Professional Documents
Culture Documents
12 Service
12.1 Introduction
This analyzer provides multiple maintenance functions for this purpose. This chapter introduces how
to use the provided functions to maintain and troubleshoot your analyzer. Preventive and corrective
maintenance procedures are required to keep the analyzer in a good operating condition.
All the analyzer components and surfaces are potentially infectious, take proper protective measures
for operation or maintenance.
Performing unauthorized maintenance procedures can damage your analyzer. Do not perform
any maintenance procedures that are not described in this chapter.
In case of problems not specified in this manual, contact Dymind customer service department
or your local agent for assistance.
Only Dymind-supplied parts can be used for maintenance. For any question, contact Dymind
customer service department or your local agent.
Exercise caution to avoid contact with the sharp sample probe when performing maintenance.
12.2 Maintenance
The analyzer provides multiple service functions helping users to perform daily maintenance.
The reagents are irritating to eyes, skin and mucosa. Wear proper personal protective
equipment (e.g. gloves, lab uniforms, etc.) and follow laboratory safety procedures when
handling them in the laboratory.
If the reagents accidentally spill on the skin, wash them off with plenty of water and if necessary,
go see a doctor; if the reagents accidentally spill into the eyes, wash them off with plenty of
water and immediately go see a doctor.
143
12 Service
After long-distance transportation, the reagent must be allowed to settle for more than one day
before use.
When you have replaced the diluent, cleansers or lyses, run a background check to see if the
results meet the requirement.
144
12 Service
4. Click the name of the reagent that needs to be replaced, such as Replace All Reagents.
After the replacement is completed, the following message box will pop up.
Figure 12-3 Reagent Replaced
12.2.2 Cleaning
Clean corresponding parts according to the actual situation:
WBC bath
When the background of WBC- and/or HGB-specific parameters exceeds the Ref. Range, you
should clean the WBC bath.
RBC bath
When the background of RBC- and (or) PLT-specific parameters exceeds the Ref. Range, you
should clean the RBC bath.
Sample probe
When the sample probe is dirty, you should clean the sample probe.
The cleaning procedures are as follows:
1. Click the Service icon in the menu page to access the Service interface.
145
12 Service
2. Click Clean in the Maintenance selection, an interface as shown in Figure 12-5 will pop up on
the screen
Figure 12-5 Cleaning
3. Click the icon of the part that needs to be cleaned, such as Clean Sample Probe.
When the system cleaning is complete, the message box will pop up to show that the cleaning is
done.
Figure 12-6 Cleaning Done
146
12 Service
12.2.3 Maintenance
Maintenance of the analyzer includes unclogg and cleanser soak.
12.2.3.1 Unclogging
If clogging is found, or it is suspected that the counting results are not accurate due to aperture
clogging, you can perform the unclogging operations.
The unclogging procedures are shown as follows:
1. Click the Service icon in the menu page to access the Service interface.
Figure 12-7 Service
147
12 Service
148
12 Service
4. Click Yes.
A dialog box as shown below will pop up.
Figure 12-11 Cleanser Soak Prompt
If insufficient aspiration volume is prompted, make sure the sample probe tip is vertically put
below the liquid surface of the cleanser. If HGB background abnormality is prompted, remove the
error as per the prompt, then redo the cleanser maintenance.
5. Present the cleanser to the sample probe as per the prompt, and press the aspirate key or click
the Aspirate button.
“Cleanser soaking…” and the soaking time will appear as shown below. As shown below.
149
12 Service
7. Click Close.
8. Perform the above procedures to perform the cleanser soak again if necessary.
150
12 Service
4. Click OK.
151
12 Service
4. Click OK.
152
12 Service
4. Click Yes.
A dialog box will pop up as shown below.
5. Remove all reagent pickup tube assemblies according to the prompt, and then click OK to start
emptying the fluidic system.
A dialog box will pop up to prompt you to power off the device after the emptying is complete.
6. Turn to [O] the [O/I] switch located on the left side of the analyzer.
7. After shutdown, empty the waste in the waste container, and dispose of it.
Be sure to dispose of reagents, waste, samples, consumables, etc. according to local legislations
and regulations.
153
12 Service
4. Click Yes.
A dialog box will pop up as shown below.
5. Remove all reagent pickup tube assemblies according to the prompt, and then click OK to start
emptying the fluidic system.
After the emptying is complete, a message box will pop up.
154
12 Service
6. Place all reagent pickup tube assemblies into the distilled water, and then click OK to start
priming.
Be sure to use distilled water in order to ensure the normal use of the device in the future. In
addition, the beaker holding the distilled water needs to be cleaned thoroughly.
The diluent pipe and lyse pipes should be stored separately in two beakers.
System performs the filling operation. After the filling is completed, the following dialog box will
pop up.
7. Take out the diluent and lyse pipes from the distilled water as per the prompt, then click OK.
After the operation, dialog box will pop up to prompt you to power off the device.
8. Turn to [O] the [O/I] switch located on the left side of the analyzer.
9. After shutdown, empty the waste in the waste container, and dispose of it.
Be sure to dispose of reagents, waste, samples, consumables, etc. according to local legislations
and regulations.
155
12 Service
Once the auto clean is performed or the analyzer is shut down, the statistical data of auto clean will
be cleared automatically.
At the Self-test or Status interface, the analyzer does not ask for confirmation to perform the
cleanser soak.
If the analyzer is running or has problems when the conditions of auto prompt for cleanser soak
is satisfied, the analyzer will prompt again after the current operation is completed or the
problems are resolved.
After cleanser soak is completed, the accumulative count values will be cleared automatically.
Cleanser soak is an important step in comprehensive device maintenance. It is recommended
not to stop soaking halfway.
You can view, set the information or do other operations besides about execution of the
sequence actions (such as runnning samples, replacing reagent, screen testing or removing
156
12 Service
If it is the time to auto sleep but the analyzer is error status, then only after the error is removed
will auto sleep start accordingly.
Different maintenances will be performed by the analyzer automatically when exiting the sleep
mode, and the exiting time depends on how long the analyzer was in the sleep mode.
If errors occur when you are trying to cancel the auto sleep of the analyzer, please refer to 13
Troubleshooting for solving the problems.
You can change the waiting time for auto sleeping as needed, see 5.3.4 Auto Maintenance.
12.3 Self-inspection
This feature is to test if some important components of the device can function properly or not,
including syringe and sampling assembly self-inspection, pressure and vacuum self-inspection,
valve self-inspection and other self-inspections.
If the testing result is abnormal, you should try again for several times; if the abnormalities persist,
please contact Dymind customer service department or your local agent.
3. Click the part that needs to be tested, e.g. Sample Syringe, and wait for the self-inspection
results.
After the self-test is completed, a dialog box will pop up to show the self-test results.
157
12 Service
158