Professional Documents
Culture Documents
Visual Physiology
Introduction to Cones are used primarily for daylight vision,
called photopic vision,
Interventional Radiology
Cones perceive small objects much better
Fluoroscope than rods.
Thomas A. Edison invented the fluoroscope Cones are also much better at detecting
in 1896 differences in brightness levels.
The fluoroscope is used primarily for dynamic Rods are used for night vision, called scotopic
studies vision
Fluoroscopy is actually a rather routine type Cones perceive color, but rods are essentially
of x-ray examination except for its color blind.
application in the visualization of vessels, Visual acuity - ability to perceive fine detail
called angiography. Contrast perception - detecting differences in
Two main areas brightness levels.
Vascular Radiology
Neuroradiology
The x-ray tube is usually hidden the patient
couch, and the image intensifier or other
image receptors are set over the patient
couch.
Some fluoroscopes are operated remotely
from outside the x-ray room.
During fluoroscopy, the x-ray tube is
operated at less than 5 mA
radiographic examination, in which the x-ray Fluoroscopic technique
tube current is measured in hundreds of mA The brightness of the fluoroscopic image
Despite the lower mA, however, the patient depends primarily on the:
dose is considerably higher during anatomy that is being examined,
fluoroscopy than during radiographic kVp,
examinations because the x-ray beam mA.
exposes the patient continuously for a The influence of kVp and mA on fluoroscopic
considerably longer time. image quality is similar to their influence on
radiographic image quality.
AUTOMATIC BRIGHTNESS CONTROL (ABC) Generally, high kVp and low mA are
Fluoroscopic equipment allows the preferred.
radiologist to select an image brightness level
that is subsequently maintained
automatically by varying the kVp, the mA, or
sometimes both.
Illumination
levels are measured in units of lumen per
square meter or lux
Radiographs are visualized under
illumination levels of 100 to 1000 lux
Flux gain
ratio of the number of light photons at the
output phosphor to the number of x-rays at
the input phosphor
The photocathode emits electrons when Flux Gain ="Number of output light
illuminated by the input phosphor photons" /"Number of input x-ray photons"
Photoemission is electron emission that
follows light stimulation. Minification gain
What is the relationship between the the image minification from input phosphor
number of electrons emitted by the to output phosphor is the ratio of the square
photocathode and intensity of light of the diameter of the input phosphor to the
reaches it? Directly Proportional square of the diameter of the output
Intensity of the incident image- phosphor
forming x-ray beam Minification gain = (di/do)²
Intensity of light where di is the diameter of input phosphor
Number of electrons emitted and do is the diameter of output phosphor.
Brightness gain
The ability of the image intensifier to
increase the illumination level of the image
The brightness gain of most image
intensifiers is 5000 to 30,000
simply the product of the minification gain
and the flux gain
Brightness Gain = Minification gain x Flux
gain
Conversion factor
the ratio of the illumination intensity at the
output phosphor, measured in candela per
meter squared (cd/m^2) to the radiation
intensity incident on the input phosphor,
measured in milligray per second (mGy/s)
approximately 0.01 times the brightness gain.
the proper quantity for expressing image
intensification.
Conversion factor = ("Output phosphor
illumination" (cd/m^2 ))/"Input exposure
rate (mGy /s) "