Standard Method of Test
for
Determining pH of Soil for Use in Corrosion Testing
1. SCOPE
L.A This test method describes pro-
cedures and apparatus for determining
pH value for corrosion testing by use of
pH meter,
1.2 The principal use of the test is
to supplement soil-resistivity measure
ments and thereby identify conditions
under which the corrosion of metals in
the soil may be sharply accentuated, This
standard is divided into two parts, The
first part involves obtaining and prepar-
ing the sample to size for testing. The
second part describes the test method for
determining the pH of soil
13° The values stated in SI units are
to be regarded as the standard
2. REFERENCED DOCUMENTS
21 AASHTO Standards:
T2 Sampling Aggregates
7248 Reducing Field Sam-
ples of Aggregates to
‘Testing Size
PART I
INITIAL PREPARATION OF
‘TEST SAMPLES
3. SCOPE
3.1 This method covers the dry prep-
aration of soil and soil-aggregate sam-
ples, as received from the field, for de-
termining a soil’s pH.
32. The following applies to all spec-
iffed limits in this standard: For the pur-
pose of determining conformance with
these specifications, an observed value
or calculated value shall be rounded off
“to the nearest unit” in the last right
hhand place of figures used in expressing
the limiting value, in accordance with
the rounding-off method R. 11, Recom:
AASHTO DESIGNATION T 289-91
mended Practice For Indicating Which
Places Of Figures Are To Be Considered
Significant In Specified Limiting Values.
4. APPARATUS.
4.1 Sieves—A series of sieves ofthe
following sizes: 63 mm (‘4 in.), 4.75
mm (No. 4), 2.00 mm (No. 10) and 2
ppan. The sieve shall conform to M 92,
Sieves for Testing Purposes (Note 1).
42. A scale, minimum capacity of
500 grams. The balance shall conform
to the requirements of M 231, for the
class of general purpose balance required
for the principal sample.
43 Drying Apparatus—Any_ suit-
able device capable of drying samples
at a temperature not exceeding 60°C
(40°F).
44 Pulverizing Apparatus—Either
mortar and rubber-covered pestle or any
device suitable for breaking up the aggre-
gations of soil particles without reducing
the size of the individual grains (Note 2).
45 Sample Spliter—A suitable rif-
fle sampler or sample spliter for propor-
tional splitting of the sample and capable
of obtaining representative portions of
the sample without appreciable loss of
fines. The width of the container used
to feed the rifle sampler spliter should
be equal fo the total combined width of
the riffle chutes. Proportional splitting
fof the sample on a canvas cloth is
also permitted.
NOTE 1—The sieve sizes which have an
opening size of 63 mm (¥, in) or lager
Shall conform to the requirements specified
in M 92 excluding column no. 7. The excl
sion of column 7 permits the use of heavier
soreens in nonstandard frames which are
NOTE 2—Other types of apparatus are
satisfactory if the aggregations of soll part
les sre broken up without reducing the size
ofthe indivigual rains
5S. SAMPLE SIZE
SAA_ The amount of soil material re-
quired to perform the testis as follows:
Sieve Size
Finer Than—
72.00 mm (No. 10)
Test Approx. Mass (9)
pu 100
6. INITIAL PREPARATION OF
‘TEST SAMPLES
6.1_The sample as received from the
field may be dried in air or a drying
apparatus not exceeding 60°C (140°F)
prior to sample selection (Note 3). It is
Fecommended thatthe soil be in a moist
condition for pH testing purposes. A rep-
resentative test sample to perform the
pH test shall then be obtained with the
sampler, or by splitting or quartering as
per T 248,
NOTE 3—Samples dried in an oven oF
other drying apparatus at a temperate not
exceeding 60°C (140°F) ae considered to be
air dred.
62. The portion of the sample se-
lected for pH testing shall be separated
into fractions by one of the following
methods:
62.1 Alternate Method Using 2.00-
mm (No. 10) Sieve—The dried saraple
shall be separated into two fractions us-
ing a 2.00-mm sieve. The fraction re-
tained on the sieve shall be ground with
pulverizing apparatus until the aggrega-
tions of soil particles are broken into
Separate grains. The ground soil shall
then be separated into two fractions using
the 2.00-mm sieve,
622 Alternate Method Using 4.75-
‘mm and 2.00-mm (No. 4 and No. 10)
Sieves—The dried sample shal first be
separated into two fractions using a 4.75-
‘mm sieve, The fraction retained on this
sieve shall be ground with a pulverizing
‘apparatus until the aggregations of soil
167768,
panicles are broken into separate grains
and again separated on the 4.75-mm
sieve. The fraction passing the 4.75-mm
sieve shall be mixed thoroughly and, by
the use of the sampler or by splitting
and quartering, a representative portion
adequate for testing shall be obtained,
‘This split-off portion shall then be sepa-
rated on the 2.00-mm sieve, and pro-
‘cessed as in Section 6.2.1
623 Altemate Method Using 63-
‘mm and 2.00-mm (‘/cin. and No. 10)
Sieves—The dried sample shall first be
separated into two fractions using a 6.3-
mm sieve. The fraction retained on this
sieve shall be ground with a pulverizing
apparatus until the aggregations of soil
panicles are broken into separate grains
and again separated on the 6.3-mm sieve,
‘The fraction passing the 6.3-mm sieve
shall be mixed thoroughly and, by the
Use of the sampler or by splitting and
quartering, a representative portion ade-
‘quate for testing shall be obtained. This
splitoff portion shall then be separated
fon the 2,00-man sieve, and processed as
in Section 6.2.1,
PART 2
DETERMINATION OF SOIL pH
SCOPE
7A This section describes the proce-
dure for determining soil pH,
8 APPARATUS
8.1 pH Meter—suitable for labora-
{ory or field analysis, with either one oF
two electrodes.
82 A -50-mL. wide-mouth glass bea-
ker or other suitable container with a
watch glass for cover. If lightweight ma-
terial is to be tested, it may be necessary
to increase the beaker size up to a maxi
‘mum of 250 mL,
83. Standard Buffer Solutions of
Known pH Values—standards to be used
are pH of 4.0, 7.0, 1000.
METHODS OF SAMPLING AND TESTING
84 Distilled water.
8S A teaspoon or small scoop.
8.6 A thermometer capable of read:
ing 25 * 10°C, to the nearest 0.1°C.
8.7 Sieves—A 2.00-mm (No. 10) for
preparing the sample and a pan. The
sieve shall conform to M 92, Sieves for
Testing Purposes.
88° A glass stirring rod,
89° A scale, minimum capacity of
500 grams. The balance shall conform
to the requirements of M 231, for the
class of general purpose balance required
for the principal sample,
PROCEDURE
9.1 Of the material selected for test-
ing, place a mass of 30.0 * 0.1 grams
of soil into the glass beaker or other
suitable container.
9.2 Add 30.0 * 0.1 grams of dis-
tilled water to the soil sample, Stir to
‘obtain a soil slurry and then cover with
2 watch glass,
9.3. The sample must stand for a min-
‘mum of one hour, stirring every 10 to
15 minutes. This is to allow the pH of
the soil slurry to stabilize,
94 Measure the temperature of the
sample and adjust the temperature con
twoller of the pH meter to that of the
sample temperature. This adjustment
should be done just prior to testing. On
‘meters with an automatic temperature
contol, follow the manufacturer's in-
structions.
95 Standardize the pH meter by
means of the standard solutions provided.
‘Temperature and adjustments must be
performed as stated under Section 9.4
(See Note 4).
9.6 Immediately before immersing
the electrode(s) into the sample, stir well
with a glass rod. Place the electrode(s)
into the soil slurry solution and gently
tum the beaker or container to make
{good contact between the solution and
the electrode(s). DO NOT place the elec-
twode(s) into the soil; only into the soil
slurry solution, (See Note 5).
9.7 The electrode(s) require immer-
sion 30 seconds or longer in the sample
before reading to allow the meter to
T 289
stabilize. If the meter has an auto read
system, it will automatically signal
When stabilized,
9.8 Read and record the pH value
to the nearest tenth of a whole number.
Wf the pH meter reads to the hundredth
Place it is necessary to round off the
result in accordance with the rounding.
off method RIL
99 Rinse off the electrode(s) well
with distilled water, then dab lightly with
tissues to remove any film formed on
the electrode(s). Caution: Do not wipe
the electrode(s) as this may result in
polarization of the electrode(s) and con-
sequent slow response. (See Note 6)
NOTE 4—To standardize the pH meter,
tse the 7.0 pH buffer standard solution plus
the other standard solution which is nearest
the estimated pH value ofthe sample to be
‘ested, I the manufacturers instvetions indi
‘ate a method other than that noted above,
then those instructions must be followed,
NOTE $—When immersing electode(s)
into the glass beaker or container car should
bbe taken not to hit the Bottom or side, causing
damage 10 electode(s),
NOTE 6—1f polarization does occur, as
indicated by a slow response, rinse the elee-
lwode(s) and dab Tighuy spain
10. PRECAUTIONS
10.1 Periodically check for damage
to electrode(s).
10.2 Electrode tip should be kept
‘moist during storage. Follow the manu-
facturer’s instructions,
1. REPORT
ALI As specified in Section 9.8, re-
port the pH value to the nearest tenth
of a whole number in accordance with
the rounding-off method AASHTO R 11
12, PRECISION AND BIAS
121 Data is not available at this
time,