Professional Documents
Culture Documents
Instruction:
Dear user, thanks for choosing our BF-6500 Automatic Hematology Analyzer.
Please read the user manual carefully before use in order to operate the instrument correctly.
Please keep the user manual safely for your any time reference.
Warning:
●An independent power supply is a must. Electromagnetism interference will affect the accuracy
of the test result.
●Don‗t pull the electrical wire with wet hand, or there is a risk of electrical shock.
●Don't stamp, twist, drag the wire and cable, or it may cause a fire. The damaged wire and cable
●The input power should conform to instrument requirement. Specified fuse should be used.
●Make sure the switch is on [O] state before connecting the power.
●DO NOT touch moving parts when the instrument is testing to avoid accident.
●Non-professionals can not open the left, right and upper cover of the instrument when the main
power is ON.
●Make sure the instrument is used under the condition that is specified in user manual. In
improper condition, the instrument may not work well, the result may be inaccurate, instrument
I
BF-6500 Automatic Hematology Analyzer User Manual
Note:
● Instrument should be operated by medical inspection specialist, physician, nurse or lab
● Maintenance plan of hospitals and inspecton bodies should be prepared and followed.
●Expired reagent can not be used. The reagent should be protected from dust, dirt and bacteria
once opened.
●Soft cloth or gauze can be used for cleaning work. A little diluted detergent and alcohol can be
used if necessary. Pine oil and benzene can not be used for outside cleaning, it may cause color
●Outdoor temperature in winter is quite low. The analyzer should be placed at room temperature
II
BF-6500 Automatic Hematology Analyzer User Manual
Biohazard Marking:
●Please dispose the reagent, waste solution, waste sample and consumable according to the
local regulation.
●Sample, Control, Calibrator and waste solution have potential biochemical infectivity that may
hurt eyes, skin and mucosa. Operator should refer to the safety regulation for lab operation.
Protective measure should be taken (Such as lab protective clothes and gloves).
●Please dispose the waste solution and instrument consumable according to the regulation of
medical waste, infective waste and industrial waste. Blood in the waste may have been
contaminated by pathogens.
●Avoid reagent contact with eyes and skin, in case of any contact, rinse immediately with plenty
●Sampling probe is sharp-pointed which may with some material have potential bio-infectivity,
such as blood sample, Control and Calibrator. Contact with sampling probe should be avoided. In
sampling, there should be a certain distance between probe tip and the wall of the container to
III
BF-6500 Automatic Hematology Analyzer User Manual
Content
Chapter 1 Brief Introduction ................................................................................................................................ 1
V
BF-6500 Automatic Hematology Analyzer User Manual
6.2 X QC ......................................................................................................................................... 93
6.3 X-B QC ........................................................................................................................................ 93
6.3.1 QC Setting ............................................................................................................................ 94
6.3.2 QC Counting......................................................................................................................... 95
6.3.3 QC Graph Review ................................................................................................................ 95
6.3.4 QC List Review .................................................................................................................... 97
VII
BF-6500 Automatic Hematology Analyzer User Manual
1.1 Overview
BF-6500 Automatic Hematology Analyzer is used in medical laboratories to
quantitative analyze blood cell, carrying out 5-differential to white blood cell(WBC) counting result,
and offer scattergram of white blood cell 4-diff, histogram of red blood cell(RBC), white blood
instrument with high-capability. It characterize in accurate test result, easy operation, low
consumable. The instrument can test quantitative analysis result of 24 blood parameters and 10
The instrument is an in vitro diagnostic medical instrument used by professionals for screening.
When making clinical judgment according to the analysis result, the doctor should taking into
CBC Mode:40μL(diluted)
Throughput: 60 T/H
WBC(differential) 1:200
WBC(differential ) 1:2000
1
BF-6500 Automatic Hematology Analyzer User Manual
Neutrophil Neu#
Lymphocyte Lym #
Eosinophil Eos#
Basophil Bas#
Hemoglobin HGB
2
BF-6500 Automatic Hematology Analyzer User Manual
Hematocrit HCT
Plateletcrit PCT
Diff Scattergram
RBC Histogram
PLT Histogram
WBC Histogram
Fuse: 250V 4A
Dimension: Mainframe:490mm(L)×540mm(W)×550mm(H)
2001+A3 2003 of EMC test standard, meeting A level requirements, should not be installed near
laser flow cytometry to obtain WBC 4-differential counting result. The result of other parameters
In whole blood mode, the instrument aspirates 20μL(CBC+DIFF mode)or 10μL(CBC mode)
In pre-dilution mode, the operator should mix 20ul peripheral blood with 180ul diluent to form a
sample with dilution ratio of 1:10, and then send it to the analyzer for aspiration. The analyzer will
WBC differential detector, WBC counting cell and RBC counting cell in order according to the test
requirement. Through the effects of different reagents while dilution process, forming samples
which are used for WBC differential testing, WBC/ hemoglobin testing and RBC/PLT testing.
For different sample, the instrument offers two different working modes - whole blood mode and
Diluent 3mL
Aspirate 50μL
Diluent 3mL
Diluent 3mL
5
BF-6500 Automatic Hematology Analyzer User Manual
6
BF-6500 Automatic Hematology Analyzer User Manual
of diluent. Under the package of sheath which is formed by diluent, the single cell flow goes
Blood cells go through the laser area after twice acceleration. Under the effect of laser beam, the
7
BF-6500 Automatic Hematology Analyzer User Manual
scattered light is related to cell size, refractive index of cell membrane and cell internal structure.
Low-angle forward scattered light reflects the size of cell. High-angle forward scattered light
reflects the internal fine structure and particulate matter. Photodiode receive the scattered light
signals and translate them into electrical pulses, according to the electrical pulses, two-dimension
map(scattergram) of cell size and cell internal information can be obtained. The abscissa reflects
the cell's internal structure. The vertical axis reflects the cell size, as shown in Figure 1-3-8:
The lymphocytes, monocytes, eosinophils and eutrophils percentage can be obtained from the
dispensed into test unit after diluted by quantitative conducting solution. The testing unit has a
test aperture. A pair of positive and negative electrodes exist beside the aperture for connecting
the constant current power supply. As the cells have the characteristic of a poor conductor, when
the cell goes through the aperture under constant negative pressure, the DC resistance between
the electrodes will change, resulting in the formation of a pulse signal which is proportional to the
cell size. A series of electrical pulse is produced when the cell continuously go through the
aperture. The number of pulses is equivalent to the cell number through the aperture. The pulse
8
BF-6500 Automatic Hematology Analyzer User Manual
Compare the amplified electric pulse with voltage range corresponding to normal WBC size range.
Calculate the electrical pulse number. All electrical pulse is classified according to different
channel voltage range. Electrical pulse number which fell in WBC channel is WBC number. Cell
number in each channel which is divided according to pulse voltage range determines the cell
size distribution.
mononuclear cells (Mon%), as well as the percentage of eosinophils (Eos%) can be obtained.
Calculate by using the WBC number, lymphocytes (Lym #), neutrophils (Neu #), mononuclear
9
cells (Mon #) as well as eosinophils (Eos #) can be obtained. Cell unit is 10 /L.
●WBC
WBC number can be obtained through testing the corresponding pulse number.
●Basophil
Basophil number can be obtained through testing the corresponding pulse number.
●Basophil Percentage
Bas #
Bas % 100%
WBC
●Lymphocyte Percentage
9
BF-6500 Automatic Hematology Analyzer User Manual
●Neutrophil Percentage
●Monocyte Percentage
●Eosinophil Percentage
●Lymphocytes
Lym#= WBC×Lym%
●Neutrophil
Neu#= WBC×Neu%
●Monocyte
Mon#= WBC×Mon%
●Eosinophil
Eos#= WBC×Eos%
method. It characterize in high hemoglobin conversion speed and no toxic substance, applicable
diluted sample is mixed with lyse, RBC dissolve, releasing hemoglobin. Hemoglobin combined
with lyse to form hemoglobin complex. At one end of the colorimetry pool, the hemoglobin
complex is irradiated by monochromatic light with a wavelength of 540nm (LED light tube).
Phototube is used at the other end to receive transmission light, and convert the light signal into
voltage signal. By comparing with the voltage produced by background transmission light
intensity before sample adding(only diluent exists), the hemoglobin concentration can be
obtained(HGB), unit is g/L. This testing and calculation process will be finished by the
10
BF-6500 Automatic Hematology Analyzer User Manual
analyzer automatically, and the result will be displayed in the counting interface.
platelet sample flow into RBC test unit after twice dilution. The testing unit has a test aperture. A
pair of positive and negative electrodes exist beside the aperture. As the cells have the
characteristic of a poor conductor, when the cell go through the aperture under constant negative
pressure, the DC resistance between the electrodes will change, resulting in the formation of a
pulse signal which is proportional to the cell size. A series of electrical pulse is produced when the
cell continuously go through the aperture. The number of pulses is equivalent to the cell number
through the aperture. The pulse amplitude is proportional to the cell size.
Compare the amplified electric pulse with channel voltage value corresponding to normal
RBC/PLT size range. Calculate the electrical pulse number. All electrical pulse is classified
according to different channel voltage value. Electrical pulse number which fell in RBC/PLT
channel is RBC/PLT number. Cell number in each channel which is divided according to pulse
voltage range determines the cell size distribution. The two-dimensional map with cell size as
abscissa and cell number as vertical axis is the histogram reflects the distribution of cell.
premise of getting accurate result. Quantitative injection pump ensures the sample volume that
goes through the testing aperture is tested. Sample volume is determined by the running steps of
motor.
Unit: 1012/L
RBC=n×l012/L
11
BF-6500 Automatic Hematology Analyzer User Manual
●RBC hematocrit, mean RBC hemoglobin content, mean RBC hemoglobin concentration
Calculate RBC HCT according to the following formula, unit %; mean RBC hemoglobin content
(MCH), unit Pg; mean RBC hemoglobin concentration (MCHC), unit g/L
RBC MCV
HCT HCT
10
HGB
MCH
Mean hemoglobin content RBC
HGB
MCHC 100
Mean hemoglobin concentration HCT
in figure 1-3-10.
Figure 1-3-10
● RBC Distribution Histogram
The RBC volume distribution histogram is offered when the result is obtained. The graph that can
indicate the distribution of cell population is RBC distribution histogram. Histogram abscissa is
RBC size (unit: fl), vertical axis is RBC relative number (unit: 10 12/L). After each counting, RBC
distribution histogram can be obtained in analysis result area of counting interface. RBC
12
BF-6500 Automatic Hematology Analyzer User Manual
●PLT Number(PLT)
PLT number is obtained through testing corresponding electrical pulse number, unit 109/L
PLT=n×l09/L
volume (10GSD)
●PCT
Calculate PCT according to the following formula, unit %; PLT unit 109/L unit ; MPV unit fl
PLT×MP V
PCT=
10000
●Big PLT Ratio(P-LCR)
indicate the distribution of cell population is PLT distribution histogram. Histogram abscissa is
PLT volume (unit: fl), vertical axis is PLT relative number (unit: 10 9/L). After each counting, PLT
distribution histogram can be obtained in analysis result area of counting interface. PLT
1.3.7 Rinsing
The instrument rinse itself automatically in each counting process, to ensure no residual sample
exists.
● Internal and external wall of sampling probe should be rinsed with diluent;
● Counting pool, quantitative pump should be rinsed with diluent;
● Flow cell should be rinsed with diluent.
13
BF-6500 Automatic Hematology Analyzer User Manual
red-power indicator from left to right) ③Sampling Probe ④Sample Aspiration Key
①Right Door ②⑥ Fan ③Upper Cover ④Back Cover ⑤Left Door ⑦BF-SLS-I Lyse
14
BF-6500 Automatic Hematology Analyzer User Manual
Inlet ⑧BF-FDO Lyse Inlet ⑨BF-FDT Lyse Inlet ⑩BF-PK Diluent Inlet ○
11
Figure 1-4-3
1.6 Symbol
The following symbols include the symbols on the analyzer, reagent, Control and Calibrator.
Symbol Meaning
15
BF-6500 Automatic Hematology Analyzer User Manual
Laser Warning
AC symbol
Storage temperature
Lot NO.
Validity
Serial number
Measurement control
Date of manufacture
Grounding
Manufacturer
16
BF-6500 Automatic Hematology Analyzer User Manual
Chapter 2 Installation
To ensure the normal working of the instrument after installation,the initial installation and set-up
Dirui.
Note:
Dedicated computer software should be used for controlling. It is recommended that the
The space, power, environment should meet the requirement prior to instrument installation. The
● The space between the wall and the right & left side of the instrument ≥50cm
● The space between the wall and the back side of the instrument≥50cm
● The space between the front of the analyzer and other equipment ≥ 100cm;
● Ensure enough space under and above the instrument for diluent, reagent and waste solution
collecting devices.
● Large load equipment like air conditioner, refrigerator, oven etc. should not be inserted in the
● Working Temperature:15℃-30℃;
● Relative Humidity:≤75%;
● Atmospheric Pressure:75kPa-106kPa;
● The instrument should be protected from dust, mechanical vibration, significant noise and
power interference.
● Do not use this instrument in close proximity to sources of strong electromagnetic radiation, as
● It should be placed far from the constant ON-OFF electrical devices like brush-type motor,
fluorescent lamp.
● It should be placed far from heat and wind source, sunlight, brush-type motor, flickering
Note:
The result will be unreliable if the room temperature or power can not meet the requirement. Or
2.2 Unpacking
damage.
● Open the accessory box, mainframe box, check the items according to the packing list. Contact
● Vibration should be avoided in removing and transportation. It should be used after checking
and debugging.
2.3.1 Connecting
Put the BF-FDT lyse、BF-FDO lyse、SLS-I lyse bottle at the back of the instrument. Connect
Diluent container should be put under the working table. Connect according to figure 2-3-1.
Note:Waste liquid should be disposed according to relevant local medical waste treatment
regulations.
(c)Computer Connecting
Connect ―Net Port‖ of computer host with ―Net Port‖ of right side analyzer(figure 1-4-2 ○
1)
19
BF-6500 Automatic Hematology Analyzer User Manual
Connect one end of supplied power line with the attaching plug on the back side of the
analyzer(③ of figure 1-4-2). And connect the power line of host, display and printer.
Note: The socket connected with the power wire should be well-grounded.
Insert one end of the bar code reader into the ―USB‖ of the computer host.
Note: The light beam of the barcode scanner may hurt eyes, therefore, staring should be avoided.
(f)Printer Connecting
Connect the printer and the computer host through data wire.
Note: Tie wires will be used for fixing sampling probe before delivery. Wire 1 and 2 should be
removed before power on, otherwise, probe may be damaged, as figure shows:
Wire1
Wire 2
Figure 2-3-2
it except abnormity occurs. In case of a must-uninstall, please follow the following steps:
computer, ―NET Framework3.5‖ install assembly will pop up, as figure 2-3-3 shows:
20
BF-6500 Automatic Hematology Analyzer User Manual
Figure 2-3-3
Select ―Accept‖ in above figure, SQL Server2005 Express Edition SP2(x86)will pop up, as figure
2-3-4 shows:
Figure 2-3-4
21
BF-6500 Automatic Hematology Analyzer User Manual
Figure 2-3-5
Click ―Install‖ in above figure, ―NET Framework3.5‖ interface will pop up, as figure 2-3-6, 2-3-7
shows:
Figure 2-3-6
Figure 2-3-7
When the above two installation have been finished, ―SQL Server 2005 Express Edition
22
BF-6500 Automatic Hematology Analyzer User Manual
Figure 2-3-8
When the above installation has been finished, if other software is running, reboot the interface
before the continuous running of popped installation program. As figure 2-3-9 shows:
Figure 2-3-9
Select ―Install‖ in above figure, the installation interface will pop up after rebooting, as figure
Figure 2-3-10
23
BF-6500 Automatic Hematology Analyzer User Manual
Figure 2-3-11
Click ―Next‖ in above figure, select the installation path, the default path is C:\Program
Figure 2-3-12
Click ―Next‖ in above figure, the confirm interface will pop up, as figure 2-3-13 shows:
24
BF-6500 Automatic Hematology Analyzer User Manual
Figure 2-3-13
Click ―Next‖ in above figure, ―Installing BF6500 ‖ interface will pop up, as figure 2-3-14 shows.
Figure 2-3-14
Click ―Next‖ after installation has been finished, figure 2-3-15 will pop up:
25
BF-6500 Automatic Hematology Analyzer User Manual
Figure 2-3-15
Click ―Close‖ in above figure to finish the installation of the software. The shortcut of the program
If the Automatic Hematology Analyzer application software need to be deleted from the current
computer, click ―Start‖ window, find ―BF-6500‖ in ―Program‖, click ―Uninstall BF-6500‖, the confirm
Figure 2-4-1
Note: Turn on the power switch of the analyzer, login the application software.
26
BF-6500 Automatic Hematology Analyzer User Manual
Double click the application software , or click ―Start‖, find the software in ―Program‖
window, enter into ―System Login‖ window, as figure 2-5-1, 2-5-2 shows:
Figure 2-5-1
Figure 2-5-2
Input username, password, the initial user name is Admin(can not be modified), the initial
password is 1). If the input username or password is wrong, the screen will display login error, as
Figure 2-5-3
27
BF-6500 Automatic Hematology Analyzer User Manual
The program will exit automatically if the username or password error occurs for continuous 3
times.
Input username, password in figure 2-5-2, click ―Login‖. Liquid pipeline, temperature and pressure
self-test will be conducted by the instrument(the screen will display ―System self-testing…..‖).
Figure 2-5-4
From left to right: network connection status, LIS system connection status, printing status:
:Printer is disconnected.
:Printer is working.
Figure 2-5-5
29
BF-6500 Automatic Hematology Analyzer User Manual
The corresponding failure information will be displayed in this area when failure occurs. Click this
area, the failure dialog box will pop up, as figure 2-5-6 shows:
Figure 2-5-6
Click the corresponding information, the detailed solution will be displayed in ―Detail Information‖
Use icon to indicate the switch status of X-B QC. With X-B is ON, without X-B is OFF.
(c) Sample Status: Display the analysis mode and test mode.
Analysis Mode: There are two modes. Whole blood mode and pre-diluted(peripheral blood).
30
BF-6500 Automatic Hematology Analyzer User Manual
Figure 2-5-7
Note:
Log off is recommended when the user is at rest. For avoiding non-user damage software or
modify the data. Periodical database backup is recommended to avoid data lose caused by
unforeseen circumstances.
Input the initial user name and password in the first login. Set the user name, permission and
31
BF-6500 Automatic Hematology Analyzer User Manual
The system parameter of the instrument has been set before delivery. The interface of the first
power on is system default. In order to meet the different needs of practical application, two
permissions are provided(user permission and administrator permission), the user can reset
some parameters.
Figure 3-1-1
Analysis Mode: There are two modes. Whole Blood mode and predilute(peripheral blood) mode.
Test Mode: There are two modes.【CBC】conduct counting without diff upon WBC, counting result
32
BF-6500 Automatic Hematology Analyzer User Manual
includes histogram and its parameter of WBC, RBC and PLT.【CBC+DIFF】conduct counting and
Sample No.: The sample No. input in this box is the next tested sample.
Note:
Sample No. with ―-‖ is accessible. ―.‖ is NOT accessible(―Sample No. Error‖ will be prompted if
there is ―.‖).
Reference Range: Click the drop down list of ―Reference‖, general, man, woman, child, infant and
Click of main interface shortcut in figure 2-5-4, ―Sample/Patient Information‖ will pop
up, as figure 3-2-1 shows:
Figure 3-2-1
Click ―Setting‖in figure 3-2-1to set the input items of patient information, displays as figure 3-2-2:
33
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-2-2
Remember: If , the last edited sample information will be remembered when editting
other information.
Click ―OK‖ when ―Default Value‖, ―Remember‖ and ―Ignore‖ have been set. The mouse will move
to the input box of sample No. in figure 3-2-1 to edit patient information.
(c)Reference:Double click the input box behind ―Reference‖ in figure 3-2-1,the following box
34
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-2-3
Input the corresponding memoni in the input box of ―System Code Selection‖, or click the line of
corresponding item(―Setting‖ of main menu→ ―Information‖ to set the commonly used ID).Click
―OK‖.The reference range of test items includes general, man, woman, child, newborn and 5 user
define.
(d)Sampling Time, deliver time: Select ―Current Time‖ or ― Empty‖ in the drop-down list of
―Sampling Time‖, ―Deliver Time‖ in figure 3-2-1. If ―Current Time‖ is selected, click the
(c) Sex: Double click the input box behind ―Reference‖ in figure 3-2-1, the following box will pop
up:
Figure 3-2-4
35
BF-6500 Automatic Hematology Analyzer User Manual
Input the corresponding memoni in the input box of ―System Code Selection‖, or click the line of
corresponding item(―Setting‖ of main menu→ ―Information‖ to set the commonly used ID).Click
―OK‖.
(d)Age:Input the patient age and double click the input box of age unit, the following box will
pop up.
Figure 3-2-5
Input the corresponding memoni in the input box of ―System Code Selection‖, or click the line of
corresponding item(―Setting‖ of main menu→ ―Information‖ to set the commonly used ID).Click
―OK‖.
(e) Department:Input the department name in department box, double click ―Department‖ input
Figure 3-2-6
Input the corresponding memoni in the input box of ―System Code Selection‖, or click the line of
corresponding item(―Setting‖ of main menu→ ―Information‖ to set the commonly used ID).Click
―OK‖.
36
BF-6500 Automatic Hematology Analyzer User Manual
(f)Bed No.:Click the input box of ―Bed No.‖in figure 3-2-1or input it directly.
(g)Deliver:Input it directly or double click the input box of ―Deliver‖, the following figure will pop
up:
Figure 3-2-7
Input the corresponding memoni in the input box of ―System Code Selection‖, or click the line of
corresponding item(―Setting‖ of main menu→ ―Information‖ to set the commonly used ID).Click
―OK‖.
Press enter to save the information when all the information have been finished.
In batch patient information input, input one patient information first and then copy it, modify part
In figure 3-2-1, select the number of the information need to be copied, the selected line will turn
Figure 3-2-8
37
BF-6500 Automatic Hematology Analyzer User Manual
Input the copy number in the input box behind ―Copy‖, and click ―OK‖, the selected information will
be copied.
The copied information need to be modified. In figure 3-2-1, select the number of the information
need to be modified, the selected line will turn into blue. The selected patient information will be
In figure 3-2-1, click the number of the information need to be deleted, click ―Delete‖, the
drop-down menu will pop up, select the item need to be deleted, as figure 3-2-9
Figure 3-2-9
3.2.5 Mask
In figure 3-2-1, select the number of the information need to be masked, the selected line will turn
into blue( in front of the number will be selected). Click ―Mask‖, the ―Mask Status‖ will be
Note:
The masked sample number(this record) will not be tested during test.
38
BF-6500 Automatic Hematology Analyzer User Manual
3.2.6 Query
Figure 3-2-10
Sample information can be queried according to ―Sample No.‖ ―Case No.‖ and ―Name‖.
3.3 Setting
Figure 3-3-1
39
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-2
Press ―OK‖ to save the changed date format. Press ―Exit‖ to exit the interface.
Note:
The changed date will be displayed in all positions with time(such as delivery time, sampling time,
etc.).
Figure 3-3-3
40
BF-6500 Automatic Hematology Analyzer User Manual
Click the drop-down menu behind ―Select Language‖, select the language, click ―Apply‖, the tip
―Save Success‖ will pop up, click ―OK‖ to finish the language setting. As figure 3-3-4 shows:
Figure 3-3-4
Figure 3-3-5
Only the password can be modified when the user login as common user. Click ―Modify
41
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-6
Input the new password and old password(the two passwords must be same), click ―OK‖ to finish
the modification.
Note:
User edition, addition and deletion can not be conducted when the user login as common user.
Single click ―Reagent Validity Setting‖ in figure 3-3-5, as figure 3-3-7 shows:
Figure 3-3-7
Click the drop down box of corresponding item, select the validity according to the reagent
instruction, press ―Apply‖, the tip ―Save Success‖ will pop up, click ―OK‖ to finish the setting. Click
Note:
The validity should be set again in the first time analyzer using or after reagent, diluent
3.3.1.5 QC Setting
Figure 3-3-8
Figure 3-3-9
Control method, calculation method and range of L-J/X-bar QC can be selected in figure 3-3-8.
43
BF-6500 Automatic Hematology Analyzer User Manual
Click ―Print Setting‖ in figure 3-3-9. Printer name, pageheader icon, title and format of sample and
Figure 3-3-10
(b)Report Default Icon: Report pageheader icon can be set. Click ―Select Picture‖, the dialog
box of picture path will pop up. Select the picture(picture size is 34*34, format can be BMP and
JPG).
Click ―Clear Picture‖ to clear the pageheader icon. The icon will not be displayed in report
preview.
(c)Sample:Headline, pagefooter and print report format can be input in this unit.
Headline:Delete the original headline and input new headline if headline need to be
changed.
Page Footer:The page footer is user-defined, e.g. ―This report is only responsible for the
delivered sample.‖
Print Report Format: Select the format in its drop-down list. Click ―Preview‖ to preview the
format.
44
BF-6500 Automatic Hematology Analyzer User Manual
(d) QC:The headline and print report format can be input in this unit.
Headline:Delete the original headline and input new headline to change it.
Print Report Format:Select the print format in its drop-down menu, and click ‖Preview‖ to
Input administrator username and password when login(figure2-5-2), click ―Setting‖, as figure
3-3-11 shows, the administrator can set the following items besides common user‘s :
Note:
Admin is system default administrator username. Default password is 1. The password can be
Figure 3-3-11
45
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-12
· Add User:
Figure 3-3-13
Input the username need to be added in the input box behind ―Username‖. Click to , which is in
front of ―Audit Permission‖ if the user needs audit permission, and click ―OK‖ to finish the user
adding.
Click ―OK‖ after username input if this user does not need audit permission.
Note:
The initial password of the new user is ―1‖, which can be modified after login.
The name of new user can not be empty or same as other username.
·Delete User:
Click the line of common user in figure 3-3-12, the selected line will turn into blue, the interface is
like 3-3-14:
46
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-14
Click ‖Delete‖ in above figure, the following prompt will pop up:
Figure 3-3-15
·Edit User:
Click the line of common user in figure 3-3-14, click ―Edit‖, the interface is like 3-3-16:
Figure 3-3-16
The audit permission of common user can be edited again. Select the radio in front of ―Audit
Permission‖ to give the permission to the user. Cancel it to cancel the permission.
· Modify Password:
47
BF-6500 Automatic Hematology Analyzer User Manual
The password of the administrator and common user can be modified by administrator.
Click the line needs to be modified in figure 3-3-14, click ―Modify Password‖, the interface is like
3-3-17:
Figure 3-3-17
Input old password, new password, press ―OK‖ to finish the operation.
shows:
(a)Local Setting:
Figure 3-3-18
The IP address and subnet mask, etc. of the computer will be displayed.
(b)Device Setting:
48
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-19
(c)LIS Setting:
Figure 3-3-20
IP and port of LIS computer can be set. ON/OFF status of auto-communication can be selected.
Figure 3-3-21
Single click the input box behind items and select the units. Click ―Apply‖ after input. Click ―OK‖
when ―Save Success‖ is prompted. Click ―Default Value‖ to restore the setting when the unit has
50
BF-6500 Automatic Hematology Analyzer User Manual
Table 3-1
Note:
51
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-22
Click the inputbox of upper and lower limit to input the values.
Click ―Apply‖ after input. Click ―OK‖ after the prompting of ―Saving succeed‖ to finish the saving.
The following box will pop up if the input lower limit is greater than the upper limit, or the input
Figure 3-3-23
Click ―Default Value‖ to restore the setting when the reference has been modified accidentally.
52
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-24
Figure 3-3-25
The saving path of ―Operation Log‖ and ―Error Log‖ can be set.
Note: The old records will be covered if the records reach the maximum saving number. Records
Click ―Information‖ in figure 3-3-25, and select ―Department Information‖ in the drop-down list of
53
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-26
· Add Department Information:Input the department name in the input box behind name. Input
the commonly used memoni in the input box behind ID, click ―Add‖.
· Delete Department Information : Select the items need to be deleted in the department
Select the ―Doctor Information‖ in the drop-down list of ―Information Type‖, as figure 3-3-27
shows:
Figure 3-3-27
54
BF-6500 Automatic Hematology Analyzer User Manual
· Add Doctor Information:Input the doctor name in the input box behind name. Input the
commonly used memoni in the input box behind ID, click ―Add‖.
· Delete Doctor Information:Select the items need to be deleted in the doctor information list,
click ―Delete‖.
(c)Reference Selection:
Select the ―Reference‖ in the drop down list of ―Information Type‖, as figure 3-3-28 shows:
Figure 3-3-28
‖User Define 2‖,‖User Define 3‖,‖User Define 4‖ and ‖User Define 5‖ in reference list. Only the
commonly used ID of reference can be modified here, but not the name or add other reference.
(d)Sex Selection:
Select ―Sex‖ in the drop down list of ―Information Type‖, as figure 3-3-29 shows:
55
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-29
Select ―Male‖ or ― Female‖ in the sex list.Only the commonly used sex ID can be changed here.
Select the ―Age‖ in the drop down list of ―Information Type‖, as figure 3-3-30 shows:
Figure 3-3-30
Select ―Years Old‖, ―Month‖, ―Day‖ or ―Hour‖ in the list of age unit. Only the commonly used age
Figure 3-3-31
(a) Backup:
Single click ―Backup‖ to backup the document into the folder. The default saving path is
―backup‖of software installation. The folder name is the current date+time+*.bak. Click ―Apply‖ for
saving.
The backup data can be used to recover the previous data if software can not be used for some
reason. Select the saving path of backup folder before recovery. And select the document
57
BF-6500 Automatic Hematology Analyzer User Manual
Figure 3-3-32
Figure 3-3-33
The prompt range of WBC, RBC/PLT abnormal alarm information can be set in this interface.
Click ―Apply‖ for saving after input, click ―Default Value‖ to recover, click ―Exit‖ to cancel the
interface.
58
BF-6500 Automatic Hematology Analyzer User Manual
This chapter describes the whole process of daily operation from start to shut down. Focusing on
the operation process of whole blood sample and pre-dilution (Peripheral Blood) sample testing.
Power On
Daily QC
Sample Preparation
Sample Analysis
Shutdown
Figure 4-1
2. Check pipeline
Check the pipeline connected with reagent, waste liquid, and also the grounding.
Check the connection between power plug and power socket of the instrument, and the
computer.
59
BF-6500 Automatic Hematology Analyzer User Manual
Check the connection between the computer and the barcode reader.
4.2 Daily QC
Note:
QC analysis should be conducted everyday before sample analyzing to ensure reliable result.
1.Use clean EDTA-K2 (1.5—2.2mg/mL blood)anti freezing vacuum tube to collect venous blood
Note:
Use clean EDTA-K2 anti freezing vacuum blood collecting tube, silicified glass/plastic tube
The sample that needs WBC differential or platelet counting should be stored at room
If 5 part of PLT, MCV, or WBC is not needed, the sample can be stored at 2 ℃ - 8 ℃ in
refrigerator for 24 hours. The stored sample should be used after placing at room
After a period of time of still placing, the sample should be mixed again before use.
· Click in shotcut area, the screen display as figure 4-3-1, 4-3-2 shows:
60
BF-6500 Automatic Hematology Analyzer User Manual
Figure 4-3-1
Figure 4-3-2
·Take a clean tube or centrifuge tube, place it under the sample probe, as figure 4-3-3 shows:
Figure 4-3-3
Press aspiration key, the diluent will be dispensed into the tube, the screen displays as figure
4-3-4 shows:
Figure 4-3-4
Ensure all the diluent(180uL)is added into the tube. Move the tube after the buzzer sound has
61
BF-6500 Automatic Hematology Analyzer User Manual
been heard.
· And put 20 uL peripheral blood into the tube follow the tube wall. Mix them.
Figure 4-3-5
· If there are some sample need to be diluted after the first one, the interface will be as figure
4-3-2, repeat the first process to finish other sample dilution. About 20 sample dilution can be
done continuously.
Note:
●The operator can also use pipette to aspirate 180uL diluent and dispense it into the tube follow
the tube wall. And add 20 uL peripheral blood into the tube follow the tube wall. Mix them.
● When ―pre-dilute‖ is selected, ―Attention! Dilute sample in ratio of 1:10‖ will be prompted in
sample information.
●After the mixing of peripheral blood and diluent, it should be placed for 3 minutes, and then
● The sample should be re-mixed after placing for a period of time before use.
●The stability of the result on pre-dilution mode should be evaluated according to their own
62
BF-6500 Automatic Hematology Analyzer User Manual
In main interface, if the current mode is ―Whole Blood‖, testing can be carried out directly. Or
Note:
The pipeline should be rinsed when ―Predilution‖ is switched into ―Whole Blood‖.
Figure 4-4-1
―CBC‖mode:conduct counting without WBC differential. The result include 14 parameters and
(d) Input the next sample No. in ―Sample No.‖ input box.
(e) Select the reference range in the drop down list of ―Reference Range‖, default as ―General‖.
Note:
63
BF-6500 Automatic Hematology Analyzer User Manual
The maximum digit of sample No. is 15. All number should not be ―0‖, otherwise, the software
Note:
Figure 4-4-2
In batch patient information input, input one patient information first and then copy the input
patient information, modify part of the information at last. Operations are same as the "3.2.2 Copy
64
BF-6500 Automatic Hematology Analyzer User Manual
Note:
● There should be a space between probe tip and test tube bottom in the process of sample
● Set the reference range of parameters in ―Setting‖ interface, otherwise, incorrect alarm will be
● The default reference range is ―Normal‖. The alarm prompted after test is according to the
Conduct whole blood sample testing according to the following steps in ―Counting Interface‖(As
Figure 4-4-3
(a)Check the working mode is ―Whole Blood Mode‖, the green indicator in the middle of the
(b)Place the sample under the sampling probe. Make sure the probe can aspirate the mixed
sample.
(c)Press ―Counting‖ on instrument panel to start the sample testing. The sampling probe
65
BF-6500 Automatic Hematology Analyzer User Manual
(d)When the buzzer is heard, the operator can remove the sample. The sampling probe injects
the sample into the counting pool. The instrument will carry out test automatically.
(e)The sampling probe reset after testing, and prepare for next testing. The test result will be
displayed in the chart review of the computer. Meanwhile, the next sample number will plus one
automatically.
(h)Click to enter into query interface, press ―Print‖ or ―Batch Print‖ to print the
report.
Note:
In the process of testing, the interface can be switched to ―Picture‖ ―Research Parameter‖ to
browse the pictures or check the study parameters. But no operation can be conducted. The
data will be saved into the data review after test automatically.
If block or air bubbles occurred, the instrument will set the relative parameter to be invalid.
The result will be inaccurate if the temperature exceeds the normal working temperature
Click ―Picture‖ in figure 4-4-3 to check the pictures of the test results, as figure 4-4-4 shows:
66
BF-6500 Automatic Hematology Analyzer User Manual
Figure 4-4-4
Histogram of RBC, PLT and WBC; histogram and scattergram of WBC 4 diff will be displayed.
Click ―Research Parameter‖ in figure 4-4-4 to check 10 research parameters, as figure 4-4-5
shows:
Figure 4-4-5
Check the working mode is ―Predilution Mode‖ in ―Counting‖ interface(as figure 4-4-3 shows),
Note:
(a)When the green indicator in the middle of the front upper cover is bright, place the diluted
sample under the sampling probe. Make sure the probe can aspirate the mixed sample.
(b)Press ―counting‖ on instrument panel to start the sample testing. The sampling probe
aspirate 80uL.
(c)When the buzzer is heard, the operator can remove the sample. The sampling probe injects
the sample into the counting pool. The instrument will carry out testing automatically.
(d)The sampling probe reset after testing,and prepare for next test. The test result will be
displayed in the pictures review of the screen. Meanwhile, the next sample number will plus one
automatically.
(h)Click to enter into query interface, press ―Print‖ or ―Batch Print‖ to print the
report.
Note:
In the process of testing, the interfaces can be reviewed. But no operation can be conducted.
The data will be saved into the data review after test automatically.
If block or air bubbles occurred, the instrument will set the relative parameter to be invalid.
The result will be inaccurate if the temperature exceeds the normal working temperature
If the result is marked with ―H‖or―L‖, that means the result exceeds the reference range.
68
BF-6500 Automatic Hematology Analyzer User Manual
If the result displayed as ―***‖,invalid result or range exceeding can be indicated. If WBC counting
result is smaller than 0.5*109/L or bigger than 999*109/L,or the pre-dilution result is smaller
than 2*109/L or bigger than 999*109/L, the system will not conduct WBC differential, the
4-1:
distribution distribution
RBC size is different. RBC size is different. RDW-SD>65or RDW-CV>20
Small RBC Smaller MCV MCV<70fL
Table 4-1
4.6 Sleeping
The analyzer enters into sleeping status after stop for 30 minutes, and the status information will
70
BF-6500 Automatic Hematology Analyzer User Manual
be displayed in failure area. Press liquid aspiration key to proceed with operation.
The analyzer will conduct automatic rinsing(sample flow parts) in each counting process to
Flow cell
When the analyzer prompt ―Aperture Clog‖, click ―Remove Clog‖ in ―Service-Maintenance‖
4.8 Shutdown
Note:
In order to ensure the stability and accurate test result of the instrument. Shutdown operation is
required after 24 hours continuous working. The operator must follow these steps to shut down
the instrument.
Shutdown includes mainframe shutdown and exit software.
4.8.1 Shutdown Mainframe
Figure 4-8-1
71
BF-6500 Automatic Hematology Analyzer User Manual
Click ―OK‖, above figure will pop up. After shutdown liquid aspirating, ―Shutdown the analyzer ‖
will pop up, the instrument will conduct pipe and counting pool rinsing and soaking.The following
Figure 4-8-2
Note:
·Click ,or select ―Menu‖→―Exit System‖,the following box will pop up:
Figure 4-8-3
Click ―OK‖ to exit the software to finish the whole shutdown process.
72
BF-6500 Automatic Hematology Analyzer User Manual
The instrument will store the result into database automatically after each test. The maximum
histogram). The operator can query the sample result, scattergram and histogram of the database.
The result will be covered if the storage number is more than 100000.
73
BF-6500 Automatic Hematology Analyzer User Manual
Click in main interface to enter into query interface, as figure 5-5-1 shows:
Figure 5-1-1
The result displayed in the base is in testing sequence(The last result is at the first place). The
If the screen can not display all result, press ‖Next‖ or ―Previous‖ to switch the pages. The position
of the current result and the total number of data base will be displayed in the form of ―Pos/Total‖
74
BF-6500 Automatic Hematology Analyzer User Manual
Figure 5-1-2
· Input the ―Start Record‖ and ―End Record‖ need to be queried in figure 5-1-2, click ―OK‖. The
Note:
The input sample position should within the range of sample lib, otherwise, ―Input range error,
·If ―Cancel Selection‖ is selected in drop-down list, the selection will be canceled.
5.2 Print
Click ―Print‖ in figure 5-1-1, if ―Preview or not before printing‖ is selected in ―Print Setting‖, as the
Figure 5-2-1
75
BF-6500 Automatic Hematology Analyzer User Manual
Figure 5-2-2
Select the test date of the report, and input the start and end sample No. If ―Audited‖ is selected
and ―Preview or not before printing‖ is selected, the report can be previewed, as the following
figure shows:
Figure 5-2-3
The selected & audited samples will be printed. If ―Audited‖ is not selected, all selected samples
will be printed.
5.3 Query
The data can be queried through different condition. Click ―Query‖ in figure 5-1-1, and select
76
BF-6500 Automatic Hematology Analyzer User Manual
Figure 5-3-1
Users can select different query conditions to query, including: sample No., barcode No., test
time, test mode, case No., name, gender, department, deliver doctor, auditor and tester. The
Input the sample No. in the box behind ―Sample No.‖ need to be queried in figure 5-3-1, if the
input sample No. is ―14‖, click ―Query‖, the sample results with No. ―14‖ will be displayed, as figure
5-3-2 shows:
77
BF-6500 Automatic Hematology Analyzer User Manual
Figure 5-3-2
Input the ID No. in the box behind ―Barcode‖ need to be queried in figure 5-3-1, click ―Query‖, the
In figure 5-3-1, select the start date and end date of the samples need to be queried in the
drop-down list behind test time, and select the corresponding ―Test Mode‖ and ―Analysis Mode‖,
Input the Case No. in the box behind ―Case No.‖ need to be queried in figure 5-3-1, click ―Query‖,
deliver doctor of the sample in the drop-down list behind ―Deliver Doctor‖ need to be queried in
Note:
If ―Perfect Match‖ is selected in above query, only perfect match results will be displayed. If
―Perfect Match‖ is not selected in above query, all results that can meet the input condition will
be displayed.
If sample with No. 11, 211, 311 exist, when ―Perfect Match‖is selected, the input query No. is 11,
only sample No.11 can be queried.11, 211, 311 will be displayed when ―Perfect Match‖ is not
selected.
5.4 CV Calculation
Users can select records among 3-500 to calculate the CV, Mean and SD. If the selected records
Figure 5-4-1
Click the No. in front of record(√), and click CV, as figure 5-4-2 shows:
79
BF-6500 Automatic Hematology Analyzer User Manual
Figure 5-4-2
The repeatability of the results can be checked. If the any parameter of the selected sample is
The user can conduct bulk audit in result query, click bulk audit in figure 5-1-1, as figure 5-5-1
shows:
Figure 5-5-1
Select the test date in the drop down list of ―Test Date‖, and input the start sample No. and end
80
BF-6500 Automatic Hematology Analyzer User Manual
sample No., click ―OK‖,the auditor information of audited sample result will be displayed behind
auditor.
5.6 Communication
Click ―LIS Comm.‖ in list review interface to conduct data transmission with LIS
system.
Figure 5-6-1
5.7 Delete
Click ―Delete‖ in figure 5-1-1, and select the records need to be deleted in its drop-down
Figure 5-7-1
5.8 Export
81
BF-6500 Automatic Hematology Analyzer User Manual
Click ―Export‖ in figure 5-3-2, select the data needs to be exported in its drop-down list ,
as the following figure shows:
Figure 5-8-1
Input the file name in the input box behind ‖File Name‖, and select the saving path, click ―Save‖,
Figure 5-8-2
Note:
User can check single record in chart form, in figure 5-1-1, double click the sample result line, as
82
BF-6500 Automatic Hematology Analyzer User Manual
Figure 5-9-1
The sample result position and the total number in sample database will be displayed in the lower
Error may occur after long-term using which may lead to unreliable result. Quality control (QC)
provides an effective way to detect error. Only by familiar with the QC theory and practical
In order to ensure the reliability of the result. The daily low, medium and high level control should
This instrument provides three QC method: L-J QC, X QC and X-B floating mean methods.
6.1 L-J QC
Under ―L-J‖ QC, the operator can carry out quality control of 24 parameters. The instrument
83
BF-6500 Automatic Hematology Analyzer User Manual
provides 12 QC documents in order to save quality control parameter and result. Each quality
control document can save up to 400 groups of quality control results. When the number of quality
control is more than 400, the new QC result will cover the old result.
Click ―QC‖ of main interface and select ―L-J/Xbar‖ in , as shown in figure 6-1-1.
Figure 6-1-1
84
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-1-2
Click the drop down list of document No. to select the document needs QC, the range is 1~12.
Input the corresponding Control lot No. in the drop-down list of ―Lot No.‖ according to Control
instruction.
(c)Validity Setting:
Click the drop down list of ―Validity‖ to input the validity according to the instruction.
Select QC level (High, Mid, Low)in drop down menu of ―Level‖. Each lot No. corresponds to one
level.
Click ―Save‖ when above input is completed. ―Save Success‖ will pop up, and click ―OK‖.
Note:
The input lot No., validity should be the validity marked on the instruction.
QC mode can be modified in ―Sample Information‖ of counting interface shortcut. For the Control
‖Test Mode‖ is “CBC+5DIFF”,if test mode is “CBC”,as shown “CBC mode can not count”when
85
BF-6500 Automatic Hematology Analyzer User Manual
counting.
If the display form of deviation limit or calculation method of deviation limit in pre-set value need to
Click ―Set‖ in main menu, select ―QC Setting‖ in its left, as figure 6-1-3 shows:
Figure 6-1-3
When‖ Absolute Value‖ calculation method is selected, the input deviation limit will be displayed in
the form of absolute value. ―Range‖ will use two times the standard deviation (2SD) or 3 times the
When―Percent‖calculation method is selected, the input deviation limit will be displayed in the
form of percent. ―Range‖will use 2 times the variation coefficient(2CV) or 3 times the variation
Click ―Apply‖ in figure 6-3, the preset value is obtained according to the set method, and it is taken
as the target value and deviation limit of current QC document. The corresponding position of the
(b)If abnormal QC point occurs, preset value can be obtained after abnormal point deleting. The
86
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-1-4
Click ―OK‖ of above figure, the screen will be switched into QC interface, as figure 6-1-5 shows:
Figure 6-1-5
· If the second point on the left is abnormal point, click it, the red cursorline is on the second point,
click ―Delete‖, the selected point will turn into ―Blue‖. Use the same method to delete the
abnormal points. The blue point will not be counted during calculation.
· Click ―Calculation‖ after the points are deleted, the screen will return to‖Setting‖ interface, and
· If the valid QC point number is less than 3, click ―Preset Value‖, figure 6-1-6 will pop up:
87
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-1-6
·If the above operation is found to be wrong, click the deleted point, and click ―Add‖, the blue point
Note:
Valid QC points are taken as reference and deviation limit when taking preset value.
6.1.2 QC Counting
Click QC count in figure 6-1-5 to enter QC counting interface, as shown in figure 6-1-7.
Figure 6-1-7
● QC Counting
Ensure the mode is ―Whole Blood‖.
Place the mixed Control under the sampling probe. Press ―Sample Aspiration‖ to start QC
counting.The sampling probe will aspirate 20ul Control automatically to conduct QC counting.
88
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-1-8
If the result is lower than the lower limit of the software or higher than the display range, the
Figure 6-1-9
● QC Data Query
The stored QC result can be queried through button Next, Previous under Position/Total Number
after QC counting.
Click QC Graph in figure 6-1-7 to enter review interface. As shown in figure 6-1-10:
89
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-1-10
Click Setting in ―QC Graph‖ interface to set the parameters or move the display position, as figure
6-1-11 shows:
Figure 6-1-11
Select the parameter name in figure 6-1-1, click ―OK‖, only the selected parameter will be saved.
90
BF-6500 Automatic Hematology Analyzer User Manual
If the parameters need to be moved, select the parameter need to be adjusted(the selected line is
blue), click ―Move Up‖, ‖Move Down‖ to move the selected parameter.
If QC counting number is less than 3, the right side of QC chart will not display the QC result.
The abscissa indicates the QC counting number. The vertical axis indicates QC counting result.
Vertical line marked for the same set of counting data. For each parameter, its QC chart can
display up to 31 points.
For each parameter, the three numbers on the left of the QC chart correspond to the three
boundaries of QC chart. From top to bottom, they represent the upper limit, target value and lower
limit.
For each parameter, the three number on the right of the QC chart represent Mean, SD and
CV% respectively.
―Green Point‖ indicates the QC result is within the range. ―Red Point‖ indicates the QC result is
(d)If the point is not within the range, conduct the following steps.
Conduct QC again if the above two points are normal. If abnormity remains, conduct QC counting
after calibration.
Contact DIRUI‘s after sales service department if abnormity remains after calibration.
Click Print in figure 6-1-10, if ―Preview before print‖ is selected in print setting, preview can be
91
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-1-12
6.1.4 QC List
Click ―QC List‖ in figure 6-1-10, and select the QC document No. in the drop down menu of
Figure 6-1-13
(a) Delete All:The operator can delete all QC results of current QC document.
92
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-1-14
Click ―Yes‖, the interface will display ―Delete Succeed‖, which indicates that all QC results are
deleted.
Select the QC data need to be deleted in figure 6-1-13, click ―Delete‖, as figure 6-1-14 shows,
click ―Yes‖, the interface displays ―Deleting Succeed‖, which indicates the selected QC results are
deleted.
LIS.
6.2 X QC
X QC:the mean value of two testing result will be one QC point in the QC chart.
6.3 X-B QC
X-B floating average method is proposed by Dr. BrianBull. Through monitoring stability of red
blood cell parameter, such as MCV, MCH, MCHC, etc. to monitor instrument performance. It
belongs to the quality control without control. Together with the quality control with control, they all
belong to instrument performance monitoring method which can reflect instrument performance
from different aspects. They are not mutually replaceable. The X-B quality control is
93
BF-6500 Automatic Hematology Analyzer User Manual
recommended when the daily sample is greater than 100 per day. This quality control method
requires the use of random sample, therefore, the classified sample is not suitable. It offers the
upper and lower limit to form a reference range. Observe the changing trend of result within the
reference range.
This instrument conduct X-B QC toward MCV、MCH、MCHC. The sample number in each group
can be set as 20-200. The sample is come from normal counting result of the instrument, without
Calculation is needed before testing, because X-B reference value is obtained through analyzing
6.3.1 QC Setting
(a)Click ―Setting‖ in main menu, and select ―QC Setting‖ in the left menu, open ‖X-B‖ interface, as
Figure 6-3-1
Sample No. selected for each X-B point can be 20-200, the recommanded number is 20.
Click ON under ―X-B QC ‖, click ―Apply‖, ―Saving succeed‖ will pop up, click ―OK‖.
(b)Click ―QC‖ in above menu, and select ―X-B‖ in , as figure 6-3-2 shows:
94
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-3-2
Target:Reference of different area may be different, so the sample number should reach certain
number(more than 500), the mean should be taken as target of X-B QC.
Note:
(c)Click ―Save‖ after setting, ―Saving Succeed‖ will pop up, click ―OK‖.
6.3.2 QC Counting
The operator can conduct X-B QC upon normal counting result. The system will conduct X-B QC
calculation after 20-200 samples(Set according to the sample number/group ). Each X-B QC
parameter will get a QC point. It will be stored in X-B QC chart and X-B QC list.
Click ―QC Graph‖ in figure 6-3-2 to enter into X-B QC graph review interface, as shown in figure
6-3-3.
95
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-3-3
The abscissa is the counting result number. The vertical axis is the result.
For each parameter, the three numbers on the left of the QC graph correspond to the three
boundaries of QC graph. From top to bottom, they represent the upper limit, target value and
lower limit.
For each parameter, the three number on the right of the QC grapch represent Mean, SD, and
CV% respectively.
―Green Point‖ indicates the QC result is within the range. ―Red Point‖ indicates the QC result is
(c)If the point is not within the range, conduct the following steps.
Contact DIRUI‘s after sales service department if abnormity remains after calibration.
The QC result can be checked through previous and next. Corresponding QC result
will be displayed under the parameter name. The position of the QC point and the total QC result
number will be displayed in the form of ―Position/Total Number‖ at the lower left side of the
interface.
Click QC list in figure 6-3-3 to enter the X-B QC list review interface, as shown in figure 6-3-4.
Figure 6-3-4
(a)Delete:Click the QC data behind corresponding No., the line will turn blue, click ―Delete‖, as
Figure 6-3-5
Click ―Yes‖ in above figure, ―Delete Succeed‖ will be displayed, the selected records will be
deleted.
97
BF-6500 Automatic Hematology Analyzer User Manual
Note:
Figure 6-3-6
Click ―Yes‖ in above figure, ―Delete Succeed‖ will be displayed, the all records will be deleted.
Figure 6-3-7
Input the document name in the input box behind ―File Name‖, and select the saving path, click
―Save‖.
(d)LIS Transmission:Click ―LIS Transmission‖, the data will be transmitted to LIS, as the following
figure shows:
98
BF-6500 Automatic Hematology Analyzer User Manual
Figure 6-3-8
Chapter 7 Calibration
Calibration of the instrument is to ensure the accuracy of the results. Calibration must be carried
The instrument calibration has been conducted in the factory. Calibration is still required in the
The instrument offers three calibration methods: manual calibration, calibrator calibration and
fresh blood calibration. In automatic calibration and fresh blood calibration, the relative calibration
will be conducted automatically by the instrument, and the calibration coefficient will be stored in
any problem.
· Check the instrument and the reagent to ensure enough reagent to finish whole calibration
process. If the reagent is finished in the process of calibration, calibration should be conducted
again.
99
BF-6500 Automatic Hematology Analyzer User Manual
· Conduct background testing. If the blank test value exceeds the background range, solution
should be seeked to ensure the result meet the requirement.
· The Control and reagent specified by Dirui company should be used. The corresponding
operation should refer to the Control and reagent instruction.
Figure 7-3-1
(b)Calibration Counting:
Put the prepared Calibrator under the sampling probe(No less than 0.5 mL) after reference value
input. Press ―Sample Aspiration‖, the sampling probe will aspirate 20uL Calibrator and start
100
BF-6500 Automatic Hematology Analyzer User Manual
Figure 7-3-2
After obtaining 3 or more counting results, the instrument will carry out CV and calibration
coefficient calculation and save the calibration coefficient result. Save after 5 times counting is
recommended. As the increase of calibration times, CV and calibration coefficient will be updated.
The following box will pop up if the results obtained beyond the scope.
Figure 7-3-3
Click ―OK‖ to close the box and clear the result of this counting. The result will be displayed on the
When saving the counting result after obtaining 3 or more counting result, if the calibration
coefficient of certain parameter is not within 75%-125%, click ―Save‖, the following box will pop
up:
Figure 7-3-4
101
BF-6500 Automatic Hematology Analyzer User Manual
Operator should check the reference value input, if the reference value input is correct, operator
If any abnormal result occurs, click ―Test Result‖ on the left after test, × will be marked, indicating
this result is not counted during CV, mean, calibration coefficient calculation. If this result deletion
is not needed after test, click ―Test Result‖ again(mark √ ), indicating this test result is valid.
Click ―Save‖ in figure 7-3-1 after 5 times counting, the following box will pop up.
Figure 7-3-5
Click ―OK‖ to save the calibration coefficient. It will be stored into calibration coefficient of ―Manual
Figure 7-3-6
Use EDTA-K2 (1.5—2.2mg/mL blood) antifreeze vacuum tube to collect venous blood sample.
Prepare 3-5 copies of normal fresh blood by using the above method.
102
BF-6500 Automatic Hematology Analyzer User Manual
Note: Fresh blood calibration need to be conducted under ―Whole Blood‖ and ―Pre-dilution‖
testing mode.
Click ―Fresh Blood Calib‖ in figure 7-3-1 to enter fresh blood calibration interface, as shown in
figure 7-4-1.
Figure 7-4-1
7.4.2.1 Whole Blood Calibration
(a)Select the number of fresh blood calibration in the drop down list behind ―No.‖ To reselect its
(b)Put the prepared fresh blood in the instrument to test three times, and calculate the mean
value. The mean value will be taken as reference value and input in figure 7-4-1.
(c)Put the fresh blood under sampling probe, and press ―Sample Aspiration‖, the sampling
probe will aspirate 20uL fresh blood and start calibration counting automatically.
103
BF-6500 Automatic Hematology Analyzer User Manual
Figure 7-4-2
(d)After obtaining 3 or more counting results, the instrument will carry out CV and calibration
The following box will pop up if the results obtained beyond the scope.
Figure 7-4-3
Click ―OK‖ to close the box and clear the result of this counting. The result will be displayed on the
When saving the counting result after obtaining 3 or more counting result, if the calibration
coefficient of certain parameter is not within 75%-125%, click ―Save‖, the following box will pop
up:
104
BF-6500 Automatic Hematology Analyzer User Manual
Figure 7-4-4
Operator should check the reference value input, if the reference value input is correct, operator
If any abnormal result occurs, click ―Test Result‖ on the left after test, × will be marked, indicating
this result is not counted during CV, mean, calibration coefficient calculation. If this result deletion
is not needed after test, click ―Test Result‖ again(mark √ ), indicating this test result is valid.
Click ―Save‖ in figure 7-4-1 after 5 times counting, the following box will pop up.
Figure 7-4-5
Click ―OK‖ to save the calibration coefficient. It will be stored into calibration coefficient of ―Manual
Calibration‖ interface.
The calibration coefficient displayed on the screen is the calibration coefficient saved after
Click ―Manual Calibration‖ in figure 7-4-1 to enter manual calibration interface. As shown in figure
7-5-1.
105
BF-6500 Automatic Hematology Analyzer User Manual
Figure 7-5-1
(a) Click the corresponding line under ―Calibration Coefficient‖ when the calibration
coefficient needs adjustment, and input the calibration coefficient, click ―Save‖, ―Save
Click ―Calib History‖ in figure 7-5-1 to enter calibration log interface, as shown in figure 7-6-1.
Figure 7-6-1
106
BF-6500 Automatic Hematology Analyzer User Manual
The date, mode, calibration method and detail test value of latest 30 times calibration will be
displayed. If calibration counting is more than 30 times, the previous calibration result will be
covered.
Query: Operator can query the calibration log according to log date, username, calibration mode
Chapter 8 Service
In order to ensure the normal running of the instrument, routine maintenance is required. The
instrument will prompt the user to conduct maintenance after testing a certain number of sample
or a continuous working period. The service menu in the instrument offers routine maintenance
methods and failure solutions, but users should make their own maintenance plan according to
daily sample number, operating environment, running time, etc. to reduce the impact of various
factors and ensure the safe, stable and effective running of the instrument.
Note:
● Improper maintenance may damage the instrument. The manual must be followed in
maintenance.
● Contact Dirui's after-sales service department for unclear answers of the manual.
107
BF-6500 Automatic Hematology Analyzer User Manual
● Daily:
If the instrument is on for 24 hours, daily ―WBC Pool Rinsing", ―RBC Pool Rinsing‖, ―DIFF Pool
● Every week:
If normal shutdown operation is conducted every day, " Detergent Soaking" (WBC Pool, RBC
● Every month:
conducted(Service-Maintenance-Rinsing).
water, conduct "Rinse Pipeline" of "Maintenance", and stop the distilled water, conduct "Empty
● When the instrument prompt "clog" failure, press ―Remove‖ to conduct manual remove or
● Carry out priming if it has not been used for a long time.
● The instrument will prompt ―Soaking with Detergent‖ if the set counting number of software has
been conducted.
●The reliable results can be obtained under the condition of normal working environment and
state.
System status is used to display the current status. Basic status, system version can be checked
108
BF-6500 Automatic Hematology Analyzer User Manual
in this interface.
Click service, and click in front of System Version, and select Software, the screen displays
as figure 8-2-1:
Figure 8-2-1
Figure 8-2-2
Version number of control board, collection board, temperature control board, sampler control
board and mechanical version can be checked by using the same method.
Select ―Instrument Information‖ in its drop down menu, as figure 8-2-3 shows:
Figure 8-2-3
Note:
Figure 8-2-4
Status explanation:
(a)Temperature:Real-time temperature of reaction pool, working environment and laser, and also
110
BF-6500 Automatic Hematology Analyzer User Manual
(c)Voltage:Display 55V, 5V, voltage of WBC aperture, RBC aperture and normal range.
(d)Current:Display the normal range and power adjust valve of laser current.
Click in front of ―Detect‖ in figure 8-2-4, and select ―Mechanical Detect‖, as shown in figure
8-3-1:
Figure 8-3-1
The following motors detection has been set in this program: Y axis motor, X axis motor, diluent
motor, lyse motor, whole blood aspiration motor and test motor. Click ―Test‖, the result will be
Valve failure will lead to abnormal work of the instrument. Therefore, value detection is an
Click the No. of the valve(No. of figure 8-4 indicate valves), the analyzer will conduct valve
detection automatically. Valve ON sound indicates normal valve. The valves with grey No. can not
be detected.
Note:
Valve detection and voltage detection can not be conducted at the same time.
In order to protect the normal and accurate running of the analyzer, the software provides a
Note:
Carry out background test after diluent, lyse replacement. Sample test can be conducted when
the background result is within normal range. The following range is the normal range of
background test:
a)RBC≤0.05×1012/L
b)WBC≤0.5×109/L
c)HGB≤2g/L
d) PLT≤10×109/L
Click in front of ―Maintain‖ in figure 8-3-1, and select ―Replace/Prime‖, as figure 8-4-1
shows:
112
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-1
Click Replace FDO when bubble exits in FDO pipeline, or FDO reagent is polluted, or FDO
Figure 8-4-2
Figure 8-4-3
113
BF-6500 Automatic Hematology Analyzer User Manual
After replacing FDO lyse, the following box will pop up:
Figure 8-4-4
Click Replace FDT when bubble exits in FDT pipeline, or FDT reagent is polluted, or FDT
reagent has been used up. The confirm interface will pop up. Click OK,pop up progress
bar ‖Replacing FDT About 3 mins…‖, After replacing FDT lyse, ―Operation finished‖ box will pop
Click Replace SLS when bubble exits in SLS pipeline, or SLS reagent is polluted, or SLS reagent
has been used up. The confirm interface will pop up. Click OK,pop up progress bar ‖Replacing
SLS About 3 mins…‖, After replacing SLS lyse, ―Operation finished‖ box will pop up, click OK to
(d)Diluent Replacement:
Click Replace Diluent when bubble exits in Diluent pipeline, or Diluent is polluted, or Diluent has
been used up. The confirm interface will pop up. Click OK,pop up progress bar ‖Replacing
Diluent About 5 mins…‖, After replacing Diluent, ―Operation finished‖ box will pop up, click OK
(e)FDO Priming:
114
BF-6500 Automatic Hematology Analyzer User Manual
Click Prime FDO when FDO Lyse being emptied. The following box will pop up:
Figure 8-4-5
Figure 8-4-6
Figure 8-4-7
Note: In the "fault information area", clear the corresponding error message, the instrument
automatically conduct FDO lyse priming when the software prompt FDO insufficiency.
Click ―Prime FDT‖ after FDT is emptied. And the operation should be conducted as the prompts.
Or: In the "fault information area", clear the corresponding error message, the instrument
automatically conduct FDT lyse priming when the software prompt FDT insufficiency.
Click ―Prime SLS‖ after SLS is emptied. And the operation should be conducted as the prompts.
Or: In the "fault information area", clear the corresponding error message, the instrument
115
BF-6500 Automatic Hematology Analyzer User Manual
automatically conduct SLS lyse priming when the software prompt SLS insufficiency.
Click ―Diluent Priming‖ after Diluent is emptied. And the operation should be conducted as the
prompts.
Or: In the "fault information area", clear the corresponding error message, the instrument
automatically conduct Diluent priming when the software prompt Diluent insufficiency.
Click ―Detergent Priming‖ after Detergent is emptied. And the operation should be conducted as
the prompts.
Or: In the "fault information area", clear the corresponding error message, the instrument
automatically conduct Detergent priming when the software prompt Detergent insufficiency.
8.4.2 Rinsing
Figure 8-4-8
(a) Conduct WBC pool rinsing when the background test value of WBC or HGB parameter is
116
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-9
Figure 8-4-10
―Operation Complete‖ will pop up after WBC pool rinsing, click OK to finish the whole process.
Click ―Cancel‖ in figure 8-4-9 to cancel WBC pool rinsing.
(b) Conduct RBC pool rinsing when the background test value of RBC or PLT parameter is
abnormal. Click Rinse RBC in figure 8-4-8. Confirm box will pop up, click OK,progress bar will
pop up. ―Operation Complete‖ will pop up after RBC pool rinsing, click OK to finish the whole
process.
(c) Conduct DIFF pool rinsing when the WBC-diff parameter is abnormal. Click Rinse DIFF in
figure 8-4-8. Confirm box will pop up, click OK,progress bar will pop up. ―Operation Complete‖ will
pop up after DIFF pool rinsing, click OK to finish the whole process.
(d) In order to avoid sampling component contamination, swab rinsing should be conducted after
Click ―Rinse Swab‖ in figure 8-4-8. The confirm box will pop up, click OK, as figure 8-4-11 shows:
117
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-11
Place the detergent used for probe rinsing under sampling probe and press sample aspiration key.
Progress bar will pop up. ―Operation Complete‖ will pop up after swab rinsing, click OK to finish
(e) Bubble may exists in flow cell when magnified cell mass exists in scattergram and the
background testing value of WBC parameter is higher than normal value. ―Sheath Pool Rinsing‖
Click Rinse Sheath Flow Pool in figure 8-4-8. Confirm box will pop up, click OK,progress bar will
pop up. ―Operation Complete‖ will pop up after rinsing, click OK to finish the whole process.
(f)If bubble exists in sample aspiration pump or the test values are lower after reagent
Click ―Debubble Sample Pump‖ in figure 8-4-8. Confirm box will pop up, click OK,progress bar
will pop up. ―Operation Completed‖ will pop up after debubble, click OK to finish the whole
process.
8.4.3 Maintenance
118
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-12
8.4.3.1 Emptying
In order to avoid liquid spill during pipeline maintenance, empty should be conducted first. Taken
(a)Click Empty WC to empty the corresponding bottle and the pipeline. The following box will pop
up.
Figure 8-4-13
Click OK in above figure to conduct empty WC. Progress bar will pop up. ―Operation Complete‖
After emptying waste pool 1, return to the interface shown in Figure 8-19.
(b) Click ―Empty Pipeline‖ in figure 8-4-12 when the analyzer will not be used for 1-2 days. Place
119
BF-6500 Automatic Hematology Analyzer User Manual
(c)Replace the reagent with distilled water when the analyzer will not be used for more than 1
week. Click ―Package Clean Pipe‖ in figure 8-4-12. Place the analyzer at a clean position after
8.4.3.2 Aperture
Aperture clog remove, zap and backflush should be conducted in order to clear the debris in
aperture.
Figure 8-4-14
Click ―OK‖ in above figure, to clean the aperture through high-voltage direct current. The progress
Click ―Backflush‖, the confirm box will pop up, click ―OK‖ to flush it. The progress bar will pop up
meanwhile.
Click ―Clog Remove‖, the confirm box will pop up, click ―OK‖ to zap and flush the aperture. The
RBC pool soaking should be conducted every other month, the operation is as follow:
120
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-15
Figure 8-4-16
Place the prepared probe detergent under sample probe, and press aspiration key. The detergent
Figure 8-4-17
Figure 8-4-18
Click ―Next‖ in figure 8-4-18 to empty RBC, as the following figure shows:
121
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-19
Operation has been finished after emptying, as the following figure shows:
Figure 8-4-20
WBC pool soaking should be conducted every other week, the operation is as follow:
Figure 8-4-21
Click OK in above figure, as 8-4-16 shows. Place the prepared probe detergent under sample
probe, and press aspiration key. The detergent will be dispensed into WBC pool, as the following
figure shows:
122
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-22
Click ―Next‖ in figure 8-4-18 to empty WBC, as the following figure shows:
Figure 8-4-23
DIFF pool soaking should be conducted every other week, the operation is as follow:
Figure 8-4-24
Click OK in above figure. The DIFF pool will be emptied. As figure 8-4-25, 8-4-26 shows:
Figure 8-4-25
123
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-26
Aspirate 1ml shutdown liquid with injector, and add it into DIFF pool, as figure 8-4-27 shows:
Figure 8-4-27
Click ―OK‖ in figure 8-4-26 after liquid adding. As figure 8-4-28 shows:
Figure 8-4-28
Figure 8-4-29 will pop up after soaking, indicates whole process is finished.
Figure 8-4-29
124
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-30
Manual barcode information input can be conducted if the external barcode scanner is not
connected.
Click barcode input box in figure 8-4-30 to input the information according to the reagent package.
Left times: Scanning a barcode information for each, the remaining number will change
Connect the barcode reader( connect the data wire of barcode reader with computer) , click the
input box behind ―Barcode‖ to turn into available status, scan the barcode outside the package
box with barcode reader, screen prompts "× × × register ok" and the barcode.
125
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-4-31
Input the information after checking if above condition occur. Contact DIRUI or his distributor for
continuous failure.
The following box will pop up if the barcode has been used:
Figure 8-4-32
Note:
Do NOT stare at the scanning beam during analyzer running to avoid eye injury.
Counting for 100 samples/day, sample aspiration pump, test pump and lyse pump(including SLS
pump, FDO pump and FDT pump)should be replaced once in 21 months. Pump used for diluent
126
BF-6500 Automatic Hematology Analyzer User Manual
Sample
Pump
Figure 8-5-1
① ④
② ⑤
③
Figure 8-5-2
①SLS Syringe Pump ②FDO Syringe Pump ③FDT Syringe Pump ④Diluent Pump
⑤Test Pump
Click ―Log‖ in main interface, and select ―Error Log‖ in its drop-down list , as
127
BF-6500 Automatic Hematology Analyzer User Manual
Figure 8-6-1
The operator can also select ―Operation Log‖ and ―Failure Log‖ in its drop-down list for checking.
128
BF-6500 Automatic Hematology Analyzer User Manual
The instrument should be protected from water, severe vibration and squeeze. Handling, loading
129
BF-6500 Automatic Hematology Analyzer User Manual
This chapter describes various types of possible failure, the reason of the failure, and the
solutions.
Note:
Test the result in case of failure may cause inaccurate result. If alarm is prompt in the process of
10.1 Overview
See the following relative failure solution if failure occur during working process.
Alarm information will be prompted in failure information area if failure occurs. Click the failure,
Figure 10-1-1
Name of failure information will be displayed according to the order of the failure.
130
BF-6500 Automatic Hematology Analyzer User Manual
131
BF-6500 Automatic Hematology Analyzer User Manual
Room over- 2. Please make sure the ambient temperature is within normal range [15,
temperature
30]℃.
3.Replace the temperature sensor if failure remains.
1.Click ―Clear‖ to solve the problem.
2. Please make sure the ambient temperature is within normal range [15,
Room low-temperature
30]℃.
132
BF-6500 Automatic Hematology Analyzer User Manual
133
BF-6500 Automatic Hematology Analyzer User Manual
134
BF-6500 Automatic Hematology Analyzer User Manual
Appendix A
Delimiter Value
Field Separator |
Discreteness Separator ^
Sub-Discreteness Separator &
Repeat Separator ~
ESC \
MSH-message head
PID-patient information
PV1-case history
MSH –message head: this message segment is required item,includes HL7 message basic
information, message separator value, message type and message coding method and so on, it
is each HL7 message‘s first message segment.
Information Example:
MSH|^~\&|BF-6500|1234567890|||20100419104618||ORU^R01|361|P^S|2.4|||||CHN|UNICODE<
cr>
Serial HL7 Advised
Field Name Length Explanation Example
NO. Length
Include the first field separator
Field after message segment, used for
1 1 1 |
Separator regulating other message field
separator value
Include discreteness separator,
Coded
2 4 4 repeat separator, ESC, ^~\&
Character
sub-discreteness separator
Send terminal apply program
Send
3 7 180 BF-6500
Program value:BF-6500
136
BF-6500 Automatic Hematology Analyzer User Manual
value: ORU^R01(Sample)
Value:S-sample、LJ-LJ /X barQC,
XB-XB QC)
PID:
PID–patient information: this information segment is optional, used for patient sample
transmission, include patient case history number, name, age, gender etc.
Message Example
137
BF-6500 Automatic Hematology Analyzer User Manual
transmission, include patient department, bed NO., deliver doctor, examiner and so on.
MEssage example:
HL7
Serial Field
Length Advice Explanation Example
NO. Name
Length
Pointed form as :department^^bed no.
3 patient 80 80 ^^clinic 235689
position
Deliver deliver doctor, character string
7 50 250 doctor Wang
doctor
8 Examiner 50 250 examiner, character string Zhang San
report information, include sample serial number, and scan No., tube rack No., deliver time and
so on.
Message example :
OBR||23|31C3F010230DFB03|0001^Count
Results||20071207080000|20071207160000|||||| |20071207083000||||2311|322<cr>
HL7
Serial Field
Length Advice Explanation Example
NO. Name
Length
Sample Sample number in testing
2 Serial 16 22 Document No. in LJ/X QC 23
Number
Barcode ID in sample testing
3 Scan No. 32 22 31C3F010230DFB03
Lot No. in LJ/X QC
Data Service ID symbol, used for sign
4 Service 200 200 on different count result type. 0001^Count Results
Type Idiographic value check the
138
BF-6500 Automatic Hematology Analyzer User Manual
conversion)
140
BF-6500 Automatic Hematology Analyzer User Manual
percentage of 2017^V_MON_p
2017 V_MON_p NM
Monocytes
The number of red blood 2018^V_RBC
2018 V_RBC NM
cells
2019 V_HGB Hemoglobin NM 2019^V_HGB
141
BF-6500 Automatic Hematology Analyzer User Manual
Enumeration Type
Data Item Value
0- CBC 1-
test mode
CBC+DIFF
0-open-whole blood 1-open-
analysis mode
pre-dilution 2-auto-whole blood
0- normal
1- M
2- F
3- Child
4- baby
reference
5- custom 1
6- custom 2
7- custom 3
8- custom 4
9- custom 5
0- high
L-J/X QC level 1- medium
2- low
1. Patient Sample
<SB> MSH|^~\&|BF-6500||||20110310150421||ORU^R01|8|P^S|2.4|||||CHN|UTF-8
<cr>
PID||1234567890|||Wang Sanqiang|||Male<cr>
142
BF-6500 Automatic Hematology Analyzer User Manual
OBX|1|IS|2001^MODE||0||||||F<cr>
OBX|2|IS|2002^MODE_EX||1||||||F<cr>
OBX|3|IS|2003^Ref||0||||||F<cr>
OBX|4|IS|2004^Age||17|age|||||F<cr>
OBX|5|ST|2005^Note||note position||||||F<cr>
OBX|6|NM|2007^V_WBC||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|7|NM|2008^V_BAS_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|8|NM|2009^V_NEU_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|9|NM|2010^V_EOS_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|10|NM|2011^V_LYM_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|11|NM|2012^V_MON_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|12|NM|2013^V_BAS_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|13|NM|2014^V_NEU_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|14|NM|2015^V_EOS_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|15|NM|2016^V_LYM_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|16|NM|2017^V_MON_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|17|NM|2018^V_RBC||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|18|NM|2019^V_HGB||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|19|NM|2020^V_MCV||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|20|NM|2021^V_MCH||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|21|NM|2022^V_MCHC||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|22|NM|2023^V_RDW_CV||4.63|10*9/L|11.00-12.00|L|||F<cr>
143
BF-6500 Automatic Hematology Analyzer User Manual
OBX|23|NM|2024^V_RDW_SD||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|24|NM|2025^V_HCT||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|25|NM|2026^V_PLT||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|26|NM|2027^V_MPV||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|27|NM|2028^V_PDW||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|28|NM|2029^V_PCT||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|29|NM|2030^V_P_LCR||4.63|10*9/L|11.00-12.00|L|||F<cr>
BASE64 code……||||||F<cr>
BASE64 code……||||||F<cr>
BASE64 code……||||||F<cr>
BASE64 code……||||||F<cr>
BASE64 code……||||||F<cr>
<EB><CR>
2.L-J/X QC
<SB>MSH|^~\&|BF-6500||||20110311091016||OUL^R21||P^LJ|2.4|||||CHN|TUF-8<cr
>
144
BF-6500 Automatic Hematology Analyzer User Manual
|||||0|0<cr>
OBX|1|IS|2006^Level||0||||||F<cr>
OBX|2|NM|2007^V_WBC||4.63||||||F<cr>
OBX|3|NM|2008^V_BAS_c||4.63||||||F<cr>
OBX|4|NM|2009^V_NEU_c||4.63||||||F<cr>
OBX|5|NM|2010^V_EOS_c||4.63||||||F<cr>
OBX|6|NM|2011^V_LYM_c||4.63||||||F<cr>
OBX|7|NM|2012^V_MON_c||4.63||||||F<cr>
OBX|8|NM|2013^V_BAS_p||4.63||||||F<cr>
OBX|9|NM|2014^V_NEU_p||4.63||||||F<cr>
OBX|10|NM|2015^V_EOS_p||4.63||||||F<cr>
OBX|11|NM|2016^V_LYM_p||4.63||||||F<cr>
OBX|12|NM|2017^V_MON_p||4.63||||||F<cr>
OBX|13|NM|2018^V_RBC||4.63||||||F<cr>
OBX|14|NM|2019^V_HGB||4.63||||||F<cr>
OBX|15|NM|2020^V_MCV||4.63||||||F<cr>
OBX|16|NM|2021^V_MCH||4.63||||||F<cr>
OBX|17|NM|2022^V_MCHC||4.63||||||F<cr>
OBX|18|NM|2023^V_RDW_CV||4.63||||||F<cr>
OBX|19|NM|2024^V_RDW_SD||4.63||||||F<cr>
OBX|20|NM|2025^V_HCT||4.63||||||F<cr>
OBX|21|NM|2026^V_PLT||4.63||||||F<cr>
145
BF-6500 Automatic Hematology Analyzer User Manual
OBX|22|NM|2027^V_MPV||4.63||||||F<cr>
OBX|23|NM|2028^V_PDW||4.63||||||F<cr>
OBX|24|NM|2029^V_PCT||4.63||||||F<cr>
OBX|25|NM|2030^V_P_LCR||4.63||||||F<cr>
code……||||||F<cr>
code……||||||F<cr>
code……||||||F<cr>
<EB><CR>
3.X-B QC
<SB>MSH|^~\&| BF-6500||||20110311091040||OUL^R21||P^XB|2.4|||||CHN|
UTF-8<cr>
OBR||||1004^ XB QC|||20071207160000||||||||||||<cr>
OBX|1|NM|2079^XB_Num||20||||||F<cr>
OBX|2|NM|2073^m_MCV_R||12.204||||||F<cr>
OBX|3|NM|2074^m_MCH_R||0.258||||||F<cr>
OBX|4|NM|2075^m_MCHC_R||12.445||||||F<cr>
OBX|5|NM|2076^m_MCV_L||45.859||||||F<cr>
OBX|6|NM|2077^m_MCH_L||1.258||||||F<cr>
OBX|7|NM|2078^m_MCHC_L||2.36||||||F<cr>
OBX|8|NM|2020^V_MCV||4.63||||||F<cr>
OBX|9|NM|2021^V_MCH||4.63||||||F<cr>
146
BF-6500 Automatic Hematology Analyzer User Manual
OBX|10|NM|2022^V_MCHC||4.63||||||F<cr>
<EB><CR>
Appendix B
The report can be modified or created through report designer setting to design the ideal report
In initial use, please save the template before modification. For the design of the template is
The following describes the specific function and use of the report designer.
B.1 Report Designer Object
Report designer is in the toolbar(left), a total of three objects:
147
BF-6500 Automatic Hematology Analyzer User Manual
B.1.1 “TextBox”
Rectangular box which contain multiple lines of text. Type, color and width of frame, font attribute,
text alignment and font direction (vertical or horizontal) can be set. Use ‖Text‖ and ‖Frame‖ tools
Figure B.1
Text box object includes: text, variables, data fields or any combination of these. Font formatting
·TextBox Modification:
Click on the left of Report Edit Designer, click the left mouse after rectangular icon appears,
Figure B.2
Clipboard operation;
Word wrap;
148
BF-6500 Automatic Hematology Analyzer User Manual
Cancel button;
OK button
Note: data included in the database includes patient information, sample test information, the
Picture can be inserted in the report. The format of the picture is BMP、WMF、ICO.
Click on the left of Report Edit Designer, click the left mouse after rectangular icon appears,
Figure B.3
Click ―Select..‖ in figure B.3, and click ―OK‖, the picture can be inserted into the report.
B.1.3 “Line”
The horizontal or vertical line can be inserted in the report. In separate statement of the report,
straight-line makes it easy to be read. The line thickness and color can be adjusted by using the
drawing toolbar.
Click , drag the mouse in the current page, the cursor will turn into a pencil to draw a straight
149
BF-6500 Automatic Hematology Analyzer User Manual
line. Click the mouse to begin the line, release it when the line is finished. The line can be
modified.
·Line modification: select the corresponding button in ―frame toolbar‖ to modify the line.
(a) “Standard”toolbar
(b)“Format”toolbar
(c)“Frame”toolbar
(d)“Alignment”toolbar
Set the page option for the current page of the report, select ―File|Page Setting‖in designer
Figure B.4
Select the paper size of current printer in the drop down list of paper size.
If the current printing support self-defined paper format, select "Self-defining", and then input the
Note: not all printer drive or printer support self-defining paper format (e.g. printer drive―HP
LaserJet 6L‖does not support 76*127mm size; printer drive―HP LaserJet 4L‖does not support all
self-defining size)
Figure B.5
B.2.3 Margins
Figure B.6
If the "Extend to the printer" option is selected, the page form of designer will not display the
border area. All regions of the page will be printed correctly. But the size is different in different
printing.
If this option is canceled, and all margin set to be 0, then the margin will be automatically set to
the selected printer's maximum print area. When designed report switch from one printer to other
151
BF-6500 Automatic Hematology Analyzer User Manual
printer (the printable area of ink jet printer is smaller than that of stylus printer), this function very
useful.
If the margin set to be non 0, margins will be reflected in a page from of the designer (marked with
gray lines). If you use a dot matrix printer, first preview whether the print content is within the print
area (some stylus printer will not print the content beyond printing scope, and other printer
prompts the beyond print scope). In this case, set the margins manually.
B.2.4 Other
Figure B.7
Set the number of columns and column spacing according to the page width. If the "print to front
page" option is selected, it allows print the remaining area in the new page.
B.3 Users create their own report sheet
The report template we provided, list all data of patient test report and L-J QC report in detail.
Open all selected objects of the report before create report sheet. The report template is in
software installation directory, \ Print \ Sample: is patient sample report template ; \ Print \ QC: is
QC report template.
Concrete action : Open select report edit Select All New Report Page SetupPaste.
In this new report, the position, font and the letter of text can be modified.
B.3.1 Title:
Set the sample test report and QC report in "System Setting". For example: × × × hospital, the title
is × × × hospital test report. The ―Test Report‖ can be set hereby. It can be modified into "Blood
test report "," LJ report ", etc. The title can be modified into static text.
152
BF-6500 Automatic Hematology Analyzer User Manual
B.3.2 Paper:
Users also can modify the size of print paper. If the user use inkjet or laser printer and A4 paper,
set paper A4. If the user use stylus printer (such as: epson 300K, 1600K, etc.), and use 80
column printing paper, set the paper to be self-defining. If a 80 column paper need to print three
reports, set the paper to be 9.34(length). If a 80 paper need to print two reports, set the paper to
Click the object need to be modified with mouse (points around the selected object will appear),
press Shift, more than one objects can be selected. Press Ctrl, and move the mouse meanwhile,
· Modify the letter of the static text: select the text need to be modified, and double click the
textbox, input the modified letter in the corresponding box, and press ―Enter‖.
· Modify the data field: If it is used to display certain data, only one textbox need to be added to
appropriate position, double-click the text box, input the data value, refer to the existing template
· Move object: select the object (more than one can be selected), then press the four buttons
around "Move" to move the object, arrow keys on the keyboard can also be used.
· Change size: select the object, press the up and down buttons of "high" or "width" to increase
· Change the font: Select the object, and then select the font size or bold, italic and so on.
· Undo: error happened during modification, undo operation for one or more times, the report will
153
BF-6500 Automatic Hematology Analyzer User Manual
· Save: press the "Save" button after all operation is completed. Note: If it is patient sample report,
please store in the software installation directory \ Print \ Sample folder; save quality control report
· Use: open the system setting in data management software, select the report in print setting.
Appendix C
Product Warranty
Dear customer:
Thank you for purchasing Automatic Hematology Analyzer of our company. We can
Address: 95, Yunhe Street, New & High Tech. Development Zone, Changchun, China
Fax :0431-85172581
Email :dirui@dirui.com.cn
Website: http://www.dirui.com.cn
Appendix D
Product Description
Belong to blood analyze system in clinic counting instrument (6840), type II in management type.
● Diluent
● BF-FDT
● BF-SLS-Ⅰ
● BF-FDO
distribution
156
BF-6500 Automatic Hematology Analyzer User Manual
Content
157
BF-6500 Automatic Hematology Analyzer User Manual
Statement
Dirui Co., LTD.is responsible for the security, reliability and capability of the product under the
following circumstance:
professionals.
The manufacturer reserves the right to make changes without prior notice.
2011/6
158