1 Applicability................................................................................................................................. 2
1.1 THALES SOFTWARE .............................................................................................................. 2
1.2 PID SOFTWARE/FIRMWARE. ................................................................................................. 2
2 PU Configuration. ........................................................................................................................ 3
3 Configuration of the SD4343 calibration software. ......................................................................... 6
3.1 Presetting the detector's Working Temperature.......................................................................... 9
4 Configuration of JUNO I software. ............................................................................................. 11
- After the connection with the PU RF4343 is established, click AcppBox Settings
-
- and check the settings according to the screenshot here below:
- The New calibration window opens. Check that the Processing Unit IP Address is
192.168.0.2 (manually input it, if needed).
During installation of the equipment and before calibrating the detector, you need to carry out the
detector working temperature preset procedure.
This procedure involves two settings in the detector temperature control system:
1) Preliminary setting, where the speed at which the fans rotate is constant until the thermal
stability of the detector is reached (i.e. after at least 3.5 hours of use).
Using the SD4343 application, access the Configuration panel and then set the parameters
below:
2) Final setting, where the automatic fan speed control is set to keep the detector temperature
constant at the thermal stability value.
The final setting must be set when the detector has reached its thermal stability (i.e. after at
least 3.5 hours of use).
1) Find the temperature reached by the detector, shown in the Pixium Manager 1:1 page;
a) Check that the proper software/firmware release has been installed: Choose Help-----> About
menu and check that the displayed release is displayed like below:
b) Check that the options of the GENERAL SETUP window are as in the following picture:
Note:
The DUAL SCREEN check box has to be checked only in case the system includes the reference
monitor.
Click to check
next menu
h) The default settings in the SINGLE IMAGE FILM SETTINGS window are as in the following
picture (check only in case the option is present):
Villa SM
l) The default settings in the DICOM STORAGE COMMITMENT SETUP window are as in the
following picture (check only in case the option is present):
Logical Mode PU Mode Left Border Right Border Top Border Bottom Border
1 3 7 7 2 2
2 7 0 0 2 2
3 15 0 0 2 2
4 27 0 0 0 0
5 2 7 7 2 2
6 7 0 0 2 2
7 22 0 0 2 2
8 27 0 0 0 0
10 17 20 20 3 3
11 4 10 10 2 2
15 17 20 20 3 3
16 17 20 20 3 3