You are on page 1of 5

DATA SHEET NO GI-2

Magnetic Analysis Corporation

FLAW DETECTION USING ENCIRCLING COIL


EDDY CURRENT SYSTEMS
PRINCIPLES OF OPERATION TESTING MAGNETIC MATERIALS
The detection of flaws such as seams, cracks, When testing materials such as carbon steel,
pits, slivers, weld-line defects and internal discon- austenitic stainless, alloy steels having a perme-
tinuities in metallic materials can be done conve- ability higher than 1, it is often necessary to satu-
niently by using an encircling coil eddy current rate the material with a magnetic field. The effect
system. This type of inspection system is most of this magnetic saturation is to even out the per-
frequently used to locate surface defects in bar meability variations in the material, thereby mak-
stock or wire products, and to detect both ID and ing the material appear to the test coil system as
OD defects in tubing. The test is usually conduct- though it were non-magnetic. The material can be
ed at production speed. saturated by using a permanent magnet or a sat-
urating coil in which D.C. current is flowing. In
The basic principles of operation in encircling or
either case, the eddy current test coil is placed
through coil systems are simple. The test coil is
within the saturating field and performs the test.
excited by an alternating current of a given fre-
quency which induces a flow of eddy currents
around the material that is passing through the ABSOLUTE OR DIFFERENTIAL TEST MODES
coil. When a flaw in the material passes through
In actual testing, a single coil system (absolute
the coil, it causes a change in the flow of eddy
mode) or a system of two or more coils that elec-
currents. It is this change that is detected by the
trically subtract from each other (differential or
electronics.
null mode) may be used to detect defects.
AC When the absolute mode is used, the output of
the coil containing acceptable material is fed into
the electronics, and variations from this norm are
detected. When the coils are connected in the dif-
ferential mode, they continually test and compare
Defects
adjacent segments of the material as it passes
Induced Eddy Currents through the coils. If there is “good” material in
both coils, the resulting difference is zero. In elec-
tronic terms, when the absolute mode is used, the
change seen by the electronics is the difference
Probable Flow of Eddy Currents Around a Defect between the biasing voltage output for good
material and the change in this voltage caused by
a defect passing through the coil. For example, if FREQUENCY SELECTION
a defect causes a change of 1 volt in the output of
The ability to select between a wide range of fre-
the coil, and the normal output of the coil is 10
quencies permits an eddy current tester to vary
volts, then the electronics senses a 10% change.
the depth of penetration and to discriminate
When using the differential mode, the same
between signals caused by conditions such as
defect would theoretically produce an infinite
noise, handling marks, and defects. The proper
change as the difference, and the electronics
frequency for testing a given material is deter-
would register 1 volt as compared to zero. The
mined by the conductivity, the type of defects to
two modes are illustrated in the diagram below.
be detected, the diameter, and in the case of tub-
ing, the wall thickness.
By properly choosing the excitation frequency for
ABSOLUTE MODE
a given material, it is often possible to generate a
phase shift between the signal for handling
marks, and the signal for a defect. When testing
tubing, if the excitation frequency is increased, the
phase angle between the signal for a defect on
10 v. 11 v.
the O.D. and the signal for a defect on the I.D. will
Meter Biased to Zero Meter Reads 1 v increase. However, the signal amplitude for the
I.D. defect will decrease as compared to the sig-
nal amplitude of the O.D. defect when the excita-
DIFFERENTIAL MODE tion frequency is increased. Thus, if the excitation
frequency is too high, the detection of I.D. defects
becomes impossible.

SENSITIVITY SELECTION
Zero Volts Actual One Volt Actual The sensitivity control permits adjustment of the
level of defect signal at which the alarm circuits
will function.

Generally, a differential mode system is more FILTER SELECTION


sensitive to intermittent defects because one sec- The speed at which a defect passes through the
tion of material is being compared to the next. coil produces a change in the flow of eddy cur-
However, with long, uniform discontinuities such rents at a very distinct rate. This rate of change
as seams, a differential mode system may indi- can be equated to a given frequency. Therefore, if
cate only the beginning and the end of the seam, circuits are used that only permit the instrument to
and nothing in between. Conversely, the absolute accept these frequencies, signals generated at
mode would signal for the complete length of the other frequencies will be ignored. The circuits that
defect. However, the ability of the differential do this are called filters.
mode to detect smaller changes and to produce a
better flaw signal-to-noise ratio makes it more
suitable for general applications.
PHASE SELECTION COIL SELECTION
The phase control allows 360º rotation of all signals. The coils can be either absolute or differential mode.
This permits the rotation of desired signals to a Some equipments permit use of both methods
place where they will exceed the threshold level and simultaneously. Coils can be wound with different
activate the alarm circuits. If a signal for an I.D. physical characteristics such as the spacing
defect is smaller in amplitude than an equivalent between the differential coils, the actual width of the
O.D. defect, it is usually possible to adjust the phase coils, and the use of multiple coils electrically con-
control so that both defects activate the alarm cir- nected in different configurations.
cuits equally. This phase rotation is illustrated in the
For some applications, such as inspecting only the
following diagram.
weld portion of a tube or pipe, a tangent or sector
coil is used instead of an encircling coil. This pro-
O.D. Defect Signal
vides greater convenience as the material does not
I.D. Defect Signal need to be threaded through the coil when changing
Threshold Level coil sizes etc. Generally, the same principles for
encircling coil eddy current testing are applicable
also to tangent coils.

Before Rotating Phase The purpose of these different coil systems is to


improve the detection of specific types of defects in
particular materials. In order to ensure equal test
results on a product that does not have a round
O.D. Defect Signal cross section, coils may also be manufactured to
I.D. Defect Signal conform to the cross section of the product.
Threshold Level

After Rotating Phase

Magnetic Analysis Corp. 103 Fairview Park Dr., Elmsford, NY 10523-1544 Tel: 914-530-2000 Fax 914-703-3790
www.mac-ndt.com e-mail: info@mac-ndt.com GI-1/06
DATA SHEET NO GI-3
Magnetic Analysis Corporation

FLAW DETECTION BY USE OF PROBE TYPE


EDDY CURRENT SYSTEMS
PRINCIPLES OF OPERATION PITCH
PROBE
EDDY
CURRENT
Surface seams and cracks in metallic materials FIELD
can be reliably detected by using eddy current
technology. Eddy currents are induced by one or
more test probes which traverse the surface of
the material under test.

To conduct a test using probe type eddy current DEFECT


instruments, the probes are excited by an alter-
nating current of a given frequency which induces Figure 1
a flow of eddy currents in the metal beneath
them. As the test probes pass over a flaw, the
In using eddy current techniques for detecting
flaw causes a change in the flow of eddy cur-
flaws, other so called “false indications” caused
rents, and it is this change which is detected by
by surface conditions and normal metallurgical
the instrument’s electronics. The change in the
variations (“noise”), may be detected. To eliminate
flow of eddy currents as the probes pass over the
these unwanted indications, various selective
defect is generally proportional to the depth of the
circuits can be added to the basic instrumenta-
defect. It is therefore possible to estimate the
tion.
depth with proper electronic calibration.

Relative motion between the test probe and the SELECTIVE CIRCUITS
material being inspected is a requirement in this Frequency Selection
type of test. Generally, the material to be tested is
The ability to select between several different fre-
rotated past a fixed probe, or the probe is rotated
quencies permits variation in the depth of pene-
around the part as it is fed through. In either case,
tration on given specimens of the same material.
the area examined by the probe is a helix with a
It is also often possible to generate a phase shift
pitch determined by the speed of rotation and the
between the signal for noise and the signal for a
linear throughput speed.
defect by changing the excitation (test) frequency.
Figure 1 illustrates that a surface defect that is
Phase Selection
shorter in length than the helical pitch will not be
consistently detected. The helical pitch is a func- The phase control permits 360 degree rotation of
tion of the rotational speed of the probes and all signals. This allows the tester to discriminate
throughput speed of the material. Increasing the between signals caused by extraneous noise and
speed of the probes, or using multiple probes those for defect signals. If there is a phase differ-
decreases the pitch, enabling shorter defects to ence between the noise and the defect signals,
be detected at the same throughput speed. the output to markers,horns etc., can be modified
Decreasing the linear throughput speed also to indicate only those signals in the selected
lessens the helical pitch. phase.
FIGURE 2 FLAW
INDICATION
NOISE NOISE
INDICATION INDICATION
FLAW
INDICATION

Thresholds

Display before phase modification where both flaw and Display after use of phase selection circuit to rotate the phase
noise indications pass above the threshold so only the flaw indication passes the threshold

Filter Selection configurations and physical characteristics to


enhance the testing of different products.
The speed at which the probe passes over the
defect causes the change in the flow of eddy cur- Practical application of this test system requires
rents to occur at a very distinct rate. This rate of that the probe and material under test be moved
change can be equated to a given frequency. in a controlled manner in relation to each other.
Therefore, if circuits are used that only permit For example, cylindrical shaped material can be
these frequencies to enter the instrument, signals inspected by rotating a probe around it, or by
generated at other frequencies will tend to be spinning the material past a stationary probe. A
ignored. The circuits that do this are called filters. suitable flaw detection system, therefore, consists
of an electronic indicator and control unit and (1)
Sensitivity Selection
a rotary mechanism to spin the probes; or (2) a
The sensitivity control permits adjustment of the
fixed probe, spin-feed mechanism which moves
level of defect signal at which the alarm circuits
the test piece. Systems available include:
will function.
Systems for Inspecting Rod and Wire:
Probe Selection
MultiMac® electronics with a Rotomac® rotary
The test probes can be connected in either the
mechanism that spins the test probes. These
absolute or differential (null) mode. When the
systems detect surface defects in magnetic or
probes are used in the absolute mode, one or
non magnetic grades of rod and wire.
more connected in parallel, the equipment is
biased on a “good” portion of material and varia- Systems for Inspecting Bar:
tions in the eddy current flow within the network MultiMac® electronics and Rotomac® rotary
are detected. When the probes are connected in mechanisms are combined with mechanical han-
the differential mode, there must be at least two dling components for moving cut lengths of round
probes, although they may appear as one. These bars through the test. Detects surface defects in
two probes electrically subtract from each other. bar stock, often as an off-line separate test sta-
In effect, they compare what one senses with tion.
what the other senses. The resulting difference is
Helitester Systems for Parts:
fed into the instrument electronics and evaluated.
Incorporates the MultiMac® electronics with a
Theoretically, probes connected in the differential stationary mechanism for holding the test probes
mode produce a better flaw signal to noise ratio. constant while the material is rotated past them.
The probes may also be designed with various This is suitable for inspecting cylindrical parts.

Magnetic Analysis Corp. 103 Fairview Park Drive., Elmsford, NY 10523-1544 Tel: 914-530-2000 Fax 914-703-3790
www.mac-ndt.com e-mail: info@mac-ndt.com GI-8/14

You might also like