You are on page 1of 1

Furnace Single point

Uncertainty Procedure
UNCERTAINITY MEASUREMENT Range of UUC/set point 1200
(The values having ( ) right to them are variables Least count of UUC 1.00
TYPE - A Uncertainity of STD. 1.45
0
C Uncertainity of indicator 0.00
1st Reading 1201.72 1200 -1.72
2nd Reading 1201.60 1200 -1.60
3rd Reading 1201.68 1200 -1.68
4th Reading 1201.63 1200 -1.63
5th Reading 1201.72 1200 -1.72

Average -1.67
Standard Deviation 0.053851648
Total readings (n) 5
Under Root (n) 2.236067977
0
Uncertainity Ua( Under Test ) 0.02408 C
TYPE - B
0
Uncertainity due to master 1.45 C
Normal Distr. K=2 2 R-type
0
U1= 0.725 C
0
Uncertainity due to master 0.00 C Master Indicator
Normal Distr. K=2 2
0
U2= 0.0000 C
0
Uncertainty due to 0.1 C Dry Block-Stability
Stability of Dry Block 1.73205
Rect.Dist. K=1.73205
0
Uncertainty due to 0.0 C Uniformity
Stability of Dry Block 1.73205
Rect.Dist. K=1.73205
0
U3= 0.034641 C
0
Uncertainty due to 0.04 C Drift of Master
Drift in Master 1.73205
Rect.Dist. K=1.73205
0
U4= 0.023094 C
0
Uncertainty due to 0.00 C Uncertainity of indicator
Indicator 1.73205
Rect.Dist. K=1.73205
0
U5= 0.000000 C
0
Uncertainty due to LC of uuc 1.00 C Least Count of Indicator
Rect.Dist. K=1.73205 1.73205*2
0
U6= 0.28868 C
0
Uncertainty due to Variation Error 0.00 C Variation error in the Ice
Rect.Dist. K=1.73205 1.73205 Point
0
U7= 0.00000 C
0
Uncertainty due to In homogeneity 0.000 C In Homogeneity of
Rect.Dist. K=1.73205 1.73205 Thermocouple
0
U8= 0.00000 C

0
Combin Uncertainity Uc 0.78 C

Effective Degrees of freedom(v) 4460493.8740

Coverage Factor 2

0
Expanded Uncertainity (+) 1.57 C

0
Final Uncertinity (+/-) 1.57 C