Professional Documents
Culture Documents
fabian HFOi
Instructions for Use SW 3.4
Warning
Carefully read the instructions of use
before using the ventilator to
familiarise yourself with the function of
the equipment
DISCLAIMER
Acutronic Medical Systems assumes no responsibility for the use or reliability of its software on equipment
that is not furnished by Acutronic Medical Systems.
Acutronic Medical Systems makes no warranty of any kind with regard to software applications that are
created by the user.
This document contains confidential and proprietary information that is protected by copyright. All rights
are reserved. Any unauthorized copying, storage, reproduction or translation of this document in any form is
strictly prohibited.
Manufacturer:
ACUTRONIC Medical Systems AG
Fabrik im Schiffli
8816 Hirzel / Switzerland
7.5 Language................................................................................................................................................................... 49
11.2 SIPPV........................................................................................................................................................................ 63
11.8 O2 Therapy mode (High and Low flow oxygen therapy) HFNC ........................................................................ 67
20.2 Monitoring............................................................................................................................................................... 89
20.3 Measuring................................................................................................................................................................ 90
These instructions for use describe equipment components and their operation. It is laid out such that
you can step by step become familiar with operation of the ventilator.
Please carefully read the instructions for use before using the ventilator.
Once you are familiar with the basic construction and operation of the ventilator you may use the
instructions for use as a reference.
The table of contents will assist you in quickly locating the respective topic.
DANGER indicates a not directly imminent but latent hazard which may result in death or
bodily injury unless prevented.
Danger
ATTENTION indicates a risk which may result in equipment damage when not avoided.
1.4 Symbols
Symbol Description
USB connection
Nebulizer Connection
Caution, refer to operator’s manual for important safety information and precautions
Disposal information
Flammability warning
Do not cover
Chargen code
Single use
Article No.
non Sterile
Latex free
Keep dry
This way up
Do not stack
The ventilator may not be used in combination with inflammable gases or narcotic agents
– there is otherwise an acute risk of fire and explosion!
An audible signal indicates a system or patient alarm and always requires action by a
medical professional.
Never connect to electrical devices not mentioned in this operator’s manual without first
consulting the manufacturer or an expert.
Never operate the ventilator while covered or set up in a way so as to negatively impact
the operation or function.
Note: the safety and health of the users is guaranteed by the fact that the products for
example do not contain any allergenic or mutagenic materials such as phthalates.
The device can only be separated from the main power supply by removing the power
cord completely. Please ensure the power socket is freely accessible at all times for
disconnection. Do not remove the power cable unless for Service purposes or transport.
Note: accessories supplied by Acutronic medical Systems AG for use on fabian ventilators
are biocompatible. Please ensure before applying non-original accessories that they are
biocompatible.
Warning: When connected to a patient do not simultaneously touch the external power
supply and the flow sensor connector cable.
Warning: if the strength of the auditory alarms are less than the ambient sound this might
impede an operator to recognize alarm conditions.
Warning: please ensure that the ventilator is never positioned in such a way that the
performance of the ventilator is adversely affected.
Warning: please ensure that Alarms are set appropriately prior to use of ventilator on
patient.
Warning: In case portions of the gas pathways through the VENTILATOR can become
contaminated with body fluids or expired gases during NORMAL CONDITION or SINGLE
FAULT CONDITION, please contact Acutronic Medical Systems.
2.2 Maintenance
This device is a ventilator classified as Class IIb according to the European directive, as such:
• Inspection as per manufacturer specifications is required every 12 months.
• Maintenance must be performed by Acutronic trained experts with access to appropriate test
and measuring equipment.
We strongly recommend ACUTRONIC Medical Systems AG exclusive representatives for service agreements
and repairs.
Only use original ACUTRONIC parts for repairs. Note chapter “Maintenance Intervals”.
The equipment is operated by a physician or at his orders by a professional with technical training in this
task, whereupon any user must be trained on this equipment and familiar with the operator’s manual and
the use of the equipment.
fabian HFOii is not approved for use in a homecare environment.
4 External bias flow port and port for nCPAP system based on flow generators (single limb systems)
2 Network jack for data management , PDMS (To network with minimum 3 KV galvanic isolation)
3 USB port for data output, Software update. Connection for Masimo SpO2 module.
8 Loudspeaker
9 Fan
Warning: Do not connect Ethernet, Nurse call, USB, R232 (CO2), Flow Sensor port to
anything other than specified devices.
The PIN for the Potential Equalisation is there for additional safety and can be connected to an
equipotential zone. Please adhere to local guidelines when using this PIN. The guidelines may vary
between countries, localities and power companies. Always keep the PIN for Potential equalisation
accessible.
4.1.1 Buttons
The key pad features 2 rows of buttons with various functions.
Manual breath
… used to apply a manual breath. Available in all the ventilation modes. In HFO mode,
it can be disabled in the ventilation menu.
O2 Flush
… used to start O2 flush. Flush concentration and time can be preset.
Menu / Calibration
… used to access the setting and calibration menus. Pressing once will open the
configuration menu; pressing again will open the calibration menu.
Nebuliser
… switches the optional medication nebuliser on and off.
The flow sensor is switched off during medication nebulisation.
Waves / Loops
…switches to ^wave-/loop display.
Home
… displays the main screen for selecting the respiration mode.
On/Off
… used to switch the device on or off.
Mute alarm
… for acknowledging and muting alarms for a maximum duration of 120 seconds.
Subsequent alarms with higher priority are only displayed optically during muting.
Start/Stop
… used to halt or commence respiration. In ventilation mode, the stop is for 2 minutes,
in nCPAP and DUOPAP mode acts as standby mode.
The rotary pulse encoder combines a push button with a rotary encoder for executing
various settings, selections and confirmation options.
Numeric field
The upper and lower limits can be manually or
automatically adjusted in the alarm limits menu.
Alarm limits
LOOP display:
• Pressure / Volume
• Volume / Flow
Warning LED
This LED will light or blink red when a system alarm
is triggered.
Battery LED
This LED will light yellow in battery operation and
green when the battery is fully charged.
The LED will blink green when the battery is
charging.
Mains LED
This LED will light green in mains operation, and
will switch off in battery operation.
When powered via the external 24 VDC connector the mains and battery LED
will not light up.
After confirming a parameter the audible alarm is automatically suppressed for 15 seconds.
The audible alarm may be immediately activated by pressing the Mute Alarm key.
Press any blank area or another parameter to not accept the parameter and keep
the previous value
Alternating alarms active at the time the mode is switched will automatically be reset.
1. Curves
Curves
2. Loops
Loops
Trends will be loaded
when pressing the trends key
3. Trends
Trends
Save loop:
1. Freeze loop
2. Push “Freeze” key
3. The loop will be saved until a new loop
is saved.
Comparing loops:
1 Switch to loops display:
2 Freeze current loop.
3 Current loop is shown in blue,
the saved loop black.
3
1
2
5 6 7
1. Parameter / selection
2. Graphics
3. Parameter values
4. Timeline
5. Adjust timeline (max. 5 days – min. 30 minutes)
6. Shift timeline
7. Update data (refresh)
The fuses used must correspond with the value and type approved for the fabian.
The particular fuse type and value is printed on the back of the fabian.
Replacement of the fuses may only be carried out by trained staff. Please replace fuses with
identical values, failing to do so, might cause fire hazards.
WARNING: To avoid risk of electric shock, this equipment must only be connected to a
mains supply with protective earth.
If there is no central gas supply, gas supply from cylinders is also possible.
The pressure gradient on the Ventilation system of the ventilator measured on the patient
connection port may increase when accessories or other components are attached to the
Ventilation system.
The labelling
TOP
must be legible upon installation. See Illustration.
Connect tube „art. No. 7067“ with the NO Flowsensor. Place the NO flow sensor on top of
the water chamber of the humidifier with a T-piece connector in between. Select in Quick
Launch external Bias Flow. With this setting will only come from the FG port, and will
halted at the Insp. Port.
2. A warning message appears Select the Confirm Box to switch to external bias flow.
When switching between HFO to a Conventional Ventilation Mode one can leave the external NO flow
sensor in this position.
Place temperature probe outside the incubator and use incubator extension line.
1 Ventilation tubes
2 Y-piece
3 Flow sensor
4 Flow sensor connection cable
5 Proximal pressure measurement tube
If the humidifier does not feature an inspiratory tube heater, water traps must be installed in the
inspiratory and expiratory tubes.
When using an inspiratory heater a water trap must be installed at the expiratory end.
We strongly suggest using an inspiratory and expiratory heated tube system in HFO mode.
For details of using nCPAP systems, refer to the operators manual of the supplier of the consumables.
(ok ≙ no error)
Gas supply Attach high pressure air and oxygen Air and oxygen supply hoses are correctly
supply hoses on the rear side of the connected
ventilator. Connect with corresponding
wall outlets.
Function test Ventilation mode: IPPV Ventilation occurs with the set frequency.
• I - flow: 10 l/min • P insp.: 20 ±2 mbar
• E - flow: 8 l/min • PEEP: 5 ±1 mbar
• PEEP: 5 mbar • O2: 30 ±2 vol.%
• Pinsp: 20 mbar
• Frequency: 30 1/ min
• I-time: 1 sec
• Oxygen: 30 vol.%
1. Calibration 7. Tools
2. Display 8. Info
3. Ventilation 9. Video ( optional )
4. Patient Data 10. PDMS ( optional )
5. Language 11. Service Mode
6. Data / Time 12. Close menu (return to previous menu)
The calibration menu can also be accessed via the calibration key.
Finally, upon ventilator start-up it you will be automatically directed to the calibration menu.
The calibration menu can be seen below.
From the calibration screen, one can calibrate the following sensors:
• Flow sensors: neonatal and pediatric
• O2 sensor
• etCO2 module
• SpO2 Module
For both types of flow sensors the calibration procedure is identical and described below.
Ensure no gas will flow through the flow sensor
during calibration.
To do so the sensor may be held occluding either
one or both ends with a sterile glove.
The flow sensor calibration needs to be performed each time a new sensor is put in place.
Sensor calibration remains valid, even if unit is switched off and after restart of it.
We recommend cleaning flow sensor once per day.
The O2 sensor is automatically calibrated when machine starts up and every 24 hours when
in use.
During calibration procedure, the FiO2 concentration of the fresh gas to patient is not
altered.
In case of a supply gas failure, oxygen sensor calibration is disabled automatically to avoid
false sensor calibration
Manual calibration:
Following successful calibration, the date of the last calibration will be displayed in the corresponding field.
Automatic sensor calibration runs every time the equipment restarts and every 24h. The
gas concentration to patient however is not altered during this calibration procedures.
For more information on the CO2 and SpO2 configurations please refer to the “fabian HFO CO2 manual” and
“fabian HFOi SpO2 Manual”.
1. Manual
2. After default time
Graphs display:
• Filled
• Not filled
filled
not filled
Trends display:
• Filled
• Not filled
filled
not filled
2. Manual breath HFO The manual breath in HFO mode can be used to perform lung
(ON / OFF) recruitment manoeuvre (sustained lung inflation) if clinically
accepted by internal guideline. Length and pressure level
adjustable.
3. Tube set Infant Flow®, Infant Selection of the nCPAP system used.
Flow LP®, Medijet®
5. Automatic oxygen
I cases where oxygen supply source does not provide 100%
calibration oxygen, this should be set to 21% only to avoid calibration at false
(21 Vol% / 21 and 100 Vol.%) value.
10. Alarm delay After visual reset of alarm, the reactivation of alarm is delayed for
the set period. Avoids an immediate re-alarm in case of values have
not stabilized yet.
11. Ppsv Parameter Select Ppsv parameter to be absolute value or above PEEP
parameter
12. Hospital settings The “Hospital Settings” button is only active if specific settings were
saved in fabian HFO.( see chapter Hospital settings )
This menu is also where the nurse call can be activated and deactivated.
Note:
Entering new patient data will delete all trend data.
1. Reset data
2. Accept
3. Change
Note:
Changing the date or time will delete all trend data.
Save of:
Trends
Trend data will be output in CSV format.
Log files
Alarm and Error log
7.8 Info
Once the alarm limit has been adjusted or there are no more active alarms, the bell symbol will turn grey.
The visual alarm message must be reset by pressing the “Alarm silence” button.
Each displayed breath data is checked for violations of the current alarm limits. Additionally the measured
pressure will be checked against the limit every 20 milliseconds with following conditions:
If an alarm condition has been detected, a dedicated alarm array is written, otherwise it is cleared.
• Date / Time
• Message
• Priority
(highlighted. See: chapter
Error – Cause – Alarms)
These values may vary depending on the parameter settings. The fabian HFOi will continuously calculate the
remaining run time and display it in the info bar. The battery charge is continuously monitored.
In the event of power failure the pneumatic system will automatically open against
atmospheric pressure so that pressure can’t build in the breathing system, allowing
spontaneous patient breathing.
When operating fabian HFOi on the internal battery the HFO amplitude is limited to 60
mbar.
Battery drained!
– Connect to mains –
– Charge battery after approx. 5 hours –
*Optional.
*Optional.
*Optional.
• IPPV
Intermittent Positive Pressure Ventilation
• SIPPV
Synchronised Intermittent Positive Pressure Ventilation
• SIMV
Synchronised Intermittent Mandatory Ventilation
• SIMV + PSV
Synchronised Intermittent Mandatory Ventilation with PSV support
• PSV
Pressure Support Ventilation
• CPAP
Continuous positive airway pressure
• NCPAP
NIV mask ventilation (Optional)
• duoPAP
NIV mask ventilation (Optional)
• HFO
High-frequency oscillation (Optional)
• O2 Therapy
High and Low Flow Oxygen Therapy HFNC (Optional)
11.2 SIPPV
Assist Controlled Ventilation (Synchronized
Intermittent Positive Pressure Ventilation) each
spontaneous patient inspiratory effort triggers a
mechanical breath of the ventilator according to
the ventilator parameters set for inspiratory period
and pressure.
Ventilation with specified pattern and frequency The synchronisation window for the next
synchronous to the patient’s independent mechanical breath is ½ Te.
breathing.
The patient may spontaneously breathe in between
breaths but receives no pressure support.
The machine breath (SIMV) is synchronised with the patient’s breathing pattern. The amount of mechanical
breath per minute is same as the preset frequency.
If patient stops spontaneous breathing, the ventilator will deliver the amount of mechanical breath set with the
Frequency button.
Every spontaneous inspiratory effort of the patient is supported with the preset Ppsv pressure level. It ends
when flow termination criteria is met.
The inspiratory pressure for the PSV breath is set using the P psv button.
If the patient stops breathing spontaneously the ventilator will assume alternate breathing with the ventilation
parameter settings.
Apnea backup ventilation will start after preset Apnea time. If this is set to OFF, backup starts after E-Time.
The inspiratory pressure for the PSV breath is set using the P insp. button.
11.6 CPAP
In CPAP ventilation the patient breathes spontaneously, the ventilator does not provide breathing. This mode
will merely produce a positive pressure during inspiration and expiration, noticeably reducing the patient’s
breathing effort. The maximum automatic leak compensation is 100% of the inspiratory flow.
In the event of an interruption in the patient’s breathing, following the default apnoea period the ventilator will
perform a default number of mechanical breaths to stimulate spontaneous breathing. Once breathing
commences stimulation will abort and will only commence with the next apnea event.
duoPAP: Same as nCPAP but with the option of positive pressure ventilation with
adjustable frequency and inspiratory period.
This mode requires a special nCPAP patient set with nCPAP generator.
Before using the nCPAP / duoPAP mode the correct system must be specified
in the specifications menu.
If Insp. time is set lower than 1 sec. the maximum pressure may not be reached.
Depending on the tube set, nCPAP generator and humidifier
11.8 O2 Therapy mode (High and Low flow oxygen therapy) HFNC
O2 Therapy is an option which allows use of a continuous flow of blended gas, from 1 – 19 LPM. Nasal cannulas
of various makes like F&P, Atom or similar can be used. There are no alarm functions active in
this mode, except for the set FiO2
Note:
This mode can also be used to put the ventilator in standby mode. By setting a flow of 4 LPM, the humidifier
dual servo temperature controls remain active, so no need to switch it off in case of short term standby mode.
When operating fabian HFOi via the internal battery HFO amplitude is limited to 60 mbar.
During HFO ventilation the distinct recognition of disconnection and measurement of the
minute volume is limited.
Use of a heated inspiratory and expiratory tubing system is strongly recommended in HFO
mode.
Changing the ventilation pattern or adjusting the volume limit the tidal volume is automatically limited.
The display will show the message: “Volume limited”
In the event of flow measurement outage ( e.g. changing the sensor) „V Limit“ will
automatically be deactivated.
Once flow measurement is available again “V Limit” will automatically be reactivated.
Pinsp will be gradually increased breath by breath to achieve preset Vte. The maximum pressure increase
breath by breath is limited to 3 mbar.
If the Pinsp setting is insufficient for applying the targeted volume, the following message is displayed: “Tidal
volume not reached”. An upward arrow in Pinsp field guides user that to achieve preset Vte, the Pinsp must be
increased.
In the event of flow measurement outage ( e.g. changing the sensor) „V Guarant“ will
automatically be deactivated.
Once flow measurement is available again “V Guarant” will automatically be reactivated.
The duration of the manual breath can be set under Configurations Menu -
> Ventilation.
This may last 2-30 seconds, then a termination of the manual breath will
be forced.
The next manual breath is only allowed after a block period of 200 ms.
Press the “O2 Flush” button in a ventilation mode to initiate the O2-Flush.
The flush concentration may now be set with the rotary pulse encoder.
Pressing the O2 button again will end the flush prematurely.
In both cases the built-in gas mixer delivers a minimal flow to prevent heat build-up inside
the ventilatory gas moistener.
Ventilation will resume following the 2 minute pause or the button is pressed again.
Air supply pressure Air supply pressure too low. Ensure pressure above 2 bar. I
Flat battery! Remaining battery time < 1 Immediately switch to mains supply
I
min.
Flow sensor not Flow sensor not connected. Check flow sensor and sensor cable
I
connected Defective sensor cable. connection.
High tube leak Upper alarm limit exceeded Adjust limit value.
II
Inspect tubing system.
Oxygen supply Oxygen supply pressure too Ensure pressure above 2 bar.
I
pressure low.
High DCO2 alarm Exceeded upper alarm limit Adjust limit value.
II
Check tubing system.
Low DCO2 alarm Value below lower alarm Adjust limit value.
II
limit Check tubing system.
Pressure too low Leak or disconnection. Check tubing system for tight connection.
Insp. flow or exp. flow set Increase flow. I
too low.
Tidal volume not Inspiratory pressure default Adjust inspiratory pressure setting.
reached is reached prematurely. Increase inspiratory period. II
Adjust volume setting
Switch on fabian.
Calibrate oxygen sensor at 21 %Vol.
and 100 %Vol.
Dispose of O2 sensors in accordance with local waste management regulations (cf. battery disposal).
Please contact local environmental or regulatory agencies and appropriate waste management companies for
information.
When installing a brand new sensor a second calibration can be necessary after about
30 minutes. If possible, expose the new sensor for about 30 minutes to the ambient air
BEFORE installing
Every 6 months
Every 12 months
Every 4 years
Maintenance and safety inspections must be performed by Acutronic trained experts with
access to suitable testing and measuring equipment.
Flow Sensor
• Do not clean sensor with compressed air or water jet, as the sensor wires could be destroyed.
• Never treat sensor in automatic cleaning-/disinfection machine.
• Rinse flow sensor element in distilled water.
• Thoroughly shake off residual water.
Steam sterilization at 132°C for 2 minutes (half cycle).
The flow sensor should be changed when it is unable to be calibrated.
Replace and clean sensor once a day. It has been shown that this increases the service life of the sensor.
Never connect an uncalibrated sensor to a patient. Always calibrate the sensor before connecting to a
patient.
Clean and disinfect wet thermally (80 - 95°C / 5 min) in the automatic cleaning/disinfection machine just
with detergent. Steam sterilization at 132°C for 2 minutes ( half cycle )
Change the membrane when the unit fails the device check – Function Test. See chapter 6.
In general, sterilize the membrane holder to ensure that residual water in the channels and control cables
dries out.
Ventilation tubes
Heating wire
Water trap
Temperature probe
Y-piece
When powering fabian HFO by internal battery HFO amplitude is limited to 60 mbar.
Step 2
Please select the appropriate Vte to be guaranteed.
Important: To ensure the VG function works appropriately, a small plateau in the tidal volume wave needs
to be present.
Triggered and non-triggered breaths are independently supported based on lung compliance. Normally,
a triggered breath has a lower Pinsp than a non-triggered one due to different lung compliance.
Step 4
VG can now be activated by selecting the “ON/OFF” button. When VG function is on this button will
turn green. Furthermore a red line will appear in the pressure wave diagram to marks the Pmax.
If the targeted volume cannot be reached with the set Pmax the warning: “Vte not reach, check settings”
will appear. An upward pointing arrow will appear in the “Pmax” field to indicate the maximum pressure
has been reached.
In case the flow sensor is deactivated, the VG will be shut off and the pre-set Pinsp will be used to ventilate.
PSV breaths as well as apnea backup breaths are volume targeted breath in this mode.
The delay to start backup ventilation is set with the Apnea time in alarm limit screen.
If Apnea is set to OFF > backup starts after E-Time
In PSV+VG mode, the patient controls the onset of inspiration as well as the end of inspiration, offering a
total synchrony with patients breathing pattern.
If patient lung compliance is improving, the ventilator automatically is using the lowest pressure, necessary
to maintain the preset target volume.
In case of a respiratory fatigue and a ceasing inspiratory effort, the ventilator automatically will provide an
apnea backup ventilation with the preset parameters for I-Time, Frequency and Pbackup.
The delay for the onset of the apnea backup ventilation is set in the alarm screen, by setting the apnea delay
time. If set to OFF, the ventilator will start apnea backup ventilation immediately after E-Time, otherwise
after preset apnea time.
It is recommended to use a delay time to avoid the ventilator to kick in unnecessarily and interfering with
patients breathing pattern.
The suction catheter system has to be placed between the flow sensor and the patient. See picture
Choose the right catheter for the suction. Otherwise, a too high vacuum is possible!
Silence the alarms before suction (2 minutes) for not disturbing the suctioning. After the suction reactivate
the alarm again.
The minimum and maximum ambient temperatures in the clinical area will not affect the accuracy of the
flow, pressure or oxygen measurement.
Depending to the accessories the environmental conditions can be different. Note Related Manuals.
20.2 Monitoring
Acoustic pressure of alarm tones. Alarm tone sequence IEC:
High priority: 49 db(A) – 65 db(A)
Medium priority: 49 db(A) – 63 db(A)
Low priority: 48 db(A) – 63 db(A)
Inspiratory O2 concentration
Alarm: Set value ± 5 Vol.% for > 1 minute at 21 Vol% below threshold value 18 Vol%
Breath frequency
Alarm: If alarm limit exceeded
Volume monitoring
Alarm lower threshold value: If the breath volume set was not supplied
Apnoea alarm
Alarm: If no breathing activity is recognized
Drift: < 1 Vol% per month. < 15% rel. over entire service life
Pressure influence: Proportional to change in partial pressure
Warm-up time: max. 3 minutes (< 30 minutes with new O2 sensor)
In addition an audible alarm can be heard on start-up. The user must check if this alarm can be heard to
verify that the speaker is functioning.
4
3
Warning: Portable and medical HF communication devices could impact MEDICAL APPLIANCES!
Warning: Never use "fabian HFOi" directly adjacent to, or stack with other apparatuses. If unavoidable, be
sure to monitor the equipment for proper operation with this set-up.
Warning: Using other accessories, cables or converters with "fabian HFOi" may result in increased
electromagnetic interference and reduce the immunity of "fabian HFOi".
Note: The key performance characteristics of the "fabian HFOi" ventilator are:
• fabian HFOi must operate within the defined specifications and the medical purpose.
Failure to meet these specifications will result in respiration being stopped.
• For this reason a second, stand-alone ventilation unit must always be available,
e.g. resuscitation bag.
The device "fabian HFOi" is intended for operation in the environment described below. The customer or
user of the "fabian HFOi" apparatus should ensure it is operated in this type of environment.
HF emission CISPR 11 Group 1 The device "fabian HFOi" uses HF energy solely for internal
operation. Its HF emission is thus very low and interference
with adjacent apparatuses is unlikely.
HF emission CISPR 11 Class A The device "fabian HFOi" is suitable for use in all other
areas except residential areas and areas directly connected
Emission of harmonics Class A
to a public mains supply which also supplies buildings
IEC 61000-3-2
used for residential purposes.
Emission of voltage Met
fluctuations / flickers
IEC 61000-3-3
The device "fabian HFOi" is intended for operation in the environment described below. The customer or
user of the "fabian HFOi" apparatus should ensure it is operated in this type of environment.
Quick electric ± 2kV for ± 2kV for power The mains quality should correspond to
transients/burst power cords cords typical business or hospital environments.
<5% UT <5% UT
(>95% drop of UT) (>95% drop of UT) for
for 5 seconds 5 seconds
Remark UT is the mains alternating voltage prior to applying the test level.
10 Veff 10 V d = 1.2√P
150 kHz to 80 MHz
within ISM bands b
REMARK 1 At 80 MHz and 800 MHz the higher frequency range applies.
REMARK 2 These guidelines may not apply to all case. The propagation of electromagnetic quantities is influenced by
absorption and reflexion of buildings, objects and people.
a The ISM bands (industrial, scientific and medical) between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.7 MHz.
b The compliance levels in ISM frequency bands between 150 kHz and 80 MHz and between 80 MHz and 2.5 GHz are defined
as such so as to decrease the probability of mobile/portable transmitters causing interference in the event they are
accidentally brought in to the vicinity of the patient. For this reason an additional factor of 10/3 is used when calculating
the recommended safety distance of transmitters in this frequency range.
c In theory the field intensity of stationary transmitters, as e.g. base units of radio telephones (mobile/cordless) and mobile
land radio devices, amateur radio stations, AM and FM radio stations and television stations cannot be predefined
precisely. A study of the location should be conducted to determine the electromagnetic environment with regard to
stationary transmitters. If the field intensity at the location where the "fabian HFOi" is used exceeds the above compliance
levels, "fabian HFOi" should be monitored for proper function. If unusual performance characteristics are observed,
additional measures may be required, as e.g. changing the direction or location of the "fabian HFOi".
d Above a frequency range of 150kHz to 80MHz the field intensity should be below 10 V/m.
100 4.0 12 12 23
For transmitters with a maximum power output not listed in the above table the recommended safety
distance d in metres (m) can be determined using the equation from the corresponding column, with P
being the transmitter’s maximum power output Watts (W) per the transmitter’s manufacturer
specifications.
REMARK 1 At 80 MHz and 800 MHz the safety distance of the higher frequency applies.
REMARK 2 The ISM bands (industrial, scientific and medical) between 150 kHz and 80 MHz are 6.765 MHz
to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.7 MHz.
REMARK 3 An additional factor of 10/3 is used when calculating the recommended safety distance for
transmitters within the ISM frequency between 150KHz and 80 MHz and between 80MHz and 2.5GHz to
reduce the probability of mobile/portable transmitters causing interference if accidentally brought into
the vicinity of the patient.
REMARK 4 These guidelines may not apply to all case. The propagation of electromagnetic quantities is
influenced by absorption and reflexion of buildings, objects and people.
Guideline 93/42/EWG