Professional Documents
Culture Documents
Electron Diffraction
31/10/2021
Micro= small/tiny + Scope= to view/to observe Electron microscopes were developed due to the
limitations of Light Microscopes which are limited
by the physics of light.
1928-1931: Knoll and Ruska, in Berlin began
development of electron lenses and built a prototype
of Electron Microscope
An instrument used for viewing very small objects, such as mineral samples or
animal or plant cells, typically magnified several hundred times.
1986: Nobel Prize in physics for Ernst Ruska
(TEM), Gerd Binnig & Heinrich Rohrer (STEM)
31/10/2021 31/10/2021
LIGHT MICROSCOPE VS TEM LIGHT MICROSCOPE VS TEM
31/10/2021
WHAT DO WE KNOW ABOUT ELECTRONS? WHAT DO WE KNOW ABOUT ELECTRONS?
Wave Characteristics of electron gives rise to coherent scattering Particle characteristics gives rise to
(diffraction) characterise X-Rays
Electron Duality:
Behave as waves Diffraction patterns when passing by crystals
Behave as particles excitation of inner shell électrons: X-Ray chemical Analysis
31/10/2021 31/10/2021
For electron microscopist, relativistic effect cannot be ignored at energies >= 100keV
31/10/2021 31/10/2021
INTERACTIONS ELECTRONS-MATTER
Multitude of signals resulting from the electron matter interaction
TEM TECHNIQUES
Crystallographic informations:
Electron diffraction
Selected Area Diffraction (SAD)
Convergent beam electron diffraction (CBED)
Imaging
EDX Low magnification mode
Dark field / Bright field imaging
High resolution (transmission) electron microscopy (HRTEM or HREM)
Chemical analysis:
X-rays
Energy dispersive X-ray spectroscopy (EDS or EDX)
Electrons
Electron energy loss spectroscopy (EELS)
EELS
Electron Diffraction
STEM/HAADF
High Resolution Electron Microscopy
31/10/2021
31/10/2021
A1.5+xSb+5−2xAl5O11.5 (A = Sr, Ba) phases A. Letrouit et al., Solid State Sciences, V10, 2008, P982-990
Homogeneous precipitation of δ-Ni2Si (orthorhombic)
A1.5+xSb+5−2xAl5O11.5 (A = Sr, Ba) phases in Cu [T. EPICIER, Z. SUN, MATEIS (2007)]
Lattice parameters, reflections condition, possible space(s) group(s) A. Letrouit et al., Solid State Sciences, V10, 2008, P982-990
control of the sample: good sample preparation, homogeneity, number of characteristic phases
Imaging Chemical Analysis
AlK SiK
A1.5+xSb+5−2xAl5O11.5 (A = Sr, Ba) phases MgK
b
a
CuL CuK
INSTRUMENTATION
Chemical Analysis
Electron Energy-Loss Spectroscopy (EELS)
Some of the electrons will undergo inelastic scattering, which means that Basic requirements:
they lose energy and have their paths slightly and randomly deflected Voltage: 200-300kV
Brillant source: W, LaB6, FEG
Essential extras
The Electron Energy Loss (EEL) spectrum can be Diffraction and image simulation software
roughly split into two different regions: the low- Image processing software
loss spectrum (up until about 50eV in energy
loss) and the high-loss spectrum. The low-loss
spectrum contains the zero-loss peak as well as Possible extras:
the plasmon(quantum of plasmon oscillation) Scanning transmission electron microscope (STEM)
peaks (quantum of plasmon oscillation) and
contains information about the band structure Energy dispersive X-Ray detector (EDS)
and dielectric properties of the sample. The Electron Energy Spectrometer (EELS)
high-loss spectrum contains the ionisation X-Ray and EELS mapping software
edges that arise due to inner shell ionisations in
the sample. These are characteristic to the Cold stage, tensile stage, heating stage…
species present in the sample, and as such can
be used to obtain accurate information about
the chemistry of a sample.
31/10/2021
INSTRUMENTATION
TEM TECHNIQUES
1: electron gun
2: anode
3: condenser(s) system
4: condenser diaphragm
5: thin object
6: objective lenses
7:objective diaphragm
8: image plan of the
objective
9: projection lenses
10: fluorenscent screen
11: vacuum pump
Specimen : Ø < 3 mm
thin thickness ≈ 10 - 100 nm
31/10/2021 31/10/2021
INSTRUMENTATION INSTRUMENTATION
31/10/2021 31/10/2021
INSTRUMENTATION
ELECTRON SOURCES
There are strict requirements to produce an electron beam with the necessary properties.
There are two types of source:
field-emission sources (or ‘guns’ as they are often called): fine tungsten (W) needle : W monocrystal
<310>//W<310> covered by a monolayer of ZrO
produces electrons when a large electric potential is applied between the thin W needle and an anode
31/10/2021 31/10/2021
31/10/2021 31/10/2021
ELECTRON SOURCES: COMPARISON BETWEEN THE MOST COMMON
SOURCES LENSES
Emission Thermionic Field emission In general:
Source W LaB6 Froide (W) Schottky (W/ZrO)
Lenses control basic operational functions in electron microscopes
Operating temperature (K) 2800 1400-2000 300 1800
Used to focus and magnify the electron beam and the images
Vacuum (Pa) 10-2 10-3 10-3.10-4 10-7 10-8 <10-6 Electromagnetic lenses are still very imperfect and cause errors
Needle radius 50 μm 10 μm 100 nm 100 nm
Very complex structure and way of working
Current density (A/cm2) 1-3 30 104 106 500 → Understanding the lens system is crucial in order to understand the basic structure of electron microscopes
Work Fonction (eV) 4.5 2.7 4.5 2.7 and being able to properly operate them.
31/10/2021 31/10/2021
LENSES LENSES
All rays from a point in the object are brought back to a point
in the image and all parallel rays (whether parallel to the optic
Simple reminder of light optics and Link between axis or not) are brought to a focus in a plane at a position
optical construct and diffraction pattern: depending on their angle to the axis.
Convex lens magnifies, focuses parallel rays to one point 3 important distances:
Electron lenses act as combination of convex glass lenses
1. the distance from the object plane to the lens (the object
(approx.)
distance do),
2. the distance from the lens to the image plane (the image
distance di),
A fundamental principle of how a lens works is that
an electron passing through the middle of the lens is 3. the distance from the lens to the back-focal plane (the
unaffected so we can draw its path as a straight line. focal length f)
All other electron paths are bent when they pass
through the lens.
Newton’s lens equation:
1/f= 1/d0+1/di
31/10/2021
OBJECTIVE LENS
CONDENSER LENS
Same structure as the condenser lenses, but producing a very strong field with a small
TEM lens systems usually consist of two or three condenser lenses and one objective lens. focal length
Performance of this lens determines the image quality (resolution, contrast, etc)
33 34
Image: Williams, Carter: Transmission electron microscopy, 1996 Image: www.jeol.co.jp/en/words/emterms/glossary_file/file/objective_lens.jpg
ELECTRON PATH THROUGH THE MAGNETIC FIELD ELECTRON PATH THROUGH THE MAGNETIC FIELD
31/10/2021
APERTURES APERTURES
An aperture is often inserted into a lens. The aperture limits the Diaphragms come in several forms
collection angle (β) of the lens as shown schematically in Figure 9 , depending on their function and
and such an aperture in the objective lens allows us to control the the particular microscope. They can
resolution of the image formed by the lens, the depth of field and the be either individual disks, each
depth of focus, the image contrast, the collection angle of the with a particular aperture diameter,
electron energy-loss spectrometer, the angular resolution of the or they can be a series of different
DP…. apertures in a single metal strip.
Physically, the aperture may reside above, in, or below the plane of
the lens.
The diameter can be as small as
Usually the apertures are circular holes in metal disks and the disks 10 μm, which is about the smallest
are made of either Pt or Mo, which are both refractory metals. circular aperture we can make
consistently or up to ≈ 0.3mm (300
μm).
31/10/2021
SPHERICAL ABERRATION CS
SPHERICAL ABERRATION CS
31/10/2021
CHROMATIC ABERRATION CC
CHROMATIC ABERRATION CC
We assumed that the electrons are monochromatic, but they are not really. However, we can make very good
high-voltage supplies and the variation of the electron energy due to the power supplies is usually smaller The objective lens bends electrons of lower
than one part in 106, which is 0.1 eV for a 100-keV beam. energy more strongly and thus electrons
from a point in the object once again are
As we already discussed before, depending on the electron source the actual energy spread in the beam may blurred to form a disk in the Gaussian
vary from 0.3 eV (cold FEG) to 1 eV(LaB6). This range is still so small that we generally do not have to worry image plane (and a smaller disk in the
about chromatic aberration affecting the image resolution. The exception is if you happen to have a Cs plane of least confusion).
corrector, in which case, after compensating for Cs, Cc is the next most-persistent aberration. Lens elements
that can correct for Cc are being developed (monochromator)
31/10/2021 31/10/2021
ASTIGMATISM
ASTIGMATISM
There are many ways to produce a TEM specimen; the one you choose will depend on the
information you need, time constraints, availability of equipment, your skill, and the
material…
One important point to keep in mind is that your technique must not affect what
you see or measure, or if it does, then you must know how. Specimen preparation
artifacts may be interesting, but they are not usually what you want to study.
The TEM specimen, when you made it, must be electron transparent (usually) and representative of the
The type of TEM specimen you prepare depends on what you are looking for so you need
material you want to study. In most cases (but not all) you would like your specimen to be uniformly thin,
stable under the electron beam and in the laboratory environment, conducting, and non-magnetic.
to think about the experiment that you are going to do before you start your preparation.
Few specimens approach the ideal and usually you have to makecompromise. In general, we can divide For example, is mechanical damage to be avoided at all costs, or can it be
specimens into two groups: self-supporting specimens and specimens resting on a support grid or thin tolerated so long as chemical changes don’t occur—or vice versa?
washer; the grid or washer is usually Cu but could be Au, Ni, Be, C, Pt, etc
Is the specimen at all susceptible to heat or radiation?
Which method you choose would depend on what you want to learn about your material. Depending on the answers to these questions, some of the following methods will
be inappropriate.
31/10/2021
SAMPLE PREPARATION REFERENCES
scienceportfolio1.weebly.com
http://www.nanoscience.gatech.edu/zlwang/image/research/tem/fig1.jpg
Physical Principle of Electron Microscopy, An Introduction to TEM, SEM and AEM, Second Edition, Egerton,
R.F. (2016), p.34
www.globalsino.com/EM/
PPT - Electron Microscopy: Lecture 1: Introduction to the Transmission Electron Microscope (TEM) PowerPoint
Figure 1: Summary of different preparation philosophies Presentation - ID:5955069 (slideserve.com)
31/10/2021