You are on page 1of 4

CLINICAL CHEMISTRY LECTURE

To know if the method is reliable and he med tech is


meticulous, the values obtained will be reasonably
LESSON 3: Quality Control and Measurements close together, but will not be identical.
CENTER OF MEASUREMENT *- suppose to have a bell shaped curved called
GAUSSIAN DISTRIBUTION, after Johann Karl
⮚ Mean
F. Gauss who developed many of these basic
⮚ Median concepts draw a bell shaped curve= the peak of the
curved should represent the mean value for the data
⮚ Mode set
if the distribution of values or set is equally
distributed in each side of the graph (symmetrical
MEAN curve)
when the curve is non symmetrical the value of
the mean is still valid, but other descriptive term
RELATIONSHIPS OF DATA will be used.
⮚ Range - largest & smallest value in data
ACCURACY
⮚ Standard Deviation - represent the average
distance from the center of the data and every ⮚ How close the answer obtained lies to the
data set “true” value (*- not obtained by statistical
method)
⮚ Coefficient variation - allows to compare SDs ⮚ Defined by comparison of measured value to
with different unit the known value obtained from the reference
material (**- other refer to it as how close it is
to the bull’s eye-target)
⮚ The closer value to the target, the more accurate
the result

PRECISION
⮚ Can be defined numerically
● Indicates how close the single value are
to one another
GAUSSIAN ● Expressed by standard deviation
● The smaller the SD the more precise
the values are
● Coefficient variation (*- parameter is
expressed as percentage - can be
DISTRIBUTION monitored to assess consistency in
precision of an assay)
● Accuracy & precision are independent
⮚ AKA normal distribution of one another
⮚ It has a symmetric ● Main goal of quality analysis is to
distribution, which means achieve a high level of both precision
that half of the values fall and accuracy in the set of
from the left of the mean measurements
and the right of the mean NOTE: Usually lies on one side of
“bell curve the graph
⮚ X-axis : values obtained
⮚ Y-axis : frequency*
GUGUA NATIONAL COLLEGES- BSMT3
QUALITY CONTROL (QC) IN CC Includes: Mislabeling the sample, pipetting errors,
improper mixing of sample and rgt, voltage
⮚ QC- process of monitoring laboratory analyses fluctuations and temperature fluctuations*
to ensure accuracy of results
● Monitor test performance, helps ⮚ Systematic - seen as trend in data**
identify problems with specific assay
system, and help health care staff violations includes: 2(2S), 4(1S) and 10(x)
assess the reliability of results Includes: Improper calibration, deterioration of
● Encompasses every aspect of reagents, sample instability, instruments drift, or
laboratory operation from initial px changes in standard material
identification, sample acquisition to
the clear reporting of final laboratory *violations includes: 1 (2S), 1 (3S), and R(4S)
results
QC in laboratory is the responsibility of to assess the situation, the sample is
laboratory data, physicians, nurses and other reassayed again using the same reagents, if a
health care providers required by the random error occurred, the same mistake may not
government. A good quality control program happen again and result will be w/in appropriate
is a part of quality assurance, w/c control limits
encompasses tge
**- control values gradually rise or fall
QC MATERIALS violations includes: 2(2S), 4(1S) and 10(x)
⮚ Sources Pool: human or animal
if reassayed does not correct the problem by
● Human: more expensive vs Bovine
bringing the control values in +/-2SD, further
serum (why?)*
analysis is necessary.
⮚ Levels: one and two
● One - somewhere in “normal range” preparing new control materials,
● Tw o- values that are expected in restandardizing the assay, checking the wavelength
common diseases that involves the or other instrument settings, or making new rgts
particular constituents
⮚ Assayed or unassayed** rgst near to their expiration dates should be
*-resembles the properties of px sample discarded and remade, the lab records, for
being analyzed, particular in enzyme assays calibration or schedule of calibration may point to a
**- assayed- more expensive need for a new standard curve. Good documentation
values falls w/in 2 +/- SD; there are will permit easier correction of the problem in the
specific methods and instruments; suited for future
test that are run infrequently
unassayed- usually have targets values, WESTGARD RULES FOR QC
w/c results falling somewhere near the target
✔ 1 (2S) - one control value exceeds +/- 2SD
values; performed in the test that are
from the mean
performed frequently
statistics can be gathered, ✔ 1(3S) - one control value exceeds +/- 3SD from
laboratories are abled to compare their data the mean
w/ other lab
✔ 2 (2S) - two consecutive values exceed the
TYPES OF ERROR same limit +/- 2SD (*- the rule can apply on
two different control materials in the same run,
⮚ Random – no tend or no means of predicting it or to two successive analyses of the same
violations includes: 1 (2S), 1 (3S), and R(4S) control material)
✔ R (4S) - one control value exceeds +2SD and
the other is -2SD (**- one material could show
GUGUA NATIONAL COLLEGES- BSMT3
a value greater than -2SD while the other may - anatomic pathology and pulmonary
provide a value in excess of +2SD) disease.
✔ 4 (1S) - four consecutive control values exceeds
either +1 SD or -1SD (***- the values must all
be in the same direction) STRICT REQUIREMENTS FOR
PROFICENCY TESTING
✔ 10 (x) - ten consecutive control results all lie on
the same side of the mean ⮚ The laboratory must incorporate proficiency
testing into its routine workflow
NOTE: THIS FOR REANALYZED DATA
⮚ Test values/samples must not be shared with
PROFICIENCY TESTING other laboratory during the testing cycle
⮚ Proficiency test - method used to validate a ⮚ Proficiency samples are tested by bench
particular measurement process technical staff who normally conduct patient
testing
⮚ Proficiency samples - spx with known
concentrations of an analyte for the test of ⮚ Testing should be completed within the usual
interest time it would take for routine patient testing
⮚ Samples are analyzed the same manner as px REFERENCE INTERVALS
spx and the result are reported to the
proficiency program ⮚ Establish by scientific community or
manufacturers (*International Federation of
REFERENCE LABORATORIES IN PHIL Clinical Chemistry (IFCC) has recommended
the used of reference range for the common
⮚ RESEARCH INSTITUTE FOR TROPICAL laboratory limits)
MEDICINE
⮚ Establishing a reference interval - no existing
- dengue, influenza, TB and other analyte or methodology in the clinical or
mycobacteria, malaria and other parasites, reference laboratory
bacterial enteric dse, measles and other viral
exanthems, mycology, enteroviruses, ● Costly; may require to test 120-700
antimicrobial resistance and engineering individuals
disease.
⮚ Verifying a reference interval (transference)
⮚ NATIONAL KIDNEY AND TRANSPLANT - confirm the validity of an existing reference
INSTITUTE interval for analyte using the same method and
instrument
- hematology, immunohematology,
immunopathology and Anatomic pathology ● Requires at least 20 individuals
⮚ SAN LAZARO HOSPITAL (SACCL) ESTABLISHING REFERENCE INTERVAL
- HIV/AIDS, hepatitis, syphilis and ⮚ Define an appropriate list of biological
other STI’s variations and analytic interference for medical
literature
⮚ EAST AVENUE MEDICAL CENTER
⮚ Choose selection and partition criteria
- environmental and occupational
health; toxicology and micronutrient assay ⮚ Complete a written consent form and
questionnaire to capture selection criteria
⮚ LUNG CENTER OF THE PHILIPPINES

GUGUA NATIONAL COLLEGES- BSMT3


⮚ Categorize the potential reference individuals Stress Interference
based on the questionnaire findings
⮚ Exclude individuals from the reference sample Food/beverage recovery
group based on exclusion criteria ingestion

⮚ Define the number of reference individuals in


consideration of desired confidence limits and
statistical accuracy
⮚ Standardize collection and analysis of reference
specimens for the measurement of given
analyte
⮚ Inspect the reference value data and prepare a
histogram to evaluate distribution of data
⮚ Identify possible data errors or outliers and then
analyze the reference values
⮚ Document all of the previously mentioned steps
and procedure

SSELECTION OF REFERENCE INTERVAL


STUDY INDIVIDUALS
⮚ Requires defining the detailed
inclusion/exclusion criteria ; age and gender
⮚ Exactly define healthy/ normal donor for
associated reference values
⮚ Donor health status and subgroups in
individuals

PREANALYTIC AND ANALYTIC


CONSIDERATION
PREANALYTIC ANALYTIC
FACTORS FACTORS

Subject Precision
preparation

Prescription Accuracy
medications

Collection time Lot to lot


reagents

Sample storage Linearity

GUGUA NATIONAL COLLEGES- BSMT3

You might also like