HISTORY RECORD 1

W
PHILIPS MEDICAL SYSTEMS
RECORD OF SYSTEM PROGRAMMING 2

DigitalDiagnost
VRïE RECORD OF DEFAULT USER PROGRAMS 3

Level 0
Documentation
RECORD OF RECOMMENDED SETTINGS 4
Part 2 (SRM)

RECORD OF MEASURED DATA 5

SERVICE INFORMATION 6

SOFTWARE RELEASE BULLETINS 7


Software, Firmware and PCB versions

GRID DATA SHEET 8

DMC Hamburg
This document and the information contained in it is proprietary
and confidential information of Philips Medical Systems
(”Philips”) and may not be reproduced, copied in whole or in part,
CABLE AND EARTH DIAGRAMS 9
adapted, modified, disclosed to others, or disseminated without
the prior written permission of the Philips Legal Department.
This document is intended to be (a.) used by customers and is
licensed to them as part of their Philips equipment purchase or
(b.) used to meet regulatory commitments as required by the
FDA under 21 CFR 1020.30 (and any amendments to it) and
other local regulatory requirements. Use of this document by
unauthorized persons is strictly prohibited.
Printed in Hamburg
PACKING LIST 10
Federal Republic of Germany

(06.0)
 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED

Manual Order No. 4512 988 02101 REV AA


4512 983 02871
11 0.1
DigitalDiagnost VR-E

Blank page

0.2 CSIP Level 0 (06.0) 4512ï983ï02871


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost VR-E

List of pages and modules (LOPAM)

4512 980 54001 Binder

Transparent plastic sheet

4512 983 02871 (TITLE) 0.1 (06.0)


4512 983 02871 (Blank page) 0.2 (06.0)
4512 983 02871 (LOPAM) 0.3 (06.0)
4512 983 02871 (Blank page) 0.4 (06.0)
4512 983 02871 (LOPAM) 0.5 (06.0)

SYSTEM REFERENCE MANUAL

Section 1: History Record


4512 983 08041 Section Page 1-1 (06.0)
4512 983 11291 General information 1-1 (06.0)

Section 2: Record of System Programming


4512 983 08051 Section Page 2-1 (04.1)
4512 983 08801 Program settings of the Optimus RAD generator 2-1 (06.0)

Section 3: Record of Default User Programs


4512 983 08061 Section Page 3-1 (04.0)

Section 4: Record of Recommended Settings


4512 983 08071 Section Page 4-1 (04.0)
4512 983 08641 Programming and jumper-setting 4-1 (06.0)

Section 5: Record of Measured Data


4512 983 08081 Section Page 5-1 (04.0)
4512 983 03141 Bucky Systems ï System status report 5-1 (04.1)
4512 983 03191 Programming and adjustment 1 (04.1)
4512 983 03201 Programming and adjustment 1 (04.0)
4512 983 03211 Programming and adjustment 1 (04.0)
4512 983 03231 General / safety aspects 1 (04.0)

Section 6: Service Information


4512 983 08091 Section Page 6-1 (05.0)
4512 985 00401 DigitalDiagnost Detector Exchange Report 1
4512 983 10351 DigitalDiagnost System labeling 2Zï10 (05.0)

4512ï983ï02871 CSIP Level 0 (06.0) 0.3


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost VR-E

Blank page

0.4 CSIP Level 0 (06.0) 4512ï983ï02871


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost VR-E

Section 7: Release Bulletins


4512 983 08101 Section Page 7-1 (05.0)
4512 983 10831 Software, Firmware and PCB versions DigitalDiagnost 7-1 (06.0)

Section 8: Grid data sheet


4512 983 11381 Section Page 8-1 (06.0)
smit röntgen grid data sheet (06.0)

Section 9: Cable and Earth Diagrams


4512 983 08111 Section Page 9-1 (05.1)
4512 982 0106. Dimensions DigitalDiagnost Workstation Zï1.1 (05.0)
4512 982 0146. Cable lenghts DigitalDiagnost VR-E Zï4.2 (03.0)
4512 982 0030. Mains connection with transformer and surge arrester Zï5.4 (05.0)
Mains connection (400 V mains) Zï5.5 (05.0)
4512 983 07051 System connection diagram Zï27 (05.1)
System earthing diagram Zï28 (05.1)
4512 983 10721 DigitalDiagnost VE / VT (moveable) CAN connections Zï30.2 (05.0)
4512 983 06851 Legend for cabling diagram Zï50 (03.0)

Section 10: Packing List


4512 983 08121 Section Page 10-1 (04.0)

Section 11:

4512ï983ï02871 CSIP Level 0 (06.0) 0.5


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
History record DigitalDiagnost VR-E with FS Fix

Section 1 History record

Contents

1. General information

2.

4512ï983ï08041 CSIP Level 0 (06.0) 1ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
History Record DigitalDiagnost

LIST OF PAGES AND MODULES MODULE CODE NUMBER: 4512ï983ï11291

1-1...1-2 (06.0)

General information

1. Introduction

This document contains all relevant programmings,


settings and configurations according to which the
DigitalDiagnost systems are preset in the factory and
delivered from Philips Medical Systems.
The programmings, settings and configurations which
are not specific to the system are described in the
System Manual Installation (Level 0 Documentation
Part 1).

2. Abbreviations

Abbreviation Explanation
General
AGenT Access Generator Tool
APR Anatomically Programmed Radiography
BIST Build In Self Test
BWLM Basic Worklist Management
BWS Bucky Wall Stand
CAN Control Area Network
COM Serial port
DICOM Digital Imaging and Communication in Medicine
DiDi DigitalDiagnost
FCO Field Change Order
FRU Field Replaceable Unit
MPPS Modality Performed Procedure Step
MSB Mini System Bus
MTP Move To Position
PCB Printed Circuit Board
PDU Power Distribution Unit
PCR Philips Computerised Radiography
PMS Philips Medical Systems
POST Power On Self Test
RGDV Registration Device (auxiliary unit)
UPS Uninterruptible Power Supply

4512ï983ï11291 CSIP Level 0 (06.0) 1ï1


 2006 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost History Record

Abbreviation Explanation
Documentation
SMI System Manual Installation (Level 0 Documentation Part 1)
SRM System Reference Manual (Level 0 Documentation Part 2)

System names
DigitalDiagnost TH table and optional VM or VS wall stand
DigitalDiagnost VM VM wall stand and optional a TH-S table or TF-M trolley
DigitalDiagnost VM Compact VM Compact wall stand and optional a TF-M or Traumob-X trolley
DigitalDiagnost VR VS wall stand and optional a TF-M trolley

Subsystem names
CS 2/4 Ceiling Suspension (2 = short longitudinal carriage / 4 = long longitudinal carriage)
FS fix Floor Stand fixed
TH Table Height adjustable
TH-S Table Height adjustable Single side suspended
TF-M Trolley Fixed Moveable
TH-M Trolley Height adjustable Moveable
VM Vertical Multipurpose stand moveable
VM Compact Vertical Multipurpose stand fixed
VS Vertical stand with fixed bucky module
VS tilting Vertical stand with tiltable bucky module

Unit names
BUF Bucky Unit Family
RO Rotalix tube (RO 1750)
SCU Segment control unit
SRO Super Rotalix tube (SRO 0951, SRO 2550, SRO 33100)

Unit lettering (documentation)


C Generator control desk
E Generator cabinet
EG High tension tank
LA Collimator
S Table
SA Bucky module (table)
U Ceiling suspension
UB Longitudinal carriage (ceiling suspension)
UZ Telescopic carriage (ceiling suspension)
V Wall stand
VA Bucky module (wall stand)
VP Electronic rack (VM wall stand) or wall connection box (VS wall stand)
VS Segment control unit (SCU)
WN Surge arrester
XA X-Ray tube

1ï2 CSIP Level 0 (06.0) 4512ï983ï11291


 2006 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of system programming DigitalDiagnost VR-E with FS Fix

Section 2 Record of system programming

Contents

1. Program settings of the Optimus RAD generator

2.

4512ï983ï08051 CSIP Level 0 (04.1) 2ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of system programming DigitalDiagnost

LIST OF PAGES AND MODULES MODULE CODE NUMBER: 4512ï983ï08801


2ï1...2ï7 (06.0)

Program settings of the Optimus RAD generator

Contents

1. Registration devices (RGDV’s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï2


1.1. Programming for DigitalDiagnost TH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï2
1.2. Programming for DigitalDiagnost VR / VM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï3
1.3. Programming for DigitalDiagnost TH with SECOND CS . . . . . . . . . . . . . . . . . . . . . 2ï4
2. Film / screen combinations (S200, S400, S800) . . . . . . . . . . . . . . . . . . . . . . . . . 2ï5
2.1. 3-Field amplimat chamber (9890 000 0267.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï5
2.2. 5-Field amplimat chamber (9890 000 70001) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï5
3. APR programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï6
3.1. System with ceiling suspension CS2 / CS4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï6
3.2. System with floor stand FS fix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2ï7

4512ï983ï08801 CSIP Level 0 (06.0) 2ï1


ã 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost Record of system programming

1. Registration devices (RGDV’s)

NOTE

These DigitalDiagnost specific program settings are already done in the factory !

1.1. Programming for DigitalDiagnost TH


If the option tomography and / or a wallstand is not part of the system the tube of the related RGDV must be set
to “none”.

RGDV 1 RGDV 2 RGDV 3 RGDV 4


Table Tomo Wallstand Free Exposure
(option) (option)
Dataset A:
Room 1 1 1 1
Tube 1 1 1 1
Release Circuit number 1 1 1 1
Enable handswitch at generator desk yes yes yes yes
Syncmaster present yes yes yes yes
Exposure switch type Double Step Double Step Double Step Double Step
Bucky format density correction 0 0 0 0
cone density correction 0 0 0 0
Dose measuring input EZX21 EZX21 EZX31 none
Dose measuring sensor Bucky Amplimat Bucky Amplimat Bucky Amplimat Bucky Amplimat
Exposure series/Tomo movement no no no no
Release delay enabled enabled enabled enabled
Mounted radiography controller BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag.
Release circuit adaptation unit none none none none
Mounted tomo extension none none none none
Medium II format kV correction 0 0 0 0
Medium II format density correction 0 0 0 0
Medium II format mAs correction 0 0 0 0
Small II format kV correction 0 0 0 0
Small II format density correction 0 0 0 0
Small II format mAs correction 0 0 0 0
Data Set B
Used for tomo no yes no no
Used for flouroscopy no no no no
CT add on no no no no
Disable time override no no no no
tube power factor [%] [100] [100] [100] [100]
kV steps Dose equiv. Dose equiv. Dose equiv. Dose equiv.
mAs steps 25% 25% 25% 25%
mA steps 25% 25% 25% 25%
time steps 25% 25% 25% 25%
Density steps 12% 12% 12% 12%
Density correction [6% steps] [0] [0] [0] [0]
Underesposure display yes yes yes yes
Tube overload protection on on on on

2ï2 CSIP Level 0 (06.0) 4512ï983ï08801


ã 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of system programming DigitalDiagnost

1.2. Programming for DigitalDiagnost VR / VM / VM Compact

RGDV 1 RGDV 2 RGDV 3 RGDV 4


Table Tomo Wallstand Free Exposure
Dataset A:
Room 1 1
Tube none none 1 1
Release Circuit number 1 1
Enable handswitch at generator desk yes yes
Syncmaster present yes yes
Exposure switch type Double Step Double Step
Bucky format density correction 0 0
cone density correction 0 0
Dose measuring input EZX31 none
Dose measuring sensor Bucky Amplimat Bucky Amplimat
Exposure series/Tomo movement no no
Release delay enabled enabled
Mounted radiography controller BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag.
Release circuit adaptation unit none none
Mounted tomo extension none none
Medium II format kV correction 0 0
Medium II format density correction 0 0
Medium II format mAs correction 0 0
Small II format kV correction 0 0
Small II format density correction 0 0
Small II format mAs correction 0 0
Data Set B
Used for tomo no no
Used for flouroscopy no no
CT add on no no
Disable time override no no
tube power factor [%] [100] [100]
kV steps Dose equiv. Dose equiv.
mAs steps 25% 25%
mA steps 25% 25%
time steps 25% 25%
Density steps 12% 12%
Density correction [6% steps] [0] [0]
Underesposure display yes yes
Tube overload protection on on

4512ï983ï08801 CSIP Level 0 (06.0) 2ï3


ã 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost Record of system programming

1.3. Programming for DigitalDiagnost TH with SECOND CS


If the option tomography is not part of the system the tube of RGDV 2 must be set to “none”.

Table tube Wallstand tube


RGDV 1 RGDV 2 RGDV 4 RGDV 5 RGDV 6
Table Tomo Free Exposure Wallstand Free Exposure
(option) Table Wallstand
Dataset A:
Room 1 1 1 1 1
Tube 1 1 1 2 2
Release Circuit number 1 1 1 1 1
Enable handswitch at generator desk yes yes yes yes yes
Syncmaster present yes yes yes yes yes
Exposure switch type Double Step Double Step Double Step Double Step Double Step
Bucky format density correction 0 0 0 0 0
cone density correction 0 0 0 0 0
Dose measuring input EZX21 EZX21 none EZX31 none
Dose measuring sensor Bucky Amplimat Bucky Amplimat Bucky Amplimat Bucky Amplimat Bucky Amplimat
Exposure series/Tomo movement no no no no no
Release delay enabled enabled enabled enabled enabled
Mounted radiography controller BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag. BuCo 1 / Dig.Diag.
Release circuit adaptation unit none none none none none
Mounted tomo extension none none none none none
Medium II format kV correction 0 0 0 0 0

Medium II format density correction 0 0 0 0 0

Medium II format mAs correction 0 0 0 0 0

Small II format kV correction 0 0 0 0 0

Small II format density correction 0 0 0 0 0

Small II format mAs correction 0 0 0 0 0

Data Set B
Used for tomo no yes no no no
Used for flouroscopy no no no no no

CT add on no no no no no

Disable time override no no no no no


tube power factor [%] [100] [100] [100] [100] [100]
kV steps Dose equiv. Dose equiv. Dose equiv. Dose equiv. Dose equiv.
mAs steps 25% 25% 25% 25% 25%
mA steps 25% 25% 25% 25% 25%
time steps 25% 25% 25% 25% 25%
Density steps 12% 12% 12% 12% 12%
Density correction [6% steps] [0] [0] [0] [0] [0]
Underesposure display yes yes yes yes yes
Tube overload protection on on no on on

2ï4 CSIP Level 0 (06.0) 4512ï983ï08801


ã 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of system programming DigitalDiagnost

2. Film / screen combinations (S200, S400, S800)

NOTE

These DigitalDiagnost specific program settings are already done in the factory !

2.1. 3-Field amplimat chamber (9890 000 0267.)


Ÿ Connect the service PC to the generator.
Ÿ Start AGenT.
Ÿ Program the sensitivity S200, S400 and S800 for the DigitalDiagnost TH (chamber 1) and/or
DigitalDiagnost VE/VT (chamber 3).
Select: Program  Dose Rate Control  Amplimat  Chamber 1 + 3  Data set 1 . .3
“Start Automatic DRC Processing ?“  yes

film FILM.TDL  PHIL. TRI-FSXD


screen SCREEN.TDL  PHIL. TRI-FSXD S200 (Data Set 1) and
S400 (Data Set 2) and
S800 (Data Set 3)
chamber CHAMBER.TDL  ap DBucky
(user side wall stand = left side of the wall stand)
or
pa Dbucky
(user side wall stand = right side of the wall stand)
cassette CASSETTE.TDL  normal Cassette (def)
system correction SYSCOR.TDL  no corr. (ISO9236-1)
correction factor  1

Ÿ Transmit the data.


Ÿ Restart the generator.

2.2. 5-Field amplimat chamber (9890 000 70001)


Ÿ Connect the service PC to the generator.
Ÿ Start AGenT.
Ÿ Program the sensitivity S200, S400 and S800 for the DigitalDiagnost TH with BUF (chamber 1) and/or
DigitalDiagnost VM / VM Compact / VS (chamber 3).
Select: Program  Dose Rate Control  Amplimat  Chamber 1+3  Data set 1 . . 3
“Start Automatic DRC Processing ?“  yes

film FILM.TDL  PHIL. TRI-FSXD


screen SCREEN.TDL  PHIL. TRI-FSXD S200 (Data Set 1) and
S400 (Data Set 2) and
S800 (Data Set 3)
chamber CHAMBER.TDL  Bucky 5 Field

cassette CASSETTE.TDL  normal Cassette (def)


system correction SYSCOR.TDL  no corr. (ISO9236-1)

correction factor  1

Ÿ Transmit the data.


Ÿ Restart the generator.
4512ï983ï08801 CSIP Level 0 (06.0) 2ï5
ã 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost Record of system programming

3. APR programs

NOTE

These DigitalDiagnost specific program settings are already done in the factory !

Ÿ Connect the service PC to the generator.


Ÿ Start the APR-Manager.
Ÿ Transfer the required file. Make sure system type, language, tube and variofocus match your system.

xx = language = GB = English DD = DigitalDiagnost


DE = German TH, VE/VT, VM, VM Compact, VS
FR = French V = variofocus
ES = Spanish 9 = SRO 0951
IT = Italian
SV = Swedish

3.1. System with ceiling suspension CS2 / CS4

DigitalDiagnost with Ceiling Suspension CS2 / CS4


APR-file variofocus tubes system type

DD15xx.tdl without RO 1750 SRO 2550 SRO 33100


DD15Vxx.tdl with RO 1750 SRO 2550 SRO 33100 DigitalDiagnost
TH, VE/VT, VM,
DD15V9xx.tdl with SRO 0951 VM Compact, VS

NOTE

For the data transfer the CAN bus cables EZX42, EZX43 and the signal bus EZX23 must be disconnected !

Ÿ After successful download reconnect the CAN bus cables EZX42, EZX43 and the signal bus EZX23.
Ÿ Restart the generator.

2ï6 CSIP Level 0 (06.0) 4512ï983ï08801


ã 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of system programming DigitalDiagnost

3.2. System with floor stand FS fix

NOTE

The APR-files in the following table can only be used in the system configuration DigitalDiagnost VR-E with
BuckyDiagnost FS fix !

DigitalDiagnost VR-E with Floor Stand FS fix


APRïfile variofocus tubes SID

DD18xx.tdl without RO 1750 SRO 2550 SRO 33100


DD18Vxx.tdl with RO 1750 SRO 2550 SRO 33100 170 – 210 cm
DD18V9xx.tdl with SRO 0951

DD22xx.tdl without RO 1750 SRO 2550 SRO 33100


DD22Vxx.tdl with RO 1750 SRO 2550 SRO 33100 210 – 260 cm
DD22V9xx.tdl with SRO 0951

NOTE

For the data transfer the CAN bus cables EZX42, EZX43 and the signal bus EZX23 must be disconnected !

Ÿ After successful download reconnect the CAN bus cables EZX42, EZX43 and the signal bus EZX23.
Ÿ Restart the generator.
J

4512ï983ï08801 CSIP Level 0 (06.0) 2ï7


ã 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of default user programs DigitalDiagnost VR-E with FS Fix

Section 3 Record of default user programs

Contents

1.

4512ï983ï08061 CSIP Level 0 (04.0) 3ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of recommended settings DigitalDiagnost VR-E with FS Fix

Section 4 Record of recommended settings

Contents

1. Programming and jumper-setting

2.

4512ï983ï08071 CSIP Level 0 (04.0) 4ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of recommended settings DigitalDiagnost

LIST OF PAGES AND MODULES MODULE CODE NUMBER: 4512ï983ï08641


4ï1...4ï9 (06.0)

Programming and jumper-setting

Contents

1. DigitalDiag. TH/VR/VM/VM Comp. sys. with one ceil. susp. CS or fl. stand FS . . 4ï2
1.1. Control handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï2
1.2. Tracking controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï2
1.3. Longitudinal motor controller CS2/4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï3
1.4. Alpha motor controller CS2/4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï3
1.5. Bucky tray motor controller TH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï4
1.6. Nicol Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï4
1.7. Bucky controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï5
2. DigitalDiagnost TH with SECOND CS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï6
2.1. Control handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï6
2.2. Tracking controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï6
2.3. Tomo controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï7
2.3.1. Longitudinal motor controller CS2/4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï7
2.3.2. Alpha motor controller CS2/4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï7
2.3.3. Bucky tray motor controller TH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï8
2.4. Nicol Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï8
2.5. Bucky controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï9
2.6. X-Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4ï9

4512ï983ï08641 CSIP Level 0 (06.0) 4ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost Record of recommended settings

1. DigitalDiagnost TH / VR / VM / VM Compact system with one ceiling suspension CS


or floor stand FS

All jumper-settings and programming are already done in the factory.

1.1. Control handle

Jumper UE2 W1
Pins 2ï3 bridged (room 1)
CAN node name: CSU_HD1

X9
UE2 H1

3
2
1

W1 W2 W3

1.2. Tracking controller

Jumper UAD X7
Pins 1, 2, 3, 4, 6, 7 bridged
Pins 5, 8 open
CAN node name: SLSCSD1

X6

Jumper
UAD
X7

1
1

4ï2 CSIP Level 0 (06.0) 4512ï983ï08641


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of recommended settings DigitalDiagnost

1.3. Longitudinal motor controller CS2/4


Only available if option “tomography”, option “automatic stitching” or option “VM hor. motorized” installed.

Jumper UBN2 X7
Pins 1, 2, 3, 4, 5, 6, 7, 8 bridged
no pins open
CAN node name: TomoDv1

X6

Jumper
UBN2
X7

1 1

1.4. Alpha motor controller CS2/4


Only available if option “tomography”, option “automatic stitching” or option “VM hor. motorized” installed.

Jumper UFN1 X7
Pins 1, 2, 4, 5, 6, 8 bridged
Pins 3, 7 open
CAN node name: AlphDv1

X6

Jumper
UFN1
X7

1 1

4512ï983ï08641 CSIP Level 0 (06.0) 4ï3


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost Record of recommended settings

1.5. Bucky tray motor controller TH


Only available if option “tomography” installed.

Jumper SAN1 X7
Pins 1, 2, 3, 4, 5, 6, 7 bridged
Pin 8 open
CAN node name: TrayDv1

X6

Jumper
SAN1
X7

1 1

1.6. Nicol Collimator

The jumper LA1 W1 is located behind the ruler side plate in the collimator.

Jumper LA1 W1
Pins 1ï2 bridged (room 1)
CAN node name: BLD1DV1

X6

1
1
2 W1
LA1 3

MP1 ï

MP2 ï
X2
MP3 ï

MP4 ï

4ï4 CSIP Level 0 (06.0) 4512ï983ï08641


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of recommended settings DigitalDiagnost

1.7. Bucky controller

Jumper SZ1 S1
Dip-switch 1 set to closed (room 1)
Dip-switch 2 set to open (CAN slave)
Dip-switch 3 set to closed (X-Scope)
CAN node name: RADCtl1

O
P
E
N
SZ1 8
7
6
5
4
VTï100 3 XïScope
slave 2 master
room 2 1 room 1

S1

4512ï983ï08641 CSIP Level 0 (06.0) 4ï5


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost Record of recommended settings

2. DigitalDiagnost TH with SECOND CS


All jumper-settings and programming are already done in the factory.

NOTE
All settings are only applicable in a DigitalDiagnost system with two ceiling suspensions (bucky controllers) !

2.1. Control handle

Ceiling suspension of the TABLE Ceiling suspension of the Wallstand


Jumper UE2 W1 Jumper UE2 W1
Pins 2ï3 bridged (room 1) Pins 1ï2 bridged (room 2)
CAN node name: CSU_HD1 CAN node name: : CSU_HD2

X9 X9
UE2 H1 UE2 H1

3 3
2 2
1 1
W1 W2 W3 W1 W2 W3

2.2. Tracking controller

Ceiling suspension of the TABLE Ceiling suspension of the Wallstand


Jumper UAD X7 Jumper UAD X7
Pins 1, 2, 3, 4, 6, 7 bridged Pins 1, 2, 3, 4, 7 bridged
Pins 5, 8 open Pins 5, 6, 8 open
CAN node name: SLSCSD1 CAN node name: SLSCSD2

X6 X6

8 8

Jumper Jumper
UAD UAD
X7 X7

1 1 1
1

4ï6 CSIP Level 0 (06.0) 4512ï983ï08641


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of recommended settings DigitalDiagnost

2.3. Tomo controller

2.3.1. Longitudinal motor controller CS2/4

Only available if option “tomography” installed.

Jumper UBN2 X7
Pins 1, 2, 3, 4, 5, 6, 7, 8 bridged
no pins open
CAN node name: TomoDv1

X6

Jumper
UBN2
X7

1 1

2.3.2. Alpha motor controller CS2/4

Only available if option “tomography” installed.

Jumper UFN1 X7
Pins 1, 2, 4, 5, 6, 8 bridged
Pins 3, 7 open
CAN node name: AlphDv1

X6

Jumper
UFN1
X7

1 1

4512ï983ï08641 CSIP Level 0 (06.0) 4ï7


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost Record of recommended settings

2.3.3. Bucky tray motor controller TH

Only available if option “tomography” installed.

Jumper SAN1 X7
Pins 1, 2, 3, 4, 5, 6, 7 bridged
Pin 8 open
CAN node name: TrayDv1

X6

Jumper
SAN1
X7

1 1

2.4. Nicol Collimator


The jumper LA1 W1 is located behind the ruler side plate in the collimator.

Ceiling suspension of the TABLE Ceiling suspension of the Wallstand


Jumper LA1 W1 Jumper LA1 W1
Pins 1ï2 bridged (room 1) Pins 2ï3 bridged (room 2)
CAN node name: BLD1DV1 CAN node name: BLD2DV1

X6 X6
1 1
1 1
2 W1 2 W1
LA1 3 LA1 3

MP1 ï MP1 ï
MP2 ï MP2 ï
X2 X2
MP3 ï MP3 ï
MP4 ï MP4 ï

4ï8 CSIP Level 0 (06.0) 4512ï983ï08641


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of recommended settings DigitalDiagnost

2.5. Bucky controller

TABLE Wallstand (SCU)


Jumper SZ1 S1 Jumper SZ1 S1
Dip-switch 1 set to closed (room 1) Dip-switch 1 set to open (room 2)
Dip-switch 2 set to open (CAN slave) Dip-switch 2 set to open (CAN slave)
Dip-switch 3 set to closed (X-Scope) Dip-switch 3 set to closed (X-Scope)
CAN node name: RADCtl1 CAN node name: RADCtl2

O O
P P
E E
N N
SZ1 8
SZ1 8
7 7
6 6
5 5
4 4
VTï100 3 XïScope VTï100 3 XïScope
slave 2 master slave 2 master
room 2 1 room 1 room 2 1 room 1

S1 S1

2.6. X-Scope
Differences in the programming to a DigitalDiagnost system with one ceiling suspension:

Bucky controller Bucky controller


TABLE WALLSTAND
SZX1 VSZX1
è Room Configuration è Table exposures è yes no
è Wallstand exposures è no yes
è Tomography exposures è yes (if installed) no
è Free Cassette è yes yes
exposures
è Vertical Tracking inst. è yes yes

4512ï983ï08641 CSIP Level 0 (06.0) 4ï9


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Record of measured data DigitalDiagnost VR-E with FS Fix

Section 5 Record of measured data

Contents

1. Read me first

2.

4512ï983ï08081 CSIP Level 0 (04.0) 5ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
System status report BuckyDiagnost Systems

LIST OF PAGES AND MODULES MODULE CODE NUMBER: 4512ï983ï03141


5-1 (04.1)
4512 983 03191 (04.1)
4512 983 03201 (04.0)
4512 983 03211 (04.0)
4512 983 03231 (04.0)

Read me first

Read me first
System status report for manufacture and field installation
Introduction:

The Bucky system is adjusted in the factory according to the actual service documentation.

The status report list gives an overview about:


ï Traceable items / system configuration.
ï Settings and adjustments during the pre-installation in the factory.
ï Values and ranges for acceptable deviations.
ï For the final handing over to the customer the status report must be completed from the field engineer.

Contents:

1. System configuration for traceable items

2. Programming and adjustments for

Only the marked text segments are included.


- BuckyDiagnost CS 2/4, FS and TraumaDiagnost (4512 983 03191) 3 pages
- BuckyDiagnost TH2 / TF and DigitalDiagnost TH (4512 983 03201) 3 pages
- BuckyDiagnost VE / VT / VR / VE2 / VT2 (4512 983 03211) 1 page

3. General / Safety aspects (4512 983 03231) 1 page

NOTE

The requirements listed in this section may differ from the requirements of national regulations (e.g. RÖV).
Measured values outside this status report must be inside the limitation of national regulations.

4512ï983ï03141 CSIP Level 0 (04.1) 5ï1


 2002 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side
2. Programming and adjustment
The mechanical installation must be done in order of the installation manual.
For adjusting and programming fill out the attached list.

2.1. BuckyDiagnost CS 2/4 , FS and TraumaDiagnost

Generator + BuckyDiagnost CS 2/4

Labelling on the keyboard - -


Generator is programmed - -
This programming must be checked especially for main power condition
U = 400 V R = 100 m -
U = .......... V R = ........... m -
Generator with tube 1 Type .......... is adapted Serial No. ............... -
Generator with tube 2 Type .......... is adapted Serial No. ............... -
All functions are in order with the operator manual - -
AMPLIMAT function is tested L C R measuring field -
Insert cassette 24 cm x 30 cm (lengthways and vertical format).
All measuring fields (L, C, R) selectable.
Insert cassette 18 cm x 24 cm (lengthways and vertical format).
Only center measuring field selectable.

Emergency switch off by < 600 mAs tested - -


Check of free exposure

Standard APR settings - -


Individual APR settings -
Programmings for individual film screen systems -
Jumper setting Basic Interface EZ150 : BuckyDiagnost : Pos. W4ï1 -
DigitalDiagnost : Pos. W4ï1 -
Software programming AMPLIMAT sensitivity BuckyDiagnost : high -
DigitalDiagnost : high -

4512ï983ï03191 CSIP Level 0 (04.1) 1/3


E 1997 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side
Collimator + BuckyDiagnost CS 2/4, FS

Type of collimator
Manual - Automatic -

For your information:

Manual + Automatic
Power supply for coll. bulb = ............ V (11...12 V) (measured on bulb) -

Coincidence of X-ray field and light field

Exposure data: SID = 100 cm


small focus
.......... kV
.......... mAs

Test images for sensing - light ï X-ray -


Enclosed this report.

The tolerances in the table are valid for an SID = 100cm.


The tolerances change linearly depending on the SID.

The maximum deviation between light and X-ray is allowed to be as follows at all four edges. (see table)

automatic manual
Tol. +5 / ï3 mm +9 / ï3 mm

The maximum deviation at the top/bottom edge and left/right edge is 5 mm.
In case the light field is smaller than the X-ray field, use minus signs.

measured deviations
Top : mm Bottom : mm Diff. : mm
Left : mm Right : mm Diff. : mm

4512ï983ï03191 CSIP Level 0 (04.1) 2/3


E 1997 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side
BuckyDiagnost CS 2/4, FS

All mechanical movements are in order to the internal checklist (factory side). -
All functions are in order to the operator manual. - -
Lamp functions ï breaks ï catches ï end stops ï cabling. - -
The detents must be programmed with X-Scope -
BuckyDiagnost CS 2/4:
The sensing function is tested
in longitudinal direction - transversal direction - - -
BuckyDiagnost FS:
The sensing function is tested - -

Tracking

BuckyDiagnost VE / VT / VR / VE2 / VT2

Function of remote control (light, collimator) during system READY -


Pick-up band (def. 300 mm) -
End-position potentiometer -

BuckyDiagnost CS 2/4

Current limit during tracking -


Pick-up band (def. 300 mm) -

TraumaDiagnost

(not implemented)

Order No.: ..........................................................................................................

Customer: ................................................... Room: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Technician/Name Signature Date


Factory
Service

4512ï983ï03191 CSIP Level 0 (04.1) 3/3


E 1997 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side
2. Programming and adjustment
The mechanical installation must be done in order of the installation manual.
For adjusting and programming fill out the attached list.

2.1. BuckyDiagnost TH2 / TF and DigitalDiagnost TH

Function in order to the operator manual - -


Programming in X-Scope customer name, room No. and basic settings
must be checked and completed -
Table height ï standard:Table height measured on
min. = ............ mm (450  5 mm) cassette surface to ground - -
max. = ............ mm (850  5 mm) digital : Sensitive layer of the flat
detector to ground - -

BuckyDiagnost CS height
(This value is depending on your room height and therefore also adjusted on field side).
min. = ............ mm -
max. = ............ mm

From both values the SID is calculated.


For testing this value the real value may differ max  2 cm.
Displayed SID = ............. cm ”Real” SID = ............ cm - -
If the deviation is too large the table height and CS height adjustment must be repeated.

NOTE

Motor current for table up (under max. load) C = c1 20mF+ c2 16mF


(50 Hz / 230 V) I = A (< 3.5 A)
without c2 I = A (< 2.5 A)

Nominal value = 120 ms "4ms -


No grid lines visible by exposure time tw3 msec.

4512ï983ï03201 CSIP Level 0 (04.0) 1/3


E 1997 Koninlijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side
Tomo + TH and Tomo 2 + TH2

Function in order to the operator manual - -


Servo assistance during tomo run o.k. - -
Image quality Tomogram:

Angle Tomo time in s Layer height in mm Layer height deviation " 3mm
40_ 1.2 45
40_ 1.2 130
40_ 1.2 199

Image resolution in images enclosed this test report.


With standard test grid (50 funk raster)
limiting resolution = 2.32lp/mm with small focus (spotcheck) -
actual resolution = .......... lp/mm -
Layer height
Measured deviation by 130 mm layer height (max. deviation is 3 mm) (spotcheck) -
nominal layer height = ........ mm actual layer height = .......... mm -

NOTE

The layer height is adjustable with ”tid” under X-Scope. tid= .......... mm -
tid nominal = 65 mm Range of adjustment = 65 ... 75 mm
tid ACL nominal= 70 mm Range of adjustment = 70 ... 80 mm
tid flat detector = 78 mm Range of adjustment = 70 ... 90 mm

4512ï983ï03201 CSIP Level 0 (04.0) 2/3


E 1997 Koninlijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side

Extent of the X-ray beam (spotcheck) - -


The extent of the X-ray beam with Tomo is tested by
layer height = 130 mm SID = 115 cm Tomo angle = 40_ 1,2 sec = Exposure time

jAj ï jBj = v 6 mm
A = .......... mm
B = .......... mm
Maximum extent of the X-ray beam
jAj + jBj = v 40 mm Center beam
A ........ mm B ........ mm

Symmetry (spotcheck) - -
The symmetry of the X-ray beam with Tomo is tested by
layer height = 130 mm SID = 115 cm Tomo angle = 40_ 1,2 sec = Exposure time

jCj ï jDj = v 5 mm
C = .......... mm
D = .......... mm
C D

..........mm ..........mm

Order No.: ..........................................................................................................

Customer: ................................................... Room: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Technician/Name Signature Date


Factory
Service

4512ï983ï03201 CSIP Level 0 (04.0) 3/3


E 1997 Koninlijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side
2. Programming and adjustment
The mechanical installation must be done in order of the installation manual.
For adjusting and programming fill out the attached list.

2.1. BuckyDiagnost VE / VT / VR / VE2 / VT2

AMPLIMAT function -
All mechanical movements and functions are in order to the operator manual. -
Nominal value = 120 ms "4ms -
No grid lines visible by exposure time tw3 msec.

Order No.: ..........................................................................................................

Customer: ................................................... Room: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Technician/Name Signature Date


Factory
Service

4512ï983ï03211 CSIP Level 0 (04.0) 1/1


E 1997 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
BuckyDiagnost Systems

factory field
side side
3. General / Safety aspects

System configuration:

Generator OPTIMUS 50 - 65 - 80 -

Room 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Room 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Room 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Emergency switch off is checked -


Fault current switch tested -
Connected earth cable and check the bucky/digital DIAGNOST TH, bucky wall stand (BWS), CS..
and measure the earth resistance test - -
The earth resistance (as measured above) must always be < 100 m
(central point = generator) -
All end stops are installed, specially on the CS rails -

Functional test of collisions switches -

Unit Down Up Remarks


TH table v 200 N completely assembled including table top and covers
CS v 100 N weight compensation performed
Motor VE / VT 70 N $ 10 N 70 N $ 10 N completely assembled and weight compensation performed

NOTE

Do not adjust the TH table collision switch SSI10 (factory calibrated).

CS covers are fixed -


Exception test according RÖV / VA / DOD or other national regulations -

Order No.: ..........................................................................................................

Customer: ................................................... Room: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Technician/Name Signature Date


Factory
Service

4512ï983ï03231 CSIP Level 0 (04.0) 1/1


E 1997 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Service information DigitalDiagnost VR-E with FS Fix

Section 6 Service information

Contents

1. DigitalDiagnost Detector Exchange Report

2. DigitalDiagnost System labeling

3.

4512ï983ï08091 CSIP Level 0 (05.0) 6ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Please send this document immediately to Hamburg (FAX) and send the original
document with the defective detector.

FAX: +49/40/5078-2348 (SERVICE SUPPORT digital DIAGNOST)

RE-USE the delivered/original packaging for the return


of the Detector !!!
Also make sure that the transport conditions are within the limits for shock, temperature
and humidity.

Country :_____________________
City :_____________________
Customer :_____________________ Room: _______________

SAP number of the System :_____________________ Refer to the Software


System Serial number :_____________________ Licence sheet.
Detector Serial # (new) :_____________________
Detector Serial # (defect,old) :_____________________ Installation Date (old Detector):___ ___ ______
Date of replacement :____ ____ _______ Time in use (old Detector) :
Years : ___ Months : ___
Wallstand-Detector † Table-Detector †

Engineer (name) :_____________________


Telephone : ____________________ E-Mail: ___________@___________________

Additional material sent with the defective DETECTOR ( mark like this : )

Films † Images / JPG Images (DOR,Floppy) †


Logfiles † Results from the IQ Tool (Floppy) †

Shipping address for the Detector:


Philips Medical Systems
Service Logistics/Return Dept.
Building G
Röntgenstr. 24
D-22335 Hamburg /GERMANY

SERVICE LOGISTIC / Returns :


FAX: +49/40/5078-1011
Tel: +49/40/5078 –1741/-1747

4512 985 00401 page 1 of 2 H


DIDI_DetectorExchangeReport.doc PMS, DMC Hamburg
Development GXR
Problem descriptions

Mark the visible artifact / location like this :


1) "One Panel failure"

The detector
consists of 4
panels.

2) "One amplifier failure"

Examples:
There are 13
readout amplifiers
on each panel.

3) Local defects 4) Local defects 5) Visible 6) Visible


"one spot" "water drops" line (horizontal) column

Sketch:
Comment
Detector Ser# : ..................
Date : ..................

† Problem in each image reproducible?


† Problem in few images, intermitting?
† Improvement with a new calibration?
† None of this Types matches !

sketch

Please make a sketch of the


localisation of the problem.

4512 985 00401 page 2 of 2 H


DIDI_DetectorExchangeReport.doc PMS, DMC Hamburg
Development GXR
OPTIMUS RAD 50/65/80
X

X
Logo Address

Type No. Electr. data CE

IEC 60601-1 IEC 60601-2-32 UL / CSA

only glued on China systems

Logo Address

Type No. Electr. data

IEC 60601-1 IEC 60601-2-32


A3/A4 05-02-17 Schr.

DigitalDiagnost
System labelling

4512 983 10351 CSIP Level 0 (05.0) 2Z-10


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Software release bulletins DigitalDiagnost VR-E with FS Fix

Section 7 Software release bulletins

Contents

1. Software, Firmware and PCB versions DigitalDiagnost

2.

4512ï983ï08101 CSIP Level 0 (05.0) 7ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Software release bulletins Digital Diagnost

LIST OF PAGES AND MODULES MODULE CODE NUMBER: 4512ï983ï10831


7-1 (06.0)

Software, Firmware and PCB versions DigitalDiagnost

NOTE

All other combinations of software, firmware and PCB are not tested / released and will lead to malfunction or
do not work at all.

All rel 1.3 system in the field were upgraded to rel 1.3.1 (FCO 712 00 005) and all rel 1.4 systems were upgraded to
rel 1.4.1 (FCO 712 00 009).

Release

Application SW Workstation 1.2.3 1.3 1.3.1 1.4 1.4.1 1.5

First delivery from factory Aug 2002 May 2003 May 2004 Oct 2004 May 2005 Xxx 2006

FW 6.2 7.2 8.1 9.1 9.2 10.2


Bucky Controller
PCB 4512 108 07506 + 07508 4512 108 07508

Optimus RAD Generator 3.5 3.6 4.1

CAN 1.5
Tracking Controller (CS + FS)
Motor 1.5

CAN 2.1 2.2 2.x


Tomo Controller (CS + FS)
Motor 2.4 2.x

Control Handle (CS + FS) 3.2 3.3

FW 1.1 2.1 2.2


IR/CAN Controller (VE/VT)
PCB 4512 131 45281 4512 131 45283

Vertical Controller (VM+VS) FW ï 1.0.1 1.1.0 / 1.1.1 2.x.x


Tilt Arm Controller (VM+VS) 4512 133
PCB ï 4512 133 02003
Horizontal Controller (VM2) 02004

BUF Controller (TH+VM+VS) ï 1.0.1 1.1.0 2.x.x

NOTE

Please be aware of the different jumper- and DIP-switch settings between the PCB- and related firmware
version !
J
4512ï983ï10831 CSIP Level 0 (06.0) 7ï1
 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Grid data sheet DigitalDiagnost VR-E with FS Fix

Section 8 Grid data sheet

Contents

1. smit röntgen grid data sheet

2.

4512ï983ï11381 CSIP Level 0 (06.0) 8ï1


 2006 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
smit röntgen

P.O. Box 218


5600 MD Eindhoven
The Netherlands
smit
Tel. +31 40 27 62707
Fax. +31 40 27 62478

SERIAL NUMBER: QUALITY CONTROL

PERFORMANCE DATA OF Contrast improvement ratio, Application limits


ANTIïSCATTER GRIDS: Selectivity, Grid exposure factor,
Grid Quality factor (K, Σ, B, Q) see
These are the focus distance limits
table 1.
In publication IEC 60627 2001 for use of a grid, where the loss of
Second Edition definitions, primary radiation at the edge of the
The primary, total and scatter
determination and indication of grid is less than 40%, compared to
transmission (Tp, Tt and Ts) are
characteristics of antiïscatter grids the primary radiation at the grid
measured according to IEC 60627,
are described. centre.
at 80 kV.

Contrast improvement ratio K: The limits are a function of grid


All our grids comply to this standard
(contrast factor) width, focusing distance and grid
Transmission of ratio.
Engraved on the grids are: primary radiation Tp
K+ +
ï number of strips per centimeter Transmission of Tt
total radiation For the various grid ratios,
ï grid ratio (r) dimensions and focusing distances
ï focusing distance (fo) in Selectivity Σ the application limits are shown in
centimeters Transmission of table 2.
primary radiation Tp
ï serial number and type number S+ +
Transmission of Ts
scattered radiation

Absorber material:
Grid expose factor B: (Buckyïfactor)

Our grids are manufactured with 1 1


B+ +
lead strips. Transmission of Tt
total radiation

Interspace material: Grid Quality factor Q:

Q+K Tp
The interspace material is fiber

CSIP Level 0 (06.0)


Table 1

Line rate/ratio Tp B (+10%) K (+10%) S (+10%) Q(+10%)

36 / 8 76% 4.5 3.4 5.9 2.6

36 / 10 76% 5.0 3.8 7.2 2.9

36 / 12 75% 5.5 4.1 8.6 3.1

40 / 8 78% 4.3 3.4 5.6 2.6

40 / 10 77% 4.7 3.6 6.4 2.8

40 / 12 75% 5.2 3.9 7.8 2.9

40 / 15 74% 5.8 4.3 9.7 3.1

44 / 8 78% 4.1 3.2 5.0 2.6

44 / 10 77% 4.7 3.6 6.4 2.9

44 / 12 75% 5.2 3.9 7.9 3.0

44 / 15 74% 5.7 4.2 9.4 3.1

60 / 10 72% 5.3 3.8 7.4 2.7

60 / 13 71% 5.7 4.1 8.5 2.9

60 / 15 70% 6.1 4.2 9.7 2.9

70 / 13 71% 5.1 4.0 8.2 2.6

70 / 17 70% 5.7 4.4 11.8 2.9

80 / 13 70% 5.0 3.5 6.4 2.5

80 / 15 70% 5.4 3.8 7.3 2.6

CSIP Level 0 (06.0)


Application Limits

Table 2
Ratio 6 : 1 Grid Width
f0 18 cm 24 cm 30 cm 36 cm 43 cm
80 cm 50 ï 196 cm 55 ï 144 cm 59 ï 124 cm 62 ï 114 cm 64 ï 106 cm
100 cm 57 ï 386 cm 64 ï 225 cm 69 ï 180 cm 73 ï 159 cm 76 ï 145 cm
140 cm 69 ï ∞ cm 79 ï 630 cm 86 ï 371 cm 92 ï 291 cm 98 ï 247 cm
180 cm 77 ï ∞ cm 90 ï ∞ cm 100 ï 90 cm 108 ï 540 cm 116 ï 407 cm

Ratio 8 : 1 Grid Width


f0 18 cm 24 cm 30 cm * 36 cm 43 cm
80 cm 55 ï 144 cm 60 ï 120 cm 63 ï 109 cm 65 ï 103 cm 67 ï 98 cm
* 110 cm 64 ï 225 cm 71 ï 171 cm 75 ï 150 cm 78 ï 138 cm 81 ï 130 cm
* 140 cm 79 ï 630 cm 88 ï 336 cm 95 ï 263 cm 101 ï 229 cm 106 ï 208 cm
180 cm 90 ï ∞ cm 103 ï 720 cm 113 ï 450 cm 120 ï 360 cm 127 ï 310 cm

Ratio 10 : 1 Grid Width


f0 18 cm 24 cm 30 cm 36 cm 43 cm
80 cm 59 ï 124 cm 63 ï 109 cm 66 ï 102 cm 68 ï 97 cm 70 ï 94 cm
100 cm 69 ï 180 cm 75 ï 150 cm 79 ï 136 cm 82 ï 129 cm 84 ï 123 cm
140 cm 86 ï 371 cm 95 ï 263 cm 102 ï 223 cm 107 ï 203 cm 111 ï 189 cm
180 cm 100 ï 900 cm 113 ï 450 cm 122 ï 346 cm 129 ï 300 cm 135 ï 271 cm

Ratio 12 : 1 Grid Width


f0 18 cm 24 cm 30 cm * 36 cm 43 cm
80 cm 62 ï 114 cm 65 ï 103 cm 68 ï 97 cm 70 ï 94 cm 71 ï 91 cm
* 110 cm 73 ï 159 cm 78 ï 138 cm 82 ï 129 cm 84 ï 123 cm 87 ï 118 cm
* 140 cm 92 ï 291 cm 101 ï 229 cm 107 ï 203 cm 111 ï 189 cm 115 ï 179 cm
* 180 cm 108 ï 540 cm 120 ï 360 cm 129 ï 300 cm 135 ï 270 cm 141 ï 250 cm

Ratio 14 : 1 Grid Width


f0 18 cm 24 cm 30 cm 36 cm 43 cm
80 cm 64 ï 107 cm 67 ï 99 cm 69 ï 94 cm 71 ï 92 cm 72 ï 90 cm
100 cm 76 ï 147 cm 81 ï 131 cm 84 ï 124 cm 86 ï 119 cm 88 ï 115 cm
140 cm 97 ï 252 cm 105 ï 210 cm 111 ï191 cm 115 ï 180 cm 118 ï 172 cm
180 cm 115 ï 420 cm 126 ï 315 cm 134 ï 274 cm 140 ï 252 cm 145 ï 237 cm

* Used grid in a DigitalDiagnost system

CSIP Level 0 (06.0)


Cable and earth diagrams DigitalDiagnost VR-E with FS Fix

Section 9 Cable and earth diagrams

Contents

1. Dimensions DigitalDiagnost Workstation

2. Cable lenghts DigitalDiagnost VR-E

3. Mains connection with transformer and surge arrester

4. Mains connection (400 V mains)

5. Connection diagrams

6. Earthing diagram

7. DigitalDiagnost VR with DigitalDiagnost VE / VT (moveable) CAN connections

8. Legend for cabling diagram

9.

4512ï983ï08111 CSIP Level 0 (05.1) 9ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
U VP

E
ME C Work- Modem Uninterruptable
station Telephone power supply
S line (for Workstation only)

S UPS
C
U (option)

D15m EZX46 dh data C300X1 D15m


Telephone
line
M2m EZX20 da supply C200X1
basic
EZX4:1

Emerg.
off
EZX4:2
MEX
L2 switched bn
T1 ENK2:3/L2
L2´ continuous power blk
T2 ENF2:4/T2
T3
N bl
N ENX1101
SX
L2 switched bn
T1 SX50
earth lekage L2´ continuous power blk
T2
circuit breaker L2 switched bn 1 SX51
T3 power supply
N bl da L2´ continuous power blk 3 SF1
N SX52
N bl 5
mains fuses Surge arrester
mains feeder (option) SX53
L1 L1 L1
ENF1:1 da power supply VPX53 M3
L2
L2 L2 M3 SX54
L3 WN ENF1:3 * A neutral conductor is not
UZX9
L3 L3 ENF1:5 required if the mains transformer M3f da power supply SX55
N 9890 000 02301 is ordered
N* N ENX1101
from * SX56

Hospital local mains 230V/115V 50/60Hz 16A


local mains 230V/115V 50/60Hz 16A

supply
tube housing
EX1100 XA 1
1 U
1 U 1 1
2 V
2 V 2 2
XA 3 W
3 db stator cable W
3 3

dc selected tube UEX5

EZX3 dc temperature switch Micro switch

O3 plugs H.V. cables XA 1


EG

X-ray tube

D25m EZX23 de signal bus SZ1X1 D25m

D9m EZX43 dd CAN SZ1X2 D9m


UAX1.1 D9f dd CAN D9m SZ1X3
M4 VPX6 Foot switch

D15m EZX31 dd amplimat chamber dd amplimat chamber VPX4 D15f

D9f SZ1X12 dd grid control VPX1 D9f

VP D9m SZ1X10 dd grid position VP1X2 D9m

D9m SZ1X11 dd position BWS VP1X3 D9m

D9m SZ1X5 dd tracking VP1X5 D9m

D9m VP3X3 dd CAN CPCX1 D9f

D9f VP3X2
R

D9f VPX100 dd communication D9m RSX2 D9f dn FPCX2 MD6m

VPB1 FOL (fibre optic link) FPCB2

digital wall stand

D9f EZX42 dd CAN CPCX11 D9m

SUN workstation
A2 05-09-15 Ost./Kö

DigitalDiagnost VR-E
with BuckyDiagnost FS Fix
OPTIMUS RAD
System connection diagram

4512 983 07051 CSIP Level 0 (05.1) Z-27


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
U VP

E
ME C Work- Modem Uninterruptable
station Telephone power supply
S line (for Workstation only)

S UPS
C
U (option)

EZX6 included in supply cable

MEX100 Surge arrester


(option)
protective earth of hospital
WN E
PE8
SX13
gn/yw PE8
UAX11
PE8
VPX11
PE8
UXA

included in supply cable

included in supply cable

Ceiling rails are always insulated,


and not connected to any earth potential.
A2 05-09-15 Ost./Kö

DigitalDiagnost VR-E
with BuckyDiagnost FS Fix
OPTIMUS RAD
System earthing diagram

4512 983 07051 CSIP Level 0 (05.1) Z-28


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
DigitalDiagnost
Workstation
BuckyDiagnost CS rear
side
CPC
X1
X11

UAD
Servo tracking

X4 X3

DigitalDiagnost
VE/VT (moveable)
UE
Control
handle
Optimus generator
LA1
X9 X10 Collimator UF
X1 X11 X3
VP3
EZ X43 X42 X3
R X2
R

S
X3 X2

SZ1 Bucky
controller
A3 05-08-24 Schr.

DigitalDiagnost VR
with DigitalDiagnost VE/ VT (moveable)
CAN connections

4512 983 10721 CSIP Level 0 (05.0) Z-30.2


C Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Heading symbols Connectors top decade plug D-SUB 25 pins D-SUB 15 pins D-SUB 9 pins D-SUB 9 pins
4512 104 0235. 2422 025 05039 2422 025 05038 2422 025 05456 2422 034 18334
C control desk gender changer 9pins male male male female
E cabinet generator
4512 104 4512 104
FS bucky DIAGNOST Floor Stand
LA collimator female male
10
MEX wcb generator 9
8
S bucky table for bucky DIAGNOST TH, 7
6
70 54 40 32 32
Segment control unit for bucky DIAGNOST VR or 5
4

Stand of duo DIAGNOST/easy DIAGNOST 3


2

SA film plane 1

30
15 15
UA column (bucky DIAGNOST Floor Stand) 15
UB ceiling crane longitudinal carriage (bucky DIAGNOST CS)
65 15 43 38 38
UZ ceiling crane transverse carriage (bucky DIAGNOST CS)
43
VE/VT vertical DIAGNOST
VP wcb vertical DIAGNOST 12
WS wallstand
XA X-ray tube crimp pin male 2422 034 11636
wcb wall connection box 11BNC-connector crimp pin female 2422 034 11632
2432 020 00272 distance piece (M3, 11mm) 2422 034 20331

connection block incl. MATE-N_LOCK 12 pins MATE-N_LOCK 3 pins MATE-N_LOCK 6 pins


Cable (text) symbols wall connection box 2422 025 02908 2422 025 02991 2422 034 16253
D9m EZX43 dd ( ) system CAN SZ1X2 D9f plug element 29 14.5 4512 103 75384 3x4 male 1x3 male 3x2 male
cable function
for entry of current cable number MATE-N_LOCK 9 pins
type of cable 2422 034 16589
designation of plug 11BNC-connector 3x3 male
m=male, f=female
2432 020 00273 T1
type of plug 2432 020 00369 T2 9 x 10mm 2 (30 width) (25 width) (20 width)
T3
cable to be laid N
N
top decade 37 30 23
M3 M3 MATE-N-LOCK 3 pins L1

24.5 14.5 L2
D37 D37 D-SUB 37 pins
28 28
L3 5 x 25mm 2 28
D25 D25 D-SUB 25 pins
D15 D15 D-SUB 15 pins
N crimp pin female crimp pin male
2422 015 14048 2422 034 17788
D9 D9 D-SUB 9 pins
MD6 MD6 MINI-DIN 6 pins

cable delivered
Cables
A68 A68 AMPLIMITE 050 68 pins
Type Description Diameter Codenumber
9f 9m gender changer 9 pins MINI-DIN 6 pin
Connection blocks da 3+1 x 1.38 mm2 8.0 mm 0722 207 03044
db 3 x 1.3 mm2 screened + yl 9.2 mm 0722 215 02054
MEX designation of connection block
1 1 dc 2 x 0.8 mm 2 6.2 mm 0722 300 01011
2 2 connection block
3 3 dd 5x2 x 0.22 mm2 screened 7.4 mm 0722 215 62001
4 4 screw junction
5 5 50 de 12x2 x 0.22 mm2 screened 9.6 mm 0722 215 51007
terminal number
df 10 x 0.5 mm 2 7.3 mm 4512 100 66451
dg 10 x 0.5 mm2 screened 7.8 mm 0722 997 07248
Earthing dh 8x2 x 0.22 mm2 screened + yl 8.4 mm 0722 215 33008
earth connection block
dk 8 x 0.75 mm 2 screened 9.6 mm
MEX designation of connection block
2 16 mm2
dm 6x2 x 0.75 mm2 screened 12.6 mm
5 25 5 earthing strip
dn 2x2 x 0.22mm2 screened com.chan.
A3 03-05-12 DP/Kö

5.6 mm 0722 215 29006


screw junction 2x35 mm2
6 10 2 6
do 40.0 mm xxxx xxx xxxxx
2
7 10 7 terminal number
4 mm 2
PE16 16.7 mm 2 9.3 mm 4512 100 66131
8 10 2 8
earth connector PE8 8.25 mm2 6.82 mm 0722 186 005..
H.V. O3-O3, 150 kV 16.5 mm 9806 402 6xx02
max. wire diameter 15
FOL fibre optic link 2.9 mm xxxx xxx xxxxx
PE protective earthing yellow/green wire
(xx = length in m)
FE functional earthing
red wire
AE antiinterference earthing
Legend for cabling diagram

4512 983 06851 CSIP Level 0 (03.0) Z-50


c Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Packing list DigitalDiagnost VR-E with FS Fix

Section 10 Packing list

Contents
1.

4512ï983ï08121 CSIP Level 0 (04.0) 10ï1


 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED