You are on page 1of 11

ARTICLE 7 ASME BPVC.

V-2015

ARTICLE 7
MAGNETIC PARTICLE EXAMINATION

ð15Þ T-710 SCOPE requirements listed in Table T-721. The written proce-
dure shall establish a single value, or range of values,
When specified by the referencing Code Section, the for each requirement.
magnetic particle examination techniques described in
this Article shall be used. In general, this Article is in con- T-721.2 Procedure Qualification. When procedure
formance with SE-709, Standard Guide for Magnetic Par- qualification is specified by the referencing Code Section,
ticle Examination. This document provides details to be a change of a requirement in Table T-721 identified as an
considered in the procedures used. essential variable shall require requalification of the writ-
ten procedure by demonstration. A change of a require-
When this Article is specified by a referencing Code
ment identified as a nonessential variable does not
Section, the magnetic particle method described in this
require requalification of the written procedure. All
Article shall be used together with Article 1, General Re-
changes of essential or nonessential variables from those
quirements. Definition of terms used in this Article are
specified within the written procedure shall require revi-
in Article 1, Mandatory Appendix I, I-121.4, MT — Mag-
sion of, or an addendum to, the written procedure.
netic Particle.

T-730 EQUIPMENT
T-720 GENERAL A suitable and appropriate means for producing the
necessary magnetic flux in the part shall be employed,
The magnetic particle examination method is applied to
using one or more of the techniques listed in and de-
detect cracks and other discontinuities on the surfaces of
scribed in T-750.
ferromagnetic materials. The sensitivity is greatest for
surface discontinuities and diminishes rapidly with in- T-731 EXAMINATION MEDIUM
creasing depth of discontinuities below the surface. Typi-
cal types of discontinuities that can be detected by this The finely divided ferromagnetic particles used for the
method are cracks, laps, seams, cold shuts, and examination shall meet the following requirements.
laminations. (a) Particle Types. The particles shall be treated to im-
part color (fluorescent pigments, nonfluorescent pig-
In principle, this method involves magnetizing an area
ments, or both) in order to make them highly visible
to be examined, and applying ferromagnetic particles (the
(contrasting) against the background of the surface being
examination’s medium) to the surface. Particle patterns
examined.
form on the surface where the magnetic field is forced
(b) Particles. Dry and wet particles and suspension ve-
out of the part and over discontinuities to cause a leakage
hicles should be in accordance with SE-709.
field that attracts the particles. Particle patterns are
(c) Temperature Limitations. Particles shall be used
usually characteristic of the type of discontinuity that is
within the temperature range limitations set by the man-
detected.
ufacturer of the particles. Alternatively, particles may be
Whichever technique is used to produce the magnetic
used outside the particle manufacturer’s recommenda-
flux in the part, maximum sensitivity will be to linear dis-
tions providing the procedure is qualified in accordance
continuities oriented perpendicular to the lines of flux.
with Article 1, T-150 at the proposed temperature.
For optimum effectiveness in detecting all types of dis-
continuities, each area is to be examined at least twice,
with the lines of flux during one examination being ap- T-740 MISCELLANEOUS REQUIREMENTS
proximately perpendicular to the lines of flux during the
other. T-741 SURFACE CONDITIONING
T-741.1 Preparation.
(a) Satisfactory results are usually obtained when the
T-721 WRITTEN PROCEDURE REQUIREMENTS surfaces are in the as-welded, as-rolled, as-cast, or as-
T-721.1 Requirements. Magnetic particle examina- forged conditions. However, surface preparation by
tion shall be performed in accordance with a written pro- grinding or machining may be necessary where surface ir-
cedure, which shall, as a minimum, contain the regularities could mask indications due to discontinuities.

192
ASME BPVC.V-2015 ARTICLE 7

Table T-721
Requirements of a Magnetic Particle Examination Procedure
Nonessential
Requirement Essential Variable Variable
Magnetizing technique X ...
Magnetizing current type or amperage outside range specified by this Article or as previously X ...
qualified
Surface preparation X . . .
Magnetic particles (fluorescent/visible, color, particle size, wet/dry) X . . .
Method of particle application X . . .
Method of excess particle removal X . . .
Minimum light intensity X . . .
Existing coatings, greater than the thickness demonstrated X . . .
Nonmagnetic surface contrast enhancement, when utilized X . . .
Performance demonstration, when required X . . .
Examination part surface temperature outside of the temperature range recommended by the X . . .
manufacturer of the particles or as previously qualified
Shape or size of the examination object . .. X
Equipment of the same type . .. X
Temperature (within those specified by manufacturer or as previously qualified) . .. X
Demagnetizing technique . .. X
Post-examination cleaning technique . .. X
Personnel qualification requirements . .. X

(b) Prior to magnetic particle examination, the surface T-750 TECHNIQUE


to be examined and all adjacent areas within at least
1 in. (25 mm) shall be dry and free of all dirt, grease, lint, T-751 TECHNIQUES
scale, welding flux and spatter, oil, or other extraneous One or more of the following five magnetization techni-
matter that could interfere with the examination. ques shall be used:
(c) Cleaning may be accomplished using detergents, or- (a) prod technique
ganic solvents, descaling solutions, paint removers, vapor (b) longitudinal magnetization technique
degreasing, sand or grit blasting, or ultrasonic cleaning (c) circular magnetization technique
methods. (d) yoke technique
(d) If nonmagnetic coatings are left on the part in the (e) multidirectional magnetization technique
area being examined, it shall be demonstrated that indica-
tions can be detected through the existing maximum coat- T-752 PROD TECHNIQUE
ing thickness applied. When AC yoke technique is used,
the demonstration shall be in accordance with Mandatory T-752.1 Magnetizing Procedure. For the prod techni-
Appendix I of this Article. que, magnetization is accomplished by portable prod type
electrical contacts pressed against the surface in the area
T-741.2 Nonmagnetic Surface Contrast Enhance- to be examined. To avoid arcing, a remote control switch,
ment. Nonmagnetic surface contrasts may be applied by which may be built into the prod handles, shall be pro-
the examiner to uncoated surfaces, only in amounts suffi- vided to permit the current to be applied after the prods
cient to enhance particle contrast. When nonmagnetic have been properly positioned.
surface contrast enhancement is used, it shall be demon-
strated that indications can be detected through the en- T-752.2 Magnetizing Current. Direct or rectified
hancement. Thickness measurement of this magnetizing current shall be used. The current shall be
nonmagnetic surface contrast enhancement is not 100 (minimum) amp/in. (4 amp/mm) to
required. 125 (maximum) amp/in. (5 amp/mm) of prod spacing
for sections 3/4 in. (19 mm) thick or greater. For sections
NOTE: Refer to T-150(a) for guidance for the demonstration re-
quired in T-741.1(d) and T-741.2.
less than 3/4 in. (19 mm) thick, the current shall be
90 amp/in. (3.6 amp/mm) to 110 amp/in.
(4.4 amp/mm) of prod spacing.
T-752.3 Prod Spacing. Prod spacing shall not exceed
8 in. (200 mm). Shorter spacing may be used to accommo-
date the geometric limitations of the area being examined
or to increase the sensitivity, but prod spacings of less

193
ARTICLE 7 ASME BPVC.V-2015

than 3 in. (75 mm) are usually not practical due to band- (d) If the area to be magnetized extends beyond 9 in.
ing of the particles around the prods. The prod tips shall (225 mm) on either side of the coil’s center, field ade-
be kept clean and dressed. If the open circuit voltage of quacy shall be demonstrated using a magnetic field indi-
the magnetizing current source is greater than 25 V, lead, cator or artificial flaw shims per T-764.
steel, or aluminum (rather than copper) tipped prods are (e) For large parts due to size and shape, the magnetiz-
recommended to avoid copper deposits on the part being ing current shall be 1200 ampere-turns
examined. to 4500 ampere-turns. The field adequacy shall be de-
monstrated using artificial flaw shims or a pie-shaped
magnetic field indicator in accordance with T-764. A
T-753 LONGITUDINAL MAGNETIZATION Hall-Effect probe gaussmeter shall not be used with encir-
TECHNIQUE cling coil magnetization techniques.

T-753.1 Magnetizing Procedure. For this technique, T-753.3 Magnetizing Current. The current required
magnetization is accomplished by passing current to obtain the necessary magnetizing field strength shall
through a multi-turn fixed coil (or cables) that is wrapped be determined by dividing the ampere-turns obtained in
around the part or section of the part to be examined. This steps T-753.2(a) or T-753.2(b) by the number of turns
produces a longitudinal magnetic field parallel to the axis in the coil as follows:
of the coil.
If a fixed, prewound coil is used, the part shall be placed
near the side of the coil during inspection. This is of spe-
cial importance when the coil opening is more than
For example, if a 5-turn coil is used and the ampere-
10 times the cross-sectional area of the part.
turns required are 5000, use
T-753.2 Magnetic Field Strength. Direct or rectified
current shall be used to magnetize parts examined by this
technique. The required field strength shall be calculated
based on the length L and the diameter D of the part in
accordance with (a) and (b), or as established in (d)
and (e), below. Long parts shall be examined in sections T-754 CIRCULAR MAGNETIZATION TECHNIQUE
not to exceed 18 in. (450 mm), and 18 in. (450 mm) shall T-754.1 Direct Contact Technique.
be used for the part L in calculating the required field (a) Magnetizing Procedure. For this technique, magneti-
strength. For noncylindrical parts, D shall be the maxi- zation is accomplished by passing current through the
mum cross-sectional diagonal. part to be examined. This produces a circular magnetic
(a) Parts With L/D Ratios Equal to or Greater Than 4. field that is approximately perpendicular to the direction
The magnetizing current shall be within 10% of the of current flow in the part.
ampere-turns’ value determined as follows: (b) Magnetizing Current. Direct or rectified (half-wave
rectified or full-wave rectified) magnetizing current shall
be used.
(1) The current shall be 300 amp/in. (12 A/mm) to
800 amp/in. (31 A/mm) of outer diameter.
(2) Parts with geometric shapes other than round
For example, a part 10 in. (250 mm) long × 2 in. with the greatest cross-sectional diagonal in a plane at
(50 mm) diameter has an L/D ratio of 5. Therefore, right angles to the current flow shall determine the inches
to be used in (1) above.
(3) If the current levels required for (1) cannot be ob-
tained, the maximum current obtainable shall be used and
the field adequacy shall be demonstrated in accordance
with T-764.
(b) Parts With L/D Ratios Less Than 4 but Not Less Than
2. The magnetizing ampere-turns shall be within 10% of T-754.2 Central Conductor Technique.
the ampere-turns’ value determined as follows: (a) Magnetizing Procedure. For this technique, a central
conductor is used to examine the internal surfaces of cy-
lindrically or ring-shaped parts. The central conductor
technique may also be used for examining the outside sur-
faces of these shapes. Where large diameter cylinders are
to be examined, the conductor shall be positioned close to
(c) Parts With L/D Ratios Less Than 2. Coil magnetiza- the internal surface of the cylinder. When the conductor is
tion technique cannot be used. not centered, the circumference of the cylinder shall be

194
ASME BPVC.V-2015 ARTICLE 7

examined in increments. Field strength measurements in


Figure T-754.2.2
accordance with T-764 shall be used, to determine the ex-
The Effective Region of Examination When
tent of the arc that may be examined for each conductor
Using an Offset Central Conductor
position or the rules in (c) below may be followed. Bars
or cables, passed through the bore of a cylinder, may be
used to induce circular magnetization. Central conductor
Effective
(b) Magnetizing Current. The field strength required region 4d
shall be equal to that determined in T-754.1(b) for a
single-turn central conductor. The magnetic field will in-
crease in proportion to the number of times the central
conductor cable passes through a hollow part. For exam-
d
ple, if 6000 amperes are required to examine a part using
a single pass central conductor, then 3000 amperes are
required when 2 passes of the through-cable are used,
and 1200 amperes are required if 5 passes are used
(see Figure T-754.2.1). When the central conductor tech-
nique is used, magnetic field adequacy shall be verified
using a magnetic particle field indicator in accordance
with T-764.
(c) Offset Central Conductor. When the conductor pas-
sing through the inside of the part is placed against an in- T-756 MULTIDIRECTIONAL MAGNETIZATION
side wall of the part, the current levels, as given in TECHNIQUE
T-754.1(b)(1) shall apply, except that the diameter used T-756.1 Magnetizing Procedure. For this technique,
for current calculations shall be the sum of the diameter magnetization is accomplished by high amperage power
of the central conductor and twice the wall thickness. packs operating as many as three circuits that are ener-
The distance along the part circumference (exterior) that gized one at a time in rapid succession. The effect of these
is effectively magnetized shall be taken as four times the rapidly alternating magnetizing currents is to produce an
diameter of the central conductor, as illustrated in Figure overall magnetization of the part in multiple directions.
T-754.2.2. The entire circumference shall be inspected by Circular or longitudinal magnetic fields may be generated
rotating the part on the conductor, allowing for approxi- in any combination using the various techniques de-
mately a 10% magnetic field overlap. scribed in T-753 and T-754.
T-755 YOKE TECHNIQUE T-756.2 Magnetic Field Strength. Only three phase,
For this technique, alternating or direct current electro- full-wave rectified current shall be used to magnetize
magnetic yokes, or permanent magnet yokes, shall be the part. The initial magnetizing current requirements
used. for each circuit shall be established using the previously
described guidelines (see T-753 and T-754). The ade-
quacy of the magnetic field shall be demonstrated using
artificial flaw shims or a pie-shaped magnetic particle
field indicator in accordance with T-764. A Hall-Effect

Figure T-754.2.1
Single-Pass and Two-Pass Central Conductor Technique

195
ARTICLE 7 ASME BPVC.V-2015

probe gaussmeter shall not be used to measure field ade- has been subjected to a major repair, periodic overhaul,
quacy for the multidirectional magnetization technique. or damage. If equipment has not been in use for a year
An adequate field shall be obtained in at least two nearly or more, calibration shall be done prior to first use.
perpendicular directions, and the field intensities shall be
balanced so that a strong field in one direction does not
overwhelm the field in the other direction. For areas T-764 MAGNETIC FIELD ADEQUACY AND
where adequate field strengths cannot be demonstrated, DIRECTION
additional magnetic particle techniques shall be used to T-764.1 Application. The use of magnetic field indi-
obtain the required two-directional coverage. cators, artificial shims, or Hall-Effect tangential-field
probes are only permitted when specifically referenced
by the following magnetizing techniques:
T-760 CALIBRATION (a) Longitudinal (T-753)
T-761 FREQUENCY OF CALIBRATION (b) Circular (T-754)
(c) Multidirectional (T-756)
T-761.1 Magnetizing Equipment.
(a) Frequency. Magnetizing equipment with an am- T-764.2 Magnetic Field Adequacy. The applied mag-
meter shall be calibrated at least once a year, or whenever netic field shall have sufficient strength to produce satis-
the equipment has been subjected to major electric re- factory indications, but shall not be so strong that it
pair, periodic overhaul, or damage. If equipment has not causes masking of relevant indications by nonrelevant ac-
been in use for a year or more, calibration shall be done cumulations of magnetic particles. Factors that influence
prior to first use. the required field strength include the size, shape, and
(b) Procedure. The accuracy of the unit’s meter shall be material permeability of the part; the technique of magne-
verified annually by equipment traceable to a national tization; coatings; the method of particle application; and
standard. Comparative readings shall be taken for at least the type and location of discontinuities to be detected.
three different current output levels encompassing the When it is necessary to verify the adequacy of magnetic
usable range. field strength, it shall be verified by using one or more
(c) Tolerance. The unit’s meter reading shall not devi- of the following three methods.
ate by more than ±10% of full scale, relative to the actual (a) Pie-Shaped Magnetic Particle Field Indicator. The in-
current value as shown by the test meter. dicator, shown in Figure T-764.2(a), shall be positioned
on the surface to be examined, such that the copper-
T-761.2 Light Meters. Light meters shall be cali- plated side is away from the inspected surface. A suitable
brated at least once a year or whenever a meter has been field strength is indicated when a clearly defined line (or
repaired. If meters have not been in use for one year or lines) of magnetic particles form(s) across the copper face
more, calibration shall be done before being used. of the indicator when the magnetic particles are applied
simultaneously with the magnetizing force. When a
ð15Þ T-762 LIFTING POWER OF YOKES clearly defined line of particles is not formed, the magne-
(a) The magnetizing power of yokes shall be verified tizing technique shall be changed as needed. Pie-type in-
prior to use each day the yoke is used. The magnetizing dicators are best used with dry particle procedures.
power of yokes shall be verified whenever the yoke has (b) Artificial Flaw Shims. One of the shims shown in
been damaged or repaired. Figure T-764.2(b)(1) or Figure T-764.2(b)(2) whose or-
(b) Each alternating current electromagnetic yoke shall ientation is such that it can have a component perpendi-
have a lifting power of at least 10 lb (4.5 kg) at the max- cular to the applied magnetic field shall be used. Shims
imum pole spacing that will be used. with linear notches shall be oriented so that at least one
(c) Each direct current or permanent magnetic yoke notch is perpendicular to the applied magnetic field.
shall have a lifting power of at least 40 lb (18 kg) at the Shims with only circular notches may be used in any or-
maximum pole spacing that will be used. ientation. Shims shall be attached to the surface to be ex-
(d) Each weight shall be weighed with a scale from a re- amined, such that the artificial flaw side of the shim is
putable manufacturer and stenciled with the applicable toward the inspected surface. A suitable field strength is
nominal weight prior to first use. A weight need only be indicated when a clearly defined line (or lines) of mag-
verified again if damaged in a manner that could have netic particles, representing the 30% depth flaw, appear
caused potential loss of material. (s) on the shim face when magnetic particles are applied
simultaneously with the magnetizing force. When a
clearly defined line of particles is not formed, the magne-
T-763 GAUSSMETERS tizing technique shall be changed as needed. Shim-type
Hall-Effect probe gaussmeters used to verify magnetiz- indicators are best used with wet particle procedures.
ing field strength in accordance with T-754 shall be cali- NOTE: The circular shims shown in Figure T-764.2(b)(2) illustration
brated at least once a year or whenever the equipment (b) also have flaw depths less and greater than 30%.

196
ASME BPVC.V-2015 ARTICLE 7

Figure T-764.2(a) Figure T-764.2(b)(1)


Pie-Shaped Magnetic Particle Field Indicator Artificial Flaw Shims

0.002 in.
A
(0.06 mm)
0.005 in.
0.75 in. (0.125 mm)
(20 mm) typical

A 0.0006 in.
(0.015 mm)
0.25 in.
(6 mm) Section A–A
Type B

A 0.002 in.
(0.05 mm)

0.75 in.
(20 mm)

A 0.0006 in.
(0.015 mm)
0.5 in.
(12.5 mm) Section A–A
Type C

2 in. (50 mm) 0.002 in.


Defect Division (0.05 mm)

(c) Hall-Effect Tangential-Field Probe. A gaussmeter 0.4 in.


(10 mm)
and Hall-Effect tangential-field probe shall be used for 0.005 in.
measuring the peak value of a tangential field. The probe (0.125 mm)
shall be positioned on the surface to be examined, such 0.2 in. typical
(5 mm)
that the maximum field strength is determined. A suitable 0.0006 in.
field strength is indicated when the measured field is Type R (0.015 mm)
within the range of 30 G to 60 G (2.4 kAm−1
GENERAL NOTE: Above are examples of artificial flaw shims used
to 4.8 kAm−1) while the magnetizing force is being ap-
in magnetic particle inspection system verification (not drawn to
plied. See Article 7, Nonmandatory Appendix A. scale). The shims are made of low carbon steel (1005 steel foil).
T-764.3 Magnetic Field Direction. The direction(s) The artificial flaw is etched or machined on one side of the foil to
a depth of 30% of the foil thickness.
of magnetization shall be determined by particle indica-
tions obtained using an indicator or shims as shown in
Figure T-764.2(a), Figure T-764.2(b)(1), or Figure should be run through the recirculating system for at least
T-764.2(b)(2). When a clearly defined line of particles 0 min to ensure thorough mixing of all particles which
are not formed could have settled on the sump screen and along the sides
(a) in the desired direction, or or bottom of the tank.
(b) in at least two nearly perpendicular directions for
the multidirectional technique T-765.1 Concentration. Take a 100-mL portion of the
suspension from the hose or nozzle, demagnetize and al-
the magnetizing technique shall be changed as needed. low it to settle for approximately 60 min with petroleum
distillate suspensions or 30 min with water-based sus-
T-765 WET PARTICLE CONCENTRATION AND pensions before reading. The volume settling out at the
CONTAMINATION bottom of the tube is indicative of the particle concentra-
tion in the bath.
Wet Horizontal Units shall have the bath concentration
and bath contamination determined by measuring its set- T-765.2 Settling Volumes. For fluorescent particles,
tling volume. This is accomplished through the use of an the required settling volume is from 0.1 to 0.4 mL in a
ASTM Test Method D 96 pear-shaped centrifuge tube with 100 mL bath sample and from 1.2 to 2.4 mL per
a 1-mL stem (0.05-mL divisions) for fluorescent particle 100 mL of vehicle for nonfluorescent particles unless
suspensions or a 1.5-mL stem (0.1-mL divisions) for non- otherwise specified by the particle manufacturer. Concen-
fluorescent suspensions. Before sampling, the suspension tration checks shall be made at least every eight hours.

197
ARTICLE 7 ASME BPVC.V-2015

Figure T-764.2(b)(2)
Artificial Flaw Shims

0.75 in. (typ) (19.05 mm) 0.75 in. (typ) (19.05 mm)
0.25 in. 0.25 in.
(6.36 mm) (6.36 mm)

0.507 in. diam. O.D. 0.507 in. diam. O.D.


(12.88 mm) (12.88 mm)

0.007 in. (typ) 0.007 in. (typ)


(0.18 mm) (0.18 mm)
Notches: Notches:
Depth: 30% 0.0006 in. Depth: 30% 0.0012 in.
(0.015 mm) (0.030 mm)
230 Shim thickness: 430 Shim thickness:
0.002 in. (0.05 mm) 0.004 in. (0.10 mm)
Shim Type CX-230 Shim Type CX-430
(a)

0.75 in. (typ) (19.05 mm) 0.75 in. (typ) (19.05 mm)
0.007 in. (type) 0.007 in. (type)
(0.18 mm) 0.507 in. diam. O.D. (0.18 mm) 0.507 in. diam. O.D.
(12.88 mm) (12.88 mm)

0.383 in. diam. O.D. 0.383 in. diam. O.D.


(9.73 mm) (9.73 mm)
0.258 in. diam. O.D. 0.258 in. diam. O.D.
(6.55 mm) (6.55 mm)

Notch depth: Notch depth:


20% 0.0004 in. 20% 0.0004 in.
(0.010 mm) O.D. (0.010 mm) O.D.
4-234 30% 0.0006 in. 2-234 30% 0.0006 in.
(0.015 mm) center (0.015 mm) center
Shim Type 3C4-234 40% 0.0008 in. Shim Type 3C2-234 40% 0.0008 in.
Shim Thickness 0.004 in. (0.102 mm) (0.020 mm) I.D. Shim Thickness 0.002 in. (0.05 mm) (0.020 mm) I.D.
(b)

0.79 in. (typ) (20.06 mm) 0.79 in. (typ) (20.06 mm)
0.235 in. (typ) 0.235 in. (typ)
(5.97 mm) (5.97 mm)

0.255 in. diam. O.D. 0.255 in. diam.


(6.48 mm) O.D. (6.48 mm)

0.006 in. (typ) 0.006 in. (typ)


(0.152 mm) (0.152 mm)
0.395 in. (typ) 0.20 in. (typ) 0.395 in. (typ) 0.20 in. (typ)
(10.03 mm) (5.08 mm) (10.03 mm) (5.08 mm)

Notch depth: Notch depth:


30% 0.0006 in. 30% 0.0012 in.
(0.015 mm) (0.030 mm)

230 430

Shim Thickness 0.002 in. (0.051 mm) Shim Thickness 0.004 in. (0.102 mm)
Shim Type CX-230 Shim Type CX4-430
(c)

198
ASME BPVC.V-2015 ARTICLE 7

T-765.3 Contamination. Both fluorescent and non- be used at these or higher amperages. The amperage val-
fluorescent suspensions shall be checked periodically ues stated shall not be exceeded in the test. If the test does
for contaminants such as dirt, scale, oil, lint, loose fluores- not reveal the required number of holes, the equipment
cent pigment, water (in the case of oil suspensions), and shall be taken out of service and the cause of the loss of
particle agglomerates which can adversely affect the per- sensitivity determined and corrected. This test shall be
formance of the magnetic particle examination process. run at least once per week.
The test for contamination shall be performed at least
once per week.
(a) Carrier Contamination. For fluorescent baths, the li- T-770 EXAMINATION
quid directly above the precipitate should be examined T-771 PRELIMINARY EXAMINATION
with fluorescent excitation light. The liquid will have a lit-
Before the magnetic particle examination is conducted,
tle fluorescence. Its color can be compared with a freshly
a check of the examination surface shall be conducted to
made-up sample using the same materials or with an un-
locate any discontinuity surface openings which may not
used sample from the original bath that was retained for
attract and hold magnetic particles because of their width.
this purpose. If the “used” sample is noticeably more
fluorescent than the comparison standard, the bath shall T-772 DIRECTION OF MAGNETIZATION
be replaced.
(b) Particle Contamination. The graduated portion of At least two separate examinations shall be performed
the tube shall be examined under fluorescent excitation on each area. During the second examination, the lines of
light if the bath is fluorescent and under visible light magnetic flux shall be approximately perpendicular to
(for both fluorescent and nonfluorescent particles) for those used during the first examination. A different tech-
striations or bands, differences in color or appearance. nique for magnetization may be used for the second
Bands or striations may indicate contamination. If the to- examination.
tal volume of the contaminates, including bands or stria-
tions exceeds 30% of the volume magnetic particles, or if
T-773 METHOD OF EXAMINATION
the liquid is noticeably fluorescent, the bath shall be The ferromagnetic particles used in an examination
replaced. medium can be either wet or dry, and may be either fluor-
escent or nonfluorescent. Examination(s) shall be done by
T-766 SYSTEM PERFORMANCE OF HORIZONTAL the continuous method.
UNITS (a) Dry Particles. The magnetizing current shall remain
The Ketos (Betz) ring specimen (see Figure T-766.1) on while the examination medium is being applied and
shall be used in evaluating and comparing the overall per- while any excess of the examination medium is removed.
formance and sensitivity of both dry and wet, fluorescent (b) Wet Particles. The magnetizing current shall be
and nonfluorescent magnetic particle techniques using a turned on after the particles have been applied. Flow of
central conductor magnetization technique. particles shall stop with the application of current. Wet
(a) Ketos (Betz) Test Ring Material. The tool steel (Ke- particles applied from aerosol spray cans may be applied
tos) ring should be machined from AISI 01 material in ac- before and/or after magnetizing current is applied. Wet
cordance with Figure T-766.1. Either the machined ring particles may be applied during the application of magne-
or the steel blank should be annealed at 1,650°F tizing current if they are not applied directly to the exam-
(900°C), cooled 50°F (28°C) per hour to 1,000°F ination area and are allowed to flow over the examination
(540°C) and then air cooled to ambient temperature to area or are applied directly to the examination area with
give comparable results using similar rings that have low velocities insufficient to remove accumulated
had the same treatment. Material and heat treatment particles.
are important variables. Experience indicates controlling
T-774 EXAMINATION COVERAGE
the softness of the ring by hardness (90 to 95 HRB) alone
is insufficient. All examinations shall be conducted with sufficient field
(b) Using the Test Ring. The test ring (see Figure overlap to ensure 100% coverage at the required sensitiv-
T-766.1), is circularly magnetized with full-wave rectified ity (T-764).
AC passing through a central conductor with a 1 in.
to 11/4 in. (25 mm to 32 mm) diameter hole located in T-775 RECTIFIED CURRENT
the ring center. The conductor should have a length great- (a) Whenever direct current is required rectified cur-
er than 16 in. (400 mm). The currents used shall be 1400, rent may be used. The rectified current for magnetization
2500, and 3400 amps. The minimum number of holes shall be either three-phase (full-wave rectified) current,
shown shall be three, five, and six, respectively. The ring or single phase (half-wave rectified) current.
edge should be examined with either black light or visible (b) The amperage required with three-phase, full-wave
light, depending on the type of particles involved. This rectified current shall be verified by measuring the aver-
test shall be run at the three amperages if the unit will age current.

199
ARTICLE 7
ð15Þ Figure T-766.1
Ketos (Betz) Test Ring

3/ in. (19 mm)


4
Typ.
125
1 2
3
4
5
11/4 in. 6 5 in.
(32 mm) (125
7 mm)
8
9

ASME BPVC.V-2015
12
D 11 10
200

7/ in.
8
(22 mm)

Hole 1 2 3 4 5 6 7 8 9 10 11 12

Diameter 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8) 0.07 (1.8)
[Note (1)]
“D” [Note (2)] 0.07 (1.8) 0.14 (3.6) 0.21 (5.3) 0.28 (7.1) 0.35 (9.0) 0.42 (10.8) 0.49 (12.6) 0.56 (14.4) 0.63 (16.2) 0.70 (18.0) 0.77 (19.8) 0.84 (21.6)
GENERAL NOTES:
(a) All dimensions are ±0.03 in. (±0.8 mm) or as noted in Notes (1) and (2).
(b) In the in-text table, all dimensions are in inches, except for the parenthesized values, which are in millimeters.
(c) Material is ANSI 01 tool steel from annealed round stock.
(d) The ring may be heat treated as follows: Heat to 1,400°F to 1,500°F (760°C to 790°C). Hold at this temperature for 1 hr. Cool to a minimum rate of 40°F/hr (22°C/h) to below 1,000°F
(540°C). Furnace or air cool to room temperature. Finish the ring to RMS 25 and protect from corrosion.
NOTES:
(1) All hole diameters are ±0.005 in. (±0.1 mm.) Hole numbers 8 through 12 are optional.
(2) Tolerance on the D distance is ±0.005 in. (±0.1 mm).
ASME BPVC.V-2015 ARTICLE 7

(c) The amperage required with single-phase (half- (e) The black light intensity shall be measured with a
wave rectified) current shall be verified by measuring black light meter prior to use, whenever the light’s power
the average current output during the conducting half cy- source is interrupted or changed, and at the completion of
cle only. the examination or series of examinations.
(d) When measuring half-wave rectified current with a T-777.3 Fluorescent Magnetic Particles With Other
direct current test meter, readings shall be multiplied by Fluorescent Excitation Wavelengths. Alternatively to
two. the requirements in T-777.2, the examinations may be
performed using alternate wavelength light sources
T-776 EXCESS PARTICLE REMOVAL which cause fluorescence in specific particle coatings.
Any alternate light wavelength light sources and specific
Accumulations of excess dry particles in examinations particle designations used shall be qualified30 in accor-
shall be removed with a light air stream from a bulb or dance with Mandatory Appendix IV. The examination
syringe or other source of low pressure dry air. The exam- shall be performed as follows:
ination current or power shall be maintained while re-
(a) It shall be performed in a darkened area.
moving the excess particles.
(b) Examiners shall be in a darkened area for at least
5 min prior to performing examinations to enable their
T-777 INTERPRETATION eyes to adapt to dark viewing. Glasses or lenses worn
by examiners shall not be photochromic or exhibit any
The interpretation shall identify if an indication as
fluorescence.
false, nonrelevant, or relevant. False and nonrelevant in-
(c) If the fluorescence excitation light source emits visi-
dications shall be proven as false or nonrelevant. Inter-
ble light intensities greater than 2 fc (20 Lx), the examiner
pretation shall be carried out to identify the locations of
shall wear fluorescence-enhancing filter glasses approved
indications and the character of the indication.
by the light source manufacturer for use with that light
ð15Þ T-777.1 Visible (Color Contrast) Magnetic Particles. source.
Surface discontinuities are indicated by accumulations of (d) The fluorescence excitation light source shall
magnetic particles which should contrast with the exam- achieve at least the minimum light intensity on the sur-
ination surface. The color of the magnetic particles shall face of the part throughout the examination as qualified
be different than the color of the examination surface. Il- in the tests of Mandatory Appendix IV.
lumination (natural or supplemental white light) of the (e) Reflectors, filters, glasses, and lenses should be
examination surface is required for the evaluation of indi- checked and, if necessary, cleaned prior to use. Cracked
cations. The minimum light intensity shall be 100 fc or broken reflectors, filters, glasses, or lenses shall be re-
(1 000 lx). The light intensity, natural or supplemental placed immediately.
white light source, shall be measured with a white light (f) The fluorescence excitation light intensity shall be
meter prior to the evaluation of indications or a verified measured with a suitable fluorescence excitation light
light source shall be used. Verification of light sources is meter prior to use, whenever the light’s power source is
required to be demonstrated only one time, documented, interrupted or changed, and at the completion of the ex-
and maintained on file. amination or series of examinations.
T-777.2 Fluorescent Magnetic Particles With Black
T-778 DEMAGNETIZATION
Light. With fluorescent magnetic particles, the process
is essentially the same as in T-777.1, with the exception When residual magnetism in the part could interfere
that the examination is performed using an ultraviolet with subsequent processing or usage, the part shall be de-
light (i.e., nominal 365 nm), called black light. The exam- magnetized any time after completion of the examination.
ination shall be performed as follows:
(a) It shall be performed in a darkened area. T-779 POST-EXAMINATION CLEANING
(b) Examiners shall be in a darkened area for at least When post-examination cleaning is required, it should
5 min prior to performing examinations to enable their be conducted as soon as practical using a process that
eyes to adapt to dark viewing. Glasses or lenses worn does not adversely affect the part.
by examiners shall not be photochromic or exhibit any
fluorescence.
(c) B l a c k l i g h t s s h a l l a c h i e v e a m i n i m u m o f
T-780 EVALUATION
1000 μW/cm2 on the surface of the part being examined (a) All indications shall be evaluated in terms of the ac-
throughout the examination. ceptance standards of the referencing Code Section.
(d) Reflectors, filters, glasses, and lenses should be (b) Discontinuities on or near the surface are indicated
checked and, if necessary, cleaned prior to use. Cracked by retention of the examination medium. However, loca-
or broken reflectors, filters, glasses, or lenses shall be re- lized surface irregularities due to machining marks or
placed immediately. other surface conditions may produce false indications.

201
ARTICLE 7 ASME BPVC.V-2015

(c) Broad areas of particle accumulation, which might T-792 RECORDING OF INDICATIONS
mask indications from discontinuities, are prohibited, T-792.1 Nonrejectable Indications. Nonrejectable in-
and such areas shall be cleaned and reexamined. dications shall be recorded as specified by the referencing
Code Section.
T-790 DOCUMENTATION T-792.2 Rejectable Indications. Rejectable indica-
T-791 MULTIDIRECTIONAL MAGNETIZATION tions shall be recorded. As a minimum, the type of indica-
TECHNIQUE SKETCH tions (linear or rounded), location and extent (length or
diameter or aligned) shall be recorded.
A technique sketch shall be prepared for each different
geometry examined, showing the part geometry, cable ar-
T-793 EXAMINATION RECORDS ð15Þ
rangement and connections, magnetizing current for each
circuit, and the areas of examination where adequate field For each examination, the following information shall
strengths are obtained. Parts with repetitive geometries, be recorded:
but different dimensions, may be examined using a single (a) the requirements of Article 1, T-190(a)
sketch provided that the magnetic field strength is ade- (b) magnetic particle equipment and type of current
quate when demonstrated in accordance with T-756.2. (c) magnetic particles (visible or fluorescent, wet or
dry)
(d) map or record of indications per T-792
(e) material and thickness
(f) lighting equipment

202

You might also like