Professional Documents
Culture Documents
Residents’ Section
Physics Minimodule
Downloaded from www.ajronline.org by 114.125.217.121 on 01/30/23 from IP address 114.125.217.121. Copyright ARRS. For personal use only; all rights reserved
Residents
T
he purpose of most radiologic im- (kilovoltage and tube current–exposure time
R. Brad Abrahams 2 aging examinations is to either product). In this article, we describe the es-
identify abnormalities in a patient sential characteristics of spatial resolution.
Huda W, Abrahams RB or to classify a patient as being Factors that influence spatial resolution, as
healthy while minimizing radiation exposure well as performance of x-ray-based medical
[1]. To achieve these goals, it is of obvious im- imaging modalities, are described [4–7].
portance to ensure that the radiologic image
quality is sufficient for a given imaging task. What Is Resolution?
Medical image quality is normally character- Blur and Sharpness
ized in terms of contrast, noise, and spatial Most radiologists have an intuitive sense
resolution. The achieved quality for any im- of what constitutes poor resolution and can
age will depend on both the intrinsic proper- easily recognize a blurred image. Resolu-
ties of the imaging system and the manner in tion, however, is sometimes taken to mean
which the images were obtained, including the ability to see objects so that when lesions
the choices of image acquisition tube voltage are seen, then the resolution must be good.
(kilovoltage) and output (tube current–expo- Such statements are incorrect, as illustrated
sure time product). by considering a single-cell molecule labeled
To be detectable, a lesion must transmit a with a radionuclide and imaged with a PET
different x-ray beam intensity compared to system. The activity will be seen, but the blur
the surrounding normal tissues. The differ- associated with this point source will be sev-
ence in x-ray beam intensity through a lesion eral millimeters, resulting in a blurry image
relative to the normal background is called with poor resolution.
subject contrast. The appearance of this sub- This example can be expanded to illus-
ject contrast in the resultant image is called trate what imaging scientists mean regard-
Keywords: focal blur, image quality, motion blur, image contrast. Noise has both random and ing spatial resolution performance. For two
radiographic imaging, receptor blur, resolution structural components, with the latter being nearby cells with radioactively labeled DNA
fixed for a given patient, such as ribs in chest in their nuclei, a system that can resolve
DOI:10.2214/AJR.14.13126
radiographs. The random noise in virtual- these two entities would show two hot spots
Received May 9, 2014; accepted after revision ly all radiographic imaging is due to quan- as distinct entities (Fig. 1, upper panel). A
May 30, 2014. tum mottle, which can be reduced by using PET system would show only one very large
more x-ray photons to create the image, but blurred object, and the activity in cell one
1
Department of Radiology and Radiological Science, which will also increase the patient dose [2]. would not be resolved from the activity de-
Medical University of South Carolina, 96 Jonathan Lucas
Contrast and noise are directly related to the picted in the nearby cell two (Fig. 1, lower
St, MSC 23, Charleston, SC 29425-3230. Address
correspondence to W. Huda (walterhuda@hotmail.com). choice of radiographic techniques, namely panel). A system with good resolution per-
x-ray tube voltage (kilovoltage) and output mits the detection (and characterization) of
2
Department of Radiology and Imaging, Georgia Regents (tube current–exposure time product) [3]. two small features, such as microcalcifica-
University, Augusta GA. The amount of blur, or unsharpness, af- tions in mammograms, even when they are
WEB
fects the appearance of both normal anat- physically close to each other.
This is a web exclusive article. omy and pathologic abnormalities. Spatial There are several terms used to describe
resolution is the technical term used to re- spatial resolution in imaging. Blur and sharp-
AJR 2015; 204:W393–W397 fer to the amount of blur in an image. Spatial ness are good descriptors that are universal-
resolution performance is an intrinsic prop- ly understood, and they simply mean that
0361–803X/15/2044–W393
erty of an imaging system that is generally a sharp edge will also appear as sharp (not
© American Roentgen Ray Society independent of the selected technique factors blurred) in an image obtained with a system
Fig. 2—Illustration on
left depicts radiographic
image being generated
of narrow line of x-rays,
such as might result when
x-ray passes through long
narrow slit in sheet of lead.
Downloaded from www.ajronline.org by 114.125.217.121 on 01/30/23 from IP address 114.125.217.121. Copyright ARRS. For personal use only; all rights reserved
Fig. 5—For scintillators that absorb x-rays and produce light, amount of image
blur is directly related to thickness of scintillator (1 mm [left] vs 0.5 mm [right]).
Thickness of scintillator can be taken to be comparable to blur introduced into
resultant image. (Illustration by Abrahams RB)
Fig. 6—Illustration of photoconductor shows Fig. 7—Photostimulable phosphor shows scattering of incident read-out narrow light beam (left) and
that charge (electrons) generated by incident emission of stimulated light from pixel of interest (right), as well as adjacent pixels, which increase image blur.
photons have minimal divergence when collected (Illustration by Abrahams RB)
by application of voltage across photoconductor.
(Illustration by Abrahams RB)
essential. By contrast, in computed radiogra- fore be obtained using smaller cassettes. The mance. With a standard 500-line television
phy, light beams are used to release the energy limiting resolution of a typical film-screen system, the nominal limiting fluoroscopy
stored in an exposed photostimulable phosphor. system used for chest radiography was about spatial resolution performance is about one
As shown in Figure 7, there will be scattering of six line pairs per millimeter, or a factor of 2 line pair per millimeter. Flat-panel detec-
the incident light, which will result in increased better than a typical digital chest radiograph- tors with a pixel size of 175 μm can achieve
image blur because of the release of light from ic imaging system. a limiting resolution of three line pairs per
adjacent pixels. For this reason, there is a limit In digital mammography, the pixel size millimeter. When these are used for per-
to the thickness of any photostimulable phos- typically ranges between 50 and 100 μm, so forming fluoroscopy, binning four pixels into
phor. When thin detectors are used, this in- that the corresponding limiting spatial reso- one large pixel is common for larger FOVs,
creases patient doses because of the reduction lution ranges between 10 and five line pairs which results in a limiting resolution of 1.5
in x-ray absorption in thinner detectors. per millimeter. A common pixel size of 70 μm line pairs per millimeter.
In fluoroscopy, the number of television corresponds to a limiting resolution of close Digital photospot images using an image
lines determines the nominal width of each to seven line pairs per millimeter. The total intensifier–based imaging chain are general-
line and thereby affects the achievable spa- number of pixels in a digital mammogram is ly obtained using a high-quality (1000 line)
tial resolution performance. When the num- about 12 million, assuming a 3000 × 4000 television system. For a 25-cm FOV, the re-
ber of television lines increases from 500 to matrix size. Most mammography worksta- sultant pixel size is 250 μm, and the corre-
1000, the limiting resolution performance tions are capable of displaying up to 5 million sponding limiting resolution will be about
will also improve by a factor of 2. In CT, it is pixels, so that seeing all the available detail in two line pairs per millimeter. Commercial
the physical size of each detector (length and a digital mammogram will require the use of 2000-line television systems are available,
width) that is directly related to the achiev- image zoom capabilities [19]. but the benefits of providing improved spa-
able spatial resolution performance [18]. Film-screen mammography typically tial resolution performance were not deemed
When corrected for image magnification, the achieved a limiting spatial resolution of 15 to be worthwhile. For flat-panel detectors,
nominal detector dimension in both direc- line pairs per millimeter, which is a factor spatial resolution for digital photospot im-
tions on a CT scanner may be taken to be a of 2 better than digital mammography. This ages is identical to that for radiographs ob-
nominal 0.5 mm or so. example illustrates that the benefits of digi- tained with these detectors—namely, three
tal mammography relate to image processing line pairs per millimeter.
Resolution Performance in rather than any resolution issue per se. Mam-
Radiographic Imaging mography film-screen systems are superior to CT
Radiography and Mammography the capabilities of the human visual system, In a head CT scan, the FOV is 250 mm,
Digital radiographs generally have a ma- which is about five line pairs per millimeter at the matrix size is 512 × 512 and the pixel
trix size of 2000 × 2500. For a 35 × 43-cm a 25-cm viewing distance [19]. This explains size is 0.5 mm. The best achievable resolu-
cassette, this digital matrix corresponds to a why mammographers in the days of film al- tion for this FOV and matrix size is thus one
pixel size of 175 μm and a limiting resolution ways had a magnifying glass in their hands for line pair per millimeter. Increasing the FOV
of three line pairs per millimeter. For a 20 × viewing mammograms on viewboxes. to 500 mm for a large patient doubles the pix-
24-cm cassette, the pixel size would be 100 el size (1 mm) and halves the limiting spa-
μm, and the limiting resolution would be five Fluoroscopy and Digital Photospot tial resolution performance (i.e., 0.5 line pair
line pairs per millimeter. Extremity radio- Fluoroscopy can be performed with image per millimeter). Spatial resolution in CT is
graphs, where good resolution is important intensifiers or flat-panel detectors and does thus an order of magnitude worse than film-
for detecting hairline fractures, should there- not generally require high-resolution perfor- screen combination and four times worse
TABLE 1: Representative Values of Limiting Resolution Performances for Digital Imaging Systems
Limiting Resolution,
Line Pairs per
Imaging System Millimeter Comments
CT ≈ 0.7 Improved by zoom reconstruction of the central region of acquired image
Downloaded from www.ajronline.org by 114.125.217.121 on 01/30/23 from IP address 114.125.217.121. Copyright ARRS. For personal use only; all rights reserved
Fluoroscopy ≈1 Improved by the use of electron magnification modes, as well as high-quality television
systems (1000 line)
Digital subtraction angiography or photospot ≈2 Improved by the use of electronic magnification modes of image intensifiers (e.g.,
magnification 1 or 2)
Radiography ≈3 Improved performance can be achieved by the use of smaller cassette sizes (5 line pairs
per millimeter for 20 × 25 cm)
Mammography ≈7 Use of smaller (50 μm) pixels would improve resolution, whereas large pixels (100 μm)
would degrade resolution
than digital chest radiography. CT has ex- that is present in the object. The key factors 8. Schueler BA. Clinical applications of ba-
cellent imaging characteristics and anatomic that influence the sharpness of an image re- sic x-ray physics principles. RadioGraphics 1998;
localization compared with any type of pro- late to the size of the source of x-rays (fo- 18:731–744; quiz, 729
jection imaging but also has inferior spatial cal spot), the physical characteristics of the 9. Villafana T. Generators, x-ray tubes, and expo-
resolution performance. x-ray detector system (area and thickness), sure geometry in mammography. RadioGraphics
In CT, the choice of the mathematic recon- and the presence of any motion blur because 1990; 10:539–554
struction filter affects the amount of blur in the of the finite duration of all radiographic ex- 10. Zink FE. X-ray tubes. RadioGraphics 1997;
resultant image [20]. Image reconstruction fil- posures. Image processing, especially in CT 17:1259–1268
ters offer reduced image noise but at the ex- through the filtered back projection recon- 11. Hogge JP, Palmer CH, Muller CC, et al. Quality
pense of more image blur. Common names for struction, can also affect spatial resolution assurance in mammography: artifact analysis.
such high-resolution reconstruction filters in- performance. The limiting resolution that is RadioGraphics 1999; 19:503–522
clude “bone,” “detail,” “high resolution,” and routinely achieved in clinical practice ranges 12. Ayyala RS, Chorlton M, Behrman RH, Kornguth
“lung.” These filters are used to get excellent from about 0.7 line pair per millimeter for PJ, Slanetz PJ. Digital mammographic artifacts on
detail for structures such as bone (vs air or tis- CT to about seven line pairs per millimeter full-field systems: what are they and how do I fix
sue) or lung (vs air), where the high intrinsic for digital mammography (Table 1). The val- them? RadioGraphics 2008; 28:1999–2008
contrast negates the importance of higher lev- ues listed in Table 1 may be compared to the 13. Barrett JF, Keat N. Artifacts in CT: recognition and
els of noise. Reconstruction filters with names human visual system, which can resolve up avoidance. RadioGraphics 2004; 24:1679–1691
such as “standard” or “soft tissue” reduce im- to five line pairs per millimeter at a normal 14. Pannu HK, Alvarez W Jr, Fishman EK. Beta-
age noise at the price of inferior spatial reso- viewing distance of 25 cm. blockers for cardiac CT: a primer for the radiolo-
lution performance and are used where the in- gist. AJR 2006; 186(suppl 2):S341–S345
trinsic lesion contrast is low. References 15. Macias CG, Chumpitazi CE. Sedation and anes-
CT spatial resolution can be improved by 1. Huda W. Understanding (and explaining) imag- thesia for CT: emerging issues for providing high-
up to a factor of about 2 by reducing the re- ing performance metrics. AJR 2014; 203:[web] quality care. Pediatr Radiol 2011; 41(suppl
construction FOV (zoom). This requires that W1–W2 2):517–522
the original projection data are used to re- 2. Huda W. Kerma area product in diagnostic radiol- 16. Haus AG. The AAPM/RSNA physics tutorial for
construct a smaller anatomic region, which ogy. AJR 2014; 203:[web]W565–W569 residents: measures of screen-film performance.
thereby differs from simple magnification, 3. Huda W, Abrahams RA. Radiographic tech- RadioGraphics 1996; 16:1165–1181
which uses reconstructed image data. Ob- niques, contrast, and noise in x-ray imaging. AJR 17. Rowlands JA, Zhao W, Blevis IM, Waechter DF,
taining the best possible resolution may also 2015; 204:W126–W131 Huang Z. Flat-panel digital radiology with amor-
require the use of a reduced focal spot (0.6 4. Bushberg JT, Seibert JA, Leidholdt EM, Boone phous selenium and active-matrix readout. Radio-
vs 1.2 mm), which will also require a reduc- JM. The essential physics of medical imaging, 3rd Graphics 1997; 17:753–760
tion of the focal spot power loading (25 vs ed. Philadelphia, PA: Lippincott, Williams & 18. Sprawls P. AAPM tutorial: CT image detail and
100 kW). However, when a small focal spot Wilkins, 2012 noise. RadioGraphics 1992; 12:1041–1046
is used with reduced power loading, the scan 5. Huda W. Review of radiologic physics, 3rd ed. 19. Samei E. AAPM/RSNA physics tutorial for resi-
time may need to be increased, which also Philadelphia, PA: Lippincott Williams & Wilkins, dents: technological and psychophysical consider-
increases the likelihood of motion blur. 2010 ations for digital mammographic displays. Radio-
6. Nickoloff EL. Radiology review: radiological physics. Graphics 2005; 25:491–501
Conclusion Philadelphia, PA: Elsevier Saunders, 2005 20. Goldman LW. Principles of CT: radiation dose
Resolution is the ability of an imaging 7. Wolbarst AB. Physics of radiology, 2nd ed. Madi- and image quality. J Nucl Med Technol 2007;
system to faithfully reproduce a sharp edge son, WI: Medical Physics Publishing, 2005 35:213–225; quiz, 226–218