Professional Documents
Culture Documents
Features
• Environmental monitoring in
calibration laboratories
• Industrial applications in
semiconductor industry, engine
testing and maritime sector
• GPS meteorology: estimating
precipitable water vapor in the
atmosphere, weather stations
Model Dimensions in mm
120 (4.72)
12 (0.47)
183 (7.20)
PTU301 for wall mounting 169 (6.65)
116 (4.57)
96 (3.78)
Ø 7 (0.28)
Model Dimensions in mm
Ø 12 (0.47)
M9x1
37.5 (1.48)
78.5 (3.09)
98.5 (3.88)
Ø 12 (0.47)
M9x1
37.5 (1.48)
79.5 (3.13)
99.5 (3.92)
PTU307 warmed probe for
Ø 3.2 (0.13)
Ø 6 (0.24)
demanding meteorological
installations
130 (5.12)
Ø 3.2 (0.13)
Ø 6 (0.24)
130 (5.12)
HMT330MIK Meteorological Installation Kit enables PTU307 to be installed outdoors to obtain reliable measurements for meteorological purposes.
Technical data
Measurement performance ∆ °C
0.7
Barometric pressure 0.6
0.5
Pressure range 500 … 1100 hPa 50 … 1100 hPa 0.4
Accuracy 500 … 1100 hPa 500 … 1100 hPa 50 … 1100 hPa 0.3
0.2
Class A Class B
0.1
°C
Linearity ±0.05 hPa ±0.10 hPa ±0.20 hPa 0
Hysteresis ±0.03 hPa ±0.03 hPa ±0.08 hPa -0.1
-0.2
Repeatability ±0.03 hPa ±0.03 hPa ±0.08 hPa -0.3
Calibration ±0.07 hPa ±0.15 hPa ±0.20 hPa -0.4
unceratinty -0.5
-0.6
Accuracy at ±0.10 hPa ±0.20 hPa ±0.30 hPa
-0.7
+20 °C / +68 °F -80 -60 -40 -20 0 20 40 60 80 100 120 140 160 180
Temperature ±0.1 hPa ±0.1 hPa ±0.3 hPa
dependence Accuracy over temperature range
Total accuracy ±0.15 hPa ±0.25 hPa ±0.45 hPa
(-40 … +60 °C /
-40 … +140 °F) Operating environment
Long-term ±0.1 hPa ±0.1 hPa ±0.2 hPa Operating temperature -40 … +60 °C (-40 … +140 °F)
stability/year
Operating temperature with optional 0 … +60 °C (+32 … +140 °F)
Response time (100 % response): display
One sensor 2s 1s 1s Humidity range Non-condensing
Pressure units hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia Measurement environment For air, nitrogen, hydrogen, argon,
Relative humidity helium, and oxygen 1)
Measurement range 0 … 100 %RH 1) Consult Vaisala if other chemicals are present. Consider safety regulations with flammable gases.
116 (4.57)
96 (3.78)
Uout 3 x 0 … 1 V / 0 … 5 V / 0 … 10 V Max. 33 mA
Ø 7 (0.28)
Iout 3 x 0 … 20 mA Max. 63 mA
Display and backlight +20 mA
During chemical purge Max. +110 mA
During probe heating +120 mA
External Loads
Current outputs RL < 500 Ω
0 … 1 V output RL > 2 kΩ
77 (3.03)
71 (2.80)
All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners.
www.vaisala.com Any reproduction, transfer, distribution or storage of information contained in this document is strictly
prohibited. All specifications — technical included — are subject to change without notice.