Professional Documents
Culture Documents
charge prevents the passage of particles far smaller in dimension than the
filter pore size. Thus, whilst the filter is capable of removing some particles
smaller than the pore size the most important characteristic is that all
particles LARGER than the pore size are positively retained on the filter
surface, and lie in a plane where they can be readily examined and counted.
FIG.1. This photo of a perforated metal screen and lumps of clay illustrates both
the absolute surface retention and the integral structure of a Millipore filter.
TABLE I
Specification of membrane filters
~
8.0 p m k l . 4 pm 950 55
SM 5.0 p m k f .2 pm 560 35
ss 3.0 p m k 0 . 9 pm 400 20
RA 1.2 p m k 0 . 3 pm 300 14
AA 0.8 pmk0.05 pm 220 9.8
DA 0.65 p m k 0 . 0 3 pm 175 8.0
HA 0.45 pmk0.02 pm 65 4.9
PH 0.30 pmk0.02 pm 40 3.7
GS 0.22 p m k 0 . 0 2 pm 22 2.5
vc 100nm+8nm 3.0 1.0
VM 50nmk3nm 1.5 0.7
VF 25 n m k 2 n m 0.5 0.3
t ml/min./sq. cm
f litre/min./sq. cm
2. Porosity
T h e pores in membrane filters are extraordinarily uniform in size. For
example the total range of pore size distribution in a type HA, 0.45 pm,
filter (Millipore) is +0.02 of a micro-meter. Each square centimetre of
filter surface contains millions of capillary pores and the pore volume occu-
pies approximately 80% of the total filter volume (Fig. 2). High porosity
results in flow rates which are of the order of 40 times faster than flow rates
through depth filters approaching the same particle siye retention capability.
For example, water will flow through a 0.45 pm, HA, filter at 38,400 litres/h/
square metre, whilst a depth filter, with the same nominal retention efficiency,
208 J . G . MULVANY
has a flow rate of only 750 litres/h/square metre of filter area (Osgood, 1967)
under the same conditions.
3 . Chemical properties
Millipore filters exhibit typical properties of cellulose esters with respect
to chemical resistance. They are not attacked by water, dilute acids and
alkalis, aliphatic or aromatic hydrocarbons, halogenated hydrocarbons or
non-polar liquids; in fact, few fluids of interest to microbiologists will
attack them. If a chemical compatibility problem exists, solvent resistant
membrane filters are available in several pore sizes.
Since the membrane filter is free of biologically inhibitory factors it
provides an optimum collection environment for micro-organisms. Any
bacteriostatic agents present in the suspending fluid may be readily flushed
free of the filter. There are no residual factors in membrane filters (Millipore)
to block enzymatic systems or to interfere with proper organism identification
by staining or culturing procedures.
4. Thermal stability
A membrane filter, made from cellulose esters, is stable in a dry state at
temperatures up to 125°C in the presence of oxygen. Gradual decomposition
takes place in air at temperatures above 125°C. They may be used in oxygen
free atmospheres at temperatures up to 200°C. Membrane filters may be
autoclaved satisfactorily at 121°C (15 p.s.i.) either individually for a period
of 15 min. or if assembled in a filter holder, for a period which is specified
for that particular holder.
5 . Colour
White membrane filters (Millipore) are available in all twelve pore size
types. Types AA (0.8 pm) and HA (0.45 pm), are also formulated in
VII. MEMBRANE FILTER TECHNIQUES 209
FIG.3. Cellulose ester Millipore filters become transparent when their pores are
filled with a fluid of matching refractive index.
the Millipore filter is 1.51 (see Fig. 3). Black membrane filters cannot be ren-
dered transparent. Selected types of filters are available with imprinted
grid marked surfaces thus facilitating statistical counts of particles or
colonies on the filter sui face.
6 . Other properties
T h e filter is an integral structure containing no fibres or particles which
can work loose to contaminate a filtrate. It produces no ionic reaction with
compatible fluids. If properly supported, Millipore filters will withstand
at least 10,000 p.s.i. differential pressure without significant distortion of
the pore structure. They require no treatment prior to use and are disposable
after use.
210 J. G . MULVANY
*8
-7
$ a 6
._
n
E
E
lc -5
d.
\
u
0
\
E .4
-
9
n
.3
.2
*I
-35 *40 *45 *50
Pore diameter - microns
FIG.4. This pore size distribution curve for the Type HA Millipore filter is
derived from high-pressure mercury intrusion test data. Ordinate shows distribution
of pores.
shows a typical result for the type HA (0.45 pm) filter. Pore sizes in the
sample vary from 0.43 pm to 0.47 p m with the mean pore peaking at
0.45 ,urn.
VII. MEMBRANE FILTER TECHNIQUES 21 1
Filtration tests are also run with contaminants of known type and size
in which the filtrate is analysed to determine the largest particle size which
can penetrate the filter.
effects of the antibiotic and in many cases the results of such tests were
meaningless. Another inherent advantage of the membrane filter technique
is now immediately obvious, in that there is a large saving on nutrient
medium as gross dilutions are not necessary.
Fortunately, new media and growth conditions are not required for culture
on membrane filters. T h e filter, bearing organisms on its surface, may be
placed directly on conventional solid media (Fig, 5) or placed on a pad
2. Apparatus
A variety of filtration apparatus is available for microbiological analysis,
however, the basic standard tool for many years has been the Pyrex Filter
Holder (Millipore) as shown in Fig. 7. This holder employs a 47 mm dia-
meter filter for liquid filtration in conjunction with a vacuum flask. Aseptic
assembly of filter and filter holder after autoclaving is normal practice, and
apparatus should be wrapped and autoclaved in the conventional manner.
214 J. G . MULVANY
is inserted in a cover port to vent the sterifil holder with sterile air (see Fig. 9).
Ports not being used are covered with gum rubber caps. The Sterifil may
be autoclaved with the filter in place; and following sample filtration may
be filled with nutrient medium and incubated with the filter in place. This
is a significant advantage in sterility testing as there is no possibility of
FIG.8. A vented cover must be placed on the funnel when autoclavingan assembled
filter holder and filter to avoid damage to the filter.
apparatus are used. Field Monitors (Fig. 15) are disposable plastic filter
holders with a membrane filter tightly sealed between the top and bottom
halves. A cellulose absorbent pad is placed beneath the filter to support it,
and, if required, to act as a reservoir for nutrient media after filtration.
Each half of the monitor is equipped with a port, covered (when not in use)
by a plastic cap. Micro-organisms will be retained on the surface of the
filter from fluids passing through and after filtration the monitor can be
capped and transported to the laboratory for examination. Or, a nutrient
solution may be introduced beneath the membrane filter by thoroughly
VII. MEMBRANE FILTER TECHNIQUES 217
FIG.10. With the cover lifted and the ports capped, test samples of biological
solutions may be poured aseptically into the Sterifil funnel.
FIG.12. After sample filtration and the introduction of broth media, the Sterifil
Holder may be incubated with the filter in place. The Swinnex vent may be left in
place if desired.
FIG.13. The covered Receiver Flask, with both connections capped, may be
used to store sterilized effluents following filtration. Side arm connections may be
used as pouring spouts.
VII. MEMBRANE FILTER TECHNIQUES 219
soaking the cellulose pad, and then the monitor may be incubated in the
conventional manner to produce visible and identifiable colonies. T h e
“Clinical Monitor” version of the “Field Monitor” incorporates a ring
between its top and bottom halves to hold the membrane filter during
FIG.14. Millipore filter wet with water to show hydrophobic edge area.
3. Culturing
T h e membrane filter technique is readily adaptable to conventional
methods of culturing. I n sterility testing of antibiotics for example, the
membrane filter is submerged in liquid thioglycollate broth for incubation
of anaerobes and aerobes. A very common technique, as mentioned above,
is to place an absorbent pad in a 47 mm plastic Petri dish and to saturate
the pad with 1.8 ml of broth medium. T h e membrane filter can then be
placed on top of this pad and the liquid medium will diffuse through the
pores of the filter so providing organisms trapped on the surface with nutri-
ent. After incubation, the top of the Petri dish is removed and the colonies
are counted. If colonies are numerous and small, it is helpful to increase
visual contrast by “shadow casting” with a low angle light, or to use a
contrast stain as mentioned below. If cultured filters are to be transported
or retained for future reference, they should be placed in a plastic Petri
dish on a pad saturated with a preserving solution of 15% glycerine, 50%
formaldehyde, 35% water. T h e dishes and filters may then be safety trans-
ported in this “wet” state, thus preserving colony morphology; or the
filters may be dried after ten minutes for storage in cellophane envelopes.
4. Staining
Since most diagnostic staining procedures decrease viability, it is prefer-
able to pick colonies for subculture prior to staining. When using the two
contrast stains described below, however, colonies may be picked up to
15 min. after staining. Using an agar plate culture this technique is not
possible.
In all staining methods it is important that solutions of dye, buffer,
rinses, etc., to be passed through the sample filter, should be first filtered
through an 0.45 ,urn filter. This step ensures that all particles larger than
0.45 p m are removed and thus will not obscure any organisms collected
on the surface of the filter.
(a) StainingJilter to contrast with colonies. A 0.01% w/v solution of Mala-
chite Green Oxalate in distilled water is employed. After incubation and
whilst still in a Petri dish, the surface of the filter is gently flooded with
2-4 ml of dye solution (Fig. 16). After 8-10 seconds the excess dye is gently
poured off (Fig. 17), leaving colonies appearing white or yellow against a
green filter background.
(b) Staining colonies to contrast with jilter. 3 g of Methylene Blue dye are
added to 300 ml of ethyl alcohol and the solution mixed with 1 litre of dilute
potassium hydroxide (0.01% by weight). A fresh absorbent pad is placed
in a clean Petri dish and soaked with 1.8 ml of dye solution. T h e filter with
VII. MEMBRANE FILTER TECHNIQUES 22 1
FIG.16. Staining filter: Gently flood surface of filter with a solution of Malachite
Green dye.
FIG.17. Staining filter: Pour off excess dye. Colonies appearwhite or yellow against
the green filter background.
222 J. G . MULVANY
clearing is performed in the holder. After filtration the filter holder and
filter are rinsed with 100 ml of filtered phosphate buffer (pH 5.2). 5 ml of
filtered 1% Gentian Violet dye solution are then pipetted on to the filter
and a contact time of 1 min. allowed before vacuum is applied to draw the
dye through the filter. Successive rinsings with 100 ml phosphate buffer,
20 ml propanol, 20 ml propanol :xylol(1 : 1 ratio), 20 ml xylol are carried
out. Finally, the filter is placed on a 2 x 3 in. glass slide, covered with immer-
sion oil, and the organisms observed using an oil immersion objective. This
method is very rapid but does not give such good resolution as the following
method, which is more lengthy.
The technique employs a dye made by adding 60 m l 5 % (w/v) aqueous
phenol, 5 ml ethanol, 4 ml glycerol, 125 mg basic fuchsin (95% (w/v) dye
VII. MEMBRANE FILTER TECHNIQUES 223
content), and 120 mg methylene blue to 1 litre of distilled water. The dye
solution is stirred, whilst being gently heated for ten minutes and then
allowed to stand for 12 h in a refrigerator. Before use, the dye solution is
filtered through a coarse grade asbestos pad followed by a 0.45 p m (HA)
filter.
After sample filtration, the filter is left in the holder and 10 ml of dye
solution is added to the funnel and a contact time with the filter of ten
minutes is allowed. Then vacuum is applied to draw the dye through the
filter which is subsequently flushed repeatedly with filtered distilled water
until the filtrate is colourless. Finally, the filter is cleared with immersion
oil for microscopy in the same manner as described for the above Gentian
Violet stain procedure.
(d) As mentioned above, the staining techniques employed with filter
collected organisms are, in principle, the same as conventional methods.
For example the staining of acid-fast bacilli collected on the filter surface
employs a modified Ziehl-Neelsen method. T h e stains are prepared as
follows-
Fuchsin stain-3 g of basic fuchsin are added to 100 ml ethanol and stirred.
10 ml of the fuchsin solution are made-up to 100 ml with a 5% (w/v) phenol
solution in distilled water. (The basic fuchsin concentration used is only
one-third of that used in the standard method.)
Counter stain-0-3 g Methylene Blue (80% (w/v) dye content) is added to
30 ml aqueous ethanol (80% (v/v)) and 100 ml distilled water (the standard
Ziehl-Neelsen method employs an alkaline counter-stain by using 0.01yo
potassium hydroxide solution instead of distilled water).
The decolourizing solution is made by adding 30 ml concentrated hydro-
chloric acid to 970 ml ethanol.
After the sample has been filtered, the filter is flushed with 10 ml filtered
physiological saline. 10 ml of fuchsin stain is then added to the funnel and
left in contact with the filter for 30 minutes, when a vacuum is applied and
the stain pulled through the filter. (In the conventional method the slide is
flooded with fuchsin stain, heated, and left in contact for five minutes. T h e
filter method employs a 30 minute contact time at ambient temperature.)
The next step is to add 50 ml of decolourizing solution, leave in contact for
2 min, and filter before flushing with decolourizing solution until the
filter appears white. (In the conventional method 20% sulphuric acid is
used for decolourizing ; this treatment would damage the filter matrix.)
The filter is then flushed with 20 ml of physiological saline and finally
20 ml of counterstain is added to the funnel and immediately filtered. T h e
filter is removed from the holder and air dried before placing on the surface
of 2 ml of immersion oil in a Petri dish. The filter is finally placed on a
2 x 3 in. slide for oil immersion microscopy.
224 J. G . MULVANY
B. Clinical microbiology
1. Introduction
No area of membrane filter technology has afforded greater opportunity,
or greater challenge, than the application to clinical fluids where the unique
characteristics of the filter are so clearly useful. T h e filter technique is
especially valuable to the pathologist in those instances where-
(a) The organisms of specific interest are of low concentration in large
volumes of fluid.
(b) The suspending fluid is bacteriostatic, as in the case of spinal fluids
and urine from antibiotic-treated patients. These agents may be
easily flushed away through the filter, leaving the organisms on the
filter surface for culturing free of inhibitory effects.
(c) The filter method allows virtually all the suspending liquid being
sampled to be washed free from the organisms present in that liquid.
In nearly every clinical test, the liquid being sampled will have some
inhibitory action on the growth of organisms. For these reasons more
rapid growth rates are possible with the membrane filter method,
VII. MEMBRANE FILTER TECHNIQUES 225
2. Pretreatment
Table I1 shows typical body fluids which have been examined by the
membrane filter technique, and also suggests the method of pretreatment
to allow rapid filtration of the sample.
TABLE I1
Treatment for various clinical fluids
Typical Treat-
Typical Fluid volume ment
~ ~~~
~ ~~
Sputum 5-1 5 ml A
Nasopharangeal washings 10-50 ml A
Gastric washings 20-30 ml B
Pleural fluids 20-30 ml A
Ascitic fluids 15-25 ml A
Bursa1fluids 10-20 ml A
Urine-24 hr sample 1-3 litres B
Urine-Fresh sample 100 ml C
Spinal fluids 2-10 ml D
Whole blood 5 ml E
FIG. 20. Typical colonies of M . tubercuZosis from sputum, after isolation and
culturing on MF-Millipore filter.
VII. MEMBRANE FILTER TECHNIQUES 229
filter which is then flushed with 30 ml of sterile normal saline. All bacteria
will be retained on the surface of the filter which is aseptically transferred
from the Sterifil holder to a sterile Petri dish containing Lowenstein
Jensen medium with 1.57/, Tween-80 (Baltimore Biological Labora-
tory, Baltimore, Maryland, U.S.A.). The Petri dish is covered, inverted,
and incubated in an atmosphere of 10% carbon dioxide at 35°C. Identi-
fiable colonies are often observable as early as the fifth day, and are illus-
trated in Fig. 20. Haley and Arch (Haley and Arch, 1957) showed that
M. tuberculosis could be positively identified using the membrane filter
technique in an average of five days; in contrast to an average of approxi-
mately 27 days when material is conventionally cultured on Lowenstein-
Jensen medium, and 18 days when it is conventionally cultured on Tarshis
blood agar (Tarshis and Frisch, 1951) (see Table 111).
(c) Viable count of uncatheterized urine. Total viable counts in urine by
conventional plating procedures require extensive dilutions to detect a
pathological condition of 10,000-100,000 organisms/ml. A membrane
filter technique employing a Clinical Monitor (Millipore) is more direct
and simple. With a flamed bacteriological loop, 0.01 ml of urine is placed in a
sterile tube containing 10 ml of sterile Brain Heart Infusion Broth and the
TABLE 111
Number of days required for growth of tubercle
bacilli from sputum
1 6 sterile sterile
2 4 23 19
3 4 24 14
4 5 24 19
5 4 23 14
6 3 sterile 15
7 6 28 22
8 4 13 13
9 6 17 17
10 5 34 26
11 6 27 21
12 6 23 13
13 7 34 20
14 6 27 19
15 5 35 14
16 5 40 19
Average 5 26.6 17.7
230 J. G . MULVANY
whole is well shaken. The plugs are removed from the Clinical Monitor
and a syringe and valve assembly is attached to the outlet of the monitor.
2 ml of the inoculated medium is drawn into a sterile volumetric pipette
which is inserted into the inlet hole of the monitor. T h e entire 2 ml of the
medium is filtered through the monitor by means of the syringe (Fig. 21)
and filtration is stopped just as the last drop of fluid disappears from the
filter surface, so that sufficient medium will remain in the underlying
absorbent pad to support organism growth. T h e monitor is then recapped,
inverted and incubated at 37°C for 24 h. A contrast stain may facilitate
colony counting, and the number should be multiplied by 500 to obtain
results for 1 ml of urine sample.
(d) Other pathogens. T h e literature contains many additional techniques
of clinical significance (Millipore Bibliography). Methods for the isolation
of Bacillus anthracis, Brucella, Coccidioides immitis, Histoplasma capsulatum,
VII. MEMBRANE FILTER TECHNIQUES 23 1
5 . Airborne Organisms
Spores from fungi and certain spore forming bacteria may be collected
without significant loss of viability by direct filtration impingement on a
0.8 p m membrane filter. However, many airborne micro-organisms-
those which do not exhibit a protective capsule or cannot revert to a protec-
tive spore stage-desiccate rapidly when impinged on a solid surface and
cannot be collected viably on a dry filter surface. Thus, sampling for all
forms is best done by first impinging the organisms into a liquid medium
FIG.23. Impinger fluid is drawn through the Monitor to concentrate the organisms.
a syringe and valve is placed in the base of the monitor and the clamp on
the sample tube released. All but 1 or 2 ml of the impingement fluid is
drawn through the filter, Fig. 23, and then the impinger is washed down with
10 ml of sterile broth medium. The remaining fluid is then drawn through
the monitor and filtration is stopped just as the last few drops of medium
disappear from the filter surface. The monitor is then capped and incubated.
C. Water microbiology
1. Introduction
T h e coliform group of bacteria originate in the intestinal tract of man
and animal. Their presence in drinking water is indicative of a potential
VII. MEMBRANE FILTER TECHNIQUES 233
FIG. 24. Basic equipment and materials required for routine coliform water
analysis in the laboratory.
2. Standard procedures
The basic equipment and materials required for routine coliform water
analysis in the laboratory are shown in Fig. 24. T h e procedure is very simple
and is detailed as follows-
(a) Place a sterile 0.45 p m white filter-grid side up-aseptically in a
VII. MEMBRANE FILTER TECHNIQUES 235
The medium is brought to boiling point and then heating is stopped; the
pH of the final medium should be between 7.1 and 7.3. The appearance of
visible green or golden metallic sheen on a colony constitutes a positive
test for coliform organisms.
Colonies not giving this reaction cannot be considered significant. T h e
broth also suppresses the growth of non-coliform colonies, thus aiding in
the differentiation between these and the coliform types which exhibit the
236 J. G . MULVANY
MFC Medium
Tryptose 10 g
Proteose Peptone 5g
Yeast Extract 3g
Sodium Chloride 5g
Lactose 12.5 g
Rile Salts No. 3 (Difco) 1.5 g
Aniline Blue Ig
Dissolve of 1 grosolic acid in 100 ml of 0.2 N sodium hydroxide. Add the ingre-
dients to distilled water containing 0.01 yo rosolic acid sodium salt prepared from
this solution to make 1 litre of medium. Heat to boiling point.
FIG.
26. Attaching a sterile sample tube.
238 J. G . MIJLVANY
3. Field procedures
When laboratory facilities are not available, microbiological examination
can be conducted in the field, using field monitors (described earlier).
Figs. 25-30 describe pictorially the sampling, filtration and introduction of
nutrient medium stages in a typical field test. Use is made of a sterile field
monitor, syringe and valve, and a sterile sampling tube (Fig. 26) to draw
the water from the sampling cup. Sterile ampouled culture medium is
introduced into the base of the monitor to fully saturate the absorbent pad
(Fig. 29). The monitor is then plugged (Fig. 30) and placed in a portable
incubator (Fig. 31) which may be operated by a car battery, for example.
T o summarize, the membrane-filter procedure is an excellent technique
for the microbiological examination of water and is capable of yielding a
more precise and rapid answer than the MPN test (Geldreich etal., 1967).
However, proper care must be paid to all aspects of the procedure.
D. Industrial microbiology
1. Introduction
The membrane-filter method is also an important standard technique
in the microbiological examination of industrial fluids. Typical of this
VII. MEMBRANE FILTER TECHNIQUES 239
FIG.28. Invert the Monitor to draw the last of the sample through the filter.
are observed for visual growth, and if growth is indicated this should be
confirmed by microscopic examination.
FIG.32. Sterifil Holders, fitted with neoprene stoppers may be used with the
Millipore Sterility Test Manifold when large numbers of samples must be run in a
short time.
E. Radiomicrobiological techniques
The membrane filter provides an excellent medium for the recovery
and subsequent detection of radioactively labelled material from liquid
244 J. G . MULVANY
4. Bubble testing
This is a non-destructive method for checking the filtration assembly
prior to operation. T h e procedure is especially advantageous since sterility
is not broken and no product is lost. T h e bubble point for each membrane
filter type, with any specific liquid, is defined as the pressure required to
force air through a liquid saturated filter. T h e air pressure has to overcome
the surface tension of the fluid in the pores before breaking through. A
filter assembly free of all leaks, therefore, will hold air pressure indefinitely
under these conditions-failure to hold the pressure specified for the filter
type and liquid, is evidence of a passage larger than the filter pore size.
The bubble point with water for a 0.22 pm filter is 55 p.s.i., and for a 0.45 p m
filter is 32 p.s.i. In general, the bubble point will be slightly higher with oils
and slightly lower with alcohol solutions. T h e filtration assembly used for
sterile filtration of biological fluids should be bubble point tested after
autoclaving and just before the filtration is commenced. A satisfactory
bubble point guarantees that fluid passing to the downstream side of the
filter is in a sterile condition.
5. Flow rate
The high porosity and pore configuration results in flow rates nearly 40
times faster than through a conventional filter approaching the same particle
size retention capability. Membrane filters may be used at high inlet pres-
sures without fear of breakthrough of bacteria, as the screen configuration
of the filter will not be modified even under pressures of several thousand
p.s.i.
C. Filtration systems
At present both the 0.45 p m and the 0.22 p m filters find use in sterile
filtration applications, although the 0-22p m has become the more widely
applied of the two, because the pore size is smaller than any known bacteria
and thus, sterility is assured. T h e 0.45 p m filter, having larger pores, yields
flow rates approximately three times those of a 0.22 p m filter but should
only be used where prior experience indicates that organisms smaller than
0.45 p m are not present in the solution to be filtered. The 0.22 p m filter
should always be used with solutions containing Sera, Plasma or Trypsin,
where species of Pseudomonas or other bacteria of a small size are known
to occur.
Selection of the proper filter holder for each application depends chiefly
on the volume of liquid to be processed, however, other factors must often
also be considered. Some liquids, because of their high viscosity, particulate
density, or molecular composition, require relatively large filtration areas
VII. MEMBRANE FILTER TECHNIQUES 247
FIG. 33. Micro-Syringe Filter Holder, when used with a Cornwall pipetting
syringe (Becton-Dickinson), makes an ideal device for dispensing identical quantities
of sterilized tissue culture medium.
to achieve practical throughputs and flow rates. These materials, serum and
plasma, for example, also typically require some preliminary processing
such as centrifugation or prefiltration prior to sterilization by filtration.
I n designing a sterilizing filtration system it is very important that positive
pressure is used to force the liquid through the filter. A negative pressure
filtration system suffers from : (i) the possibility of leaks drawing contam-
inated air into the sterile products; (ii) a maximum pressure differential
of 1 atmosphere, and (iii) the necessity to transfer the sterile liquid from
the vacuum flask to the point of use. When filtering protein solutions
it is imperative to use positive pressure to reduce foaming which could
result in denaturation. Providing the membrane filter is correctly supported,
there is no possibility of forcing bacteria larger than the pore size through
the filter. The above disadvantages of negative pressure filtration systems
Prefilter
MF
FIG.35. A Microfibre Glass prefilter directly before the MF helps to extend filter
life.
are overcome by employing positive pressure, and also higher flow rates are
obtainable from a relatively small area of filter by using high pressures.
Fig. 33 shows positive pressure filtration on a small scale using a luer-lok
syringe, and Fig. 34 shows a larger scale, using a prlessurized vessel and a
142 mm filter holder.
If the liquid to be sterilized has a high burden of suspended particulate
matter which will rapidly clog the membrane filter then it is advantageous
to use a fibre glass prefilter directly before the membrane filter. T h e pre-
filter is a depth type filter, as described in an earlier section, and is capable
of removing a high percentage of particulate matter and so extending the life
of the membrane filter (Fig. 35). Fibre glass prefilters are extremely inert,
will not contribute particulate contamination to the filtrate and may be
sterilized by steam under pressure in place with the membrane filter.
T h e assembled Filter Holder System and connecting hoses should be
sterilized together (Fig. 36) and a final aseptic connection to a sterile receiv-
ing vessel made. If the autoclaving schedule for each assembled filter holder
is carefully followed, complete and positive reliability may be expected.
VII. MEMBRANE FILTER TECHNIQUES 249
FIG.36. Hose ends are wrapped with good quality Kraft paper for autoclaving.
Hoses must not be pinched OR.
(f) If the bubble point is satisfactory disconnect the nitrogen line from
B and vent excess pressure. Close valves B and D.
(g) Open valve A and allow liquid into the filter holder and use vent
valve to bleed off any entrained air.
(h) Open valve C and continue with filtration.
The bubble point test is a most critical method of determining the integrity
of a filtration system just prior to the commencement of filtration.
IV. SUMMARY
I n this chapter it has been possible only to scratch the surface of the many
biological applications for membrane filters. T h e characteristic which makes
a membrane filter such a valuable tool is that all particles both biological
and non-biological which are larger than the pore size will be positively
retained on the filter surface. Membrane filters were originally used strictly
as an analytical tool in the microbiological analysis of potable waters.
However, due to their “absolute” nature, the membrane filter has gradually
become an important device for the process sterilization of fluids.
REFERENCES
Baltimore Biological Laboratory, Baltimore, Maryland, U.S.A.
Berlin, M., and Rylander, R. (1964). J . Hyg., Camb., 61, 307-315.
Bowman, F. W. (1966). J. Pharm. Sci., 55, 818-821.
Danielsson, D. (1965). Acta. path. microbiol. Scand., 63, 597-603.
Danielsson, D., and Lurell, G. (1965). Acta. path. microbiol. Scand., 63, 604-608.
Difco Laboratories, Michigan, U.S.A.
Geldreich, E. E., Clark, H. F., Huff, B. H., and Best, L. C. (1965). J. Am. W a t . W k s
ASS., 57,208-213.
Geldreich, E. E., Jeter, H. L., and Winter, J. A. (1967). H.L.S., 4, 113-125.
Haley, L. D., and Arch, R. (1957). Am.J. din. Path., 27, 117-121.
Lightbrown, J . W. (1963). Proc. Round Table Conference on Sterility Testing,
London.
McClatchy, J. K., and Rosenblum, E. D. (1963).J. Bact., 86, 1211-1215.
McLeod, R. A., Light, M., White, L. A., and Currie, J. F. (1966). Appl. Microbiol.,
14, 979-984.
Millipore Bibliography, Millipore (U.K.) Ltd., Heron House, 109 Wembley Hill
Rd., Wembley, Middlesex.
Osgood, G. (1967). “Sheet Filtration and Filter Sheets.” Paper presented to the
Filtration Society, 4th April, 1967 (London).
Rohm and Haas Product. Lennig Chemicals, 26 Bedford Row, London, W.C.1.
Rose, R. E. (1966). W a t . Sewage Wks., 113, 1 4 .
Shanahan, A. J., Fornwald, R. E., Chapman, D. M., and Bleachey, A. N. (1956).
J . Bact., 71, 499-500.
Tarshis, M. S., and Frisch, A. W. (1951). Am.J. clin. Path., 21, 101-113.
Wallerstein Lab. Comm. (1950). 13, 357.
Wallerstein Lab. Comm. (1951). 14, 169.
Weissman, C. (1965). Proc. natn. Acad. Sci. U.S.A., 54, 202-207.
Winn, R. W., White, M. L., Carter, W. T., Miller, A. B., and Finegold, S. M.
(1966). J . Am. med. Ass., 197, 539-548.