Professional Documents
Culture Documents
MODEL #CM350
Model CM350
Model CM350
Catalog # 425-CM350
No. CAT CM35018Ae
Genesis BPS
465 Route 17, PO Box 331
Ramsey, NJ 07446 U.S.A.
Tel: 201-708-1400, 866-712-5663
Fax: 201-708-1104
info@genesisbps.com
www.genesisbps.com
Contents
Page
Important Notes 3
2. Specification 6
3. Configuration 7
3-1. Appearance 7
3-2. Control Keys and Indicator Lamps 8
3-3. Display Panel 9
4. Preparation for Use 10
4-1. Installation 10
4-2. Automatic Weight Calibration 11
4-3. Manual Calibration Adjustment 12
5. Operating Procedures for Collection 13
6. Operational Tips 15
8. Battery Recharging 18
9. Maintenance 20
9-1. Troubleshooting 20
9-2. Cleaning Instructions 20
9-3. Storage 20
9-4. Disposal of Waste Products or Residues 20
9-5. Service Call 21
1
Description of Safety Symbols
WARNING
“Warning” indicates that inappropriate use of instrument might result serious injury to the user and
damage to the instrument. User must follow the procedures as specified in this manual.
CAUTION
“Caution” indicates the presence of a hazard that could result in minor injury, or property damaged if
ignored.
NOTE
“Note” describes information for the installation, operation, or maintenance which is important for
proper use.
CLASSIFICATION
2
.
Important Notes
WARNING
This equipment has been tested and found to comply with the limits of EMC requirements in
EN 61326-1:2006. These limits are designed to provide reasonable protection against harmful
interference from adjacent devices.
CAUTION
1. Do not hang excessive weight on or pull the bag hook downward in a forceful manner.
(This avoids damage to the weight-sensitive device above the bag hook.)
Do not hang weights over 1 kg (2.2 lb) on the bag hook.
2. Hold the stand while in motion if the main unit is mounted on.
CLASSIFICATION
Battery Capacity
Turn power off to save the battery when the machine is not to be used for a ling while. Unit power
will turn off automatically if a control key has not been touched for 30 minutes.
A machine with a fully charged battery can operate continuously for approximately 50 hours.
Keeping
Avoid using or leaving the machine in a location where the temperature is higher than 104 F (40 C).
Keep it away from any heating source, direct sunlight, where the equipment is exposed to water
vapor, where the humidity is extremely high or there is a ventilation problem, and where the
equipment is subject to excessive shocks or vibration. Don’t operate the equipment with a wet hand
as well.
3
1. General Scope and Features
The Blood Collection Monitor TM CM350 monitors blood flow to measure blood volume collected in
the blood bag, and clamps the tubing automatically when the collected blood volume reaches the
preset target volume.
To ensure error-free operation, diverse audible and visual alarms are activated:
- When a collection procedure is properly completed. ("End collection alarm")
- When one minute has elapsed since the preceding mixing. ("No mix alarm")
- When 30 seconds have elapsed since the bag was removed from hook. ("Bag off alarm")
- When blood flow rate is lower than 30 mL/min. ("Low flow alarm")
- When 20 minutes have elapsed from collection start. ("20 minute alarm")
- When battery voltage is low. ("Low battery alarm")
Audible alarms are also generated to provide the following alert and indication messages:
- Beeps for 10 seconds when 10 minutes have elapsed from collection start. ("10 minute alarm")
- 3 beeps when [POWER] key is accidentally pressed in a collection cycle. ("Wrong key alarm")
- A short (0.2 sec) beep when any control key is pressed. ("Key touch indication")
- A long (0.5 sec) beep when device power is turned on or off. ("Power on/off indication")
4
A standard set of CM350 package consists of the following, which are contained in a carton:
The following are not included in a CM350 package, but are available as optional accessories:
500-gram Stand
Aluminum carrying case (*)
Calibration weight (Pole + Pedestal)
(*) Carrying case can contain: main unit, pole mount, charging adaptor, calibration weight.
Front Rear
5
2. Specification
(1) Collection volume display : 0 ~ 1,199 mL, with 1 mL increments
(2) Total weight display : 0 ~ 1,199 gram, with 1 gram increments
(3) Elapsed time display : 0:00 ~ 19:59 min/sec
(4) Target volume set : adjustable between 100 ~ 600 mL, with 5 mL steps
(5) Preset target volume : 2 ~ 4 figures are factory-set upon request
(6) Accuracy and Tolerance:
a. Volume/Weight display : 1 ~ +3 mL or -1~+3 gram
b. Elapsed time display : 1 second
c. mL-gram conversion factor (Specific gravity of blood) : 1.057
(7) Control keys: [POWERCAL], [TIMEBEEP], [5 ml ▼], [PRESET], [5 ml ▲], [START], [CLAMP]
(8) Indication panel : custom-designed composite LCD display
(9) Visual indicator lamps:
a. START lamp [ ] turns on when collection starts, and flashes when collection is completed.
b. CLAMP lamp [ ] turns on when the clamp is closed, and turns off when the clamp is open.
c. MIX lamp [ ] flashes after one minute has elapsed since the preceding mixing.
(10) Audible and visual alarms:
a. End collection alarm : when a collection procedure is properly completed.
b. No mix alarm : when one minute has elapsed since the preceding mixing.
c. Bag off alarm : when 30 seconds have elapsed since bag was removed from hook.
d. Low flow alarm : when blood flow rate is lower than 30 mL/min.
(*)
e. 20 minute alarm : when 20 minutes have elapsed from collection start.
f. Low battery alarm : when battery voltage is low.
(11) Audible alarms:
a. 10 minute alarm (*) : when 10 minutes have elapsed from collection start.
b. Wrong key alarm : when [POWER] key is accidentally pressed in a collection cycle.
c. Key touch indication : when any control key is pressed.
d. Power on/off indication : when device power is turned on or off.
(12) Dimension - Main unit (width-height-depth): 162-120-60 mm (6.4-4.8-2.4 inch)
(13) Weight: a. Main unit only : 750 gram (1.7 lb)
b. Standard package set : 1,400 gram (3.1 lb)
c. Stand (pole + pedestal) : 3,000 gram (6.6 lb)
d. Carrying case only : 1,800 gram (4.0 lb)
e. Total package (including calibration weight): 6,700 gram (14.8 lb)
(14) Battery pack : 7.2V/1.8AH rechargeable cells, NiMH
(15) Charging adaptor : Elpac Power Systems, model MW1212
Input: 100~240V, 50/60 Hz, 0.5A, Output: 12V, 1.0A, 12W
Or, Bridge Power, model JMW110KA12
Input: 100~240V, 50/60 Hz, 0.5A, Output: 12V, 1.25A, 15W
(16) Power cord set in the U.S.A. and Canada :
For 120 volt applications, use only UL Listed detachable power cord with NEMA
configuration 5-15P type (parallel blades) plug cap.
For 240 volt applications, use only UL Listed detachable power cord with NEMA
configuration 6-15P type (tandem blades) plug cap.
Note : For other countries, use a grounded 3-prong power cord set.
(17) Temperature range - Operating : 0 ~ 40 oC (32 ~ 104 oF)
Storage : 20 ~ 70 oC (4 ~ 158 oF)
(Note*) : This function may be disabled by the request of customer.
6
3. Configuration
CM350 unit
Display panel
Tube clamp
Bag hook
7
3-2. Control Keys and Indicator Lamps (Figure 2)
(1) Display panel - Displays weight, volume, elapsed time and operation symbols.
(2) [POWERCAL] - Dual function key:
Short press (a click for 1 second) - Turns device power on or off.
Long press (hold for 3 seconds) - Calibration procedure starts.
(*) Remark:
If the key is pressed during a collection cycle, power does not turn off but a warning alarm
(Wrong key alarm: 3 short beeps) sounds. This is a measure not to make the device turned
off by an unintentional or accidental press during a collection cycle.
If the key is pressed and held for 20 seconds continuously, device power does turn off
regardless of the procedure. This is a measure to make the device turned off during a
collection cycle, in case that user wants to stop a collection before completion.
(3) [TIMEBEEP] - Dual function key:
Short press (one click) - Displays elapsed time from start (min : sec).
Long press (hold for 3 seconds) - Turns beep sound on or off.
(4) [5 ml ▼] - Decreases target collection volume by 5 mL increments.
(5) [PRESET] - Displays preset target collection volumes.
(6) [5 ml ▲] - Increases target collection volume by 5 mL increments.
(7) [START] - Starts a collection procedure cycle. The green lamp turns on.
(8) [CLAMP] - Opens or closes clamp.
(9) CLAMP lamp [ ] - Turns on when the tube clamp is closed.
(10) START lamp [ ] - Turns on once collection procedure has started, and flashes at
completion.
(11) MIX lamp [ ] - Flashes after one minute has elapsed since the preceding mixing.
8
3-3. Display Panel (Figure 3)
(1) Main display area - Displays weight, volume and elapsed time.
(2) [GM] - Indicates the displayed weight is in gram.
(3) [ML] - Indicates the displayed volume is in milliliter (mL).
(4) [ ] - Indicates the audible alarms are muted.
(5) [ ] - Indicates collection procedure is in progress. The indicator turns on after [START] is
pressed. Indicator flashes either when the collected volume reaches target volume, or
when a low flow alarm is activated.
(6) [LOW] - Turns on when battery charge is low (LOW ), or when flow rate is low (LOW ).
(7) [ ] - Turns on during battery recharging. Flashes when the battery has been fully charged.
LOW is turned on when battery charge is low.
(8) [ALARM] - Turns on when any alarm is activated.
(9) [PRESET] - Indicator for target collection volume.
It turns on when [PRESET], [5 ml ▲] or [5 ml ▼] key is pressed.
(10) [CAL] - Indicator for calibration mode.
9
4. Preparation for Use
CM350 unit
Thumbscrew
10
4-2. Automatic Weight Calibration
It is recommended that the weight calibration procedure be performed prior to the first donor
collection of the day. It is also recommended when collecting on a mobile vehicle, to repeat the
weight calibration procedure each time the vehicle is moved to a new location. Precise leveling of
the unit or the site is not required. The weight calibration procedure can be done automatically
using the following steps.
11
4-3. Manual Calibration Adjustment
As in the Section 4-2, weight calibration is done automatically. However, automatic calibration can
be done when calibration deviation remains within ±2 % range (490-510 grams). If the calibration
deviation falls outside the range, a manual adjustment is necessary to get calibration deviation back
into automatic adjustable range. It could happen if CM350 has had severe shocks.
If A d J appears during calibration process in the Section 4-2-(2), remove the calibration weight
from hook, and follow the manual calibration adjustment procedure below.
Manual calibration adjustment is done by simply turning a potentiometer located on the left side.
In the hole next to the charger jack is a potentiometer. Turning the potentiometer changes load cell
circuit ratio to adjust the calibration deviation back into automatic adjustable range.
12
5. Operating Procedures for Collection
The purpose of the Blood Collection Monitor TM CM350 monitors the volume of blood collected
in the blood bag by measuring the weight, and clamps the tubing automatically when the
collected blood volume reaches the preset target volume. It also tells when to mix blood with
coagulant by blinking the MIX indicator lamp.
(3) Place the primary blood bag on the bag hook. (e.g.)
Display panel shows the weight of the primary blood bag.
13
REMARK: A collection cycle begins only when the bag (NOTE)
weight is between 30-500 grams. If the key is pressed when
the weight hung on the bag hook is either below 29 grams or
over 501 grams, an error message ("E r r") is displayed.
The display panel shows the current weight in 3 seconds.
Check the blood bag to adjust what's wrong.
Press [START] key again.
NOTE: The unit ignores an abrupt change of the weight while
the unit is in a collection cycle. Any weight change of +/ 30
grams or greater in 1 second interval will be ignored so the
display remains unchanged.
NOTE: The following alarm may sound.
- Low Flow alarm if blood flow rate is below 30 mL/min.
- 10 Minute alarm if 10 minutes have elapsed.
- 20 Minute alarm if 20 minutes have elapsed.
(9) The MIX indicator lamp (MIX ) blinks after one minute has
elapsed since preceding mixing ("No Mix alarm"). MIX
Mix the bag periodically as required by your facility's protocol.
NOTE: When the blood bag is removed for mixing, collected
volume figure displayed on the display panel freezes until the
bag is placed back to the bag hook.
REMARK: If the bag is removed from the bag hook for more
than 30 seconds, an alarm will sound ("Bag Off alarm"). If the
bag remains off the hook for 1 minute, the unit will return to
the initial power-on state and all collection data will be lost.
CAUTION: Do not disturb the collection bag while the last 20
mL are collected. The weight of the bag is being verified.
The MIX indicator lamp will not turn on during this period.
(10) When collected volume reaches target volume, collection ends. (e.g.)
Clamp closes automatically.
The START indicator lamp and symbol flash.
End Collection alarm sounds. (3 long beeps)
14
(14) Remove the blood bags.
Go to step (2) above for the next collection, or press
[POWERCAL] key to turn the unit power off.
NOTE
If equipment is not operated properly, see the troubleshooting section(9-1.) of this manual.
6. Operational Tips
(e.g.)
(1) To record collected blood volume:
See Operating procedure shown in step (10) above.
(e.g.)
(2) To record total weight of blood bag and collected blood:
See Operating procedure shown in step (12) above.
(e.g.)
(3) To record elapsed time of collection:
Press [TIMEBEEP] key.
(e.g.)
(6) To change the preset target volume during collection:
Adjust the target volume anytime by pressing
[PRESET], [5 ml ▲], or [5 ml ▼] key.
15
Remove the bag from the hook and wait for 1 minute.
All collection data will be lost.
If [POWERCAL] key is pressed and held for 20 seconds,
device power forcibly turns off. All collection data will be
lost.
Pressing CLAMP key terminates collection procedure, and turns off the alarm.
Tube clamp closes. The CLAMP indicator lamp turns on. The START indicator lamp and
symbol flash.
B. No Mix alarm:
This alarm is activated after one minute has elapsed since the preceding mixing.
Three short beeps sound at the beginning of this alarm.
The MIX indicator lamp (MIX ) flashes.
When the blood bag is removed for mixing, collected volume figure displayed on the display
panel freezes until the bag is placed back to the bag hook.
16
To prevent losing collection data, the blood bag must be placed back on the hook within 30
seconds.
NOTE
If the bag is off the hook for longer than 1 minute during the collection, the unit will
return to the initial power-on state and all collection data will be lost.
Be sure to place the bag back on the hook before this alarm ends.
E. 10 Minute alarm:
This alarm is activated after 10 minutes have elapsed from the start of a collection procedure.
Short beeps sound for 10 seconds, only once:
F. 20 Minute alarm:
This alarm sounds after 19:59 minutes have elapsed from the start of a collection procedure.
Short beeps sound until the unit returns to its initial state:
NOTE
- To end the collection, remove collection bag from hook and wait for 1 minute.
This will erase collection data. All collection data will be lost.
- To prevent the loss of any collection data, press [5 ml ▼] key to decrease the selected
target volume to the current volume collected.
17
H. Low Battery alarm:
This alarm sounds when the battery charge is depleted.
On the display panel, LOW turns on. One long beep sounds once.
Recharge the battery. If this alarm begins, the battery pack still has sufficient charge to
continue operating for several hours.
Battery symbol ( ) flashes when the battery pack is fully charged. Stop recharging.
8. Battery Recharging
A. Battery pack:
The unit is powered by a rechargeable battery pack, consisting of six Nickel Metal Hydride
(NiMH) cells, 7.2 volts/1.8 AH, AA-size.
B. Battery operation:
Once the battery pack is fully charged, the unit will run for approximately 50 hours.
If the unit was stored for an extended period, recharge it before use.
When battery charge is low and needs recharging, LOW symbol will appear in the display
panel. Recharge the battery.
NOTE
Even when LOW symbol is initially displayed, the battery pack will run the unit for
approximately 5 hours.
C. Battery recharging:
For recharging, use only the charging adaptor supplied with the unit.
Plug the circular plug of charging adaptor into the battery charger jack on the unit.
Then plug the AC cord plug of charging adaptor into an AC wall outlet.
AC cord plug
Circular plug
Recharging continues whether unit power is on or off. Battery symbol ( ) turns on.
Recharging takes approximately 15 hours if the battery is fully depleted.
If there is a residual charge remaining in the battery, recharging will take less time.
When the battery pack is fully charged, the symbol ( ) starts flashing.
18
Remove the AC cord plug from AC outlet, and the circular plug from battery charger jack.
Excessive recharging will not damage the battery pack.
CAUTION
Be sure to use the charging adaptor supplied with the CM350 unit. Using any other
charging adaptor may result in permanent damage to the battery pack and the CM350
unit.
NOTE
You can find the detailed distributor information in the last page of this manual.
19
9. Maintenance
9-1. Troubleshooting
The following table covers common problems and suggested solutions:
9-3. Storage
If the equipment is not to be used for a long time, it is recommended to protect it from dust, dirt
or humidity, etc.
CAUTION
To dispose the equipment, accessories, and components, follow local governing ordinances
and recycling plans regarding disposal or recycling of device components. Especially a lithium
battery may pollute the environment if the instrument or a lithium battery is abandoned.
When disposing packing materials, sort them by the materials and follow local governing
ordinances and recycling plans.
20
9-5. Service Call
Genesis BPS
465 Route 17, PO Box 331
Ramsey, NJ 07446 U.S.A.
Tel: 201-708-1400, 866-712-5663
Fax: 201-708-1104
info@genesisbps.com
www.genesisbps.com
Manufactured by:
21