You are on page 1of 8

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/351750112

Performance characterization of non-contact infrared thermometers (NCITs)


for forehead temperature measurement

Article  in  Medical Engineering & Physics · May 2021


DOI: 10.1016/j.medengphy.2021.05.007

CITATIONS READS

4 26

7 authors, including:

Nathanael Seay Liang Zhu


University of Maryland, Baltimore County 101 PUBLICATIONS   1,397 CITATIONS   
3 PUBLICATIONS   14 CITATIONS   
SEE PROFILE
SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Medical Countermeasures View project

Development of test methods to evaluate the risk of cancer-cell permeation through tissue containment bags during laparoscopic power morcellation of uterine
fibroids View project

All content following this page was uploaded by Jean E Rinaldi on 20 June 2022.

The user has requested enhancement of the downloaded file.


Medical Engineering and Physics 93 (2021) 93–99

Contents lists available at ScienceDirect

Medical Engineering and Physics


journal homepage: www.elsevier.com/locate/medengphy

Performance characterization of non-contact infrared thermometers


(NCITs) for forehead temperature measurement
Stacey JL Sullivan a, Nathanael Seay b, Liang Zhu b, Jean E Rinaldi a, Prasanna Hariharan a,
Oleg Vesnovsky a, LD Timmie Topoleski a, b, *
a
Office of Science and Engineering Laboratories, USFDA, 10903 New Hampshire Avenue, Building 62, Silver Spring, MD 20993, United States
b
Department of Mechanical Engineering, UMBC, 1000 Hilltop Circle, ENGR225K, Baltimore, MD 21250, United States

A R T I C L E I N F O A B S T R A C T

Keywords: The ability to assess the performance of a non-contact infrared thermometer (NCIT) may be limited due to the
Thermometer algorithms necessary to predict a reference site temperature (e.g., oral) from a measurement of the forehead skin
Non-contact temperature. The algorithm not only adjusts for the difference between the reference site temperature and
Infrared
forehead temperature, but may also account for hardware corrections, bias adjustments and emissivity settings.
Blackbody
These algorithms are proprietary to the manufacturer and can be unique for each device. ASTM E1965-98 (2016)
Performance
is a standard test method for the evaluation of NCITs. It includes forehead thermometers; however, the algorithm
must be known or an unadjusted calibration mode must be accessible. This study evaluates 6 NCIT models (10
units of each) against the ASTM standard error criterion using a blackbody source. Units were tested within the
manufacturer’s operating and temperature measurement range specification. A method to evaluate measurement
outliers and characterize each model’s performance when the adjustment algorithm is unknown is proposed.
Using this method, 5 of the 6 models had a predicted error > 0.3◦ C.

1. Introduction site temperature (an internal body temperature), especially when the
ambient environment is cold and windy [7].
Non-contact infrared thermometers (NCITs) used for forehead skin NCITs include compensation circuitry to correct for detector prop­
temperature measurement are widely-accessible and convenient for erties, surface emissivity, and ambient conditions. In addition, the
determining a reference body site temperature (e.g., oral temperature) manufacturers incorporate an algorithm to adjust for the difference
and assessing fevers, particularly with parents and caregivers in the between the forehead skin surface temperature and the expected target
home. NCITs have gained popularity in pediatrics, especially for infants reference site temperature (the “adjusted mode,” typically referred to as
and children under five years of age for whom oral or rectal temperature “subject mode” for most NCITs). Most models also include an unadjusted
measurement is invasive, uncomfortable, difficult, time-consuming, and mode (“object mode” or “surface mode”). These NCIT software and
may be contraindicated [1–3]. In addition, non-contacting temperature design features are proprietary to each manufacturer. NCIT temperature
measurement reduces the risk of infectious disease transmission among measurement can also be influenced by the device’s distance and angle
individuals [4]. Such design features make the use of NCITs attractive from the target surface and air movement across the target surface.
for screening at ports of entry during infectious disease outbreaks [5]. Thus, the manufacturer provides instructions for the distance from the
NCITs derive oral temperature based upon detection of infrared surface and the proper positioning to obtain the temperature measure­
radiant energy from the surface of the skin. Measurements are taken ment in a prescribed environment. A feature that is notably missing from
most commonly at the forehead, although the temporal artery may also most NCITs is a means for the user to verify if the NCIT is working
be used [6]. Temperature measured using an NCIT is affected by the properly and providing an accurate measurement.
infrared detector properties, the emissivity of the surface and skin The accuracy and reproducibility of the temperature measured using
characteristics, and the ambient environment. The skin surface tem­ a specific NCIT model or among different models of NCITs are not well-
perature may deviate several degrees Celsius from the reference body characterized. Prior to initiating a clinical trial of multiple NCIT models,

* Corresponding author.
E-mail addresses: stacey.sullivan@fda.hhs.gov (S.J. Sullivan), topoleski@umbc.edu (L.T. Topoleski).

https://doi.org/10.1016/j.medengphy.2021.05.007
Received 26 October 2020; Received in revised form 24 April 2021; Accepted 10 May 2021
Available online 21 May 2021
1350-4533/© 2021 Published by Elsevier Ltd on behalf of IPEM.
S.J. Sullivan et al. Medical Engineering and Physics 93 (2021) 93–99

we decided to evaluate each model in a laboratory setting. One of the where j = 1…8 are the repeated measurements for each TBBS and the
test methods prescribed in ASTM E1965-98:2016 Standard Specification DUT = 1…10 for each model: A, B, C, D, E or F. To allow for the required
for Infrared Thermometers for Intermittent Determination of Patient Tem­ rest time between measurements as prescribed by the manufacturer in
perature is to assess whether an individual NCIT model meets the mea­ the instructions for use, a single temperature measurement was taken by
surement error acceptance criterion of ± 0.3 ◦ C (± 0.5 ◦ F) when each DUT in the set (e.g., 1A, 1B, 1C, 1D, 1E, and 1F) before repeating
evaluated within the manufacturer’s specified operating ambient tem­ the process for a total of 8 measurements per DUT per TBBS. During this
perature conditions and measurement range using a laboratory black­ time, the ambient room temperature was constant within ±0.5◦ C. The
body [8]. The standard requires that the temperature adjustment average difference for a single TBBS for each DUT was then calculated:
algorithm be disclosed or the model operates in an optional unadjusted (∑ )/
mode. ΔDUTave = ΔDUT,j n (2)
The goal of this study was to evaluate the performance of NCITs
using a blackbody source when the adjustment algorithm is unknown by where j = 1…8 are the repeated measurements for each given TBBS, DUT
predicting the offset algorithm and minimizing operator bias. Using a = 1…10 for each model A, B, C, D, E or F, and n is the total number of
controlled, uniform, horizontally-mounted infrared blackbody radiation measurements.
source (prescribed in ASTM E1965), we assess the in-vitro NCIT mea­ The manufacturer’s instructions for use included a measurement
surement variability and reproducibility within a specific NCIT model range and its accuracy (Table 1). All NCIT models had a manufacturer’s
and across different models of NCITs when tested in the adjusted mode. specified measurement resolution of 0.1◦ C.
The measurement difference between the NCIT and the blackbody was The “measurement range” stated in the manufacturer’s instructions
determined; however, the error remains unknown because the algorithm for use is the range of temperatures for which the NCIT will accurately
for each model is not accessible. To evaluate the error, a post-processing measure temperature. An NCIT’s adjusted measurement output is dis­
and analysis method is proposed to evaluate NCIT performance when played using an algorithm to estimate the oral temperature from the
the manufacturer’s algorithm is unknown. measured forehead skin temperature. The temperature of the forehead
skin is always less than the oral temperature [7], therefore, the NCIT
2. Materials and methods adjusts the display output by adding to the forehead skin temperature
measurement. The TBBS represents the forehead skin temperature and
2.1. Procedure not the oral temperature, therefore the TBBS is always lower than the
NCIT display output.
The temperature measurement performance of 60 commercially
available NCITs (6 models, each from a different manufacturer, 10 units
2.2. Measurement exclusions
of each model) was evaluated using a commercially available blackbody
source (BBS). All NCITs were purchased from commercial vendors. The
Any displayed temperature measurement outside of the NCIT man­
thermometers are designed to display the same temperature that would
ufacturer’s measurement range was excluded from analysis. If more than
be measured by an oral thermometer by measuring the skin temperature
half of all measurements for a single model at a single TBBS were
at the center of the forehead (adjusted mode). Ten NCIT sets were
excluded, the TBBS was declared outside of the NCIT’s allowable range
assembled and each NCIT was uniquely identified. Every set contained 6
and all were excluded (noted in Table 2 as a “0”). All remaining data sets
thermometers, one of each NCIT model (e.g., Set 1 contained device
comprised a minimum of 6 data points.
under test (DUT): 1A, 1B, 1C, 1D, 1E, and 1F). Thermometers were
The 8th measurement for DUT 1F was discarded due to an incomplete
cleaned and prepared according to the manufacturer’s instructions for
reference temperature reading.
use and had fresh batteries installed prior to testing.
A single BBS (Kambic, Model OB-15/2 BBLT; Semic, Slovenia) was
2.3. Data Post-Processing and Analysis
used to provide a uniform, stable source of infrared radiation to assess
each DUT. The BBS blackbody cavity was horizontally mounted in the
The original difference calculation (ΔDUT, Eq. (1)) includes the
BBS (for measurement access from the side of the BBS, as opposed to the
manufacturer’s temperature offset (Δoffset) as determined by their algo­
top of the BBS) and fully immersed in a stirred water bath. The manu­
rithm. In addition, any error introduced by the difference between the
facturer’s specifications stated stability and uniformity of the bath was
device emissivity value set by the manufacturer and the actual emis­
at least ± 0.007◦ C and that the blackbody source conformed to ASTM
sivity of the measured surface (δemissivity) is also present in the DUT
E1965-98:2016 Annex A1 [8]. Each NCIT was tested in adjusted mode at
measurement difference, ΔDUT. The remaining portion of ΔDUT is
BBS target temperatures (TBBS) of 30, 32, 34, 36, 37, 38, 40, and 42◦ C.
attributed to measurement error (δDUT) for the DUT.
The BBS was set to the target temperature, TBBS, for at least 30 minutes,
( )
until it reached equilibrium as verified by a reference thermometer ΔDUT = Δoffset + δemissivity + δDUT (3)
(TREF). The reference thermometer was a calibrated NIST traceable
thermistor probe (VRW digital thermometer 89369-140, Radnor, PA). where:
The reference thermometer measurement accuracy was ±0.05◦ C with a
resolution of 0.0001◦ C and an uncertainty of ± 0.014◦ C for the mea­ Δoffset = intentional temperature adjustment by manufacturer’s
surement range of 0◦ C to 100◦ C. Prior to testing, the room temperature algorithm
and relative humidity (%RH) were recorded to verify whether the δemissivity = error due emissivity mismatch [8]
environmental conditions were within the operating range of all NCIT δDUT = measurement errors (imperfect positioning, lens degradation,
models. Measurements for each DUT at a specific TBBS were completed etc.)
on the same day by the same operator. All DUTs were stored in the
laboratory where the testing was performed. Eight temperature mea­ The algorithm used to determine the adjusted value was not given by
surements were collected from each DUT at each TBBS. The reference any of the manufacturers, and 5 of the 6 manufacturers did not provide
temperature, TREF,j, was recorded for every individual temperature their NCIT emissivity setting. Although these values are unknown, the
measurement for each DUT (TDUT,j) and used to determine the individual output of each device is still representative of its performance. For each
measurement difference for that DUT (ΔDUT,j) as: combination of TBBS and DUT, the BBS set up was the same and room
⃒ ⃒ temperature and relative humidity were considered constant. (All DUTs
ΔDUT,j = ⃒TDUT,j − TREF,j ⃒ (1)
were tested within a room temperature range: 21.54◦ C - 24.98◦ C and not

94
S.J. Sullivan et al. Medical Engineering and Physics 93 (2021) 93–99

Table 1
Manufacturer’s labeling for measurement range and accuracy.
Model A Model B Model C Model D Model E Model F

Adjusted mode accuracy, C ◦


0.2 0.2 0.3 (34.4-35.9), 0.2 (36-39), 0.3 (39.1-42.2) 0.2 0.3 0.3 (34-35.9), 0.2 (36-39), 0.3 (39.1-42.5)
NCIT Measurement Range, ◦ C 34-43 32-42.9 34.4-42.2 34.4-42.2 32-42.5 34-42.5

(∑ )/
Table 2 ΔDUTave,i = ΔDUTi,j ni (6)
Number of data points per NCIT model per TBBS within the manufacturer’s
measurement range.
δDUT,i,j = ΔDUT,i,j − ΔDUTave,i (7)
Model A Model B Model C Model Model E Model F
D ⃒( )⃒
δDUTmax,i = max⃒ δDUT,i,j ⃒ (8)
TBBS ¼ 80 80 80 0 79 76
30◦ C
where j = 1…8 are the repeated measurements for each given TBBS and n
TBBS ¼ 80 80 80 80 80 80
32◦ C = the number of measurements made for a DUT at a specific TBBS.
TBBS ¼ 80 80 80 80 80 80 The average difference is a representation of the repeated measure­
34◦ C ment of a DUT at a specific TBBS (ΔDUTave,i , Eq. (6)). To reduce the in­
TBBS ¼ 80 80 80 80 80 80 fluence of individual δDUT,i,j (Eq. (7))data points suspected of being
36◦ C
TBBS ¼ 80 80 80 80 80 80
anomalous, we calculated the absolute value of the difference between
37◦ C two adjacent data points in the ordered set for a single DUT at a single
TBBS ¼ 80 80 80 0 80 80 TBBS (Fig. 1). If the difference was greater than 0.3◦ C, the data point was
38◦ C removed from the set and the standard deviation of the remaining points
TBBS ¼ 80 0 0 0 80 0
was calculated. If the data point was outside 3 times the standard de­
40◦ C
TBBS ¼ 0 0 0 0 0 0 viation, the point was excluded from the set. If it was within 3 times the
42◦ C standard deviation of the set, the data point was returned to the set. All
Total 560 480 480 320 559 476 remaining data sets comprised a minimum of 6 data points.
The error for a single model can then be characterized by the
maximum absolute difference, δDUTmax, of all DUTs a single model for a
exceeding ±0.5◦ C for each set of measurements. %RH range: 23% –
single TBBS (Eq. (9)).
52%. Note: the %RH for the TBBS = 30◦ C was not taken until the
following day). Given these conditions were constant, the contribution
from the offset algorithm (Δ offset) and the error contribution from the 2.4. Statistical analysis
emissivity setting (δemissivity) remained constant during a single test
period. An analysis of variance (ANOVA) combined with post-hoc pairwise
The initial measurement difference (ΔDUT) is a sum of the intended analysis (Tukey’s Test) was performed. P-values less than 0.05 were
offset value (Δoffset), the error associated with the emissivity mismatch (δ considered statistically significant.
emissivity) and measurement error (δDUT) (Eq. (3)). To characterize the
error of the device, the combined value of the intended offset value and 3. Results
emissivity mismatch (Δoffset + δ emissivity) must be predicted and removed.
The remaining variation can then be used to characterize the measure­ A total of 3,840 temperature measurements were collected, of which
ment error of the DUT (δDUT). 2,875 resulted in temperature performance data that were evaluated.
To obtain the best characterization of the δDUT, the best estimate of The majority of excluded measurements was attributed to falling outside
(Δoffset + δ emissivity) must be calculated. For each data set, ΔDUT,j (Eq. (1)), of the DUT’s measurement range (Table 2). The average difference,
collected for a single DUT at a single TBBS, we assume that (Δoffset + δ ΔDUTave, is shown in Fig. 2 as a function of TBBS.
The test for outliers identified 43 data points > 0.3◦ C from their
emissivity) (from Eq. (3)), is a constant (A) because all influencing factors
(ambient temperature, ambient humidity, TBBS, BBS, power source) are adjacent data point. Of these, 41 were outside of 3 times the standard
kept constant and the remaining variable is the measurement error, deviation of the group and removed from the data set (Table 3).
δDUT. Using the maximum error Eq. (6) across all TBBS for each DUT, 5 of
the 6 NCIT models did not meet the ASTM E1965 criterion specifying
ΔDUT = A + δDUT (4) that the error should be no greater than 0.3◦ C when tested within the
manufacturer’s specified operating ambient conditions over the device’s
where the constant A = (Δ offset + δemissivity) for a single data set (j = 1...8) displayed temperature range [8]; these values are highlighted in orange

and from Eq. (2) (where ΔDUTave = ( ΔDUT,j) /n): in Table 4. Model B met the ASTM criterion both with and without ex­
(∑ ) / clusions applied. However, error values highlighted in purple exceeded
ADUT,i = ΔDUTave,i = ΔDUT,j i ni (5)
their manufacturer’s specification while remaining within the ASTM
criterion (Table 1).
where i is the TBBS (30, 32, 34, 36, 37, 38, 40, and 42◦ C), and n is the To determine if different models reported a different output display
total number of measurements for the specific DUT (e.g., 1A). temperature for the same input temperature (BBS set temp), a pairwise
Therefore, to quantify the individual measurement error (δDUT,i,j, Eq. ∑
comparison between the means of the difference ([ (TDUT-Tref)]/n) of
(7)), the individual measurement difference (ΔDUT,j Eq. (1)), is reduced all 10 samples for each model (A, B, C, D, E, and F) at each specific BBS
by the value of the average difference (ΔDUTave, Eq. (6)) for a single DUT set temperature (TBBS = 30, 32, 34, 36, 37, 38, 40, or 42◦ C) was made
for a single TBBS. The removal of ΔDUTave,i represents the removal of the (Table 5).
offset algorithm and emissivity error contributions. The maximum ab­
solute value of all individual DUT measurement error represents the
4. Discussion
device error (δDUTmax,i, Eq. (8)).
For a specific TBBS (i = 30, 32, 34, 36, 37, 38, 40, and 42◦ C):
This study assessed the performance of consumer-grade NCITs using

95
S.J. Sullivan et al. Medical Engineering and Physics 93 (2021) 93–99

Idenficaon of Outliers
Example: Model A, TBBS = 37°C
Outlier
0.4 adjacent value difference ≤ 0.3°C
δDUTj = ΔDUTj - ΔDUTave, j = 1...8

0.3
0.2
0.1
0
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
1 2 3 4 5 6 7 8 9 10
DUT Number

Fig. 1. Example of identification of outliers for a single model (model A) at a single TBBS (37◦ C).

7
Average Difference
per NCIT Model per TBBS
Δave = ∑|TDUTj – TBBj| /n, C°

6
A B C D E F
5

0
30 32 34 36 37 38 40
Blackbody Set Temp, °C
Fig. 2. Average difference, for each NCIT model per BBS set temperature, TBBS.

a blackbody source (BBS) in a typical user environment verified to be temperature was the measured forehead temperature adjusted to the
within the NCIT operating range. This assessment was carried out to oral temperature value using the offset algorithm determined by the
characterize and understand the performance of each device in a manufacturer. Although the output may not appear accurate (because it
controlled laboratory setting. is different from the reference temperature), the measured difference
Using the proposed evaluation method, this study found that 5 of the was a repeatable value with an associated variation. Our post-processing
6 NCIT models did not meet the ASTM acceptance criterion for error. All and analysis isolated this variation to assess the error associated with
laboratory measurements produced a calculated difference (Eq. (1)) that repeated temperature measurement.
was outside of the allowable standard error, however this difference A limitation of using δDUTmax to estimate error is that it may under­
alone cannot be interpreted as non-compliance. This is because every state or overstate the error value if there was a bias in the measurement.
NCIT was tested in its adjusted mode and therefore every output display It is possible that, if the DUT error (δDUT) was consistently biased, the

96
S.J. Sullivan et al. Medical Engineering and Physics 93 (2021) 93–99

Table 3
Data exclusions and outliers.
Model A Model B Model C Model D Model E Model F

Excluded TBBS sets (n¼80); (TDUT outside the 42 C



40 C,

40 C,

30 C,40 C, 42 C
◦ ◦ ◦
42 C

40◦ C, 42◦ C
meas. range.) 42◦ C 42◦ C
Outliers per TBBSDUT; #Excluded| TBBS ¼ 1A; 1|14A; 0| (No data)
#Tested 30◦ C 110A; 2|2
TBBS ¼ 2E; 1|1
• 6 NCIT models 32◦ C
• 10 DUTs per model TBBS ¼ 4D; 1|16D; 2|2
“4A” = sample #4 of model A 34◦ C
TBBS ¼ 2A; 2|29A; 1|1 4D; 1|16D; 2|27D; 1|1 7F; 1|1
36◦ C
TBBS ¼ 3A; 1|17A; 1|1 5C: 1|1 2D; 1|13D; 1|16D; 1|28D; 1| 10E; 1|1
37◦ C 19D; 2|2
TBBS ¼ 1A; 1|1 5A; 1|1 7C; 1|1 (No data) 2E; 1|15E; 1|19E; 3F; 3|3*5F; 2|27F;
38◦ C 1|1 3|3*
TBBS ¼ 2A; 1|16A; 1|1 (No data) (No data) (No data) (No data)
40◦ C
TBBS ¼ (No data) (No data) (No data) (No data) (No data) (No data)
42◦ C
Number of outliers tested 13 0 2 14 5 9
Number of outliers excluded 12 0 2 13 5 9
Data points per modelᵻ 560 480 480 320 559 476
% of excluded outliers 2.1% 0.0% 0.4% 4.1% 0.9% 1.9%

*indicates that the 3 exclusions resulted in an n<6 and the exclusion of the entire data set for that TBBS.

Number of data points per model prior to exclusions.

Table 4
Maximum error (δDUTmax ; Eq. (8)) per DUT.

bias would be included in the constant, A. This would incorrectly average of their data set, indicating that the outliers were biased to read
decrease δDUT,j if the bias was to increase the temperature measurement. a lower temperature than the majority of readings. The purpose of
Conversely, a bias that decreased the temperature measurement would removing the outliers was to reduce the influence of bias when evalu­
incorrectly increase δDUT,j. ating the error. Removing the outliers decreased the number of DUTs
The test for outliers used in this study identified 43 data points, 41 of that did not meet the ASTM criterion of δDUTmax ≤ 0.3◦ C. However, when
which were excluded. Of the 41 exclusions, 38 (92%) were less than the the data were pooled from all ten DUTs of each model, 5 of the 6 models

97
S.J. Sullivan et al. Medical Engineering and Physics 93 (2021) 93–99

Table 5
Significant differences between NCIT models. Evaluation of whether different models reported a different output display temperature for the same input (TBBS)
temperature.

still did not meet the criterion. Model B had no outliers identified. In found the predictive algorithms deviated from a NIST calibrated refer­
contrast, Model D had the highest percentage of outlier exclusions, at ence by more than 1◦ C [9]. In another study by Tsai comparing infrared
over 4%. Models A and F had ~ 2% of their readings excluded and the ear thermometers, 4 of 14 DUTs did not meet the ASTM criterion when
exclusions for Models C and E were < 1%. Loss of infrared energy evaluated using a blackbody source with a calibrated reference [10].
detected by the device would result in a lower temperature read by the It would be expected that each unit of every model has the same
device, therefore, these outliers may be an indicator of device usability display output for the same TBBS, just as a temperature measurement
issues. For example, an NCIT that is not easy to grip and operate could taken by two different NCITs of a single forehead would be expected to
result in misalignment with the targeted measurement area. be the same. However, the results showed the display temperatures to be
When taking a subject’s forehead temperature, losses may occur if significantly different for different NCITs evaluated at the same TBBS
external conditions are not controlled (e.g., excessive air movement), if (Table 5) and that no model fully replicated the output of another. The
the distance to the target area is not compliant with the manufacturer’s inter-model variability in the measured temperature could be attributed
instructions, or if the surface is obscured (sweat, make-up or lotion on a variety of sources: (a) usability or model-to-model variability
the skin). These conditions can affect the NCIT measurement of the (Table 3), (b) differences between manufacturer’s algorithms and
forehead surface temperature and introduce error into the offset algo­ emissivity settings, and (c) differences in the calibration set up used by
rithm that assumes clean dry skin of a given emissivity. For our labo­ the manufacturers (for estimating the offsets) and those used in this
ratory test, the design of the orifice of the horizontally-mounted conical study.
cavity blackbody source naturally isolates the NCIT sensor from unin­
tended radiation sources. The orifice of the blackbody is a single port, 5. Conclusions
such that when the face of the NCIT is placed against the port, the sensor
is exposed only to the blackbody radiation. Overall, many, but not all, of This study elucidates the challenges of performing NCIT measure­
the error sources in a typical use environment are mitigated or elimi­ ment evaluations. The method used in this study allows for the evalu­
nated by the use of the blackbody cavity. ation of NCIT error without knowing the manufacturer’s algorithm;
No other previous studies were discovered to specifically evaluate however, this method is cumbersome as it still requires the use of a
the performance of NCIT forehead thermometers against an ASTM reference infrared radiation source (e.g., BBS). Five of the 6 models
blackbody source. This is likely due to the limited interpretations that evaluated had multiple outlier measurements identified and removed to
can be made if the offset algorithm is unknown. ASTM E1965 provides a reduce the influence of measurement bias. Nonetheless, these same 5
test method for the evaluation of NCIT performance; however, this models still did not meet the ASTM criterion, even after the outliers were
method requires a controlled infrared radiation source and the adjust­ removed. Ideally, a clear means for the consumer to verify whether an
ment algorithm to be known. For 5 of the 6 models, an object mode is a NCIT is working properly and providing an accurate measurement is
feature of the NCIT; however, it is intended for measuring objects or preferred. Model-to-model variability in displayed temperature is a
fluids. In each case, the object mode has a wider accuracy specification major source of concern and should be properly evaluated to ensure that
(± 0.5 to 1.0◦ C) for the same measurement range. In addition, the NCIT NCIT displayed temperatures are credible and consistent between
may use a different emissivity setting from the adjusted mode that is models. A clinical study using the same sets of DUTs will be conducted to
intended specifically for the forehead skin, making it inappropriate for further evaluate their performance.
verification of the adjusted mode.
The findings of this study are consistent with past studies comparing 6. Disclaimer
predictive thermometers to a reference thermometer. In a prior inves­
tigation, an evaluation of 3 fast read (contact) digital thermometers The mention of commercial products, their sources, or their use in

98
S.J. Sullivan et al. Medical Engineering and Physics 93 (2021) 93–99

connection with material reported herein is not to be construed as either References


an actual or implied endorsement of such products by the Department of
Health and Human Services. [1] Fortuna EL, et al. Accuracy of non-contact infrared thermometer versus rectal
thermometry in young children evaluated in the emergency department for fever.
J Emerg Nurs 2010;36(2):101–4.
Funding [2] Franconi I, et al. Digital axillary and non-contact infrared thermometers for
children. Clin Nurs Res 2018;27(2):180–90.
[3] Wang K, et al. Non-contact infrared thermometers for measuring temperature in
None. children: primary care diagnostic technology update. The British journal of general
practice: the journal of the. R College General Practitioners 2014;64(627):e681–3.
Ethical approval [4] Ng DK, et al. Non-contact infrared thermometry temperature measurement for
screening fever in children. Ann Trop Paediatr 2005;25(4):267–75.
[5] Bitar D, Goubar A, Desenclos JC. International travels and fever screening during
Not required. epidemics: a literature review on the effectiveness and potential use of non-contact
infrared thermometers. Euro Surveill 2009;(6):14.
[6] Pompei F, Pompei M;. Non-invasive temporal artery thermometery: physics,
Declaration of Competing Interests
physiology and clincial accuracy. In medical thermometry for SARS detection, SPIE
defency and security symposium. 2004. Orlando, FL.
None declared; Funding: None; Ethical approval: Not required. [7] Vesnovsky O, et al. Modeling of differences between body core and forehead
temperatures measured by infrared thermometers. In: 2017 Design of Medical
Devices Conference; 2017.
Acknowledgements [8] ASTM. In: Standard Specification for Infrared Thermometers for Intermittent
Determination of Patient Temperature; 2016. E1965–98. 2016ASTM International
This project was supported through an FDA Medical Countermea­ West Conshohocken, PA.
[9] Vesnovsky O, et al. Performance testing of fast read digital thermometers. J Med
sures Initiative (MCMi) challenge grant and the Oak Ridge Institute for Devices-Trans Asme 2014;8(3):3.
Science and Education (ORISE) Research Participation Program at [10] Tsai SF. Comparison measurements of infrared ear thermometers against three
USFDA/CDRH. The authors thank FDA researchers Jon Casamento and types of blackbody sources. J Int J Thermophys 2010;31(8):1821–31.
Guha Suvajyoti for manuscript review.

99

View publication stats

You might also like