 19-10-2022
CUSTOMER Address
624001
Date of Issue 21-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 19SNP09012J2065
Make/Model No. KO-MED/M12 Location CCU
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229.3 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 229.2 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.4 3 0 V P
4 EQUIPMENT CURRENT 0.4 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.4 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.7 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.6 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.5 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.4 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.6 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.5 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.2 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.6 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.2 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD031
Date of Calibration 20-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 19-10-2022
CUSTOMER Address
624001
Date of Issue 21-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 19SNP09012J2131
Make/Model No. KO-MED/M12 Location CCU
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229.4 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.4 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.2 3 0 V P
4 EQUIPMENT CURRENT 0.3 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.7 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.3 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.5 1 0 MΩ P
  5.LEAD 5 0.4 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.5 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.4 1 0 MΩ P
  5.LEAD 5 0.1 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD032
Date of Calibration 20-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 19-10-2022
CUSTOMER Address
624001
Date of Issue 21-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 19SNP09012J2063
Make/Model No. KO-MED/M12 Location CCU
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.2 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.4 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.4 1 0 MΩ P
  4.LEAD 4 0.2 1 0 MΩ P
  5.LEAD 5 0.2 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD049
Date of Calibration 20-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 19-10-2022
CUSTOMER Address
624001
Date of Issue 21-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. DE67IH3864
Make/Model No. PHLIPS Location CCU
Asset Barcode No. SJH/BME/MUT/14 Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229.4 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.4 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.2 3 0 V P
4 EQUIPMENT CURRENT 0.3 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.7 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.3 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.5 1 0 MΩ P
  5.LEAD 5 0.4 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.5 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.4 1 0 MΩ P
  5.LEAD 5 0.1 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD111
Date of Calibration 21-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 20-10-2022
CUSTOMER Address
624001
Date of Issue 22-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. CN12001403
Make/Model No. PHILIPS/EFFICIA CM12 Location OT-1
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.2 1 0 MΩ P
  2.LEAD 2 0.1 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD115
Date of Calibration 21-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 20-10-2022
CUSTOMER Address
624001
Date of Issue 22-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. US12556126
Make/Model No. PHILIPS/SURESIGN VM8 Location OT-2
Asset Barcode No. SJH/BME/MOT/07 Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229.4 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.4 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.2 3 0 V P
4 EQUIPMENT CURRENT 0.3 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.7 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.3 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.6 1 0 MΩ P
  4.LEAD 4 0.5 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.5 1 0 MΩ P
  3.LEAD 3 0.4 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD119
Date of Calibration 21-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 20-10-2022
CUSTOMER Address
624001
Date of Issue 22-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. US12556167
Make/Model No. PHILIPS/SURESIGN VM8 Location OT-3
Asset Barcode No. SJH/BME/MOT/08 Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.4 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.4 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.7 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.6 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.3 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.2 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.5 1 0 MΩ P
  4.LEAD 4 0.6 1 0 MΩ P
  5.LEAD 5 0.4 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.5 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.4 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD125
Date of Calibration 21-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 20-10-2022
CUSTOMER Address
624001
Date of Issue 22-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. SNF10P1902
Make/Model No. L&T/PLANET 50 Location OT-4
Asset Barcode No. SJH/BME/MOT/09 Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229.3 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 229.2 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.4 3 0 V P
4 EQUIPMENT CURRENT 0.4 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.4 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.7 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.6 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.5 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.4 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.6 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD184
Date of Calibration 22-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 21-10-2022
CUSTOMER Address
624001
Date of Issue 23-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 200331091100168
Make/Model No. YONKER/YK-8000C Location EMERGENCY
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.5 1 0 MΩ P
  5.LEAD 5 0.4 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.5 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.4 1 0 MΩ P
  5.LEAD 5 0.1 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD185
Date of Calibration 22-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 21-10-2022
CUSTOMER Address
624001
Date of Issue 23-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 200331091100131
Make/Model No. YONKER/YK-8000C Location EMERGENCY
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.6 1 0 MΩ P
  5.LEAD 5 0.4 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.5 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.4 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD186
Date of Calibration 22-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 21-10-2022
CUSTOMER Address
624001
Date of Issue 23-10-2021

DUC Details
DUC Name ECG MONITOR S No. U99N 1338
Make/Model No. L&T Location EMERGENCY
Asset Barcode No. SJH/BME/EMR/05 Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.6 1 0 MΩ P
  3.LEAD 3 0.3 1 0 MΩ P
  4.LEAD 4 0.2 1 0 MΩ P
  5.LEAD 5 0.5 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.6 1 0 MΩ P
  2.LEAD 2 0.2 1 0 MΩ P
  3.LEAD 3 0.6 1 0 MΩ P
  4.LEAD 4 0.4 1 0 MΩ P
  5.LEAD 5 0.4 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD193
Date of Calibration 22-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 21-10-2022
CUSTOMER Address
624001
Date of Issue 23-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 200331091100115
Make/Model No. YONKER/YK-8000C Location SICU
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229.4 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.4 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.2 3 0 V P
4 EQUIPMENT CURRENT 0.3 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.7 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.3 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.6 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD194
Date of Calibration 22-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 21-10-2022
CUSTOMER Address
624001
Date of Issue 23-10-2021

DUC Details
DUC Name MULTIPARA MONITOR S No. 200331091100103
Make/Model No. YONKER/YK-8000C Location SICU
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 229.4 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.4 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.2 3 0 V P
4 EQUIPMENT CURRENT 0.3 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.7 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.3 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.4 1 0 MΩ P
  3.LEAD 3 0.6 1 0 MΩ P
  4.LEAD 4 0.5 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.4 1 0 MΩ P
  2.LEAD 2 0.5 1 0 MΩ P
  3.LEAD 3 0.4 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER
Customer Name ST. JOSEPH HOSPITAL Certificate No. SJHD220
Date of Calibration 22-10-2021
 NEHRUJI NAGAR, DINDIGUL, TAMIL NADU Valid Upto 21-10-2022
CUSTOMER Address
624001
Date of Issue 23-10-2021

DUC Details
DUC Name ECG MONITOR S No. U00N2424
Make/Model No. L&T Location MEDICAL WARD
Asset Barcode No. … Visual Inspection OK
ELECTRICAL SAFETY TEST VALUES

TEST
NO. NAME OF THE TEST READING HIGH LIMIT LOW LIMIT UNIT STATUS
1 MAIN VOLTAGE -LIVE TO NEUTRAL 228.6 250 225 V P
2 MAIN VOLTAGE -LIVE TO EARTH 228.5 240 220 V P
3 MAIN VOLTAGE-NEUTRAL TO EARTH 0.3 3 0 V P
4 EQUIPMENT CURRENT 0.2 16 0.1 mA P
5 PROTECTIVE EARTH RESISTANCE 0.3 2 0 Ω P
6 EARTH LEAKAGE CURRENT - NC 0.6 1 0.5 mA P
7 EARTH LEAKAGE CURRENT - OC 0.5 1 0.5 mA P
8 ENCLOSURE LEAKAGE CURRENT -NC 0.4 0.5 0.1 mA P
9 ENCLOSURE LEAKAGE CURRENT -OC 0.5 0.5 0.1 mA P
11 PATIENT AUXILERY -NC          
  1.LEAD 1 0.6 1 0 MΩ P
  2.LEAD 2 0.3 1 0 MΩ P
  3.LEAD 3 0.5 1 0 MΩ P
  4.LEAD 4 0.3 1 0 MΩ P
  5.LEAD 5 0.2 1 0 MΩ P
12 PATIENT AUXILERY -OC          
  1.LEAD 1 0.3 1 0 MΩ P
  2.LEAD 2 0.6 1 0 MΩ P
  3.LEAD 3 0.1 1 0 MΩ P
  4.LEAD 4 0.2 1 0 MΩ P
  5.LEAD 5 0.3 1 0 MΩ P
13 INSULATION RESITANCE          
  1.MAINS-PROTECTIVE EARTH  OL 100 0 MΩ P
  2.APPLIED PART TO PROTECTIVE EARTH  NA 100 0 MΩ P
  3.MAINS TO APPLIED PART  NA 100 0 MΩ P
  4.MAINS TO NEUTRAL  OL 100 0 MΩ P
  5.APPLIED PART TO NEUTRAL  NA 100 0 MΩ P
*NOTE ( NC -NORMAL CONDITION, OC-OPEN CONDTION). REMARKS- STANDARD USED IS IEC 60601-1
DUC:- DEVICE UNDER CALIBRATION

Calibrated By Issued and Approved By


SARATH K B
TECHNICAL MANAGER

You might also like