Professional Documents
Culture Documents
DELTASCOPE® FMP30
ISOSCOPE® FMP30
DUALSCOPE® FMP40
DELTASCOPE® FMP30
ISOSCOPE® FMP30
DUALSCOPE® FMP40
Operators Manual
Non-destructive measurement on
magnetic and non-magnetic
metallic base materials
© 2008 Copyright by
Helmut Fischer GmbH
Institut für Elektronik und Messtechnik, Sindelfingen.
All rights reserved. No part of this manual may be reproduced by any
means (print, photocopy, microfilm, or any other method) or processed,
multiplied or distributed by electronic means without the written consent
of Helmut Fischer GmbH Institut für Elektronik und Messtechnik.
1 Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.1 Trademarks and Liabilities . . . . . . . . . . . . . . . . . . . . . . . 9
1.2 Symbols and Conventions Used in the Manual . . . . . . . 9
1.3 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.4 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.5 Requirements on the Operating Personnel . . . . . . . . . 10
1.6 Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.7 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . 12
1.8 Probe Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
1.9 Handling, Storage and Transport of Calibration
Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.10 Instrument Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.11 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5 Probe Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1 Handling During Measurements . . . . . . . . . . . . . . . . . 49
5.2 Assigning a New Probe . . . . . . . . . . . . . . . . . . . . . . 50
5.3 Setting Up the Dual Mode for DUAL Probes . . . . . . . 52
6 Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.1 Setting Up an Application . . . . . . . . . . . . . . . . . . . . . . 54
6.1.1 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.1.2 Setting Up an Application Using a Dual Probe . . . . 56
6.2 Selecting the desired Application . . . . . . . . . . . . . . . . 57
6.3 Deleting an Application . . . . . . . . . . . . . . . . . . . . . . . . 58
6.4 List of Set Up Applications . . . . . . . . . . . . . . . . . . . . . 59
6.5 Assigning Application Designations . . . . . . . . . . . . . . 61
6.6 Application Specific Settings . . . . . . . . . . . . . . . . . . . . 62
6.6.1 Tolerance Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6.6.2 Measurement Display Resolution . . . . . . . . . . . . . . 65
6.6.3 Automatic Block Creation . . . . . . . . . . . . . . . . . . . . . 66
6.6.4 Auto-Averaging Mode . . . . . . . . . . . . . . . . . . . . . . . . 67
6.6.5 Outlier rejection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
6.6.6 Measurement Programs . . . . . . . . . . . . . . . . . . . . . 70
6.7 Linking Applications . . . . . . . . . . . . . . . . . . . . . . . . . . 72
6.7.1 Application Linking Procedure . . . . . . . . . . . . . . . . . . 72
6.7.2 Enabling and Disabling the Linking Mode . . . . . . . . . 73
6.7.3 Example for Linked Applications . . . . . . . . . . . . . . . . 74
6.7.4 Linking Dual Probes . . . . . . . . . . . . . . . . . . . . . . . . . 76
8 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
8.1 Preparing for a Measurement . . . . . . . . . . . . . . . . . 108
8.2 Influencing Parameters . . . . . . . . . . . . . . . . . . . . . . . 108
8.3 Making a Measurement . . . . . . . . . . . . . . . . . . . . . . 109
8.3.1 Measurement Acquisition . . . . . . . . . . . . . . . . . . . 112
8.3.2 Measurements With External Start Enabled . . . . . . 113
9 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
9.1 Evaluation of the Current Block “Block Result” . . . . 142
9.1.1 Documenting the Block Result With a Printer . . . . . 145
9.1.2 Printout of the List of Single Readings . . . . . . . . . . . 146
9.1.3 Computed parameters - Block result . . . . . . . . . . . . 147
9.2 Evaluation of the Open Application “Final Result” . . 148
12 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251
Important Information
1.1 Trademarks and Liabilities
DUALSCOPE®, DELTASCOPE® and ISOSCOPE® are registered trade-
marks of Helmut Fischer GmbH Institute for Electronics and Metrology.
The fact that the trademark characters ® and ™ may be missing
does not indicate that a name is a free trademark.
Great care has been exercised in creating this operator's manual. The Helmut Fis-
cher GmbH Institute for Electronics and Metrology assumes no liability for po-
tentially remaining erroneous or incomplete statements and their results. We
would, however, appreciate if you can make us aware of potentially existing er-
rors or incomplete information.
z Listing.
ENTER Writing convention for instrument keys and command buttons on the
display.
The instrument is to be used for coating thickness measurements on steel and iron
base materials as well as on non-ferromagnetic, electrically conducting base ma-
terials.
Only accessories recommended and approved by Helmut Fischer GmbH ( be-
ginning on Page 25) may be connected to this instrument.
The values shown for the measured coating thicknesses and the texts of
the information lines of the LCD display serve as examples for possible
displays. It is entirely possible that different values appear on the LCD
display or in the printout without having made any mistakes.
Important Information
To avoid damage to the instrument or wrong measurement
results due to a wrong line voltage, the instruments must be
connected to a power outlet only via the AC adapter pro-
vided by Helmut Fischer GmbH Institute for Electronics and
Metrology. The line voltage must agree with the line voltage
rating stated on the nameplate of the AC adapter.
EMC
The instrument complies with the laws concerning electromagnetic compatibility
of instruments (2004/108/EC). The measured coating thickness values are not in-
fluenced by the highest level of interference mentioned in the EN 61000-6-2
Standard (which references the Standards EN 61000-4-2, EN 61000-4-3 and EN
61000-4-4).
In particular, the instrument is shielded effectively from strong electromagnetic
fields (e.g., motors, power lines, radio transmission towers).
Low Voltage
The instrument adheres to the Low Voltage Directive 2006/95/EC.
Important Information
To avoid breakage of the wiring, do not bend the probe con-
nector cable! The radius of rolled up probe connector cables
should always be at least 50 mm!
Probe-
connector ca-
ble
R ≥ 50 mm !
During the measurement, the magnetic poles of the probes are placed directly
onto the specimen. Observe the following to keep wear of the magnetic poles dur-
ing the contacting measurement to a minimum:
Observe the following to ensure the proper condition of the calibration standards:
To keep wear of the calibration standards during the contacting mea-
surements to a minimum, use the calibration standards for the calibration
only and not for test measurements!
Important Information
The Helmut Fischer GmbH Institute for Electronics and Metrology will assume
no warranties in the following instances:
z Use of instrument or accessories for purposes other than the intended use.
1
1 Probe connector socket,
Page 38
10 Battery compartment,
9
Page 36
10
Bell:
Tolarance limits are enabled ( beginning on
Page 117).
Padlock:
restricted operating mode is enabled, i.e., the keys
MENU, ZERO and CAL are not active, the service func-
tions cannot be retrieved, applications cannot be deleted
Arrow circle:
“Free-running display” is enabled, measurements are
displayed continuously when the probe is placed on the
specimen
( beginning on Page 127).
Alternatively:
Display for area measurement
Display for automatic measurement
Arrow up:
Upper tolerance limit is exceeded.
Arrow down:
Measurement below lower tolerance limit.
Battery:
The battery must be replaced or the rechargeable bat-
tery must be charged because the voltage dropped be-
low a minimum value ( beginning on Page 34).
Hourglass:
Measurements are currently not possible because an in-
strument-internal routine is running.
Chain:
Applications that have been set up using the same probe
are linked to each other, i.e., the same normalization
and/or corrective calibration is used for determining the
measurements for these Applications ( beginning on
Page 72).
Sheets:
Matrix measuring mode is enabled ( beginning on
Page 137).
Key:
Measurement block is closed.
Information lines:
...SCOPE ... Instrument type:
FKA... Instrument-internal software version
The overview on the following pages provides a brief description of the functions
of the individual control panel keys:
Key Function
.. during normalization:
1x DEL - Deletes the last reading,
2x DEL - Deletes the measurement series of the base material
.. during calibration:
1x DEL - Deletes the last reading,
2x DEL - Deletes the measurement series of the current cali-
bration standard.
Repeated pressing of DEL: Deletes the measurement series
of the previous calibration standards ( beginning on
Page 125).
.. in all menus:
DEL - Returns to the previous menu or cancels the procedure.
.. and then ∧:
Ends the display of the block result (return to the measurement
screen) without closing the open measurement block (the cur-
rent measurement series can be continued).
.. and then ∨:
Displays the block result of the previous or following measure-
ment block of the open Application.
All block results of the open Application can be retrieved in
succession through repeated pressing of ∨.
.. and then ∧ or ∨:
Selects the desired Application.
tolerance Limits
Resolution
Block size
i individual values
Outlier
.. and then ∧ or ∨:
Selects the settings to be edited
5 x ENTER:
Calls the service functions
The instrument settings in the Service Functions menu are
password-protected. “157” will be displayed after pressing EN-
TER 5 times. Press ∧, 2 times to increase this value to the fac-
tory-default password “159” and confirm the entry with
ENTER.
2.3 Accessories
2.3.1 Probes
Various probe models are available for measurements on objects with different
shapes and different surface properties. Special probes with different measure-
ment ranges are available for the following areas of application, for example:
A probe-specific calibration standard set for the master calibration (can be or-
dered as an option) and a probe-specific calibration standard set for the corrective
calibration (included with the probe) are available for each probe model and have
been compiled specifically for this probe model.
You can obtain additional calibration standards of various thicknesses on request
from Helmut Fischer GmbH Institute for Electronics and Metrology or from your
authorized supplier.
Reference area
Order
number
of the cali
bration foil
Foil thick- 75.0
Guaranteed Masterfolie 2
error limit
Fig. 2-4 Calibration foil (example) and master foil (example)
When measuring the thickness of foils with the same thickness but made
of different materials (e.g., a 12 µm thick CuBe foil and a 12 µm thick
plastic foil) on rough specimens, the values measured on these two foils
may deviate by 1 - 2 µm from each other, depending on the roughness.
This deviation is caused by the greater rigidity of the CuBe foil. (The rigid
CuBe foil rests relatively plane on the roughness peaks while the soft
plastic foil adjusts somewhat to the roughness due to the pressure force
of the probe tip.)
For this reason, the same foil material that is used for the calibration
should also be used for the test measurements!
2.3.3 Printer
For an overview of printers suitable to be connected to the instrument see the
menu Service Functions / PRINT ( Chapter 10.5.1 ‘Print Output’, begin-
ning on Page 165).
After receiving the shipment, packaging and contents should be checked for po-
tential damage. If the packaging, the contents or the accessories show signs of
damage, retain the packaging. It might be needed to assert a claim for damages
versus the shipping company.
It is also advisable to keep the packaging for future transport.
Also verify that all components of the standard contents of the shipment and all
ordered options are present. Notify your authorized supplier or the Helmut Fis-
cher GmbH Institute for Electronics and Metrology if this is not the case.
2.5.2 Options
Available options are:
z AC adapter
z Various measurement probes
z Foils - also with certificate - for the calibration and for verifying the
calibration
z Probe-specific calibration standard set for the master calibration -
also with certificate
z 1.5 V NiMh rechargeable battery (4 each/unit)
z Charger for NiMh battery
z Support stand V12 for reproducible positioning of measurement
probes on the specimen
ness Measurements
The following table shows an exemplary listing of FMP instruments and probes
as well as the measuring application for which they can be used and the measure-
ment method employed to measure the coating thickness.
plastic coatings.
Determination of the thickness of Amplitude-Sensitive Eddy
electrically non-conducting, non- Current Method According to
magnetic coatings on non-ferro- DIN EN ISO 2360.
magnetic electrically conducting
base materials. Paint, lacquer or
plastic coatings on, for example,
aluminum, copper, zinc, etc as
well as anodized coatings on alu-
minum.
ing
To connect the instrument to the line voltage via the AC adapter, the AC adapter
must be connected to the instrument and to the line power outlet. The instrument
must be switched off for this purpose!
1. Use ON/OFF ( Page 41) to switch the instrument off (if not yet done).
2. Place the instrument with its back pointing up on the table. Open and
remove the battery compartment cover on the rear of the instrument as
depicted below.
3. If old batteries are in the instrument, remove them from the unit. Other-
wise, install new batteries directly; observe the correct polarity of the bat-
teries
Disposal: Do not dispose of batteries with regular household waste!
Place damaged or used batteries / rechargeable batteries in designated
collection containers! Please observe the guidelines in your region con-
cerning proper handling of waste electrical and electronic equipment and
accessories.
- - -
-
LR6 1.5V
LR6 1.5V
+
LR6 1.5V
LR6 1.5V
+ + +
Fig. 4-2 Inserting the batteries and closing the battery compartment cover
1. Use ON/OFF ( Page 41) to switch the instrument off (if not yet done).
3. Plug the probe plug of the new probe into the probe connector socket of
the instrument.
1. 2.
Probe connector plug
Connector socket
Instrument
Exception: A flashing symbol for the measurement method on the LCD display
indicates that the instrument does not recognize the connected probe. In such a
case:
z A new Application must be set up for the connected probe
( 6.1 ‘Setting Up an Application’, beginning on Page 54) and/or
z The probe must be assigned to the instrument or to the respective Applica-
tion, respectively ( 5.2 ‘Assigning a New Probe’, beginning on Page 50)
or
z The probe that has been linked to the respective Application thus far must be
reconnected.
A new corrective calibration must be performed after the probe has been
assigned ( Page 86)!
Display Explanation
Alternatively to the LCD displays presented above, the following displays may
appear after power-up:
The current Application has not yet been set up. To make
coating thickness measurements, an Application must be
set up using the connected probe ( 6.1 ‘Setting Up an
Application’, beginning on Page 54).
z a normalized countrate Xn
z a countrate X
z Xn and Xs
.
Wipe off dirt immediately to avoid it from drying onto the surface!
Probe Handling
5.1 Handling During Measurements
Example:
Probe
FGAB1.3
z Always hold the probe at its grip sleeve (right fig- Grip sleeve
ure).
Specimen
z Always place the probe gently and at a right angle
on the specimen surface.
z Slide the grip sleeve to the specimen surface such
that the sleeve rests on the specimen (center and bot-
tom Figure, right).
z With the default setting, a beep will signal the mea-
surement capture.
z Lift the probe off the specimen before making the
next measurement.
The instrument recognizes if the probe connected to the unit is different than the
one expected according to the probe identification in the current Application.
Reason: Each individual probe has a name consisting of the identification num-
ber and the model designation (e.g., FGAB 1.3). The probe must be “registered”
in the instrument under this name.
z If a user has more than one probe of the same probe model, a problem occurs
if a not yet assigned probe is connected to the instrument. In such a case, it is
advisable to identify the probes and/or instruments with numbers.
If the measurements of an opened Application have not been made with the con-
nected probe, the symbol identifying the measurement method (e.g., [NF/Fe] or
[Iso/NF]) flashes.
If a DUAL probe is connected to the instrument, the menu Dual method will
appear additionally in the menu Service Functions / Measurement
( 11.7.6 ‘Dual Method’, beginning on Page 205). Select the measurement
method from this menu:
Probe Handling
up will be displayed on the LCD display after power-up ( 4.3.1 ‘Measurement
Method of the Connected Probe’, beginning on Page 41):
The dual method can be set up separately for each Application. The
dual method settings of the other Applications will not be affected.
If the dual method is set to both, both measurement methods can be used to make
measurements. When the probe is placed on the specimen, the correct measure-
ment method will be selected automatically.
If the dual method is set to NC/NF or NF/Fe, only the respective selected mea-
surement method can be used to make measurements. This can ensure, for exam-
ple, that only the coating of interest of a multi-coating system is determined.
For example, to measure the thickness of the Iso coating from Fig. 5-1, the NF
coating must have a certain minimum thickness (e.g., for zinc
about 60 µm) and the dual method must be set to NC/NF:
If the setting is NF/Fe, the ferromagnetic base material will be taken into account
and the thickness of the Iso coating plus the NF coating will be determined.
However, if the setting is both, either the Iso coating alone (for relative thick NF
coatings) or the thickness of the Iso coating plus the NF coating (for relatively
thin NF coatings) will be determined depending on the thickness and electrical
conductivity of the NF coating.
All relevant settings and parameters for a measuring application as well as the
captured measurement data are stored in a file - we refer to this file as an Appli-
cation.
Once the linking mode is enabled (indicated by: on the LCD display),
an automatic check will be carried out to see whether one or more Applications
have already been set up with the connected probe. If this is the case, no normal-
ization is required when setting up the Application. Instead, the normalization
and corrective calibration of the Application(s) that has/have been set up with this
probe will be used.
As long as the restricted operating mode is enabled (indicated by
on the LCD display), only Applications that have already been
set up can be selected, i.e., it is not possible to set up new Applica-
tions ( 11.6.1 ‘Restricted operating mode’, beginning on
Page 190).
Applications
Key sequ./ Detail of the LCD Explanation
Action display
Please note that it is essential to use uncoated reference parts from the
same production as the actual specimens for the normalization or the
corrective calibration!
Applications
To obtain coating thickness readings, prior to making a measurement, a probe
must be connected and an Application must be selected that has been set up using
the connected probe.
The following options exist if an Application has not yet been set up using the
connected probe:
z Setting up a new Application using the connected probe ( 6.1 ‘Setting Up
an Application’, beginning on Page 54),
z Overwriting an existing Application using the connected probe
( 6.3 ‘Deleting an Application’, beginning on Page 58),
z Connecting a probe that has already been used to set up an Application
( 4.2 ‘Connecting Probes’, beginning on Page 38).
Applications
Key sequ./ Detail of the LCD Explanation
Action display
Applications
A customer-specific designation that can be comprised of a max. of 16 ASCII
characters can be assigned to every Application.
Assigning the Application designation can be carried out in the following man-
ner:
z Use of the optional software MPNAME (the software is available from your
authorized supplier or directly from the Helmut Fischer GmbH Institute for
Electronics and Metrology.)
z Transmitting the command “SAN” via the USB port ( 10.4.2 ‘Control
Commands’, beginning on Page 160)
When making measurements in the matrix measuring mode, each block can be
assigned a designation as well ( Page 141).
Once Application or block designations have been assigned, the
designations appear in the information lines of the LCD display,
where applicable, alternating with the respective numbers. On print-
outs, the Application or block designation will appear in place of the
Application or block number.
The following settings apply only to the settings of the open Application, i.e.,
they are Application-specific:
z Settings that are entered after pressing the MENU key.
Applications
Having the tolerance limits enabled allows for a simple and quick determination
if the measured coating thicknesses are within specified limits
( 8.6 ‘Measurements in the Free-Running Display Mode’, beginning on
Page 127).
It is also possible to set the upper and lower tolerance limits by mak-
ing a measurement on a coating in the stipulated thickness range.
If the upper and lower tolerance limits are reversed when entering the settings,
the instrument will automatically select the lower value as the lower tolerance
limit and the higher value as the upper tolerance limit.
Applications
The measurement display resolution specifies the resolution for displaying the
readings on the LCD display.
Resolution 0.0 ...999 1.0 ...9.99 10 ...99.99 100 ...999.9 1000 ...9999
The block size can be modified even if automatic block creation has already been
enabled. However, the prompt for deleting previously made measurements will
appear again because the subsequent statistical evaluation does not permit blocks
containing different numbers of readings.
If the “i” value is modified during a test series, the prompt to delete
all readings obtained before will appear here as well - for statistical
reasons. If this is not carried out, it will not be possible to modify the
“i” value.
Applications
With outlier rejection enabled, recognized outliers will be indicated on the LCD
display and announced through an audible signal ( 8.3.9 ‘Measurements With
Outlier Rejection Enabled’, beginning on Page 121).
The following methods are available for the outlier rejection function:
z Grubbs Test ( Glossary)
z normalized countrate Xn
z Countrate X
z th and Xs
z Xn and Xs
Explanation of the types of measurement data: Glossary
The measurement program can be set differently in each Applica-
tion. The settings of the measurement programs of the other Appli-
cations will not be affected.
With every modification of the type of measurement data, a prompt
for deleting readings already stored in the Application will appear
because it is not possible to evaluate different types of measure-
ment data statistically. This prevents erroneous interpretations of
the measurement results.
Example:
If a separate Application has been set up for each batch of a component (using
the same probe each time), it is useful to link the Applications to each other in
order to compare the Applications to each other. Because these are the same kind
of components, only from different batches, the same normalization and/or cor-
rective calibration should be used every time.
Applications that have been set up with different probes of the same
probe model (indicated by the same model designation but different
serial numbers) will not be linked.
Applications
The linking mode is enabled or disabled from the Service functions of the instru-
ment ( 11.6.4 ‘Linking Applications’, beginning on Page 196).
After disabling the linking mode, all Applications will again be inde-
pendent of each other!
Each Application can again be normalized and/or calibrated sepa-
rately.
2 F
3
4 7
5 8
6 9
F F F F F F
F F F F F F
With the normalization, a new zero point is established for the calibration curve
of the open Application and stored in the open Application. The coefficients of
the master characteristic stored in the EEPROM of the probe plug are not affect-
ed.
Required materials
Reference part: Uncoated part from the production.
Page 84
Operators Manual FMP30 / FMP40
z if the obtained mean value is with the guaranteed error limits stated on the
z If the mean value is not within the guaranteed error limits, perform a correc-
tive calibration.
With a corrective calibration, a new zero point and one additional point (one-
point calibration with one calibration standard) or two additional points (two-
point calibration with two calibration standards) are established for the calibra-
tion curve of the open Application and are stored in the open Application. The
coefficients of the master characteristic stored in the EEPROM of the probe plug
are not affected.
Required materials
z Reference part (uncoated part from the production)
z Calibration foils
Calibration standards are suitable for corrective calibrations only if their normal-
ized countrates Xn are between the following Xn ranges:
z The normalized countrate Xn of the first calibration standard must be
between 0.02 and 0.9 (0.02 ≤ Xn calibration standard 1 ≤ 0.9).
z The normalized countrate of the second calibration standard (in case of a
two-point calibration) must be between 0.12 and 0.98 (0.12 < Xn calibration
standard ≤ 0.98).
z The difference of the normalized countrates of the two calibration standards
that are used for the two-point calibration must be greater than 0.1 (Xn cali-
bration standard 2 - Xn calibration standard 1 = DXn > 0,1).
Required materials
z Reference part (coated part from the production)
Measurements on
Iso/NF: 23.70 µm Calibration stan-
th.= 23.77 µm s= 0.206 µm dard (Correction
foil)
Fig. 7-2 Print form of a corrective calibration (example)
The EEPROM in the probe plug contains memories for the coefficients of two
master calibrations. These two master calibrations differ from each other in the
following manner:
During the user master calibration, the master characteristic can be determined
only if suitable calibration standards are used. The calibration standards are suit-
ed for the user master calibration only if their normalized countrates Xn are with-
in pre-established probe-specific Xn ranges. The Xn values can be displayed after
the 1st step of the user master calibration (normalization) (Query of the probe-
specific limits of the Xn ranges for the connected probe during the master cali-
bration: Page 104.)
z The master calibration applies to all Applications that have been set up with
the same probe that has been used for the master calibration (because the
coefficients have been stored in the probe plug of the EEPROM)! This
applies to both an enabled and a disabled linking mode.
z After a master calibration is completed, the normalizations and corrective
calibrations of all Applications that have been set up with the same probe
that has been used for the master calibration will be deleted automatically.
z If a dual probe is connected, only the normalizations and corrective calibra-
tions of the measuring method in which the new master calibration has been
performed will be deleted.
Required materials
z Uncoated specimen with base material and shape that correspond to those of
the actual parts to be measured
z Probe-specific calibration standard set (master foils)
Pressing FINAL-RES:
Determination of the Normalized Countrate
Xn of a Calibration Standard During a Mas-
ter Calibration Page 107.
Pressing BLOCK-RES:
Displaying Xn Ranges for Calibration Stan-
dards for the Master
Calibration Page 104.
Measuring
curvature, surface roughness, base material thickness)
z Magnetic permeability or electrical conductivity of the specimen
z To avoid erroneous readings, do not allow the probe to hover above the spec-
imen.
z How high the probe should be lifted off depends on the measurement range
of the probe. To obtain a correct air value, the distance to the specimen
should be at least 3 to 4 times the max. measurable coating thickness.
z To allow sufficient time for a measurement acquisition, the time between
individual measurements must be greater than 0.5 seconds.
1. Place the probe: 2. Lift the probe off:
Measurement Object
Measurement Object
Measuring
probes for the measurements:
z With cylindrical specimens:
Place the probe tips parallel to the longitudinal axis!
z Place both probe tips simultaneously and with the same pressure on the spec-
imen!
z For specimens with a base material exhibiting a preferred magnetic direc-
tion, it is recommended to generate the measured value from two averaged
readings with the probe being rotated by 90° after the first measurement. A
preferred magnetic direction may be created through specimen processing
(e.g., rolling or drawing of sheet metal).
90°
6. Rotate the probe
90°
With the “free-running” display mode on, measurement acquisition can be trig-
gered in the following manner ( 8.6 ‘Measurements in the Free-Running Dis-
play Mode’, beginning on Page 127):
z Pressing the ENTER key
z Transmitting the control character “ESC?” via the USB port (ESC =
ASCII27)
( 10.4.2 ‘Control Commands’, beginning on Page 160)
Regardless of how the measurement acquisition occurred, the reading will appear
on the LCD display following the measurement acquisition.
Measuring
When an automatic measurement acquisition is not desired, e.g., when making
measurements in pipes, boreholes or grooves, external start should be enabled
and automatic measurement acquisition disabled. (Enabling externally triggered
measurement acquisition and disabling automatic measurement acquisition:
Page 199)
With external start enabled, the measurement acquisition can be triggered manu-
ally in the following ways once the probe is positioned on the measurement loca-
tion:
z Press the ∧ key
(not with a normalization or calibration)
z Press the FINAL-RES key
(only during a normalization or calibration)
z Transmitting the “ES” command via the USB port
( 10.4.2 ‘Control Commands’, beginning on Page 160)
Measuring
With automatic measurements, a specified number of readings is taken with a se-
lectable time interval between 2 measurements. In contrast to the area measure-
ment, all single readings will be stored ( Page 113).
This measuring mode can be advantageous, for example, when the coating thick-
ness distribution shall be determined along a line. If the probe is moved by hand
or a suitable mechanical device along a line at a consistent speed, then the mea-
surement points will be at equal distances.
Signal Meaning
Measuring
In addition, with dual probes, the measurement method used for the last per-
formed measurement will be indicated.
Display Explanation
[NF/Fe NC/NF] The magnetic induction method was used for the mea-
surement with the dual probe.
[NF/Fe NC/NF] The eddy current method was used for the measurement
with the dual probe.
As long as the tolerance limits are enabled, will appear on the LCD
display.
If the measurement acquisition signal has been disabled, only the signal, or the
signals, for violating the tolerance limits will sound.
With outlier rejection enabled, the audible signals indicating an outlier measure-
ment will sound when an outlier is recognized. In this case, the violation of the
tolerance limit will not be indicated acoustically.
z appears on the LCD display after the last measurement of the block has
been taken, and
z the closure of the block is indicated by a long audible signal (if the measure-
ment acquisition signal has been disabled, only the signal to indicate the
block closure will sound).
With a connected and powered up printer, it is now possible to print the block re-
sult automatically after the block closure or upon request ( 11.5.4 ‘Block Re-
sult’, beginning on Page 184).
In the auto-averaging mode and with outlier rejection enabled, the single readings
that are recognized as outliers will not be integrated in the measurement calcula-
tion! For example, if an outlier is recognized at i = 5, only the 4 remaining read-
ings will be used to generate the mean value.
Measuring
When making measurements with outlier rejection enabled:
z The outlier measurements recognized by the instrument will appear on the
LCD display and will be indicated acoustically ( 6.6.5 ‘Outlier rejection’,
beginning on Page 69),
z two long audible signals will sound to indicate the outlier measurement in
addition to the audible signal for the measurement acquisition,
z recognized outliers will not be integrated in the evaluation of the current
block or the current Application.,
z only the signals for indicating the outlier measurements will sound if the
measurement acquisition signal has been disabled,
z outlier measurements recognized in the auto-averaging mode will not be
integrated in the measurement calculation, and
z only the audible signals indicating an outlier measurement will sound when
an outlier is recognized (a violation of the tolerance limits will not be indi-
cated acoustically).
Detail of the LCD Explanation
display
Outlier! appears briefly in the information lines of the
LCD display in addition to the audible signal if a past
reading with the number n (in the example, n = 1) is rec-
ognized as an outlier measurement (thereafter, the mea-
surement just taken will appear).
If tolerance limits are enabled when making measurements, the readings will be
entered between or next to the tolerance limits on the printout ( Fig. 8-5).
For measurements in auto-averaging mode, only the mean value computed from
the single readings will be printed and stored. The individual readings are not
printed.
Measuring
ginning on Page 66), the readings and the block results of all blocks of the open
Application will be printed in succession:
Measuring
8.5.1 Deleting Erroneous Readings
If an erroneous reading is recognized directly after measurement acquisition, the
reading can be deleted by pressing DEL.
The deleted reading will then not be included in the generation of the block and
final results.
Repeated pressing of DEL will delete all readings of the current block in succes-
sion.
8.5.2 Deleting All Readings of a Block That Has Not Been Closed
Pressing DEL during an evaluation of an open block will delete all readings that
are stored in the open block that has not been closed ( 9.1 ‘Evaluation of the
Current Block “Block Result”’, beginning on Page 142).
The instrument does not carry out outlier rejection when overwriting
stored measurement data (even if outlier rejection is enabled).
z the data are output continuously via the USB port only if [Free-running
transmit on] has been selected from the Service functions.
Measuring
To enable the free-running display mode: Press the ∨ arrow key.
z The “free-running” display mode remains enabled until it is disabled; i.e., it
does not need to be enabled every time the instrument is powered up.
To disable the free-running display mode: Press the arrow key ∨ again.
The magnetic poles are subject to increased wear when moving the
probe across a surface.
When making measurements in the “free-running” display mode, the analog dis-
play facilitates a quick recognition of tendencies in the coating thickness chang-
es. If analog display is enabled and measurements are made in the “free-running”
display mode, the analog display with the set limits will appear in place of the
information lines. The reading will appear between the limits as an analog bar.
(To enable analog display: Page 192).
As long as the tolerance limits have not been enabled, the measurement range
limits of the connected probe will be displayed as the limits for the analog dis-
play. Once the tolerance limits have been enabled, the set tolerance limits will be
displayed as the limits for the analog display. ( 6.6.1 ‘Tolerance Limits’, be-
ginning on Page 63).
With the free-running mode enabled, the tolerance limits cannot be set
through a measurement but only via the arrow keys.
If a reading violates the upper or lower tolerance limit when making measure-
ments in the “free-running” display mode and the measurement acquisition signal
is enabled, the respective signal for indicating the violation of the tolerance limit
will sound ( 8.3.4 ‘Audible signals After the Measurement Acquisition’, be-
ginning on Page 116 or 8.3.6 ‘Measurements With Tolerance Limits En-
abled’, beginning on Page 117).
If the measurement acquisition signal is disabled during measurements in the
“free-running” display mode, the audible signal for indicating a violation of the
tolerance limits is automatically disabled as well ( 11.7.1 ‘Audible Signal’, be-
ginning on Page 197).
If the magnetic induction method was used before the “free-running” display
mode was turned on, the measurements in the free-running mode will be made
according to the magnetic induction method as well. The dual probe will then
measure only on ferromagnetic base materials. The probe will not measure on
other base materials, i.e., [- - - -] will remain on the LCD display.
If the eddy current method was used before the “free-running” display mode was
turned on, the measurements in the free-running mode will be made according to
the eddy current method as well. The dual probe will make correct measurements
only on non-ferromagnetic base materials; erroneous measurements will be taken
on ferromagnetic base materials.
If no measurements were made before the “free-running” display mode is turned
on, the measurements in the free-running mode will be made according to the
magnetic induction method.
Measuring
mode
The preparations required for making the actual measurement are independent of
the selected measuring mode and are described in the Chapters
“ 8.1 ‘Preparing for a Measurement’, beginning on Page 108 and
8.3 ‘Making a Measurement’, beginning on Page 109.
Measuring modes can be changed from the service functions only
( 11.6.3 ‘Matrix Mode’, beginning on Page 193).
The instrument is re-initialized automatically when changing from the
standard to the matrix measuring mode. During a re-initialization, all
Applications and all measurement data stored in them are deleted and
all settings of the service functions are reset to their default settings.
After the re-initialization, i.e., also after changing the measuring mode,
the required Applications will need to be set up anew and the desired
settings need to be established again in the service functions!
In the standard measuring mode, readings can be stored only in the respective lat-
est not yet closed block of an Application.
The individual applications may contain different numbers of blocks and differ-
ent numbers of readings may be stored in a block. However, if automatic block
creation is enabled in an Application, only blocks of the set block size can be cre-
ated ( 6.6.3 ‘Automatic Block Creation’, beginning on Page 66 and
8.3.7 ‘Measurements With a Fixed Block Size’, beginning on Page 118).
Fig. 8-8 Example for the configuration of the Application in the Standard measuring mode
Measuring
When changing to the matrix measuring mode, the number of Applications and
the number of blocks must be entered. The same number of blocks is set up for
each Application. Each block can receive the same maximum number of read-
ings.
After entering the number of Applications and blocks, the instrument will auto-
matically compute and display the maximum number of readings. For example,
if HMatrix mode on (3/20/318)G appears when the instrument configuration is
displayed, a maximum of 3 Applications each with 20 blocks each with a maxi-
mum number 318 readings can be set up.
Fig. 8-9 Example for the configuration of the Application in the matrix measuring mode
The matrix measuring mode is suited for coating thickness measurements when
the measurements are to be made on different specimens of the same type in suc-
cession, always at certain reference areas, and the measurements of correspond-
ing areas are to be combined into blocks. The matrix measuring mode is used, for
example, in the automotive and in the steel construction industries.
When selecting a reference area where measurements in an Application
are to be made, it should be observed that the same normalization and
corrective calibration is underlying the determination of the coating thick-
ness value. Thus, only such reference areas should be selected for
which no other normalization or corrective calibration is required.
Automatic block creation is not available in the matrix measuring mode. Corre-
spondingly, it is not possible to set a block size for automatic block creation dur-
ing the application-specific settings using MENU ( 6.6 ‘Application Specific
Settings’, beginning on Page 62).
Measuring
appears on the LCD display when the matrix measuring mode is
enabled.
In the matrix measuring mode, it is possible to change the blocks, i.e., to select
the block where the next reading is to be stored manually or automatically
( 11.6.3 ‘Matrix Mode’, beginning on Page 193).
A block will be closed automatically when during the measurement in the matrix
measuring mode the maximum number of readings that can be stored in a block
is reached. appears on the LCD display.
In addition, a long acoustical signal indicating the block closures sounds after the
audible signal for the measurement acquisition.
If the measurement acquisition signal has been disabled, only the signal indicat-
ing the block closure will sound.
Measuring
If additional measurements are made in the matrix measuring mode
while appears on the LCD display, these measurements will not be
stored, printed or taken into account during the evaluation. After the
measurement, The LCD display will briefly show the error message:
“E024 - Block is full”.
To store additional readings, select the block where the readings are to
be stored (i.e., change the blocks manually) or delete the readings.
The following options are available for evaluating the measured coating thick-
ness values:
z Evaluation of the current block (block result)
When making measurements with a fixed block size and the last block has not
been concluded, the next reading will be added to this block (even if ENTER is
used to end the evaluation). In this case, the block will be concluded only after
the number of readings in this block corresponds to the set block size
( 8.3.7 ‘Measurements With a Fixed Block Size’, beginning on Page 118).
Notes:
[Change block: ∨] or [Change block:
∧∨]: Use ∧ or ∨ to display the block result
of the next or previous block.
[Single meas.: MENU]:
Use MENU to display the single readings
of the evaluated block (then use ∧ or ∨ to
display the readings in succession, use
MENU to return to the block result).
[Delete block: DEL]:
Use DEL to delete the readings of the last
not yet closed block (appears only during
the evaluation of not closed blocks).
[Exit: ENTER]: Ends the block result dis-
play. The current block will be closed.
appears on the display once the block
is closed.
Evaluation
If outlier rejection is enabled, recognized outlier measurements are
not included in the evaluation.
If you change during the display of the block result of the current block to the
block result of a previous block (by pressing ∨), then the block result of this
block can be printed as well using PRINT. The printout will then include a list
of the readings of this block as well.
Evaluation
FISCHER DUALSCOPE Instrument type:
FMP40
2008-06-27 Current date
Appl. No. Application name/Application number (appears only if
a designation has been assigned ( 6.5 ‘Assigning
Application Designations’, beginning on Page 61)
-- Block result -- Result type
Block No. Current measurement block number/block name (ap-
pears only if a designation has been assigned
6.5 ‘Assigning Application Designations’, begin-
ning on Page 61)
Date Time Date and time of the last measurement of the evalu-
ated block or closed block, respectively
Mean value th. Mean value with a confidence interval of 95%
Std. dev. s Standard deviation
n In the list of the readings: Sequential number of the
reading
In the block result: Number of the evaluated readings
C.O.V. Coefficient of variation
lowest meas. Lowest block mean value (Minimum)
highest meas. Highest block mean value (Maximum)
LSL/USL Upper / lower specification limit (tolerance limit)
* Reading is within the tolerance limits
<< / >> Reading violates the tolerance limits
!/!! The measurement was recognized as an outlier
The final result combines the results of the evaluation of all readings contained
in the open Application.
Taking a new measurement or pressing the ENTER key during the evaluation of
the open Application ends the final result display.
When making measurements with a fixed block size and the last block has not
been concluded, the next reading will be added to this block (even if ENTER is
used to end the evaluation). In this case, the block will be concluded only after
the number of readings in this block corresponds to the set block size
( 8.3.7 ‘Measurements With a Fixed Block Size’, beginning on Page 118).
When making measurements with a fixed block size, only the mea-
surements of the closed blocks will be included in the evaluation of
the open Application.
[Exit: ENTER]:
To end the final results function
without deleting the readings
use ENTER.
ENTER
Use ENTER to finish the final result func-
tion.
Evaluation
Corresponding to the settings in Service menu/Print/Final result (
11.5.5 ‘Final Result’, beginning on Page 186), pressing FINAL-RES allows for
the block result of the current block to be printed with date and time stamp “au-
tomatically” or “on request” if a printer is connected and powered up.
Only if the tolerance limits are enabled will the number of measurements that are
violating the lower or upper tolerance limits be specified on the printout of the
final result( Fig. 9-5).
Product ..............................
Name ..............................
Appl.No. 0 -- Final result --
from 27.06.08 11:26 to 27.06.08 11:27
Product ..............................
Name ..............................
Appl.No. 0 -- Final result --
from 27.06.08 11:12 to 27.06.08 11:23
Evaluation
FISCHER DUAL- Instrument type:
SCOPE FMP40
2008-06-27 Current date
Product .....
Name .....
Appl. No. Number of the Application
-- Final result -- Result type
from ... to ... Date and time of the last block closure of the first and
the last block or the current date and the current time
(if the last block is not yet closed).
Mean value th./th.. Mean value/Mean value of the block mean values
with a 95% confidence interval
Std. dev. s Standard deviation
Est. Std. dev. Estimated standard deviation (only specified with
fixed block sizes).
n Number of evaluated readings
ni Block size (number of readings per block)
Std. dev. sa Standard deviation sa (is specified only if the spread
of the mean values cannot be explained from the
spread within the blocks according to a variance anal-
ysis).
nBl Number of evaluated blocks
Estimated COV Coefficient of variation
smallest value/ Lowest reading (Minimum)/
smallest block lowest block mean value
largest value/ Highest reading (Maximum)/
largest Block highest block mean value
LSL/USL Upper / lower specification limit (tolerance limit)
cp / cpk Process capability index
HISTOGRAM
n
37.9|
1 |*
38.2|
3 |*
38.3|
1 |*
38.4|
2 |*
38.5|
1 |*
38.7|
Normal distribution
NORMAL PROBABILITY CHART
Upper limit Norm. Prop. Chart
38.2 12.50 %
38.3 50.00 %
38.4 62.50 %
38.5 87.50 %
38.7 100.00 %
Fig. 9-7 Printout of a histogram with normal distribution (example)
Evaluation
normal distribution (with a normal distribution, skewness and kurtosis are Zero!):
HISTOGRAM
n
-3.0|
3 |***
-2.0|
1 |*
-1.1|
0|
0.68|
0|
34.6|
3 |***
39.6|
Distribution not normal
Skewness= 0.30 Kurtosis= -1.9
NORMAL PROBABILITY CHART
Upper limit Norm. Prop. Chart
-2.0 42.86 %
-1.1 57.14 %
0.68 57.14 %
34.6 57.14 %
39.6 100.00 %
Fig. 9-8 Printout of a histogram without normal distribution (example)
The USB port for the instrument is on the unit's rear side. Bi-directional data ex-
change occurs via the USB interface.
The following operations are possible when the USB port is connected to a com-
puter (PC):
z Transfer of the readings and the characteristic statistical data from the instru-
ment to the PC.
z Remote control of the instrument by sending commands from the PC to the
instrument.
z Requesting measurement data and other data (e.g., the name of the current
Application) by sending commands from the PC to the instrument.
z Transfer of data (e.g., designations for Applications) from the PC to the
instrument by sending commands from the PC to the instrument.
Both commercial or one's own data processing programs can be used to process
the data exported by the instrument. Information regarding import and processing
of the data using such programs may be obtained from the respective manuals for
these programs.
Fig. 10-1 Side view of the instrument with the USB port
1. Connect the instrument to the USB port of your PC. The “Found New
Hardware Wizard” opens.
2. Follow the instructions of the Windows wizard. If the driver is not found
automatically, select or enter the source to search for the USB driver (e.g.,
CD-ROM drive. removable media (CD, diskettes, ...) or local path).
Windows XP®:
Ignore the message for the Windows Logo Test (Window “Hardware
Installation”). Click the Continue Installation button and continue the
installation.
The successful installation of the USB driver can be verified in the Windows De-
vice Manager.
z Open the Device Manager: Start/Control Panel/System, Hardware tab,
Device Manager button.
2. Select the data for online export: 11.6 ‘Instrument Mode’, beginning
on Page 190.
Command Function
PT2 Sets the measurement date output mode “Single readings with
group separator”, i.e., during the measurement data output, the
single readings are transferred and a group separator (ASCII
GS) is transmitted between each of the measurement blocks.
PT3 Sets the measurement data output mode
“only block mean values”, i.e., during the measurement data
output, only the block mean values are transferred.
ESC0 Operates the DEL key
ESC1 Operates the FINAL-RES key
ESC2 Operates the BLOCK-RES key
ESC3 Operates the ON/OFF key
ESC4 Operates the ZERO key
ESC5 Operates the CAL key
ESC6 Operates the ∧ key
ESC7 Operates the ∨ key
ESC8 Operates the APPL No key
ESC9 Operates the MENU key
ESC: Operates the PRINT key
ESC; Operates the ENTER key
AN Requests the maximum possible number of Applications that
can be set up in the instrument.
GAA Requests the number of the current Application.
SWA0 Selects the Application with the number 0 (... 99) in the instru-
... ment.
SWA99 Example: The result of the command “SWA2” is that the Appli-
cation with the number 2 is selected in the instrument.
IEX0 Checks if the Application with the number 0 (... 99) is set up in
... the instrument. The instrument outputs “1” if the Application is
IEX99 set up and “0” if it is not set up.
DAT0 Requests the date and time of the block closure of the block
... with the number 1 (... 1000).
DAT999 Example: The command “DAT2” outputs the date and time of
the block closure of the third block of the current Application.
GNB Requests the number of measurement blocks that are stored in
the current application.
GG Requests the block size of the current Application, i.e., the
number of readings that are to be combined during automatic
block creation. “0” is output if automatic block creation is dis-
abled.
Enables automatic block creation and sets the block size of the
current Application: 6.6.3 ‘Automatic Block Creation’, begin-
ning on Page 66.
GMX Requests the measuring mode. The instrument outputs “1” if
the matrix mode is enabled. A “0” output indicates that the stan-
dard measuring mode is enabled.
Enabling the matrix measuring mode: 11.6.3 ‘Matrix Mode’,
beginning on Page 193.
GGS Requests the group separator mode.
The instrument outputs “1” if group separator is enabled (i.e., a
group separator is transmitted between each measurement
block during the output via the USB port). “0” is output if the
group separator is disabled.
Setting the group separator mode: Command “SGS1” or
10.3.3 ‘Transferring Data With a Group Separator’, begin-
ning on Page 159.
SGS0 Disables the group separator in the instrument. If the group
separator is disabled, no group separator will be transmitted
between the individual measurement blocks during the output
via the USB port.
SGS1 Enables the group separator in the instrument. If the group sep-
arator is enabled, a group separator will be transmitted be-
tween the individual measurement blocks during the output via
the USB port.
PE Requests the ASCII character that
the instrument uses as the group separator.
The instrument outputs “GS” via the USB port.
SER Requests the serial number of the instrument.
Fig. 10-2 Side view of the instrument with the USB port
z You can obtain an overview of printers that are suitable for connection to the
instrument from the Helmut Fischer GmbH Institute for Electronics and
Metrology or from your authorized supplier.
For information about operation, maintenance and care of the printer,
consult the instruction manual of the printer.
tion Menu
The instrument settings described below are configured from the Service func-
tion menu.
How to access the “Service functions” menu:
Start on
Service menu Functions
Page
11.2 System
11.2.1 Language
Do not select a language for which you do not understand the charac-
ters, e.g., Cyrillic! You might have difficulty returning to a language that is
familiar to you!
11.2.2 Time
z “on w. power unit”: The light is on only with the power unit connected.
∧ or ∨ always off
Enter After confirming the selection “always off”
with ENTER you are immediately returned
to the menu in order to enter additional set-
tings.
∧ or ∨ always on
Enter After confirming the selection “always on”
with ENTER, you are immediately returned
to the menu in order to enter additional set-
tings.
∧ or ∨ on w. power unit
Enter After confirming the selection “on w. power
unit” with ENTER, you are immediately re-
turned to the menu in order to enter addi-
tional settings.
11.3 Evaluation
ENTER
Select the desired group separator mode
by pressing the arrow key ∧ or ∨ and con-
firm the selection with ENTER.
ENTER Enter the value for the left margin width us-
ing the arrow keys ∧ or ∨ and confirm the
entry with ENTER.
The restricted operating mode will remain enabled even after the instrument is
switched off and on.
To ensure that the Applications linked to each other use the same
normalization and corrective calibration, a normalization or correc-
tive calibration, respectively, must be performed after linking for all
probes for which more than one application has been set up.
12.1 Malfunctions
([Delete Appl.?
Yes:DEL N.:EN-
TER]: DEL to deleted
the Application.
[To port?
Yes: DEL N.:EN-
TER], DEL to output
the measurement data
of this Application via
the USB port to a print-
er or a PC.)
Instrument corrected -- --
W 003 faulty settings autono-
mously.
Options
corrected !
Glossary
Terms and Formulas
Adjustments Calibration
Application
Measurement Application of the User.
The instrument memory, where the coefficients from the corrective cali-
bration/normalization (adjustment of the measurement system to the spec-
imen / coating/base material adjustment, system adjustment), the param-
eter settings and the measurement data for a certain measuring application
of the user are stored is called an application.
Application Selection
Menu page(s) in the instrument, where all Applications are listed that have
been set up thus far. Use the command button APPL or use
File/Open to select an Application.
Application Memory
The instrument memory contains all data and measurements that are rele-
vant to a measuring application.
Outliers
Readings that are significantly lower or higher than the other readings of
a measurement series and that can, therefore, be considered unexpected
and not acceptable.
Outlier Rejection
Is used to prevent the distortion of measurement results by Outliers. Out-
lier rejection can be carried out using the Grubbs-Test or Sigma- Outlier
Rejection (specification of a known spread). Measurements that are rec-
ognized as outliers during outlier rejection are not included in the statisti-
cal evaluations.
Evaluation
Computation of statistical parameters such as mean value, standard devi-
ation, etc. as well as the graphical presentation of the measurements, e.g.,
in a sum frequency chart.
Baud
Unit of the speed for transferring information (data). 1 Baud corresponds
to a transfer rate of one bit per second.
Baud Rate
Data transfer rate. Used mainly in connection with terminal programs for
data transfer. Since data are transferred via a serial port, the transfer rate
is calculated in bits per second.
Bit
(Binary Digit), binary number. 1 bit is the smallest unit in the binary num-
ber system. The value of a bit is 0 or 1. Being the smallest unit of infor-
mation in a computer, a bit forms the basis of every computer system. 8
bits are combined to a byte and several bytes to a word.
Block
Grouping of single readings. Several measurements are combined into a
block. A key symbol on the display indicates the end of each block (con-
clusion of a block).
End of Block
Mark after n single readings. As a rule, a key symbol on the display indi-
cates the end of a block.
Block Result
Statistical evaluation of the measurement data of a block. E.g., mean val-
ue, standard deviation, coefficient of variation, range, minimum value,
maximum value, number of single readings per block.
Glossary
Number of single readings that are combined to form a block.
Block, open
Group of single readings for which the block has not yet been ended.
Chi-Squared-Test
Statistical mathematical test method to determine an existing normal dis-
tribution of the measurements (for more than 30 measurements).
Cu
Copper
S1
d..
Mean value of the block mean values of selected blocks. Analogous to the
arithmetic mean value (d.), the block mean values are added up and divid-
ed by the number of evaluated blocks nBl.
n Bl
∑ d.i
1
d.. = ------- ⋅
n Bl d.. Mean value of the block mean values
i=1
nBl Number of blocks
d.i Block mean values
DUALSCOPE
Protected brand name of Helmut Fischer GmbH for the measuring instru-
ment.
Single Reading
Measurement result that is displayed or printed after a single measurement
at the measurement location.
Final Result
Evaluation of all measurements or selected blocks of an Application (mea-
surement application memory).
External Start
A setting in the menu function File/Properties/Measurement Accep-
tance.
Measurement Acceptance can be initiated by tapping the menu function
Meas/Initiate External Start, the command button Ext, by transmitting
commands from a connected PC.
Excess Curvature
Glossary
The process capability is evaluated using the factors Cp and Cpk.
The capability index Cp is a measure for the spread of a process and its
ability for continuously producing parts according to specifications.
The capability index Cpk takes the position of the mean value in relation
to set specification limits into account.
OGW – UGW ⎧ OGW – d.. d.. – UGW ⎫
Cp = ----------------------------------- and Cpk = ⎨ --------------------------- ; --------------------------- ⎬
6 ⋅ σ̂ ⎩ 3 ⋅ σ̂ 3 ⋅ σ̂ ⎭
∑ si
1 2
σ̂ : Estimated value for the theoretical standard deviation σ̂ = ------- ⋅
nBl: Number of blocks n Bl
i=1
si: Standard deviation of the individual blocks
USL: Upper specification limit
LSL: Lower specification limit
FDD® Evaluation
Graphical presentation of the mean values of the measurement blocks (=
Features) in an ascending rank order.
Application:
By using the FDD®, existing systematic differences of coatings can be
shown quickly and clearly in a graphical format
Example 1:
When painting car bodies, the desired homogeneous coating distribution
is not achieved if the settings of one or more spray robots is wrong or dis-
turbed. If the readings of one feature (e.g, the hood) are combined into one
measurement block, then the systematic differences between various fea-
tures (hood, roof, doors, trunk lid, etc.) can be recognized quickly using
FDD®, and corrective measures can be initiated promptly.
Example 2:
When electroplating racks, parts may be coated inadequately due to erro-
neous current contacting or shadowing of flux lines. If the readings of ad-
jacent parts of the rack are combined into one measurement block, then
the groups, which statistically do not “fit” together with the other groups,
i.e., which exhibit systematic differences, are identified clearly in the
FDD® evaluation. This allows for easy “localization” of the source of the
error.
Fe
Magnetizable material made of steel or iron.
Accuracy
Qualitative designation for the degree of approximation of a measurement
result to the true value. Usually, the accuracy is divided into Trueness
and Precision.
Specification Limits
The upper specification limit (USL) is the highest reading and the lower
specification limit (LSL) is the lowest reading allowed at the measurement
location.
Maximum Value
Maximum value measured in a test series.
Grubbs Test
Test method for outlier rejection. A method developed by Grubbs to test,
whether the highest or lowest single reading should be considered an out-
lier.
Population
All pieces or specimens to be measured. In practical applications, for ex-
ample, all parts of a production unit, batch, etc.
Group Separator
Mark for the end of a block that can be transferred together with the mea-
surement data to the PC. Default setting ASCII character 29.
Glossary
Graphical presentation of all readings of an Application (measuring appli-
cation memory) according to their portions in classes (e.g., coating thick-
ness ranges), where the class frequencies are illustrated by the contents of
rectangles. The class widths should be of equal size. Among other criteria,
the informational value of statistical results depends on the shape of this
distribution curve.
Interface Interface,
Iso
Insulating material, electrically non-conducting, non-magnetizable.
Calibration Curve
(Characteristic, master characteristic) Quantitative relationship between
the signal of the probe and a scale for the coating thickness as represented
by the calibration standards.
log d
The coating thickness d is presented as a
calibration curve as a function of the nor-
malized count rate Xn. The mid portion of
the calibration curve approaches a
straight line and constitutes the range
with the lowest relative measurement er-
ror.
Xn
0 1
Calibration
In this manual, the term calibration is used as a comprehensive term for
the adjustment and the calibration: Adjustment of the instrument using
calibration standards to adapt the measuring system (instrument and
probe) to the measuring application.
Class
Range between a lower and an upper class boundary (limit values). The
readings of a measurement series can be sorted according to such classes
if they cover the entire measurement range without gaps. The class con-
tents (frequency or number of measurements per class) plotted over the
classes is called a histogram.
Minimum Value
Minimum value measured in a test series.
System Check
A significant part of monitoring the measurement devices. Calibration
standards or better yet, reference samples are used to check the calibra-
tion and to ensure the measurement stability.
Corrective calibration
One-point or two-point calibration. Adjustment of the instrument using 1
or 2 calibration standards. The corrective calibration includes calibration
and adjustment.
During the corrective calibration, the master characteristic (calibration
curve) is adjusted to the individual measuring application. The obtained
coefficients are stored in the active Application (measuring application
memory). The master characteristic itself remains unchanged.
Kurtosis Curvature
LF Line Feed
Glossary
LF line feed. Advances the printer paper by one line. Is usually used to-
gether with the CR (carriage return) character to start the next line at the
beginning.
Unit of Measurement
Unit for displaying measurement data. In coating thickness measurement,
the common units are µm and mils. 100 µm = 3.9 mils.
Master characteristic
(Characteristic probe output function) Original characteristic of the mea-
surement system. The master characteristic is the basis for determining
the measurements because it represents the relationship between the probe
signal and the coating thickness. The coefficients of the master character-
istic are stored in the probe plug.
Max
Highest measured value of a test series.
Maximum Max
Measuring Application
Properties of the test specimen with regard to measurement quantity, ge-
ometry, geometric dimensions, permeability, etc. The test method, the
measurement display mode, the probe and the type of instrument are all
determined by the measurement application.
Measurement Range
The range between two limit values within which a measurement is pos-
sible at a specified trueness and precision. In a narrower sense, it refers to
the range of the scale of an analog instrument. The measurement range de-
pends on the measurement method, the design of the probe and the mea-
suring application.
Measurement Error
The difference between the actual and the measured value of a measured
quantity. For measuring instruments, there is a distinction between ran-
dom (unpredictable) and systematic (correctable) measurement errors.
Random errors determine the repeatability precision. Systematic errors
affect the trueness and the reproducibility. Systematic measurement er-
rors are far more prevalent in practical applications. Systematic measure-
ment errors can be traced back to 1. faulty calibration, 2. operating or per-
sonal errors and 3. deviations in the test conditions (inhomogeneities,
instabilities, material aging, etc.). They tend to lean in one direction. With
appropriate care, the influences 1. and 2. can usually be avoided or cor-
rected or taken into account in the result.
Glossary
A quality assurance task. It consists of ensuring at specified intervals that
the test equipment (instrument, probe) is still operating properly and cal-
ibrated correctly, and to take corrective measures, if necessary (recalibra-
tion or repair of the instrument). System Check
Measuring Mode
This is the condition in which the instrument can capture and display mea-
surement data. The display mode is determined by the respective mea-
surement display settings.
Measurement Object
Object on whose surface the measurements are performed to determine
the coating thickness.
Measurement Series
A series of single readings between two block or final results.
Measurement Location
A limited and clearly defined location within a reference area of the spec-
imen, where the coating thickness is to be determined.
Measurement Uncertainty u
Measurement Method
A procedure and process for obtaining information from the specimen
concerning its properties. The measurement method is based on scientific
knowledge and is determined by the measuring application.
Reading
Numeric reading of an instrument supplemented by the unit of measure-
ment. The measurement can be obtained from the result of a single read-
ing or from the arithmetic mean of several single readings (for example
for the averaged display value (i individual values)).
Min
Lowest measured value of a test series.
Mean Value d. or
Mean Range R.
NF Metals
Non-ferrous, non-magnetizable metals.
NF
Non-magnetizable material.
Ni
Nickel
Normal Distribution
Gaussian normal distribution, Gaussian distribution, bell curve.
Probability distribution discovered by C. F. Gauß in 1794.
If a quantity X is classified as having normal distribution, 68.3 % of the
observed values X are within the σ-interval of the spread, around the
mean value μ of the quantity X. I.e., the following applies to 68.3 % of the
observed values: μ – σ ≤ X ≤ μ + σ . In the following figure, this interval is
identified by the gray areas underneath the curve.
Probability distribution
P(X) of a quantity X
with a normal distribu-
Probability P(X)
tion.
μ-σ μ μ+σ X
Glossary
μ of the quantity X with normal distribution. Skewness and Curvature
equal zero for the normal distribution.
The populations that are examined for technical purposes often can be
classified as having normal distribution. However, the following fact is
of great significance: If several random samples of equal extent are
drawn from any population (blocks) and their mean values (block mean
values) are determined, these mean values will always have normal dis-
tribution (central limits theorem).
The mean value of these random sample mean values (block mean val-
ues) is an estimated value ( μ̂ = d.) for the mean value μ of the popula-
tion. Due to the normal distribution, the measurement uncertainty u can
be calculated using the standard deviation of the random sample mean
values.
The sum frequency chart shows, whether a quantity has normal distribu-
tion, with a straight line indicating normal distribution.
In the instrument, the test, whether the measurements at hand (random
samples) have a normal distribution is carried out using the Kolmogoroff
Smirnoff Test (for up to 30 readings) and the Chi-Squared Test (30 read-
ings and up).
Normalization
Adjustment of the instrument to the material properties of the coating and/
or the base material (probe dependent). Thus, the normalization essential-
ly defines the zero point or end point, respectively. A normalization is cru-
cial for correct measurements due to the electrical conductivity and the
permeability of the specimen materials. The coefficients of the adapted
calibration plot are stored in the active Application (measurement appli-
cation memory).
Offline
State of a peripheral device (printer or PC) connected to the instrument
that does not allow it to receive data.
Open Block
Group of readings for which the block has not yet been closed.
USL
The upper specification limit (USL) is the highest reading allowed at the
measurement location.
Parity)
An error check method for data transmission, where the cross-sum of all
error-free transmitted bit groups must always be even or odd. During data
transfer, the parity bits are linked to the data bits of each character or byte
to be transferred. In every word, this bit is set such that the sum of the
Ones of a byte are always an even or odd number. This corresponds to an
even or odd parity. The type of parity requirement must be defined prior
to data transfer. By checking the parity, the recipient can determine if bit
transfer errors occurred.
Precision
Agreement between the individual measurement results under precisely
defined test conditions; the precision is comprised of reproducibility and
repeatability precision.
Quality Assurance
All measures taken in a plant that are concerned with ensuring that a con-
trolled production within established quality requirements can take place.
One partial aspect of it is quality monitoring of which coating thickness
measurement is a part.
R
The range R equals the difference between the highest reading (Maxi-
mum) dmax and the lowest reading (Minimum) dmin of a measurement se-
ries.
R = dmax - dmin
Glossary
Mean range of all block ranges.
R.: Mean range
R. = σ̂ ⋅ d 2
σ̂ : Estimated standard deviation σ of the population
d2: Factor, depends on the random sample size, can be obtained
from popular published tables.
Range R
Reference Area
A defined partial area of the specimen surface with a known coating thick-
ness.
Trueness
Agreement between the “true” value and the mean value of a measure-
ment result generally obtained under practical circumstances. The “true”
value is considered a value known based on mathematical theoretical ap-
proaches. Since such values are rarely available, a value traced to national
or international standards is generally assumed to be “correct”. This “cor-
rect” value is often called the “true” value.
USB Port
Interface that is used for connecting instruments with PCs, printers, USB
sticks and USB keyboards.
σ̂ SS
Estimated value for the standard deviation σ of the population. Is output
only in the final result for Applications (measurement application memo-
ries), for which automatic block creation has been enabled.
σ̂ : Estimated value of the standard deviation σ of the population
σ̂ = R.
----- R. Mean range
d2
d2: Factor, depends on the random sample size, can be obtained from
popular published tables.
s
The standard deviation s is a measure for the spread of the single readings
of a measurement series from their common mean value. It is equal to the
mean square deviation of the single readings from the mean value and is
calculated in the following manner:
s: Standard deviation
n
∑
1 2 d.: Mean value across all single readings
s = ------------ ⋅ ( d. – d i )
n–1
i=1 n: Number of single readings
d i: Single readings
The following figure points out that two very different measurement
series can have different standard deviations even with the same mean
value.
sa
Spread of the mean values of various groups (“Blocks” Block), corrected
with regard to the spread of the single readings. To be able to calculate sa,
the spread of the group mean values SII must be significantly greater than
the spread of the single values SI within the group.
If, for example, the same number of measurements per measurement spot
is performed at several measurement spots, and the measurements per
spot are combined in one group (block), then SI is a measure for the instru-
Glossary
spread.
F beo – 1
sa = S I ⋅ ------------------
-
n Bl
2
S II
- (
F beo = -------
2
Variance-analytical evaluation)
SI
Skewness
Measure for the asymmetry of a single-peak probability distribution
around its mean value. A positive skewness indicates a distribution with a
peak that stretches more toward values that are greater than the mean val-
ue. A negative skewness indicates a distribution with a peak that stretches
more toward values that are smaller than the mean value. The skewness
for a symmetric distribution (normal distribution) is zero.
Interface
Transfer or connecting point between components, circuits or programs.
A data exchange is carried out via the interface. With serial interfaces, the
data are transferred in individual bits (i.e., one bit after another), with par-
allel interfaces, several bits are transferred simultaneously.
Probe
Transducer that delivers an electrical signal to the instrument based on a
particular measurement method. This signal is proportional to the coating
thickness and is converted in the instrument into a corresponding coating
thickness value according to the master characteristic and the normaliza-
tion and calibration coefficients of the open Application (measurement
application memory). All probes of Helmut Fischer GmbH have been
master-calibrated in the plant.
Probe Frequency
A generator feeds the measurement probe with an alternating current of a
certain frequency. The applied frequency is determined by the measuring
application and the probe type.
Range R
Standard Deviation s
Start Bit
With asynchronous serial data transfer, a start bit is transmitted before the
data word to be transferred. With the logic One to logic Zero transition of
the start bit, the receiver can be synchronized to the subsequent data bits.
Statistics
The result of a measurement series, i.e., the compilation of a large number
of single readings into a few characteristic quantities (e.g., mean value,
standard deviation, etc.)
Statistical Evaluation
Calculation using the measurement data according to statistical mathe-
matical methods.
Glossary
Some parts of the population. In practical applications, a small part of a
production batch taken from the production according to random sam-
pling methods; the results of the random sample are extrapolated to the en-
tire batch (lot, production unit).
Stop Bit
With asynchronous serial data transfer, the stop bit is added to the data
word to be transferred. 1 to 2 bit logic Ones are used. After the stop bit,
the transmitter remains at logic One until the start bit of the next character
arrives.
Student Factor t
Sum Frequency
The sum frequency is that portion of parts (in percent), where the coating
thickness is smaller or equal to a particular measurement. In a sum fre-
quency chart, the sum frequency can be viewed referenced to the coating
thickness. Example: One realizes that 9% of the parts exhibit a coating
thickness of less than or equal to 39 µm (1.56 mils).
t
The student factor t can be obtained from popular published tables (e.g.,
Graf, Henning, Stange, Wilrich: Formeln und Tabellen der angewandten
mathematischen Statistik [Formulas and Tables in Applied Mathematical
Statistics by Graf, Henning, Stange and Wilrich]; Springer-Verlag) and is
stated as follows:
t⎛ α Example for a confidence level of 95 % and n > 200 (and thus, de-
1 – ---⎞ ;f
⎝ 2⎠ gree of freedom 199, since f = n - 1) the student factor is
t97.5; 199 = 1.96.
Part Specimen
u
Every instrument is subject to random measurement errors
( Accuracy). With a certain probability (the confidence level), the pre-
sumed “true” value (μ) of the measured quantity lies within an interval
around the measured mean value d. of a measurement series. The interval
is also refereed to as confidence interval. The boundaries of this interval
are at a distance u, the measurement uncertainty, from the mean value μ.
d. – u ≤ μ ≤ d. + u
LSL
Lower specification limit; is the smallest reading allowed at the measure-
ment location.
u-Scale
Scale on the right ordinate in the printout of the sum frequency chart. Lin-
ear transformation of the measurements into standardized features u. The
transformation serves comparison and analysis purposes. The standard-
ized feature values are without dimension; their arithmetic mean u is Zero
and their standard deviation σ(u) is always 1.
V
Coefficient of variation. The spread of a measurement series in percent,
i.e., the standard deviation in reference to the mean value. V [%] is a char-
acteristic process constant. A sudden change in V [%] indicates a change
in the process conditions. V is calculated in the following manner:
s V: Coefficient of variation
V = ---- ⋅ 100 [ % ]
d. s: Standard deviation
d. Mean value
Vˆ VDach
Estimated value of the coefficient of variation.
σ̂ Vˆ : Estimated value of the coefficient of variation
Vˆ = ------ ⋅ 100 [ % ]
d.. σ̂ : Estimated value of the standard deviation σ of the popula-
tion
d.. Mean value across the block mean values
Variance
Mean squared deviation. The square root of the variance is called standard
deviation (s).
n s2: Variance
∑
2 1 2
s = ------------ ⋅ ( d i – d. ) d.: Mean value of the single readings
n–1
i=1
d i: Single readings
n Number of measurements
Variance-Analytical Evaluation
Statistical method for checking the mean values of various random sam-
ples to determine, whether they are comparable or exhibit significant dif-
ferences. The spreads of the group mean values is compared to the mean
spread of the single readings within the groups.
The check value Fbeo is determined ( sa) and based on a comparison
k
SI2: Mean value of the squared group spreads sj2
∑ sj
2 1 2
S I = ---
k k: Number of random samples
j=1
SII2: Squared spread of the group mean values
k
∑ ( xj – x )
2 n 2 x j: Group mean value of the group
S II = -----------
k–1 n Number of single readings per group
j=1
x Mean value of the group mean values
k
∑
1
with: x = ---
k
xj
j=1
2
S II
F beo = -------
- Fbeo Check value for deciding the answer to the question
2
SI (in the instrument possibly designated with Fb)
If the condition Fbeo ≤ FTab is met, the mean values of the random samples
belong to a common population. If Fbeo > FTab, then the mean values are
significantly different. The characteristic value sa ( sa) states the
spread of the mean values corrected with regard to the spread of the single
readings.
FTab Table value of the F-distribution with Ff1, f2, 1- α
f Degrees of freedom with f1 = (k-1) and f2 = k (n-1)
α Significance level
Coefficient of Variation V
Reproducibility
Term for the differences of the individual measurement results under re-
producibility conditions. Reproducibility conditions refer to measure-
ments on a specimen according to a specified method, e.g., at different
times or with different instruments or with different observers or at differ-
ent locations. Measurement results that have been obtained by different
persons using different instruments at different locations on the identical
specimen must be comparable. The reproducibility is the basis for com-
puting the confidence interval for the expected value.
Confidence Interval u
Glossary
An area of the sum frequency chart, where the sum plot can be found.
With 95% certainty (confidence level), the true portion following below
the respective feature value (e.g., coating thickness) can be found within
these borders. For a normal distribution, the confidence borders ptop and
pbottom are calculated as follows:
⎛ ⎛ d-------------
– d.⎞ 2⎞
d – d. ⎜ 1 ⎝ σ̂ ⎠ ⎟
p unten = ------------- – ⎜ 1,96 ⋅ --- + ----------------------⎟
σ̂ ⎜ n 2n – 2 ⎟
⎝ ⎠
⎛ ⎛ d-------------
– d.⎞ 2⎞
d – d. ⎜ 1 ⎝ σ̂ ⎠ ⎟
p oben = ------------- + ⎜ 1,96 ⋅ --- + ----------------------⎟
σ̂ ⎜ n 2n – 2 ⎟
⎝ ⎠
⎛ 1 ( y p ( i ) ) -⎞
2
p oben ( i ) = y p ( i ) + ⎜ 1,96 ⋅ --- + ------------------ ⎟
⎝ n 2n – 2 ⎠
Confidence Level u
Repeatability Precision
Repeated measuring under consistent conditions at the same measurement
location leads to random deviations of the measurement data. Consistent
conditions means the same observer performing measurements according
to a specified method on an identical specimen within short periods using
the same instrument at the same location. The standard deviation of the
measurement data obtained under repeatability conditions is a measure for
the repeatability precision. A large standard deviation or measurement un-
certainty of an instrument indicates a poor repeatability precision. The
smaller the standard deviation, the better the repeatability precision. The
repeatability precision is dependent on the measurement method and the
properties of the instrument but also on the properties of the specimen.
The repeatability precision can be improved by generating mean values of
the measurement series (i single readings).
d base material
Glossary
The curvature is a measure for how pointed (Excess) or how flat (Kurto-
sis) a distribution is compared to a normal distribution. A positive curva-
ture is an indication for a relatively narrow and pointed distribution. A
negative curvature indicates a relatively flat and wide distribution. The
curvature for a normal distribution is Zero.
X Count Rate
XN Count Rate
Count Rate
X (Phi). Probe signal displayed as a number of electrical impulses. Phi
corresponds directly to the measured quantity. The Phi values range be-
tween the two extremes Xmin and Xmax.
In general, displaying the count rate X serves the purpose of determining
whether a noticeable measurement effect is present for a particular mea-
suring application.
The numeric values for the normalized count rate XN are between 0
and 1 and are calculated according to the following equation:
V
v 245
Variance 245
Variance Analytical Evaluation
245
Voltage supply 28
W
Warning messages 215
Warrantee 15
Weight 28
X
Xn 107
DKD-D-33101
Accredited acc. to
DIN EN ISO/IEC 17025
SWITZERLAND
HELMUT FISCHER AG
Moosmattstrasse 1 · Postfach · CH-6331 Hünenberg
: (+41) 41 785 08 00 · Fax: (+41) 41 785 08 01
switzerland@helmutfischer.com
Valid for Helmut Fischer AG and
www.helmutfischer.com Branch Offices