Professional Documents
Culture Documents
Vicencio, RMT
DMMC Institute of Health Sciences
▪ is an instrument used to see objects that are too small to be seen by
the naked eye.
1. Lens system
- Primary components of the lens system are the oculars, the
objectives, and the coarse- and fine adjustment knobs
2. Illumination system
- contains the light source, condenser, and field and iris
diaphragms.
Objects to be examined are placed on a platform, referred to as the
mechanical stage.
1. Light Microscopy
1. Brightfield
2. Darkfield
3. Phase Contrast
4. Polarizing
5. DIC
6. Fluorescence microscope
7. Confocal laser scanning
2. Electron Microscopy
1. Scanning
2. Transmission
▪ Microscopes used in clinical practice are light microscopes.
▪ consists of two lens systems (combination of lenses) to magnify the image.
▪ A compound light microscope with a single eye-piece is called monocular; one with
two eye-pieces is said to be binocular.
▪ A compound light microscope is the most common microscope used in microbiology
▪ The specimen is placed between two polarizers crossed at 90 o to each other (one in
condenser and one in objective).
▪ Only light from the plane of focus reaches the detector. The scanned image (an optical
section) is digitally recorded.
▪ Images from consecutive focal planes can be recorded, and composite or 3-D images
can be digitally created.
7. Confocal
laser
scanning
- krypton/argon laser
- high resolution,
sharp image
- high sensitivity
- Can be used in
reflectance or
fluorescence mode
- eliminates
background
interference
7. Confocal laser scanning
2. Electron Microscopy
▪ The electron beam is absorbed or deflected by the heavy metal stains and shadows
are cast onto film or a phosphorescent plate (image is a shadow) at the bottom of the
column.
▪ 2-D image
▪ reveals internal cell structure
▪ high resolution, high magnification
▪ electron beam is focused by magnetic field
2. Transmission Electron
Microscopes
Focusing Specimen
1. Always start with the SCANNING OBJECTIVE.
2. Focus using the COARSE ADJUSTMENT KNOB to bring the
object into focus.
3. Use only the FINE ADJUSTMENT KNOB when using the OIO.
4. If the specimen is too light or too dark, try adjusting the iris
diaphragm and the light intensity (illuminator) knob.
5. Keep both eyes open to reduce eye strain