Professional Documents
Culture Documents
255
256 Noninvasive Diagnostic Instrumentation
a = cfP
Characteristic
Temperature Density Velocity Impedance x 10^
terial rc) (g/cm'J (m/sec) (kg/mVsec) ^f(MHz) a (per cm)
ter 40 0.992 1529 1.517 1 2 0.00025
ne, 0.9*^0 normal 40 0.998 1539 1.537
tor oil 40 0.941 1411 1.328 1.67 0.037
in, average 37 1.03 1510 1.56 0.11
5 0.44
rtical gray matter 37 1.03 0.08
Ileal white matter 37 1.03 0.14
scle, skeletal 37 1.07 1570 1.68 0.13
37 0.97 1440 1.40 0.05
le, skull 37 1.77 3360 6.00 0.5 1.7 0.37
1.2
1.5 2.5
2 4.0
2.5 5.9
3 8.1
3.5 10.5
n 1.63
er 1.08 1510 1.63 2 0.19
od 1.01 1550 1.56 2 0.04
:ite 3.23
tumor
lin
Meningioma 5 0.73
jlioblastoma 5 0.38
Metastatic 5 0.50
Iney 1.04 1560 1.62 2 0.27
om W. Welkowitz and S. Deutsch, Biomedical Instruments: Theory and Design, Academic Press, New York,
6; by permission.
257
,
amplitude at point X _ n
amplitude at A" + 1 unit distance
As shown in Table 9. 1
a (per cm) = cf^
This formula shows that attenuation increases with some power of the
frequency, which means that the higher the frequency, the less distance it
can penetrate into the body with a given amount of ultrasonic energy. For
this reason, lower ultrasound frequencies are used for deeper penetration.
However, lower frequencies are incapable of reflecting small objects. As a
rule, a solid object surrounded by water or saline must be at least a quarter-
wave thick in order to cause a usable reflection. Thus, for finer resolution,
higher frequencies must be used. Ultrasound frequencies of 1 to 15 MHz
are usually used for diagnostic purposes. At 2 MHz, distinct echoes can be
recorded from interfaces 1 mm apart. Higher-frequency ultrasound is also
more subject to scattering than ultrasound at lower frequencies. However,
the high-frequency ultrasound beam can be focused for greater resolution
at a given depth.
1^1^
9.2 Principles of Ultrasonic Measurement 2S9
A
well-known characteristic of ultrasound frequently utihzed in bio-
medical instrumentation is the Doppler effecty in which the frequency of the
reflected ultrasonic energy is increased or decreased by a moving interface.
The amount of frequency shift can be expressed in the formula:
A/= 2_K
X
where / = shift in frequency of the reflected wave
V = velocity of the interface
X = wavelength of the transmitted ultrasound
The frequency increases when the interface moves toward the trans-
ducer and decreases when it moves away. With an ultrasound frequency of
3 MHz, about 40 Hz for each cm/sec of interface velocity.
the shift is
A way
to understand this in general terms is to consider what
useful
happens if an automobile with its horn sounding passes by on the street.
The pitch or perceived frequency of the sound seems higher as the car is
approaching but seems lower as it goes away. This is an example of the
Doppler frequency shift. When ultrasound is reflected from a moving object,
the measured frequency shift is proportional to velocity.
(a)
(b)
M-scan
K A
Corresponding A-scan
9.3 Ultrasonic Diagnosis 263
trace lies immediately adjacent to the one preceding it, the dot
representing each target will trace a line on the paper as shown
in Figure 9.9. A stationary target will trace a straight hne,
whereas a moving target will trace the pattern of its movement
with respect to time. A light-pen recorder in which the intensity
of the light source can be controlled may be used instead of an
oscilloscope to produce a chart record of themovement of echoes
with respect to time. An example of an M-scan recording is the
echocardiogram shown in Figure 9.12.
3. B-scan display: While the M-scan is used to display the movement
of targets with respect to time, the B-scan presents a two-dimen-
sional image of a stationary organ or body structure. As in the
M-scan, the brightness of the oscilloscope or light-pen beam is
controlled by returning echoes; however, in the B-scan the trans-
ducer is moved with respect to the body while the vertical deflec-
tion of the oscilloscope or movement of the chart paper is made
to correspond to the movement of the transducer. The movement
may be linear, circular, or a combination of the two, but where
it is anything other than Unear, the sweep must be made to
compensate for the variations in order to provide a true two-
dimensional display of the segment being scanned. Examples
of B-scan displays are shown in Figure 9.15.