@ Note: This document changes TSB 117-031C Improved -110 CMIA Optics Board to Active Complete
status. All field active instruments have been completed per attached documentation. Changes to this
document are indicated with an @ sign.
I. DISTRIBUTION:
Worldwide.
II. PURPOSE:
This TSB provides instruction to install the improved -110 CMIA Optics board.The -110 Optics
Abbott Confidential
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB - 117-031D - Improved -110 CMIA Optics Board
TSB - 117-031D - Improved -110 CMIA Optics Board Page 2
board eliminates the optics board initialization issue that can occur when the -109 optics board is
combined with a specific Quad board on an i2000 or i2000SR system. This issue cannot occur on
the i1000SR systems because this Quad board is not present in the power supply. The -110
optics board provides the same improvements as the -109 board. The -110 optics board is
equivalent to the -109 and meets all the corrective action requirements associated with the -109
board.
The new Optics board increases the optics pulse data read rate to every 5 milliseconds (msec).
The current optics board has a read rate of 100 milliseconds. The 20X faster read rate will assist
in identifying a static spike that can then be filtered from the data set. Eliminating these spikes will
provide an additional level of robustness to the optics subsystem and reduce the number of
erroneous 1007 errors reported by the system.
This TSB contains a software update that turns off the static detection algorithm that was
implemented in the v7 software release. This algorithm identifies static spikes within a result
which generates a 1007 error and sends the test to exceptions. This algorithm would also
misidentify some fast reacting assays as static spikes which would also generate a 1007 error
and send the test to exceptions. Because of this issue, the algorithm is disabled by the software
update in this TSB. This static detect algorithm has been disabled in the v8 software release and
therefore the software update in this TSB is no longer required. Any instrument with the new
optics board, that experiences 1007 errors when running fast reacting assays should be
upgraded to v8 software. The software files attached to this TSB should be used for disaster
recovery on instruments that have v7 software and have had this TSB previously installed.
Summary of Changes:
Added cut in serial numbers for the new optics board to the instruments requiring
modifications field.
Added information in the purpose section stating the software portion of this TSB is no longer
required with the release of v8 software.
Added note to Attached Files table and Software Installation procedure.
Removed information in the Administrative notes regarding v8 software and new installs.
Revised the Administrative notes to provide clarifying information regarding instruments
requiring modification.
Removed repetitive information in Administrative notes regarding the software upgrade in this
TSB and new installs.
Added clarifying information to the Administrative notes regarding instruments requiring
modification.
Abbott Confidential
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB - 117-031D - Improved -110 CMIA Optics Board
TSB - 117-031D - Improved -110 CMIA Optics Board Page 3
assigned as a result of TSB 117-027 may be voided/cancelled as per the applicable call
documentation procedure (i.e. CMSNext Version 3.0 User Guide, Navigator Reference Guide
(rev 2.0), or Pathfinder Reference Guide (rev 1.0).
ATTACHED FILES
The installation of these files are no longer required. Installing the software update should be done for
disaster recovery on instruments where v7 software and this TSB have been previously installed. Any
instrument that experiences 1007 errors when running fast reacting assays should be upgraded to v8
software.
Title Attachment Filename Filesize, Date & Time Intended Use
Installation File TSB117-031.zip 169 KB (173,253 Software update required for
bytes) new optics board. Eliminates
06/09/2011 the software driven spike
11:34:46 am GMT-6 detection.
IV. PARTS:
PARTS DISPOSITION
NEW Inventory Part Description OLD Inventory OLD Inventory Notes
Number Number Disposition
7-76185-04 Improved CMIA Optics 7-76185-03 The -01 and -02
Board 7-76185-02 boards in all
7-76185-01 Warehouse, Depot,
and Field Service
inventory locations
should be
scrapped.
Field service
inventory for the
-03 should be
returned to Dallas
Abbott Confidential
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB - 117-031D - Improved -110 CMIA Optics Board
TSB - 117-031D - Improved -110 CMIA Optics Board Page 4
for rework.
Abbott Confidential
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB - 117-031D - Improved -110 CMIA Optics Board
TSB - 117-031D - Improved -110 CMIA Optics Board Page 5
Tools and Blank CD-R obtained locally for Backup Estimated Procedure: 00:15 minutes
Materials Antistatic (ESD) Work Kit, Time Verification: 00:30 minutes
cross tip screw driver, Total: 00:45 minutes
7-76185-04 Improved -110 CMIA Optics
Board
Abbott Confidential
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB - 117-031D - Improved -110 CMIA Optics Board
TSB - 117-031D - Improved -110 CMIA Optics Board Page 6
Software Installation:
Note: Installing the software update should be done for disaster recovery on instruments where v7
software and this TSB have been previously installed. Any instrument that experiences 1007 errors
when running fast reacting assays should be upgraded to v8 software.
This section is not required for instruments at v8.00 or higher.
Do not perform this software update if the SCC is already at v8.00 software.
Action Steps Reference
Prepare 1. Obtain a blank CD-R CD.
Media
2. Label the media as follows:
Optics Board System File
TSB 117-031 207060-101
(current date)
Disk 1 of 1
Copy the 1. Create a temporary folder on your local
Files To CD laptop hard drive.
Abbott Confidential
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB - 117-031D - Improved -110 CMIA Optics Board
TSB - 117-031D - Improved -110 CMIA Optics Board Page 7
Abbott Confidential
** Potential Biohazard & Voltage Hazard. Observe Proper Safety Precautions at all Times.**
TSB - 117-031D - Improved -110 CMIA Optics Board
TSB - 117-031D - Improved -110 CMIA Optics Board Page 8
Trademark: ARCHITECT, i2000, i2000SR, i1000SR, ci4100, c8000, ci8200, c16000 and ci16200 are registered trademarks
of Abbott Laboratories. i4000, i6000, i8000, c4000, iSystem, cSystem and ARCHITECT ARM are trademarks of
Abbott Laboratories.
All other trademarks, brands, product names and trade names are the property of their respective
companies. All rights reserved. This document is Confidential for use by Abbott trained personnel only.
Admin Maintenance History:
Document # History:
TSB 117-031 moved to production with a status of Active by Patricia W Dockery on 07/27/2011 at 02:48:17 PM.
Status of TSB 117-031 set to Draft by David W Pitts on 08/09/2011 at 06:34:07 PM.
TSB 117-031A moved to production with a status of Active by Mark D. Roubadeaux on 08/23/2011 at 11:52:08 AM.
Status of TSB 117-031A set to Draft by David W Pitts on 02/13/2012 at 01:48:04 PM.
TSB 117-031B moved to production with a status of Active by Mark D Lien on 03/22/2012 at 12:58:16 PM.
Status of TSB 117-031B set to Draft by David W Pitts on 07/24/2012 at 12:58:02 PM.
TSB 117-031C moved to production with a status of Active by Jack D Sutter on 08/06/2012 at 09:59:46 AM.
Status of TSB 117-031C set to Draft by Daniel J. Gerdt on 08/24/2012 at 02:14:05 PM.
TSB 117-031D moved to production with a status of Active by Jack D Sutter on 08/29/2012 at 02:55:41 PM.
Electronic Record History: TSB 117-031D moved to production by Jack D Sutter on 08/29/2012 at 02:55:41 PM.
END OF DOCUMENT