You are on page 1of 11

Bioelectromagnetics

Design and Dosimetric Analysis of an Exposure


Facility for Investigating Possible E¡ects of
2.45 GHz Wi‐Fi Signals on Human Sleep
Gernot Schmid ,1* Rene Hirtl,1 Ana Bueno‐Lopez,2 Hans Dorn,2
Torsten Eggert,2 and Heidi Danker‐Hopfe 2
1
Seibersdorf Laboratories, Seibersdorf, Austria
2
Competence Center of Sleep Medicine, Charité‐Universitaetsmedizin,
Berlin, Germany
A new head exposure system for double‐blind provocation studies investigating possible effects of
2.45 GHz Wi‐Fi exposure on human sleep was developed and dosimetrically analyzed. The
exposure system includes six simultaneously radiating directional antennas arranged along a circle
(radius 0.6 m) around the test subject's head, and enables a virtually uniform head exposure, i.e.
without any preferred direction of incidence, during sleep. The system is fully computer‐controlled
and applies a real wireless local area network (WLAN) signal representing different transmission
patterns as expected in real WLAN scenarios, i.e. phases of “beacon only” as well as phases of
different data transmission rates. Sham and verum are applied in a double‐blind crossover study
design and all relevant exposure data, i.e. forward and reverse power at all six antenna inputs, are
continuously recorded for quality control. For a total antenna input power (sum of all antennas) of
220 mW, typical specific absorption rate (SAR) in cortical brain regions is approximately
1–2 mW/kg (mass average SAR over respective brain region), which can be seen as a realistic
worst‐case exposure level in real WLAN scenarios. Taking into account variations of head positions
during the experiments, the resulting exposure of different brain regions may deviate from the given
average SAR levels up to 10 dB. Peak spatial 10 g average SAR in all brain and all head tissues is
between 1.5–3.5 and 10.4–25 mW/kg, respectively. Bioelectromagnetics. © 2020 Bioelectromag-
netics Society.
Keywords: Wi‐Fi; WLAN; EEG; sleep study; human study

INTRODUCTION 385 MHz [Danker‐Hopfe et al., 2019]. In most of these


studies signal characteristics and intensities typical for
Possible effects of radiofrequency (RF) electro-
using a handset close to the head were used, i.e. the
magnetic fields (EMF) on sleep have been investigated
resulting exposure in terms of peak spatial 10 g averaged
in numerous studies in the past. Such studies are of
specific absorption rate (psSAR10g) inside the test
particular interest because human sleep is commonly
subjects’ heads was typically in the range between 0.1
seen as a sensitive indicator for the impact of various
and 10 W/kg. Moreover, exposure of the head was
environmental factors on central nervous processes
applied either laterally or frontally or from the top in the
[Basner and McGuire, 2018; Harding et al., 2019].
However, so far RF EMF‐related sleep studies have
mainly focused on signals emitted by cellular mobile Grant sponsor: FSM—Swiss Research Foundation for Electricity
communication systems, such as the global system for and Mobile Communication.
mobile communication in frequency ranges around 900
and 1,800 MHz [Borbely et al., 1999; Huber et al., 2000; Conflicts of interest: None.
Loughran et al., 2005, 2012; Hinrichs et al., 2005; Hung
*Correspondence to: Gernot Schmid, Seibersdorf Laboratories,
et al., 2007; Regel et al., 2007; Arnetz et al., 2007; A‐2444 Seibersdorf, Austria. E‐mail: gernot.schmid@seibersdorf-
Danker‐Hopfe et al., 2010, 2011, 2016, 2019; Lowden laboratories.at
et al., 2011; Schmid et al., 2012a,b; Lustenberger
et al., 2013, 2015], the universal mobile telecommunica- Received for review 30 September 2019; Accepted 12 February
tions system in the frequency range between 1,900 and 2020
2,000 MHz [Danker‐Hopfe et al., 2011, 2016; Nakatani‐ DOI:10.1002/bem.22256
Enomoto et al., 2013; Lowden et al., 2019], and the Published online 2020 in Wiley Online Library
terrestrial trunked radio in the frequency range around (wileyonlinelibrary.com).

© 2020 Bioelectromagnetics Society


2 Schmid et al.

mentioned studies, i.e. there was always a predominant the UE to synchronize to the WLAN. In WLAN in the
direction of the incoming radiation, and hence, only a 2.45 GHz frequency band there is no transmit power
part of the brain could be considered exposed during the control, and maximum transmit power is limited to
experiments. A relatively small fraction of RF‐related 100 mW equivalent isotropic radiated power (EIRP).
sleep studies were designed as field studies, i.e. the As a consequence of the above, the time averaged
experiments took place at the residences of the test RF power Pavg emitted by an AP can vary between
subjects (e.g. Leitgeb et al., 2008; Danker‐Hopfe approximately 1% (beacon transmission only, see
et al., 2010), using RF fields emitted from a real or Fig. 1a) and approximately 90% (heavy and prolonged
experimental base station, respectively, as source of DL transmission, see Fig. 1b) of the burst power PBurst.
exposure, resulting in significantly less psSAR10g inside A Pavg of a single station (AP or UE) of substantially
the head. more than 90% of PBurst is not possible due to the
Interestingly, so far no study has been available timing characteristics of the WLAN technology
concerning possible effects of the RF EMF emitted by (remaining time is needed to acknowledge packets
wireless local area network (WLAN) signals on from the receiving station and for timing and
human sleep, although this technology has been synchronization purposes within the network). As
tremendously deployed in recent years not only modulation scheme WLAN transmitters operating in
in the public but also in the private environment. It the 2.45 GHz band use either direct sequence spread
can be expected that the majority of households and spectrum (DSSS) or orthogonal frequency division
areas in the public domain (e.g. hotels, airports, multiplex (OFDM), both result in stochastic signal
schools, universities, train stations, etc.) are presently characteristics during the radiated RF bursts.
equipped with this technology for internet access and In order to address the question of possible effects
similar purposes. WLAN technology has been stan- of WLAN exposure on human sleep, a corresponding
dardized in IEEE 802.11 standard series, and devices study was created by the Swiss Research Foundation for
certified to comply with IEEE 802.11 standards are Electricity and Mobile Communication (FSM). In the
also commonly known as Wi‐Fi devices. Starting study, subjective (total sleep time, sleep onset latency,
more than 20 years ago several different IEEE wake after sleep onset, and number of awakening) and
801.11x standards have been published, specifying a objective information on sleep were collected. From
variety of different WLAN transmission methods in polysomnography, 24 parameters describing sleep macro-
different frequency bands. Presently, WLAN ac- structure were extracted, including those 14 variables that
cording to IEEE 802.11‐2007, with operating fre- are mandatory for describing patients according to the
quency bands around 2.45 GHz, are most common in standards of the American Academy of Sleep Medicine
Europe. In such networks, various kinds of user [Iber et al., 2007], in which the analysis of microstructure
equipment (UE; e.g. mobile phones, computers, etc.) comprises arousals, spindles, and power spectra for 19
can connect to the WLAN via an access point (AP). electrode sites.
All the UE connected to an AP share the same This paper describes the design and dosimetric
transmission channel (of typically 20 MHz bandwidth) analysis of a new head exposure system for sleeping
in time domain by a carrier sense multiple access/ subjects specially developed for this study.
collision avoidance (CSMA/CA) access scheme, i.e.
UE connected to an AP (operating at the same WLAN
frequency channel) cannot transmit RF simulta- MATERIALS AND METHODS
neously, but only consecutively. A specific timing
algorithm and maximum length of data packets ensure Design Goals for the Exposure Setup
that the probability of getting access to the channel is The basic requirement for the exposure system was
approximately the same for all devices in the network. to enable RF exposure of the brain at realistic worst‐case
Moreover, both transmission from the UE to the AP levels as they can be expected in real WLAN exposure
(uplink, UL) and transmission from the AP to the UE scenarios. Moreover, exposure should not be focused on a
(downlink, DL) take place in the same frequency specific brain region but should rather be distributed over
band. Consequently, WLAN devices (AP as well as the cortex in the sense that there should be no preferred
UE) are transmitting short RF bursts during their direction of incoming radiation. The resulting exposure
operation and the burst lengths and burst repetition should be well‐defined and reproducible for sleeping
rates are highly dependent on the actual amount and subjects, and intra‐ and inter‐subject variability of
direction of data (UL or DL) transmitted in the exposure, due to variations of head orientation during
network. In case of no data traffic at all, only the AP sleep, should be kept low (no particular measures for
transmits a short beacon signal, every 100 ms, to allow restricting head movements during sleep would be
Bioelectromagnetics
2.45 GHz Wi‐Fi Exposure Setup for Sleep Studies 3

Fig. 1. Time domain characteristics of a typical wireless local area network (WLAN) signal
according to IEEE 802.11‐2007 during “beacon only” transmission (a) and maximum data
transmission rate (b), measured at the radiofrequency (RF) output of the access point.

applied). The applied exposure signal should be in the time domain (FDTD)‐based full wave solver of the
2.45 GHz Wi‐Fi frequency band and should represent SEMCAD X V14.8.6 simulation platform (Schmid &
both “beacon only” transmission and other realistic time Partner Engineering, Zurich, Switzerland) and the male
domain transmission patterns. anatomical body model Duke of the virtual family
Exposure and Sham condition must be applied [Christ et al., 2010]. In these simulations the head of
double‐blind in a randomized crossover study design, the body model (uniform spatial resolution 1 mm) was
and applied exposure data must be continuously exposed by a plane wave with varying angles of
recorded during the experiments for quality control. incidence (varied in steps of 15° around the head) and
In order to avoid any noise or vibrations vertical as well as horizontal polarization (Fig. 2a). The
perceptible by the subjects, all active components of dielectric properties of body tissues were assigned
the exposure facility (signal generation, amplifier, according to an online resource provided by the IT'IS
switches, control unit, etc.) had to be located outside Foundation [Hasgall et al., 2015].
the sleeping room. Finally, the entire exposure On the basis of these simulations, the SAR inside
apparatus must be suitable for a clinical environment the different brain regions in dependence on the angle of
(sleep laboratory). incidence was analyzed (Fig. 2b). The considered brain
regions were distinguished according to the stereotactic
brain atlas system of Talairach and Tournoux [1988].
Multi‐antenna Approach From this analysis a number of six antennas, positioned
To address the requirement of virtually uniform every 60° along a circle of radius 0.6 m around the head,
incoming radiation around the head, the concept of were found to be a reasonable compromise (Fig. 3). As
multiple simultaneously radiating directional antennas antennas, commercial WLAN patch antennas (Model
was chosen. In order to determine a suitable number of Digitus, Gain 10 dBi; Assmann Electronics, Lüdensc-
antennas, a series of simplified numerical computations heid, Germany) with 10 dBi gain and a 3 dB beamwidth
were carried out in first step, using the finite difference of approximately 75° were chosen (Figs. 4 and 7).

Bioelectromagnetics
4 Schmid et al.

Fig. 2. Numerical computations with 2.45 GHz plane wave exposure with different angle of
incidence (varied in steps of 15°) and horizontal and vertical E‐field polarization (a).
Analysis of brain region specific absorption was carried out in order to find a reasonable
number of antennas for the exposure setup. Only example data of some brain regions in
right hemisphere are shown in the figure (b).

Electrical Design characteristics, i.e. WLAN standard 802.11b, frequency


A schematic block diagram of the exposure channel, and transmit power of this AP was controlled
facility is shown in Figure 5. A commercial WLAN by software. The AP was connected to the system
AP (Model WRT54GL v1.1; Linksys, Irvine, CA) was computer (with control and recording software) via
used as primary signal source. The main transmission ethernet. In order to use real WLAN signals for

Fig. 3. Arrangement of the six antennas around the head in a carrier frame at the head end
of the bed. Left: numerical model; center: as installed at the sleep laboratory (antennas
hidden behind a cotton cloth); right: detail of antenna mounting (cotton cloth removed).
Bioelectromagnetics
2.45 GHz Wi‐Fi Exposure Setup for Sleep Studies 5

Fig. 4. A 2.45 GHz patch antenna as used (left), with cover removed (center), and
corresponding numerical model (right).

exposure of the test subjects, a short distance wireless the downlink antenna connector of the AP was
link between the AP and a Raspberry PI client (Model decoupled and fed to a 10 W linear power amplifier
3b; Raspberry PI Foundation, Cambridge, UK) placed with 40 dB gain (Model RF23002600‐10; RFPA,
immediately next to the AP was automatically Artigues pres Bordeaux, France). The microwave
established after system start, and a controlled data signal from the output of the power amplifier was
transmission from the system computer to the client via then fed through a power divider (Model PD1060;
the AP was initiated. Part of the microwave signal from INSTOCK Wireless Components, Boonton, NJ) to the

Fig. 5. Schematic block diagram of the exposure facility.


Bioelectromagnetics
6 Schmid et al.

six patch antennas via bidirectional couplers (Model varying burst lengths and burst repetition times, as they
CS06‐05‐436/10; Pulsar Microwave, Clifton, NJ) appear in real WLAN operation during data transmission,
enabling simultaneous forward and reverse power were included in the exposure signal. For this reason the
measurement on all six antenna inputs, using diode exposure signal was composed of periodically repeated
detectors (Model DM 211; S‐Team Lab, Bratislava, 6‐min intervals. Each of the 6‐min intervals were
Slovakia). The insertion loss, amplitude balance, and subdivided into six 1‐min timeslots, in which different
phase balance of the power divider in the frequency data transmission conditions were created (controlled by
range of interest were <1.1, <0.4 dB, and <6°, the system software). In the first 1‐min timeslot “beacon
respectively. Insertion loss and coupling directivity of only” transmission took place, while in each of the
the directional couplers were <0.2 and >22 dB, remaining five 1‐min timeslots the signal corresponded to
respectively. Flexible double‐shielded coaxial cables the continuous transmission of data files of a particular
of type RG‐400/U (attenuation 0.8 dB/m) were used as file size. The file size of the transmitted data files
cables between the power divider and directional increased from timeslot 2 to timeslot 6. Figure 6 shows
couplers immediately next to the antennas. the corresponding time course of the antenna input power
over a 24‐min time interval.
Mechanical Design
The six antennas were mounted in a carrier frame Control and Recording Software
made of white acrylic glass, which could be placed at the The control and recording software, developed
head end of the bed (Fig. 3). The selected antenna under LabView (Version 2010; National Instruments,
arrangement provided a minimum distance of approxi- Austin, TX), was installed at the system computer and
mately 40–50 cm between the subject's head and the provided an easy‐to‐understand graphical user interface
closest part of the exposure facility, and therefore avoided (GUI), which allowed the study instructor to create/
significant impairment of the subjects when lying in the recall identifiers for the test subjects and to start,
bed. In order to avoid visual impairment and temptation to interrupt/continue, and stop test sessions. After starting a
touch the antennas, the inner side of the carrier frame was test session, the appropriate exposure condition was
covered by a thin cotton cloth. The directional couplers automatically applied in a double‐blind manner, con-
and diode detectors for forward and reverse power trolled by the software, i.e. the SHAM/VERUM switch
measurement were mounted directly behind the antennas was electrically set to the appropriate status by the
inside the carrier frame. All microwave antenna feeding software without the possibility of acoustically or
cables (from the output of the power divider to the visually identifying the status of the switch. The
directional couplers) were guided within the carrier frame. appropriate condition (SHAM or VERUM) was auto-
The main microwave feed line originating from the power matically and pseudorandomly selected by the control
amplifier output was fed via the computer‐controlled software based on a balanced randomization scheme.
SHAM/VERUM switch into the shielded sleep laboratory Peak and RMS values (of 300 ms intervals) of forward
room to the input of the power divider (located below the and reverse microwave power were continuously (every
bed). The 12 analog signals delivered from the diode 3 s) measured and stored in the system computer, in an
detectors were fed via coaxial lines (guided inside the encoded form, during each experimental session.
carrier frame) to a shielded connector box (placed below
the bed), and from there via a shielded 12 twisted pair Dosimetry and Exposure Uncertainty Analysis
cable to A/D conversion electronics (2 pieces of model
Detailed RF dosimetry and exposure uncertainty
USB‐201; Measurement Computing, Norton, MA) con-
estimation was carried out based on FDTD computations
nected via USB‐interface to the system computer. The
and experimental methods. All numerical computations
WLAN AP, Raspberry PI, power amplifier, SHAM/
were performed using the SEMCAD X simulation
VERUM switch, and A/D conversion electronics were
platform (Schmid & Partner Engineering). In the first
housed inside a 19″ rack, installed together with the
step, a numerical model of the used patch antenna was
system computer, at the study instructor's place outside
developed and validated by free space measurements at
the experimental room.
2.45 GHz in an anechoic chamber (Fig. 7). Second, the
impact of the wooden bed frame, mattress, and pillow on
T|me Domain Characteristics of Applied Exposure the signals radiated from the two antennas located below
Signal the bed was analyzed by measurements of incident
According to the study design, phases with “beacon field strength in the volume of 20 × 20 × 25 cm3 above
only” transmissions (i.e. approximately 80–90 µs long the pillow where the head of test subject was located. For
bursts, repeated every 100 ms) as well as phases with the measurements, a broadband electric field meter
Bioelectromagnetics
2.45 GHz Wi‐Fi Exposure Setup for Sleep Studies 7

Fig. 6. Radiofrequency (RF) forward power measured at the antenna input during a 24‐min
interval, i.e. four periods of the 6‐min signal sequence, starting with a 1‐min interval of “beacon
only” followed by another five 1‐min time intervals with stepwise changed data transmission rate.
Interestingly, the burst rms during “beacon only” is approximately 0.5 dB less than the bursts rms
during data transmission, which obviously is a characteristic (unspecified) of the access
point used.

(Model EMR 300 with probe type 18C; Narda Safety and uncertainty analysis were carried out using the anatomical
Test Solutions, Pfullingen, Germany) was used. Measure- body models Duke, Ella, and Louis [Christ et al., 2010;
ments were taken repeatedly during separate transmission IT'IS Foundation, 2012] with tissue properties according
of each antenna at constant net antenna input power at to Hasgall et al. [2015], taking into account the variations
grid steps of 5 cm in the volume indicated above. of head position during sleep and impact of the
In the next step, a validation of the entire antenna electroencephalography (EEG) electrodes and cables
arrangement was carried out by SAR measurements present during exposure. During all computations, the
(Fig. 8b) and corresponding numerical computation computational grid resolution was kept constant at 1 mm
(Fig. 8a) using the homogeneous SAM phantom (tissue inside the anatomical model and the SAM head phantom,
simulating liquid with εr = 39.0, σ = 1.80 S/m). For the respectively.
measurements, both the whole carrier frame (including
antennas, couplers, power detectors, and cabling) and bed
were flipped by 90° (head end of bed downwards) in Suppression of Potential Interference of Applied RF
order to enable insertion of the SAR probe into the SAM Signal With EEG Measurements
phantom via the neck (Fig. 8b). Finally, computational When EEG measurements are carried out
brain region specific exposure analysis and corresponding during exposure of the head to RF signals with a

Fig. 7. Results of the antenna model validation in free space (1 m distance).


Bioelectromagnetics
8 Schmid et al.

Fig. 8. Numerical model (a) and the corresponding experimental setup (b) with
homogeneous SAM head phantom used for validation of the exposure system. The
comparison of measured and computed specific absorption rate (SAR) inside the Sam
head phantom shows reasonable agreement within 1.8 dB (c, d).

non‐constant envelope, significant electromagnetic (LF) components of the RF signal envelope are
interference (EMI) on the EEG recording may occur. demodulated into a lower frequency band, and the
This effect, even when purely passive EEG elec- recorded EEG signal spectrum may be significantly
trodes are used, can be explained by (unwanted) corrupted by the RF exposure, and may therefore
demodulation of the RF signal (coupled into the EEG also break the double blinding.
electrode leads) at nonlinear components of the EEG Therefore, utmost care had to be taken to
amplifier stages. Once this occurs, the low frequency strictly avoid infiltration of the RF signal at any

Fig. 9. Parameters and corresponding ranges considered for brain region specific
exposure and uncertainty analysis. (a): considered variations in lateral and longitudinal
head position and axial rotation; (b): considered electroencefalography electrodes on the
scalp as arranged during the experiments.
Bioelectromagnetics
2.45 GHz Wi‐Fi Exposure Setup for Sleep Studies 9

TABLE 1. Result of Brain Region Specific Absorption Analysis Taking Into Account Uncertainties Due to Variations of Head
Position and Rotation as Well as Electroencephalography Electrodes Present on the Scalp

Mass Averaged SAR [mW/kg]


During Burst @ PBurst = 220 mW
Talairach Level (Sum of All Antennas)

Level 1 Level 2 Level 3 Level 4 Level 5 Mean Max. Min.


Left cerebrum Occipital lobe Inferior occipital gyrus Gray matter Brodmann area 17 2.06 3.23 0.46
Left cerebrum Frontal lobe Middle frontal gyrus Gray matter * 1.99 4.44 0.27
Left cerebrum Occipital lobe Inferior occipital gyrus * * 1.78 4.51 0.66
Left cerebrum Occipital lobe Superior occipital gyrus * * 1.71 5.36 0.41
Left cerebrum Occipital lobe Fusiform gyrus Gray matter Brodmann area 18 1.63 3.37 0.45
Left cerebrum Occipital lobe Cuneus Gray matter * 1.62 2.86 0.23
Left cerebrum Occipital lobe Lingual gyrus Gray matter Brodmann area 17 1.61 2.51 0.16
Left cerebrum Temporal lobe Angular gyrus * * 1.58 3.75 0.21
Left cerebrum Occipital lobe Middle occipital gyrus Gray matter Brodmann area 18 1.47 3.37 0.57
Left cerebrum Occipital lobe Middle occipital gyrus * * 1.45 2.91 0.61
Left cerebrum Occipital lobe Inferior occipital gyrus Gray matter Brodmann area 18 1.45 4.61 0.32
Left cerebrum Frontal lobe Inferior frontal gyrus Gray matter * 1.43 4.15 0.24
Left cerebrum Frontal lobe Sub‐gyral Gray matter Brodmann area 10 1.42 4.36 0.15
Left cerebrum Occipital lobe Superior occipital gyrus Gray matter Brodmann area 19 1.42 4.65 0.26
Left cerebrum Occipital lobe Inferior temporal gyrus Gray matter Brodmann area 19 1.42 2.59 0.22
Left cerebrum Temporal lobe Fusiform gyrus Gray matter Brodmann area 19 1.41 2.27 0.11
Left cerebrum Frontal lobe Inferior frontal gyrus Gray matter Brodmann area 10 1.40 4.35 0.14
Left cerebrum Temporal lobe Angular gyrus Gray matter Brodmann area 39 1.34 3.51 0.29
Left cerebrum Temporal lobe Middle temporal gyrus Gray matter Brodmann area 19 1.33 4.53 0.08
Left cerebrum Parietal lobe Precuneus Gray matter Brodmann area 19 1.31 3.38 0.37
Left cerebrum Occipital lobe Superior occipital gyrus White matter * 1.29 4.55 0.14
Left cerebrum Occipital lobe Inferior occipital gyrus White matter * 1.28 3.72 0.40
Left cerebrum Frontal lobe Inferior frontal gyrus Gray matter Brodmann area 46 1.28 2.58 0.16
Left cerebrum Occipital lobe Inferior temporal gyrus Gray matter Brodmann area 37 1.26 2.45 0.17
Left cerebrum Frontal lobe Middle frontal gyrus Gray matter Brodmann area 46 1.26 2.33 0.13
Left cerebrum Temporal lobe Inferior temporal gyrus Gray matter Brodmann area 19 1.25 2.40 0.20
Left cerebrum Occipital lobe Middle occipital gyrus Gray matter Brodmann area 19 1.21 2.80 0.36
Left cerebrum Frontal lobe Sub‐gyral Gray matter * 1.18 3.23 0.06
Left cerebrum Parietal lobe Inferior parietal lobule Gray matter Brodmann area 39 1.17 2.54 0.33
Left cerebrum Occipital lobe Cuneus Gray matter Brodmann area 19 1.16 1.93 0.40
psSAR10g (all brain tissues) 2.06 3.50 1.36
psSAR10g (all head tissues) 15.7 25.0 9.46
Data for the 30 brain regions of left hemisphere with highest exposure are shown. *All subregions contained within the next lower level.

stage of the EEG signal amplification by appropriate RESULTS AND DISCUSSION


filtering techniques. In this particular study, this was
done by double‐shielding the analogue electronics Validation of the Numerical Model of the Exposure
of the biosignal recording system, and by feeding all Facility
bio signal lines into this double‐shielded box via The numerical antenna model showed reason-
fifth order RF‐suppressing feedthrough filters. In able agreement (within 2 dB) in main lobe direction
order to test the effectiveness of these measures, the between simulations and measurements in free space
EEG recordings using a passive head dummy (Fig. 7). As the measurements showed systematically
(watermelon) were carried out once with the RF (approximately 1–2 dB) lower field strength in main
signal switched on and once with the RF signal lobe direction compared with the simulations, this
switched off. Comparison of the recorded EEG deviation was corrected by a corresponding correction
spectrum obtained during “RF ON” and “RF OFF” factor of 1.5 dB in the power budget.
did not show any significant difference, proving that Concerning the impact of the wooden bed frame,
no relevant EMI from the exposure setup had to be mattress, and pillow on the signal radiated from the two
expected in the EEG recordings. antennas located below the bed, comparison of volume

Bioelectromagnetics
10 Schmid et al.

TABLE 2. Summary of Most Important Dosimetric System Parameters


Frequency range 2.40–2.48 GHz
Antenna system 6 patch antennas (10 dBi) arranged along a circle (r = 0.6 m) around the test
subjects head
Signal/modulation Real WLAN‐Signal (IEEE 802.11b)
Duty cycle (Pavg/PBurst) Systematically (computer‐controlled) varied between 1% and 90%, cyclic
repetition every 6 minutes
Total input power (sum of all antennas) 220 mW
maxSAR10g in head tissues <25 mW/kg (during burst), <6.4 mW/kg (6 minutes average)
maxSAR10g in brain tissues <3.5 mW/kg (during burst), <0.9 mW/kg (6 minutes average)
Mean mass averaged SAR in cortical brain regions typ. >1–2 mW/kg (during burst), typ. >0.3–0.5 mW/kg (6 minutes average)
SAR variations (max/mean)a typ. 2–5 dB
SAR variations (mean/min)a typ. 2–10 dB
Minimum mass averaged SAR in cortical brain regions typ. >0.1 mW/kg (during burst), typ. >0.03–0.05 mW/kg (6 minutes average)
SAR = specific absorption rate.
a
Mainly caused by variations of head position and rotation.

average and maximum field strength in the head region represents the overall uncertainty when taking into
indicated systematically lower incident field strengths account all the abovementioned uncertainty contributors.
from the two antennas located below the bed, which was The data show that SAR in the brain region may vary
finally compensated by appropriately adjusting the considerably (up to 10 dB around the mean) during the
antenna input power such that the deviation in incident experiments, mainly caused by changes in the head
field strength over all antennas was less than 0.5 dB. This position and rotation. The impact of the EEG electrodes
adjustment of antenna input power for the two antennas is comparably small (typ. <4 dB). Minimum absorption
located below the bed was achieved by shorter cables in cortical brain regions is >0.1 mW/kg at a total antenna
between the power divider and these two antennas, input power PBurst of 220 mW. The psSAR10g in all
compared with the other antenna cables (1 m shorter, brain and all head tissues varied between 1.4–3.5 and
corresponding to 0.8 dB less attenuation). 9.5–25 mW/kg, respectively.
The comparison of measured and computed SAR Table 2 gives a summary of the most important
inside the SAM head phantom shows reasonable dosimetric system parameters of the developed exposure
agreement within 1.8 dB (Fig. 8c and d). facility.
Although the described exposure setup was
designed for a particular transmission scheme (i.e.
Exposure Uncertainty Analysis considering a specific timing for data and beacon‐only
Based on the validated numerical model of the transmission), it can easily be adopted to other RF
exposure facility, numerical computations with the exposure schemes in the same frequency range. In such a
anatomical body models Duke, Ella, and Louis case, all given SAR values can be linearly scaled with
were carried out to quantify brain region specific exposure antenna input power.
and its uncertainty due to variations of head position and
head rotation, considering the head positions and
orientations depicted in Figure 9a, i.e. head rotations up
CONCLUSION
to ±90° in steps of 30° and axial and left/right shifts up to
±150 mm in steps of 50 mm. The indicated variation A newly developed computer‐controlled exposure
ranges are considered realistic, because movements/ facility for double‐blind provocation studies investigating
rotations over wider ranges are not possible due to the possible effects of 2.45 GHz Wi‐Fi exposure on human
“natural” restriction given by the EEG cables. Moreover, sleep has been described and dosimetrically analyzed. Six
the impact of the EEG electrodes on the test subjects’ simultaneously radiating antennas around the head
scalps was analyzed (Fig. 9b). enabled virtually uniform head exposure, i.e. without
Table 1 summarizes the results for the 30 brain any preferred direction of incidence. The time domain
regions of highest absorption (data shown for left characteristics of the exposure signal were representing
hemisphere only; corresponding values in right hemi- both idle phases (“beacon only” transmission) and phases
sphere are typically within 3 dB) in terms of mass of different data transmission rates. A total RF input
averaged SAR in the respective brain region for a total power of 220 mW yielded a typical SAR in cortical brain
antenna input power PBurst of 220 mW (sum of all regions of approximately 1–2 mW/kg (mass average SAR
antennas). The range MAX‐MIN indicated in Table 1 over respective brain region). Variations of head positions
Bioelectromagnetics
2.45 GHz Wi‐Fi Exposure Setup for Sleep Studies 11

during the experiments could lead to deviations from the Huber R, Graf T, Cote KA, Wittmann L, Gallmann E, Matter D,
indicated SAR values of up to 10 dB. Schuderer J, Kuster N, Borbely AA, Achermann P. 2000.
Exposure to pulsed high‐frequency electromagnetic field
during waking affects human sleep EEG. Neuroreport
11:3321–3325.
ACKNOWLEDGMENTS Hung CS, Anderson C, Horne JA, McEvoy P. 2007. Mobile phone
'talk‐mode' signal delays EEG‐determined sleep onset.
The authors gratefully acknowledge FSM— Neurosci Lett 421:82–86.
Swiss Research Foundation for Electricity and Iber C, Ancoli‐Israel S, Chesson AL, Jr, Quan SF. 2007. The
Mobile Communication—for funding this work. AASM Manual for the Scoring of Sleep and Associated
Moreover, thanks goes to Dr. Jürg Fröhlich, Institute Events: Rules, Terminology and Technical Specifications.
for Electromagnetic Fields, ETH Zürich, for valuable Westchester, IL: American Academy of Sleep Medicine.
IT'IS Foundation. 2012. The virtual population. https://itis.swiss/
discussions during the project design phase. virtual‐population/virtual‐population/overview/ [Last accessed
20 March 2019].
Leitgeb N, Schröttner J, Czech R, Kerbl R. 2008. EMF‐protection
REFERENCES sleep study near mobile phone base stations. Somnologie
(Somnology) 12:234–243.
Arnetz BB, Akerstedt T, Hillert L, Lowden A, Kuster N, Wiholm C. Loughran SP, Wood AW, Barton JM, Croft RJ, Thompson B,
2007. The effects of 884 MHz GSM wireless communication Stough C. 2005. The effect of electromagnetic fields
signals on self‐reported symptom and sleep (EEG)—An emitted by mobile phones on human sleep. Neuroreport
experimental provocation study. PIERS Online 3:1148–1150. 16:1973–1976.
Basner M, McGuire S. 2018. WHO environmental noise guidelines for Loughran SP, McKenzie RJ, Jackson ML, Howard ME, Croft RJ.
the european region: A systematic review on environmental 2012. Individual differences in the effects of mobile phone
noise and effects on sleep. Int J Environ Res Public Health exposure on human sleep: Rethinking the problem.
15:519. Bioelectromagnetics 33:86–93.
Borbely AA, Huber R, Graf T, Fuchs B, Gallmann E, Achermann Lowden A, Akerstedt T, Ingre M, Wiholm C, Hillert L, Kuster N,
P. 1999. Pulsed high‐frequency electromagnetic field Nilsson JP, Arnetz B. 2011. Sleep after mobile phone
affects human sleep and sleep electroencephalogram. exposure in subjects with mobile phone‐related symptoms.
Neurosci Lett 275:207–210. Bioelectromagnetics 32:4–14.
Christ A, Kainz W, Hahn EG, Honegger K, Zefferer M, Neufeld E, Lowden A, Nagai R, Åkerstedt T, Hansson Mild K, Hillert L.
Rascher W, Janka R, Bautz W, Chen J, Kiefer B, Schmitt P, 2019. Effects of evening exposure to electromagnetic fields
Hollenbach H‐P, Shen J, Oberle M, Szczerba D, Kam A, Guag emitted by 3G mobile phones on health and night sleep
JW, Kuster N. 2010. The virtual family—Development of EEG architecture. J Sleep Res 28:e1281.
surface‐based anatomical models of two adults and two children Lustenberger C, Murbach M, Dürr R, Schmid MR, Kuster N,
for dosimetric simulations. Phys Med Biol 55:N23–N38. Achermann P, Huber R. 2013. Stimulation of the brain with
Danker‐Hopfe H, Dorn H, Bornkessel C, Sauter C. 2010. Do mobile radiofrequency electromagnetic field pulses affects sleep‐
phone base stations affect sleep of residents? Results from an dependent performance improvement. Brain Stimul 6:805–811.
experimental double‐blind sham‐controlled field study. Am J Lustenberger C, Murbach M, Tushaus L, Wehrle F, Kuster N,
Hum Biol 22:613–618. Achermann P, Huber R. 2015. Inter‐individual and intra‐
Danker‐Hopfe H, Dorn H, Bahr A, Anderer P, Sauter C. 2011. individual variation of the effects of pulsed RF EMF exposure
Effects of electromagnetic fields emitted by mobile phones on the human sleep EEG. Bioelectromagnetics 36:169–117.
(GSM 900 and WCDMA/UMTS) on the macrostructure of Nakatani‐Enomoto S, Furubayashi T, Ushiyama A, Groiss SJ,
sleep. J Sleep Res 20:73–81. Ueshima K, Sokejima S, Simba AY, Wake K, Watanabe SI,
Danker‐Hopfe H, Dorn H, Bolz T, Peter A, Hansen ML, Eggert T, Nishikawa M, Miyawaki K, Taki M, Ugawa Y. 2013. Effects
Sauter C. 2016. Effects of mobile phone exposure (GSM 900 of electromagnetic fields emitted from W‐CDMA‐like mobile
and WCDMA/UMTS) on polysomnography based sleep phones on sleep in humans. Bioelectromagnetics 34:589–598.
quality: An intra‐ and inter‐individual perspective. Environ Regel SJ, Tinguely G, Schuderer J, Adam M, Kuster N, Landolt
Res 145:50–60. HP, Achermann P. 2007. Pulsed radio‐frequency electro-
Danker‐Hopfe H, Dorn H, Sauter C, Schmid G, Eggert T. 2019. magnetic fields: dose‐dependent effects on sleep, the sleep
An experimental study on effects of radiofrequency EEG and cognitive performance. J Sleep Res 16:253–258.
electromagnetic fields on sleep in healthy elderly males Schmid MR, Murbach M, Lustenberger C, Maire M, Kuster N,
and females: Gender matters! Environ Res 183:109181. Achermann P, Loughran SP. 2012a. Sleep EEG alterations:
Harding EC, Franks NP, Wisden W. 2019. The temperature Effects of pulsed magnetic fields versus pulse‐modulated radio
dependence of sleep. Front Neurosci 13:336. frequency electromagnetic fields. J Sleep Res 21:620–629.
Hasgall PA, Neufeld E, Gosselin M, Klingenböck A, Kuster N. Schmid MR, Loughran SP, Regel SJ, Murbach M, Bratic
2015. IT'IS database for thermal and electromagnetic Grunauer A, Rusterholz T, Bersagliere A, Kuster N,
parameters of biological tissues. http://www.itis.ethz.ch/ Achermann P. 2012b. Sleep EEG alterations: Effects of
database [Last accessed 30 September 2019]. different pulse‐modulated radio frequency electromagnetic
Hinrichs H, Heinze HJ, Rotte M. 2005. Human sleep under the fields. J Sleep Res 21:50–58.
influence of a GSM 1800 electromagnetic far field. Talairach J, Tournoux P. 1988. Co‐Planar Stereotaxic Atlas of the
Somnologie (Somnology) 9:185–191. Human Brain. Stuttgart, Germany: Thieme Publishers. p 132.

Bioelectromagnetics

You might also like