DS451 (TI-1232) Scale key operation detail

KEY OPERATIONS :

 S-on Mode: Press [0 ] + [T] [] [] keys

 Technical specification Mode: Enable S-on mode


Press [0 ] + [] [T] [] keys.

 Customer specification Mode: Press [0 ] + [] [] [] keys.

 Count mode: Enable S-on mode, Press [0 ] +[] [][T] keys

 Calibration mode: Enable S-on mode, Press [0 ] + [] [T] [T]
keys.

 Set point Programming mode: Press [0 ] + [] [] [] keys.

Entering W&M specifications


S-on mode to be enabled to enter W&M Specifications (Technical Specification)

Sl.No OPERATION SEQUENCE DISPLAY REMARKS


1 Go to S-On mode. S – on Press [Rezero] key and Press
To Enter S-on mode Press [Tare],[Digit Increment] [Digit
[0 ] + [T] [] [] Select] Keys

2 Press T10 Depress Press[REZERO] key and


press [DIG-SEL] key, [TARE]
[0 ] +[] [T] [] key & [DIG-SEL] key in
XX sequence. The display toggles
between the messages Spec No
and the current value
3 T10 Using [DIG-SEL] and [DIG-INC]
[] [] key
Enter the required value
00
4 T11 Press * key to store spec value
entered and advance to next Spec
Press  Key
XX
5 T11 Follow the procedure mentioned
[] [] in SL.No 3 & 4. and Repeat the
same Up to T19
01
6 [T] S - on Press [TARE] key, to exit from
Technical specification mode
Entering Customer specifications
Sl.No OPERATION SEQUENCE DISPLAY REMARKS
1 Press C10 Depress [Rezero] key and Press
XX [Digit Select] Key 3times.
[0 ] + [][][] The display toggles between the
message Spec No and the
Current Value.
2 C10 Using [DIG-SEL] and [DIG-INC]
[] [] key
Enter the required value
02
4 C11 Press * key to store spec value
Press  key entered and advance
to next Spec
XX

5 C11 Follow the procedure mentioned in


[] [] SL.No 2 & 3. and Repeat the same
up to C19
11
6 0.000 Press [TARE] key, to exit from
[ T  ] Technical specification mode

Internal count access


This function is provided for checking of internal count of A/D

Sl.No OPERATION SEQUENCE DISPLAY REMARKS


1 Go to S-On mode. S – on Press [Rezero] key and Press
To Enter S-on mode Press [Tare],[Digit Increment] [Digit
[0 ] + [T] [] [] Select] Keys

2 Press XXXXXX Depress [REZERO] key and press


[DIG-SEL] key twice, [TARE] key
[0 ] + [][][T] once in sequence. The display
indicates the Span count.

3 Press [TARE] key, to exit from


0.000 Count mode and switch for
Press [T] Key Weighing mode.

Calibration procedure :
Calibration using full scale machine capacity / 1/3 rd of Machine Capacity

Sl.No OPERATION SEQUENCE DISPLAY REMARKS


1 Go to S-On mode. S – on Press [Rezero] key and Press
To Enter S-on mode Press [Tare],[Digit Increment] [Digit Select]
[0 ] + [T] [] [] Keys

2 Press FS Depress [REZERO] key and press


[DIG-SEL] key, [TARE] key twice in
[0 ] + [] [T] [T] sequence. The display toggles between
XXXX.XXX the messages FS and the current value
/ Capacity

3 [] [] FS Using [DIG-SEL] and [DIG-INC] key


Enter the required Capcity / value
0015.000
4 CAL 00 Press * key to accept Capacity entered
Press  Key and to go to next step
XXXXXX
5 CAL SP Press * key to accept the Zero
Press  Key Calibrated value and to go to next
step. Place the entered Capacity
XXXXXX weight on platter.
6 600000 Press * key to accept span
Press  Key Calibration and display will
read span count.
7 Press [TARE] key, to exit from
15.000 Calibration mode and switch for
Press [T] Key Weighing mode.

Note: During calibration process, Press [TARE] key to exit from calibration mode.

Digital Tare Programming:


Sl.No OPERATION SEQUENCE DISPLAY REMARKS
1 Go to weighing mode. 0.000

2 Press [DIG INC] key.


Press [] key Tare The display toggles between the
message Tare and current
value
XXXX.XXX
3 Using [DIG-SEL] and [DIG-INC]
[] [] 1.000 key
Enter the required value

4 Press [TARE] key, to Tare weight


Press [T] -1.000 reading entered

5 Press [Tare] key to exit from the


Press [T] 0.000 mode / switch for
Weighing more

1.CUSTOMER SPECIFICATION:
SPEC MOST SIGNIFICANT DIGIT(MSD) LEAST SIGNIFICANT DIGIT(LSD)
C10 RS232/RS485 DATA BITS & PARITY: RS232/RS485 Baud Rate (Stop bit =1):
1-> 8 Data Bits & NO Parity 1 -> 1200
2 -> 7 Data Bits & NO Parity 2 -> 2400
3 -> 8 Data Bits & ODD Parity 3 -> 4800
4 -> 7 Data Bits & ODD Parity 4 -> 9600
5 -> 8 Data Bits & EVEN Parity 5 -> 19200
6 -> 7 Data Bits & EVEN Parity Note 3
Note 3
C11 RS232/RS485 Weight Data Transfer: RS232 Data Transfer Mode:
1 -> All 1 -> Stream
2 -> Net only 2 -> Manual
3 -> Gross only 3 -> Command (CTRL E)
4 -> Auto Weight Transfer
5 -> Postal(Meghdooth)
6 -> Milkotronics (W1 to W9)
7 -> Manual Serial Print
8 -> Auto Serial Print
C12 RS232/RS485 Weight Header: RS232/RS485 Transfer Stable
1 -> No Weight/Manual Serial Print:
2 -> Numeric 0 -> No
3 -> Alpha 1 -> Yes
C13 RS232 One Touch Tare from PC On RS232/RS485 Data :
Stable(T/t): 0 ->Variable
0 -> No 1 -> Fixed ( 8 characters)
1 -> Yes
C14 RS232 Stream Data Transfer Mode Skip “STX” Character In RS232 Output:
Stable Status Flag: 0 -> No
0 -> No 1 -> Yes
1 -> Yes
C15 Setpoint Output Type: Setpoint Output Logic:
1 > Active Low 1 -> Logic I (Bar Mode)
2 -> Active High 2 -> Logic II (Dot Mode)
C16 Negative Setpoint Output: VFD Standby Mode (In Battery):
0 -> No 0 -> No
1 -> Yes 1 -> 5 Minute
2 -> 10 Minute
C17 No. of CR & LFs after Total Print: Serial Printout Format:
0 -> FORM FEED 0 -> 80 col All Wt.
1 -> 1 CR LF 1 -> 40 col All Wt.
~ 2 -> Net Wt only.
9 -> 9 CR LF
C18 RS485 Weighing Scale ID:
00 -> Reserved
01 – 99
C19 Not used Not used

2.TECHNICAL SPECIFICATION:
SPEC MOST SIGNIFICANT DIGIT(MSD) LEAST SIGNIFICANT DIGIT(LSD)
T10 Dual Accuracy (Upto 50% of Decimal Point Position:
Capacity): 0 -> No Decimal point
0 - No 1 -> 1 DP
1 - Yes 2 -> 2 DP
DISPLAY RESOLUTION SHOULD BE LESS THAN 3 -> 3 DP
OR EQUAL TO 1/15000 & SPEC T11 IS AMONG 4 -> 4 DP
02, 05, 10, 20 AND 50 VALUES for dual range.
T11 Accuracy (2 digits):
T12 Capacity:
T13 Power Fail Return: Tare Limit:
0 -> No 1 -> Capacity
1 -> Yes 2 -> 1/2Capacity – 1
100% TARE LIMIT IS NOT POSSIBLE IF DUAL
RANGE IS WORKING.
T14 Power On Start Range: Forced Zero Tracking:
1 -> 25% of capacity 0 -> ¼ d
2 -> Unlimited 1 -> 1d
2 -> 2 d
2 -> 2d
3 -> 3d
4 -> 4d
5 -> 5d
T15 RS232 Minimum Weight Transfer Serial Interface Option:
Limit: 0 -> No
0 -> No limit 1 -> RS232
1 -> Net >/ 1d 2 -> RS485
2 -> Net >/ 5d Note 3
3 -> Net >/ 20d
T16 S-On Through Key Board: Second Serial Interface Option:
0 -> Yes 0 -> No
1 -> No 1 -> RS232
Note 3
T17 ADC Gain: Unit Conversion:
1 -> UNITY 0 -> No
2 -> Double 1 -> Carat ( 5.0000 )
Note 3 2 -> Tola ( 0.085734)
3 -> Litre (Cow Milk) ( 0.9724 )
4 -> Litre (Soya Oil) (1.0989 )
5 -> Milkotronics Litre constant
Range (0.9000 to 1.0000)
T18 Not used Not used
T19 Stable Indicator On: Display Brightness Level:
0 - No 3 -> 29% Blue LED
1 - Yes 4 -> 71% Red LED
5 -> 77%
6 -> 86% VFD
7 -> 93%
8 -> 100%
Note 3
Note 1 : Default is indicated by Highlighting for applicable specs.
Note 2 : If a Spec is set to a value other than specified, then default function is used.
Note 3 : The Specs will take effect only at power on. So on/off is required whenever changed.