CERTIFICATE OF CALIBRATION
uxccezer22000000884°
ISSUE DATEag/o1/2022
CERTIFICATE NO. : PC - 01-22 / ATL / 0069-01
CUSTOMER : M/s. ON LOAD GEARS., CAL DATE: 07/01/2022
ADDRESS: UNIT- IL AMBATTUR CHENNAI.
CUST REF — : DC.NO: 70481 , DATED: 06/01/2022
DEPT : MECHANICAL LAB
EXTERNAL MICROMETER
DATE OF RECEIPT
07/01/2022 ‘TEMPERATURE : 203°C
PHYSICAL CONDITION : SATISFACTORY HUMIDITY:
INSTRUMENT DETAILS
MAKE MITUTOYO TYPE: ANALOG
RANGE =: 25-50mm SL.NO. 67056285
Le 0.01 mm INO. OMM- 04
‘STANDARD USED ‘SLNO/ID.NO. CERT.NO ‘VALIDITY
[SLIP GAUGE SET PCCPL/SG -10 KCP/01/21-22/8229, (02/12/2023
[OPTICAL FLATS PCCPL/OF/O% TRPM-CCTR-DIN-2021-0472, 24/01/2023
REFERENCE STANDARD: __1S:2967-1983 CALIBRATION PROCEDURE: WI-M-EMM-02(L)
CALIBRATION RESULTS : (ALL VALUES in mm)
REFERENCE | MEASURED | OBSERVED
SLNO. Wait vail aoe! FLATNESS ( Opto flat checking )
T 25.0 25.000 (i 0.000 [SPINDLE 2 FRINGES
2 275 27.500 0.000 JANvIL 1 FRINGES
3 301 30.100 ‘0.000 [PARALLELISM 2_ FRINGES
4 327 32.700 0.000
5 35.3 35.300 0.000
6 379 37.900 0.000
7 400 40.000 0.000
8 426 42599
9 452 45.199
10 478 47.799
ul 50.0 49.999
PERMISSIBLE ERROR AS PER IS 0-50 mm-#4 ym
UNCERTAINTY OF MEASUREMENT = 26.3 un,
‘THE REPORTED EXPANDED MEASUREMENT UNCERTAINITY 1S STATED AS THE COMBINED STANDARD MEASUREMENT
UNCERTAINITY MULTIPLIED BY THE COVERAGE FACTOR K=2 SUCH THAT THE COVERAGE PROBABILITY CORRESPONDS
‘TO APPROXIMATELY 95%.
++ END OF CERTIFICATE ***