You are on page 1of 14

Medical & Biological Engineering & Computing (2022) 60:1497–1510

https://doi.org/10.1007/s11517-022-02548-6

ORIGINAL ARTICLE

Total hip replacement monitoring: numerical models for the acoustic


emission technique
Remya Ampadi Ramachandran1,2 · Christine Lee1,3 · Lu Zhang4,5 · Supriya M. H2 · Divya Bijukumar3 · P. Srinivasa Pai6 ·
Kharma Foucher7 · Sheng‑Wei Chi4 · Didem Ozevin4 · Mathew T. Mathew1,3

Received: 26 June 2021 / Accepted: 6 March 2022 / Published online: 21 March 2022
© International Federation for Medical and Biological Engineering 2022

Abstract
Any mechanical instability associated with total hip replacement (THR) excites elastic waves with different frequencies and
propagates through the surrounding biological layers. Using the acoustic emission (AE) technique as a THR monitoring tool
provides valuable information on structural degradations associated with these implants. However, several factors can com-
promise the reliability of the signals detected by AE sensors, such as attenuation of the detected signal due to the presence of
biological layers in the human body between prosthesis (THR) and AE sensor. The main objective of this study is to develop
a numerical model of THR that evaluates the impact of biological layer thicknesses on AE signal propagation. Adipose tis-
sue thickness, which varies the most between patients, was modeled at two different thicknesses 40 mm and 70 mm, while
the muscle and skin thicknesses were kept to a constant value. The proposed models were tested at different micromotions
of 2 µm, 15–20 µm at modular junctions, and different frequencies of 10–60 kHz. Attenuation of signal is observed to be
more with an increase in the selected boundary conditions along with an increase in distance the signals propagate through.
Thereby, the numerical observations drawn on each interface helped to simulate the effect of tissue thicknesses and their
impact on the attenuation of elastic wave propagation to the AE receiver sensor.

Keywords  Total hip replacements (THR) · Non-invasive techniques · Acoustic emission (AE) · Finite element (FE)
analysis · Numerical modeling

1 Introduction
* Remya Ampadi Ramachandran
rampad2@uic.edu The acoustic emission (AE) technique, one of the non-
invasive testing methods, offers the competence to moni-
* Sheng‑Wei Chi
swchi@uic.edu tor the structural degradations of orthopedic implants both
passively and in real-time [1, 2]. Numerous studies are
1
Department of Biomedical Engineering, University reported in the literature on structural monitoring of total
of Illinois Colleges of Engineering and Medicine, Chicago, hip replacement (THR) implants by the AE technique [3].
IL, USA
Though several factors can compromise the signals detected
2
Department of Electronics, Cochin University of Science by the AE sensors, the presence of several layers of soft tis-
and Technology, Kochi, India
sue with different thicknesses and densities between the hip
3
RMDR Lab, Department of Biomedical Sciences, University implant and sensor can affect the wave propagated from the
of Illinois College of Medicine, Rockford, IL, USA
source to a great extent [4, 5]. Consequently, the initial wave
4
Department of Civil, Materials, and Environmental at the implant becomes distorted by the time it is detected
Engineering, University of Illinois, Chicago, IL, USA
by the AE sensor placed on the skin. As the wave traverses
5
College of Civil Engineering and Architecture, Guilin through different thicknesses, it can become attenuated and
University of Technology, Guilin, China
dispersed before it reaches the AE receiver sensor or skin
6
Department of Mechanical Engineering, NMAM Institute surface. Besides, the wave may also pick up noises as it
of Technology, Nitte, Karnataka, India
propagates through different layers, affecting the magnitude
7
Department of Kinesiology and Nutrition, University of the wave received by the sensor.
of Illinois, Chicago, IL, USA

13
Vol.:(0123456789)
1498 Medical & Biological Engineering & Computing (2022) 60:1497–1510

Gao et al. [6] identified a function to determine the result- Development and modeling of hip prosthesis mainly con-
ing AE signal after it propagates through biological layers sider the mechanical aspects, an FE modeling study per-
between the implant and sensor; where the pathway of AE formed by Malau et al. [25] considers stress analysis towards
signal attenuation is determined by the transfer function the development of porous stemmed hip implant for the
(H) of the whole propagation route obtained by dividing the pediatric patients, and Çelik et al. [26] considers mechani-
detected AE signal (G) from the initial/source AE signal (F) cal stability as well as reduced risk of aseptic loosening
in its frequency domain. That work also shows that the layers towards hip prosthesis development. FE modeling accurately
between a total hip replacement (THR) to the sensor to be in depicts what happens in the real world by simply mocking
the order of a connective part (α), soft tissue, bone, perios- up the environment and applying the physical effect to the
teum, subcutaneous tissue, skin, and a connective part that model with the help of modeling software such as COMSOL
attaches the sensor to the skin with Tn representing the sensi- Multiphysics [27]. Chen et al. [28] consider the FE model
tivity of the receiving AE sensor. Thus, the transfer function approach to analyze different restoration practices of Van-
(H) of the whole propagation route (Eq. 1) can be obtained couver B1 periprosthetic fractures that occurred after THR
by multiplying transfer functions associated with these lay- surgery and identified a better method with reasonable fixa-
ers, where Ha ;a = 1, ..., n, corresponds to the transfer func- tion power to treat the fracture.
tion of each of the above-mentioned layers, respectively. Since FE models suggested so far in literature mainly con-
sider the stress and structural analysis to determine implant
G
failure, here comes the significance of the devised study.

H=
F
= Tn ⋅ Ha (1)
a=1 The proposed scheme is a proof-of-concept study involv-
The influence of biological layers on wave propagation ing a numerical modeling approach to identify the impact
can be assessed by evaluating the attenuation [7] which of soft tissues on acoustic emission signal attenuation and
mainly arouses due to scattering and absorption of a wave to determine its significance in obtaining reliable AE data.
during its propagation through a medium. Ultrasonic stud- Also, a better understanding can be achieved on how AE
ies on tissues [8] showed that frequency and the propaga- techniques should be altered to cater to patients with varying
tion distance have a direct impact on the attenuation of the tissue thicknesses to accurately monitor hip replacements.
signal, while Pan et al. [9] considers the effect of skin on This study also paves the way for the development of an
wave attenuation. The impact of muscle, adipose tissue, and FE numerical model that addresses the effect of acoustic
skin on signal attenuation was addressed in [10, 11] by plac- nonlinearity, elastic modulus, density characteristics on the
ing sensors both externally on the skin and directly onto propagation of an elastic wave through the biological lay-
the femoral stem implant. Literature also confirms various ers, and post-processing of these simulation results with the
numerical studies on THR evaluation to simulate the implant diagnostic data to provide anatomically rich and clinically
behavior under different conditions. A numerical model sug- relevant visualizations [29].
gested by Li et al. [12] predicts the progressive failure of
thick laminated composite femoral components, whereas
Chethan et al. [13] presents a finite element (FE) analysis 2 Materials and methods
demonstrating its efficiency for fatigue tests to predict the
implant life under dynamic conditions. FE analysis of the 2.1 Basic concepts
vibrational behavior of the prosthesis-femur system was car-
ried out by Pastrav et al. [14] to identify the influence of Longitudinal sound waves travel through any medium,
changing contact conditions on resonance frequency. How- whereas transverse sound waves travel only through a rela-
ever, a two-dimensional FE model investigation identified tively solid medium [30, 31]. Acoustic speed is greater in
the influence of surface roughness parameters and soft tis- more rigid or less compressible materials and is five times
sue thickness on the ultrasonic response of the bone-implant greater in soft tissue and bone when compared to air. When
interface [15]. a sound wave travels through the tissue, a continuous loss
Aseptic loosening of THR implants is more likely to be in its energy can be observed and it is termed attenuation.
linked to its stress shielding behavior [16, 17]. Most of the Divergence, deflection, hysteresis, and absorption are the
FE modeling studies work on loading events to perform main causes of attenuation which mainly arise due to the
implant evaluations such as stress–strain analysis [18–20], distance from the source or viscosity of the propagation
fatigue monitoring [21], and the effect of day-to-day life medium [32]. The proposed study considers the stress
activities [22]. FE models were also employed to assess wave propagation in the human body through different soft
the stress shielding behavior of hip prosthesis [23, 24] and tissue layers before it reaches the skin surface. The main
obtained a clear picture of the implant’s stress distribution. aim of this work is to address the influence of adipose tis-
sue thickness in wave propagation which is initiated at the

13
Medical & Biological Engineering & Computing (2022) 60:1497–1510 1499

implant interface. Since the adipose tissue has a significant


role related to load transfer due to any bodily movements;
its resistance to deformation and its residual lengthening a
tissue maintains on stretching and relaxation must also be
carefully studied [33]. As a baseline study, this numerical
model assumes the selected tissues as linear elastic materi-
als which in turn paves the way to the development of a
linear and then non-linear viscoelastic model that addresses
the viscous nature of adipose tissue and correspondingly its
hysteresis characteristics.
Computational modeling helps to overcome the restric-
tions on analytical approaches such as problems of simple
geometries or non-realistic assumptions [29]. In this study,
FE modeling (FEM) [34, 35] is employed to compute the
displacement field obtained at each interface point [13]. The
Fig. 1  Two-dimensional COMSOL model representing the main bio-
accuracy of an FE model depends on computationally fea-
logical layers between total hip replacement and a AE receiver sensor
sible and adequately fine mesh selection. Numerical disper- placed on the skin surface. P1 represents the muscle-adipose tissue
sion and dissipation errors in a dynamic model can be con- interface point, P2 represents the adipose tissue-skin interface point,
trolled by selecting at least 15–20 elements per wavelength and P3 represents the skin-outer boundary interface point
[36]. The following equations (Eqs. 2, 3, and 4) were consid-
ered in this work to calculate the element size for meshing: selected points. The point evaluations helped to analyze how
the initial wave attenuates as it propagates through materials
𝜆
h≤ (2) with different properties.
n
For the first case, a representative modular junction dis-
h denotes the element size, n denotes the number of ele- placement of 2-µm micromotion is studied and the signal
ments, and 𝜆 denotes the wavelength attenuation and energy loss over the medium are estimated.
Whereas in the second model, the focus is on the initiated
1 signal frequency variation ranging from 10 to 60  kHz.
Δt ≤ (3)
20f Frequencies were selected assuming that the hip implant
Δt denotes the time-step requirement, and f denotes the loosening initiates a wave with a frequency around 20 kHz,
wave frequency. whereas fatigue crack may initiate elastic waves or AE sig-
Also, for a signal with wavelength λ, and frequency f fol- nals with very high frequencies [38, 39]. Adipose tissue
lows the equation: thickness of 40 mm and 70 mm was taken because this fac-
tor fluctuates the most between 40 and 70 mm [40, 41] from
c=
𝜆
orc = 𝜆.f (4) patient to patient. Muscle and skin thickness of 45 mm and
T 10 mm, respectively, were considered seeing that they do not
where c represents the wave speed. vary as significantly as adipose tissue.
The speed of sound in any medium is determined by
the medium’s stiffness and density. In a stiffer medium,
2.2 Model setup the wave will be faster, and if the density ( 𝜌 ) of a medium
is greater, then the wave will be slower [42, 43]. Here, we
As shown in Fig. 1, COMSOL Multiphysics® modeling have considered the same propagation medium but with
software was utilized to create a cross-sectional two-dimen- two different thickness characteristics for adipose tissue
sional (2D) model with three linear elastic layers represent- layer. Apart from the soft tissue thickness parameters,
ing muscle, adipose tissue, and skin [37]. In this numerical other characteristics such as density, bulk modulus, shear
model, elastic wave initiation at the center CoCrMo head, modulus, and thereby Young’s modulus and Poisson’s ratio
where the implant would be located, and its propagation are also important. It will help in mimicking a real-world
through the three selected biological layers of different thick- experimental scenario in this case to simulate the signal
nesses were studied for different conditions such as modu- propagation from the prosthesis through a real biological
lar junction displacement, and frequency wave instigation. medium to the AE receiver sensor. Such material charac-
A point on each interface: muscle and adipose tissue (P1), teristics were selected based on the previously established
adipose tissue and skin (P2), and skin and outer boundary studies as illustrated in Table 1 [38, 44–46], assuming pop-
(P3), were evaluated to analyze the received wave on these ulation age between 40 and 60. Parameters in Table 2 were

13
1500 Medical & Biological Engineering & Computing (2022) 60:1497–1510

Table 1  Material characteristics of the model To deal with computational efficiency, effectiveness, and
Biological layer Bulk modulus Shear modulus Density (kg/m )3 model accuracy, a mesh convergence study has been per-
formed with 22 ≤ n ≤ 16 . Final models considered n = 20
Muscle 0.00329 (GPa) 2.60e − 5 (GPa) 1090 of Eq. 2 as a representative based on the similarity in the
Adipose-tissue 2.0 (GPa) 1.84 (GPa) 911 displacement recorded. Point evaluations of this model are
Skin 2.0 (GPa) 5.43e − 3 (GPa) 1109 defined by associating the interpolation function concerning
Implant metal Young’s modu- Poisson’s ratio Density (g/cm3) the applied frequency parameter which in turn enables the
lus
developed models to reproduce the selected data points for
CoCrMo 240 (GPa) 0.29 8.4
evaluating the wave propagation.
Figure 2 represents the FE 2D-model created with a tis-
sue thickness of 40 mm and mesh statistics corresponding
Table 2  Parameters selected for the model to a micromotion of 2μm at the modular junction specific to
Name Expression Value
20 kHz frequency [48]. Since the selection of mesh size has
a great impact on the quality of the model being developed,
Muscle thickness (th1) 45 (mm) 0.045 m a keen focus must be given to its calculation. Considering
Adipose tissue thickness (th2) 40 (mm) 0.04 m Eq. 2, the element size of the selected mesh has been limited
70 (mm) 0.07 m to guarantee the accuracy and efficiency of the generated
Skin thickness (th3) 10 (mm) 0.01 m model.
Frequencies 10 (kHz) 1E4 Hz
20 (kHz) 2E4 Hz
30 (kHz) 3E4 Hz 3 Results
40 (kHz) 4E4 Hz
50 (kHz) 5E5 Hz One of the best ways to identify the effect of thickness of the
60 (kHz) 6E5 Hz propagation medium on stress wave propagation is to ana-
lyze the attenuation effect of the wave received at desired
points. This attenuation which refers to a decrease in its signal
selected to consider the variation of model statistics on the strength over the propagation distance mainly results from (a)
signal frequency as well as its tissue thickness values in geometric spreading of the elastic wave; (b) energy loss, (c)
the FE model. Model development parameters are limited absorption, scattering, and diffraction; (d) noise, friction, or
to Tables 1 and 2, as focus of this study is to identify the other obstacles; and (e) dispersion due to its frequency depend-
effect of soft tissues on AE wave propagation. However, ence [49, 50]. Thus, an effort is taken to determine how the
the efficiency of AE for continuous THR monitoring in amplitude of the signal at the selected point differs due to
an in-vitro experimental has been studied [47]. Studies the thickness of the layer as it propagates over the medium,
are still progressing to implement in tissue models, ani- especially on varying adipose tissue thickness factors. Using
mal models, and finally to real patient populations and to this information, the impact of the biological layers between
benefit society. a theoretical THR and the AE sensor placed on the skin is
The governing equation for the implemented study can analyzed. Numerical model building parameters were based
be derived based on the conservation of momentum. on already established studies [38, 44–46] and on one of our
in vitro experimental models [47]. Since with properly cali-
D2 u brated model parameters, a finite element model offers a reli-
𝜌 = ∇.𝜎 + FV (5)
Dt2 able solution for mechanics of materials and structures and
where 𝜎 is the Cauchy stress, and 𝜌 is the density of the allows us to calculate various measurements under various
material, and FV is the body force per unit volume, FV  . A conditions, such as displacement, strain energy, kinetic energy,
default boundary condition (free node) where the outer sur- and stresses, we have taken the named measurements and
face of the model was selected with no constraints or loads depicted them here to discuss the same model but from vari-
acting on this boundary. The four domains or layers were ous perspectives. With current AE technology, the likelihood
discretized with 2D linear triangular elements under plane of placing AE receiver sensor on the skin surface is higher;
strain assumption. The modular junction micromotions unfortunately, this may lead to excessive noise or signal attenu-
considered a representative 2μm, and15 − 20μm displace- ation as the AE signal propagates through different biological
ment conditions with different frequencies of 10 − 60kHz . layers. Therefore, signal attenuation caused by the presence of
Also, the maximum element size was selected according to soft tissue must be analyzed to deploy AE as a structural moni-
the frequency parameter selected with proper time steps. toring tool in medical implant health evaluations. Accordingly,

13
Medical & Biological Engineering & Computing (2022) 60:1497–1510 1501

Fig. 2  Schematic diagram representing a hip implant with an effect face point, with its output parameters, c von-Mises stress experienced
of 2 µm micromotion at the modular junction, b FE stress model with at the implant-muscle interface, and d total surface displacement
its 2D-cross sectional mesh generated for tissue thickness 40  mm at experienced at the implant-muscle interface
20 kHz frequency and 2 µm displacement at the implant-muscle inter-

the developed models consider two extreme values of tissue


thickness at varying frequencies and loads.
A mesh convergence [51] study is performed to confirm the
selection of n (Eq. 2) so as to obtain an accurate FE model,
since mesh convergence and mesh independence analyses help
to develop an accurate and efficient FE model. Here, mesh
convergence study also considers responses at the selected
points to confirm its impact on the true solution [52, 53]. This
process works iteratively with an increase in the number of
elements. Figure 3 shows the displacement data obtained at the
tissue-skin interface point P2 for different mesh sizes accord-
ing to n = 22, 20, 18, and 16 in Eq. 2 and for a frequency of
20 kHz. The trend of the displacement graphs is almost similar
with minimal amplitude fluctuation. However, in this study,
mesh size and corresponding time-step relative to n = 20 are
Fig. 3  Displacement field obtained at the adipose tissue-skin interface
selected as a representative to examine the effect of adipose point P2 for different mesh sizes where n = 22, 20, 18, and 16 in Eq. 2
tissue thickness in AE wave propagation.

3.1 Models based on micromotion (displacement initiation and propagation over the biological layers. Proper
parameter) at the modular junction consideration has been given to the nearly incompressible
mechanical property of human soft tissues [38, 39] where
This model assumes a representative 2-µm micromotion the COMSOL FE model employs a mixed formulation with
[54] at the modular junction and the subsequent elastic wave pressure as an independent variable.

13
1502 Medical & Biological Engineering & Computing (2022) 60:1497–1510

3.1.1 Amplitude loss due to geometric spreading with an increase in thickness for a 2μm micromotion at the
at interface points implant interface. The stress experienced at the interface
points P1, P2, and P3 are found to be lower for the increased
Simulated waveforms at P1, P2, and P3 for an applied thickness condition (Fig. 5b) associating a decreased mate-
boundary condition of 2μm micromotions at the modu- rial deformation.
lar junction with a frequency of 20 kHz for adipose tissue
thickness of 40 mm and 70 mm, respectively, are plotted in 3.1.3 Signal attenuation over the propagation distance
Fig. 4. In this figure, the blue plots depict the signal varia-
tion received at P1, while red and yellow plots represent the To understand how a signal is attenuated with a variation
signal variation received at points P2, and P3, respectively. in its propagation distance, a signal is recorded at the outer
For the same 20 kHz frequency, the signals received at P1, boundary where the receiver sensor is placed and plotted
P2, and P3 of adipose-tissue thickness of 70 mm (Fig. 4b) for different thicknesses characteristics. In Fig. 6, the black
show a considerable reduction in its magnitude compared plot shows the signal received for 40 mm and the red plot
to the respective plots of adipose-tissue thickness of 40 mm for 70 mm thickness at the adipose tissue layer. This figure
(Fig. 4a). This reduction in magnitude represents the impact is a clear indicator of how the propagated waveform loses its
of adipose tissue thickness in stress wave propagation and strength as the traveling distance increases and encounters
the possibility of receiving a stronger or weaker signal at the more obstacles or friction over the propagation distance.
sensor placed on the skin surface.
3.1.4 Energy distribution at the interface points
3.1.2 Attenuation analysis based on von Mises stress
A decrease in amplitude can be viewed as the energy loss
The von-Mises stress represents the effective stress experi- over the propagation channel. Since this study considers
enced at a specific point (time) after the beginning of wave isotropic layers [58–60] with finite propagation distances,
excitation, which in turn is representative of strain energy; the attenuation of the wave over the propagation medium
the material undergoes due to the load impact. Analyzing is non-negligible. Elastic strain energy or elastic potential
von-Mises stress also shows the energy attenuation brought energy on an object represents the stored energy at each
by varying the thickness of the propagation medium [55–57]. element when a force is applied to deform the object [61].
Figure 5a and b shows how the stress distribution varies Figure 7a and b represents the elastic strain energy variation

Fig. 4  Time vs displacement
field recorded at the muscle-
tissue interface (P1), tissue-skin
interface (P2), and skin surface
(P3) at an adipose tissue thick-
ness of a 40 mm and b 70 mm
for a 2-µm micromotion at the
modular junction

Fig. 5  Time vs von-Mises stress


recorded at the muscle-tissue
interface (P1), tissue-skin inter-
face (P2), and skin surface (P3)
at a adipose tissue thickness
of 40 mm and b adipose tissue
thickness of 70. The increase in
adipose tissue thickness dimin-
ishes the stress rate experienced
in the selected points

13
Medical & Biological Engineering & Computing (2022) 60:1497–1510 1503

obstacles specifically at the adipose tissue-skin interface


point P2. Kinetic energy (KE) is mainly generated because
of the vibration of material particles and radiated as elas-
tic waves. Also, a large part of the total energy comprises
the KE and the measured AE energy was observed to be
scaled with the incident KE of the particles. Analyzing the
variations of KE received at the selected points can reveal
the energy loss obtained at each point and thus the signal
attenuation [62–65]. Figure 8 a and b represents the KE
due to the increase in adipose-tissue thickness or increase
in the distance the wave propagates. In this figure, higher
energy loss and lower KE can be observed for the model
with adipose tissue thickness of 70 mm when compared to
the 40-mm model.

Fig. 6  Time vs amplitude of the signal recorded at the outer layer 3.2 Attenuation characteristics
with an adipose tissue thickness of 40 mm, 70 mm, and 2 µm micro- concerning the frequency
motion at the modular junction
To understand the effect of frequency variations in elastic
experienced for an adipose-tissue thickness of 40 mm and wave transmission, a range of frequencies from 10 to 60 kHz
70 mm, respectively. In other words, each group in Fig. 7a have been considered and its energy loss observations are
and b corresponds to the different load conditions due to depicted in Fig. 9 a and b. As expected, an increase in fre-
various thickness characteristics, and we can observe a quency corresponds to greater energy release good enough
lesser deformation with increased thickness resulting in to deform the object and radiated as elastic waves which
more loss due to a longer propagation distance and other in turn depicts the signal loss effect due to an increase

Fig. 7  Time vs elastic strain


energy recorded at the muscle-
tissue interface (P1), tissue-skin
interface (P2), and skin surface
(P3). a Adipose tissue thickness
of 40 mm; b adipose tissue
thickness of 70 mm, when 2 µm
displacement is applied at the
modular junction. Frequency
parameter considered in this
model is 20 kHz

Fig. 8  Time vs kinetic energy


recorded at the muscle-tissue
interface (P1), tissue-skin
interface (P2), and skin surface
(P3). a Adipose tissue thickness
of 40 mm; b adipose tissue
thickness of 70 mm, when 2 µm
displacement is applied at the
modular junction. Frequency
parameter considered in this
model is 20 kHz

13
1504 Medical & Biological Engineering & Computing (2022) 60:1497–1510

in frequency [66, 67]. Since more amplitude peaks are thickness is increased from 40 to 70 mm because they
expected as frequency increases an increase in frequency become heavily scattered. In [68], elastic wave propaga-
from 10 to 60 kHz, more cycles are generated. The wave- tion studies in human femur bone samples also confirmed
length decreases as the frequency increases, indicating that the effect of distance in wave attenuation by placing four
the distance between the cycles decreases. As a result, with broadband AE sensors on the femur diaphysis separated
more cycles occurring, more signal scattering will occur by a specific distance of 10 mm.
and lead to an attenuation of the amplitude. That is, more To verify the effect of thickness of the biological layers in
energy loss and magnitude reduction are experienced for wave propagation, a parametric sweep study with an adipose
higher frequencies of the initiated wave as well as with an tissue thickness of 40 to 70 mm has been conducted for a
increase in the distance the wave propagates through. This frequency of 20 kHz. For each of these trials, the amplitude
directly reveals the effect of frequency and propagation dis- values of the received signals were found to be attenuated
tance on signal dampening. The same tendency is observed (Fig. 10) and von Mises stress appeared to be decreasing
for an adipose tissue thickness of 40 mm as well. So proper (Fig. 11) with an increase in adipose tissue thickness. The
consideration must be taken to address different frequencies rate at which a signal is attenuated can be identified with its
occurring due to various damages or events associated with attenuation coefficient (α) which is directly related to the
a hip implant failure. path length (L), frequency (f), and attenuation (a) of the sig-
nal. Wave attenuation at each point was calculated assuming
3.3 Attenuation studies with different thickness the average attenuation coefficient (db∕cm∕MHz) of muscle,
parameters tissue, and skin as 1.09, 0.54, and 0.8, respectively [7, 68].
Shear wave propagation related to MRI studies [70] also
The amplitude value obtained at the muscle interface is shows the effect of tissue elasticity in wave attenuation.
generally the highest, followed by the skin and/or the adi-
pose tissue. The highest magnitude values at the muscle
may generally be due to it being the first layer in the 3.4 Convergence study based on micromotion
numerical model. It does not mean that the signal propa- at modular junction
gated from the center of the model travels through the
muscle layer and scatters less compared to the following Auxiliary sweep study in the COMSOL Multiphysics enables
layers. Due to the smaller distance the wave propagates, the resolution of the developed simulation model with a set
scattering may be less, and hence, the signal is less atten- of parameters rather than changing the parameters manually.
uated and displays higher amplitude when compared to For the numerical model where micromotion at the modular
the subsequent layers. Although the skin layer is the last junction is taken as the boundary condition, the parametric
interface in the numerical model, the amplitude of the sig- sweep considers 15μm, 16μm, 17μm, 18μm, 19μm, and20μm
nals at this layer is not attenuated as significantly as the micromotions for adipose tissue thicknesses of 40 mm and
previous layers due to it being the thinnest layer. There- 70 mm, respectively [71, 72]. Figure 12 a and b shows the
fore, the signals propagate through the skin layer with- amplitude variation that occurred at the interface point P2
out heavy attenuation and scattering. The adipose tissue on varying micromotion boundary conditions at the modu-
layer shows the lowest amplitude in most models might lar junction. These findings match with the results obtained
be due to its higher thickness properties. Also, the signal for load acting on the implant interface and thus confirm
amplitude significantly decreases as the adipose tissue the focus of this study where an increase in adipose tissue

Fig. 9  a Elastic strain-energy


and b kinetic energy recorded
at the skin-AE receiver sensor
(P3) for an adipose tissue thick-
ness of 70 mm for frequencies
10–60 kHz. Representative plots
indicates that energy loss, and
thus, wave dispersion increases
with an increase in frequency
from 10 to 60 kHz, same as the
energy loss effect observed with
an increase in adipose tissue
thickness from 40 to 70 mm

13
Medical & Biological Engineering & Computing (2022) 60:1497–1510 1505

4 Discussion

In this study, mainly two different numerical model cat-


egories have been developed that consider the effect of
adipose tissue thickness in stress wave transmission of a
THR prosthesis [73]. The initial signal propagates through
different layers of varying densities and thicknesses to
reach the AE sensor placed on the skin surface, where
the sensor would theoretically be located. As the thick-
ness increases, the signal must travel through a greater
distance and have to interact with additional noise. This
noise can be caused by the scattering of waves across dif-
ferent biological interfaces, or by the distance between the
AE sensor and skin interface. Thus, an increase in tissue
thickness increases signal attenuation, and a decrease in
Fig. 10  Wave attenuation evaluated at the adipose tissue-skin inter- signal amplitude when detected at the AE sensor.
face point P2 for frequency of 20 kHz for adipose tissue thicknesses Some of the guided wave propagation studies pre-
ranges between 40 and 70 mm
sented in the literature determine the effect of soft tissue
on detecting degenerative disorders. Vavva et al. [67] and
Protopappas et al. [68] employed the first-arriving signal
in ultrasonic evaluation to show the influence of overlying
soft tissue in detecting the received wave velocity. Vibra-
tional mode evaluation [76] also depicts the effect of vary-
ing soft tissue thickness and other properties with vary-
ing frequencies. This paper also delivers the same results
depicting the adverse impact of soft tissue in the received
signal. In a real-time scenario, proper attention should be
given to the tissue characteristics since it varies for each
person even though some studies [77] consider a constant
thickness for the tissue, and thus excludes its effect for the
measurement of ultrasound wave velocity.
The signal attenuation (Fig. 10) and von-Mises stress
(Fig. 11) at the adipose tissue-skin interface point P2 for
Fig. 11  von Mises stress recorded at the adipose tissue-skin inter- the different thickness values show the attenuation effect
face point P2 for frequency of 20 kHz for adipose tissue thicknesses
ranges between 40 and 70 mm occurred due to the distance the wave needs to propagate
or thickness of the medium when the same frequency
parameter is considered. It shows that with an increase
thickness degrades the elastic waves as it propagates over in tissue thickness, the signal undergoes more absorption
the biological medium. and scattering, thereby causing a reduction in its amplitude

Fig. 12  Amplitude values
recorded at the adipose tissue-
skin interface point P2 for fre-
quency of 20 kHz for an adipose
tissue thickness of a 40 mm and
b 70 mm with varying displace-
ment parameters at the modular
junction interface

13
1506 Medical & Biological Engineering & Computing (2022) 60:1497–1510

and intensity characteristics. The same trend was observed material properties before arriving at the sensor. AE sen-
with an increase in frequency value as well. This in turn sor can also pick up noise at the skin-to-sensor interface,
leads to the fact that as the distance between the AE source depending on how it is attached to the skin. Since signal
and receiver sensor increases, the arrived signal may be attenuation does not allow accurate evaluation of the damage
diminished in strength resulting in an imprecise structural source, methods to reduce or account for attenuation should
assessment. Apart from the distance and frequency char- be in place to utilize AE clinically.
acteristics, the density of the medium also has a direct
impact on the signal decay rate. The density of the skin is 4.2 Challenges
generally greater than that of the adipose tissue, so propa-
gated signal suffers scattering and absorption even in the Even though AE evaluation can non-intrusively analyze the
thinner skin medium might contribute to increased attenu- damage mechanisms associated with orthopedic implants,
ation effect. several factors can deter its effectiveness in real-time clini-
Comparing the models developed for different micromo- cal applications. The main challenges are (i) hip being
tions, and frequencies, the AE wave was found to be attenu- one of the most complex biological structures, variations
ated as the thickness or distance increases, which corrobo- of the anatomical structures of patients with hip implant
rates the focus of this study and matches with the results having varying biomechanics and mobility factors are key
presented in the research [6, 8, 10, 11] that addresses the challenges while considering AE as an effective diagnostic
impact of biological layers over the stress wave propaga- tool. (ii) There are additional biological factors like syno-
tion route. Also, an increase in frequency does significantly vial fluid and tissue thickness of each patient that must be
impact the attenuation characteristics of the signal like vary- considered as a prerequisite before deploying AE as a non-
ing the adipose tissue thickness does. Hip implant loosening invasive early structural monitoring tool in orthopedics. (iii)
initiates a wave with a frequency around 20 kHz, whereas It is imperative that the experimental and numerical data
fatigue crack may initiate elastic waves with very high fre- should be validated in a human motion lab with real patient
quencies of AE signals [38, 39]. It may also be noted that the groups. (iv) A cloud-based or on-premise database is not
frequency of the stress wave generated is entirely depending available to correlate AE signatures with hip implant abnor-
on the damage level associated with the implant, such as malities and malfunction and provide clinical advice to the
mechanical instability, crack initiation, or severity of tissue physician and patients. The featuring studies address these
infection. Considering all these facts, the AE signal received challenges by considering multiple AE sensor locations to
at the receiver sensor may vary based on the propagation dis- obtain accurate signals and building a large set of numerical
tance, frequency of the generated event, energy loss over the models to mimic the biological variabilities integrated with
propagation medium, and hence the accuracy of the signal. deep neural network models. Accurate positioning the AE
Thus, to implement AE as an implant monitoring tool focus sensor is derived considering the biomechanics and mobil-
must be given to identify the effect of tissue thickness as ity aspects in numerical modeling performed, and the same
well as to receive the accurate high-frequency elastic waves will be applied in experimental and clinical settings. Imple-
by identifying a proper place to position the AE receiver menting the same concept for monitoring other prosthetic
sensor. interventions is also under consideration but with different
databases for each case.
4.1 Clinical implications
4.3 Future perspective
In clinical applications, the distance between the THR and
the AE sensor is significant because it can heavily impact An enhancement to this FE study, considering the nonlin-
the quality of AE signals detected. In other words, it can earity, viscosity, and anisotropic characteristics to under-
compromise the effectiveness of utilizing AE as a tech- stand the effect of attenuation over the variable thickness
nique to detect material degradations. When examining the medium, is ongoing [78–80]. The attenuation of signals
cause of signal attenuation in a clinical setting, the source caused by the presence of biological layers must be bet-
of stress waves, AE sensor placement, transducer proper- ter understood. This knowledge can be applied to create
ties, and complex geometry of the structure being evaluated a standard evaluation curve that can clinically predict the
must be considered. As the strategic sensor placement on AE signal emitted from specific failure modes. To elabo-
the skin can reduce noise and attenuation of the recorded rate, estimated signal attenuation at the implant-to-sensor
signals to obtain more accurate information on the damage interface can be predicted based on the patients’ biologi-
progression of the implant. For example, attenuation results cal layer properties and thus the original signal. Since
from the scattering and absorption of waves generated from we could not replicate the entire clinical set up, continu-
the implant across several biological layers with different ing studies deal with the simulation of a wide range of

13
Medical & Biological Engineering & Computing (2022) 60:1497–1510 1507

biological structures from tissue models to animal models References


and patient models, and its success metric evaluation by
computational time (~ 1 h per simulation) using reduced 1. Browne M, Roques A, Taylor A (2005) The acoustic emission
finite element models. Also, numerical model complexity technique in orthopaedics-a review. J Strain Anal Eng Design
40:59–79
will be adjusted to minimize the significant differences 2. FitzPatrick AJ, Rodgers GW, Hooper GJ, Woodfield TBF (2017)
in experimental and numerical models (if present) and to Development and validation of an acoustic emission device to
mimic the complex biological structures. By better under- measure wear in total hip replacements in-vitro and in-vivo.
standing the specific signals emitted from different failure Biomed Signal Process Control 33:281–288
3. Ampadi R., Remya, Vishwash B., Christine Lee, Srinivasa Pai
modes in correlation with the impact of the tissue-skin P., Alejandro A. Espinoza Oras, Didem Ozevin, and Mathew T.
interface, AE can be utilized in a clinical setting to evalu- Mathew (2020). Hip implant performance prediction by acous-
ate the material progression of an implant despite attenu- tic emission techniques: a review. Med Biol Eng Comput 1–14.
ation or noise produced. 4. Banerjee, Shiladitya, Kazage JC Utuje, and M. Cristina Mar-
chetti (2015). Propagating stress waves during epithelial expan-
In summary, the current study is purely a proof of concept sion. Phys Rev Lett 114.
with few limitations but highlighting the impact of adipose- 5. Benoit, M., J.H. Giovanola, A. Curnier, Kossi Agbeviade, and
tissue thickness on AE signal transmission and detectability Marcel Donnet (2009). Experimental characterization of pres-
of AE signals at the skin surface or receiver sensor. Future sure wave generation and propagation in biological tissues. In
13th Int Conference Biomed Eng, Springer, Berlin, Heidelberg
research will address various aspects to correlate the sig- 1623–1626.
nal received at AE receiver sensor and signal sources at 6. Gao XJ, Murota K, Ono Tomita YM, Nunomura Higo YS (1990)
the implant interfaces. This will assist in developing an Evaluation of the fixation of artificial hip joint by acoustic emis-
enhanced AE-based monitoring system for the orthopedic sion. Jpn J Appl Phys 29(S1):215
7. Roshni. Interaction Of Sound and Media. Roshni’s Physics
application. E-Portfolio. https://​sites.​google.​com/​site/​roshn​isphy​sicse​portf​
olio/​inter​action-​of-​sound-​and-​media. Accessed 13 April 2020
8. Amin VR (1989) Ultrasonic attenuation estimation for tissue
characterization. (Iowa State University). Accessed 13 April
2020
5 Conclusions 9. Pan L, Zan L, Stuart Foster F (1998) Ultrasonic and viscoelastic
properties of skin under transverse mechanical stress in vitro.
The main findings of this study include the following: Ultrasound Med Biol 24:995–1007
10. Mavrogordato M, Taylor M, Taylor A, Browne M (2011) Real
time monitoring of progressive damage during loading of a
The variation in adipose tissue thickness in the numerical simplified total hip stem construct using embedded acoustic
model impacted the amplitude of the signal propagated emission sensors. Med Eng Phys 33:395–406
through the biological layers between the THR and the 11. Rodgers GW, Young JL, Fields AV, Shearer RZ, Woodfield
AE receiver sensor. Thus, an increase in signal-to-sensor TBF, Hooper GJ, Geoffrey Chase J (2014) Acoustic emission
monitoring of total hip arthroplasty implants. IFAC Proceed
distance does have an impact on the quality of the signals 47:4796–4800
being detected by an AE sensor due to its attenuation 12. Li C, Granger C, Del Schutte H, Biggers SB, Kennedy JM,
effect. Therefore, biological layers must be accounted for Latour RA (2003) Failure analysis of composite femoral com-
when utilizing AE in a clinical setting for the accurate ponents for hip arthroplasty. J Rehabilitation Res Develop
40:131–146
evaluation of material degradations. 13. Chethan KN, Zuber M, Shenoy S, Kini CR (2019) Finite ele-
Depending on the load or the micromotion the THR expe- ment analysis of different hip implant designs along with femur
riences, the strength of the initiated AE signal will vary under static loading conditions. J Biomed Phys Eng 9:507
and proper evaluations must be in place to identify the 14. Pastrav LC, Devos J, Van der Perre G, Jaecques SVN (2009) A
finite element analysis of the vibrational behaviour of the intra-
events happening in the implant. operatively manufactured prosthesis–femur system. Med Eng
A better understanding of these layer properties can help Phys 31:489–494
predict their impact on AE signals. Through this, AE can 15. Hériveaux Y, Vu-Hieu N, Vladimir B, Cyril G, Guillaume H
be utilized to non-invasively and non-destructively moni- (2019) Reflection of an ultrasonic wave on the bone−implant
interface: effect of the roughness parameters. J Acoustical Soc
tor the material degradations of a THR through strategic Am 145:3370–3381
AE sensor placement on the skin. 16. Abu-Amer, Y., Darwech, I. and C Clohisy, J. (2007) Aseptic
loosening of total joint replacements: mechanisms underlying
osteolysis and potential therapies. Arthritis Res Therapy 9.
17. Lewis JL, Askew MJ, Wixson RL, Kramer GM, Tarr RR (1984)
The influence of prosthetic stem stiffness and of a calcar collar
on stresses in the proximal end of the femur with a cemented
Declarations  femoral component. J Bone Joint Surg 66:280–286

Conflict of interest  The authors declare no competing interests.

13
1508 Medical & Biological Engineering & Computing (2022) 60:1497–1510

18. Colic K, Sedmak A, Grbovic A, Tatic U, Sedmak S, Djordjevic 06_​Linea​rElas​ticity/​06_​Linear_​Elast​icity_​01_​Elast​ic_​Model.​


B (2016) Finite element modeling of hip implant static loading. pdf. Accessed 25 April 2020
Procedia Eng 149:257–262 38. Pilarczyk B (2019) Contributions of muscle, skin, and adipose
19. Anguiano-Sanchez J, Martinez-Romero O, Siller HR, Diaz-Eli- tissue to indentation response, assessed with computational arm
zondo JA, Flores-Villalba E, Rodriguez CA (2016) Influence of model under quasi static conditions. (Master's thesis, University
PEEK coating on hip implant stress shielding: a finite element of Waterloo)
analysis. Comput Math Methods Med 39. Unger AC, Cabrera-Palacios H, Jürgens Schulz APC, Paech A
20. Chethan KN, Zuber M, Shenoy S, Kini CR (2019) Static structural (2009) Acoustic monitoring (RFM) of total hip arthroplasty
analysis of different stem designs used in total hiparthroplasty results of a cadaver study. Eur J Med Res 14(6):264
usingfinite element method. Heliyon 5:e01767 40. Krupinski, E. (Ed. ) (2008). Digital mammography: 9th Interna-
21. Senalp AZ, Kayabasi O, Kurtaran H (2007) Static, dynamic and tional Workshop. vol. 5116 (Springer).
fatigue behavior of newly designed stem shapes for hip prosthesis 41. Otte JW, Merrick MA, Ingersoll CD, Cordova ML (2002) Subcu-
using finite element analysis. Mater Des 28:1577–1583 taneous adipose tissue thickness alters cooling time during cryo-
22. Kumar KN, Tandon T, Silori P, Shaikh A (2015) Biomechanical therapy. Arch Phys Med Rehabil 83:1501–1505
stress analysis of a human femur bone using ANSYS. Materials 42. Speed of sound, frequency, and wavelength. speed of sound,
Today: Proceed 2:2115–2120 frequency, and wavelength — Physics https://​cnx.​org/​conte​nts/​
23. Gross ST, Abel EW (2001) A finite element analysis of hollow zOZP3​vRI@​13.6:​n4Yub​baZ@​11/​14-1-​Speed-​of-​Sound-​Frequ​
stemmed hip prostheses as a means of reducing stress shielding ency-​and-​Wavel​ength. Accessed 25 April 2020
of the femur. J Biomech 34:995–1003 43. Speed of sound, frequency, and wavelength. speed of sound, fre-
24. Sabatini AL, Goswami T (2008) Hip implants VII: Finite ele- quency, and wavelength | Physics https://​cours​es.​lumen​learn​ing.​
ment analysis and optimization of cross-sections. Mater Des com/​physi​cs/​chapt​er/​17-2-​speed-​of-​sound-​frequ​ency-​and-​wavel​
29:1438–1446 ength/. Accessed 25 April 2020
25. Malau DP, Utomo MS, Annur D, Asmaria T, Prabowo Y, Rahy- 44. Chi SW, Hodgson J, Chen JS, Edgerton VR, Shin DD, Roiz RA,
ussalim AJ, Supriadi S, Amal MI (2019) Finite element analysis Sinha S (2010) Finite element modeling reveals complex strain
of porous stemmed hip prosthesis for children. AIP Conference mechanics in the aponeuroses of contracting skeletal muscle. J
Proceed AIP Publishing LLC 2193:050020 Biomech 43(7):1243–1250
26. Çelik T, Kişioğlu Y (2019) Evaluation of new hip prosthesis 45. McKee CT, Last JA, Russell P, Murphy CJ (2011) Indentation
design with finite element analysis. Australas Phys Eng Sci Med versus tensile measurements of Young’s modulus for soft biologi-
42:1033–1038 cal tissues. Tissue Eng Part B Rev 17:155–164
27. Özgun N, Busse DW-IM, Trejo IEG (2013) Finite element method 46. Madeti BK, Srinivasa Rao C, BSiva Rao Bollapragada SS (2018)
(FEM) & COMSOL - A (very) brief introduction.Systems Neuro- Force evaluation and stress distribution at possible weight and
science & Neurotechnology Unit, Neurocenter, Saarland Univer- structure of femur bone in pelvis frame while standing. Int J Med
isty Hospital. Accessed 12 May 2020 Eng Inform 10:235–251
28. Chen DW, Lin C-L, Hu C-C, Wu J-W, Lee MS (2012) Finite 47. Lee C, Zhang L, Morris D, Cheng KY, Ramachandran RA, Barba
element analysis of different repair methods of Vancouver M, Bijukumar D, Ozevin D, Mathew MT (2021) Non-invasive
B1 periprosthetic fractures after total hip arthroplasty. Injury early detection of failure modes in total hip replacements (THR)
43:1061–1065 via acoustic emission (AE). J Mechanic Behavior Biomed Mater
29. Cerrolaza M (2004) Computational bioengineering: current trends 118:104484
and applications. Imperial College Press. 48. Rafael Mello T, Maru MM (2017) Importance of preclinical eval-
30. Categories of waves. The physics classroom https://​www.​physi​ uation of wear in hip implant designs using simulator machines.
cscla​ssroom.​com/​class/​waves/​Lesson-​1/​Categ​ories-​of-​Waves#:​~:​ Revista Brasileira de Ortopedia (English Edition) 52:251–259
text=T​ ransv​ erse waves are always characterized, direction that the 49. Mayrosh J (2001) Experimental study of the attenuation of acous-
wave moves. Accessed 13 April 2020 tic emission signals in welded steel structures
31. Transverse Waves. Transverse and longitudinal waves http://h​ yper​ 50. Zakharov D, Ptichkov SN, Shemyakin VV (2010) Acoustic emis-
physi​cs.​phy-​astr.​gsu.​edu/​hbase/​Sound/​tralon.​html. Accessed 25 sion signal attenuation in the waveguides used in underwater AE
April 2020 testing. In 10th Eur Conference on Non-Destructive Test 50
32. Okuno, E. Physics of ultrasound. https://​human​health.​iaea.​org/​ 51. Trott, D. W. and Matthias, K. G. (2010) Conducting finite element
HHW/​Medic​alPhy​sics/​TheMe​dical​Physi​cist/​Stude​ntsco​r ner/​ convergence studies using COMSOL 4.0. In Proceedings of the
Handb​ookfo​rTeac​hersa​ndStu​dents/​Chapt​er_​12.​pdf. Accessed 13 COMSOL Conference.
April 2020 52. Patil H, Jeyakarthikeyan PV (2018) Mesh convergence study and
33. Geerligs M, Peters GW, Ackermans PA, Oomens CW, Baaijens estimation of discretization error of hub in clutch disc with inte-
F (2008) Linear viscoelastic behavior of subcutaneous adipose gration of ANSYS. In IOP Conference Series: Materials Science
tissue. Biorheology 45:677–688 and Engineering, vol. 402, no. 1, p. 012065. IOP Publishing
34. Pradhan T (2015) Finite element modeling of impact-generated 53. Stolk J, Verdonschot N, Huiskes R (1998) Sensitivity of failure
stress wave propagation in concrete plates for non-destructive criteria of cemented total hip replacements to finite element mesh
evaluation. (Lehigh University). density. J Biomech 1001:165
35. Trivedi S (2014) Finite element analysis: a boon to dentistry. J 54. Falkenberg A, Drummen P, Morlock MM, Huber G (2019) Deter-
Oral Biol Craniofacial Res 4:200–203 mination of local micromotion at the stem-neck taper junction of
36. Kamali N, Mahdavi A, Sheng-Wei C (2020) Numerical study on a bi-modular total hip prosthesis design. Med Eng Phys 65:31–38
how heterogeneity affects ultrasound high harmonics generation. 55. Amirkhizi AV, Tehranian A, Nemat-Nasser S (2010) Stress-wave
Nondestructive Test Evaluation 35:158–176 energy management through material anisotropy. Wave Motion
37. The Linear Elastic Model. Solid mechanics part I: an introduc- 47:519–536
tion to solid mechanics http://h​ omep​ages.​engin​eering.​auckl​and.​ 56. Yu J, Liu Z, He Z, Zhou X, Ye J (2020) Fluctuation characteristic
ac.​nz/​~pkel0​15/​Solid​Mecha​nicsB​ooks/​Part_I/​BookSM_​Part_I/​ test of oblique stress waves in infilled jointed rock and study of
the analytic method. Advances in Civil Engineering 2020

13
Medical & Biological Engineering & Computing (2022) 60:1497–1510 1509

57. Abbey T (2016) Stress in FEA: part 3. https://w ​ ww.d​ igita​ lengi​ neer​ 78. Rus G, Faris IH, Torres J, Callejas A, L Melchor (2020) Why are
ing247.​com/​artic​le/​stress-​in-​fea-​part-3/. Accessed 12 May 2020 viscosity and nonlinearity bound to make an impact in clinical elas-
58. Kalra A, Lowe A,Al-Jumaily AM (2016) Mechanical behaviour tographic diagnosis? Sensors (Basel) 20.
of skin: a review. J Material Sci Eng 5 79. Kim B, Kwon S, Lee M, Kim QH, An S, Jhe W (2015) Probing
59. Aaron R, Huang M, Shiffman CA (1997) Anisotropy of human nonlinear rheology layer-by-layer in interfacial hydration water. Proc
muscle via non-invasive impedance measurements. Phys Med Biol Natl Acad Sci U S A 112:15619–15623
42:1245–1262 80. Benoit M, Giovanola JH, Curnier A, Agbeviade K, Donnet M (2009)
60. Farina D, Rainoldi A (1999) Compensation of the effect of sub- Characterization of pressure wave propagation in biological tissues.
cutaneous tissue layers on surface EMG: a simulation study. Med DYMAT-International Conference on the Mechanical and Physical
Eng Phys 21:487–497 Behaviour of Materials under Dynamic Loading, vol 1. EDP Sci-
61. Romenskii EI, Lys EB, Cheverda VA, Epov MI (2017) Dynamics of ences. https://​doi.​org/​10.​1051/​dymat/​20091​24
deformation of an elastic medium with initial stresses. J Appl Mech
Tech Phy 58:914–923 Publisher's note Springer Nature remains neutral with regard to
62. Wei J, Zhu W, Guan K, Zhou J, Song JJ (2019) An acoustic emission jurisdictional claims in published maps and institutional affiliations.
data-driven model to simulate rock failure process. Rock Mechanics
and Rock Engineering 1–17.
63. Sim E, Larkin J, Burke K, Bock CW (2003) Testing the kinetic
energy functional: kinetic energy density as a density functional. J Remya Ampadi Ramachandran  : Ph.D. in Technology, postdoc
Chem Phys 118:8140–8148 researcher and research specialist at the Department of Biomedical
64. Droubi MG, Reuben RL (2016) Monitoring acoustic emission (AE) Engineering, University of Illinois at Chicago. Research interests
energy of abrasive particle impacts in a slurry flow loop using a include acoustics and speaker recognition, biomedical implant monitor-
statistical distribution model. Appl Acoust 113:202–209 ing through acoustic emission, signal processing, computational meth-
65. Droubi MG, Reuben RL, White G (2012) Statistical distribution ods for biomedical applications, machine learning, and deep learning
models for monitoring acoustic emission (AE) energy of abra- models for biomedical applications, biotribocorrosion.
sive particle impacts on carbon steel. Mech Syst Signal Process
30:356–372 Christine Lee  : M.S. in Bioengineering, University of Illinois at Chi-
66. Han X, Zheng L, Chen C, Shi H (2018) Velocity and attenuation of cago, IL, USA. The research focused on noninvasive early detection
elastic wave in a developed layer with the initial inner percolation of mechanical failure in total hip replacements via acoustic emission
in the pores. J Petrol Explor Prod Technol 8:1079–1088 technique.
67. Moebs W, Ling SJ, Sanny J (2016) 16.4 Energy and power of a wave.
in University Physics Volume 1 (OpenStax). Accessed 12 May 2020 Lu Zhang  : Ph.D. in Civil Engineering, assistant professor at College of
68. Strantza M, Louis O, Demosthenes P, Frans B, Van Hemelrijck Civil Engineering and Architecture, Guilin University of Technology,
D, Aggelis DG (2014) Wave dispersion and attenuation on human Guilin, China. Research interests include quantitative acoustic emission
femur tissue. Sensors 14:15067–15083 (AE) method, computational methods for biomedical applications, and
69. Augusta University, Medical college of Georgia, (2020) Attenua- AE monitoring in biomedical implants.
tion math. Resident Physics Lectures. https://​www.​augus​ta.​edu/​mcg/​
radio​logy/​resid​ency/​docum​ents/​atten​uatio​nmath.​ppt. Accessed 12 Supriya M. H  : Ph.D. in Electronics, professor and head at the Depart-
May 2020 ment of Electronics, Cochin University of Science and Technology,
70. Bishop J, Poole G, Leitch M, Plewes BD (1998) Magnetic resonance India. Research interests include Signal processing, underwater acous-
imaging of shear wave propagation in excised tissue. J Magn Reson tics target identification, and biometric recognition.
Imaging 8:1257–1265
71. Haschke H, Konow T, Huber G, Morlock MM (2019) Influence of Divya Bijukumar  : Ph.D. in Biotechnology, assistant professor at the
flexural rigidity on micromotion at the head-stem taper interface of Department of Biomedical Sciences, University of Illinois College of
modular hip prostheses. Med Eng Phys 68:1–10 Medicine at Rockford. Research interests include nanotechnology, sur-
72. Farhoudi H, Fallahnezhad K, Oskouei RH, Taylor M (2017) A finite face modification, drug delivery, and toxicity.
element study on the mechanical response of the head-neck interface
of hip implants under realistic forces and moments of daily activi- P Srinivasa Pai  : Ph.D. in Mechanical Engineering, professor at Depart-
ties: part 1, level walking. J Mech Behav Biomed Mater 75:470–476 ment of Mechanical Engineering, NMAM Institute of Technology,
73. Weightman BO, Paul IL, Rose RM, Simon SR, Radin EL (1973) A Nitte,India. Research interests are tool condition monitoring acoustic
comparative study of total hip replacement prostheses. J Biomech emission,vibration monitoring, artificial neural networks maintenance
6:299–311 engineering andmanagement, wavelet and its applications to machin-
74. Vavva MG, Protopappas VC, Gergidis LN, Charalambopoulos A, ing related areas, high-speedmachining–related areas including wear
Fotiadis DI, Polyzos D (2008) The effect of boundary conditions studies, and surface roughness analysis.
on guided wave propagation in two-dimensional models of healing
bone. Ultrasonics 48:598–606 Kharma Foucher  : Ph.D. in Bioengineering, associate professor at
75. Protopappas VC, Fotiadis DI, Malizos KN (2006) Guided ultrasound the Department of Kinesiology and Nutrition. Research interests
wave propagation in intact and healing long bones. Ultrasound Med include on biomechanics, motion analysis, musculoskeletal modeling
Biol 32:693–708 of people with and without hip implants, and clinical outcomes.
76. Saha S, Lakes RS (1977) The effect of soft tissue on wave-propa-
gation and vibration tests for determining the in vivo properties of Sheng‑Wei Chi  : Ph.D. in Civil and Environmental Engineering, asso-
bone. J Biomech 10:393–401 ciate professor at Department of Civil, Materials, and Environmen-
77. Lowet G, Van der Perre G (1996) Ultrasound velocity measurement tal Engineering, University of Illinois, Chicago, IL, USA. Research
in long bones: measurement method and simulation of ultrasound interests include image-based computational methods for biomedical
wave propagation. J Biomech 29:1255–1262 applications; multiscale material modeling; advanced numerical meth-
ods, including meshfree methods, collocation methods, and generalized

13
1510 Medical & Biological Engineering & Computing (2022) 60:1497–1510

finite element methods; large deformation and contact mechanics, frag- such as acoustic emission and low-frequency ultrasounds. Acoustic
ment; and impact simulations emission and numerical models for biomedical applications.

Didem Ozevin  : Ph.D. in Materials Engineering, associate professor Mathew T Mathew  : Ph.D. in Mechanical Engineering, associate
at Department of Civil, Materials, and Environmental Engineering, professor at Department of Biomedical Science, Director of Faculty
University of Illinois, Chicago, IL, USA. Research interests include Research, UIC School of Medicine, Rockford, IL, USA. Leading the
sensor development to combine a variety of sensing elements and to Regenerative Medicine and Disability Lab at UIC Rockford. Research
develop new structural evaluation methods for understanding the dam- interests include biomaterials, biomedical implants, corrosion, tribo-
age states using a variety of structural health-monitoring techniques, corrosion, and biotribocorrosion.

13

You might also like