Humidity 95 % RH at 25 Degree C

Uncertainty Procidure : -
UNCERTAINITY MEASUREMENT
(The values having ( ) right to them are variables
Range 95.00
TYPE - A LC UUC 0.01
% RH Uncertainity of std. 0.43
1st Reading 95.26 95.00 -0.26
2nd Reading 95.09 95.00 -0.09
3rd Reading 95.13 95.00 -0.13
4th Reading 95.21 95.00 -0.21
5th Reading 95.25 95.00 -0.25

Average -0.19
-0.19
Standard Deviation 0.074966659 %RH

Total readings (n) 5

Under Root (n) 2.236067977

Uncertainity Ua( Under Test ) 0.03353 %RH UUC

TYPE - B

Uncertainity due to master 0.43 %RH Master Indicator


Normal Dis. K=2 2.000
U1= 0.215 %RH

Uncertainity due to Stability 0.09 %RH Stability of Chamber


Chamber 1.732
Rect. Dis. K=1.73205
U2= 0.05 %RH

Uncertainity due to Uniformity 0.0 %RH Uniformity of Chamber


Chamber 1.732
Rect. Dis. K=1.73205
U3= 0.00 %RH

Uncertainity due to 0.12 %RH Drift of master


Drift of master 1.732
Rect. Dis. K=1.73205
U4= 0.07 %RH

Uncertainty due to L.C. of UUC 0.01 %RH Least Count


Rect. Dis.K=1.73205 1.732
U5= 0.0029 %RH

Combin Uncertainity 0.234 %RH


Degrees of freedom(v) 9426.8
Coverage Factor 2
Expanded Uncertainity 0.47 %RH
Final Uncertinity (+/-) 0.47 %RH