You are on page 1of 8

1074 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 OCTOBER 2005 www.chromatographyonline.

com

Multidimensional and
Comprehensive Liquid
Chromatography
Column
Watch
M ultidimensional-chromatogra-
phy has been practiced for
many years, both in off-line
and on-line modes. Basically, the experi-
ured cleanly provided no other components
in the fraction happen to elute at the same
time on column 2. A simplified schematic
of a coupling experiment, a size-exclusion
In this month’s “Column ment involves transferring a fraction or column with an affinity separation column,
fractions from one chromatographic is depicted in Figure 1.
Watch,” Ron Majors
medium (usually a column) to a secondary When the demand requires the complete
discusses advances in (or additional) chromatographic medium characterization of complex mixtures, then
multidimensional (column or columns) for further separation. the term comprehensive chromatography is
chromatography and The technique can be used for further reso- used. The late Prof. Calvin Giddings of the
lution of complex mixtures that cannot be University of Utah, Salt Lake City, coined
comprehensive
separated entirely on a single medium, for the term comprehensive multidimensional
multidimensional sample cleanup by removing matrix or separation (10). Unlike multidimensional
chromatography. These interfering compounds, for increased sam- chromatography, in comprehensive chro-
techniques are well suited ple throughput, and for trace enrichment of matographic systems, every fraction from
minor compounds of interest. The most the primary column is subjected to the sec-
for the separation of
popular version of multidimensional chro- ond dimension. Normally, fractions are
complex samples. matography is two-dimensional (2-D) diverted from the primary column (10) to
Although GC  GC is fairly chromatography. Table I provides examples the secondary column (20) at defined time
well developed, LC  LC is of popular combined modes that have been intervals.
used. In a recent special issue of LCGC Proteomics is an area in which extremely
in its infancy. But with
Europe, several of these approaches were complex samples are encountered. All pro-
fast LC columns, based on reviewed in detail (8,9). In the column teins within a cell may be in need of char-
monoliths, superficially mode, 2-D chromatography also has been acterization to determine how they are
porous particles, and short referred to as column switching, multiphase affected in disease states. Because there are
chromatography, coupled column chro- potentially tens of thousands of proteins or
sub-2-m particle matography, boxcar chromatography, and more, a 1-D separation is inadequate to
columns, it has a bright sequential analysis. characterize this complex sample. However,
future, especially when In conventional 2-D chromatography, if the proteins (or tryptic peptides) are bro-
coupled with MS usually one or only a small number of com- ken into smaller fractions, which in turn
pounds are of interest. One example would are broken into even smaller fractions by a
techniques. Some of the be the high performance liquid chromatog- second (or sometimes even a third) separa-
latest information from raphy (HPLC) separation and determina- tion dimension, and if they are coupled to
HPLC 2005 provides tion of a drug in a biological fluid such as a powerful detection measurement tech-
examples of the potential urine. The matrix (for example, proteins, nique such as mass spectrometry (MS)
uric acids, and so forth) is of no interest yet (which is really an added dimension), then
of LC  LC. may interfere with the analysis and identifi- it is conceivable that individual proteins (or
cation of the drug compound in a one- peptides) could be identified at these trace
dimensional (1-D) experiment. Other drug levels. Then, perhaps, the biochemists can
metabolites and small molecular weight measure those individual proteins that can
compounds might also be of little or no lead to new drugs or other treatments that
interest. By directing a fraction containing could block or modify their behavior.
the drug peak plus any possible overlapping Comprehensive 2-D gas chromatography
Ronald E. Majors contaminants from the primary column to (GC) (abbreviated GC  GC) is a mature
Column Watch Editor a secondary column, the drug can be meas- technique with commercial products
www.chromatographyonline.com OCTOBER 2005 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 1075

plex the samples that can be resolved. multidimensional GC, a cold trap might be
Mixture
When a sample is separated using two dis- used; in HPLC, a trapping column could
similar columns, the maximum peak capac- be employed to minimize this band disper-
ity max will be the product of the individ- sion. Another approach might be to use a
ual column’s peak capacity n (as depicted narrow bore or capillary column for the
in Equation 1). first dimension and a wider bore column
Size separation
for the second dimension.
max  1  2 [1] Obviously, if the speed of the first
dimension is faster than the speed of the
For example, if each separation mode second dimension, something must be
generates peak capacities of 100 and 200, done to halt or slow up the primary separa-
Affinity separation respectively, the theoretical peak capacity of tion that will affect the overall time. An
the 2-D experiment will be 20,000, a huge alternative approach could be to speed up
gain in separation space. To achieve this the separation in the second dimension by
gain, however, the two techniques should using a shorter column, faster gradient
be totally orthogonal, that is, based upon (LC), faster temperature program (GC),
Figure 1: Simplified schematic of 2-D multidi- completely different principles. For exam- and higher flow rate. Schoenmakers (13)
mensional chromatography LC separation. The ple, in HPLC, if the primary separation pointed out how Poppe plots (14) can be
first and second dimension separation modes
were size-exclusion and affinity chromatogra- column were based upon chirality while the used to predict the optimum resolution at
phy, respectively. second separation column separated on the reasonable separation times using currently
basis of hydrophobicity, the individual peak available pressures. Combinations of iso-
already on the market. The technique has capacity of the combined system would be cratic and gradient elution analyses for the
tremendous separation power, uses simple their multiplicand. two dimensions might also help to alleviate
robust hardware, and has similar analysis There are many practicalities to be real- the speed dilemma.
times to temperature-programmed high-res- ized when coupling two different chro- The compatibility issue is important for
olution capillary chromatography. On the matographic techniques. In a plenary lec- the practical coupling of the techniques
other hand, comprehensive 2-D LC (LC  ture during the opening session at HPLC outlined in Table I. An example of the
LC) is still in its infancy and is more com- 2005, Peter Schoenmakers of the University incompatibility of the two orthogonal tech-
plex to perform. However, it is driven by of Amsterdam, The Netherlands, a member niques in HPLC would be the application
user needs, especially in proteomics, and of the Permanent Scientific Committee,dis- of normal-phase chromatography using
the topic is getting more attention. cussed the state-of-the-art in multidimen- nonpolar solvent such as hexane as the first
The purpose of this month’s installment sional chromatography and LC  LC com- dimension. Solvent compatibility difficul-
of “Column Watch” is to discuss multidi- prehensive 2-D separations (13). He ties will be encountered if the second
mensional chromatography and compre- detailed some of the factors that govern dimension is reversed-phase chromatogra-
hensive 2-D chromatography. I will briefly multidimensional techniques. The most phy which uses a polar solvent such as
discuss GC  GC, but because much has important factors are: water. In GC, if a packed column is used
been written on this approach (3,11), I will ● peak dispersion for the first dimension and a capillary col-
focus mainly on LC  LC. I will confine ● relative speeds, and umn is used for the second dimension, the
my discussion to the use of 2-D techniques ● mode compatibility. sample capacities of the two columns may
in GC and HPLC. It is obvious that one Let’s examine these factors. As the sample not be mutually compatible. In coupling
can extend the experiment to “n” dimen- components move down the primary sepa- LC with GC, the HPLC mobile phase sol-
sions, but for the most part, many of them ration column, they become diluted, influ- vent can be detrimental to the GC station-
are difficult to put into practice. I will also encing the injection dispersion of the sec- ary phase, especially if a large volume is
summarize several papers presented on ond dimension. Ideally, their bandwidth injected.
multidimensional chromatography and should be very small so that they start out
comprehensive multidimensional HPLC as narrow band on column two. If not, the Off-Line Multidimensional
from the recent HPLC 2005 Symposium in sensitivity and resolution may be compro- Chromatography
Stockholm, summarized in last month’s mised. In many cases, an intermediate Off-line 2-D techniques are performed
“Column Watch” (12). “refocusing” step might be required. In every day in a normal laboratory operation.
Table I: 2-D chromatographic techniques
Background in Multidimensional
Chromatographic Separations Primary Technique Secondary Technique Off-Line On-Line Reference
By the combination of multiple chromato-
TLC (or HPTLC) TLC (or HPTLC)   1
graphic steps, there is one great advantage: Preparative TLC GC or LC  X 1
an increase in peak capacity. Peak capacity GC GC   2,3
is, simply stated, the maximum number of LC LC   4
peaks that can be resolved in a given time- LC GC   5
frame. The more peaks that a combination SPE or SBSE LC or GC   6
SFC LC   7
of techniques can handle, the more com-
1076 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 OCTOBER 2005 www.chromatographyonline.com

Something as simple as a solid-phase


extraction (SPE) sample cleanup using a
packed cartridge followed by a GC separa- Valve
tion is really an off-line 2-D experiment. injector
Pump Gradient
Because of the ease of collecting and han- HPEC LC
for
dling liquids, off-line LC–LC and LC–GC HPEC columns Six-port
techniques are popular. Although less con- sampling valve
venient, off-line 2-D GC can be performed Detector(s) Detector(s) Waste
by cold trapping the effluent gas stream
Analytical
from a nondestructive detector such as column
Waste (reversed phase)
thermal conductivity at the appropriate Recorder Recorder
retention time for the analyte of interest.
An alternative method would be to cold
trap the analyte from a GC instrument Figure 2: Schematic of an on-line multidimensional LC–LC system. The first and second dimen-
sion columns were high performance size exclusion and reversed-phase chromatography columns,
equipped with a postcolumn, predestructive respectively.
detector (e.g., flame ionization detector)
splitter. Although conceivably comprehen-
sive LC  LC and GC  GC could be technique has many variations, including: achieve miscibility (for example, addition of
performed off-line, the handling, concen- heartcutting to improve separation in a third solvent, miscible with solvent 1 and
tration, and transfer of multiple fractions selected regions; backflushing to remove solvent 2).
for subsequent reinjection could pose some unwanted higher mass components; and The practical application of on-line
practical problems. Table II lists the advan- packed-to-capillary switching to maximize LC–LC dates back to the early days of
tages and disadvantages of off-line LC solute sensitivity. For those interested in the HPLC (18–20) when reliable high-pressure
techniques. history and varied applications of GC–GC multiport valves first became available. In
as well as GC  GC, an excellent two-part LC–LC, six- and 10-port high-pressure
On-Line Multidimensional review on these subjects was published by valves with one or two sample loops are
Chromatography Bertsch (16,17). used to interface the two separate chro-
From a convenience and automation view- On-line Multidimensional LC (LC–LC): In matographic systems. Depending upon the
point, the on-line coupling of two (or HPLC, the great potential of the on-line requirements of the experiment, a number
more) chromatographic techniques is pre- technique lies not only in the selection of of valving configurations can be plumbed
ferred. Table III provides some advantages different modes for the first and second to switch flows to employ various tech-
and disadvantages of on-line multidimen- dimensions but also in the ability to use niques such as heartcutting or backflushing.
sional chromatography. different mobile phases–additives to pro- The technique used most often in multidi-
On-line 2-D GC (designated GC–GC): The vide the needed separation selectivity and mensional LC is heartcutting, in which all
coupling of GC columns via multiport adjustment of capacity factors. Even more or a portion of the analyte of interest plus
switching valves has been used for many important than off-line LC–LC, a prime coeluted compounds from the primary col-
years. However, the metal valves tended to requisite is that the mobile phases must umn are selectively diverted to the second-
have catalytic activity, unswept volumes, have some degree of compatibility. Table IV ary column. For example, Figure 2 illus-
and cold-trapping effects and thus gave suggests the different LC modes that can be trates an multidimensional chromatography
lower efficiency and undesirable band conveniently coupled based upon the most system used to couple a high-pressure size-
broadening (2). Valveless Dean’s switching commonly used mobile phases. Obviously, exclusion chromatography (SEC) column
was introduced (15), but the technique was normal-phase techniques using nonpolar with a reversed-phase chromatography col-
never well accepted until 30 years later with solvents such as hexane or diethyl ether are umn for heartcutting. In this approach, the
the advent of modern electronic pneumatic difficult to combine with modes using pre- effluent from the primary SEC column is
control that can be integrated as part of the dominantly aqueous mobile phases. Thus, passed continuously through one sample
GC method. Now GC–GC is as conven- Table IV serves only as a guideline because loop of a 10-port valve to waste. At the
ient and routine as single-column GC. The conditions could always be developed to appropriate time, when the desired compo-
nent(s) of the sample enters the loop, its
contents are injected onto the second col-
Table II: Off-line 2-D LC techniques umn. The volume injected is governed by
Advantages Disadvantages the size of the sample loop volume that is
fixed. Another plumbing configuration
Easy to collect column effluent Difficult to automate; more cumbersome and
could allow the effluent from the first
inconvenient
Can concentrate trace solutes Greater chance of sample loss dimension to be directed to waste while the
from large volume (e.g. evaporation, adsorption, oxidation) second dimension experiment is taking
Can work with two LC modes with More time consuming place. In the configuration shown in Figure
incompatible solvents (e.g. adsorption More difficult to quantitate and reproduce 2, two chromatographs were employed: a
using hexane → reversed-phase
single-pump isocratic system for the SEC
chromatography using water) by
evaporation and reconstitution column and a dual-pump gradient system
1078 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 OCTOBER 2005 www.chromatographyonline.com

area of interest in the initial separation separation of fractions on a capillary or


could be handled, allowing good selectivity. nanocolumn eluted by a continuous trifluo-
(a) (b) The disadvantage is that, because only a roacetic acid-acetonitrile gradient. MS was
BioTrap C8 portion of the sample is introduced into the used for detection and identification of sep-
secondary column, sensitivity is reduced. arated peptides. One example of this
Alternatively, the effluent from the SEC approach in a off-line mode was demon-
SPS C8 column could be directed to the secondary strated in the presentation of Jenny Sam-
4 mAU reversed-phase column and all of the initial skog and colleagues (GE Healthcare, Upp-
ISRP C8 peak captured but at the expense of sala, Sweden) (21). She studied
Switching time decreased sample dispersion on the phosphopeptides in brain tissue that are
LiChrospher
1 mV
RP-4 ADS reversed-phase column — unless some refo- present at the femtomole level. A microbore
0 5 10 15
cusing is done. If the initial column was an strong cation exchange column with a 2.1-
0 5 10 15 20 25
Time (min) Time (min) SEC column using tetrahydrofuran as the mm inner diameter with a salt gradient was
mobile phase, a relatively strong solvent in used for the first dimension. Fractions were
Figure 3: Use of a coupled-column reversed-phase chromatography, then it collected off-line and reinjected into a trap
RAM–reversed-phase chromatography system might be difficult to divert a large volume column where they were desalted. Finally,
for the analysis of drugs in plasma (23). (a) Elu-
tion profile of a 100-L plasma sample. First onto the reversed-phase column. A large nanoliquid chromatography using a 75-m
dimension column: LiChrospher RP-4 alkyl diol injection volume of such a strong solvent i.d. reversed-phase LC column was used for
silica ADS RAM (Merck, Darmstadt, Germany) would cause partial migration of injected the final separation step. To improve
using water as an eluent at 0.5 mL/min. Second fractions down the reversed-phase column throughput, a dual column setup was used
dimension column: 250 mm  4.6 mm LiChros-
pher 60 RP Select B. (b) Separation of epirubicin causing band dispersion thereby limited res- so that while one column set was perform-
and its metabolites after prefractionation on olution. On the other hand, if the SEC col- ing the analysis, the second column set
four RAM materials. Mobile phase: 70:30 (v/v) umn was used with an aqueous mobile could be regenerated. Using MS3, phospho-
acetonitrile (0.1% trichloroacetic acid)–tri- phase, a large volume could be diverted to peptides were identified, triggered by their
ethanolamine (pH 2); flow rate: 1.0 mL/min;
detection: fluorescence (excitation wavelength: the reversed-phase column and the ana- neutral loss of phosphoric acid.
445 nm, emission wavelength: 560 nm). Peaks: lyte(s) refocused easily because water is a In another HPLC 2005 presentation,
1  epirubicinol, 2  epirubicinal aglycone, 3  weak solvent in this mode. In this case, David Lubman and colleagues from Uni-
epirubicin, 4  epirubicin aglycone, 5  7- something as simple as a three-way diverter versity of Michigan, Ann Arbor, Michigan
deoxyepirubicinal aglycone. Compounds were
in the 5.6–8.2 ng/mL concentration range. valve could be used as the interface between (22) used a different and novel 2-D liquid
the two chromatographs. mass mapping technique to find protein
biomarkers in tumor cells from ovarian
Selected HPLC 2005 Applications cancer patients. Their group has developed
for the reversed-phase column. This setup of LC–LC a 2-D liquid separation of intact proteins
allowed the flow on the primary column to One of the main applications that stood based upon pI using chromatofocusing in
be stopped and additional components on out in HPLC 2005 was the increased use of the first dimension and by hydrophobicity
the SEC column handled at a later time. multidimensional chromatography, not using nonporous reversed-phase HPLC in
Because diffusion coefficients in HPLC only for proteomics research but for other the second dimension. The eluent of the
mobile phases are rather small, the band applications as well. For proteomics, a typi- HPLC can be directed into an electrospray-
dispersion resulting from stopping the flow cal approach reported was to use a narrow- time-of-flight (TOF) MS for analysis of the
for a short period of time is rather minimal. bore cation exchange column (first dimen- intact molecular weight of the proteins to
Once the systems are isolated, the second sion) to concentrate and perform a rough produce a mass map of the protein content
chromatograph then could perform the separation of peptides from a tryptic digest. of the cell. They were able to use this
reversed-phase separation. The advantage of This step is followed by a desalting step method to study over 25 tumor samples
the heartcutting technique is that only the and/or a reversed-phase (second dimension) with high reproducibility in analysis based
upon pI and MW value to allow compar-
isons between samples for biomarker dis-
Table III: On-line MDC
covery. This method can be used to profile
Advantages Disadvantages tumor protein expression and search for
Easy to automate, especially Requires more complex hydraulics or pneumatics, markers specific to subtypes of cancer.
with modern chromatographs switching valves, more expense One ever-popular application of column
Less chance of sample loss because Difficult to handle trace compounds because very switching is the determination of drugs in
experiment performed in dilute and can be in large volume; can compensate biological fluids. If a biological fluid such as
closed system by trapping, trace enrichment or cyro-focusing
blood plasma is directly injected onto a
Can configure switching system Solvents from first and second dimensions must
which best suits needs (e.g. backflush, be compatible both from miscibility and strength reversed-phase column, hydrophobic pro-
heartcutting, on-column requirements (LC); LC solvents can affect GC teins and other plasma components can
concentration) stationary phases adsorb irreversibly on the column, chang-
More reproducible ing its characteristics and eventually causing
Can increase sample throughput;
a rise in pressure. By choice of a restricted
decreased total analysis time
access medium (RAM), columns in the first
1080 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 OCTOBER 2005 www.chromatographyonline.com

dimension, and a C18 reversed-phase col-


umn in the second dimension, plasma can
be injected directly, without incident (23). (a)

RAM packings are designed to exclude pro- 350

teins and other biomolecules yet allow 300

Absorbance (mAU)
small molecules such as drugs to diffuse 250

into the pores and be retained by 200

hydrophobic interactions. The plasma and 150

other unwanted excipients can be diverted 100

via a switching valves to waste before the 50

drugs are eluted from the RAM column. 0

The analytes, epirubicin and its metabo- 10 20 30 40 50


lites, were diverted to the reversed-phase
column and separated using isocratic elu- 100

tion. Figure 3a shows the elution profile of 80

Absorbance (mAU)
a plasma sample from a LiChrospher RP-4 60

alkyl diol silica (ADS) RAM column 40

(Merck, Darmstadt, Germany) using water 20

as an eluent. Within 5 min, the entire pro- 0

tein matrix was eluted to waste but to play -20

it safe, a time period of 7 min was used -40

before a switching valve was diverted to the 0.2 0.4 0.6 0.8
second dimension column, a LiChrospher 40
Time (min)

60 RP Select B (250 mm  4.6 mm). Fig-


(b)

ure 3b shows the separation of epirubicin


Second dimension (s)

and its metabolites after prefractionation on 30


four different RAM materials. Each RAM
gave slightly different recoveries of the ana-
lytes and the ADS column gave the best 20

overall performance.
Other multidimensional chromatography
applications at HPLC 2005 were reported. 10
15 20 25 30 35 40 45 50
Egidijus Machtejevas and colleagues, First dimension (min)
Johannes Gutenberg-University, Mainz,
Germany; Merck, Darmstadt; and Astra- Figure 4: (a) Comprehensive 2-D-LC Mode: normal-phase LC–reversed-phase LC-UV-MS. (Top)
First dimension normal-phase LC. Column: 250 mm  1 mm, 5-m dp Betasil diol (Thermo Electron
Zeneca, Molndal, Sweden (24) developed a
Corporation, Waltham, Massachusetts; mobile phase: 85:15 hexane–1-butanol  0.2%
novel sample cleanup column to study ethanolamine, isocratic; fl gradient: 25% B for 3 s, to 50% B in 3 s, to 100% B in 15 s, 100% B for
endogenous peptides in biofluids by direct 9 s, back to 25% B in 3 s; flow rate: 5 mL/min (170 bar); 30 °C; sample: pharmaceutical compounds
injection. The packing consisted of a RAM B-mix; response time: 0.1 s. (b) Comprehensive 2-D-LC-UV chromatogram.
with strong cation-exchange functionality.
By adjusting the pore size of the packing,
size exclusion could be used to determine compounds. Using bisphenol A as a model system with partially correlated selectivities,
the molecular weight fractionation range compound, trace levels were determined in allowing to predict the range of the struc-
and the selective ion exchange chromatog- serum combining column switching and tural units in a 2-D sample (according to
raphy could be performed within the pores. LC–MS. the Giddings’ notation) that can be
For added selectivity, Jan Haginaka and H. From a more theoretical side, Pavel Jan- resolved in a given time range for the two
Sambe of the Mukogawa Women’s Univer- dera and colleagues from the University of dimensions for not fully orthogonal sys-
sity, Nishinimiya, Japan (25) developed a Pardubice, Czech Republic (26) studied tems. The group also investigated the possi-
combined RAM and molecular imprinted structural effects on the separation selectiv- bility of phase transfer between a reversed-
polymer (MIP). MIPs are used for the very ity and performance optimization in 1-D phase system (1-D) and a normal-phase
selective extraction of target compounds, and 2-D LC systems. The goal of the work system (2-D). A hydrophilic interaction LC
while RAMs have been used for the direct was to characterize the degree of orthogo- (HILIC) system was proposed for the sec-
injection assays of drugs and metabolites in nality between the first and the second ond dimension in a reversed phase–normal
biological fluids. Although MIP template dimension systems in 2-D LC using the phase LC  LC system. The HILIC system
leakage is still a problem, by the use of an correlation characterizing the similarity of was more resistant to aqueous solvents
isotopologue form of the template molecule selectivity in the two separation systems, transferred from the first dimension than
and MS detection, the template leakage will allowing more efficient use of the peak classical NP systems with nonaqueous
not interfere because MS can discriminate capacity theoretically available in 2-D mobile phases. This system was used for
between the labeled and unlabeled target HPLC. A model was derived for the 2-D comprehensive 2-D separation of ethylene
www.chromatographyonline.com OCTOBER 2005 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 1081

oxide–propylene oxide (co)oligomers. Table IV: HPLC modes easily coupled


Comprehensive GC (GC  GC): GC 
10 Mode
GC is the most developed of the compre-
hensive chromatographic techniques by far. 20 Mode LSC NPC RPC IEX GPC GFC
Not only has the technique been widely LSC ++ ++ ++
applied to complex mixtures (3,5,16,17) NPC ++ ++ ++
but commercial instrumentation is avail- RPC ++ ++ + ++
IEX ++ ++ ++
able. The key to successful application of
GPC ++ ++ + ++
GC  GC is the ability to trap or ther- GFC ++ ++ ++
mally modulate at the juncture of columns
LSC  liquid–solid (adsorption) chromatography (silica, alumina, etc.)
1 and 2 (11), as this defines the amount of NPC  normal-phase chromatography (cyano, amino, diol, etc.)
time provided for the second separation if RPC  reversed-phase chromatography (C8, C18, phenyl, etc.)
“wrap around” peaks are to be avoided. Ide- IEX  ion exchange chromatography (cation, anion)
ally, the first dimension should be relatively GPC  gel permeation chromatography (size exclusion chromatography with nonaqueous mobile phases)
GFC  gel filtration chromatography (size exclusion chromatography with aqueous mobile phases)
slow and the second dimension much
faster. In theory, one could stop the flow of
the first column but since diffusion coeffi- to interpret. that have been separated and identified in a
cients of volatile solutes in a typical GC Comprehensive LC (LC  LC): As typical experiment is a fraction of the tens
carrier gas are quite high, some band Schoenmakers pointed out (13), LC  LC of thousands that are present in a typical
spreading would be expected. However, the has a way to go before it reaches the level of cell extract. Its established competitor, 2-D
ability of modern instrumentation to rap- GC  GC. Currently, peak capacity for a gel electrophoresis, also requires many
idly cool (via cryojets) trapping columns typical LC  LC experiment rarely hours to accomplish similar tasks, is very
and to rapidly heat these same columns has approaches 1000, even with separation labor intensive, lacks reproducibility, and is
made GC  GC practical to perform rou- times in the 1-h timeframe. Typical pro- difficult to quantitate over a wide dynamic
tinely. Throughput is an order of magni- teomics LC  LC  MS experiments still range. So, there is a bright future for LC 
tude better than 1-D GC and modern data require many hours whether performed off- LC. With the advent of monolithic
systems provide 2-D outputs that are easy line or on-line. The number of proteins columns (27,) in conventional and capil-

Circle 17
1082 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 OCTOBER 2005 www.chromatographyonline.com

lary–nanocolumn–chip formats, superfi- the use of LC  LC as well as the combi- dimension but with a different composi-
cially porous packings (28), and short, fast nation of supercritical fluid chromatogra- tion. A modulating device was constructed
sub-2-m columns, the ultrahigh-speed phy and HPLC (SFC  LC) as applied to from a 10-port PC-controlled valve and
separations of small and large molecules, complex mixtures of natural products. In two 50-L sampling loops. The modula-
especially in the second dimension, should the latter technique, the use of supercritical tion time was only 1.0 min. Because of the
improve the situation. carbon dioxide as the mobile phase for the low flow rate in the first dimension and the
first dimension offers the advantage of easy rapid flow rate in the second dimension, it
Selected HPLC 2005 Applications transfer and excellent compatibility since was possible to transfer every fraction to
of LC  LC the carbon dioxide can be evaporated dur- completely characterize the organic acid
A number of papers presented at HPLC ing the process. They also combined nor- content of the aerosol. The use of UV and
2005 showed that LC  LC could be put mal phase chromatography (1o) with time-of-flight TOFMS aided in the identi-
to practical use. reversed-phase chromatography (2o) and fication of the separated acids.
In an oral lecture at HPLC 2005, Tanaka overcame the solvent compatibility problem
and colleagues of the Kyoto Institute of by using a 1-mm i.d. normal phase column Conclusion
Technology, Kyoto, Japan reported on their operating at 50-L/min coupled to a 4.6- Multidimensional chromatography has long
continued work with silica monoliths (29). mm reversed-phase column operated at a been used for resolving selective compo-
Their emphasis has now turned to compre- high flow rate to give a 1-min analysis. The nents from poorly resolved samples in both
hensive 2-D applications of monoliths,in small amount of nonpolar solvent from col- off-line and on-line applications. The
which columns of different dimensions are umn 1 could be accommodated by the advent of modern low-volume, high-pres-
coupled via a switching valve. Using 2-D larger volume of column 2 without detri- sure multiport switching valves conve-
contour plots, they looked at the types of mental effects. niently packaged that can be precisely con-
monolithic column and mobile phase con- Y. Zhao and colleagues (30) presented a trolled and programmed by computerized
ditions (isocratic and gradient elution and pharmaceutical application of LC  LC instrumentation, on-line multidimensional
high flow rates) to maximize the time reso- combined with UV and MS detection. chromatography should receive more atten-
lution of multicomponent mixtures. Obvi- They assembled a fully automated multidi- tion as a way to cleanup difficult samples
ously, when both monolith columns were mensional LC system that could be used and improve analytical results. Comprehen-
of the same types such as C18, there were for LC–LC (heartcutting) and LC  LC. sive multidimensional chromatography
many blank spaces in the contour plots, This initial system was used to ensure that should see even more widespread applica-
meaning that the resolution per unit time there were no coeluted compounds in their tion as more difficult samples that must be
was not optimal. By using different mobile formulations and to characterize impurities. fully characterized are encountered. GC 
phase compositions, water-tetrahydrofuran Coupling normal phase chromatography GC has already proved itself in handling
in the first dimension and water–methanol (diol) as the first dimension using 1-mm difficult volatile samples such as hydrocar-
in the second dimension, they achieved a i.d. columns with a 4.6-mm i.d. C18 bons in petrochemical samples (32). In
semi-orthogonal separation that gave an monolith column as the second dimension, HPLC, with full characterization, require-
increased peak capacity. But, when they they were able to analyze a fraction every ment could dictate the use of longer
switched to a truly orthogonal set of 60 s. Figure 4a shows the separations columns with small particles to give
columns using a strong cation exchange achieved on the two columns for a propri- enhanced chromatographic resolution but
microparticulate column (1-D) coupled to etary drug mixture. Note the good orthog- at the expense of increased pressure. How-
a reversed-phase monolith (2-D), the peak onal separations achieved on the two differ- ever, because the increase in peak capacity
capacity was increased dramatically. By the ent systems. Selectivity was quite different will increase only with N1/2 with the longer
use of these orthogonal columns combined on the two columns. Finally, Figure 4b pro- column, such a system will have a difficult
with a gradient run in the first dimension vides a colored contour plot resembling a time competing with the significant
and isocratic elution in the second dimen- typical 2-D profile resulting from the LC increase of peak capacity with LC  LC
sion, they achieved peak capacity of 700.  LC  UV coupling experiment. performed at more moderate pressures.
Tanaka then spelled out a set of ideal con- An environmental application of LC 
ditions for the second dimension column. LC was reported by Riekkola and col- References
One of the problems with conventional leagues from the University of Helsinki, (1) S. Nyiredy, Recent Applications in Multidimen-
silica monolith columns (4.6-mm i.d.) is Finland (31). Aerosol particulates contain- sional Chromatography, December 2003, 52–59.
the high solvent usage. These columns are ing carboxylic acids are typically very com- (2) K. MacNamara, R. Leardi, and A. Hoffmann,
often used at 4–5 mL/min to achieve high- plex, and it is not possible to separate all of LCGC Eur. 16(12), 14–22.
speed separations. Reduction of the column the analytes by traditional chromatographic (3) P. J. Marriott, P.D. Morrison, R.A. Shellie, M.S.
dimensions to capillary sizes (0.25-mm i.d. methods. To solve this problem, these Dunn, E. Sari, and D. Ryan, Recent Applica-
or smaller) helps to cut down on solvent researchers were able to combine cation- tions in Multidimensional Chromatography,
use but still provides sufficient capacity for exchange chromatography (10) using a nar- LCGC Eur., December 2003, 23–31.
the sample sizes encountered. The applica- row-bore column (150 mm  1.0 mm, 50- (4) E. Hogendoorn, P. van Zoonen, and F. Hernán-
tion of capillary polystyrene–divinylbenzene L/min flow rate) with a silica-based dez, Recent Applications in Multidimensional
monoliths in the second dimension was monolithic reversed-phase column (100 Chromatography, LCGC Eur., December 2003,
also a popular topic at HPLC 2005. mm  4.6 mm, 3.5-mL/min flow rate) 44–51.
Sandra and colleagues (7) reported on (20) using the same mobile phase for each (5) P. Dugo, G. Dugo, and L.Mondello, On-line
1084 LCGC NORTH AMERICA VOLUME 23 NUMBER 10 OCTOBER 2005 www.chromatographyonline.com

Coupled LC–GC: Theory and Applications, LCGC M. Oreans, J. Chromatogr. 83, 267 (1973) K. Hosoya, K. Cabrera, and N. Tanaka, HPLC
Eur., December 2004, 35–43. (20) E.L. Johnson, R. Gloor, and R.E. Majors, J. 2005, Stockholm, Sweden, Lecture TuK4, June
(6) B. Tienpont, F. David, A.Stopforth, and P. San- Chromatogr. 149, 571–585 (1978). 27, 2005.
dra, Recent Applications in Multidimensional (21) J. Samskog, H. Wadensten, and J. Flensburg, (30) Y. Zhao, A. De Villers, P. Sandra, T. Baumgart-
Chromatography, LCGC Eur., December 2004, HPLC 2005, Stockholm, Sweden, paper TuL7: ner, B. Zhang, and J. Kofman, HPLC 2005,
5–13. 1, June 27, 2005. Stockholm, Sweden, paper WeL3:3, June 29,
(7) I. Francois, A. de Villiers, C. Brunelli, F. David, (22) D.M. Lubman, Y. Zhu, H-Y Kim, Y. Wang, R. 2005.
and P. Sandra, HPLC 2005, Stockholm, Swe- Wu, K. Cho, M. Pal, A. Sreekmar, and A. Chin- (31) R. Edam, J. Blomberg, H. -G. Janssen, and P.J.
den, paper P10:10, June 29, 2005. naiyan, HPLC 2005, Stockholm, Sweden, paper Schoenmakers, J. Chromatogr., A 2086, 12–20
(8) Recent Applications in Multidimensional Chro- TuL2: 1, June 27, 2005. (2005).
matography, LCGC Eur. 16(12ª), 2003. (23) A. Rudolphi and K.-S. Boos, LCGC 15(9),
(9) http://www.lcgceurope.com/lcgceurope/issue 813–823 (1997).
/issueDetail.jsp?id=2886. (24) E. Machtejevas, K.K. Unger, D. Lubda, R. Hen-
(10) J.C. Giddings, J. High Resolut. Chromatogr. 10, driks, and M. Miliotis, HPLC 2005, Stockholm, Ronald E. Majors
319–323 (1987). Sweden, paper TuL7: 1, June 27, 2005. “Column Watch” Edi-
(11) J.V. Hinshaw, LCGC 22(1), 32–40 (2004). (25) J. Haginaka and H. Sambe, HPLC 2005, Stock- tor Ronald E. Majors
is business develop-
(12) R.E. Majors, LCGC, 23(9) 1016–1023 (2005). holm, Sweden, paper MoL7: 3, June 26, 2005.
ment manager, Con-
(13) P. Schoenmakers, HPLC 2005, Stockholm, Swe- (26) P. Jandera, K Novotna, J. Fischer, M. Halama, sumables and Acces-
den, paper SuPL:1, June 25, 2005. and L. Kolarova, HPLC 2005, Stockholm, Swe- sories Business Unit,
(14) H. Poppe, J. Chromatogr., A 778, 3–21 (1997). den, paper TuL1: 2, June 26, 2005. Agilent Technologies,
Wilmington,
(15) D. Deans, Chromatographia 1, 18–22 (1968). (27) N. Tanaka, H. Kimura, D. Tokuda, K. Hosoya,
Delaware, and is a
(16) W.J. Bertsch, J. High Resolut. Chromatogr. 22, T. Ikegami, N. Ishizuka, H. Minakuchi, K. member of LCGC’s
647–665 (1999). Nakanishi, Y. Shintani, M. Furuno, and K. editorial advisory board. Direct correspon-
(17) W.J. Bertsch, J. High Resolut. Chromatogr. 23, Cabrera, Anal. Chem. 76, 1273–1281 (2004). dence about this column to “Sample Prep Per-
spectives,” LCGC, Woodbridge Corporate
167–181 (2000). (28) R. Ricker, C. Woodward, and R.E. Majors, Am.
Plaza, 485 Route 1 South, Building F, First
(18) L.R. Snyder, J. Chromatogr. Sci. 8, 692 (1970). Lab., July 2004, 25–32. Floor, Iselin, NJ 08830, e-mail lcgcedit@lcgc-
(19) J.F.K. Huber, R. van der Linden, E. Ecker, and (29) T. Ikegami, H. Kimura, T. Hara, H. Kobayashi, mag.com.

Circle x

You might also like