Professional Documents
Culture Documents
Note: 51258
Biodiesel (FAME) Analysis by FT-IR
Mike Bradley, Ph.D., Thermo Fisher Scientific, Madison, WI, USA
Figure 3: Spectrum of FAME collected on the 45° and 60° ARK plates
The key concern for testing labs is that the variability The methods developed thus far have focused only on In addition to these
in petroleum diesels from region to region must be the FAME content, in keeping with the work being outlined offices, Thermo Fisher
modeled for the analysis to be useful. Initial predictions by the ASTM D02 committee. However, FT-IR can be used Scientific maintains
(using a calibration based on one petroleum stock) of a to analyze for many other components, including free fatty a network of represen-
set of biodiesel blends made with a different petroleum acids or glycerol, with proper calibration. A generalized tative organizations
base gave poor results, as shown in Table 1, due to the method could be developed given a proper set of standards. throughout the world.
un-modeled behavior of the petroleum base. Incorporation
of only three samples with the different petroleum base Conclusions
improved these predictions dramatically, as seen in column Biodiesel is appearing more commonly in the media, and
three of the table. None of the biodiesel stocks examined is evoking interest at many levels. The need for a rapid
thus far shows extreme variation. and simple analysis tool for FAME is filled very effectively Australia
by the Nicolet FT-IR spectrometer family coupled with the +61 2 8844 9500
Actual B% Initial Prediction Inoculated Method Austria
unique Smart ARK ATR accessory. Our product offering +43 1 333 50340
18.00 15.95 17.92 includes a booklet that covers the analysis of biodiesel, Belgium
including a step-by-step procedure for data collection. +32 2 482 30 30
20.00 14.36 20.01 Canada
With the addition of the OMNIC and TQ Analyst +1 800 532 4752
2.00 1.21 2.03 software packages, Thermo Scientific’s biodiesel analysis China
package provides the scientific knowledge and +86 10 5850 3588
Table 1: Data showing effect of training a PLS method with standards
Denmark
covering petroleum base variability. instrumental capabilities needed for this analysis. +45 70 23 62 60
France
Transfer of the calibration between spectrometers can +33 1 60 92 48 00
Germany
be addressed with the Thermo Scientific ACLS algorithm, +49 6103 408 1014
developed by Sandia National Laboratories. Figure 6 India
shows data taken from two spectrometers, one being used +91 22 6742 9434
Italy
to predict the other. A small offset is visible. The ACLS +39 02 950 591
algorithm removes this offset cleanly. Thus, once a Japan
calibration is built on one spectrometer, methods can be +81 45 453 9100
Latin America
transferred to another spectrometer by “inoculating” the +1 608 276 5659
method with data from the second unit, as is frequently Netherlands
done in FT-NIR applications. +31 76 587 98 88
South Africa
+27 11 570 1840
Spain
+34 91 657 4930
Sweden / Norway /
Finland
+46 8 556 468 00
Switzerland
+41 61 48784 00
UK
+44 1442 233555
USA
+1 800 532 4752
www.thermo.com