You are on page 1of 7

Solar Energy Vol. 35. No. 2, pp. 167-173, 1985 0038-092X/85 $3.00 + .

00
Printed in the U.S.A. ~ 1985 Pergamon Press Ltd.

DIRECT MEASUREMENT AND ANALYSIS OF


THERMOSIPHON FLOW~

A. I. KUDISH. P. SANTAMAURA-~and P. BEAUFORT§


Solar Energy Laboratory, Department of Chemical Engineering, Ben-Gurion University of The
Negev, Beer-Sheva, Israel

(Received 31 May 1984; accepted 11 January 1985)


Abstract--The thermosiphon flow rate has been measured directly by adapting a simple and well-
known laboratory technique, a constant level device, to a solar collector operating in the thermosiphon
mode. The measurements were performed as a function of thermosiphon head and inlet temperature.
The thermosiphon flow rate data were correlated both with the temperature change across the solar
collector (A T) and the global insolation rate. The minimum A T threshold value, necessary to initiate
thermosiphon flow in the morning was determined as a function of thermosiphon head and the cor-
responding thermosiphon pressure heads were observed to be invariable. The thermosiphon flow data
were utilized to construct a standard efficiency test curve, thus showing that this technique can be
applied for testing collectors in the thermosiphon mode. The instantaneous collector efficiency was
also determined as a function of time of day.

1. INTRODUCTION them. We have isolated the solar collector and have


developed a relatively simple technique for direct
Thermosiphon solar D H W systems are the most
measurement of thermosiphon flow.
widely used of all man-made solar energy thermal
convension devices, the reason being that they are
cost competitive with the alternative conventional 2. DESCRIPTION OF EXPERIMENTAL SYSTEM
energy D H W systems available in many regions of
The experimental technique used in this study
the world. There are today in excess of 600,000 such
differed from those used previously in that the ther-
units operating in Israel[l], which are capable of
mosiphon flow rate was measured directly. This
supplying, in most cases, about 70% of a consum-
was possible since we isolated the solar collector
er's annual D H W requirement.
by eliminating the storage tank. Thermosiphon flow
The performance of thermosiphon systems has
was simulated by means of a constant level device
been investigated extensively, both experimentally
attached to the inlet pipe of the solar collector (Fig.
and analytically, by numerous researchers[2-12]. A
1). The constant level device consisted of an outer
comprehensive review on this subject has been pub-
cylinder (height 30 cm, diameter 14 cm) open on
lished recently[13]. In particular, the flow rate of
top to the atmosphere and closed at the bottom. The
the heat exchange fluid, almost exclusively water,
bottom plate had two holes drilled for (1) a ~ in.
has intrigued many investigators. Due to its nature,
fitting flush with its outer surface for connecting it
viz., a very low pressure head, thermosiphon flow
to the collector inlet pipe, and (2) an adjustable fit-
is amenable only with difficulty to measurement by
ting for insertion of the internal drain pipe (2 in.).
the standard laboratory flow rate measuring de-
Soldered onto the cylinder surface 20 cm above the
vices. It can be measured accurately only by those
bottom plate was a a in. inlet tube for mains (make-
techniques that do not add any flow resistance to
up) water. The latter was maintained in all cases
the system. Thermosiphon flow has been measured
- 5 cm below the water surface in the cylinder so
previously by indirect techniques such as dye
that the kinetic energy or the make-up water would
injection[4], thermal dissipation tracing[7], laser
be dissipated radially. The water level in the con-
Doppler anemometer[8, 9] and rate of change of
stant level device was set equal to that in the outlet
storage tank temperatures[12].
pipe from the collector (i.e. h3) by fine adjustment
In all the previously reported studies, thermo-
of the height of the internal drain pipe. The flow
siphon flow was investigated on a complete D H W
rate of the mains water entering the constant level
system comprised of solar collector(s), storage tank
device was adjusted to exceed (by - 5 0 % ) the max-
and the auxiliary plumbing connections between
imum measured thermosiphon flow rate, thereby
ensuring a constant water level on the inlet side of
+ Presented, in part, ASES Annual Meeting, Minne- the system. Thus, the hydrostatic pressure across
apolis, MN June (1983). the solar collector was maintained equal to zero and
Current address: 181 Dunvegan Rd.. Toronto, On-
tario, Canada. any flow measured was caused by the thermosiphon
§ Current address: Gen. Dibbetslaan 83, 5623 JM Eind- pressure head only. The solar collector in this study
hoven, The Netherlands. operated in a single-pass mode, which is typical of

167
168 A. I. KUDISH et al.

CONSTANTLEVEL
B

TOUT

h2 h3

TIN ~ - - F
h~
t
Fig. I. Schematic diagram of experimental simulated thermosiphon system.

most Israeli thermosiphon D H W systems. A some- flow rate. The rate of thermosiphon flow was meas-
what different application of such a single-pass ured periodically throughout the day during the
thermosiphon system has been reported re- sunshine hours.
cently[14]. The threshold AT (i.e. Tout - Ti,) values nec-
The direct measurement of the thermosiphon essary for the initiation of thermosiphon flow in the
flow rate was accomplished by simply collecting the morning, corresponding to the different thermo-
solar heated water effluent from the solar collector siphon heads studied, were also determined. This
in a graduated cylinder during a prescribed time in- was achieved by continuously monitoring the inlet
terval (usually 1 min.). The inlet and outlet water and outlet temperature to and from the solar col-
temperatures to and from the solar collector were lector. The threshold A T value corresponds to the
monitored continuously by means of type T (cop- instant when thermosiphon flow is initiated and is
per-constantan) thermocouples inserted into 4-mm evidenced graphically when the inlet temperature
copper tubing probes and connected to a multipoint drops to the main's water temperature and the out-
recorder (Chino) (see Fig. 1). The inlet and outlet let temperature rises rapidly (cf Fig. 2). The meas-
legs (viz. auxiliary plumbing) of the solar collector
were insulated with glass wool molded shells, which
were wrapped with aluminum laminate. The latter
I

prevented external heat gain via solar radiation.


The global insolation was measured at the tilt 12 I

angle of the solar collector (i.e. 30°) by means of an I

Eppley pyranometer, Model PSP, connected to a I :j ,.S:"


calibrated recorder (Cole-Parmer). The latter pro-
vided us with instantaneous insolation values nec-
essary for calculating instantaneous solar collector IC I

efficiencies. Toui
The solar collector used in this study was a stan- v

-- b

dard Israeli commercial type. It consists of eight ~- L~ 9


in.-O.D, copper tube risers connected to two l~-in.- I

O.D. copper tube leaders. The solar absorption sur- 8 ii


face being of a honeycomb structure, viz. aluminum
I

fins spaced at 4-ram intervals perpendicular to the f

riser tubes. Its overall dimensions are 220 x 95 x


10 cm (aperture area = 1.89 m 2) and it is intended - - i !

to operate in conjunction with a 120-1 storage tank. 6 I I I


The thermosiphon head was altered by raising 20 40 60 80
(lowering) both the constant level device and the
TEMPERATURE (°C)
outlet pipe by the same amount (i.e. the overall
length of the auxiliary plumbing, vertical pipes, was Fig. 2. Experimental plot of temperature vs time for Sept.
17, 1982 (h3 - h~ = 1.98 m). Arrow indicates the rapid
changed). This enabled us to study the effect of the temperature change corresponding to initiation of ther-
height of the thermosiphon head (h3 - h~) on the mosiphon flow.
Direct measurement and analysis of thermosiphon flow 169

FLOW RATE VS DELTA T H=O 7M A . F l o w rate a n d a T


200
The results for the correlation between the ther-
ca 180 ill I "_.
mosiphon flow rate and A T are shown in Fig. 3 for
,~o m m
it .,.
(h3 - h2) = 0.7 m and in Fig. 4 for ( h 3 - h 2 ) =
140 • i l
ml 0 m. The former is also representative of the data
._J 120 ~o I
f o r ( h 3 - h 2 ) = 0.35 and 1 m in that two distinct
I00 • oe •
flow rate ranges, a high and low, were observed.
LId 8C • •
<~ They corresponded to the time of day the meas-
60
urements were performed. The high range was
[g 40 •
© measured during the time period 10:00-14:00 (solar
20 •
LL time) and the flow rate was observed to be fairly
o i i
55 45 55 constant. During this time period the range of/x T
TOuT-TIN ( DEG £ ) values measured was quite narrow (see Fig. 3) and
the global insolation rate was also relatively con-
Fig. 3. T h e r m o s i p h o n f l o w r a t e vs A T f o r (h3 - h2) = 0.7
stant. The low flow rate range corresponded to
m. • 10:00-14:00; • <10:00 and >14:00.
those measurements performed either prior to 10:00
or after 14:00. During these time periods both the
urements were performed for four values of ( h 3 - range of A T values measured and the global inso-
hi), viz. 1.98, 1.68, 1.33, and 0.98 m. lation rate were changing (increasing for < 10:00 and
The experimental system was adapted, by the decreasing for >14:00). The data in this range are
addition of heating tape and powerstat, for meas- therefore scattered. The average thermosiphon
urement of the collector efficiency curve while op- flow rates in the high range are reported in Table 1
erating in the thermosiphon mode. In general, col- for ( h 3 - h 2 ) = 0 . 3 5 , 0 . 7 0 a n d 1 m .
lector efficiency tests are performed in the forced The results for ( h 3 - h 2 ) = 0 m are distinct from
circulation mode. The heating tape was wrapped the others in that there appears to be a linear cor-
around the inlet pipe (vertical section) and then in- relation between the thermosiphon flow rate and
sulated. The instantaneous efficiency was calcu- AT. The measurements in this case were all per-
lated from the inlet and outlet temperatures, in- formed during the time period 10:00-14:00. (The
stantaneous insolation and measured flow rates. reason for this will be discussed in section 4). The
The inlet temperature was varied by means of the measurements in the case of ( h 3 - h2 = 0 m were
powerstat to which the heating tape was connected. also performed as a function of the collector inlet
temperature range. The linear correlation between
thermosiphon flow rate and A T for five different Ti,
3. R E S U L T S
ranges are listed in Table 2. The relatively poorer
The measured thermosiphon flow rate data were correlation factor for the lowest reported T~, range
analyzed by attempting to correlate them to the may be explained by the fact that the corresponding
temperature increase across the collector (i.e. A T measurements were performed in the winter ( N o v . -
= Tout - Tin) and to the instantaneous global in- Dec. 1982), whereas the other four sets were meas-
solation rate, in the different thermosiphon heads ured in the summer (July-Aug. 1983). The former
studied. being less amenable to outdoor measurements (i.e.

FLOW RATE VS DELTA T H=OM


I00 -

90-
80-
Z
70-

..J 60-
ill •

v 50-
LU 40-
H-
<~ El • •

135 :.'50-

20~-
0 I0
_J
ii
0 I I I
55 65 75 85

TOU T - TIN (DEG .C)


F i g . 4. T h e r m o s i p h o n f l o w r a t e vs A T f o r (h~ - h2) = 0 m .
170 A. I. KUDISH et al.

Table 1. A v e r a g e t h e r m o s i p h o n flow rate v s (h3 - h2) days are not as clear and wind gusts are much more
during the time period 10:00-14:00 frequent).

(h3 - h2) A v e r a g e t h e r m o s i p h o n flow rate


B. F l o w rate and global insolation
(m) (ml/min.m 2) The correlation between the thermosiphon flow
0.35 85.0 _+ 15.5 r a t e a n d t h e g l o b a l i n s o l a t i o n w a s o b s e r v e d to b e
0.70 157.1 _+ 16.6 l i n e a r in all c a s e s s t u d i e d . T h e r e s u l t s f o r (h3 - h2)
1. 177.3 +_ 20.6
= 0 . 7 m a r e s h o w n in Fig. 5. T h e l i n e a r c o r r e l a t i o n s
f o r all c a s e s a r e r e p o r t e d in T a b l e 3.

Table 2. Linear correlation b e t w e e n t h e r m o s i p h o n flow C. Collector efficiency and time o f day


rate and A T for ( h 3 - hD = 0 m T h e i n s t a n t a n e o u s c o l l e c t o r e f f i c i e n c y w a s cal-
culated as a function of time of day by simply ap-
T~n range Correlation plying
(°C) y = mx + b a factor

18-26 y = 2.0x - 86 0.81 "q = thCvA T/G, (1)


29-35 y = 3.6x - 150 0.90
35-39 y = 4.3x - 152 0.95 w h e r e ~q is t h e i n s t a n t a n e o u s e f f i c i e n c y ; & , t h e t h e r -
40-44 y = 4.6x - 142 0.91
48-53 y = 5.8x - 152 0.93 m o s i p h o n f l o w r a t e ( m l / m i n . m 2 ) ; Cp, t h e h e a t c a -
pacity of water (KJ/kg.°C); and G, the instantaneous
ay - t h e r m o s i p h o n flow rate (ml/min.m2); x - AT global insolation rate (KJ/min.m2). The results for
(°C) (h3 - h2) = 0.35 a r e r e p o r t e d in Fig. 6 a n d t y p i f y

Table 3. Linear correlation between t h e r m o s i p h o n flow rate and


global insolation rate

(h3 - h2) Tin Range


(m) (°C) y = mx + b a Correlation factor

1 -- y = 4.2x - 66.5 0.88


0.70 -- y = 4.1x - 74.9 0.97
0.35 -- y = 2.6x - 58.1 0.94
0 18-26 y = 2.4x - 73.4 0.81
0 29-35 y = 3.2x - 73.7 0.98
0 35-39 y = 3.2x - 68.8 0.96
0 40-44 y = 3.2x - 73.6 0.92
0 48-53 y = 4.6x - 140.3 0.88

a y - t h e r m o s i p h o n flow rate (ml/min.m2); x - global insolation rate


( K J / m i n . m 2)

FLOW RATE VS INSOLATION H=O7M


20O

180

160 .U
140
Ld
I-- 120

rY I00

8O
0 60
/
%
LL
4O

2O

0 I I I I I
20 SO 40 50 60 70

INSOLATION ( K J / M I N * M * ~2)
Fig. 5. T h e r m o s i p h o n flow rate vs global insolation rate for (h3 - h2) = 0.7 m.
Direct measurement and analysis of thermosiphon flow 171

EFFICIENCY VS TIME H =0 5 5 M

>.-
¢.D ,,,¢,, • . • • • ,~
Z
LLI 3 . /v \ . \

13_ 2 /~ ~..~
LL
LLI

0 I I I I I I, I
9 tO II 12 13 14 15 16

TIME (HR)

Fig. 6. Instantaneous collector efficiency vs time for (h~ - h2) = 0.35 m.

those for all cases. The instantaneous collector ef- ciency testing of solar collectors in the thermo-
ficiency increases with increasing thermosiphon siphon mode (in which they are intended to operate)
head [viz. ( h 3 -- hj)]. This is expected; since rh as opposed to the widely used forced circulation
increases more rapidly, then A T decreases with in- mode of testing. The results are presented graphi-
creasing thermosiphon head. Also the early morn- cally in Fig. 7 and the linear regression curve for
ing (<10:00) efficiency values are greater than the these data is
late afternoon (>14:00) values and that the latter
decrease quite rapidly. It is of interest to note that ( T i n - Tamb)
the maximum instantaneous efficiency (i.e. the "q = FR('ra)e - FRUL (2)
G
range) decreases with increasing values of Tm (see
Table 4). This is caused by the increased temper- where FR('ra)e = 0.79, FRUL = 22.4 W/m2'°C and
ature gradient (Tin - Tamb) and consequent faster the correlation factor = 0.95.
rate of heat loss.

D. Thermosiphon threshold A T values 4. A N A L Y S I S OF THE RESULTS

The minimum magnitude of A T (threshold value) A. Thermosiphon flow rates


required to initiate thermosiphon flow in the morn- The experimental system, on which the ther-
ing was measured, as described in section 2, as a mosiphon flow measurements were performed, dif-
function of the thermosiphon head. The average fered from those studied previously [2-12], the
values are reported in Table 5. These results will present one being an open system. We have
be further analyzed in section 4. adapted the analytical approach first used by
Close[2] to the conditions prevalent in our system,
E. Standard efficiency test viz. no storage tank, and it is significantly simpli-
The data from the measurement of thermosiphon fied. The thermosiphon head equation (cf eqn (11)
flow rate as a function of Tin were utilized to con- in [2]) reduces to
struct a standard efficiency curve for the case of
(h3 - h2) = 0 m. It is important to note that these hr = ½(Sin - - Sout)(h3 - hi), (3)
measurements were performed over a period of
about 5 weeks and not under the stringent restric- where S is the specific gravity. Utilizing the para-
tions (viz. climatic conditions) required for such bolic relationship between specific gravity and tem-
testing. It is our purpose to show that this experi- perature[2], rather then the linear one assumed in
mental technique can be used for standard effi-

Table 5. Threshold AT and thermosiphon pressure head


Table 4. Maximum instantaneous efficiency range as a values as a function of the thermosiphon head
function of inlet temperature range
(h3 - hi) (h3 - h2) AT pghT
Tin R a n g e (m) (m) (°C) (kg/s 2"m)
(°C) "q.... Range
1.98 1 12.7 + 0.6 63.2
29-35 0.46-0.54 1.68 0.70 15.5 _+ 0.6 59.2
35-39 0.44-0.52 1.33 0.35 19.1 +_ 0.3 59.6
40-44 0.44-0.50 0.98 0 33.3 _+ 3.5 59.3
48-53 0.41-0.49 pghr = 60.3 -+ 1.9
172 A.I. KUDISHet al.

Io[-
07!

0.50 , "~"

025-

0 I I I I 1
O[ 012 015 04 05

(Tin -Tomb)/G [°C/(KJ/ml.m,n)]


Fig. 7. Standard efficiency test curve for (h3 - h2) = 0 m.

eqn (3), we arrive at factor (i.e. the sum of the individual section friction
factors adjusted by the continuity equation, p~ VIA~
hr = Ti, - Tout (2AT - B)(h3 - hl), (4) = p2V2A2). The fluid temperature does not vary in
2 transit between the Tou, probe position and the pipe
outlet (B) due to both the insulation and its rela-
where A = - 1.25 × 10 -6, B = 5.83 x 10 -5 and tively short residence time. Thus, the value of 9B
= (Tin + Tom)/2. (The temperature units for eqn corresponds to that at T = Tou t and the value of V8
(4) are °F as per the parabolic equation fit for the is the volumetric flow rate measured divided by the
specific gravity in [2]). outlet pipe cross-sectional area. We have utilized
The thermosiphon flow through the collector the continuity equation to calculate the linear flow
was analyzed by means of applying Bernoulli's rate within the various sections of the experimental
equation between A and B, on Fig. 1, i.e. system. The temperature (density) was assumed
constant within the inlet and outlet legs of the solar
collector. The temperature within the riser was as-
PA + gZAPA + - ~ DA sumed to increase linearly and the density as per
the parabolic relationship in [2]. The friction factor,
v~ _ v~ skin and fittings, was determined for each section
= PB Jr- gZBPB "~- T OB -~ E l~'i T Pi, (5)
of the system. The values for the thermosiphon
pressure head and overall system friction factor
where P is the pressure; Z, height above reference; were inserted into eqn (7) which was then solved
p, density; V, linear velocity; and F, friction factor. for VB. The agreement between the calculated and
In our system Pa = PB, ZA = ZB, and VA = 0 (i.e. experimental values was about 10%, which is quite
in the constant level device). Thus eqn (5) reduces satisfactory considering the lack of precision in de-
to termining friction factors for any particular system.

v~ v~ B. Thermosiphon threshold values


gZ(Pa - 9n) = --~ PB + ~ F,-~- p,. (6)
A priori, it is reasonable to assume that the mag-
nitude of the thermosiphon pressure head required
The term in the left-hand side of eqn (6) is the ther- to initiate thermosiphon flow in the morning should
mosiphon pressure head which can be substituted be constant, since it must overcome what may be
for by use of eqn (4), viz. termed the system inertia.It Thus, the increase in
the threshold A T values (cf Table 5) with decrease
v~ v~ in the thermosiphon head was as expected, since
gpBh~ = T OB + (~, F;) T OR, (7)

If The change in the overall system skin friction factor


where B corresponds to the conditions at the col- with the change in the length is negligible relative to the
lector outlet, and F~ is the overall system friction overall system friction factor.
Direct measurement and analysis of thermosiphon flow 173

they should be inversely proportional if the ther- 3. The correlation between thermosiphon flow
mosiphon pressure head is, indeed, invariable. This and global insolation rate was observed to be
was verified by calculating the corresponding ther- linear for all cases studied.
mosiphon pressure heads, using the left-hand side 4. The thermosiphon pressure head required to in-
of eqn (7), and the results are presented in Table 4, itiate flow in the morning was shown to be
column 4. The standard deviation (-3.2%) is quite constant.
small considering the uncertainty involved in de- 5. The friction factor analysis of the system ver-
termining the threshold temperatures (cf. Fig. 2). ified the experimental technique.
The constant thermosiphon pressure head is evi-
dence of the accuracy of our technique of insuring
that the fluid levels (inlet and outlet) were equal (i.e. Acknowledgement--The authors wish to express their
zero hydrostatic pressure head) prior to measuring gratitude to the International Association for the Ex-
the thermosiphon flow rate. change of Students for Technical Experience (IAESTE)
for enabling two of us (P.S. and P.B.) to participate in this
research project.
C. Solar collector efficiency
The increase in efficiency with increasing ther-
mosiphon head is an important consideration when
dealing with single-pass systems. In practice, while REFERENCES
there is a minimum thermosiphon head necessary
to prevent reverse thermosiphon flow during the 1. Y. Nowarski, Israel Ministry of Energy and Infra-
structure, Department of Energy Conservation, pri-
night, the thermosiphon head is generally deter- vate communication. 1984.
mined by where the system is positioned and its 2. D. J. Close, Performance of solar water heaters with
aesthetic and architectural interaction with its sur- natural circulation. Solar Energy 6, 33 (1962).
roundings. 3. G. L. Gupta and H. P. Garg, System design in solar
water heaters with natural circulation. Solar Energy
The commencement of thermosiphonic flow is 12, 163 (1968).
also a function of thermosiphon head, in that the 4. K. S. Ong, A finite-difference method ot evaluate the
greater (other parameters being constant) the ear- thermal performance of a solar water heater. Solar
lier the onset of thermosiphon flow and consequent Energy 16, 137 (1974).
5. K. S. Ong, An improved computer program for the
solar energy conversion. This is because the greater
thermal performance of a solar water heater. Solar
the thermosiphon head the lower the threshold A T Energy 18, 183 (1976).
value required to overcome system inertia. Thus, 6. Y. Zvirin, A. Shitzer and H. Grossman, The natural
both system efficiency and total daily hours of op- circulation solar heater models with linear and non-
eration are proportional to the thermosiphon head. linear temperature distribution. Int. J. Heat Mass
Transfer 20, 999 (1977).
This explains why the onset of flow in the case of 7. A. Shitzer, D. Kalmanoviz, Y. Zvirin and G. Gross-
(h3 - h2) = 0 was relatively late, i.e. after 10:00. man, Experiments with a fiat plate solar water heating
system in thermosiphone flow. Solar Energy 22, 27
(1979).
5. CONCLUSIONS 8. G. L. Morrison and D. B. J. Ranatunga, Transient
response of thermosiphon solar collectors. Solar En-
1. We have shown how a very simple and well- ergy 24, 55 (1980).
known laboratory technique can be adapted to 9. G. L. Morrison and D. B. J. Ranatunga, Thermo-
measure the rate of thermosiphon flow directly siphon circulation in solar collectors. Solar Energy
and thereby be used to perform standard effi- 24, 191 (1980).
10. B. J. Huang, Similarity theory of solar water heater
ciency tests on solar collectors operating in the with natural circulation. Solar Energy 25, 105 (1984).
thermosiphon mode. 11. M. S. Sodha and G. N. Tiwari, Analysis of natural
2. The thermosiphon flow rate was found to sta- circulations solar water heating systems. Energy Con-
bilize during the mid-day hours as did the A T version and Management 21, 288 ( 1981).
values across the solar collector. The exception 12. M. F. Young and J. B. Bergquam, Performance char-
acteristics ofa thermosiphon solar domestic hot water
to this observation was in the case of (h3 - h2) system. ASME J. Solar Energy Engng. 103, 193
= 0, where the correlation between thermo- (1981).
siphon flow and A T was linear, independent of 13. B. Norton and S. D. Probert, Natural circulation
Ti, values. The latter may be a result of the solar-energy stimulated systems for heating water.
Applied Energy 11, 167 (1982).
greater sensitivity of the thermosiphon flow to 14. Y. F. Wang, Z. L. Liand and X. L. Sun. A "'once-
A T since the thermosiphon pressure head is rel- through" solar water heating system. Solar Energy 29,
atively low. 541 ~1982).

You might also like