Professional Documents
Culture Documents
Influence of Grid Control and Object Detection On Radiation Exposure and Image Quality Using Mobile C-Arms - First Results
Influence of Grid Control and Object Detection On Radiation Exposure and Image Quality Using Mobile C-Arms - First Results
1
Institute Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Leipzig AöR
2
Ziehm Imaging GmbH, Nürnberg
Gosch D et al. Influence of Grid… Fortschr Röntgenstr 2007; 179: 896 – 900
Rapid Communication 897
unabhängig von der Objektpositionierung eine optimale Belich- obtained independently of object positioning. Object detection
tung erreicht wird. Die Objekterkennung kann zu einer Reduk- may lead to a reduction in radiation exposure for the patient
tion der Strahlenexposition von Patient und Untersuchungsper- and operating staff.
sonal führen.
Gosch D et al. Influence of Grid… Fortschr Röntgenstr 2007; 179: 896 – 900
898 Rapid Communication
linear relationship between the number of pulses and dose area Table 2 Dose reduction
product was checked experimentally.
country dose reduction (in %)
To quantitatively assess the dose reduction that can be achieved
with object detection, examination data from various C-arms average minimum maximum standard
used in 5 different countries were collected. The units were value value value deviation
used in a wide variety of OR applications, including orthopedic, China 22 17 25 3
traumatology, and vascular and neurosurgical procedures. The Denmark 30 17 35 6
selection of C-arms and procedures occurred accidentally. Only England 26 14 57 11
data from examinations with a pulse frequency of 25 pulses/sec Germany 19 1 60 13
and object detection were evaluated. Scotland 29 12 41 10
all 21 1 60 13
Results
! distribution of patients and shows the average, minimum, and
●" Fig. 1a shows an image of a skull phantom. The measuring maximum values as well as the standard deviation of the cor-
circle is only partially covered by the object, whereas the re- responding dose area product values. ● " Table 2 summarizes
maining area receives unattenuated useful radiation. Conse- the dose reduction that was achieved. ● " Fig. 3 is a graphical
quently, the exposure rate control system sets insufficient representation of the frequency distribution of the dose reduc-
fluoroscopy values, and the object is clearly underexposed. tion.
The effect of grid control on image quality is visible in
●" Fig. 1b, which shows the same object, however with grid
control. In this case the skull is correctly exposed, although po- Discussion
sitioned off-center. Exposure control was based solely on the !
measuring cells represented in the image. The measuring circle The use of mobile C-arms for intraoperative imaging has in-
and measuring cells are normally not displayed. They are creased over the last few years as a result of minimally invasive
drawn in the figures as an addition. An example from clinical procedures [7]. To achieve optimum image quality regardless
practice is illustrated in ●
" Fig. 2, which shows a foot in off-cen- of the object to be screened, the C-arm must provide automatic
ter position captured with grid control. The image quality is exposure rate control. In Germany, automatic exposure rate
good, although the object is positioned on the periphery of the control is required by law for fluoroscopic X-ray examination
FOV. of humans (Art. 26 of the German X-Ray Ordinance) [8]. In the
The possible reduction of radiation exposure resulting from event of inadequate positioning, especially of small objects that
object detection and gradual pulse rate reduction in static ima- only partially cover the image intensifier's FOV, the use of con-
ges was examined on the basis of a total of 146 patients from 5 ventional exposure rate control systems with a central measur-
different countries. ● " Table 1 breaks down the geographical ing circle may result in bad image quality due to an insufficient
Gosch D et al. Influence of Grid… Fortschr Röntgenstr 2007; 179: 896 – 900
Rapid Communication 899
Gosch D et al. Influence of Grid… Fortschr Röntgenstr 2007; 179: 896 – 900
900 Rapid Communication
The generalizability of the results of the present study is limited 4 Rampersaud YR, Foley KT, Shen AC et al. Radiation exposure to the spine
surgeon during fluoroscopically assisted pedicle screw insertion.
due to the use of a static phantom and the small patient sample
Spine 2000; 25: 2637–2645
size. Further investigations using dynamic phantoms and larger 5 Theocharopoulos N, Perisinakis K, Damilakis J et al. Occupational expo-
patient groups are needed to confirm the first results regarding sure from common fluoroscopic projections used in orthopaedic sur-
image exposure optimization and dose reduction as a result of gery. J Bone Joint Surg Am 2003; 85: 1698–1703
grid control and automatic object detection. In addition, possi- 6 Lewall DB, Riley P, Hassoon AA et al. A fluoroscopy credentialling
programme for orthopaedic surgeons. J Bone Joint Surg 1995; 77:
ble dose reduction depending on the type of examination and
442–444
selected pulse rate should be explored. 7 Fuchs M, Modler H, Schmidt A et al. Messung der intraoperativen Strah-
Our study has shown that the grid control implemented on lenexposition des Unfallchirurgen. Unfallchirurg 1999; 102: 371–376
Ziehm Vision mobile C-arms is able to produce good image 8 Röntgenverordnung. Köln Berlin Bonn München: Carl Heymanns Ver-
quality (correct exposure) regardless of the position of the sta- lag KG, 2002
9 Herrmann K, Helmberger T, Waggershausen T et al. Erste Erfahrungen
tic object under examination, thus facilitating operation of the
mit gepulster Durchleuchtung an einer multifunktionalen Durch-
unit. Object detection and automatic pulse rate reduction ap- leuchtungsanlage. Fortschr Röntgenstr 1996; 165: 475–479
plied to static objects in the pulsed fluoroscopy mode using 10 Alonso JA, Shaw DL, Maxwell A et al. Scattered radiation during fixation
25 pulses/sec yielded an average dose reduction of 21 % in a of the hip fractures. J Bone Joint Surg 2001; 83: 815–818
11 von Boetticher H, Lachmund J, Hoffmann W et al. Optimierung des
total of 146 examinations.
Strahlenschutzes für das Personal in der Radiologie auf der Grundlage
der effektiven Dosis. Fortschr Röntgenstr 2006; 178: 287–291
References 12 Eder H, Panzer W, Schöfer H. Ist der Bleigleichwert zur Beurteilung der
1 Wieners G, Pech M, Beck A et al. Vergleich von Strahlenexposition und Schutzwirkung bleifreier Röntgenschutzkleidung geeignet? Fortschr
Gosch D et al. Influence of Grid… Fortschr Röntgenstr 2007; 179: 896 – 900