You are on page 1of 4

No.

INF-102 Page: 1/3


TECHNICAL BULLETIN Date: 28-Sep-2006

CATEGORY PARTS REQUIRED


Top urgent Attached
As soon as possible Separately sent to you
Next time you visit the customer Prepare at your side
When the symptom appears Apply to ALOKA office for your territory when necessary
Information Sales item. Order to the sales person in charge of your country

TITLE:
IPC-1531 problem of chroma, image size, image position

MODEL AFFECTED:
IPC-1531

SERIAL NUMBERS AFFECTED:


IPC-1531: M00101C ~ M02085C (Japan/Export)
Specially M00496C ~ M00755C (produced in Nov. 2005),
M01056C ~ M01155C (produced in Apr. 2006.04),
M01756C ~ M02085C (produced in Nov. 2005) are high risk model.

SSD-4000: M05336C ~ M05765C、M06076C ~ M06160C、M05846C ~ M05895C,


M05541 ~ M06036

SSD-3500 and SSD-ALPHA5 are not affected since those have other type of TV monitor
IPC-1530(U) for SSD-3500 and IPC-1530(U-1) for SSD-ALPHA5.

BEGINNING OF PRODUCTION CHANGE:


IPC-1531:S/N M02086C and after
(Indication label “A” pasted near the AC inlet, or printed “A” in the S/N label)

PURPOSE:
Inform that both temporary solution (resetting) and permanent solution (exchanging the IC).

SYMPTOM:
¾ When turn on the power, the monitor get
1) wrong chroma
2) incorrect size of display area
3) wrong position of display

¾ The symptom may continue until the Factory


preset is recalled.

¾ Although it is temporally fixed by recalling


Factory preset, please exchange IC301.

¾ As there might be the other fact causing the


similar phenomenon, please refer the following Incorrect size of display area
technical bulletin.
TB_SSD-4000-35_Display problem caused by EP52590*CC (VPU-R)
VPU-R : EP525900CC (NTSC) / EP525901CC (PAL).
No. INF-102 Page: 2/3
TECHNICAL BULLETIN Date: 28-September-2006

CAUSE:
IPC-1531 has an MPU. Due to the quality of MPU: IC301, “Auto input signal detect” function
which is not used for ALOKA system works and change the setup data (chroma, display position,
display size, ..etc.) during the boot up.

[Technical term]
IC301 : The MTV212M32 micro-controller is an 8051 CPU core embedded device
specially customized for monitor application. As ALOKA factory preset data keeps in the
other IC. You will not loose it when you exchange IC301.
The IC has three functions, MPU, ROM, and RAM.

Factory preset data : It is the data of chroma, display position, display size, and including
OSD setting adjusted for ALOKA. It is saved in the different IC than IC301.

Factory preset recall : Restores "ALOKA factory preset data" and copied into an RAM
table in IC301.

RAM table in IC301 : It is a table contains the data for MPU to set an initial condition
during boot-up. ALOKA Factory setting data is copied into this table by the master data on
the other IC. However, the wrong data could be saved if “Auto input signal detect”
unexpectedly worked.

SOLUTION:
1: (Temporary solution/ Cause confirmation: Factory preset data recall)
If you don’t have PM-IPC-1531-1, or judge the cause of trouble which from IC301-monitor side
or main unit SSD side. Remove the monitor rear cover, do the “Factory preset data recall” follow
the attached sheet. (TB_INF-102_Proceduer of the Reset).

2: (Permanent solution/ exchange the IC by PM kit: PM-IPC-1531-1)


Use the PM-IPC-1531 and follow the attached IC301replacement proceduer_MS4-2470

The location of IC301 shows right.

High voltage line is not near of IC301.


But Please be cheerfully to avoide electrical
shock hazards.

During modification,
Please pay attention to Electro Static Discharge.

Note.
This monitor has a MPU, after apply power,
MPU start up then cooling fan start running,
there is a time lag.

Please follow the IC301Replacement


procedure (MS4-2470) basically,

Location of IC 301(Top view)


However, you can follow (MS4-2469),
if you can put your hand and arm from a top side, and modification is possible without giving a
damage to an electricity components. If it’s not sure, you should follow the (MS4-2470).
No. INF-102 Page: 3/3
TECHNICAL BULLETIN Date: 28-September-2006

ITEMS OF REQUIRED PARTS:


PM-IPC-1531-1 (Contents of kit: IC-301(HC1512 VER1.1A), modified indication label “A” )
Please request the modification kit to Aloka office with S/N of IPC-1531 and S/N of SSD-4000.

ATTACHED DOCUMENTS:
1) Factory preset data recall (Attachment of INF-102)
2) INF-102_IC301Replacement proceduer_MS4-2469(20pages)
3) INF-102_IC301Replacement proceduer_MS4-2470(28pages)
4) Product Update Report, 1 sheet

ESTIMATED LABOR TIME:


1.0 hour

SPECIAL TOOLS AND TEST EQUIPMENT REQUIRED:


* Screw driver: No Magnetic type 3mm Phillips screwdriver (+).
* Screw driver: No Magnetic type Flat-bladed screwdriver (-).
* Tweezers.
* Pastern for fixing screws. (1401B or its equivalent)
* Cushion : for protect the monitor front cover during modification
* Wrist strap with conductive sheet For ESD. (3M 8501 (98-0798-1751-0) or its equivalent)

--- Not necessary but useful Items ---


* IC lifter(IC remover) is useful tool to avoide a brake the IC socket.
* Monitor cable “CO-PSC-135-G-21C0” is useful to connect the J50 (monitor video signal) on
SSD-4000 to the other IPC-monitor. S(Y/C) cable and BNC cable also useful.

3M Field Service Kit 8501 CO-PSC-135-G-21C0


ESSENTIAL CONDITION:
None

PRODUCT UPDATE REPORT


After the modification, please send “Product Update Report” by fax to ALOKA TOKYO office
Quality Assurance Department directly. Fax : +81(Japan) 428 32 5903
Fax Number is also in the report sheet.
Approved by

General ManagerMedical Ultrasound


Engineering Dept.

Written by

AlokaTechnical Service
Product Update Report
Copy this sheet to fill in.

To ALOKA Co., Ltd. Fax +81(Japan) 428 32 5903


Your company name Your Fax No. Start from Country code
From
Fax +

This reports that the work has completed according to ALOKA Technical Bulletin INF-102.

Title: IPC-1531_IC301 problem.

Report: Please fill the followings and fax to ALOKA TOKYO.


Customer
Name or
Date - - Hospital
(yyyy-mm-dd) Name

Country City

Serial Number
Model Name SSD- of SSD M0

Serial Number
Model Name IPC-1531 of IPC M0

Company

Modification done by. (BLOCK LETTER)


Name of
Technician.

Remarks
IC-301 exchanged to V1.1A by PM-IPC-1531-1

Please fill this form and fax to ALOKA


TOKYO office.

Signature of technician

You might also like