You are on page 1of 19

Philips Healthcare

Release Bulletin DigitalDiagnost 2.0.2 SP1

Level 0 Documentation

This document and the information contained in it is proprietary and confidential information of Philips Healthcare ("Philips") and may not be
reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the
Philips Legal Department. This document is intended to be (a.) used by customers and is licensed to them as part of their Philips equipment
purchase or (b.) used to meet regulatory commitments as required by a FDA under 21 CFR 1020.30 (and any amendments to it) and other
local regulatory requirements. Use of this document by unauthorized persons is strictly prohibited.
Printed in Hamburg
Federal Republic of Germany
© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED

Manual Order No. 4512 988 03891 REV AB


Release Bulletin DigitalDiagnost 2.0.2 SP1

Proprietary Notice:

This document and the information contained in it is proprietary and confidential information of
Philips Healthcare (”Philips”) and may not be reproduced, copied in whole or in part, adapted,
modified, disclosed to others, or disseminated without the prior written permission of the Philips
Legal Department. Use of this document and the information contained in it is strictly reserved
for current Philips personnel and Philips customers who have a current and valid license from
Philips for use by the customer’s designated in house service employee on equipment located
at the customer’s designated site. Use of this document by unauthorized persons is strictly
prohibited. Report violation of these requirements to the Philips Legal Department. This docu‐
ment must be returned to Philips when the user is no longer licensed and in any event upon
Philips’ first written request.

Warranty Disclaimer Language:

Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including,
but not limited to, the implied warranties of merchantability and fitness for a particular purpose.

Limitations of Liability Language:

Philips has taken care to ensure the accuracy of this document. However, Philips assumes no
liability for errors or omissions and reserves the right to make changes without further notice to
any products herein to improve reliability, function, or design. Philips may make improvements
or changes in the product(s) or program(s) described in this document at any time.

2 CSIP Level 0 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

SERVICE MANUAL - Unit

Release Bulletin DigitalDiagnost 2.0.2 SP1

Type no:

Manual Order No. 4512 988 03891


released: 08/2009
List of Pages and Drawings

1 ...3

A4
4 ...18 (09.0)

4512 988 03891 REV AB CSIP Level 0 3


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

RELEASE BULLETIN DIGITALDIAGNOST 2.0.2 SP1

Contents
1 What's new ............................................................................................................................ 5
1.1 New Features ....................................................................................................................................... 5
1.1.1 Acquisition Workspot (AWS) ................................................................................................................ 5
1.1.1.1 Application software ............................................................................................................................. 5
1.2 Improvements / Solved Problems ........................................................................................................ 5
1.2.1 Geometry .............................................................................................................................................. 5
1.2.2 Acquisition Workspot (AWS) ................................................................................................................ 6
1.2.2.1 Application Software ............................................................................................................................. 6
1.2.2.2 EVA ...................................................................................................................................................... 7
1.2.2.3 iSite ...................................................................................................................................................... 7
1.2.3 RIO ....................................................................................................................................................... 7
1.2.4 CSP Star SW ........................................................................................................................................ 7
1.2.5 Hot keys ............................................................................................................................................... 8
2 Further important information ................................................................................................ 9
2.1 Known Restrictions ............................................................................................................................... 9
2.1.1 Geometry .............................................................................................................................................. 9
2.1.2 Acquisition workspot (AWS) ............................................................................................................... 10
2.1.2.1 Application Software ........................................................................................................................... 10
2.1.2.2 Operating system ............................................................................................................................... 11
2.1.2.3 EEC .................................................................................................................................................... 11
2.1.2.4 EVA .................................................................................................................................................... 12
2.1.2.5 FSF ..................................................................................................................................................... 12
2.1.3 AGentT ............................................................................................................................................... 13
2.2 Compatibility & requirements ............................................................................................................. 13
2.2.1 Release version levels ....................................................................................................................... 13
2.2.2 Configuration ...................................................................................................................................... 17
2.3 Manpower, estimated time, tools ........................................................................................................ 17
3 Appendix ............................................................................................................................. 18
3.1 Configuration of the Table Related VM Settings ................................................................. 19

4 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

1 WHAT'S NEW

1.1 NEW FEATURES


The following features have been newly implemented with the current DigitalDiagnost release.

1.1.1 Acquisition Workspot (AWS)

1.1.1.1 Application software

Title Description
Finnish language available The Finnish language has been added to the UI language options with
DigitalDiagnost 2.0.2 SP1.

1.2 IMPROVEMENTS / SOLVED PROBLEMS


The following problems have been solved with the current DigitalDiagnost Release.

1.2.1 Geometry
Title Description
CS error during tracking with There were some occurrences of a CS error with a "Call Service" user guidance
THS (GXRCQ 19443) popping up during downwards tracking with the table THS. This is now solved.
VS non-tiltable: failure in SID lat‐ During the adjustment for the VS non-tiltable (FSF> Adjustment -> System
eral adjustment (GXRCQ 18896) -> SID lat) the message "Unable to get the VS tilt monitoring data from the VS
component" appeared after pressing <Apply>; the completion of the adjust‐
ment was not possible. This problem is now solved.
Tracking possible outside of Tracking is now possible for a vertical beam even when CS is not in lateral or
locks (GXRCQ 17450) longitudinal locks and when tube and auxiliary are not aligned.
Error message due to large num‐ An error message showed up on the CSUI when an exam with more than
ber of vies on CS UI (GXRCQ twenty views was initiated. This problem is now solved, the number of views
18859) that can be displayed is no longer limited to 20.
Failure of longitudinal-range ad‐ For a VM that is positioned at the frontside of the room, the longitudinal range
justment for frontside mounted adjustment failed when it was started at the right side, as requested in the FSF
VM2 (GXRCQ 19419) procedure. After this adjustment, the longitudinal axis could move motorized
at all. This problem is now solved.
Permanent error messages from After a wrong calibration of the beta axis, the problem occurred that the CSP
CSP (GXRCQ 20237) permanently generated error messages and a re-calibration was not possible.
This problem is now solved with CSP SW R10.x.5 Build 090406.
CSP vertical collision retraction The backwards movement of the CSP following a collision was inhibited. This
problem is now solved. The backwards movement will work as required with
CSP R10.x.5 Build 090406, provided the CS height drive controller is updated
to SW version 3.4.0.
CSP movement slow when used The longitudinal and transversal park and APR movements of the CSP were
with VM anti-collision system too slow, when used with a VM (due to interferences with the VM anti-collision
system). This problem has been solved with CSP R10.x.5 Build 090406.

4512 988 03891 REV AB CSIP Level 0 (09.0) 5


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Title Description
CSP database download error The CSP stopped sending records when the first empty APR data was detect‐
ed during the APR database download. This problem has been solved with
CSP R10.x.5 Build 090406.
Unintended tracking movement When free cassette was selected as auxiliary, it occasionally occurred that a
with free cassette (GXRCQ tracking movement was started after a change of mode (e.g. from diagnostic
19641&21534; FPR DMC mode to QA mode) or after a change of view. This has been solved with
00002783, -2786,-2751,-2838, DigitalDiagnost 2.0.2 SP1.
-2775)

1.2.2 Acquisition Workspot (AWS)

1.2.2.1 Application Software

The Application SW version was updated to 2.0.2 SP1.

Title Description
Store-image-to-service function Storing images to the service partition created a lot of undesired data, which
creates undesired data (GXRCQ could reduce the remote service performance. This problem is now solved.
18333)
Incorrect patient type Sporadically, the patient type newborn/baby/child did not come up correctly in
(BST_DMC0002458/GXRCQ correlation to the patient date of birth and the default patient type was used
18605) instead. This is now solved.
Wrong x-ray generator data set Sporadically, after storing some x-ray generator data via the "hand/disk" icon
stored (GXRCQ 19282) on the exam screen, a wrong exam parameter set was stored. This problem is
now solved.
Mirror icon missing in rotated im‐ When an acquired image was freely rotated the "Mirror icon" (mirrored letter
ages (GXRCQ 18016) R) was not visible within the PMS DICOM Viewer for some image export for‐
mats. This problem is now solved
System stops during image ac‐ There were some occurrences where the progress bar stopped when an image
quisition (GXRCQ 19011) was coming in and the system then had to be rebooted. This problem is now
solved.
MPPS cache file corrupt There were some occurrences of corrupted MPPS cache files. This is now
(GXRCQ 18535) solved, corrupted MPPS cache files will be replaced automatically when de‐
tected by the system.
Printer definitions missing Some printer definitions were missing from the system, these have now been
(GXRCQ 19177) added.
Store manual override, also for The "Store Override" button was shown on the UI following a manual change
receptor sensitivity, workspot, of the receptor sensitivity, workspot or amplimat rotation. This was not correct,
amplimat rotation (GXRCQ as these values can be overridden, but overridden values cannot be stored in
19758) the EPX database. The button was removed.
Image Area Dose Product not in‐ When DX images were exported as CR images the Image Area Dose Product
cluded in CR export for DR im‐ was not in the DICOM header. This DICOM information was added.
ages (GXRCQ 19724)
Stitching views could not be For copied examinations, the stitching editor was disabled and images could
stitched manually (GXRCQ not be stitched. This is now solved.
19439)
Missing dose per image for free There was no dose per image information available to free cassette examina‐
cassette examinations (GXRCQ tions. This was corrected, the dose indication is now displayed after exposure.
19015)
TH-S system freeze (GXRCQ Systems with a THS experienced some system slow-downs and system
19669) freezes due to a problem with the CAN message not being processed. This
problem is now solved.

6 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Title Description
Error message after free expo‐ The error message "CALL SERVICE - Digital Flat Detector error" sometimes
sure (GXRCQ 19516) popped up after a free exposure. This happened due to interference with a dark
image sequence. This problem is now solved.
Ready light for next image ac‐ There was a ready light and image acquisition was possible after the message
quisition after flat detector error "CALL SERVICE - Digital Flat Detector error" had appeared following an ex‐
(GXRCQ 18552 &19532) posure, even though image acquisition should not be possible in this situation.
this problem is now solved.
Wrong patient type after error Following the error message "CALL SERVICE - Could not select exam, no
message (GXRCQ 18299) matching APR data found", the patient types were for incorrect for all views.
This problem is now solved.
Export filter calibration has no ef‐ Performing an export filter calibration had no effect. This has been solved with
fect (GXRCQ 17097, 17272, DigitalDiagnost 2.0.2 SP1.
20414, & 20862; FPR
DMC00002710)

1.2.2.2 EVA

Title Description
No display of views on CSCH af‐ In some cases, after a successful validation of the EVA database, all views
ter successful validation of EVA had the status "Skipped" and were not displayed on the CSCH. This is now
database (GXRCQ 19262) solved.
EPX database validation failed The EPX database validation took a long time and sometimes ended with a
due to timeout (GXRCQ 19647) time-out and could not be completed. This problem is now solved.

1.2.2.3 iSite

Title Description
System crash when pressing A system crash occurred when the queue button in iSite was pressed. This
queue button (GXRCQ 20023) problem is now solved.

1.2.3 RIO
Title Description
Safety circuit selftest failure The safety circuit selftest failed with FW version 1.0.4. This problem is solved
(GXRQC 19847) with FW version 1.0.5.

1.2.4 CSP Star SW


Title Description
Wrong system type in CSP Star In the CSP Star software 2R10.x.4 Build 081217, the system type setting was
set to "Bucky TH" when the configuration data was sent to the CSP without
completing the configuration dialogue. This setting has now been changed to
"SYS_DiDi2".

4512 988 03891 REV AB CSIP Level 0 (09.0) 7


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

1.2.5 Hot keys


Key Description
<Ctrl> + <Alt> + <t> Start the Touch Screen Tool
<Shift> + <F11> Take snapshot of full desktop
<Ctrl> + <Alt> + <s> Start the Field Service Framework
<Ctrl> + <Alt> + <F4> Switches on the EEC/Enables digital monitor output (DVI mode)
<Ctrl> + <F10> Grab viewport: Take snapshot of patient image (=image A); to be used for
image comparison
<Ctrl> + <F9> Image comparison:
Show/Hide snapshot image A beside actual viewport (= patient image B)

8 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

2 FURTHER IMPORTANT INFORMATION

2.1 KNOWN RESTRICTIONS


The following restrictions exist for the current DigitalDiagnost release. A workaround procedure is described for
handling each restriction.

2.1.1 Geometry
Title Description Workaround
Move-to-position speed The speed of the Move-to-position motion is NA
(BST_DMC00002513) slightly slower than for the DigitalDiagnost
1.x system due to the fact that the move‐
ments take place in serial order.
CS longitudinal axis reference For systems with a TH table and a VS locat‐ Move the CS into the 'CSOverTh'
ed at the rear or front side of the table, the lock after the system start-up.
motorized CS longitudinal axis can only be
referenced (after system switched on) if the
CS is moved into the 'CSOverTh' lock. This
will be improved in the near future.
CS motor running continuously When the CS moves backwards after a col‐ Press the height brake button
after collision (GXRCQ 19480) lision, it can happen that the motor does not (yellow button) on the CS control
stop running. handle.
Wrong cable labeling for VS tilt For some VS configurations, the cable for NA
poti (GXRCQ 17881) the tilt potentiometer is labeled X9, but
should be X8. Mixing up the cables can lead
to a malfunction of the tilt movement.
Adjusting the wallstand remote From time to time, problems occur when When adjusting the wallstand re‐
control channel number teaching the remote control. mote control channel number,
(GXRQC 18579) the frequency should be 2433
MHz or higher to ensure smooth
operation, and the power should
be set to 0dBm.
Table stitching function not run‐ The stitching does not work when CS in a NA
ning while CS in catch without catch without programmed SID.
SID (GXRQC 20121)
SW download not possible for For some BUF controllers, the embedded Exchange the EPROM U20 (re‐
BUF controller (GXRQC 19256) software download is not possible. fer to Upgrade Manual
DigitalDiagnost 2.0.2 4512 988
0371x).
Beam alignment - change of ori‐ With the upper and lower beam alignment Refer to the application notes.
entation leads to change in field‐ and tracking activated, the following prob‐
size (GXRQC 20456 & 20635) lem occurs: first, the fieldsize is set, then the
orientation is changed. After changing the
orientation, the fieldsize suddenly changes
to an incorrect size.
CSP: some functions do not work Some functions of the CSP are disabled af‐ Restart the system using the
after restart triggered by FSF ter a restart done via the FSF. EEC switch-off and switch-on
(GXRQC 20181) buttons.
De-activation of tracking with re‐ It is not possible to de-activate the tracking De-activate the tracking process
mote control not possible with process via the CSP remote control or the using the CSP control handle.
CSP (GXRQC 18540) wallbox, as was the case with
DigitalDiagnost release 1.x.

4512 988 03891 REV AB CSIP Level 0 (09.0) 9


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Title Description Workaround


CSP: changing of servo The user has the possibility to adjust the ser‐ NA
strengths has no effect (GXRQC vo strengths for beta and transversal
18840) (strength can vary between 0 and 255). Cur‐
rently, changing the servo strength has no
noticeable effect.
CSP does not indicate a collision In the event of a collision, the CSP does give NA
properly (GXRQC 20941) a proper collision indication. There is only a
beep and the message "motor vert" on the
control handle.
CSP: vertical calibration with ta‐ When using the table as reference at the Refer to the newest version of
ble as reference does not work vertical calibration no drive positions ap‐ the CSP Star manual to obtain
(GXRQC 20783) pear. A consequence of this defect is that the detailled information on how to
vertical calibration cannot be performed if perform the vertical calibration.
the user has only a TH2 and a CSP. The
calibration works when you use the wall‐
stand as reference.
CSP: Alpha & beta movement When setting the movement ranges, the The movement ranges for alpha
ranges incorrect (GXRCQ minimum value for the alpha and beta move‐ and beta should not be adjusted,
18841) ment ranges should be a negative value and but should simply be left on de‐
the maximum value should be a positive val‐ fault value.
ue. However, the system currently shows
these values vice versa.

2.1.2 Acquisition workspot (AWS)

2.1.2.1 Application Software

Title Description Workaround


Incorrect monitor solution after After performing a backup and restore, the Perform a second system re‐
restore (GXRCQ 17928) resolution of the EECA monitor (19") is start.
sometimes incorrect after the first restart.
Switching on of AWS immediate‐ When the AWS is switched on again shortly Press the EEC shut-down button
ly after shutting down creates after being shut-down, it sometimes hap‐ and then start up the system
problems (GXRCQ 16679) pens that the geometry an generator are on, again.
but the AWS itself stays switched off.
Stitching images - no matching There were some rare occurrences of the NA
APR data found (GXRCQ error message "CALL SERVICE - Could not
18098) select exam - no matching APR data found"
popping up when a stitching view was se‐
lected.
Exposure after offset calibration Performing an exposure just after the offset This can be rectified by either
leads to system error (GXRCQ calibration has successfully run through and performing a restore of the de‐
19545) while the offset calibration view is selected tector calibration data (if availa‐
will result in a system error; there will be no ble) or by re-booting the system
system-ready available after this. several times.
Virtual keyboard: only letter keys When caps lock is activated, only the letter NA
active with caps lock (GXRCQ keys can be used. All other keys are deacti‐
19356) vated.
Export and print destination have In the Italian language setting, the export Refer to the application notes.
the name in Italian (GXRCQ and print destination both have the same
19533) name.

10 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Title Description Workaround


System hang-up during shut- Some sites have the problem that their RIS Perform a hardware shut-down
down (GXRCQ 20547) cannot handle DICOM compliant "double by pressing the switch-off button
queries", which can lead to a system hang- on the EEC for at least 8 sec‐
up during shut-down. onds.
Mirror icon placed in anatomical For images that are exported or burnt on CD, NA
region of patient (GXRCQ the mirror icon is sometimes overlapping the
20461) anatomy of the patient.
Stitching takes too long for pe‐ Autostitching examinations might not be NA
diatric examinations (GXRCQ suitable for pediatric examinations due to
20596) the length of exposure time.
Waiting time after high dosage After taking an exposure with a high dosage, Refer to application note.
exposures it takes some time until the next exposure
can be taken, as the system takes dark im‐
ages.
Annotations not sent to PACS When data is sent to a second PACS desti‐ NA
(GXRCQ 19693) nation, the annotations and overlays are not
displayed.
Printer error after data restore When upgrading DigitalDiagnost release Before doing an upgrade from
(GXRCQ 20724) 2.0.0 or 2.0.1 to DigitalDiagnost release DigitalDiagnost release 2.0.0 or
2.0.2 SP1, some printers are missing and 2.0.1 to release 2.0.2 SP1, install
have to be reconfigured after a restore of the the service pack "Printer Patch
backup files. Tool DiDi 2.0.0 & 2.0.1".
The patch needs to be installed
before a backup is performed.
The service patch CD
(4512-134-06171) is delivered
with FCO upgrade kit 71200037.

2.1.2.2 Operating system

Title Description Workaround


AWS automatic switch-on It sometimes happens that the AWS com‐
This problem can be solved by
(BST_DMC00002400/GXRCQ puter switches on automatically when the
changing the BIOS settings as
18087) main power supply is switched on. follows:
Chipset - South Bridge Chipset
Configuration:
DiDi 2.x: Restore ON AC Power
Loss = Last State
Canadian Keyboard with French The OS does not accept the Canadian key‐ The solution to this problem is to
layout not accepted (GXRCQ board with French layout. install an English (US) or a
19687) French keyboard.

2.1.2.3 EEC

Title Description Workaround


Touch screen function partly dis‐ Sporadically, it can happen during touch Use the mouse.
abled (GXRCQ 17850) screen adjustment that the touch screen
functionality is partly disabled due to a EEC
hardware problem.

4512 988 03891 REV AB CSIP Level 0 (09.0) 11


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

2.1.2.4 EVA

Title Description Workaround


FSF and EVA to be closed during The FSF and the EVA database need to be NA
examinations closed during patient examinations to pre‐
vent a malfunction of the SW.
Stitching views rejected after val‐ In some cases, the DR stitching views are Validate rejected views sepa‐
idation but license present rejected during a validation of the EVA da‐
rately using "Validate online
(GXRCQ 19644) tabase (using "Validate online force all"),
force all". Afterwards, the valida‐
even though a stitching license is present.
tion for the whole database
should be successful
mAs calculation value incorrect Within the generator settings for PCR stitch‐ Please select “calc” for mAs to
for PCR stitching exams ing spine views, the mAs values are not cal‐ obtain the correct values and
(GXRCQ 19540 & 19559) culated correctly. save and activate the database
as usual.
Variofocus only works for 50/50 EVA offers several variofocus configurations NA
option (GXRCQ 15869) to choose from, but only the 50/50 selection
has an effect.

2.1.2.5 FSF

Title Description Workaround


Working position adjustment For systems with a VM, the <Cancel> oper‐ NA
VM: cancel button does not work ation during the working position adjustment
correctly (GXRCQ19198) in the FSF (FSF: Adjustment ➜ System ➜
Working positions CS-VM) does not work
correctly. When the <Cancel> button is
pressed, the original wallstand adjustment
data is deleted.
Grid correction instruction for KV The instructions for the grid correction for KV Refer to the following corrected
and mAs is misleading (GXRCQ and mAs in the FSF (FSF: Configuration ➜ paragraphs:
19526) Generator ➜ Grid correction steps) are in‐ kV steps: Based on the dose
correct. The kV and mAs steps are dose equivalent of about 25%. Possi‐
equivalent, no configuration is required. ble values: 0 ... 10 (half step in‐
crement: 0.5, 1.0, 1.5, 2.0,
2.5 ...).
mAs steps: Based on the dose
equivalent of about 25%. Possi‐
ble values: 0 ... 10 (decimals pos‐
sible but the system accepts only
0.25 steps, otherwise an error
message appears at the bottom
of the screen).

12 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Title Description Workaround


VS geometry configuration incor‐ For systems with a VS, the geometry con‐ Check and, if necessary, correct
rect after restore figuration is not displayed correctly after athe geometry configuration via
restore of backup data (applicable for an up‐FSF: Configuration ➜ System
grade from DigitalDiagnost 2.0 to 2.0.2). A components ➜ Geometry com‐
VM is displayed instead of a VS. ponents ➜ System components.
If the settings had to be correct‐
ed, activate them and restart the
system.
Error in Configuration of the Ta‐ The section 'Configuration of the Table Re‐ Refer to the corrected procedure
ble Related VM Settings- Under lated VM Settings - Under TH (S) Distance' attached to this document (see
TH (S) Distance under FSF: Configuration ➜ Wallstand chapter Appendix, page 18).
VM➜ Table Distances is currently incorrect.

2.1.3 AGentT
Title Description Workaround
UI does not function when When AGenT is running, the UI cannot be Exit AGentT and perform a re-set
AGenT is running used. of the generator.

2.2 COMPATIBILITY & REQUIREMENTS

2.2.1 Release version levels


The following table shows an overview of the HW and SW release version levels (RVLs) for the DigitalDiagnost
2.0.2. An overview of the versions currently installed on the system can be obtained in the FSF via System infor‐
mation/Surveys ➜ Release version levels.

Name Description and Digital Diagnost Release 2.0.1 Digital Diagnost Release 2.0.2
Circuit board ID (February 2008) SP1 (August 2009)
Ceiling Suspension CS2/4 and CSP
CS Collimator Automatic collimator SW Version:01.04.01
NICOL V2 (LA)
PCB: 9896 010 22161
CS Collimator Automatic collimator SW Version:01.00.02
NICOL V3 (LA)
(limited field monitoring)
CS Control Handle Control handle con‐ SW Version: 01.06.013
(CSCH) troller (UE2)
PCB: 4512 201 07189
CS beta drive (CAN) CS beta drive con‐ SW Version: 11.22.001 6 (CD: 4512 201 05594)
(CSP only) troller (URN1)
PCB: 4512 130 234936
CS beta drive (motor)
(CSP only)
CS height drive Tracking 2 controller SW Version: 03.04.001 (CD: 4512 201 05594)
(CAN) (CS height) (UAP)
PCB: 4512 133 02005
CS height drive (mo‐
tor)
CS longitudinal drive Longitudinal Drive SW Version: 11.22.001 (CD: 4512 201 05594)
(CAN) Controller (longitudi‐
PCB: 4512 130 23492 (until remaining stock is depleted) / 4512 130
CS longitudinal drive nal) (UBN2)
23493
(motor)
CS tilt drive (CAN) Alpha Drive Control‐ SW Version: 11.22.001 (CD: 4512 201 05594)
ler (Tilting) (UFN1 )

4512 988 03891 REV AB CSIP Level 0 (09.0) 13


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Name Description and Digital Diagnost Release 2.0.1 Digital Diagnost Release 2.0.2
Circuit board ID (February 2008) SP1 (August 2009)
CS tilt drive (motor) PCB: 4512 130 23492 (until remaining stock is depleted) / 4512 130
23493

CS trans drive (CAN) CS trans drive con‐ SW Version: 11.22.001 6 (CD: 4512 201 05594)
(CSP only) troller (UYN2)
PCB: 4512 130 234936
CS trans drive (mo‐
tor) (CSP only)
CSP controller SW Version: 10.x.4 - 090107 SW Version: 10.x.5 (Build
(4512 201 06751) 7 090406) (CD: 4512 201 06752)7
CSP Star SW SW Version: 10.x.5 (Build
090406) (CD: 4512 201 06752)8
EPX/EVA
EPX server database EPX Application da‐ SW Version: 02.23.02.6-1510 SW Version: 02.23.03.6-1610
tabase
(CD: 4512 133 46731 (CD: 4512 133 46732 EPX Da‐
EPX Database) tabase)
EPX Service data‐ SW Version: 02.23.02.6-8 SW Version: 02.23.02.6-9
base (not included in
(CD: 4512 133 46731 (CD: 4512 133 46732
RVL overview)
EPX Database) EPX Database)
Eva offline tool (not SW Version: 4.2.8.15 V605
included in RVL over‐
(CD: 4512 133 46732
view)
EPX Database)
Firewall
Firewall commercial mShield SW Version: 3.1.01
version
(CD: 4512 133 54583)

HW: 9890 010 04163


Firewall software ver‐ SW Version: 3.1.0.65171
sion
(CD: 4512 133 54583)

HW: 9890 010 04163


Acquisition Workspot (AWS)
Operating system SW Version: 2.0.41
version
(DVD: 4512 133 46721
- Operating System Recovery DVD AWS Rel. 2.0.4)
Software version Application software SW Version: 2.0.1(Incr. 13.7.4)1 SW Version: 2.0.2 SP1 (Incr.
13.7.12)1,
(CD: 4512 201 05592 (CD: 4512 201 05594 - CD Ap‐
- CD Applications Software plications Software
DigitalDiagnost Rel. 2.0.1) DigitalDiagnost Rel. 2.0.2 SP1
BIOS Version (not in‐ SW Version:8.0.11
cluded in RVL over‐
view)
BIOS ID (not includ‐ EEA1C907
ed in RVL overview)

14 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Name Description and Digital Diagnost Release 2.0.1 Digital Diagnost Release 2.0.2
Circuit board ID (February 2008) SP1 (August 2009)
Acquisition Workspot 4512 201 04836
(AWS) PC (not in‐
cluded in RVL over‐
view)
Eleva Examination 9890 010 87393
Control (EEC) (not in‐
cluded in RVL over‐
view)
Generator
Optimus CXA (not in‐ 9890-000-70361 OPTIMUS CXA 50 1T 400V (low speed)
cluded in RVL over‐ 9890-000-70371 OPTIMUS CXA 50 1T 480V (low speed)
view) 9890-000-70501 OPTIMUS CXA 50 1T 480V (dual speed)
9890-000-70381 OPTIMUS CXA 65/80 1T 400V (dual speed)
9890-000-70391 OPTIMUS CXA 65/80 1T 480V (dual sp)
9890-000-70401 from Q1/2009 OPTIMUS CXA 65/80 2T 400V
(dual sp)
9890-000-70411 from Q1/2009 OPTIMUS CXA 65/80 2T 480V
(dual sp)
RIO embedded ver‐ Room-Input-Output SW Version: 01.00.041 (CD: SW Version: 01.00.051 (CD:
sion (RIO) embedded ver‐ 4512 201 05592) 4512 201 05594)
sion FW (TRB)
PCB: 4512 201 06731 PCB: 4512 201 06732
X-ray Generation X-ray Generation SW Version:01.03.03 2

(EZ139 - CU)
Tables TH-S/TH
TH
TH Bucky unit (BU2) BUF controller (TH) SW Version: 02.03.02 9 (CD: SW Version: 02.03.02 BL. 1.0.9
FW version (SAVAC1) 4512 201 05592) 9
(CD: 4512 201 05594)

PCB: 4512 133 14943 PCB: 4512 133 14943; EPROM:


4512 201 06891
TH Table FW version Bucky controller SW Version: 01.00.01 (4512 114 32001 exchange of 3 ICs)
3

(SZ1)
PCB: 4512 108 07509
TH Tray Drive CAN Tray drive controller SW Version: 11.22.001 (CD: 4512 201 05594)
FW version (SAN1)
PCB: 4512 130 23492 (until remaining stock is depleted) / 4512 130
TH Tray Drive Motor
23493
FW version
TH-S
TH-S Controller TH-S Controller SW Version: 03.00.011 (CD: 4512 201 05594)
(SZ1)
PCB: 4512 133 26543
Wallstands VS/VM/VM2/VM Compact
Wall stand Bucky unit BUF controller (Wall‐ SW Version: 02.03.02 9 (CD: SW Version: 02.03.02 BL. 1.0.9
(BU2) stand) (VAC1) 4512 201 05592) 9
(CD: 4512 201 05594)

PCB: 4512 133 14943 PCB: 4512 133 14943; EPROM:


4512 201 06891
Wallstand Control‐ VS
lers Tilt arm controller SW Version: 03.00.071 (CD: 4512 201 05594)
(VT1)
PCB: 4512 133 02005

4512 988 03891 REV AB CSIP Level 0 (09.0) 15


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

Name Description and Digital Diagnost Release 2.0.1 Digital Diagnost Release 2.0.2
Circuit board ID (February 2008) SP1 (August 2009)
Wallstand controller SW Version: 03.00.07 (CD: 4512 201 05594)
1

vertical (VP1)
PCB: 4512 133 02005
VM
Tilt arm controller SW Version: 03.00.071 (CD: 4512 201 05594)
(VP2)
PCB: 4512 133 02005
Wallstand controller SW Version: 03.00.071 (CD: 4512 201 05594)
vertical (VP1)
PCB: 4512 133 02005
VM 2
Tilt arm controller SW Version: 03.00.071 (CD: 4512 201 05594)
(VP2)
PCB: 4512 133 02005
Wallstand controller SW Version: 03.00.071 (CD: 4512 201 05594)
vertical (VP1)
PCB: 4512 133 02005
Horizontal control‐ SW Version: 03.00.071 (CD: 4512 201 05594)
ler(longitudinal)
PCB: 4512 133 02005
(VL2)
VM compact
Tilt arm controller SW Version: 03.00.071 (CD: 4512 201 05594)
(VP2)
PCB: 4512 133 02005
Wallstand controller SW Version: 03.00.071 (CD: 4512 201 05594)
vertical (VP1)
PCB: 4512 133 02005
Wallstand Remote Remote control SW Version: 01.00.00
Controllers (transmitter)
Transmitter: 4512 133 02912
Remote controller SW Version: 01.01.023
(receiver)
PCB: 4512 133 02902
Virus Guard
Virus guard definition SW Version:4.0.5182
Virus guard engine SW Version:5.2.00
Virus guard product SW Version:8.0.0.x
PCR
PCR Reader - Inte‐ PCR Reader SW Version: 1.7 or higher4
gration (Eleva S CR-
▪ 4512 201 0428x (Eleva S)
IR 357) or Eleva S
Plus CR-IR 356) (not ▪ 4512 201 0429x (Eleva S-Plus)
included in RVL over‐
view)
PCR System - Read‐ Application SW version: 1.2
er Sharing Mode (not
included in RVL over‐
view)

1. Software download can be done via FSF:General functions ➜ Install. Software download versions are part of
the CD Applications Software DigitalDiagnost Rel. 2.0.2 (4512 201 05594). Folder on the CD: Embedded
Software for download. The labels on the EPROMS do not necessarily show the firmware versions currently

16 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

in use for DigitalDiagnost 2.x. due to software download of embedded software. To check the software and
hardware versions of the components refer to FSF: System information/Surveys ➜ Release version levels.
2. Software download can be done via AGenT (version 5.1 or higher) ➜ Program\Update generator software.
3. No Software download option, only PROM exchange.
4. The SW version of the PCR reader can be checked via FSF: Configuration ➜ PCR ➜ Start Maint. Utility for
PCR Reader (PC-MUTL).
5. The current version of the EVA database can be V60 or higher, this does not affect the functionality.
6. PCB must be version 4512 130 23493, otherwise the SW download will not work. If PCB is a lower version,
the update of the software can be done via a PROM exchange, but this will prevent future SW downloads.The
preferred solution would therefore be to exchange the PCB for a newer version (4512 130 23493).
7. SW update possible via Flash Tools 3 or in DOS mode with a bootable CD.
8. SW to be installed on service PC.
9. EPROM exchange required to upgrade SW for controllers delivered before release 2.0.2. Controllers delivered
with release 2.0.2 do not require an EPROM exchange.
10. If an older version of the EPX database is converted and used, the version level will not change to the current
EPX database release, but the converted database is still compatible with the current release of the
DigitalDiagnost system.

2.2.2 Configuration
The following configuration options are available for the DigitalDiagnost 2.0.2:

DigitalDiagnost 2.0.2 Single Detector System

Components:

▪ Wallstand BuckyDiagnost VS digital/VS digital tilting


▪ Wallstand DigitalDiagnost VM/VM2/VM Compact
▪ DigitalDiagnost Table height adjustable TH (BUF)
▪ BuckyDiagnost Table height adjustable single sided suspended TH-S
▪ BuckyDiagnost Ceiling Suspension CS 2/4
▪ Ceiling Suspension Programmable CSP-V3 (only in combination with DigitalDiagnost CS4)
▪ Generator Optimus CXA

DigitalDiagnost 2.0.2 Dual Detector System

Components:

▪ Wallstand BuckyDiagnost VS digital/VS digital tilting


▪ Wallstand DigitalDiagnost VM/VM2
▪ DigitalDiagnost Table height adjustable TH (BUF)
▪ BuckyDiagnost Ceiling Suspension CS 2/4
▪ Ceiling Suspension Programmable CSP-V3 (only in combination with DigitalDiagnost CS4)
▪ Generator Optimus CXA

2.3 MANPOWER, ESTIMATED TIME, TOOLS


NA

4512 988 03891 REV AB CSIP Level 0 (09.0) 17


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Release Bulletin DigitalDiagnost 2.0.2 SP1

3 APPENDIX
Below you will find the corrected FSF procedure Configuration of the Table Related VM Settings - Under TH (S)
Distance.

18 CSIP Level 0 (09.0) 4512 988 03891 REV AB


© 2009 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Configuration of the Table Related VM Settings
Under TH (S) Distance
Defines the distance between the cover of the Bucky unit and the upperside of the tabletop (see figure). The system will take the
thickness of the tabletop into account automatically.

Examples
Measured value Thickness of the tabletop Distance Bucky unit <> underside of the
Table
[mm] [mm] tabletop [mm]
TH-S 80* 55 25
TH
53* 28 25
750mm
TH
63* 38 25
850mm
* Measured at the head- or foot-end side of the tabletop center (refer to the figures below)

NOTE
If the distance between Bucky unit and the underside of the tabletop is less than 25mm the detector and the tabletop may collide
during MTP and alignment movements.
Correct the distance by checking the following adjustments: FSF: Adjustment > Ceiling suspension 2/4 > CS height;
Adjustment > Table TH/TH-S > Table height; Adjustment > Wallstand VS/VM > VS/VM height.

VM under TH (S) distance

TH (S) Pick-up Band


Within this band the VM Bucky unit will follow the tabletop according to the configured value "Under TH (S) Distance" whenever the
table height changes.

VM/TH (S) Pick-up band

You might also like