Measuring device list

Digital manometer (or digital calibrator)


(Pressure measuring device)
Pressure gauge, pressure transmitter, pressure
switch, for calibration
The measurement range:0 ~ 3.5 MPa

CA150 calibrator
(DC voltage current generator, DC voltage current
measuring device)
Digital Panel Meter, transducers, for calibration
DC 4~20 m A、DC1~5V、RTD(Pt100Ω)
Output of thermocouple (K type), input possible

Digital multimeter
(Voltage, current, resistance measuring instrument)
For checking the contact of the pressureswitch (Ω,
buzzer)
For signal measurement of pressure transmitter
(DCmA)
For confirmation of continuity of thermocouple
/ RTD (Ω)
Digital thermometer
Thermometer, temperature switch for calibration

Insulation resistance meter (commonly known as


megger)
RTD, thermocouple insulation resistance
measurement (range 250V)
How to use digital multimeter

Digital multimeter, code Insert the cord clip into the measuring point

In case of current measurement, mA · COM


Normally OFF For voltage and resistance measurement, V · Ω
and COM

Voltage range (thermocouple inspection) Buzzer (Pressure switch test)


Resistance range (continuity test) Current range (pressure transmitter, signal converter)

How to use pressure manometer and pressure pump

Pressure manometer, air hose Pressure pump set

Fine adjustment valve


UP、DOWN swich
(Used when raising
(Used when measuring
and lowering pressure
range change)
a little)

Closing valve
(Close at
pressurization)
MENU switch
(Use at 0 tone)

Pressure pump body


Combination of pressure pump, pressure Before pressurization / After pressurization
manometer, air hose

How to use digital thermometer

Digital thermometer body

Thermosensitive part (K type sensor)


Insert the sensor connector into the
socket

Connector slot Sensor


connector

Power switch
(Use at zero tone)

Select sensor type


(Choose K type here)

How to use insulation resistance meter (megger)


Insulation resistance meter body
Connect the red lead wire to the LINE
Connect black lead wire to
EARTH

Connect the lead wires to the connector on the back of the megger

Set the measurement range to 250 V Ready for measurement

Pull up MEAS lever and


measure
How to use Calibrator (CA 150)

Connect red wire toH side of


OUTPUT terminal and black
lead wire to L side

Portable Calibrator (CA 150) Body OUTPUT terminal


INPUT terminal
Lead wire
The opposite side (clip side) of the lead
wire is connected to the measuring
instrument

DC voltage
DC voltage current
Press power switch Ω
TC(Thermocouple emf)
RTD(Resistance
temperature detector)
・Select a type with the FUNCTION button
・Press SOURCE button Output with UP, DOWN button
(Note the number of digits)

List of Tools

Pressure pump Use it in the test of 0 to 1.5 Electric pot used during thermometer test
Mpa Pressure gauge, pressure switch, (Anything is acceptable as long as
pressure transmitter calibration heating is possible)

Air hose (Used for pressure related Needle withdrawal


instrument test) (Used when adjusting pressure gauge)
Lubricant, Cleaning solution
Fittings(Used when testing pressure (Clean the instrument, lubricate the shaft)
related instruments)
Closure plug
Closure plug
(When removing the instrument,
plug it so as not to leak water or oil)