1. Main parameters:
1. A/D conversion:∑-Δ,resolution is 1,000,000,10 times/second。Signal input range:-19mV ~ +19mV
2. Load cell excitation voltage:DC +3V,capable of connecting 1 load cell of 350Ω, 4-wire.
+E -IN +IN ○
8 +IN
○
5 Shield
3.Calibration & inspection
Correctly connecting weighing indicator with load cell, power on the indicator, it begins self-inspection. When finishes, it enters
weighing mode (first power-on, it’s better to pre-heat weighing indicator and load cell for 15~30 minutes, then begin to calibrate).
Turn the calibration head up, then it enters calibration mode, as per the steps listed below you can finish the calibration:
Select the division by pressing Setup of division:press the key again and again to choose the division among
2 the key , then press press 1、2、5、10、20、50、100, 200, If confirmed, press then
[d 2]
to confirm
It enters to next step. I.e.: 2
Press to choose Setup of decimal point:Press again and again to choose the bit of decimal
the decimal point, then press point among 0、1、2、3、4. I.e.: 2
3 to confirm
[d P 2]
(Note:If add the bit of decimal point 5,it can stop the indicator entering the saving power
mode)
Set full capacity by Setup of full capacity:
pressing , Pressing to make flash bit move to right side, and pressing
4 [ F030.00]
To make flash bit move to left side. The number on flash bit can be added by pressing ,
Press to confirm then input full capacity in this way, later press to enter to next step. I.e.: 3000
After it gets stable, press Setup of zero point:Make sure there is no load on the scale。When the stable light
5 to confirm [ noLoAd ] is on,press to confirm the zero point,the indicator displays“ ---- ”about two seconds, then
4.Review inner code: Under weighing mode,press the two keys & together ,can enter or exist inner mode。(1d=20 inner codes)