You are on page 1of 6

OX/110 - 5

TECHNICAL DATA
GENERAL INFORMATIONS
Voltage When mounting tube inserts adopt proper caution, in order to avoid glass bulb breaking
and fragments projection. Please use protective gloves and glasses.
nominal 120
kV Tube insert connected to H.V. supply is a radiation source: be sure to take all necessary
maximum for test 135 safety cautions

Inverse voltage
nominal 120  Wash thoroughly with alcohol the external surface of tube insert (care of fire risk).
kV Avoid contact of dirty surfaces with cleaned tube insert.
maximum for test 135
 Clamp system inside housing or self-contained units must not mechanically stress the
Focal spots (IEC 60336:2005) tube.
small 0.5
mm  After installation, check the right working of the tube (no fluctuation of tube current
large 1.5 nor crackling)
Filament characteristics  Comply with insert thermal parameters, planning and programming the exposure
small 2.4 ÷ 4.0 V 3.0 ÷ 4.0 A parameters and cooling pauses. Housing or self-contained units must be provided with
4.0 ÷ 8.0 V 3.5 ÷ 4.5 A an adequate thermic protection.
large
 Voltages indicated in charts are valid for transformer supplied with ground center.
Anode material tungsten
 Tube inserts contain environment polluting materials, particularly lead liner tubes.
Target angle 12 °
Please apply to qualified operator for waste disposal, according to local regulation
Anode heat storage capacity 30000* J requirements.
Maximum anode cooling rate 270* W
Nominal anode input power at 0.1 s (DC)
small 780
W INCIDENT REPORT ACCORDING TO 93/42/EEC MEDICAL DEVICES DIRECTIVE
large 4200
In order to comply with CE marking requirements, end users have to report to local Competent
Inherent filtration 0.5 mm Al Authority all the informations about possible incidents involving the device, regarding any
deterioration in its characteristics and performances, as well as any inaccuracies in this
Maximum diameter 50 mm documentation, which might lead to or might have led to the death of patient / user or a
mm deterioration in his state of health. This information must be promptly reported also to C.E.I. in order
Overall length 155* to start manufacturer reporting, as per above mentioned directive.

* with standard shank. On every kind of shank it is possible to make customized millings for
the allignement of the tube.

this mark assures device conformity to EC Directive 93/42 on Medical Devices Safety
0051
Rev. N – 0816 – Original text: Italian Pag. 1 di 6 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/110 - 5
DIMENSIONS AND CONNECTIONS

STD SHANK
(Anode heat storage capacity 30 kJ, Maximum anode cooling rate 270 W)

large focus - red

common - black

Small focus-yellow

raggi
X

THERMAL CURVES

INSULATING CYLINDER Heat storage (J)

LEAD Time (s)

Rev. N – 0816 – Original text: Italian Pag. 2 di 6 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/110 - 5
DIMENSIONS AND CONNECTIONS

LONG SHANK
(Anode heat storage capacity 33 kJ, Maximum anode cooling rate 280 W)

large focus - red

common - black

Small focus-yellow

raggi
X

THERMAL CURVES

Heat storage (J)


INSULATING CYLINDER
LEAD Time (s)

Rev. N – 0816 – Original text: Italian Pag. 3 di 6 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/110 - 5
DIMENSIONS AND CONNECTIONS

RADIATOR
(Anode heat storage capacity 40 kJ, Maximum anode cooling rate 500 W)

large focus - red

common - black

Small focus-yellow

raggi
X

Heat storage (J)

INSULATING CYLINDER Time (s)


LEAD

Rev. N – 0816 – Original text: Italian Pag. 4 di 6 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/110 - 5
EMISSION CHARACTERISTICS FILAMENT CHARACTERISTICS DC
20 24 4.2
SMALL FOCAL SPOT (0.5 IEC 336)

4.1

16 20
4.0

3.9
16
12
Anode current (mA)

Anode current (mA)

Filament current (A)


3.8
70 kV 70 kV
12 3.7
8
3.6
8
3.5
4
3.4
4
3.3
0 0 3.2
2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 2.4 2.8 3.2 3.6 4.0
Filament voltage (V) Filament voltage (V) Filament voltage (V)

100 120
4.5
LARGE FOCAL SPOT (1.5 IEC 336)

4.4
80 100
4.3

80 4.2
60
Anode current (mA)

Anode current (mA)

Filament current (A)


70 kV 4.1
70 kV
60 4.0
40
3.9
40
3.8
20
3.7
20
3.6
0
0 3.5
4.0 5.0 6.0 7.0 8.0 4.0 5.0 6.0 7.0 8.0
4.0 5.0 6.0 7.0 8.0
Filament voltage (V) Filament voltage (V) Filament voltage (V)

Rev. N – 0816 – Original text: Italian Pag. 5 di 6 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy
OX/110 - 5
MAXIMUM RATING CHARTS DC

SMALL FOCAL SPOT (0.5 IEC 336) LARGE FOCAL SPOT (1.5 IEC 336)

40 kV
40 kV
50 kV 50 kV
60 kV 60 kV
70 kV 70 kV
80 kV 80 kV
90 kV 90 kV
100 kV
100 kV 110 kV
110 kV 120 kV
120 kV

Rev. N – 0816 – Original text: Italian Pag. 6 di 6 Skan-X Radiology Devices S.p.A. – via della Tecnica 3 – San Lazzaro di Savena – BOLOGNA - Italy