JUNO I CONFIGURATION CHECK.

1 Applicability................................................................................................................................. 2
1.1 THALES SOFTWARE .............................................................................................................. 2
1.2 PID SOFTWARE/FIRMWARE. ................................................................................................. 2
2 PU Configuration. ........................................................................................................................ 3
3 Configuration of the SD4343 calibration software. ......................................................................... 6
3.1 Presetting the detector's Working Temperature.......................................................................... 9
4 Configuration of JUNO I software. ............................................................................................. 11

Juno I General Preset 11-15 Rev. 25 Nov 2015 1/21


1 Applicability.
The described settings applies to the following SW configuration set:

1.1 THALES SOFTWARE


PU SIDE
Release 3.0.2.1
HOST SIDE
PixDemoApp Release 3.0.2.

1.2 PID SOFTWARE/FIRMWARE.

SD4343 Release 1.3.0.2


HIRIS RF43 Release 8.0.8.7
HIRIS RF43 MC Release 2.63 [chksum 33AF]
PLD HIC #31

Juno I General Preset 11-15 Rev. 25 Nov 2015 2/21


2 PU Configuration.
a) Log-on to the system as Administrator (see section 2, paragraph 7.3 of the Juno I Technical
Manual)
User name: Admin2 Password: ats_PED.24

b) Run the PixDemoApp program and perform the following steps:


Check the correct PixDemoApp release
By choosing Help--->About Menu, the following popup will appear:

Right value for PixDemoApp Version is 3.0.2

Juno I General Preset 11-15 Rev. 25 Nov 2015 3/21


In the Setup tab, check the IP address according to the screenshot here below and click
Open.

- After the connection with the PU RF4343 is established, click AcppBox Settings

-
- and check the settings according to the screenshot here below:

Juno I General Preset 11-15 Rev. 25 Nov 2015 4/21


Offset Ref Life Time can also be set to 3600s
- Click Apply and then Close.
- Select the Diagnostic tab and check the PU SW release.
- Sw release must be 3.0.2.1.

- Close the PixDemoApp program.

Juno I General Preset 11-15 Rev. 25 Nov 2015 5/21


3 Configuration of the SD4343 calibration software.
Run the SD4343 program and perform the following steps:
Check the correct SD4343 release:

Right value for SD4343 FileVersion is 1.3.0.2.

Open the Options menu

Check the settings, modify them if


needed, and click OK.

Juno I General Preset 11-15 Rev. 25 Nov 2015 6/21


- Open the File menu and select New.

- The New calibration window opens. Check that the Processing Unit IP Address is
192.168.0.2 (manually input it, if needed).

- Select the option Online using the current PU configuration

Juno I General Preset 11-15 Rev. 25 Nov 2015 7/21


- After the connection between PU RF4343 and PID videoprocessor is established, from the
Logical Mode Manager menu select the Configuration tab and check that the settings are
according to the following picture:

Juno I General Preset 11-15 Rev. 25 Nov 2015 8/21


3.1 Presetting the detector's Working Temperature.
The following procedure is used when installing the system and it is advisable whenever a new calibration
of the detector is to be performed.

During installation of the equipment and before calibrating the detector, you need to carry out the
detector working temperature preset procedure.

This procedure involves two settings in the detector temperature control system:

1) Preliminary setting, where the speed at which the fans rotate is constant until the thermal
stability of the detector is reached (i.e. after at least 3.5 hours of use).

Using the SD4343 application, access the Configuration panel and then set the parameters
below:

Fan mode: set Specified speed


Target temperature: not used
Fan 1 speed: set 2200
Fan 2 speed: set 0

This is the default setting by the manufacturer


when
the equipment is first supplied.

2) Final setting, where the automatic fan speed control is set to keep the detector temperature
constant at the thermal stability value.

The final setting must be set when the detector has reached its thermal stability (i.e. after at
least 3.5 hours of use).

Juno I General Preset 11-15 Rev. 25 Nov 2015 9/21


After this period:

1) Find the temperature reached by the detector, shown in the Pixium Manager 1:1 page;

2) Set this value in the Target Temperature


field in the Configuration panel.

3) Set Pixbox controlled in the Fan Mode field.

In this mode, fields Fan 1 speed and


Fan 2 speed are not used.

Juno I General Preset 11-15 Rev. 25 Nov 2015 10/21


4 Configuration of JUNO I software.
Run JUNO I application and perform the following steps:

a) Check that the proper software/firmware release has been installed: Choose Help-----> About
menu and check that the displayed release is displayed like below:

b) Check that the options of the GENERAL SETUP window are as in the following picture:

Click to check next


menu

Note:
The DUAL SCREEN check box has to be checked only in case the system includes the reference
monitor.

Juno I General Preset 11-15 Rev. 25 Nov 2015 11/21


c) Check that the settings in the MAIN CONTROLLER RAD SETUP window are as in the following
picture:

d) Exit the MAIN CONTROLLER RAD SETUP window;

Juno I General Preset 11-15 Rev. 25 Nov 2015 12/21


e) Click button "Customize RX Filter"

Click to check
next menu

Juno I General Preset 11-15 Rev. 25 Nov 2015 13/21


Define collimator filters as follows:
o FILTER 0 (i.e. NO Filter)

o FILTER 1 (2mm Al Filter)

Juno I General Preset 11-15 Rev. 25 Nov 2015 14/21


o FILTER 2 (1mm Al/0,1mmCu Filter)

o FILTER 3 (1mm Al/0,2 mmCu Filter)

Juno I General Preset 11-15 Rev. 25 Nov 2015 15/21


f) The default settings in the DICOM SETUP window are as in the following picture:

Juno I General Preset 11-15 Rev. 25 Nov 2015 16/21


g) The default the settings in the Q/R SETTINGS window are as in the following picture (check
only in case the option is present):

h) The default settings in the SINGLE IMAGE FILM SETTINGS window are as in the following
picture (check only in case the option is present):

Villa SM

Juno I General Preset 11-15 Rev. 25 Nov 2015 17/21


i) The default settings in the IHE SETUP window are as in the following picture (check only in case
the MPPS option is present):

Juno I General Preset 11-15 Rev. 25 Nov 2015 18/21


j) The default settings in the DICOM WORKLIST SETUP window are as in the following picture
(check only in case the option is present):

Juno I General Preset 11-15 Rev. 25 Nov 2015 19/21


k) The default settings in the MPPS SETUP window are as in the following picture (check only in
case the option is present):

l) The default settings in the DICOM STORAGE COMMITMENT SETUP window are as in the
following picture (check only in case the option is present):

Juno I General Preset 11-15 Rev. 25 Nov 2015 20/21


m) Check that in the MAIN CONTROLLER FLUOROSCOPY SETUP window the Polarity
parameter (polarity of the control signal of the fluoroscopy dose) is set to “+”, as in the following
picture:

n) Close the archive and go to the LOGICAL MODE SETUP.


For each Logical Mode, check that the BORDER values are set according to the following table.

Logical Mode PU Mode Left Border Right Border Top Border Bottom Border

1 3 7 7 2 2

2 7 0 0 2 2

3 15 0 0 2 2

4 27 0 0 0 0

5 2 7 7 2 2

6 7 0 0 2 2

7 22 0 0 2 2

8 27 0 0 0 0

10 17 20 20 3 3

11 4 10 10 2 2

15 17 20 20 3 3

16 17 20 20 3 3

Juno I General Preset 11-15 Rev. 25 Nov 2015 21/21